A miniature X-ray diffraction setup on ID20 at the European Synchrotron Radiation Facility

crossmark logo

aESRF – The European Synchrotron, 71 Avenue des Martyrs, 38043 Grenoble Cedex 9, France, bDepartment of Physics, University of Helsinki, PO Box 64, 00014 Helsinki, Finland, cHelsinki Institute of Physics, University of Helsinki, 00014 Helsinki, Finland, and dIstituto per lo Studio dei Materiali Nanostrutturati (ISMN)-CNR, UOS Palermo, via Ugo La Malfa 153, Palermo 90146, Italy
*Correspondence e-mail: sahle@esrf.fr, alessandro.longo@esrf.fr

Edited by H. Tolentino, Brazilian Synchrotron Light Laboratory, Brazil (Received 5 July 2024; accepted 18 September 2024; online 25 October 2024)

We describe an ultra-compact setup for in situ X-ray diffraction on the inelastic X-ray scattering beamline ID20 at the European Synchrotron Radiation Facility. The main motivation for the design and construction of this setup is the increasing demand for on-the-fly sample characterization, as well as ease of navigation through a sample's phase diagram, for example subjected to high-pressure and/or high-temperature conditions. We provide technical details and demonstrate the performance of the setup.

1. Introduction

The advent of fourth-generation synchrotron radiation facilities has led to a considerable increase in data and sample throughput, even for photon-hungry X-ray techniques such as inelastic X-ray scattering (IXS). This development has allowed more and more complex samples to be studied by IXS and ever more complex sample environments to be used (Petitgirard et al., 2019[Petitgirard, S., Sahle, C. J., Weis, C., Gilmore, K., Spiekermann, G., Tse, J. S., Wilke, M., Cavallari, C., Cerantola, V. & Sternemann, C. (2019). Geochem. Persp. Lett. pp. 32-37.]; Cerantola et al., 2023[Cerantola, V., Sahle, C. J., Petitgirard, S., Wu, M., Checchia, S., Weis, C., Di Michiel, M., Vaughan, G. B. M., Collings, I. E., Arató, R., Wilke, M., Jones, A. P., Hanfland, M. & Tse, J. S. (2023). Commun. Earth Environ. 4, 67.]; Kumar Das et al., 2024[Kumar Das, S., D'ooghe, L., Srinath, N. V., Theofanidis, S.-A., Longo, A., Sahle, C., Van Geem, K., Poelman, H., Poelman, D. & Galvita, V. (2024). ACS Catal. 14, 1311-1323.]).

Unlike experimental techniques that are based on elastic scattering or photoelectric absorption, IXS experiments are still very time consuming. This implies that samples cannot simply be screened, for example for crystallinity, on-the-fly. X-ray diffraction (XRD) is one of the most powerful and popular techniques used within and outside of synchrotron radiation facilities and is an absolutely standard technique for the characterization of materials at atomic length scales (Warren, 1990[Warren, B. E. (1990). X-ray Diffraction. Courier Corporation.]; Guinier, 2013[Guinier, A. (2013). X-ray Diffraction in Crystals, Imperfect Crystals, and Amorphous Bodies. Courier Corporation.]). XRD poses relatively small challenges for instrumentation at entry level and the time scales are short compared to IXS experiments (Krywka et al., 2007[Krywka, C., Sternemann, C., Paulus, M., Javid, N., Winter, R., Al-Sawalmih, A., Yi, S., Raabe, D. & Tolan, M. (2007). J. Synchrotron Rad. 14, 244-251.]; Vaughan et al., 2020[Vaughan, G. B. M., Baker, R., Barret, R., Bonnefoy, J., Buslaps, T., Checchia, S., Duran, D., Fihman, F., Got, P., Kieffer, J., Kimber, S. A. J., Martel, K., Morawe, C., Mottin, D., Papillon, E., Petit­demange, S., Vamvakeros, A., Vieux, J.-P. & Di Michiel, M. (2020). J. Synchrotron Rad. 27, 515-528.]; Wright et al., 2020[Wright, J., Giacobbe, C. & Majkut, M. (2020). Curr. Opin. Solid State Mater. Sci. 24, 100818.]). The latter often take multiple hours and are sensitive to the local electronic and atomic structure, whereas XRD is more sensitive to the long-range atomic structure (Petitgirard et al., 2022[Petitgirard, S., Sahle, C. J., Malfait, W. J., Spiekermann, G., Blanchard, I., Jennings, E. S., Cotte, M. & Murakami, M. (2022). Phys. Rev. B, 105, 134106.]). It is often imperative to confirm local symmetries and a sample's integrity prior to time-consuming IXS experiments.

