Demonstration of neutron time-of-flight diffraction with an event-mode imaging detector

crossmark logo

aMaterial Science and Technology Division, Los Alamos National Laboratory, Los Alamos, NM 87545, USA, bInstitut für Kernphysik, TU Darmstadt, Schlossgartenstrasse 9, 64289 Darmstadt, Germany, cForschungs-Neutronenquelle Heinz Maier-Leibnitz, TU München, Lichtenbergstrasse 1, 85748 Garching, Germany, dEuropean Spallation Source ERIC, Data Management and Software Centre, Asmussens Allé 305, 2800 Lyngby, Denmark, eSecond Target Station, Oak Ridge National Laboratory, 1 Bethel Valley Road, Oak Ridge, TN 37830, USA, fMaterials Science & Engineering, University of Virginia, 395 McCormick Road, Charlottesville, VA 22904, USA, gNeutron Optics and Polarization Group, Oak Ridge National Laboratory, 1 Bethel Valley Road, Oak Ridge, TN 37830, USA, hEuropean Spallation Source ERIC, PO Box 176, SE-22100 Lund, Sweden, iPhysics Division, Los Alamos National Laboratory, Los Alamos, NM 87545, USA, and jDepartment of Industrial Engineering, University of Trento, 38123 Trento, Italy
*Correspondence e-mail: sven@lanl.gov

Edited by G. J. McIntyre, Australian Nuclear Science and Technology Organisation, Lucas Heights, Australia (Received 10 October 2023; accepted 13 May 2024; online 11 July 2024)

This article is part of a collection of articles from the IUCr 2023 Congress in Melbourne, Australia, and commemorates the 75th anniversary of the IUCr.

Neutron diffraction beamlines have traditionally relied on deploying large detector arrays of 3He tubes or neutron-sensitive scintillators coupled with photomultipliers to efficiently probe crystallographic and microstructure information of a given material. Given the large upfront cost of custom-made data acquisition systems and the recent scarcity of 3He, new diffraction beamlines or upgrades to existing ones demand innovative approaches. This paper introduces a novel Timepix3-based event-mode imaging neutron diffraction detector system as well as first results of a silicon powder diffraction measurement made at the HIPPO neutron powder diffractometer at the Los Alamos Neutron Science Center. Notably, these initial measurements were conducted simultaneously with the 3He array on HIPPO, enabling direct comparison. Data reduction for this type of data was implemented in the MAUD code, enabling Rietveld analysis. Results from the Timepix3-based setup and HIPPO were benchmarked against McStas simulations, showing good agreement for peak resolution. With further development, systems such as the one presented here may substantially reduce the cost of detector systems for new neutron instrumentation as well as for upgrades of existing beamlines.

1. Introduction

Neutron scattering is a widely used technique for investigating the crystal structure (e.g. lattice parameters and atom positions) and microstructure (e.g. texture, defects, phase fractions) of materials. Covering large areas with neutron detectors is key to efficient scattering experiments (Diawara, 2023[Diawara, Y. (2023). Neutron Detectors for Scattering Applications. Geneva: Springer Nature.]). Most modern neutron scattering instruments rely on 3He gas, 10B or 6Li doped scintillators as the neutron-active component (Kirstein et al., 2014[Kirstein, O., Hall-Wilton, R., Stefanescu, I., Etxegarai, M., Anastasopoulos, M., Fissum, K., Gulyachkina, A., Höglund, C., Imam, M., Kanaki, K. & Khaplanov, A. (2014). arXiv:1411.6194.]; Stefanescu et al., 2017[Stefanescu, I., Christensen, M., Fenske, J., Hall-Wilton, R., Henry, P. F., Kirstein, O., Müller, M., Nowak, G., Pooley, D., Raspino, D., Rhodes, N., Šaroun, J., Schefer, J., Schooneveld, E., Sykora, J. & Schweika, W. (2017). J. Instrum. 12, P01019.]; Diawara, 2023[Diawara, Y. (2023). Neutron Detectors for Scattering Applications. Geneva: Springer Nature.]). Because of the scarcity of 3He (Zeitelhack, 2012[Zeitelhack, K. (2012). Neutron News, 23(4), 10-13.]) and the number of detector and readout modules required by those systems, the neutron detectors can become a major cost factor for scattering beamlines for new projects as well as upgrades of existing ones. Thus an active search for suitable alternatives is ongoing (Kirstein et al., 2014[Kirstein, O., Hall-Wilton, R., Stefanescu, I., Etxegarai, M., Anastasopoulos, M., Fissum, K., Gulyachkina, A., Höglund, C., Imam, M., Kanaki, K. & Khaplanov, A. (2014). arXiv:1411.6194.]; Pietropaolo et al., 2020[Pietropaolo, A., Angelone, M., Bedogni, R., Colonna, N., Hurd, A. J., Khaplanov, A., Murtas, F., Pillon, M., Piscitelli, F., Schooneveld, E. M. & Zeitelhack, K. (2020). Phys. Rep. 875, 1-65.]), e.g. for planned diffraction beamlines such as PIONEER (Liu et al., 2022[Liu, Y., Cao, H., Rosenkranz, S., Frost, M., Huegle, T., Lin, J. Y., Torres, P., Stoica, A. & Chakoumakos, B. C. (2022). Rev. Sci. Instrum. 93, 073901.]) at the Spallation Neutron Source or BEER (Fenske et al., 2016[Fenske, J., Rouijaa, M., Šaroun, J., Kampmann, R., Staron, P., Nowak, G., Pilch, J., Beran, P., Šittner, P., Strunz, P., Brokmeier, H. G., Ryukhtin, V., Kadeřávek, L., Strobl, M., Müller, M., Lukáš, P. & Schreyer, A. (2016). J. Phys. Conf. Ser. 746, 012009.]) at the European Spallation Source.

