High-transmission spectrometer for rapid resonant inelastic soft X-ray scattering (rRIXS) maps

crossmark logo

aInstitute for Photon Science and Synchrotron Radiation (IPS), Karlsruhe Institute of Technology (KIT), Kaiserstr. 12, 76131 Karlsruhe, Germany, bInstitute for Chemical Technology and Polymer Chemistry (ITCP), Karlsruhe Institute of Technology (KIT), Kaiserstr. 12, 76131 Karlsruhe, Germany, and cDepartment of Chemistry and Biochemistry, University of Nevada, Las Vegas (UNLV), 4505 Maryland Parkway, Las Vegas, NV 89154-4003, USA
*Correspondence e-mail: lothar.weinhardt@kit.edu

Edited by D. Bhattacharyya, Bhabha Atomic Research Centre, India (Received 22 March 2024; accepted 15 August 2024; online 30 September 2024)

The design and first results of a high-transmission soft X-ray spectrometer operated at the X-SPEC double-undulator beamline of the KIT Light Source are presented. As a unique feature, particular emphasis was placed on optimizing the spectrometer transmission by maximizing the solid angle and the efficiencies of spectrometer gratings and detector. A CMOS detector, optimized for soft X-rays, allows for quantum efficiencies of 90% or above over the full energy range of the spectrometer, while simultaneously offering short readout times. Combining an optimized control system at the X-SPEC beamline with continuous energy scans (as opposed to step scans), the high transmission of the spectrometer, and the fast readout of the CMOS camera, enable the collection of entire rapid resonant inelastic soft X-ray scattering maps in less than 1 min. Series of spectra at a fixed energy can be taken with a frequency of up to 5 Hz. Furthermore, the use of higher-order reflections allows a very wide energy range (45 to 2000 eV) to be covered with only two blazed gratings, while keeping the efficiency high and the resolving power E/ΔE above 1500 and 3000 with low- and high-energy gratings, respectively.

1. Introduction

Soft X-ray emission spectroscopy (XES) and, when performed with resonant excitation, resonant inelastic soft X-ray scattering (RIXS) are very powerful techniques to probe many different aspects of the electronic structure of materials. Both XES and RIXS are `local' techniques due to the selection of a specific core orbital, allowing for a variety of studies, including on chemical bonding (O'Bryan & Skinner, 1940[O'Bryan, H. M. & Skinner, H. W. B. (1940). Proc. R. Soc. London, 176, 229-262.]; Meisel et al., 1968[Meisel, A., Steuer, I. & Szargan, R. (1968). At. Spectrosc. 23, 527-533.]; Meyer et al., 2014[Meyer, F., Blum, M., Benkert, A., Hauschild, D., Nagarajan, S., Wilks, R. G., Andersson, J., Yang, W., Zharnikov, M., Bär, M., Heske, C., Reinert, F. & Weinhardt, L. (2014). J. Phys. Chem. B, 118, 13142-13150.]; Seitz et al., 2022[Seitz, L. C., Doronkin, D. E., Hauschild, D., Casapu, M., Zengel, D., Zimina, A., Kreikemeyer-Lorenzo, D., Blum, M., Yang, W., Grunwaldt, J.-D., Heske, C. & Weinhardt, L. (2022). J. Phys. Chem. C, 126, 20998-21009.]; Weinhardt, Hau­schild et al., 2021[Weinhardt, L., Hauschild, D., Steininger, R., Jiang, N., Blum, M., Yang, W. & Heske, C. (2021). Anal. Chem. 93, 8300-8308.]; Li et al., 2023[Li, B., Zhuo, Z., Zhang, L., Iadecola, A., Gao, X., Guo, J., Yang, W., Morozov, A. V., Abakumov, A. M. & Tarascon, J.-M. (2023). Nat. Mater. 22, 1370-1379.]), band structure (Johnson & Ma, 1994[Johnson, P. D. & Ma, Y. (1994). Phys. Rev. B, 49, 5024-5027.]; Carlisle et al., 1995[Carlisle, J. A., Shirley, E. L., Hudson, E. A., Terminello, L. J., Callcott, T. A., Jia, J. J., Ederer, D. L., Perera, R. C. C. & Himpsel, F. J. (1995). Phys. Rev. Lett. 74, 1234-1237.]; Eich et al., 2006[Eich, D., Fuchs, O., Groh, U., Weinhardt, L., Fink, R., Umbach, E., Heske, C., Fleszar, A., Hanke, W., Gross, E. K. U., Bostedt, C., Von Buuren, T., Franco, N., Terminello, L. J., Keim, M., Reuscher, G., Lugauer, H. & Waag, A. (2006). Phys. Rev. B, 73, 115212.]), vibrational structure in molecular systems (Werme et al., 1973[Werme, L., Grennberg, B., Nordgren, J., Nordling, C. & Siegbahn, K. (1973). Nature, 242, 453-455.]; Hennies et al., 2010[Hennies, F., Pietzsch, A., Berglund, M., Föhlisch, A., Schmitt, Th., Strocov, V., Karlsson, H. O., Andersson, J. & Rubensson, J.-E. (2010). Phys. Rev. Lett. 104, 193002.]; Rubensson et al., 2015[Rubensson, J.-E., Söderström, J., Binggeli, C., Gråsjö, J., Andersson, J., Såthe, C., Hennies, F., Bisogni, V., Huang, Y., Olalde, P., Schmitt, T., Strocov, V. N., Föhlisch, A., Kennedy, B. & Pietzsch, A. (2015). Phys. Rev. Lett. 114, 133001.]; Ertan et al., 2018[Ertan, E., Savchenko, V., Ignatova, N., Vaz da Cruz, V., Couto, R. C., Eckert, S., Fondell, M., Dantz, M., Kennedy, B., Schmitt, T., Pietzsch, A., Föhlisch, A., Gel'mukhanov, F., Odelius, M. & Kimberg, V. (2018). Phys. Chem. Chem. Phys. 20, 14384-14397.]), as well as elementary excitations (charge transfer, dd, magnons, phonons) (Chiuzbăian et al., 2005[Chiuzbăian, S. G., Ghiringhelli, G., Dallera, C., Grioni, M., Amann, P., Wang, X., Braicovich, L. & Patthey, L. (2005). Phys. Rev. Lett. 95, 197402.]; Ghiringhelli, Matsubara et al., 2006[Ghiringhelli, G., Matsubara, M., Dallera, C., Fracassi, F., Tagliaferri, A., Brookes, N. B., Kotani, A. & Braicovich, L. (2006). Phys. Rev. B, 73, 035111.]; Ament et al., 2011[Ament, L. J. P., van Veenendaal, M., Devereaux, T. P., Hill, J. P. & van den Brink, J. (2011). Rev. Mod. Phys. 83, 705-767.]; Schlappa et al., 2012[Schlappa, J., Wohlfeld, K., Zhou, K. J., Mourigal, M., Haverkort, M. W., Strocov, V. N., Hozoi, L., Monney, C., Nishimoto, S., Singh, S., Revcolevschi, A., Caux, J.-S., Patthey, L., Rønnow, H. M., van den Brink, J. & Schmitt, T. (2012). Nature, 485, 82-85.]; Vale et al., 2019[Vale, J. G., Dashwood, C. D., Paris, E., Veiga, L. S. I., Garcia-Fernandez, M., Nag, A., Walters, A., Zhou, K.-J., Pietsch, I.-M., Jesche, A., Gegenwart, P., Coldea, R., Schmitt, T. & McMorrow, D. F. (2019). Phys. Rev. B, 100, 224303.]; Lu et al., 2021[Lu, H., Rossi, M., Nag, A., Osada, M., Li, D. F., Lee, K., Wang, B. Y., Garcia-Fernandez, M., Agrestini, S., Shen, Z. X., Been, E. M., Moritz, B., Devereaux, T. P., Zaanen, J., Hwang, H. Y., Zhou, K.-J. & Lee, W. S. (2021). Science, 373, 213-216.]).

Fueled by this versatility, XES and RIXS have gained an important role in synchrotron spectroscopy, with many new instruments coming online in the past two decades. Generally, the experimental setups have been developed in two opposing directions: high resolution and high transmission. The former is particularly important for studying elementary excitations that can require a resolution of a few tens of meV. This has been achieved by building very long (>5 m) spectrometers (Ghiringhelli, Piazzalunga et al., 2006[Ghiringhelli, G., Piazzalunga, A., Dallera, C., Trezzi, G., Braicovich, L., Schmitt, T., Strocov, V. N., Betemps, R., Patthey, L., Wang, X. & Grioni, M. (2006). Rev. Sci. Instrum. 77, 113108.]; Chiuzbăian et al., 2014[Chiuzbăian, S. G., Hague, C. F., Avila, A., Delaunay, R., Jaouen, N., Sacchi, M., Polack, F., Thomasset, M., Lagarde, B., Nicolaou, A., Brignolo, S., Baumier, C., Lüning, J. & Mariot, J.-M. (2014). Rev. Sci. Instrum. 85, 043108.]; Dvorak et al., 2016[Dvorak, J., Jarrige, I., Bisogni, V., Coburn, S. & Leonhardt, W. (2016). Rev. Sci. Instrum. 87, 115109.]; Zhou et al., 2022[Zhou, K.-J., Walters, A., Garcia-Fernandez, M., Rice, T., Hand, M., Nag, A., Li, J., Agrestini, S., Garland, P., Wang, H., Alcock, S., Nistea, I., Nutter, B., Rubies, N., Knap, G., Gaughran, M., Yuan, F., Chang, P., Emmins, J. & Howell, G. (2022). J. Synchrotron Rad. 29, 563-580.]), but the price to pay is a comparably low transmission that leads to long exposure/acquisition times.

