High-flux ptychographic imaging using the new 55 µm-pixel detector `Lambda' based on the Medipix3 readout chip
aUniversity of Göttingen, Institute for X-ray Physics, Friedrich-Hund-Platz 1, 37077 Göttingen, Germany, and bDeutsches Elektronen-Synchrotron DESY, Notkestrasse 85, 22607 Hamburg, Germany
*Correspondence e-mail: rwilke@gwdg.de
Suitable detection systems that are capable of recording high photon count rates with single-photon detection are instrumental for coherent X-ray imaging. The new single-photon-counting pixel detector `Lambda' has been tested in a ptychographic imaging experiment on solar-cell nanowires using Kirkpatrick–Baez-focused 13.8 keV X-rays. Taking advantage of the high count rate of the Lambda and 5 photons nm−2 s−1 or a of ∼1010 photons s−1 on the sample. By comparison with data taken without the semi-transparent central stop, an increase in resolution by a factor of 3–4 is determined: from about 125 nm to about 38 nm for the nanowire and from about 83 nm to about 21 nm for the illuminating wavefield.
expansion by the semi-transparent central stop, a high-dynamic-range diffraction signal covering more than seven orders of magnitude has been recorded, which corresponds to a density of about 10Keywords: phase retrieval; ptychography; coherent diffractive imaging; semi-transparent central stop; semi-transparent beam stop; Lambda detector; Medipix3 chip; solar-cell nanowires.
1. Introduction
Coherent diffractive imaging (Miao et al., 1999; Chapman & Nugent, 2010; Thibault & Elser, 2010) and, in particular, its variant ptychography (Faulkner & Rodenburg, 2004; Thibault et al., 2008) needs to deal with diffraction patterns spanning orders of magnitude which can be problematic for current photon-counting detectors. In the case of ptychography, this high-dynamic-range problem is more accentuated because both the strong signal of the primary beam and the weak diffracted signal of the probed object need to be recorded. Intended as high-resolution X-ray imaging techniques, the potential resolution is considered to be limited by the invested dose (Howells et al., 2009). Consequently, a high primary-beam intensity is needed to achieve a high resolution. Numerical phasing of the diffraction data thus crucially depends on modern detectors which provide a high An alleviation of the high-dynamic-range problem can be achieved by giving up strict For instance, the MM-PAD (Vernon et al., 2007; Tate et al., 2013) achieved a count rate of 108 photons pixel−1 s−1 at 8 keV X-rays in a recent ptychography study (Giewekemeyer et al., 2014). Another important development in this direction is the Adaptive Gain Integrating Pixel Detector (Potdevin et al., 2009). In addition to the high fast readout, small pixels and low noise are also important. Single-photon-counting pixel detectors such as Pilatus (Kraft et al., 2009) and Maxipix (Llopart et al., 2002) have the advantage of effectively zero readout noise and can reach maximum count rates in the range of 105–107 photons pixel−1 s−1 (Toyokawa et al., 2010; Trueb et al., 2012). The new Large Area Medipix-Based Detector Array (Lambda) is being developed (Pennicard & Graafsma, 2011; Pennicard et al., 2011, 2012) in order to better meet experimental demands. Based on the Medipix3 chip (Ballabriga, Campbell et al., 2011), the Lambda is constructed with two counters for deadtime-free readout. In the standard mode it can operate at 2000 frames s−1. In combination with its small pixel size and the increased maximum photon count rate of about 3 × 105 photons s−1 pixel−1 (Ballabriga, Blaj et al., 2011), in comparison to its predecessor with the Medipix2 chip, the Lambda thus becomes a valuable device in synchrotron applications such as ptychography. Another important detector development, albeit with a larger pixel area (factor 1.86), for while framing at kHz rates is the 75 µm-pixel detector `Eiger' (Dinapoli et al., 2011; Johnson et al., 2012) which, very recently, was successfully operated in a ptychographic imaging experiment (Guizar-Sicairos et al., 2014).
In terms of robust and efficient coherent X-ray diffractive imaging (CDI) algorithms (Gerchberg & Saxton, 1972; Fienup, 1978, 1982; Sayre, 1980; Bauschke et al., 2002; Elser, 2003; Giewekemeyer et al., 2011; Rodriguez et al., 2013) that iteratively invert oversampled diffraction patterns of non-crystalline specimens (Sayre, 1952; Miao et al., 1998), the ptychographic method originating from the field of (Hoppe, 1969a,b; Hoppe & Strube, 1969; Nellist et al., 1995; Nellist & Rodenburg, 1998) has stirred significant interest in the past (Faulkner & Rodenburg, 2004, 2005; Rodenburg et al., 2007; Rodenburg, 2008; Thibault et al., 2008). Ptychography uses the so-called overlap constraint: the object is successively illuminated in such a way that neighbouring illuminated spots share common regions of the object (Bunk et al., 2008). Importantly, this overlap constraint is powerful enough to estimate not only the object but also the illumination function (probe) even in non-ideal experimental conditions, e.g. partial coherence, broad bandpass illumination or a thick object (Maiden & Rodenburg, 2009; Thibault et al., 2009; Thibault & Menzel, 2013; Claus et al., 2013; Edo et al., 2013; Suzuki et al., 2014; Batey et al., 2014; Clark et al., 2014; Enders et al., 2014). Hence, ptychography is also an ideal tool for analysing X-ray wavefields (Schropp et al., 2010; Kewish, Guizar-Sicairos et al., 2010; Kewish, Thibault et al., 2010; Hönig et al., 2011; Vila-Comamala et al., 2011; Wilke et al., 2012; Schropp et al., 2013; Huang et al., 2013). However, achieving optimum resolution in ptychographic imaging remains a challenge. On the algorithmic side, the results can be improved by e.g. correcting for positional errors (Guizar-Sicairos & Fienup, 2008; Maiden et al., 2012; Beckers et al., 2013; Tripathi et al., 2014) but on the experimental side this problem is currently mainly approached by increasing the fluence (e.g. Takahashi et al., 2011, 2013; Guizar-Sicairos et al., 2012; Holler et al., 2012; Schropp et al., 2012; Wilke et al., 2013).