Recently, Losko et al. (2021[Losko, A. S., Han, Y., Schillinger, B., Tartaglione, A., Morgano, M., Strobl, M., Long, J., Tremsin, A. S. & Schulz, M. (2021). Sci. Rep. 11, 21360.]) demonstrated the use of a Timepix3-based event-mode imaging camera to detect individual photons emitted from a scintillator screen, and its advantages in cold and thermal neutron imaging. In this previous work, it was established that superior spatial and temporal resolution can be achieved through careful analysis of spatial and temporal photon distributions emitted from a scintillator screen (event clustering). Moreover, background from scintillator afterglow and incident gammas can be subtracted by particle discrimination based on simple scintillation photon multiplicity thresholds and temporal photon distributions. One promising advantage of this system is its use of optics for scintillation photon detection, which allows for adjustable fields of view that can be tailored for any given experimental setup, such as neutron diffraction detectors which require large-area fields of view. Additionally, the entire detector is read out by one chip and one software, avoiding the necessity for thousands of photomultipliers and acquisition channels. While this approach was originally developed and demonstrated for neutron imaging, here we present a first demonstration of the event-mode technique for neutron scattering data acquisition, in particular neutron time-of-flight diffraction, including the adaptation of the MAUD (Materials Analysis Using Diffraction) Rietveld code (Lutterotti et al., 1999[Lutterotti, L., Wenk, H. & Matthies, S. (1999). Proceedings of the Twelfth International Conference on Textures of Materials (ICOTOM-12), Vol. 2, pp. 1599-1604. NRC Research Press.]) to refine the required calibration parameters during the Rietveld analysis.

To compare this novel diffraction detection method utilizing imaging detectors with a conventional neutron time-of-flight diffraction detector system, a silicon powder diffraction experiment was conducted on the High Pressure–Preferred Orientation (HIPPO) beamline (Wenk et al., 2003[Wenk, H. R., Lutterotti, L. & Vogel, S. (2003). Nucl. Instrum. Methods Phys. Res. A, 515, 575-588.]; Vogel et al., 2004[Vogel, S., Williams, D., Zhao, Y., Bennett, K., Von Dreele, B. & Wenk, H. R. (2004). APS March Meeting Abstracts, pp. R1-225.]; Takajo & Vogel, 2018[Takajo, S. & Vogel, S. C. (2018). J. Appl. Cryst. 51, 895-900.]) at the Los Alamos Neutron Science Center (Lisowski & Schoenberg, 2006[Lisowski, P. W. & Schoenberg, K. F. (2006). Nucl. Instrum. Methods Phys. Res. A, 562, 910-914.]) Spallation Neutron Source, and simulations with the Monte Carlo code McStas (Lefmann & Nielsen, 1999[Lefmann, K. & Nielsen, K. (1999). Neutron News, 10(3), 20-23.]) were carried out.