Here, we describe a compact setup for XRD based on an Advacam Minipix pixelated area detector that allows us to survey and assess sample integrity and crystal structure, and/or obtain complementary sample characterization, while performing (in situ and ex situ) non-resonant inelastic X-ray scattering (NIXS) and X-ray emission spectroscopy (XES).

2. Technical details

The design of the compact diffraction setup, hereafter referred to as miniXRD, is extremely simple: a small pixelated area detector is mounted on a linear translation stage and attached to the large-solid-angle spectrometer for NIXS on ID20 (Huotari et al., 2017[Huotari, S., Sahle, C. J., Henriquet, C., Al-Zein, A., Martel, K., Simonelli, L., Verbeni, R., Gonzalez, H., Lagier, M.-C., Ponchut, C., Moretti Sala, M., Krisch, M. & Monaco, G. (2017). J. Synchrotron Rad. 24, 521-530.]) at short sample-to-detector distances (typically 90–120 mm). All non-commercial parts are based on 3D-printing technology. Three-dimensional renderings of the design are shown in Fig. 1[link] in (a) a vertical configuration or (b) a horizontal arrangement. Fig. 1[link](c) shows a photograph of the detector in its vertical configuration; a high-temperature in situ sample stage is visible in the foreground.

[Figure 1]
Figure 1
Three-dimensional renderings of the miniXRD setup installed in (a) the vertical and (b) the horizontal scanning configuration. (c) Photograph of the setup installed within the large-solid-angle spectrometer on ID20 at the ESRF. (d) Direct beam positions (vertical and horizontal) on the detector versus vertical detector translation, including linear fits. Along the scan direction, the slope of the linear fit suggests a pixel size of 0.055 mm per pixel, in perfect agreement with the hardware specifications.

As detector we use an Advacam Minipix device based on hybrid single photon counting Timepix3 sensor technology (256 × 256 pixels, pixel size 55 µm × 55 µm, sensor thickness 500 µm Si) (Granja et al., 2022[Granja, C., Jakubek, J., Soukup, P., Jakubek, M., Turecek, D., Marek, L., Oancea, C., Gohl, S., Bergmann, B., Pospisil, S., Malich, M., Vuolo, M., Owens, A., Zach, V., Stursa, J., Chvatil, D., Olsansky, V., Rucinski, A., Gajewski, J., Stasica, P., Vykydal, Z. & Solc, J. (2022). J. Instrum. 17, C03019.]). This ultra-compact detector has recently been shown to be useful in a number of applications, ranging from Compton scattering to the characterization of pulsed ultra-high-dose electron beams (Granja et al., 2018[Granja, C., Kudela, K., Jakubek, J., Krist, P., Chvatil, D., Stursa, J. & Polansky, S. (2018). Nucl. Instrum. Methods Phys. Res. A, 911, 142-152.]; Turecek et al., 2020[Turecek, D., Jakubek, J., Trojanova, E. & Sefc, L. (2020). J. Instrum. 15, C01014.]; Oancea et al., 2022[Oancea, C., Bălan, C., Pivec, J., Granja, C., Jakubek, J., Chvatil, D., Olsansky, V. & Chiş, V. (2022). J. Instrum. 17, C01003.]). The small size of this detector enables the compact design of the setup, which makes it possible to integrate it easily in elaborate beamline end stations such as the large-solid-angle spectrometer on ID20 (Huotari et al., 2017[Huotari, S., Sahle, C. J., Henriquet, C., Al-Zein, A., Martel, K., Simonelli, L., Verbeni, R., Gonzalez, H., Lagier, M.-C., Ponchut, C., Moretti Sala, M., Krisch, M. & Monaco, G. (2017). J. Synchrotron Rad. 24, 521-530.]; Sahle et al., 2023[Sahle, Ch. J., Gerbon, F., Henriquet, C., Verbeni, R., Detlefs, B., Longo, A., Mirone, A., Lagier, M.-C., Otte, F., Spiekermann, G. & Petitgirard, S. (2023). J. Synchrotron Rad. 30, 251-257.]; Moretti Sala et al., 2018[Moretti Sala, M., Martel, K., Henriquet, C., Al Zein, A., Simonelli, L., Sahle, C., Gonzalez, H., Lagier, M.-C., Ponchut, C., Huotari, S., Verbeni, R., Krisch, M. & Monaco, G. (2018). J. Synchrotron Rad. 25, 580-591.]), even if complicated sample environments such as diamond anvil cells or high-temperature stages are used. Data transfer and power supply are provided via a micro-USB connection.