In this work, we investigate the performance of the Lambda detector in a ptychographic imaging experiment on solar-cell nanowires using 13.8 keV photons at the high-brilliance synchrotron radiation source PETRAIII. In order to make use of the full, highly coherent et al., 2012). The ptychographic probe reconstruction enables us to monitor the systematic variation of the beam size due to a change of the numerical aperture in front of the Kirkpatrick–Baez (KB) mirrors. In addition, the fluence on the sample can be accurately determined which, in turn, is used to achieve a fluence-optimized setting. The paper closes with a short discussion of the results in view of the Lambda detector.
of the Kirkpatrick–Baez (KB) focus we test the Lambda in combination with the semi-transparent central stop (Wilke2. Materials and methods
2.1. Experimental setup
The experiments with the Lambda detector were performed at the P10 coherence beamline of the PETRAIII synchrotron located at DESY, Hamburg in Germany. A monochromatic X-ray beam with an energy of 13.8 keV was selected by using a channel-cut monochromator. After passing a set of beam-defining slits, the beam was focused by a pair of X-ray mirrors in the KB geometry (cf. Fig. 1a). The elliptically shaped mirrors (Pd coating) were used at 4.05 mrad incident angle (centre). The focal distances are 302 mm and 200 mm of the vertical mirror (WinlightX) and horizontal mirror (JTEC), respectively (for details see Kalbfleisch et al., 2011; Kalbfleisch, 2012). A soft-edge, rectangular aperture was placed 12 mm upstream of the nominal focal plane to suppress side lobes of the X-ray beam (Takahashi et al., 2013). The scanning of the sample through the X-ray beam was carried out by a high-precision piezo-electric stage (PI, Germany). Both the KB X-ray mirror system and the sample-positioning stages used are part of the Göttingen Instrument for Nano-Imaging with X-rays (GINIX). An evacuated flight tube of 5 m length was installed behind the sample stage to minimize scattering effects of X-rays in air. A positionable semi-transparent central stop (STCS) was installed in the flight tube. The STCS is a piece of germanium with lateral dimensions of about 4 × 4 mm, which was cut from a wafer using a dicing saw Disco dad 321 (Disco, Japan). Additionally, the Ge STCS was ground down to an average nominal thickness of 100 µm using SiC grinding paper with successive grain sizes 2000 and 4000 on a grinding machine LaboPol-21 (Struers). The Lambda detector was placed a few centimetres behind the exit window of the flight tube. The distance between the focal plane and Lambda was measured to be 5.07 m.
2.2. Samples
The nanowires in this work (Fig. 2) were grown and processed as described in Wallentin et al. (2013) and Wallentin (2013). Briefly, (i) gold seed particles were formed on an InP substrate by nano-imprint lithography (Mårtensson et al., 2004), (ii) InP nanowires were grown with vapour-phase (iii) the gold was removed by wet etching, (iv) SiO2 deposition (insulator) by atomic layer deposition and (v) sputter deposition of an optically transparent and conducting layer of indium tin oxide (ITO). After fabrication, the nanowires were cut off the substrate by gently wiping the surface with a soft tissue. Nanowires on the tissue were then allowed to adhere on an Si3N4 membrane (1 µm thickness) (Silson, UK). This process yields randomly distributed nanowires (Fig. 2b), which occasionally are oriented perpendicular to the membrane (Fig. 2c). After the X-ray experiment the sample was sputtered with a 5 nm gold layer for (SEM) imaging.
2.3. Data recording and data treatment
At first, the attenuation profile of the STCS was measured with the Lambda by using an open slit setting of S1 = S2 = 300 × 300 µm and sufficient beam attenuation (Fig. 1b). The rescaling of STCS-attenuated diffraction patterns is exemplified in Fig. 1(c).
The first ptychographic scan (data set 1) covers an area of 3 × 3 µm around the nanowire that was oriented perpendicular to the membrane or `standing' (Fig. 2c). Here, 31 ×31 scan points were taken on a rectangular grid with step sizes of 100 nm. At each point the exposure time was 1 s. Including movements of motors and readout, the total scanning time was about 29 min. The scan of data set 2 was recorded on a nanowire being oriented parallel to the membrane (`lying') with the same parameters as data set 1, except only 26 ×26 scan points were distributed equidistantly over a square of 2.5 × 2.5 µm. For both data sets a small beam-defining slit gap S1 of 50 µm was chosen and the impinging on the sample was reduced by molybdenum foils to about 107 photons s−1 (cf. Table 1). The ptychographic data set 3 was taken on the same nanowire with identical scan parameters as data set 2. The total scanning time was about 20 min in both cases. In contrast, the beam-defining slit gap S1 was opened to 100 µm, which corresponds to an increase in by a factor of about 4. Importantly, no beam attenuation in front of the sample was used and the higher (increase by a factor of ) of the KB beam on the detector was compensated by inserting the STCS in front of the Lambda detector.
|
The design of the Lambda detector is such that the gaps between the chips are filled with pixels larger than the chip pixels (55 × 55 µm). There are 2 ×2 pixels in the intersections of horizontal and vertical gaps with a size of 165 × 165 µm. The remaining pixels in the gaps are 55 × 165 µm (h × v) and 165 × 55 µm (h × v) in the horizontal and the vertical gaps, respectively. The data in the gap pixels were rebinned to the size of the chip pixels.