While this is affordable for many questions in fundamental research, this can be a severe limitation for the investigation of applied material systems. Many samples, in particular organic materials, can already be damaged by soft X-rays after short exposure (Zubavichus et al., 2004[Zubavichus, Y., Fuchs, O., Weinhardt, L., Heske, C., Umbach, E., Denlinger, J. D. & Grunze, M. (2004). Radiat. Res. 161, 346-358.]; Weinhardt et al., 2008[Weinhardt, L., Blum, M., Bär, M., Heske, C., Cole, B., Marsen, B. & Miller, E. L. (2008). J. Phys. Chem. C, 112, 3078-3082.], 2019[Weinhardt, L., Benkert, A., Meyer, F., Blum, M., Hauschild, D. G., Wilks, R., Bär, M., Yang, W., Zharnikov, M., Reinert, F. & Heske, C. (2019). Phys. Chem. Chem. Phys. 21, 13207-13214.]; Wilks et al., 2009[Wilks, R. G., MacNaughton, J. B., Kraatz, H.-B., Regier, T., Blyth, R. I. R. & Moewes, A. (2009). J. Phys. Chem. A, 113, 5360-5366.]; Léon et al., 2017[Léon, A., Fiedler, A., Blum, M., Benkert, A., Meyer, F., Yang, W., Bär, M., Scheiba, F., Ehrenberg, H., Weinhardt, L. & Heske, C. (2017). J. Phys. Chem. C, 121, 5460-5466.]; Roychoudhury et al., 2021[Roychoudhury, S., Zhuo, Z., Qiao, R., Wan, L., Liang, Y., Pan, F., Chuang, Y., Prendergast, D. & Yang, W. (2021). Appl. Mater. Interfaces, 13, 45488-45495.]), which requires the minimization of the radiation dose. Furthermore, the element of interest might occur in a very low concentration, which requires an efficient detection. And finally, the past decade has seen a growing interest in operando studies using soft X-rays, and suitable sample environments have been developed (Heske et al., 2003[Heske, C., Groh, U., Fuchs, O., Weinhardt, L., Umbach, E., Schedel-Niedrig, T., Fischer, C. H., Lux-Steiner, M. C., Zweigart, S., Niesen, T. P., Karg, F., Denlinger, J. D., Rude, B., Andrus, C. & Powell, F. (2003). J. Chem. Phys. 119, 10467-10470.]; Blum et al., 2009[Blum, M., Weinhardt, L., Fuchs, O., Bär, M., Zhang, Y., Weigand, M., Krause, S., Pookpanratana, S., Hofmann, T., Yang, W., Denlinger, J. D., Umbach, E. & Heske, C. (2009). Rev. Sci. Instrum. 80, 123102.]; Nagasaka et al., 2010[Nagasaka, M., Hatsui, T., Horigome, T., Hamamura, Y. & Kosugi, N. (2010). J. Electron Spectrosc. Relat. Phenom. 177, 130-134.]; Schwanke et al., 2014[Schwanke, C., Golnak, R., Xiao, J. & Lange, K. M. (2014). Rev. Sci. Instrum. 85, 103120.]). In these experiments, quick measurements are required to monitor changes of the sample as a function of time. This has motivated the development of high-transmission spectrometers, for which several strategies have been used. While earlier designs (Nordgren & Nyholm, 1986[Nordgren, J. & Nyholm, R. (1986). Nucl. Instrum. Methods Phys. Res. A, 246, 242-245.]; Callcott et al., 1986[Callcott, T. A., Tsang, K. L., Zhang, C. H., Ederer, D. L. & Arakawa, E. T. (1986). Rev. Sci. Instrum. 57, 2680-2690.]; Shin et al., 1995[Shin, S., Agui, A., Fujisawa, M., Tezuka, Y., Ishii, T. & Hirai, N. (1995). Rev. Sci. Instrum. 66, 1584-1586.]; Dallera et al., 1996[Dallera, C., Puppin, E., Trezzi, G., Incorvaia, N., Fasana, A., Braicovich, L., Brookes, N. B. & Goedkoop, J. B. (1996). J. Synchrotron Rad. 3, 231-238.]) used an entrance slit to define the source spot, current spectrometers significantly increase the solid angle by using slit-less designs with the source spot defined by the beam spot on the sample (Cocco et al., 2004[Cocco, D., Zangrando, M., Matteucci, M., Bondino, F., Platè, M., Zacchigna, M., Parmigiani, F., Nelles, B. & Prince, K. C. (2004). AIP Conf. Proc. 705, 873-876.]; Chuang et al., 2005[Chuang, Y.-D., Pepper, J., McKinney, W., Hussain, Z., Gullikson, E., Batson, P., Qian, D. & Hasan, M. Z. (2005). J. Phys. Chem. Solids, 66, 2173-2178.]; Tokushima et al., 2006[Tokushima, T., Harada, Y., Ohashi, H., Senba, Y. & Shin, S. (2006). Rev. Sci. Instrum. 77, 063107.]; Ghiringhelli, Piazzalunga et al., 2006[Ghiringhelli, G., Piazzalunga, A., Dallera, C., Trezzi, G., Braicovich, L., Schmitt, T., Strocov, V. N., Betemps, R., Patthey, L., Wang, X. & Grioni, M. (2006). Rev. Sci. Instrum. 77, 113108.]; Fuchs et al., 2009[Fuchs, O., Weinhardt, L., Blum, M., Weigand, M., Umbach, E., Bär, M., Heske, C., Denlinger, J., Chuang, Y. D., McKinney, W., Hussain, Z., Gullikson, E., Jones, M., Batson, P., Nelles, B. & Follath, R. (2009). Rev. Sci. Instrum. 80, 063103.]; Chiuzbăian et al., 2014[Chiuzbăian, S. G., Hague, C. F., Avila, A., Delaunay, R., Jaouen, N., Sacchi, M., Polack, F., Thomasset, M., Lagarde, B., Nicolaou, A., Brignolo, S., Baumier, C., Lüning, J. & Mariot, J.-M. (2014). Rev. Sci. Instrum. 85, 043108.]; Dvorak et al., 2016[Dvorak, J., Jarrige, I., Bisogni, V., Coburn, S. & Leonhardt, W. (2016). Rev. Sci. Instrum. 87, 115109.]; Qiao et al., 2017[Qiao, R., Li, Q., Zhuo, Z., Sallis, S., Fuchs, O., Blum, M., Weinhardt, L., Heske, C., Pepper, J., Jones, M., Brown, A., Spucces, A., Chow, K., Smith, B., Glans, P.-A., Chen, Y., Yan, S., Pan, F., Piper, L. F. J., Denlinger, J., Guo, J., Hussain, Z., Chuang, Y.-D. & Yang, W. (2017). Rev. Sci. Instrum. 88, 033106.]; Zhou et al., 2022[Zhou, K.-J., Walters, A., Garcia-Fernandez, M., Rice, T., Hand, M., Nag, A., Li, J., Agrestini, S., Garland, P., Wang, H., Alcock, S., Nistea, I., Nutter, B., Rubies, N., Knap, G., Gaughran, M., Yuan, F., Chang, P., Emmins, J. & Howell, G. (2022). J. Synchrotron Rad. 29, 563-580.]). This must be used with some caution in case of radiation-sensitive samples, since smaller sample spots also increase the radiation dose.