In order to establish the quality and reproducibility of the setup, the pixel positions (vertical and horizontal) of the direct X-ray beam are plotted versus the detector translation in Fig. 1[link](d). The linear fits of these curves are proof of the sufficient performance of this setup and confirm the pixel size of 55 µm × 55 µm as expected from the Timepix3 chip. We attribute the small horizontal offset of 5 pixels over the translation range of 14 mm to the imperfect mounting of the translation stage (angular offset between the laboratory horizontal plane and the detector translation direction of approximately 1.1°). The single detector images resulting from a typical detector scan can be combined to form a single diffraction image such as is obtained with a conventional large-area detector. Currently, we do not account for the small angular offset when stitching the individual detector images together, which degrades the final angular resolution slightly. For geometry fitting and azimuthal integration we use pyFAI (Ashiotis et al., 2015[Ashiotis, G., Deschildre, A., Nawaz, Z., Wright, J. P., Karkoulis, D., Picca, F. E. & Kieffer, J. (2015). J. Appl. Cryst. 48, 510-519.]) or Fit2d (Hammersley et al., 1996[Hammersley, A. P., Svensson, S. O., Hanfland, M., Fitch, A. N. & Hausermann, D. (1996). High. Press. Res. 14, 235-248.]). Data reduction can be achieved directly using the XRStools program package (Sahle et al., 2015[Sahle, Ch. J., Mirone, A., Niskanen, J., Inkinen, J., Krisch, M. & Huotari, S. (2015). J. Synchrotron Rad. 22, 400-409.]), which adopts the conventions introduced by Boesecke (2007[Boesecke, P. (2007). J. Appl. Cryst. 40, s423-s427.]). At the sample-to-detector distances used here, the full translation range of the detector covers an angular range of approximately 45°, comparable with what is available with the Pilatus 300Kw detector at a working distance of ca 240 mm.

3. Examples

3.1. Comparison with Pilatus 300Kw

In order to assess the quality of the XRD patterns obtainable with the miniXRD setup, we performed comparative measurements on a pellet comprised of a mixture of WO3 [95410 Sigma–Aldrich tungsten(VI) oxide, purity 99.9%] and α-Al2O3 (NIST SRM 676) using both our miniXRD setup and a Pilatus 300Kw hybrid photon counting large-area detector (Eikenberry et al., 2003[Eikenberry, E. F., Brönnimann, C., Hülsen, G., Toyokawa, H., Horisberger, R., Schmitt, B., Schulze-Briese, C. & Tomizaki, T. (2003). Nucl. Instrum. Methods Phys. Res. A, 501, 260-266.]). We used a test bench available on ID20 for these measurements and an X-ray wavelength of 1.21 Å.

The sample-to-detector distances [\overline{SD}] and the three detector tilt angles (ϕ1, ϕ2, ϕ3) were estimated based on XRD patterns of CeO2 standards (NIST, CAS No. 1306–38-3) to be [\overline{SD}_{\rm M}] = 90.3 mm for the miniature XRD setup and [\overline{SD}_{\rm P}] = 244.9 mm for the Pilatus detector; the tilt angles were all found to be <1°. These conventions follow the descriptions presented by Boesecke (2007[Boesecke, P. (2007). J. Appl. Cryst. 40, s423-s427.]) and were obtained using the pyFAI program package.

Exposure times for the large-area detector were 0.1 s per diffraction pattern. For the miniXRD setup, we used exposure times of 0.1 s for each of the 170 frames (total exposure time 17 s). This resulted in an overall duration of 45 s per diffraction pattern for the miniXRD setup due to the detector translation movement. The miniXRD setup is therefore orders of magnitude slower than conventional diffraction cameras, but it is sufficiently fast on timescales appropriate for IXS experiments, which often take several hours. The individual frames were interpolated onto a final grid of 256 × 1801 pixels of 55 µm × 55 µm using the XRStools program package.