2. Methods

2.1. Experimental setup

As a first test of imaging-based neutron diffraction analysis, a silicon powder sample in a vanadium can with 9 mm diameter and 50 mm height was measured at the HIPPO beamline. Two event-mode camera systems were deployed for this test (see Fig. 1[link]): one in the forward transmission direction at about 10 mm distance to the sample for radiography, and one more to the left of the sample (when viewed along the beam direction) at approximately 90° angle to the incident beam and 11.3 cm distance to the sample. Fig. 2[link] shows a schematic of the event-mode imaging system employed in this diffraction experiment. The neutron-sensitive area of the detector was a 12 × 12 cm scintillator screen with a 400 µm-thick mixture of silver-doped zinc sulfide (ZnS:Ag) and lithium-6-enriched lithium fluoride [RC Tritec AG, 6LiF/ZnS:Ag (ratio 1/2), 450 nm peak emission (Neuwirth et al., 2020[Neuwirth, T., Walfort, B., Sebold, S. & Schulz, M. (2020). J. Imaging, 6, 134.]; RC Tritec, 2004[RC Tritec (2004). 6LiF-Based Scintillators for Neutron Imaging, https://www.rctritec.com/en/isotopes/li-f-based-scintillators-2.]). The scintillator screen was shielded from air scatter neutrons by a 10B-containing plastic shielding (orange in Fig. 1[link]). The light from this scintillator screen was projected on a dual multi-channel plate (in chevron configuration) image intensifier [Photonis, Cricket Hi-QE Green, 5 × 105 gain, P47 phosphor (Exosens Products, 2024[Exosens Products (2024). Cricket Advanced Image Intensifier Adapter, https://exosens.com/products/crickettm.])] via a mirror and an optical lens [Navitar, DO-2595, 25 mm focal length, F/0.95 (Navitar Lenses, 2024[Navitar Lenses (2024). (Navitar) High Speed Lenses, https://www.navitar.com/products/imaging-solutions/high-speed.])]. The image intensifier was viewed by a Timepix3-based camera [Amsterdam Scientific Instruments, TPX3Cam (Zhao et al., 2017[Zhao, A., van Beuzekom, M., Bouwens, B., Byelov, D., Chakaberia, I., Cheng, C., Maddox, E., Nomerotski, A., Svihra, P., Visser, J., Vrba, V. & Weinacht, T. (2017). Rev. Sci. Instrum. 88, 113104.]; TPX3Cam Product, 2024[TPX3Cam Product (2024). TPX3Cam - Single Photon Counting Camera, https://quarkphotonics.au/product/tpx3cam-single-photon-counting-camera/.])]. All these components were enclosed in a light-tight box.

[Figure 1]
Figure 1
(a) Schematic of the experimental setup inside the HIPPO sample chamber with three event-mode cameras (the transmission detector for radiography is much smaller than the large-field-of-view scattering detectors). The camera to the right was removed before the experiment described here because it was broken. (b) Photograph of the actual setup from the top of the HIPPO sample chamber (the camera to the right was removed before the experiment described here).
[Figure 2]
Figure 2
Schematic setup of the camera system used for the diffraction experiment. The active field of view is oriented such that it detects scattered neutrons.

The scintillator screen of this setup could have a maximum size of 11.5 × 11.5 cm. By changing the optical lens and its distance to the scintillator, scintillator screens larger than 11.5 × 11.5 cm can be utilized, but this requires a larger box. Furthermore, the scintillator material and thickness can also be changed readily as needed for any given experiment. For the diffraction setup described here, a field of view of 15.5 × 15.5 cm was used, thus exceeding the physical size of the scintillator to contain the field of view of the image intensifier. Data from the TPX3Cam were processed following the event centroiding procedures described by Losko et al. (2021[Losko, A. S., Han, Y., Schillinger, B., Tartaglione, A., Morgano, M., Strobl, M., Long, J., Tremsin, A. S. & Schulz, M. (2021). Sci. Rep. 11, 21360.]). The TPX3Cam has 256 × 256 pixels, achieving an effective resolution of ∼50 µm for the 11.5 × 11.5 cm screen, due to the event centroiding [see below and Losko et al. (2021[Losko, A. S., Han, Y., Schillinger, B., Tartaglione, A., Morgano, M., Strobl, M., Long, J., Tremsin, A. S. & Schulz, M. (2021). Sci. Rep. 11, 21360.])].