The development of a short, slit-less spectrometer with spherical mirror, plane variable-line-space (VLS) grating, and a CCD detector in normal incidence by some of the authors achieved a significant improvement in spectrometer transmission collecting spectra with acquisition times as short as 10 s (Fuchs et al., 2009[Fuchs, O., Weinhardt, L., Blum, M., Weigand, M., Umbach, E., Bär, M., Heske, C., Denlinger, J., Chuang, Y. D., McKinney, W., Hussain, Z., Gullikson, E., Jones, M., Batson, P., Nelles, B. & Follath, R. (2009). Rev. Sci. Instrum. 80, 063103.]). This allowed the development of the soft X-ray RIXS map (Fuchs et al., 2009[Fuchs, O., Weinhardt, L., Blum, M., Weigand, M., Umbach, E., Bär, M., Heske, C., Denlinger, J., Chuang, Y. D., McKinney, W., Hussain, Z., Gullikson, E., Jones, M., Batson, P., Nelles, B. & Follath, R. (2009). Rev. Sci. Instrum. 80, 063103.]; Weinhardt et al., 2009[Weinhardt, L., Fuchs, O., Fleszar, A., Bär, M., Blum, M., Weigand, M., Denlinger, J. D., Yang, W., Hanke, W., Umbach, E. & Heske, C. (2009). Phys. Rev. B, 79, 165305.]), where a RIXS spectrum is measured for every point in a regular X-ray absorption scan; the emission intensity is then plotted color-coded as a function of excitation and emission energies. Typical measurement times for such a RIXS map are of the order of a few tens of minutes. Other spectrometer designs use transmission gratings (Hatsui et al., 2005[Hatsui, T., Setoyama, H., Shigemasa, E. & Kosugi, N. (2005). J. Electron Spectrosc. Relat. Phenom. 144-147, 1059-1062.]; Yamane et al., 2013[Yamane, H., Kosugi, N. & Hatsui, T. (2013). J. Electron Spectrosc. Relat. Phenom. 188, 155-160.]) or transmission off-axis zone plates (Marschall et al., 2017[Marschall, F., Yin, Z., Rehanek, J., Beye, M., Döring, F., Kubiček, K., Raiser, D., Veedu, S. T., Buck, J., Rothkirch, A., Rösner, B., Guzenko, V. A., Viefhaus, J., David, C. & Techert, S. (2017). Sci. Rep. 7, 8849.]; Schunck et al., 2021[Schunck, J. O., Döring, F., Rösner, B., Buck, J., Engel, R. Y., Miedema, P. S., Mahatha, S. K., Hoesch, M., Petraru, A., Kohlstedt, H., Schüssler-Langeheine, C., Rossnagel, K., David, C. & Beye, M. (2021). Optica, 8, 156-160.]) to further increase the solid collection angle. By opening the beamline exit slit and moving the beamline monochromator focus from the exit slit to the sample, some additional degree of parallelization can be achieved, allowing an excitation energy range of a few eV to be collected simultaneously with suitable imaging optics of the spectrometer (Warwick et al., 2014[Warwick, T., Chuang, Y.-D., Voronov, D. L. & Padmore, H. A. (2014). J. Synchrotron Rad. 21, 736-743.]; Chuang et al., 2020[Chuang, Y.-D., Feng, X., Glans-Suzuki, P.-A., Yang, W., Padmore, H. & Guo, J.-H. (2020). J. Synchrotron Rad. 27, 695-707.]; Marschall et al., 2017[Marschall, F., Yin, Z., Rehanek, J., Beye, M., Döring, F., Kubiček, K., Raiser, D., Veedu, S. T., Buck, J., Rothkirch, A., Rösner, B., Guzenko, V. A., Viefhaus, J., David, C. & Techert, S. (2017). Sci. Rep. 7, 8849.]; Schunck et al., 2021[Schunck, J. O., Döring, F., Rösner, B., Buck, J., Engel, R. Y., Miedema, P. S., Mahatha, S. K., Hoesch, M., Petraru, A., Kohlstedt, H., Schüssler-Langeheine, C., Rossnagel, K., David, C. & Beye, M. (2021). Optica, 8, 156-160.]). Combining this parallel collection scheme with transmission off-axis zone plates, a section of a RIXS map can be collected `in one shot' within approximately 10 min (Marschall et al., 2017[Marschall, F., Yin, Z., Rehanek, J., Beye, M., Döring, F., Kubiček, K., Raiser, D., Veedu, S. T., Buck, J., Rothkirch, A., Rösner, B., Guzenko, V. A., Viefhaus, J., David, C. & Techert, S. (2017). Sci. Rep. 7, 8849.]). The energy window of this RIXS map section is limited in excitation energy by the bandwidth of the beamline (of a few eV) and the energy window in focus for one setting of the spectrometer (also a few eV) (Marschall et al., 2017[Marschall, F., Yin, Z., Rehanek, J., Beye, M., Döring, F., Kubiček, K., Raiser, D., Veedu, S. T., Buck, J., Rothkirch, A., Rösner, B., Guzenko, V. A., Viefhaus, J., David, C. & Techert, S. (2017). Sci. Rep. 7, 8849.]). A very different strategy for high transmission is the use of an energy-dispersive design instead of the wavelength-dispersive approaches discussed above. This promises very high efficiency with large solid angle and is insensitive to the spot size on the sample. In recent years, this approach has been steadily improved and a first instrument is in operation (Lee et al., 2019[Lee, S.-J., Titus, C. J., Alonso Mori, R., Baker, M. L., Bennett, D. A., Cho, H.-M., Doriese, W. B., Fowler, J. W., Gaffney, K. J., Gallo, A., Gard, J. D., Hilton, G. C., Jang, H., Joe, Y. I., Kenney, C. J., Knight, J., Kroll, T., Lee, J.-S., Li, D., Lu, D., Marks, R., Minitti, M. P., Morgan, K. M., Ogasawara, H., O'Neil, G. C., Reintsema, C. D., Schmidt, D. R., Sokaras, D., Ullom, J. N., Weng, T.-C., Williams, C., Young, B. A., Swetz, D. S., Irwin, K. D. & Nordlund, D. (2019). Rev. Sci. Instrum. 90, 113101.]), but its energy resolution of 1.5 eV is still inferior to their wavelength-dispersive counterparts and maximal count rates are limited due to long dead-times. A second, energy-dispersive strategy is the `photoelectron spectrometry for analysis of X-rays' (PAX) that uses a material converting X-rays into photoelectrons, which are then detected with a conventional electron analyzer; however, only a first demonstration experiment has been reported to date (Dakovski et al., 2017[Dakovski, G. L., Lin, M.-F., Damiani, D. S., Schlotter, W. F., Turner, J. J., Nordlund, D. & Ogasawara, H. (2017). J. Synchrotron Rad. 24, 1180-1186.]).

In this paper, we present a very high transmission soft X-ray spectrometer at the X-SPEC double-undulator beamline (Weinhardt, Steininger et al., 2021[Weinhardt, L., Steininger, R., Kreikemeyer-Lorenzo, D., Mangold, S., Hauschild, D., Batchelor, D., Spangenberg, T. & Heske, C. (2021). J. Synchrotron Rad. 28, 609-617.]) at the KIT Light Source that allows single spectra to be measured in a fraction of a second, full RIXS maps in half a minute (henceforth termed rRIXS, for `rapid RIXS'), and achieves a resolving power of 1500 or better over its wide energy range from 45 to 2000 eV.

2. Design considerations and optical layout

The presented soft X-ray spectrometer is aimed for use in the investigation of applied material systems under ultra-high vacuum (UHV), in situ and operando conditions. As outlined above, this requires maximizing the spectrometer transmission while still maintaining the required experimental resolution. For chemical bonding properties and electronic band structure information, an energy resolution of 0.2–0.5 eV is sufficient in most cases, which can also resolve dd and vibrational excitations in some cases. As the final requirement in our case, the X-ray spectrometer needs to cover energies from 45 to 2000 eV to include the entire soft X-ray energy range of the X-SPEC beamline (Weinhardt, Steininger et al., 2021[Weinhardt, L., Steininger, R., Kreikemeyer-Lorenzo, D., Mangold, S., Hauschild, D., Batchelor, D., Spangenberg, T. & Heske, C. (2021). J. Synchrotron Rad. 28, 609-617.]) and allow for Li K XES (i.e. with non-resonant excitation). The X-SPEC beamline has two experimental stations for studies under UHV and in situ/operando conditions at atmospheric pressures or above, respectively. Each of these endstations is equipped with one X-ray spectrometer of the design discussed in this paper.

To fulfill all of the discussed requirements, our spectrometer uses a slit-less design with spherical VLS gratings (one for low and one for high energies) as the single optical element, coupled with a soft X-ray CMOS camera. The gratings for low and high energies have central groove densities of 600 and 1200 lines mm−1, with energy ranges of 45–800 and 450–2000 eV, respectively. To cover these large energy ranges efficiently, we make extensive use of higher diffraction orders (Fuchs et al., 2009[Fuchs, O., Weinhardt, L., Blum, M., Weigand, M., Umbach, E., Bär, M., Heske, C., Denlinger, J., Chuang, Y. D., McKinney, W., Hussain, Z., Gullikson, E., Jones, M., Batson, P., Nelles, B. & Follath, R. (2009). Rev. Sci. Instrum. 80, 063103.]), as discussed below. In general, three main factors can be exploited to optimize the X-ray spectrometer transmission: the use of parallelization, the acceptance angle, and the efficiency of the optics and the detector.