The results of the two simultaneous measurements are compared in Fig. 2[link](a) which shows the pattern obtained using the miniXRD setup, while Fig. 2[link](b) shows the data from the commercial Pilatus detector. We performed a Rietveld refinement on both data sets using the GSAS-II software package (Toby & Von Dreele, 2013[Toby, B. H. & Von Dreele, R. B. (2013). J. Appl. Cryst. 46, 544-549.]). The best fits are shown as thin dashed grey lines, the fitted background function as thin dashed–dotted grey lines and the residuals as thin solid grey lines in Fig. 2[link].

[Figure 2]
Figure 2
Raw data after azimuthal integration and results of our Rietveld refinement of a mixed WO3–Al2O3 polycrystalline powder sample measured using (a) the miniXRD setup and (b) a Pilatus 300Kw large-area detector.

Owing to the considerably closer sample-to-detector distance in the case of the new miniature setup, the covered momentum transfer range is comparable between the two detectors. The two patterns are in close agreement despite slightly different peak ratios, which we attribute to the different physical positions of the two detectors and the poor powder averaging due to the use of a static polycrystalline powder pellet. Besides the main constituent α-Al2O3, we find a considerable fraction of θ-Al2O3, which serves as a precursor phase for the manufacture of α-Al2O3.

The weighted profile R factors were 4.51% and 4.36% for the miniXRD and Pilatus 300Kw setups, respectively. The lattice parameters, grain sizes and phase fractions resulting from the two Rietveld refinements are summarized in Table 1[link] and are compared with values from the literature (Loopstra & Boldrini, 1966[Loopstra, B. O. & Boldrini, P. (1966). Acta Cryst. 21, 158-162.]; Finger & Hazen, 1978[Finger, L. W. & Hazen, R. M. (1978). J. Appl. Phys. 49, 5823-5826.]; Zhou & Snyder, 1991[Zhou, R.-S. & Snyder, R. L. (1991). Acta Cryst. B47, 617-630.]). All extracted parameters are in remarkable agreement between the two detectors and establish the miniXRD setup as a viable online sample characterization tool of sufficient quality for basic sample characterization, including Rietveld refinements.

Table 1
Summary of refinement parameters for the different sample constituents based on Rietveld refinement of 1559 (miniXRD) and 1602 (Pilatus 300Kw) observations

The weighted profile R factors were 4.51% and 4.36% for the miniXRD and Pilatus 300Kw setups, respectively.

  a (Å) b (Å) c (Å) β (°) Grain size (nm) Phase fraction (%)
WO3 (P21/c)            
MiniXRD 7.661 ± 0.016 7.493 ± 0.001 10.532 ± 0.027 136.17 ± 0.03 82 ± 2 7.9 ± 0.5
Pilatus 300Kw 7.673 ± 0.023 7.507 ± 0.001 10.552 ± 0.040 136.21 ± 0.04 92 ± 1 7.8 ± 0.5
Loopstra & Boldrini (1966[Loopstra, B. O. & Boldrini, P. (1966). Acta Cryst. 21, 158-162.]) 7.673 7.507 10.559 136.24
             
α-Al2O3 ([R {\overline 3} ch])            
MiniXRD 4.738 ± 0.001 4.738 ± 0.001 12.937 ± 0.001 1.22 ± 0.07 84.0 ± 0.5
Pilatus 300Kw 4.747 ± 0.001 4.747 ± 0.001 12.970 ± 0.001 1.11 ± 0.04 84.5 ± 0.6
Finger & Hazen (1978[Finger, L. W. & Hazen, R. M. (1978). J. Appl. Phys. 49, 5823-5826.]) 4.748 4.748 12.972
             
θ-Al2O3 (C2/m)            
MiniXRD 11.803 ± 0.020 2.906 ± 0.001 5.572 ± 0.005 104.84 ± 0.03 300 ± 28 8.1 ± 0.5
Pilatus 300Kw 11.806 ± 0.018 2.898 ± 0.002 5.619 ± 0.011 104.16 ± 0.09 277 ± 3 7.7 ± 0.6
Zhou & Snyder (1991[Zhou, R.-S. & Snyder, R. L. (1991). Acta Cryst. B47, 617-630.]) 11.86530 2.89945 5.62693 104.50