The ptychographic reconstruction for both object and probe function was carried out using the `ePIE' algorithm (using ; cf. Maiden & Rodenburg, 2009). The algorithmic settings are the same for all three data sets. The reconstruction procedure was initialized using a Gaussian (FWHM = 200 nm) amplitude distribution as probe guess and a unity, pure amplitude plane as object. The number of ptychographic iterations was 600. The final results of the probe and object were obtained by averaging the current estimate over the last 200 iterations.
3. Results
3.1. Ptychographic reconstruction of KB wavefields
The results of the ptychographically reconstructed complex wavefield incident on the sample of data set 1 (S1 gap 50 µm) are presented in Fig. 3. In a first step, the reconstructed probe was numerically propagated along the optical axis by mm. The focal plane was determined to be 240 µm upstream of the sample by using the `sharpness' criterion (e.g. Guizar-Sicairos, 2010):
where denotes the free space propagation of the reconstructed probe field P by . The maximum of determines the focal plane. The complex wavefield of the focal plane is shown in Fig. 3(a). It can be seen that side lobes are more pronounced in the vertical direction. The phase representation in the image reveals an effectively flat phase of the central peak. Line cuts through the maximum of the peak can be seen in Figs. 3(b), 3(c). Here, the difference in side-lobe formation between the horizontal and vertical direction is visible, too. Using a Gaussian fit to determine the size of the focus yields FWHM = 292 nm in the horizontal direction. For the vertical direction FWHM = 393 nm is obtained. A horizontal and a vertical slice through the focus and along the optical axis is shown in Figs. 3(e) and 3(f), respectively. The depth of focus (DOF) extends visibly over a few millimetres. The `sharpness' can also be used to quantify the DOF. In the case of a Gaussian intensity distribution
along the radial direction with Rayleigh length x0 and beam waist w0 the sharpness can be analytically expressed:
Therefore, the normalized sharpness is a measure of the Rayleigh length x0 of the beam, which is equal to half of the DOF. A fit of to the sharpness of the reconstructed probe field yields DOF = 2x0 = 7.32 mm (cf. Fig. 3d).
The reconstructed wavefield of data set 3 (S1 gap 100 µm) is analysed in the same way. In this case, the focal plane almost coincides with the plane of the sample. The distance between sample plane and focal plane is determined to be 16 µm. The intensity distribution of the focal plane reveals a smaller peak with reduced side lobes (cf. Fig. 4a). In the line cuts through the central peak it can be seen that the intensity of the side lobes is reduced by a factor of about 2 in comparison to the other probe field. Gaussian fits to the curves yield a beam width of FWHM = 217 nm and FWHM = 136 nm in the vertical (Fig. 4b) and horizontal (Fig. 4c) direction, respectively. Hence, the decrease in size is by a factor of 1/2.15 in the horizontal direction and by a factor of 1/1.81 in the vertical direction. Moreover, the decrease in size of the beam can also be observed in the slices through the focus and along the optical axis as a decrease in the depth of focus (cf. Figs. 4e, 4f). A fit of to the sharpness of the reconstructed probe field yields DOF = 2x0 = 1.52 mm (cf. Fig. 4d).
3.2. Ptychographic imaging of nanowires
At first, we list the results of the ptychographic reconstructions using a (coherent) 7 photons s−1 and a small beam-defining slit gap S1 of 50 µm (Fig. 5). The image of the reconstructed phase of data set 1 (Fig. 5a) shows the phase shift of the standing nanowire (oriented perpendicular with the membrane, Fig. 2c). A line cut through the phase reconstruction of the nanostructure is drawn in Fig. 5(c). It is indicated in the original figure as a dashed white line. The maximum phase shift of the nanostructure in this graph is about −0.24 rad (after background subtraction). A Gaussian fit to the curve yields a width estimate of 305 nm (FWHM). A typical diffraction pattern of the KB beam imprint on the detector of this data set is shown in Fig. 5(b). The maximum count rate in the diffraction pattern is 91 442 photons s−1 per pixel. It should be noted that the count rate is still below the maximum count rate of 3 × 105 photons s−1 per pixel of the Lambda detector. In Fig. 5(d) the phase of the ptychographic reconstruction of a lying nanowire (oriented parallel to the membrane) can be clearly seen. The length is about 2.6 µm. The extent of the nanowire along the minor axis is in the range of 400–500 nm. The phase shift was estimated from the reconstruction by averaging the phase map over small regions. An offset phase (black frame) was subtracted. On the body of the nanowire the average phase shift yields = rad (white frame) and on the central part of the head the phase shift is increased in magnitude to rad. The pixel size of both reconstructions is 52 nm.
of about 10Next, we consider the result of the ptychographic reconstructions using a (coherent) 9 photons s−1 and a beam-defining slit gap S1 of 100 µm (Fig. 6). A representative diffraction pattern (after correction for the STCS attenuation) of data set 3 is shown in Fig. 6(a). In this setting the Lambda detector can record the diffracted intensity over more than seven orders of magnitude. The KB imprint is obviously larger in comparison to the small S1 gap setting (cf. Fig. 5b). It can be seen that borders of the central part of the beam have been placed close to the intermediate regions between neighbouring chips of the Lambda. Although rescaled, the regions between the chips can be identified in regions of low count rates. However, a smooth rescaling can be obtained if the count rate is sufficiently high. For instance, this can be seen where the tails of the KB beam extend over two chips (cf. Fig. 6b). In addition, the high of the signal reveals the fine features of the tails of the KB beam whose signal lies three to seven orders of magnitude below the central beam signal. Small oscillations in the range of a few pixels are characteristic and are highlighted by an enlarged representation (cf. Fig. 6b).
of 9.8 × 10The phase reconstruction of data set 3 shows the lying nanowire of data set 2 (Fig. 6c). Here, the pixel size is 21 nm. In comparison to the reconstruction of data set 2, the nanowire appears smoother with less background. The resolution is visibly improved. In particular, the shape of the head of the nanowire resembles more the shape that is observed in the SEM images (Fig. 2). A quantification of the resolution follows in the next section. The measured phase shift on parts of the head (dashed, white frame) and the body (white frame) are rad and rad, respectively. A line profile along the minor axis indicates the thickness of the nanowire (Fig. 6d). The shape is well fitted by a Gaussian function with an FWHM = 247 nm.