To measure all photon energies of a spectrum in parallel with an area detector, care must be taken that aberrations are sufficiently corrected in the relevant energy window. To achieve this, the VLS parameters and radii of the gratings were optimized numerically using the procedure by Strocov et al., as implemented in the GeneVLS package (Strocov et al., 2011[Strocov, V. N., Schmitt, T., Flechsig, U., Patthey, L. & Chiuzbăian, G. S. (2011). J. Synchrotron Rad. 18, 134-142.]), and checked by ray-tracing using the RAY program (Schäfers, 2008[Schäfers, F. (2008). Modern Developments in X-ray and Neutron Optics, Vol. 137 of Springer Series in Optical Sciences, p. 9. Berlin/Heidelberg: Springer.]). All optical parameters of the spectrometer are listed in Table 1[link]. For the low-energy grating (LEG), parameters were optimized for an energy of 180 eV, and for the high-energy grating (HEG) for an energy of 1000 eV. For other center energies on the detector, aberrations can be minimized by changing the grating angle and grating–detector distance to values optimized using TraceVLS (Strocov et al., 2011[Strocov, V. N., Schmitt, T., Flechsig, U., Patthey, L. & Chiuzbăian, G. S. (2011). J. Synchrotron Rad. 18, 134-142.]). With this, aberrations do not play any role over the full energy range of the LEG, and only influence the overall resolving power for the HEG at lower energies. This can be seen in Fig. 1[link], where the contributions of aberrations, source spot size and CMOS pixel size are illustrated. Depending on the beamline settings, the resolving power is limited by the source spot on the sample (with a minimum of ∼5 µm for narrow beamline exit slits) or the effective pixel size (approximated with 20 µm) of the CMOS detector. It decreases with increasing photon energy. For the LEG, source sizes of 30 µm are sufficient to keep the resolving power E/ΔE above 1500. This allows large exit slits of the beamline to be used to maximize the photon flux. The smaller source sizes and thus beamline exit slits required for the HEG match the need to keep the resolving power of the beamline in line with that of the spectrometer of E/ΔE above 3000 for RIXS experiments. Limited by the mechanical movement of the detector, the LEG covers an energy range from 45 eV to 270 eV in first diffraction order, while the center energy of the HEG in first diffraction order can be tuned from 450 to 1350 eV. The remaining energies are covered in higher orders (up to fourth order for the LEG and second order for the HEG), as illustrated by the respective energy scales on the abscissa in Fig. 1[link]. Besides limiting the necessary detector movement, using higher orders allows to avoid the decrease in resolving power at higher energies due to source and detector broadening. For the LEG, the detector can be operated in normal incidence, while it needs to be tilted to more grazing incidence for the HEG to reduce the effective pixel size and be parallel to the focal plane.

Table 1
Optical parameters of the spectrometer

Element Parameter Low-energy grating (LEG) High-energy grating (HEG)
Housing Entrance arm length 400 mm
  Exit arm length 1387–985 mm 1393–1002 mm
 
Grating Dimensions 25 mm × 150 mm
  Radius of curvature 19828 mm 25322 mm
  Slope error <0.5 µrad
  Grazing incidence angle 2.9–1.8° 1.0°
  Grazing exit angle 10.3–4.6° 4.8–2.8°
  Blaze angle 1.7° 0.9°
  Micro-roughness <0.6 nm RMS
  Central groove density a0 = 600 mm−1 a0 = 1200 mm−1
  VLS parameters a1 = 0.72306 mm−2 a1 = 0.63798 mm−2
    a2 = 7.9087 × 10−6 mm−3 a2 = 5.5736 × 10−4 mm−3
    a3 = 3.889 × 10−6 mm−2 a3 = 3.9074 × 10−6 mm−2
  Groove density n(x) = a0 + a1x + a2x2 + a3x3
  Maximum line density error <0.1 mm−1
 
Detector Dimensions 22.5 mm × 22.5 mm
  Pixel size 11.0 µm × 11.0 µm
[Figure 1]
Figure 1
The different contributions to the spectrometer resolving power for (a) LEG and (b) HEG as a function of photon energy (given in different orders on the abscissa). Variations of source spot size (blue) and CMOS incidence angle (red) are also shown. Note that aberrations (yellow) do not play any role for the LEG and thus do not appear in the shown resolving power range in (a).

As described in the Introduction[link], spectrometer designs exist that also measure a (small) range of incoming photon energies in parallel. However, this needs additional optical elements and only works with very homogeneous samples. The latter is often not the case for applied material systems, and thus this option is not used for our spectrometer.

To optimize the acceptance angle in the non-dispersive direction, the total length of the spectrometer was minimized (1.4–1.8 m depending on the center energy). For maximal acceptance in the dispersive direction, the gratings were placed as close as possible to the sample spot (400 mm) with their length maximized (150 mm), while still meeting the requirements on energy resolution. Some spectrometer designs furthermore use collection mirrors to increase the acceptance angle in the non-dispersive direction (Tokushima et al., 2011[Tokushima, T., Horikawa, Y. & Shin, S. (2011). Rev. Sci. Instrum. 82, 073108.]; Dvorak et al., 2016[Dvorak, J., Jarrige, I., Bisogni, V., Coburn, S. & Leonhardt, W. (2016). Rev. Sci. Instrum. 87, 115109.]; Zhou et al., 2022[Zhou, K.-J., Walters, A., Garcia-Fernandez, M., Rice, T., Hand, M., Nag, A., Li, J., Agrestini, S., Garland, P., Wang, H., Alcock, S., Nistea, I., Nutter, B., Rubies, N., Knap, G., Gaughran, M., Yuan, F., Chang, P., Emmins, J. & Howell, G. (2022). J. Synchrotron Rad. 29, 563-580.]). The gain of this approach is highest for very long spectrometers, while short spectrometers will not profit as much. For this reason, the additional cost, and the more complex alignment, no collection mirrors were used in our design. Overall, our spectrometer has, depending on energy, acceptance angles (non-dispersive × dispersive) between 12.5 mrad × 22.7 mrad and 16.1 mrad × 12.2 mrad for the LEG, and 12.5 mrad × 6.9 mrad and 16.1 mrad × 6.6 mrad for the HEG.

With only one optical element, reflection losses are reduced compared with designs with several optical elements. For the LEG coating, we chose nickel, which gives the highest and most uniform reflectivity for energies up to 800 eV, with incident angles between 2.9° (around 50 eV) and 1.8° (around 270 eV) in first order. For the HEG, the incidence angle is close to 1° for all energies, and a gold coating was chosen. The reflectivity, in particular for the higher orders, is further maximized by using blazed gratings. The blaze angles of the two gratings were optimized to 1.7° (LEG) and 0.9° (HEG) by simulating the grating efficiency with the REFLEC software (Schäfers, 2008[Schäfers, F. (2008). Modern Developments in X-ray and Neutron Optics, Vol. 137 of Springer Series in Optical Sciences, p. 9. Berlin/Heidelberg: Springer.]) to give the best performance for the respective energy ranges. The simulated grating efficiencies are shown in Fig. 2[link] as a function of photon energy, grating and diffraction order. Using higher orders allows to stay close to the optimal blaze angle over the full energy range. For the LEG, this leads to very high efficiencies in all orders, while going to second order costs a factor of about four for the HEG (necessary when a high resolving power is required, see above). Gratings were manufactured by Diffraction Optics Solutions (DIOS) with the parameters listed in Table 1[link], using mechanical ruling and ion milling to adjust the blaze angle.

[Figure 2]
Figure 2
Simulated grating and CMOS sensor quantum efficiencies as a function of photon energy. Grating efficiencies are given for the LEG (black) and the HEG (red) in different orders, as indicated in the graph. The CMOS quantum efficiency (blue) is given for three different incidence angles, with 90° corresponding to normal incidence.

As detector, we used a back-illuminated CMOS sensor with 2048 × 204 pixels of 11 µm × 11 µm from Gpixel Inc. optimized for EUV and soft X-rays (Harada, Teranishi, Watanabe, Zhou, Bogaerts et al., 2019[Ament, L. J. P., van Veenendaal, M., Devereaux, T. P., Hill, J. P. & van den Brink, J. (2011). Rev. Mod. Phys. 83, 705-767.]). For energies below 1000 eV, this sensor has a quantum efficiency of 90% or above in normal incidence, as has been experimentally characterized and simulated (Harada, Teranishi, Watanabe, Zhou, Bogaerts & Wang, 2019[Harada, T., Teranishi, N., Watanabe, T., Zhou, Q., Yang, X., Bogaerts, J. & Wang, X. (2019). Appl. Phys. Expr. 12, 082012.]). Using the same simulation approach with the thickness values of the `dead-layer' (5 nm) and the active layer (9.5 µm) (from Harada, Teranishi, Watanabe, Zhou, Bogaerts & Wang, 2019[Harada, T., Teranishi, N., Watanabe, T., Zhou, Q., Yang, X., Bogaerts, J. & Wang, X. (2019). Appl. Phys. Expr. 12, 082012.]), we have simulated the quantum efficiency for photon energies up to 2000 eV and different incidence angles in Fig. 2[link]. In this approach, the quantum efficiencies are estimated by the product of the transmission of the `dead layer' and the absorption of the active layer (Harada, Teranishi, Watanabe, Zhou, Yang et al., 2019[Harada, T., Teranishi, N., Watanabe, T., Zhou, Q., Bogaerts, J. & Wang, X. (2019). Appl. Phys. Expr. 13, 016502.]). X-ray attenuation lengths were retrieved from the CXRO X-ray database (Henke et al., 1993[Henke, B. L., Gullikson, E. M. & Davis, J. C. (1993). At. Data Nucl. Data Tables, 54, 181-342.]). For normal incidence, the quantum efficiency drops from close to 100% at 1000 eV to ∼50% just below the Si K-edge at ∼1.8 keV. This is mitigated by using the sensor in more grazing incidence with the HEG (which is also necessary to achieve the desired resolving power, as discussed above) with simulated quantum efficiencies above 90% for energies above 600 eV and 15° incidence. Currently, an AXIS-SXRF-EUV camera with the EUV optimized CMOS sensor is installed. In this camera, the sensor is fixed in the flange plane, and we can vary the incidence angle only between ∼75 and 90° incidence. For optimal operation with the HEG, we will upgrade to a custom-built camera with the same sensor that can be rotated over the full range between 0° and 90°. Besides the very high quantum efficiency, the CMOS sensor allows for much higher frame rates (up to 48 Hz) as compared with a CCD camera. As will be shown below, this is essential for using the full potential of the presented spectrometers.