We would like to stress, however, that the small size of this detector poses limitations compared with dedicated diffraction instruments. Detailed characterization of samples exhibiting preferred orientation or texture or single-crystal studies will be challenging. Recent work shows that characterization of samples within high-pressure sample environments is possible (Sahle et al., 2024[Sahle, C. J., Petitgirard, S., Spiekermann, G., Sakrowski, R., Suomalainen, N., Gerbon, F., Jacobs, J., Watier, Y., Sternemann, C., Moretti Sala, M. & Cerantola, V. (2024). High. Press. Res. 44, 337-360.]).

3.2. Tandem XRS and XRD

We next demonstrate the utility of the miniXRD setup and the added benefit of enabling sample characterization via XRD while measuring NIXS spectra at the Ce N4,5 edge of the CeO2-based catalyst support material (Scavini et al., 2012[Scavini, M., Coduri, M., Allieta, M., Brunelli, M. & Ferrero, C. (2012). Chem. Mater. 24, 1338-1345.]). High-temperature conditions were reached with a small furnace provided by the ESRF sample environment pool, similar to what is described elsewhere (Longo et al., 2023[Longo, A., Mirone, A., De Clermont Gallerande, E., Sahle, C. J., Casaletto, M. P., Amidani, L., Theofanidis, S. A. & Giannici, F. (2023). Cell. Rep. Phys. Sci. 4, 101699.]; Kumar Das et al., 2024[Kumar Das, S., D'ooghe, L., Srinath, N. V., Theofanidis, S.-A., Longo, A., Sahle, C., Van Geem, K., Poelman, H., Poelman, D. & Galvita, V. (2024). ACS Catal. 14, 1311-1323.]). The XRD patterns were recorded at an incident beam energy of 9.69 keV, while for the measurement of the Ce N4,5 edge the incident energy was scanned between 9.73 and 9.82 keV in order to create energy losses in the vicinity of the Ce N4,5 edge relative to the Si(660) analyser energy of 9.69 keV.

In principle, the XRD measurements could be performed while simultaneously measuring the XRS data of the Ce N4,5 edge, if a variation in X-ray wavelength (here between 1.275 and 1.263 Å) is acceptable for the XRD measurements. In the current case this variation is smaller than the bandwidths provided by, for example, dedicated diffraction instruments utilizing pink beams.

Ceria (CeO2) is a critical catalyst due to its unique capability to generate oxygen vacancies during redox reactions. In the context of anaerobic carbon monoxide oxidation, ceria exhibits the formation of oxygen vacancy clusters within its bulk structure. This phenomenon and its interplay with the orbital hybridization of Ce3+ 4f and 5d states have been recently investigated using advanced techniques such as in situ XRS spectroscopy at the O K and Ce N4,5 edges, along with in situ X-ray diffraction (Longo et al., 2023[Longo, A., Mirone, A., De Clermont Gallerande, E., Sahle, C. J., Casaletto, M. P., Amidani, L., Theofanidis, S. A. & Giannici, F. (2023). Cell. Rep. Phys. Sci. 4, 101699.]).

The utilization of in situ XRS spectroscopy at specific energy levels allows the characterization of the oxygen vacancy clusters and their impact on the electronic structure of cerium (Ce) ions. In situ X-ray diffraction is employed to monitor the transformation of Ce4+ to Ce3+ as a consequence of oxygen vacancy creation during carbon monoxide oxidation. These combined analytical approaches provide deeper insights into the catalytic mechanisms facilitated by ceria under specific redox conditions (Longo et al., 2023[Longo, A., Mirone, A., De Clermont Gallerande, E., Sahle, C. J., Casaletto, M. P., Amidani, L., Theofanidis, S. A. & Giannici, F. (2023). Cell. Rep. Phys. Sci. 4, 101699.]; Longo et al., 2022[Longo, A., Giannici, F., Casaletto, M. P., Rovezzi, M., Sahle, C. J., Glatzel, P., Joly, Y. & Martorana, A. (2022). ACS Catal. 12, 3615-3627.]; Longo et al., 2012[Longo, A., Liotta, L. F., Pantaleo, G., Giannici, F., Venezia, A. M. & Martorana, A. (2012). J. Phys. Chem. C, 116, 2960-2966.]; Deganello et al., 2006[Deganello, G., Giannici, F., Martorana, A., Pantaleo, G., Prestianni, A., Balerna, A., Liotta, L. F. & Longo, A. (2006). J. Phys. Chem. B, 110, 8731-8739.]). The possibility of coupling XRS with XRD allows us to evaluate both the changes in the electronic structure and the concomitant lattice distortions in the bulk of the material.