The maximum event rate of the detector system depends on parameters such as the aperture and the focus of the lens, the image intensifier gain, and the type of scintillator. If these parameters are tuned properly, maximum neutron detection rates of 5 MHz are feasible with ZnS:Ag scintillators. For comparison, the incident flux on the sample for HIPPO is ∼1 × 107 neutrons s−1 cm−2 at a typical proton current of the linear accelerator of 100 µA (Ino et al., 2004[Ino, T., Ooi, M., Kiyanagi, Y., Kasugai, Y., Maekawa, F., Takada, H., Muhrer, G., J. Pitcher, E. & J. Russell, G. (2004). Nucl. Instrum. Methods Phys. Res. A, 525, 496-510.]), providing an upper bound for the total neutron detection rate of the entire detector system in the case that all neutrons are scattered and detected. Assuming isotropically scattered intensity and a scattering detector system consisting of six cameras with the scintillator screens forming a cube around the sample, each camera would have to detect one-sixth of 1 × 107 neutrons s−1 = 1.7 × 106 neutrons s−1. The maximum detection rate of 5 MHz of the current system is therefore suitable for time-of-flight measurements at HIPPO. The data used in this paper were acquired within 1 h at a nominal proton current of 90 µA, i.e. adjusted for fluctuations in the proton current.

2.2. Data processing with GSAS

The event-mode data were processed following procedures described by Losko et al. (2021[Losko, A. S., Han, Y., Schillinger, B., Tartaglione, A., Morgano, M., Strobl, M., Long, J., Tremsin, A. S. & Schulz, M. (2021). Sci. Rep. 11, 21360.]). To obtain diffraction spectra, event data from stripes consisting of pixels of five columns (perpendicular to the neutron beam) were integrated and output in GSAS data format (Larson & Von Dreele, 2004[Larson, A. C. & Von Dreele, R. B. (2004). General Structure Analysis System (GSAS). Report LAUR 86-748. Los Alamos National Laboratory, New Mexico, USA. ]). Using a script written in gsaslanguage (Vogel, 2011[Vogel, S. C. (2011). J. Appl. Cryst. 44, 873-877.]), a Rietveld (1969[Rietveld, H. M. (1969). J. Appl. Cryst. 2, 65-71.]) time-of-flight to d-spacing conversion factor (DIFC in GSAS), background parameters and the peak width parameter σ1 of the GSAS time-of-flight profile function #1 (Von Dreele et al., 1982[Von Dreele, R. B., Jorgensen, J. D. & Windsor, C. G. (1982). J. Appl. Cryst. 15, 581-589.]) were refined with the silicon lattice parameter held constant at 5.43086 Å. The isotropic atomic displacement parameter Uiso for the silicon atom was fixed at 0.006719 Å2, corresponding to a Biso value for room tem­pera­ture reported by Peng et al. (1996[Peng, L.-M., Ren, G., Dudarev, S. L. & Whelan, M. J. (1996). Acta Cryst. A52, 456-470.]), since the limited d-spacing range did not allow for a reliable refinement. The incident intensity function as calibrated for the regular 90° HIPPO detector bank was used for the normalization of the diffraction data for the TPX3Cam. The resulting σ1 parameter allowed us to find the column with the minimum σ1 which is assumed to be at 2ϑ = 90° and these are the data reported here.