4. Discussion, conclusion and outlook
Firstly, we discuss the ptychographic reconstructions of the wavefields. The probe reconstruction allows us to compare the KB wavefield between two different optical settings. A change of the slit gap setting causes a change of the size of the KB focus (Matsuyama et al., 2006). Here, the increase of the gap width of the beam-defining slit by a factor of 2 could be shown to result in a decrease in size of the focus approximately by a factor of 2. The result is in agreement with Fourier optics where the focus size decreases linearly with an increase of the numerical aperture indicating full or a high degree of spatial coherence of the beam. In comparison to preceding experiments (Giewekemeyer et al., 2013, 2014; Wilke et al., 2013) where typically a pinhole was used in front of the focal plane, we can show here the effectively pure KB beam, which is only slightly cleaned by the soft-edge aperture. Moreover, the application of the STCS makes it possible to reconstruct the probe from high-dynamic-range intensity distributions, which cover about seven orders of magnitude. The of 9.8 × 109 photons s−1 is still below the maximum coherent of about 1011 photons s−1 for larger slit gaps at the GINIX (cf. Salditt et al., 2011) and thus limited by the geometric dimensions of the STCS which define the maximum slit gaps (cf. Fig. 1b).
Next, the ptychographic reconstructions of the nanowires need to be considered. The phase shift of the lying nanowire should be comparable in both data sets 2 and 3 (Figs. 5d, 6c). Indeed, we find a good agreement of the overall phase shift between both data sets. Moreover, the average phase shift over small parts of the body deviates less than 4% between both data sets. The slight deviation in the head region can be explained by the fact that the phase shift of a rod or cylindrically shaped object is not homogeneous, which makes it difficult to define an average phase shift. In brief, the reconstructions are very consistent. The agreement in phase shift between the data sets holds also for the projection of the standing nanowire. From the SEM images we infer a length of about 2.1 µm and a thickness of about 0.3 µm. Using the phase shift as determined from the ptychographic reconstruction of data set 3, an expected phase shift of rad = −0.29 rad is obtained for a model nanowire that consists of 4/5 of body and 1/5 of head (both of thickness 0.3 µm), which is in good agreement with the reconstruction (Fig. 5a). However, we note a discrepancy in comparison to tabulated values. For instance, the typical width of the InP core of the nanowires shown here is not below 150 nm, which already yields a phase shift of −0.048 rad ( g cm−3) at 13.8 keV (Henke et al., 1993). It is known that a wrong rescaling of the STCS can induce errors in the phase shift (Wilke et al., 2013). On the other hand, the consistency of the two phase maps (with and without STCS) is a good quality control of the STCS rescaling. Therefore we are confident about the quantitativeness of the ptychographic reconstructions with respect to the STCS rescaling.
Finally, the improvement of resolution due to the gain in ) that give about 300–350 nm. A comparison of the two-dimensional power spectral densities (PSDs) of the reconstructed phase maps (cf. Fig. 7) reveals that most information on the nanowire (the shape) is contained within a frequency ring that corresponds to a resolution of about 125 nm. In the case of data set 3, the inner frequency part appears to be slightly larger but there are also contributions for structure sizes of 38 nm (cf. white arrow in Fig. 7b). Another estimate of the resolution that also estimates the correctness of the phasing of the algorithm can be obtained by calculating the so-called `phase-retrieval transfer function' (PRTF) (cf. Shapiro et al., 2005; Chapman et al., 2006; Giewekemeyer et al., 2010):
from data set 2 to 3 needs to be addressed. Here we could achieve a visible improvement of the reconstruction in terms of resolution and background. However, we also note that the diameter of the nanowire in the reconstruction is larger than expected from SEM images (Fig. 2Here, IRj and IMj denote the reconstructed and measured intensity at position j (IRj corresponds to the object and probe being averaged over the last iterations), respectively. indicates azimuthal averaging, i.e. averaging over = const. The PRTFs of data set 1 (red) and 2 (green) are very similar. Both curves drop below 0.5 at about µm−1 (half-period) indicating that the resolution of these data sets is not better than nm. In contrast, the PRTF of data set 3 (blue) remains close to unity without any significant descent. It should be noted that this only indicates a resolution of the probe field in the range of the pixel size (dx = 21 nm), i.e. the contributions of the object are `hidden' (Dierolf et al., 2010; Giewekemeyer et al., 2010). The resolution of the object cannot be better than 38 nm as the PSD analysis revealed. As can be seen by comparing the images in Figs. 5(d) and 6(c), little more information about the structure of the nanowires could be obtained. For instance, in the optimal case one would be able to observe a deviation from the smooth Gaussian curve across the nanowire rod (cf. Fig. 6d) due to the different layers of material.