3. Control system

To profit from the very short acquisition times possible with our spectrometer, a very efficient and fast data collection is necessary. In particular, the dead-time per data point must be kept as short as possible, which is challenging for the control system and, during energy scans, the beamline mechanics. The spectrometer is integrated with a Tango server (Götz et al., 2003[Götz, A., Taurel, E., Pons, J. E., Verdier, P., Chaize, J. M., Meyer, J., Poncet, F., Heunen, G. & Götz, E. (2003). Proceedings of the 2003 International Conference on Accelerator and Large Experimental Physics Control Systems (ICALEPCS2003), 13-17 October 2003, Gyeongju, Korea, pp. 220-222. MP705.]) into the X-SPEC control system, which is based on SPEC (Certified Scientific Software, Cambridge, USA). SPEC allows measurements with acquisition dead-times of currently 150–170 ms at X-SPEC, and the use of macros and easy interfacing to other software makes it possible to automatize complex measurement programs. However, during standard stepwise energy scans, dead-times are dominated by the undulator and monochromator movements, which can take a few seconds per energy step. At X-SPEC, we have thus implemented energy scans in which undulator and monochromator are run continuously and in parallel with an adjustable energy change per time. With this, we can achieve speeds of up to 2 eV per second (depending on the energy range), only limited by the need of precise motions with the long lever arms of the monochromator mechanics.

4. Rapid RIXS maps of boron nitride

Since their introduction (Fuchs et al., 2008[Fuchs, O., Zharnikov, M., Weinhardt, L., Blum, M., Weigand, M., Zubavichus, Y., Bär, M., Maier, F., Denlinger, J. D., Heske, C., Grunze, M. & Umbach, E. (2008). Phys. Rev. Lett. 100, 249802.], 2009[Fuchs, O., Weinhardt, L., Blum, M., Weigand, M., Umbach, E., Bär, M., Heske, C., Denlinger, J., Chuang, Y. D., McKinney, W., Hussain, Z., Gullikson, E., Jones, M., Batson, P., Nelles, B. & Follath, R. (2009). Rev. Sci. Instrum. 80, 063103.]; Weinhardt et al., 2009[Weinhardt, L., Fuchs, O., Fleszar, A., Bär, M., Blum, M., Weigand, M., Denlinger, J. D., Yang, W., Hanke, W., Umbach, E. & Heske, C. (2009). Phys. Rev. B, 79, 165305.]), soft X-ray RIXS maps have become a powerful tool, giving – arguably – the most complete electronic structure depiction possible with a single measurement. RIXS maps have been used for fundamental studies of solids (Weinhardt et al., 2009[Weinhardt, L., Fuchs, O., Fleszar, A., Bär, M., Blum, M., Weigand, M., Denlinger, J. D., Yang, W., Hanke, W., Umbach, E. & Heske, C. (2009). Phys. Rev. B, 79, 165305.]; Schlappa et al., 2012[Schlappa, J., Wohlfeld, K., Zhou, K. J., Mourigal, M., Haverkort, M. W., Strocov, V. N., Hozoi, L., Monney, C., Nishimoto, S., Singh, S., Revcolevschi, A., Caux, J.-S., Patthey, L., Rønnow, H. M., van den Brink, J. & Schmitt, T. (2012). Nature, 485, 82-85.]; Huang et al., 2017[Huang, H. Y., Chen, Z. Y., Wang, R.-P., de Groot, F. M. F., Wu, W. B., Okamoto, J., Chainani, A., Singh, A., Li, Z.-Y., Zhou, J.-S., Jeng, H.-T., Guo, G. Y., Park, J.-G., Tjeng, L. H., Chen, C. T. & Huang, D. J. (2017). Nat. Commun. 8, 15929.]), liquids (Weinhardt et al., 2010[Weinhardt, L., Fuchs, O., Blum, M., Bär, M., Weigand, M., Denlinger, J. D., Zubavichus, Y., Zharnikov, M., Grunze, M., Heske, C. & Umbach, E. (2010). J. Electron Spectrosc. Relat. Phenom. 177, 206-211.]; Jeyachandran et al., 2014[Jeyachandran, Y. L., Meyer, F., Nagarajan, S., Benkert, A., Bär, M., Blum, M., Yang, W., Reinert, F., Heske, C., Weinhardt, L. & Zharnikov, M. (2014). J. Phys. Chem. Lett. 5, 4143-4148.]; Yin et al., 2015[Yin, Z., Rajkovic, I., Thekku Veedu, S., Deinert, S., Raiser, D., Jain, R., Fukuzawa, H., Wada, S., Quevedo, W., Kennedy, B., Schreck, S., Pietzsch, A., Wernet, P., Ueda, K., Föhlisch, A. & Techert, S. (2015). Z. Phys. Chem. 229, 1855-1867.]; Kunnus et al., 2016[Kunnus, K., Zhang, W., Delcey, M. G., Pinjari, R. V., Miedema, P. S., Schreck, S., Quevedo, W., Schröder, H., Föhlisch, A., Gaffney, K. J., Lundberg, M., Odelius, M. & Wernet, P. (2016). J. Phys. Chem. B, 120, 7182-7194.]) and gases (Weinhardt et al., 2012[Weinhardt, L., Benkert, A., Meyer, F., Blum, M., Wilks, R. G., Yang, W., Bär, M., Reinert, F. & Heske, C. (2012). J. Chem. Phys. 136, 144311.]; Fouda et al., 2020[Fouda, A. E. A., Seitz, L. C., Hauschild, D., Blum, M., Yang, W., Heske, C., Weinhardt, L. & Besley, N. A. (2020). J. Phys. Chem. Lett. 11, 7476-7482.]; Kjellsson et al., 2021[Kjellsson, L., Ekholm, V., Agåker, M., Såthe, C., Pietzsch, A., Karlsson, H. O., Jaouen, N., Nicolaou, A., Guarise, M., Hague, C., Lüning, J., Chiuzbăian, S. G. & Rubensson, J.-E. (2021). Phys. Rev. A, 103, 022812.]), but have also been used to study applied material systems like solar cells (Hauschild et al., 2021[Hauschild, D., Seitz, L., Gharibzadeh, S., Steininger, R., Jiang, N., Yang, W., Paetzold, U. W., Heske, C. & Weinhardt, L. (2021). Appl. Mater. Interfaces, 13, 53202-53210.]; Wilks et al., 2021[Wilks, R. G., Erbing, A., Sadoughi, G., Starr, D. E., Handick, E., Meyer, F., Benkert, A., Iannuzzi, M., Hauschild, D., Yang, W., Blum, M., Weinhardt, L., Heske, C., Snaith, H. J., Odelius, M. & Bär, M. (2021). J. Phys. Chem. Lett. 12, 3885-3890.]) or batteries (Wu et al., 2020[Wu, J., Yang, Y. & Yang, W. (2020). Dalton Trans. 49, 13519-13527.]; Li et al., 2023[Li, B., Zhuo, Z., Zhang, L., Iadecola, A., Gao, X., Guo, J., Yang, W., Morozov, A. V., Abakumov, A. M. & Tarascon, J.-M. (2023). Nat. Mater. 22, 1370-1379.]).

For applied material systems, the required measurement times for these maps of a few ten minutes up to a few hours still limits the practical use of this approach. This limitation is now overcome by the new spectrometer design (in combination with the X-SPEC beamline) presented in this paper. Fig. 3[link](a) shows the RIXS map of a pressed h-BN pellet, measured with our spectrometer in a conventional step scan with an exposure time of ten seconds per spectrum, which leads to a total measurement time of slightly below half an hour. As can be seen, this map shows little to no noise, which suggests that (much) shorter measurement times are feasible. In a step scan, the practical gain of this, however, is limited by the time needed to move undulator and monochromator to the next energy, which is around 3.5 s in the present case. As discussed above, at X-SPEC we can use continuous scans to avoid this dead-time and, in combination with the short readout times of the CMOS sensor and the high spectrometer transmission, the rRIXS maps shown in Figs. 3[link](b)–3(d) become possible. These maps are collected in second diffraction order with total measurement times of 2, 1 and 0.5 min and exposure times of 1, 0.5 and 0.25 s per spectrum, respectively. At the top of each map, the corresponding XES spectrum, extracted from the map at an excitation energy of 410.2 eV, is shown. The X-SPEC beamline exit slit was set to 200 µm, giving an excitation energy width of ∼0.5 eV and a photon flux of ∼2 × 1013 photons s−1 at 100 mA ring current as estimated from the gold mesh current. Flux values and resolving powers of the X-SPEC beamline as a function of photon energy at smaller slit values are given by Weinhardt, Steininger et al. (2021[Weinhardt, L., Steininger, R., Kreikemeyer-Lorenzo, D., Mangold, S., Hauschild, D., Batchelor, D., Spangenberg, T. & Heske, C. (2021). J. Synchrotron Rad. 28, 609-617.]). Table 2[link] lists the measurement parameters of the RIXS maps presented in Fig. 3[link]. Switching from step scans to continuous scans strongly reduces the dead-time per point, and, even though the rRIXS maps in Figs. 3[link](b) and 3(c) are measured much faster, the dead-time fraction is lower than that of the step scan. Even for the fastest map, the dead-time still only accounts for 40% of the total measurement time.