2.3. Data processing with MAUD

Rietveld refinement of the diffraction data with MAUD followed the same procedure as described above for the GSAS analysis. However, the Rietveld analysis with MAUD also comprises the data reduction including calibration of instrument parameters. Reduction of 2D time-of-flight data into histograms analyzable by Rietveld codes can be accomplished with codes such as MANTID (Arnold et al., 2014[Arnold, O., Bilheux, J. C., Borreguero, J. M., Buts, A., Campbell, S. I., Chapon, L., Doucet, M., Draper, N., Ferraz Leal, R., Gigg, M. A., Lynch, V. E., Markvardsen, A., Mikkelson, D. J., Mikkelson, R. L., Miller, R., Palmen, K., Parker, P., Passos, G., Perring, T. G., Peterson, P. F., Ren, S., Reuter, M. A., Savici, A. T., Taylor, J. W., Taylor, R. J., Tolchenov, R., Zhou, W. & Zikovsky, J. (2014). Nucl. Instrum. Methods Phys. Res. A, 764, 156-166.]). However, since for example the reproducibility of the sample alignment cannot be guaranteed, being able to refine calibration parameters such as distance between sample and detector simultaneously with other parameters of the Rietveld refinement has proven crucial in neutron (Wenk et al., 2010[Wenk, H. R., Lutterotti, L. & Vogel, S. C. (2010). Powder Diffr. 25, 283-296.]) and synchrotron data analysis (Lutterotti et al. 2007[Lutterotti, L., Bortolotti, M., Ischia, G., Lonardelli, I. & Wenk, H. R. (2007). Z. Kristallogr. Suppl. 2007, 125-130.]; Wenk et al., 2014[Wenk, H. R., Lutterotti, L., Kaercher, P., Kanitpanyacharoen, W., Miyagi, L. & Vasin, R. (2014). Powder Diffr. 29, 220-232.]). To implement such an approach for 2D time-of-flight data, MAUD was extended to process event-mode data after reduction to location on the scintillator screen and time of flight. Similarly to constant-wavelength 2D data, such as in a synchrotron diffraction experiment, few parameters are required to describe the detector position and orientation. Since neutron data are much more sparse compared with X-ray data, with the additional complication that time-of-flight bins each require sufficient counts for meaningful analysis, the concept of superpixels is introduced in MAUD. Superpixels rebin the entire field of view into small segments of e.g. 8 × 32 or 1 × 8 pixels, with the short distance being along the axis along the neutron beam (assuming the detector is aligned with the scintillator plane approximately parallel to the incident neutron beam, as is the case here) or the greatest change in 2ϑ, while the superpixel's long axis is along the azimuthal direction, i.e. approximately circumferential to the Debye–Scherrer cones. This resembles the arrangement of 3He detector tubes in neutron time-of-flight diffractometers. After defining the superpixel dimensions and starting values for the detector geometry (see below), all events in an event-mode data file are processed and binned in time-of-flight histograms for the superpixels. Using the spatial arrangement of the superpixels within the field of view of the detector and the known diffraction pattern of a calibration substance, in our case silicon powder, a few parameters describing the detector geometry such as sample-to-detector distance, center pixel of the detector, tilt and rotation of the detector plane are refined against all histograms of the superpixels. The superpixels can then be integrated into histograms, similarly to the data reduction for 3He tubes for a detector panel or ring, for further refinement.

2.4. McStas simulation

Simulations with the Monte Carlo code McStas (version 3.3) were carried out with a sample-to-detector distance L2 = 11.3 cm as in the experimental setup described above. Furthermore, simulations were run at L2 = 83 cm, the nominal distance for the 90° 3He detector panels (Wenk et al., 2003[Wenk, H. R., Lutterotti, L. & Vogel, S. (2003). Nucl. Instrum. Methods Phys. Res. A, 515, 575-588.]), to validate the reliability of the simulations. McStas' Source_simple component was used to generate 8 × 1010 neutrons with wavelengths between 0.4 and 3.5 Å emitted from a source area of 2 × 2 cm, which are collimated to a 1.2 × 1.2 cm window at 8.5 m distance, approximately the location of the HIPPO collimation system at the entrance to the sample chamber. The source-to-sample distance of 9 m matches HIPPO's configuration. After traveling L1 = 9 m through a vacuum, the neutrons arrive at an air-filled cylinder with r = L2 around a cylindrical silicon powder sample with the same dimensions as in the experiment (9 mm diameter, 50 mm tall). The sample was simulated using the Union components Union_master, Union_cylinder and Union_make_material (Bertelsen, 2017[Bertelsen, M. (2017). PhD thesis, The Niels Bohr Institute, Faculty of Science, University of Copenhagen, Denmark.], 2022[Bertelsen, M. (2022). The Union Components, https://mads-bertelsen.github.io/tutorial/Union_tutorial_1_processes_and_materials.html.]) and NCrystal (Cai & Kittelmann, 2020[Cai, X. X. & Kittelmann, T. (2020). Comput. Phys. Commun. 246, 106851.]) to create the silicon powder and the surrounding air.