Next, let us compare the increase in fluence, which can be used to estimate the potential gain in resolution due to the increase in dose. The dose D depends on the photon energy E, the μ, the and the fluence F (Howells et al., 2009):
where denotes the two-dimensional photon density and is the characteristic function of the probed area. In order to account for the overlap of the ptychographic data is obtained by scaling the normalized intensity of the probe at each scan position j to the actual photon number Nj (Wilke et al., 2012):
Here we define the fluence on the sample per single recorded diffraction pattern by taking only the photons of the reconstructed single probe intensity IP around the central peak into account, i.e. and χ corresponds to an ellipse whose minor and major axes are defined by the half width at half-maximum (FWHM/2) of the intensity distribution. The fluence of the low-flux data set 2 is estimated to be photons nm−2 and the high-flux data set 3 yields photons nm−2 (equivalent to a of 105 photons nm−2 s−1). On the other hand, the overall average fluence over the area of the sample which accounts for the overlap [equation (3)] yields an even higher fluence of photons nm−2 for data set 3. Hence, the STCS in combination with the beam resizing allows the fluence on the sample to be increased by three to four orders of magnitude. At first, we note that the fluence obtained here is quite high. For instance, 2.75 ×104 photons nm−2 has been reported by Schropp et al. (2012) using 15.25 keV X-rays focused down to ∼80 × 80 nm (smaller by a factor of 5). Schropp and co-workers estimated their resolution on a 500 m-thick binary tantalum test structure to be about 10 nm. Hence, at least on test structures with sharp edges, the potential resolution should be in the range of 10 nm, which is in good agreement with preceding experiments (Wilke et al., 2013). One way of explaining the discrepancy between expected and observed resolution would be to attribute the result to non-ideal experimental parameters such as the effects of partial coherence. However, the PRTF of data set 3 indicates that at least the probe reconstruction appears to be very good up to the highest frequency. Hence, vibrations of the sample as discussed by e.g. Clark et al. (2011) may explain our results, also with respect to the observed deviation of the phase shift of the nanowire from tabulated values. Another degrading factor could be fluorescence radiation from the Ge STCS. In addition, a supposedly minor effect may be small sensitivity variations of the pixels of the Lambda detector (Schavkan et al., 2013), which could be compensated by a flatfield correction.
In conclusion, we have shown a powerful ptychographical experimental setting that paves the way for experiments with unprecedented K edges such as Ni, Cu, Zn and Br in comparison to earlier experiments using 7.9 keV. In addition, the reconstructions clearly benefit from the effectively non-missing data problem between adjacent chips of the detector, which is a common problem for many other pixel detectors. The high count rate of the Lambda detector in combination with the STCS and the reconstruction of the probe field has enabled us to set up a fluence-optimized ptychographic setting. As far as we could deduce from other works in this field, the of about 105 photons nm−2 s−1 is the highest one that has been reported at synchrotron storage rings for ptychographic imaging experiments in combination with photon-counting pixel detectors (cf. Table 2). In addition, there are two more features of the Lambda detector that have not been addressed in these experiments but which may become important for following experiments. Firstly, the small pixel size of 55 µm in comparison to e.g. 172 µm (Pilatus, Dectris) relaxes oversampling requirements. In simple terms, oversampling limits the ratio between beam size and field of view (FOV) in the sample plane per single diffraction pattern (Miao et al., 1998). According to the small-angle approximation the FOV per single diffraction pattern depends on the wavelength λ, the distance between sample and detector X and the pixel size of the detector D, i.e. . The smaller pixel size thus allows a large field of view (here ∼8 × 8 µm), which in principle opens the possibility for large beam experiments. Consequently, the total scanned area of the sample can be increased while the number of scan points is kept constant. The small pixel size also alleviates the use of higher energies (or smaller distances between sample and detector). Secondly, the 1010 photons s−1 of the beam can also be distributed in favour of shorter exposure times. Thus reduced scanning times would presumably increase the quality of reconstructions due to a decrease of thermal drifts and vibrations. The Lambda can be operated in a continuous read–write mode and it provides a read-out of about 2000 frames s−1. Hence, the detector is very well suited for short exposure times, and it may become important in experiments where deadtime is crucial.