Table 2
Measurement parameters for the h-BN RIXS maps presented in Fig. 3[link]

  Total time (s) Exposure time per point (s) Dead-time per point (s) Dead-time fraction (%) Step size (eV) Number of spectra
Fig. 3[link](a) 1540 10 3.63 27 0.10 113
Fig. 3[link](b) 118 1 0.168 14 0.11 101
Fig. 3[link](c) 60 0.5 0.167 25 0.13 90
Fig. 3[link](d) 30 0.25 0.167 40 0.16 72
[Figure 3]
Figure 3
N K RIXS maps and non-resonant (hν = 410.2 eV) XES spectra of h-BN measured in second diffraction order (a) with a step scan and (b)–(d) in continuous energy scans. The exposure times and total measurement times are given next to the spectra and RIXS maps, respectively.

In Fig. 4[link], we show cuts from the RIXS maps at selected excitation energies. For comparison and benchmark, similar energies as in a recent experimental and theoretical RIXS study on h-BN (Vinson et al., 2017[Vinson, J., Jach, T., Müller, M., Unterumsberger, R. & Beckhoff, B. (2017). Phys. Rev. B, 96, 205116.]) were chosen. Apart from some changes in relative peak intensities caused by the different scattering angle, our data agree well with that of Vinson et al. (2017[Vinson, J., Jach, T., Müller, M., Unterumsberger, R. & Beckhoff, B. (2017). Phys. Rev. B, 96, 205116.]). Remarkably, our data also exhibit a similar signal-to-noise ratio with exposure times of 0.5 s when compared with 1800–3600 s in Vinson et al. (2017[Vinson, J., Jach, T., Müller, M., Unterumsberger, R. & Beckhoff, B. (2017). Phys. Rev. B, 96, 205116.]). The bottom-most spectrum is excited below the absorption edge with some weak contributions from higher order/harmonics (∼5% of the first order/harmonic intensity) excitation and the elastically scattered photons. Evaluating the elastic peak, we find a combined full width at half-maximum (FWHM) of 0.67 eV, which includes both broadening from the beamline and the spectrometer. With a beamline contribution of ∼0.5 eV, the spectrometer resolution is estimated to 0.45 eV, limited by the spot size on the sample.

[Figure 4]
Figure 4
N K RIXS spectra extracted at selected energies from the maps in Figs. 3[link](a) (bottom-most spectrum, 399.34 eV) and 3(c). The bottom-most spectrum is excited below the N K absorption edge, only showing the elastically scattered line and emission excited with higher orders/harmonic of the beamline.

In addition to fast rRIXS maps, measurements at fixed energies can be performed with frequencies up to 5 Hz, limited by the dead-time of the control system. This can be useful for non-resonant excitation or at an excitation energy selected for best sensitivity to study a desired sample property in a time-resolved way. To further speed up the data collection, we are currently optimizing the control system reducing the acquisition dead-time of the SPEC scan significantly to be only limited by the maximal frequency of the CMOS camera (48 Hz), if the rRIXS signal strength is sufficiently high.

5. Summary

The design of a high-transmission spectrometer for rapid RIXS (rRIXS) maps is presented, together with a demonstration experiment of hexagonal boron nitride. With a compact design, maximized solid angle, optimized grating parameters, the use of a CMOS sensor tailored for soft X-rays, and a streamlined integration into the X-SPEC beamline control system, this spectrometer allows the total measurement time for a full rRIXS map to be reduced to half a minute in a continuous energy scan, with an exposure time of 0.25 s per spectrum. At the same time, the spectrometer achieves a resolving power E/ΔE of 1500 or better with its low-energy grating, and above 3000 with its high-energy grating. Furthermore, the spectrometer covers a very large energy range from 45 to 2000 eV with only two gratings by making use of higher-order reflections. With measurement times of well below 1 s per spectrum, the spectrometer opens the route to study chemical and electronic processes under operation conditions using rapid RIXS.

Acknowledgements

Open access funding enabled and organized by Projekt DEAL.

Funding information

D. Hauschild, L. Weinhardt and C. Heske thank the Deutsche Forschungsgemeinschaft (DFG) for funding in projects GZ:INST 121384/65-1 FUGG and GZ:INST 121384/66-1 FUGG.