Fig. 3[link] shows an example of the simulated intensity versus scattering angle at 3.72 ms time-of-flight (TOF), corresponding to an arbitrary d spacing of approximately 1 Å, for an L2 of 83 cm, measured with the Cyl_monitor_TOF component. To ensure the detector resolution is not limiting the simulated d-spacing resolution, Cyl_monitor_TOF was set to nr = 800 radial bins, a height of 4 mm and nt = 1000 temporal bins ranging from tmin = 3.72 ms to tmax = 8.69 ms in the case of L2 = 83 cm and tmin = 1.00 ms to tmax = 8.06 ms in the case of L2 = 11.3 cm. To determine the resolution achievable at this distance, all diffraction peaks were fitted with a Gaussian as depicted in Fig. 3[link]. Assuming the uncertainty in scattering angle 2ϑ as [2{{\Delta}}\vartheta] = FWHM = [\left[ {8\ln \left(2 \right)} \right]^{1/2} {{\sigma}}] of the Gaussian, the d-spacing resolution is calculated as

[{{{{\Delta}}d} \over d} = \left\{ {{{\left({{{{{\Delta}}L} \over L}} \right)}^2} + {{\left({{{{{\Delta}}t} \over t}} \right)}^2} + {{\left[{{\rm cotan}\left(\vartheta \right)\Delta \vartheta } \right]}^2}} \right\}^{1/2}. \eqno (1)]

ΔL is the uncertainty of the flight path length resulting from the neutron moderation process, scattering from different points in the sample and beam divergence, assumed to be 2.5 cm. The total flight path of the neutrons is L = 900 cm + 83 cm = 983 cm or L = 900 cm + 11.3 cm = 911.3 cm. Δt is associated with different moderation times creating uncertainty in the neutron's TOF. Russell et al. (1985[Russell, G. J., Bowman, C. D., Whitaker, E. R., Robinson, H. & Meier, M. M. (1985). ICANS-VIII: Proceedings of the Eighth Meeting of the International Collaboration on Advanced Neutron Sources, pp. 272-293. Report RAL-85-110, Vol. 2. Rutherford Appleton Laboratory, Didcot, Oxford, UK. ]) report a neutron pulse width of ∼20 µs for neutron energies between 1 and 50 meV, corresponding to neutron wavelengths from 1.3 to 9 Å and for neutron detectors at 90° a d-spacing range from 0.9 to 6 Å. This matches our d-spacing range of interest and we therefore assumed Δt = 20 µs for our simulation.

[Figure 3]
Figure 3
Intensity in arbitrary units versus 2ϑ at an arbitrary TOF of 3.72 ms and L2 = 83 cm simulated with McStas is shown in blue. One of the Gaussian fits used to derive the resolution as a function of 2ϑ is depicted in orange.

3. Results and discussion

Fig. 4[link](a) shows the background-corrected intensity measured with only five pixel columns of the imaging detector setup (in red) versus d spacing at [2\vartheta = {90^ \circ }] and L2 = 11.3 cm, including the GSAS Rietveld fit in green. As for the 3He tubes too, the background was determined by fitting it with GSAS. Comparing the results with [2\vartheta = {90^ \circ }] data measured simultaneously with a single 3He detector panel consisting of 24 3He tubes at L2 = 83 cm, depicted in Fig. 4[link](b), and the McStas simulation carried out for L2 = 11.3 cm, shown in Fig. 4[link](c), indicates good agreement of the positions of the 422, 331, 400, 311 and 220 silicon diffraction peaks at 1.108, 1.246, 1.358, 1.637 and 1.920 Å, respectively. Because, for a medium-resolution diffractometer such as HIPPO, the constant and quadratic terms σ0 and σ2 for the peak broadening are zero in practice, the expression for the d-spacing-dependent width for the TOF profile function #1 in GSAS reduces to σ = σ1d (Larson & Von Dreele, 2004[Larson, A. C. & Von Dreele, R. B. (2004). General Structure Analysis System (GSAS). Report LAUR 86-748. Los Alamos National Laboratory, New Mexico, USA. ]). Using the relationship t = DIFC × d between TOF t and d spacing d yields equation (2[link]) for the Δd/d resolution:

[{{{{\Delta}}d} \over d} = {{{{\Delta}}t} \over {{\rm DIFC}\, \times d}} = \left[ {8\ln \left(2 \right)} \right]^{1/2} \,{{{{{\sigma}}_1}} \over {\rm DIFC}}. \eqno (2)]

[Figure 4]
Figure 4
Background-subtracted neutron counts per µs versus d spacing at [2\vartheta = {90^ \circ }] scattering angle from a ∅ = 9 mm Si powder sample: (a) measured in 1 h with only five (out of 256) pixel columns of the Timepix3 imaging detector, (b) measured in 45 min with 24 3He tubes in HIPPO's bank 17, and (c) simulated with the Monte Carlo code McStas. The data are depicted in red, with the GSAS fit in green, and the difference between fit and data is shown in purple below. Tick marks indicate calculated peak positions for silicon; the additional peaks observed in (b) and Fig. 6[link] result from the incident beam diffracting off an aluminium component in the imaging detector, which is to be eliminated in future designs.