We have presented first ptychographic reconstructions using the new Lambda detector. The use of a higher energy of 13.8 keV is an improvement, as future experiments investigating elemental constituents of the sample will take advantage of the wider range of availableAcknowledgements
We are grateful to Mike Kanbach for the preparation of the STCS and it is our pleasure to thank Damir Asoli, Maria Huffman, Ingvar Åberg and Martin Magnusson for the nanowire solar-cell sample and Klaus Giewekemeyer for discussions. Financial support by SFB 755 Nanoscale Photonic Imaging of the Deutsche Forschungsgemeinschaft is gratefully acknowledged and J. Wallentin acknowledges a fellowship by the KA Wallenberg Foundation.
References
Ballabriga, R., Blaj, G., Campbell, M., Fiederle, M., Greiffenberg, D., Heijne, E. H. M., Llopart, X., Plackett, R., Procz, S., Tlustos, L., Turecek, D. & Wong, W. (2011). J. Instrum. 6, C01052. Web of Science CrossRef Google Scholar
Ballabriga, R., Campbell, M., Heijne, E., Llopart, X., Tlustos, L. & Wong, W. (2011). Nucl. Instrum. Methods Phys. Res. Sect. A, pp. 633 Supplement 1, S15-S18. Google Scholar
Batey, D. J., Claus, D. & Rodenburg, J. M. (2014). Ultramicroscopy, 138, 13–21. Web of Science CrossRef CAS PubMed Google Scholar
Bauschke, H. H., Combettes, P. L. & Luke, D. R. (2002). J. Opt. Soc. Am. A, 19, 1334–1345. Web of Science CrossRef Google Scholar
Beckers, M., Senkbeil, T., Gorniak, T., Giewekemeyer, K., Salditt, T. & Rosenhahn, A. (2013). Ultramicroscopy, 126, 44–47. Web of Science CrossRef CAS PubMed Google Scholar
Bunk, O., Dierolf, M., Kynde, S., Johnson, I., Marti, O. & Pfeiffer, F. (2008). Ultramicroscopy, 108, 481–487. Web of Science CrossRef PubMed CAS Google Scholar
Chapman, H. N., Barty, A., Marchesini, S., Noy, A., Hau-Riege, S. P., Cui, C., Howells, M. R., Rosen, R., He, H., Spence, J. C. H., Weierstall, U., Beetz, T., Jacobsen, C. & Shapiro, D. (2006). J. Opt. Soc. Am. A, 23, 1179–1200. Web of Science CrossRef Google Scholar
Chapman, H. N. & Nugent, K. A. (2010). Nat. Photon. 4, 833–839. Web of Science CrossRef CAS Google Scholar
Clark, J. N., Huang, X., Harder, R. J. & Robinson, I. K. (2014). Phys. Rev. Lett. 112, 113901. Web of Science CrossRef PubMed Google Scholar
Clark, J. N., Putkunz, C. T., Curwood, E. K., Vine, D. J., Scholten, R., McNulty, I., Nugent, K. A. & Peele, A. G. (2011). Opt. Lett. 36, 1954–1956. CrossRef PubMed Google Scholar
Claus, D., Robinson, D. J., Chetwynd, D. G., Shuo, Y., Pike, W. T., Garcia, J. J. D. J. T. & Rodenburg, J. M. (2013). J. Opt. 15, 035702. Web of Science CrossRef Google Scholar
Dierolf, M., Thibault, P., Menzel, A., Kewish, C. M., Jefimovs, K., Schlichting, I., von König, K., Bunk, O. & Pfeiffer, F. (2010). New J. Phys. 12, 035017. Web of Science CrossRef Google Scholar
Dinapoli, R., Bergamaschi, A., Henrich, B., Horisberger, R., Johnson, I., Mozzanica, A., Schmid, E., Schmitt, B., Schreiber, A., Shi, X. & Theidel, G. (2011). Nucl. Instrum. Methods Phys. Res. Sect. A, 650, 79–83. Web of Science CrossRef CAS Google Scholar
Edo, T. B., Batey, D. J., Maiden, A. M., Rau, C., Wagner, U., Pešić, Z. D., Waigh, T. A. & Rodenburg, J. M. (2013). Phys. Rev. A, 87, 053850. Web of Science CrossRef Google Scholar
Elser, V. (2003). J. Opt. Soc. Am. A, 20, 40–55. Web of Science CrossRef Google Scholar
Enders, B., Dierolf, M., Cloetens, P., Stockmar, M., Pfeiffer, F. & Thibault, P. (2014). Appl. Phys. Lett. 104, 171104. Web of Science CrossRef Google Scholar
Faulkner, H. M. & Rodenburg, J. M. (2004). Phys. Rev. Lett. 93, 023903. Web of Science CrossRef PubMed Google Scholar
Faulkner, H. M. & Rodenburg, J. M. (2005). Ultramicroscopy, 103, 153–164. Web of Science CrossRef PubMed CAS Google Scholar
Fienup, J. R. (1978). Opt. Lett. 3, 27–29. CrossRef PubMed CAS Web of Science Google Scholar
Fienup, J. R. (1982). Appl. Opt. 21, 2758–2769. CrossRef CAS PubMed Web of Science Google Scholar
Gerchberg, R. W. & Saxton, W. O. (1972). Optik (Jena), 35, 237–246. Google Scholar
Giewekemeyer, K., Krüger, S. P., Kalbfleisch, S., Bartels, M., Beta, C. & Salditt, T. (2011). Phys. Rev. A, 83, 023804. Web of Science CrossRef Google Scholar
Giewekemeyer, K., Philipp, H. T., Wilke, R. N., Aquila, A., Osterhoff, M., Tate, M. W., Shanks, K. S., Zozulya, A. V., Salditt, T., Gruner, S. M. & Mancuso, A. P. (2014). J. Synchrotron Rad. 21, 1167–1174. Web of Science CrossRef IUCr Journals Google Scholar
Giewekemeyer, K., Thibault, P., Kalbfleisch, S., Beerlink, A., Kewish, C. M., Dierolf, M., Pfeiffer, F. & Salditt, T. (2010). Proc. Natl Acad. Sci. USA, 107, 529–534. Web of Science CrossRef CAS PubMed Google Scholar