References

First citationAment, L. J. P., van Veenendaal, M., Devereaux, T. P., Hill, J. P. & van den Brink, J. (2011). Rev. Mod. Phys. 83, 705–767.  Web of Science CrossRef CAS Google Scholar
First citationBlum, M., Weinhardt, L., Fuchs, O., Bär, M., Zhang, Y., Weigand, M., Krause, S., Pookpanratana, S., Hofmann, T., Yang, W., Denlinger, J. D., Umbach, E. & Heske, C. (2009). Rev. Sci. Instrum. 80, 123102.  Web of Science CrossRef PubMed Google Scholar
First citationCallcott, T. A., Tsang, K. L., Zhang, C. H., Ederer, D. L. & Arakawa, E. T. (1986). Rev. Sci. Instrum. 57, 2680–2690.  CrossRef CAS Web of Science Google Scholar
First citationCarlisle, J. A., Shirley, E. L., Hudson, E. A., Terminello, L. J., Callcott, T. A., Jia, J. J., Ederer, D. L., Perera, R. C. C. & Himpsel, F. J. (1995). Phys. Rev. Lett. 74, 1234–1237.  CrossRef PubMed CAS Web of Science Google Scholar
First citationChiuzbăian, S. G., Ghiringhelli, G., Dallera, C., Grioni, M., Amann, P., Wang, X., Braicovich, L. & Patthey, L. (2005). Phys. Rev. Lett. 95, 197402.  Web of Science PubMed Google Scholar
First citationChiuzbăian, S. G., Hague, C. F., Avila, A., Delaunay, R., Jaouen, N., Sacchi, M., Polack, F., Thomasset, M., Lagarde, B., Nicolaou, A., Brignolo, S., Baumier, C., Lüning, J. & Mariot, J.-M. (2014). Rev. Sci. Instrum. 85, 043108.  Web of Science PubMed Google Scholar
First citationChuang, Y.-D., Feng, X., Glans-Suzuki, P.-A., Yang, W., Padmore, H. & Guo, J.-H. (2020). J. Synchrotron Rad. 27, 695–707.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationChuang, Y.-D., Pepper, J., McKinney, W., Hussain, Z., Gullikson, E., Batson, P., Qian, D. & Hasan, M. Z. (2005). J. Phys. Chem. Solids, 66, 2173–2178.  Web of Science CrossRef CAS Google Scholar
First citationCocco, D., Zangrando, M., Matteucci, M., Bondino, F., Platè, M., Zacchigna, M., Parmigiani, F., Nelles, B. & Prince, K. C. (2004). AIP Conf. Proc. 705, 873–876.  CrossRef CAS Google Scholar
First citationDakovski, G. L., Lin, M.-F., Damiani, D. S., Schlotter, W. F., Turner, J. J., Nordlund, D. & Ogasawara, H. (2017). J. Synchrotron Rad. 24, 1180–1186.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationDallera, C., Puppin, E., Trezzi, G., Incorvaia, N., Fasana, A., Braicovich, L., Brookes, N. B. & Goedkoop, J. B. (1996). J. Synchrotron Rad. 3, 231–238.  CrossRef CAS Web of Science IUCr Journals Google Scholar
First citationDvorak, J., Jarrige, I., Bisogni, V., Coburn, S. & Leonhardt, W. (2016). Rev. Sci. Instrum. 87, 115109.  Web of Science CrossRef PubMed Google Scholar
First citationEich, D., Fuchs, O., Groh, U., Weinhardt, L., Fink, R., Umbach, E., Heske, C., Fleszar, A., Hanke, W., Gross, E. K. U., Bostedt, C., Von Buuren, T., Franco, N., Terminello, L. J., Keim, M., Reuscher, G., Lugauer, H. & Waag, A. (2006). Phys. Rev. B, 73, 115212.  Web of Science CrossRef Google Scholar
First citationErtan, E., Savchenko, V., Ignatova, N., Vaz da Cruz, V., Couto, R. C., Eckert, S., Fondell, M., Dantz, M., Kennedy, B., Schmitt, T., Pietzsch, A., Föhlisch, A., Gel'mukhanov, F., Odelius, M. & Kimberg, V. (2018). Phys. Chem. Chem. Phys. 20, 14384–14397.  Web of Science CrossRef CAS PubMed Google Scholar
First citationFouda, A. E. A., Seitz, L. C., Hauschild, D., Blum, M., Yang, W., Heske, C., Weinhardt, L. & Besley, N. A. (2020). J. Phys. Chem. Lett. 11, 7476–7482.  Web of Science CrossRef CAS PubMed Google Scholar
First citationFuchs, O., Weinhardt, L., Blum, M., Weigand, M., Umbach, E., Bär, M., Heske, C., Denlinger, J., Chuang, Y. D., McKinney, W., Hussain, Z., Gullikson, E., Jones, M., Batson, P., Nelles, B. & Follath, R. (2009). Rev. Sci. Instrum. 80, 063103.  Web of Science CrossRef PubMed Google Scholar
First citationFuchs, O., Zharnikov, M., Weinhardt, L., Blum, M., Weigand, M., Zubavichus, Y., Bär, M., Maier, F., Denlinger, J. D., Heske, C., Grunze, M. & Umbach, E. (2008). Phys. Rev. Lett. 100, 249802.  Web of Science CrossRef PubMed Google Scholar
First citationGhiringhelli, G., Matsubara, M., Dallera, C., Fracassi, F., Tagliaferri, A., Brookes, N. B., Kotani, A. & Braicovich, L. (2006). Phys. Rev. B, 73, 035111.  Web of Science CrossRef Google Scholar
First citationGhiringhelli, G., Piazzalunga, A., Dallera, C., Trezzi, G., Braicovich, L., Schmitt, T., Strocov, V. N., Betemps, R., Patthey, L., Wang, X. & Grioni, M. (2006). Rev. Sci. Instrum. 77, 113108.  Web of Science CrossRef Google Scholar
First citationGötz, A., Taurel, E., Pons, J. E., Verdier, P., Chaize, J. M., Meyer, J., Poncet, F., Heunen, G. & Götz, E. (2003). Proceedings of the 2003 International Conference on Accelerator and Large Experimental Physics Control Systems (ICALEPCS2003), 13–17 October 2003, Gyeongju, Korea, pp. 220–222. MP705.  Google Scholar
First citationHarada, T., Teranishi, N., Watanabe, T., Zhou, Q., Bogaerts, J. & Wang, X. (2019). Appl. Phys. Expr. 13, 016502.  Web of Science CrossRef Google Scholar
First citationHarada, T., Teranishi, N., Watanabe, T., Zhou, Q., Yang, X., Bogaerts, J. & Wang, X. (2019). Appl. Phys. Expr. 12, 082012.  Web of Science CrossRef Google Scholar
First citationHatsui, T., Setoyama, H., Shigemasa, E. & Kosugi, N. (2005). J. Electron Spectrosc. Relat. Phenom. 144–147, 1059–1062.  Web of Science CrossRef CAS Google Scholar
First citationHauschild, D., Seitz, L., Gharibzadeh, S., Steininger, R., Jiang, N., Yang, W., Paetzold, U. W., Heske, C. & Weinhardt, L. (2021). Appl. Mater. Interfaces, 13, 53202–53210.  Web of Science CrossRef CAS Google Scholar
First citationHenke, B. L., Gullikson, E. M. & Davis, J. C. (1993). At. Data Nucl. Data Tables, 54, 181–342.  CrossRef CAS Web of Science Google Scholar
First citationHennies, F., Pietzsch, A., Berglund, M., Föhlisch, A., Schmitt, Th., Strocov, V., Karlsson, H. O., Andersson, J. & Rubensson, J.-E. (2010). Phys. Rev. Lett. 104, 193002.  Web of Science CrossRef PubMed Google Scholar
First citationHeske, C., Groh, U., Fuchs, O., Weinhardt, L., Umbach, E., Schedel-Niedrig, T., Fischer, C. H., Lux-Steiner, M. C., Zweigart, S., Niesen, T. P., Karg, F., Denlinger, J. D., Rude, B., Andrus, C. & Powell, F. (2003). J. Chem. Phys. 119, 10467–10470.  Web of Science CrossRef CAS Google Scholar
First citationHuang, H. Y., Chen, Z. Y., Wang, R.-P., de Groot, F. M. F., Wu, W. B., Okamoto, J., Chainani, A., Singh, A., Li, Z.-Y., Zhou, J.-S., Jeng, H.-T., Guo, G. Y., Park, J.-G., Tjeng, L. H., Chen, C. T. & Huang, D. J. (2017). Nat. Commun. 8, 15929.  Web of Science CrossRef PubMed Google Scholar
First citationJeyachandran, Y. L., Meyer, F., Nagarajan, S., Benkert, A., Bär, M., Blum, M., Yang, W., Reinert, F., Heske, C., Weinhardt, L. & Zharnikov, M. (2014). J. Phys. Chem. Lett. 5, 4143–4148.  Web of Science CrossRef CAS PubMed Google Scholar
First citationJohnson, P. D. & Ma, Y. (1994). Phys. Rev. B, 49, 5024–5027.  CrossRef CAS Web of Science Google Scholar
First citationKjellsson, L., Ekholm, V., Agåker, M., Såthe, C., Pietzsch, A., Karlsson, H. O., Jaouen, N., Nicolaou, A., Guarise, M., Hague, C., Lüning, J., Chiuzbăian, S. G. & Rubensson, J.-E. (2021). Phys. Rev. A, 103, 022812.  Web of Science CrossRef Google Scholar
First citationKunnus, K., Zhang, W., Delcey, M. G., Pinjari, R. V., Miedema, P. S., Schreck, S., Quevedo, W., Schröder, H., Föhlisch, A., Gaffney, K. J., Lundberg, M., Odelius, M. & Wernet, P. (2016). J. Phys. Chem. B, 120, 7182–7194.  Web of Science CrossRef CAS PubMed Google Scholar
First citationLee, S.-J., Titus, C. J., Alonso Mori, R., Baker, M. L., Bennett, D. A., Cho, H.-M., Doriese, W. B., Fowler, J. W., Gaffney, K. J., Gallo, A., Gard, J. D., Hilton, G. C., Jang, H., Joe, Y. I., Kenney, C. J., Knight, J., Kroll, T., Lee, J.-S., Li, D., Lu, D., Marks, R., Minitti, M. P., Morgan, K. M., Ogasawara, H., O'Neil, G. C., Reintsema, C. D., Schmidt, D. R., Sokaras, D., Ullom, J. N., Weng, T.-C., Williams, C., Young, B. A., Swetz, D. S., Irwin, K. D. & Nordlund, D. (2019). Rev. Sci. Instrum. 90, 113101.  Web of Science CrossRef PubMed Google Scholar
First citationLéon, A., Fiedler, A., Blum, M., Benkert, A., Meyer, F., Yang, W., Bär, M., Scheiba, F., Ehrenberg, H., Weinhardt, L. & Heske, C. (2017). J. Phys. Chem. C, 121, 5460–5466.  Google Scholar