Fig. 3[link](a) shows the 2D XRD pattern of ambient CeO2 after stitching of the individual miniXRD detector images. Fig. 3[link](b) presents the azimuthally integrated XRD patterns at ambient temperature (RT) and during annealing at T = 700°C. A shift in the reflections is clearly observable. This shift has been shown to be larger than the thermal structure expansion only and is, instead, related to defect-induced chemical expansion due to the formation of oxygen vacancies and Ce3+ ions (Marrocchelli et al., 2012[Marrocchelli, D., Bishop, S. R., Tuller, H. L. & Yildiz, B. (2012). Adv. Funct. Mater. 22, 1958-1965.]; Longo et al., 2023[Longo, A., Mirone, A., De Clermont Gallerande, E., Sahle, C. J., Casaletto, M. P., Amidani, L., Theofanidis, S. A. & Giannici, F. (2023). Cell. Rep. Phys. Sci. 4, 101699.]). Rietveld refinement of the high-temperature curve yields a lattice constant of (5.5095 ± 0.0005) Å and a mean crystallite size of 8.3 µm, well in accordance with the literature (Longo et al., 2023[Longo, A., Mirone, A., De Clermont Gallerande, E., Sahle, C. J., Casaletto, M. P., Amidani, L., Theofanidis, S. A. & Giannici, F. (2023). Cell. Rep. Phys. Sci. 4, 101699.]). We attribute spurious peaks around 2θ = 20°, 21° and 49° to scattering from the sample holder.

[Figure 3]
Figure 3
Simultaneously taken XRD and XRS data of a polycrystalline CeO2 sample in pellet form. (a) Stitched miniXRD detector image from the ambient temperature data shown in panel (b). (b) XRD data for CeO2 at ambient temperature (RT) and T = 700°C after azimuthal integration and the result of a Rietveld refinement at T = 700°C. (The small peaks around 2θ = 20°, 21° and 49° are due to diffraction from the sample holder used.) (c) Ce N4,5 excitation spectra measured using X-ray Raman scattering spectroscopy at a momentum transfer of q = 9.3 Å−1.

The measured Ce N4,5 edges are shown in Fig. 3[link](c). Arrows mark the most prominent alterations to the N4,5 multiplet spectrum upon temperature increase, which are directly related to Ce3+ ion formation as discussed in depth by Longo et al. (2023[Longo, A., Mirone, A., De Clermont Gallerande, E., Sahle, C. J., Casaletto, M. P., Amidani, L., Theofanidis, S. A. & Giannici, F. (2023). Cell. Rep. Phys. Sci. 4, 101699.]).

This example demonstrates that the miniXRD setup can be used as a sample characterization tool while performing sophisticated in situ NIXS experiments at the typically used incident photon energies of approximately 10 keV.

4. Summary and conclusion

We have presented a new and ultra-compact X-ray diffraction setup for on-the-fly sample characterization on the large-solid-angle spectrometer for non-resonant inelastic X-ray scattering spectroscopy on ID20. We have established the quality of the obtained XRD data by comparison of patterns and results of Rietveld analyses of a WO3–Al2O3 powder pellet using both a large-area Pilatus detector and our miniXRD setup. We have further demonstrated the quasi-parallel use of the miniXRD setup with our large-solid-angle spectrometer by studying temperature-induced electronic configuration changes via XRS and the resulting lattice distortions within a CeO2 powder sample.

We hope the new setup will simplify sample characterization and navigation through the temperature/pressure phase space, therefore increasing efficiency for the often time-consuming IXS experiments on ID20 and elsewhere.