Giewekemeyer, K., Wilke, R. N., Osterhoff, M., Bartels, M., Kalbfleisch, S. & Salditt, T. (2013). J. Synchrotron Rad. 20, 490–497. Web of Science CrossRef CAS IUCr Journals Google Scholar
Guizar-Sicairos, M. (2010). PhD thesis, University of Rochester, USA. Google Scholar
Guizar-Sicairos, M. & Fienup, J. R. (2008). Opt. Express, 16, 7264–7278. Web of Science PubMed Google Scholar
Guizar-Sicairos, M., Holler, M., Diaz, A., Vila-Comamala, J., Bunk, O. & Menzel, A. (2012). Phys. Rev. B, 86, 100103. Google Scholar
Guizar-Sicairos, M., Johnson, I., Diaz, A., Holler, M., Karvinen, P., Stadler, H.-C., Dinapoli, R., Bunk, O. & Menzel, A. (2014). Opt. Express, 22, 14859–14870. Web of Science PubMed Google Scholar
Henke, B. L., Gullikson, E. M. & Davis, J. C. (1993). Atom. Data Nucl. Data, 54, 181–342. CrossRef CAS Web of Science Google Scholar
Holler, M., Raabe, J., Diaz, A., Guizar-Sicairos, M., Quitmann, C., Menzel, A. & Bunk, O. (2012). Rev. Sci. Instrum. 83, 073703. Web of Science CrossRef PubMed Google Scholar
Hönig, S., Hoppe, R., Patommel, J., Schropp, A., Stephan, S., Schöder, S., Burghammer, M. & Schroer, C. G. (2011). Opt. Express, 19, 16324–16329. Web of Science PubMed Google Scholar
Hoppe, W. (1969a). Acta Cryst. A25, 495–501. CrossRef IUCr Journals Web of Science Google Scholar
Hoppe, W. (1969b). Acta Cryst. A25, 508–514. CrossRef IUCr Journals Web of Science Google Scholar
Hoppe, W. & Strube, G. (1969). Acta Cryst. A25, 502–507. CrossRef IUCr Journals Web of Science Google Scholar
Howells, M., Beetz, T., Chapman, H., Cui, C., Holton, J., Jacobsen, C., Kirz, J., Lima, E., Marchesini, S., Miao, H., Sayre, D., Shapiro, D., Spence, J. & Starodub, D. (2009). J. Electron Spectrosc. 170, 4–12. Web of Science CrossRef CAS Google Scholar
Huang, X., Yan, H., Nazaretski, E., Conley, R., Bouet, N., Zhou, J., Lauer, K., Li, L., Eom, D., Legnini, D., Harder, R., Robinson, I. K. & Chu, Y. S. (2013). Sci. Rep. 3, doi:10.1038/srep03562. Google Scholar
Johnson, I., Bergamaschi, A., Buitenhuis, J., Dinapoli, R., Greiffenberg, D., Henrich, B., Ikonen, T., Meier, G., Menzel, A., Mozzanica, A., Radicci, V., Satapathy, D. K., Schmitt, B. & Shi, X. (2012). J. Synchrotron Rad. 19, 1001–1005. Web of Science CrossRef CAS IUCr Journals Google Scholar
Kalbfleisch, S. (2012). PhD thesis, Georg-August Universität, Göttingen, Germany. Google Scholar
Kalbfleisch, S., Neubauer, H., Krüger, S. P., Bartels, M., Osterhoff, M., Mai, D. D., Giewekemeyer, K., Hartmann, B., Sprung, M. & Salditt, T. (2011). AIP Conf. Proc. 1365, 96–99. CrossRef Google Scholar
Kewish, C. M., Guizar-Sicairos, M., Liu, C., Qian, J., Shi, B., Benson, C., Khounsary, A. M., Vila-Comamala, J., Bunk, O., Fienup, J. R., Macrander, A. T. & Assoufid, L. (2010). Opt. Express, 18, 23420–23427. Web of Science CrossRef PubMed Google Scholar
Kewish, C. M., Thibault, P., Dierolf, M., Bunk, O., Menzel, A., Vila-Comamala, J., Jefimovs, K. & Pfeiffer, F. (2010). Ultramicroscopy, 110, 325–329. Web of Science CrossRef CAS PubMed Google Scholar
Kraft, P., Bergamaschi, A., Broennimann, Ch., Dinapoli, R., Eikenberry, E. F., Henrich, B., Johnson, I., Mozzanica, A., Schlepütz, C. M., Willmott, P. R. & Schmitt, B. (2009). J. Synchrotron Rad. 16, 368–375. Web of Science CrossRef CAS IUCr Journals Google Scholar
Llopart, X., Campbell, M., Dinapoli, R., San Segundo, D. & Pernigotti, E. (2002). IEEE Trans. Nucl. Sci. 49, 2279–2283. Web of Science CrossRef Google Scholar
Maiden, A. M., Humphry, M. J., Sarahan, M. C., Kraus, B. & Rodenburg, J. M. (2012). Ultramicroscopy, 120, 64–72. Web of Science CrossRef CAS PubMed Google Scholar
Maiden, A. M. & Rodenburg, J. M. (2009). Ultramicroscopy, 109, 1256–1262. Web of Science CrossRef PubMed CAS Google Scholar
Mårtensson, T., Carlberg, P., Borgström, M., Montelius, L., Seifert, W. & Samuelson, L. (2004). Nano Lett. 4, 699–702. Google Scholar
Matsuyama, S., Mimura, H., Yumoto, H., Sano, Y., Yamamura, K., Yabashi, M., Nishino, Y., Tamasaku, K., Ishikawa, T. & Yamauchi, K. (2006). Rev. Sci. Instrum. 77, 103102. Web of Science CrossRef Google Scholar
Miao, J., Charalambous, P., Kirz, J. & Sayre, D. (1999). Nature (London), 400, 342–344. Web of Science CrossRef CAS Google Scholar
Miao, J., Sayre, D. & Chapman, H. N. (1998). J. Opt. Soc. Am. A, 15, 1662–1669. Web of Science CrossRef Google Scholar
Nellist, P. D., McCallum, B. C. & Rodenburg, J. M. (1995). Nature (London), 374, 630–632. CrossRef CAS Web of Science Google Scholar
Nellist, P. D. & Rodenburg, J. M. (1998). Acta Cryst. A54, 49–60. Web of Science CrossRef CAS IUCr Journals Google Scholar
Pennicard, D. & Graafsma, H. (2011). J. Instrum. 6, P06007. Google Scholar
Pennicard, D., Lange, S., Smoljanin, S., Becker, J., Hirsemann, H., Epple, M. & Graafsma, H. (2011). J. Instrum. 6, C11009. Web of Science CrossRef Google Scholar
Pennicard, D., Lange, S., Smoljanin, S., Hirsemann, H. & Graafsma, H. (2012). J. Instrum. 7, C11009. Web of Science CrossRef Google Scholar