First citationLi, B., Zhuo, Z., Zhang, L., Iadecola, A., Gao, X., Guo, J., Yang, W., Morozov, A. V., Abakumov, A. M. & Tarascon, J.-M. (2023). Nat. Mater. 22, 1370–1379.  Web of Science CrossRef CAS PubMed Google Scholar
First citationLu, H., Rossi, M., Nag, A., Osada, M., Li, D. F., Lee, K., Wang, B. Y., Garcia-Fernandez, M., Agrestini, S., Shen, Z. X., Been, E. M., Moritz, B., Devereaux, T. P., Zaanen, J., Hwang, H. Y., Zhou, K.-J. & Lee, W. S. (2021). Science, 373, 213–216.  Web of Science CrossRef CAS PubMed Google Scholar
First citationMarschall, F., Yin, Z., Rehanek, J., Beye, M., Döring, F., Kubiček, K., Raiser, D., Veedu, S. T., Buck, J., Rothkirch, A., Rösner, B., Guzenko, V. A., Viefhaus, J., David, C. & Techert, S. (2017). Sci. Rep. 7, 8849.  Web of Science CrossRef PubMed Google Scholar
First citationMeisel, A., Steuer, I. & Szargan, R. (1968). At. Spectrosc. 23, 527–533.  CrossRef CAS Web of Science Google Scholar
First citationMeyer, F., Blum, M., Benkert, A., Hauschild, D., Nagarajan, S., Wilks, R. G., Andersson, J., Yang, W., Zharnikov, M., Bär, M., Heske, C., Reinert, F. & Weinhardt, L. (2014). J. Phys. Chem. B, 118, 13142–13150.  Web of Science CrossRef CAS PubMed Google Scholar
First citationNagasaka, M., Hatsui, T., Horigome, T., Hamamura, Y. & Kosugi, N. (2010). J. Electron Spectrosc. Relat. Phenom. 177, 130–134.  Web of Science CrossRef CAS Google Scholar
First citationNordgren, J. & Nyholm, R. (1986). Nucl. Instrum. Methods Phys. Res. A, 246, 242–245.  CrossRef Web of Science Google Scholar
First citationO'Bryan, H. M. & Skinner, H. W. B. (1940). Proc. R. Soc. London, 176, 229–262.  CAS Google Scholar
First citationQiao, R., Li, Q., Zhuo, Z., Sallis, S., Fuchs, O., Blum, M., Weinhardt, L., Heske, C., Pepper, J., Jones, M., Brown, A., Spucces, A., Chow, K., Smith, B., Glans, P.-A., Chen, Y., Yan, S., Pan, F., Piper, L. F. J., Denlinger, J., Guo, J., Hussain, Z., Chuang, Y.-D. & Yang, W. (2017). Rev. Sci. Instrum. 88, 033106.  Web of Science CrossRef PubMed Google Scholar
First citationRoychoudhury, S., Zhuo, Z., Qiao, R., Wan, L., Liang, Y., Pan, F., Chuang, Y., Prendergast, D. & Yang, W. (2021). Appl. Mater. Interfaces, 13, 45488–45495.  Web of Science CrossRef CAS Google Scholar
First citationRubensson, J.-E., Söderström, J., Binggeli, C., Gråsjö, J., Andersson, J., Såthe, C., Hennies, F., Bisogni, V., Huang, Y., Olalde, P., Schmitt, T., Strocov, V. N., Föhlisch, A., Kennedy, B. & Pietzsch, A. (2015). Phys. Rev. Lett. 114, 133001.  Web of Science CrossRef PubMed Google Scholar
First citationSchäfers, F. (2008). Modern Developments in X-ray and Neutron Optics, Vol. 137 of Springer Series in Optical Sciences, p. 9. Berlin/Heidelberg: Springer.  Google Scholar
First citationSchlappa, J., Wohlfeld, K., Zhou, K. J., Mourigal, M., Haverkort, M. W., Strocov, V. N., Hozoi, L., Monney, C., Nishimoto, S., Singh, S., Revcolevschi, A., Caux, J.-S., Patthey, L., Rønnow, H. M., van den Brink, J. & Schmitt, T. (2012). Nature, 485, 82–85.  Web of Science CrossRef CAS PubMed Google Scholar
First citationSchunck, J. O., Döring, F., Rösner, B., Buck, J., Engel, R. Y., Miedema, P. S., Mahatha, S. K., Hoesch, M., Petraru, A., Kohlstedt, H., Schüssler-Langeheine, C., Rossnagel, K., David, C. & Beye, M. (2021). Optica, 8, 156–160.  Web of Science CrossRef Google Scholar
First citationSchwanke, C., Golnak, R., Xiao, J. & Lange, K. M. (2014). Rev. Sci. Instrum. 85, 103120.  Web of Science CrossRef PubMed Google Scholar
First citationSeitz, L. C., Doronkin, D. E., Hauschild, D., Casapu, M., Zengel, D., Zimina, A., Kreikemeyer-Lorenzo, D., Blum, M., Yang, W., Grunwaldt, J.-D., Heske, C. & Weinhardt, L. (2022). J. Phys. Chem. C, 126, 20998–21009.  Web of Science CrossRef CAS Google Scholar
First citationShin, S., Agui, A., Fujisawa, M., Tezuka, Y., Ishii, T. & Hirai, N. (1995). Rev. Sci. Instrum. 66, 1584–1586.  CrossRef CAS Web of Science Google Scholar
First citationStrocov, V. N., Schmitt, T., Flechsig, U., Patthey, L. & Chiuzbăian, G. S. (2011). J. Synchrotron Rad. 18, 134–142.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationTokushima, T., Harada, Y., Ohashi, H., Senba, Y. & Shin, S. (2006). Rev. Sci. Instrum. 77, 063107.  Web of Science CrossRef Google Scholar
First citationTokushima, T., Horikawa, Y. & Shin, S. (2011). Rev. Sci. Instrum. 82, 073108.  Web of Science CrossRef PubMed Google Scholar
First citationVale, J. G., Dashwood, C. D., Paris, E., Veiga, L. S. I., Garcia-Fernandez, M., Nag, A., Walters, A., Zhou, K.-J., Pietsch, I.-M., Jesche, A., Gegenwart, P., Coldea, R., Schmitt, T. & McMorrow, D. F. (2019). Phys. Rev. B, 100, 224303.  Web of Science CrossRef Google Scholar
First citationVinson, J., Jach, T., Müller, M., Unterumsberger, R. & Beckhoff, B. (2017). Phys. Rev. B, 96, 205116.  Web of Science CrossRef PubMed Google Scholar
First citationWarwick, T., Chuang, Y.-D., Voronov, D. L. & Padmore, H. A. (2014). J. Synchrotron Rad. 21, 736–743.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationWeinhardt, L., Benkert, A., Meyer, F., Blum, M., Hauschild, D. G., Wilks, R., Bär, M., Yang, W., Zharnikov, M., Reinert, F. & Heske, C. (2019). Phys. Chem. Chem. Phys. 21, 13207–13214.  Web of Science CrossRef CAS PubMed Google Scholar
First citationWeinhardt, L., Benkert, A., Meyer, F., Blum, M., Wilks, R. G., Yang, W., Bär, M., Reinert, F. & Heske, C. (2012). J. Chem. Phys. 136, 144311.  Web of Science CrossRef PubMed Google Scholar
First citationWeinhardt, L., Blum, M., Bär, M., Heske, C., Cole, B., Marsen, B. & Miller, E. L. (2008). J. Phys. Chem. C, 112, 3078–3082.  Web of Science CrossRef CAS Google Scholar
First citationWeinhardt, L., Fuchs, O., Blum, M., Bär, M., Weigand, M., Denlinger, J. D., Zubavichus, Y., Zharnikov, M., Grunze, M., Heske, C. & Umbach, E. (2010). J. Electron Spectrosc. Relat. Phenom. 177, 206–211.  Web of Science CrossRef CAS Google Scholar
First citationWeinhardt, L., Fuchs, O., Fleszar, A., Bär, M., Blum, M., Weigand, M., Denlinger, J. D., Yang, W., Hanke, W., Umbach, E. & Heske, C. (2009). Phys. Rev. B, 79, 165305.  Web of Science CrossRef Google Scholar
First citationWeinhardt, L., Hauschild, D., Steininger, R., Jiang, N., Blum, M., Yang, W. & Heske, C. (2021). Anal. Chem. 93, 8300–8308.  Web of Science CrossRef CAS PubMed Google Scholar
First citationWeinhardt, L., Steininger, R., Kreikemeyer-Lorenzo, D., Mangold, S., Hauschild, D., Batchelor, D., Spangenberg, T. & Heske, C. (2021). J. Synchrotron Rad. 28, 609–617.  Web of Science CrossRef IUCr Journals Google Scholar
First citationWerme, L., Grennberg, B., Nordgren, J., Nordling, C. & Siegbahn, K. (1973). Nature, 242, 453–455.  CrossRef CAS Web of Science Google Scholar
First citationWilks, R. G., Erbing, A., Sadoughi, G., Starr, D. E., Handick, E., Meyer, F., Benkert, A., Iannuzzi, M., Hauschild, D., Yang, W., Blum, M., Weinhardt, L., Heske, C., Snaith, H. J., Odelius, M. & Bär, M. (2021). J. Phys. Chem. Lett. 12, 3885–3890.  Web of Science CrossRef CAS PubMed Google Scholar
First citationWilks, R. G., MacNaughton, J. B., Kraatz, H.-B., Regier, T., Blyth, R. I. R. & Moewes, A. (2009). J. Phys. Chem. A, 113, 5360–5366.  Web of Science CrossRef PubMed CAS Google Scholar
First citationWu, J., Yang, Y. & Yang, W. (2020). Dalton Trans. 49, 13519–13527.  Web of Science CrossRef CAS PubMed Google Scholar
First citationYamane, H., Kosugi, N. & Hatsui, T. (2013). J. Electron Spectrosc. Relat. Phenom. 188, 155–160.  Web of Science CrossRef CAS Google Scholar
First citationYin, Z., Rajkovic, I., Thekku Veedu, S., Deinert, S., Raiser, D., Jain, R., Fukuzawa, H., Wada, S., Quevedo, W., Kennedy, B., Schreck, S., Pietzsch, A., Wernet, P., Ueda, K., Föhlisch, A. & Techert, S. (2015). Z. Phys. Chem. 229, 1855–1867.  Web of Science CrossRef CAS Google Scholar
First citationZhou, K.-J., Walters, A., Garcia-Fernandez, M., Rice, T., Hand, M., Nag, A., Li, J., Agrestini, S., Garland, P., Wang, H., Alcock, S., Nistea, I., Nutter, B., Rubies, N., Knap, G., Gaughran, M., Yuan, F., Chang, P., Emmins, J. & Howell, G. (2022). J. Synchrotron Rad. 29, 563–580.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationZubavichus, Y., Fuchs, O., Weinhardt, L., Heske, C., Umbach, E., Denlinger, J. D. & Grunze, M. (2004). Radiat. Res. 161, 346–358.  Web of Science CrossRef PubMed CAS Google Scholar