Acknowledgements

We acknowledge the European Synchrotron Radiation Facility for provision of synchrotron radiation on beamline ID20 within proposals IH-MA-396 and BLC-15572. We are grateful to Y. Watier and the ESRF sample environment laboratory for support and access to 3D-printing facilities and for fruitful discussions.

References

First citationAshiotis, G., Deschildre, A., Nawaz, Z., Wright, J. P., Karkoulis, D., Picca, F. E. & Kieffer, J. (2015). J. Appl. Cryst. 48, 510–519.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationBoesecke, P. (2007). J. Appl. Cryst. 40, s423–s427.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationCerantola, V., Sahle, C. J., Petitgirard, S., Wu, M., Checchia, S., Weis, C., Di Michiel, M., Vaughan, G. B. M., Collings, I. E., Arató, R., Wilke, M., Jones, A. P., Hanfland, M. & Tse, J. S. (2023). Commun. Earth Environ. 4, 67.  CrossRef Google Scholar
First citationDeganello, G., Giannici, F., Martorana, A., Pantaleo, G., Prestianni, A., Balerna, A., Liotta, L. F. & Longo, A. (2006). J. Phys. Chem. B, 110, 8731–8739.  Web of Science CrossRef PubMed CAS Google Scholar
First citationEikenberry, E. F., Brönnimann, C., Hülsen, G., Toyokawa, H., Horisberger, R., Schmitt, B., Schulze-Briese, C. & Tomizaki, T. (2003). Nucl. Instrum. Methods Phys. Res. A, 501, 260–266.  Web of Science CrossRef CAS Google Scholar
First citationFinger, L. W. & Hazen, R. M. (1978). J. Appl. Phys. 49, 5823–5826.  CrossRef ICSD CAS Web of Science Google Scholar
First citationGranja, C., Jakubek, J., Soukup, P., Jakubek, M., Turecek, D., Marek, L., Oancea, C., Gohl, S., Bergmann, B., Pospisil, S., Malich, M., Vuolo, M., Owens, A., Zach, V., Stursa, J., Chvatil, D., Olsansky, V., Rucinski, A., Gajewski, J., Stasica, P., Vykydal, Z. & Solc, J. (2022). J. Instrum. 17, C03019.  CrossRef Google Scholar
First citationGranja, C., Kudela, K., Jakubek, J., Krist, P., Chvatil, D., Stursa, J. & Polansky, S. (2018). Nucl. Instrum. Methods Phys. Res. A, 911, 142–152.  CrossRef CAS Google Scholar
First citationGuinier, A. (2013). X-ray Diffraction in Crystals, Imperfect Crystals, and Amorphous Bodies. Courier Corporation.  Google Scholar
First citationHammersley, A. P., Svensson, S. O., Hanfland, M., Fitch, A. N. & Hausermann, D. (1996). High. Press. Res. 14, 235–248.  CrossRef Web of Science Google Scholar
First citationHuotari, S., Sahle, C. J., Henriquet, C., Al-Zein, A., Martel, K., Simonelli, L., Verbeni, R., Gonzalez, H., Lagier, M.-C., Ponchut, C., Moretti Sala, M., Krisch, M. & Monaco, G. (2017). J. Synchrotron Rad. 24, 521–530.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationKrywka, C., Sternemann, C., Paulus, M., Javid, N., Winter, R., Al-Sawalmih, A., Yi, S., Raabe, D. & Tolan, M. (2007). J. Synchrotron Rad. 14, 244–251.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationKumar Das, S., D'ooghe, L., Srinath, N. V., Theofanidis, S.-A., Longo, A., Sahle, C., Van Geem, K., Poelman, H., Poelman, D. & Galvita, V. (2024). ACS Catal. 14, 1311–1323.  CrossRef CAS Google Scholar
First citationLongo, A., Giannici, F., Casaletto, M. P., Rovezzi, M., Sahle, C. J., Glatzel, P., Joly, Y. & Martorana, A. (2022). ACS Catal. 12, 3615–3627.  CrossRef CAS Google Scholar
First citationLongo, A., Liotta, L. F., Pantaleo, G., Giannici, F., Venezia, A. M. & Martorana, A. (2012). J. Phys. Chem. C, 116, 2960–2966.  CrossRef CAS Google Scholar