Potdevin, G., Trunk, U. & Graafsma, H. (2009). Nucl. Instrum. Methods Phys. Res. Sect. A, 607, 51–54. Web of Science CrossRef CAS Google Scholar
Rodenburg, J. (2008). Adv. Imag. Electron Phys. 150, 87–184. Web of Science CrossRef Google Scholar
Rodenburg, J. M., Hurst, A. C., Cullis, A. G., Dobson, B. R., Pfeiffer, F., Bunk, O., David, C., Jefimovs, K. & Johnson, I. (2007). Phys. Rev. Lett. 98, 034801. Web of Science CrossRef PubMed Google Scholar
Rodriguez, J. A., Xu, R., Chen, C.-C., Zou, Y. & Miao, J. (2013). J. Appl. Cryst. 46, 312–318. Web of Science CrossRef CAS IUCr Journals Google Scholar
Salditt, T., Kalbfleisch, S., Osterhoff, M., Krüger, S. P., Bartels, M., Giewekemeyer, K., Neubauer, H. & Sprung, M. (2011). Opt. Express, 19, 9656–9675. Web of Science CrossRef CAS PubMed Google Scholar
Sayre, D. (1952). Acta Cryst. 5, 843. CrossRef IUCr Journals Web of Science Google Scholar
Sayre, D. (1980). Imaging Processes and Coherence in Physics, Vol. 112, edited by M. Schlenker, M. Fink, J. P. Goedgebuer, C. Malgrange, J. Ch. Vieenot & R. H. Wade, pp. 229–235. Berlin: Springer. Google Scholar
Schavkan, A., Westermeier, F., Zozulya, A., Bondarenko, S., Grübel, G., Schroer, C. & Sprung, M. (2013). J. Phys. Conf. Ser. 425, 202004. CrossRef Google Scholar
Schropp, A. et al. (2010). Appl. Phys. Lett. 96, 091102. Web of Science CrossRef Google Scholar
Schropp, A. et al. (2013). Sci. Rep. 3, doi:10.1038/srep01633. Web of Science CrossRef Google Scholar
Schropp, A., Hoppe, R., Patommel, J., Samberg, D., Seiboth, F., Stephan, S., Wellenreuther, G., Falkenberg, G. & Schroer, C. G. (2012). Appl. Phys. Lett. 100, 253112. Web of Science CrossRef Google Scholar
Shapiro, D., Thibault, P., Beetz, T., Elser, V., Howells, M., Jacobsen, C., Kirz, J., Lima, E., Miao, H., Neiman, A. M. & Sayre, D. (2005). Proc. Natl Acad. Sci. USA, 102, 15343–15346. Web of Science CrossRef PubMed CAS Google Scholar
Suzuki, A., Furutaku, S., Shimomura, K., Yamauchi, K., Kohmura, Y., Ishikawa, T. & Takahashi, Y. (2014). Phys. Rev. Lett. 112, 053903. Web of Science CrossRef PubMed Google Scholar
Takahashi, Y., Suzuki, A., Furutaku, S., Yamauchi, K., Kohmura, Y. & Ishikawa, T. (2013). Appl. Phys. Lett. 102, 094102. Web of Science CrossRef Google Scholar
Takahashi, Y., Suzuki, A., Zettsu, N., Kohmura, Y., Senba, Y., Ohashi, H., Yamauchi, K. & Ishikawa, T. (2011). Phys. Rev. B, 83, 214109. Web of Science CrossRef Google Scholar
Tate, M. W., Chamberlain, D., Green, K. S., Philipp, H. T., Purohit, P., Strohman, C. & Gruner, S. M. (2013). J. Phys. Conf. Ser. 425, 062004. CrossRef Google Scholar
Thibault, P., Dierolf, M., Bunk, O., Menzel, A. & Pfeiffer, F. (2009). Ultramicroscopy, 109, 338–343. Web of Science CrossRef PubMed CAS Google Scholar
Thibault, P., Dierolf, M., Menzel, A., Bunk, O., David, C. & Pfeiffer, F. (2008). Science, 321, 379–382. Web of Science CrossRef PubMed CAS Google Scholar
Thibault, P. & Elser, V. (2010). Annu. Rev. Condens. Matter Phys. 1, 237–255. Web of Science CrossRef CAS Google Scholar
Thibault, P. & Menzel, A. (2013). Nature (London), 494, 68–71. Web of Science CrossRef CAS PubMed Google Scholar
Toyokawa, H., Broennimann, C., Eikenberry, E., Henrich, B., Kawase, M., Kobas, M., Kraft, P., Sato, M., Schmitt, B., Suzuki, M., Tanida, H. & Uruga, T. (2010). Nucl. Instrum. Methods Phys. Res. Sect. A, 623, 204–206. Web of Science CrossRef CAS Google Scholar
Tripathi, A., McNulty, I. & Shpyrko, O. G. (2014). Opt. Express, 22, 1452–1466. Web of Science CrossRef PubMed Google Scholar
Trueb, P., Sobott, B. A., Schnyder, R., Loeliger, T., Schneebeli, M., Kobas, M., Rassool, R. P., Peake, D. J. & Broennimann, C. (2012). J. Synchrotron Rad. 19, 347–351. Web of Science CrossRef CAS IUCr Journals Google Scholar
Vernon, W., Allin, M., Hamlin, R., Hontz, T., Nguyen, D., Augustine, F., Gruner, S. M., Xuong, N. H., Schuette, D. R., Tate, M. W. & Koerner, L. J. (2007). Proc. SPIE, 6706, 67060U. CrossRef Google Scholar
Vila-Comamala, J., Diaz, A., Guizar-Sicairos, M., Mantion, A., Kewish, C. M., Menzel, A., Bunk, O. & David, C. (2011). Opt. Express, 19, 21333–21344. Web of Science PubMed Google Scholar
Wallentin, J. (2013). PhD thesis, Lund University, Lund, Sweden. Google Scholar
Wallentin, J., Anttu, N., Asoli, D., Huffman, M., Aberg, I., Magnusson, M. H., Siefer, G., Fuss-Kailuweit, P., Dimroth, F., Witzigmann, B., Xu, H. Q., Samuelson, L., Deppert, K. & Borgström, M. T. (2013). Science, 339, 1057–1060. Web of Science CrossRef CAS PubMed Google Scholar
Wilke, R. N., Priebe, M., Bartels, M., Giewekemeyer, K., Diaz, A., Karvinen, P. & Salditt, T. (2012). Opt. Express, 20, 19232–19254. Web of Science CrossRef CAS PubMed Google Scholar
Wilke, R. N., Vassholz, M. & Salditt, T. (2013). Acta Cryst. A69, 490–497. Web of Science CrossRef CAS IUCr Journals Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.