Figure 1
Electrostatic field-map simulations of the XIBPM. (a) Original design showing strong local electric-field distortion caused by the MCP high-voltage lead. (b) Redesigned configuration with a conductive shielding cover biased at the MCP potential, effectively suppressing the field distortion and providing more uniform electric field near the detector. |