First citationLongo, A., Mirone, A., De Clermont Gallerande, E., Sahle, C. J., Casaletto, M. P., Amidani, L., Theofanidis, S. A. & Giannici, F. (2023). Cell. Rep. Phys. Sci. 4, 101699.  CrossRef Google Scholar
First citationLoopstra, B. O. & Boldrini, P. (1966). Acta Cryst. 21, 158–162.  CrossRef ICSD CAS IUCr Journals Web of Science Google Scholar
First citationMarrocchelli, D., Bishop, S. R., Tuller, H. L. & Yildiz, B. (2012). Adv. Funct. Mater. 22, 1958–1965.  CrossRef CAS Google Scholar
First citationMoretti Sala, M., Martel, K., Henriquet, C., Al Zein, A., Simonelli, L., Sahle, C., Gonzalez, H., Lagier, M.-C., Ponchut, C., Huotari, S., Verbeni, R., Krisch, M. & Monaco, G. (2018). J. Synchrotron Rad. 25, 580–591.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationOancea, C., Bălan, C., Pivec, J., Granja, C., Jakubek, J., Chvatil, D., Olsansky, V. & Chiş, V. (2022). J. Instrum. 17, C01003.  CrossRef Google Scholar
First citationPetitgirard, S., Sahle, C. J., Malfait, W. J., Spiekermann, G., Blanchard, I., Jennings, E. S., Cotte, M. & Murakami, M. (2022). Phys. Rev. B, 105, 134106.  CrossRef Google Scholar
First citationPetitgirard, S., Sahle, C. J., Weis, C., Gilmore, K., Spiekermann, G., Tse, J. S., Wilke, M., Cavallari, C., Cerantola, V. & Sternemann, C. (2019). Geochem. Persp. Lett. pp. 32–37.  CrossRef Google Scholar
First citationSahle, Ch. J., Gerbon, F., Henriquet, C., Verbeni, R., Detlefs, B., Longo, A., Mirone, A., Lagier, M.-C., Otte, F., Spiekermann, G. & Petitgirard, S. (2023). J. Synchrotron Rad. 30, 251–257.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationSahle, Ch. J., Mirone, A., Niskanen, J., Inkinen, J., Krisch, M. & Huotari, S. (2015). J. Synchrotron Rad. 22, 400–409.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationSahle, C. J., Petitgirard, S., Spiekermann, G., Sakrowski, R., Suomalainen, N., Gerbon, F., Jacobs, J., Watier, Y., Sternemann, C., Moretti Sala, M. & Cerantola, V. (2024). High. Press. Res. 44, 337–360.  CrossRef Google Scholar
First citationScavini, M., Coduri, M., Allieta, M., Brunelli, M. & Ferrero, C. (2012). Chem. Mater. 24, 1338–1345.  Web of Science CrossRef ICSD CAS Google Scholar
First citationToby, B. H. & Von Dreele, R. B. (2013). J. Appl. Cryst. 46, 544–549.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationTurecek, D., Jakubek, J., Trojanova, E. & Sefc, L. (2020). J. Instrum. 15, C01014.  CrossRef Google Scholar
First citationVaughan, G. B. M., Baker, R., Barret, R., Bonnefoy, J., Buslaps, T., Checchia, S., Duran, D., Fihman, F., Got, P., Kieffer, J., Kimber, S. A. J., Martel, K., Morawe, C., Mottin, D., Papillon, E., Petit­demange, S., Vamvakeros, A., Vieux, J.-P. & Di Michiel, M. (2020). J. Synchrotron Rad. 27, 515–528.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationWarren, B. E. (1990). X-ray Diffraction. Courier Corporation.  Google Scholar
First citationWright, J., Giacobbe, C. & Majkut, M. (2020). Curr. Opin. Solid State Mater. Sci. 24, 100818.  Web of Science CrossRef Google Scholar
First citationZhou, R.-S. & Snyder, R. L. (1991). Acta Cryst. B47, 617–630.  CrossRef ICSD CAS Web of Science IUCr Journals Google Scholar

This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.

Journal logoJOURNAL OF
SYNCHROTRON
RADIATION
ISSN: 1600-5775
Follow J. Synchrotron Rad.
Sign up for e-alerts
Follow J. Synchrotron Rad. on Twitter
Follow us on facebook
Sign up for RSS feeds