Figure 3
Trapezoidal parameters tuning performed with DANTE+ DPP. (a) Energy resolution (quoted as FWHM of the Mn-Kα peak) as a function of the trapezoidal filter's flat top (FT) for varying values of peaking time (PT) parameter. (b) Energy resolution as a function of the trapezoidal filter's PT for varying ICR. (c) Schematic representation of trapezoidal filter parameters. |