Inserting σ1 and DIFC refined with GSAS, the imaging detector's d-spacing resolution was determined to be Δd/d = 2.1%, which is in good agreement with the GSAS analysis of the simulation results (at 90° for L2 = 11.3 cm) yielding 2.2% resolution. The resolution obtained with the imaging detector system is far from what the detector's spatial resolution of 50 µm and its temporal resolution of 1.6 ns would allow, which is consistent with the simulation result that the resolution at such distances is mainly determined by sample broadening due to the proximity of the detector to the sample. This means that even larger scintillators can easily be viewed with one camera since the spatial resolution of the camera is not the limiting factor. The significant deviations between simulation data and the Rietveld model for the McStas simulation are due to our rather simple model intended to study the effect of instrument and sample geometry (distances, angles etc.) on the broadening, not to predict detailed line profiles such as asymmetric broadening caused by the pulsed source etc.

To compare the number of diffracted neutrons detected by each detector system, we calculate the solid angles of each. The diffraction data from the TPX3Cam [Fig. 4[link](a)] were obtained with an active area of 3.48 cm2 (5 × 190 out of 256 × 256 pixels viewing 15.5 × 15.5 cm at a distance of 11.3 cm) covering 0.027 sr, whereas the 24 3He (1/2 inch diameter, 12 inches tall, ∼83 cm from sample) tubes of bank 17 [Fig. 4[link](b)] cover 929.03 cm2 or 0.135 sr. To minimize the influence of the higher background detected by the Timepix3 setup, caused by its proximity to the sample, the incident neutron beam and the second imaging detector, only the events in the 1.920 Å diffraction peak were integrated for comparison (after background subtraction). In 60 min of beam at a nominal proton current of 90 µA the Timepix3 detector recorded 1.536 × 103 events, while the 3He tubes recorded 1.029 × 104 events in 45 min at the same nominal current. After normalization by solid angle and count time, the imaging detector collected 3.16 events sr−1 s−1 and the 3He tubes detected 5.65 events sr−1 s−1. Hence, the imaging detector recorded 44.1% fewer events per steradian and second than the 3He detector system of the HIPPO instrument. In part, this can be explained by the event-mode readout software for the imaging detector, which caused ∼20% dead time in the version used for this experiment. Ongoing code optimizations will mitigate this in future versions. Furthermore, the scintillator used in this experiment was a standard radiography scintillator not optimized for this diffraction experiment. Gamma rejection using the Timepix3 system is achieved in the software during the photon event processing utilizing the different event shapes of neutron and gamma events (Sykora et al., 2012[Sykora, G. J., Schooneveld, E. M., Rhodes, N. J. & Van Eijck, L. (2012). 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC), pp. 1567-1571. IEEE.]; Johnson & Losko, 2023[Johnson, M. S. & Losko, A. S. (2023). IEEE Trans. Nucl. Sci. 70, 1413-1415.]). The parameters for the gamma rejection depend on scintillator properties and detector configuration, and their optimization for a given experiment is an ongoing development. In the present configuration gamma events are negligible due to the low interaction probability of gammas with the silicon sample. Our observation from preliminary data analysis of TPX3Cam transmission data on HIPPO and the adjacent flight path 5 (FP5), viewing the direct beam including its gamma component, is that compared with previously used detector technology (Tremsin et al., 2017[Tremsin, A. S., Losko, A. S., Vogel, S. C., Byler, D. D., McClellan, K. J., Bourke, M. A. M. & Vallerga, J. V. (2017). AIP Adv. 7, 015315.]) the background in both the thermal (<0.4 eV) and epithermal (>0.4 eV) energy ranges for Bragg-edge and neutron absorption resonance analysis, respectively, is extremely low with the Timepix3-based detector setup presented here. Since the beamline characteristics, especially for the thermal/epithermal neutron imaging beamline FP5, are identical to previous work, we conclude that the gamma background rejection in the event-mode processing is very promising. We will report these results in detail in the near future. Therefore, a slightly more gamma-sensitive 6Li glass scintillator (∼1 mm thickness) could be used to achieve 90.4% neutron absorption at 2.72 Å compared with the 45.1% the 0.4 mm ZnS:Ag/6Li used here provides. While a quantitative comparison of the efficiencies will be the subject of future work, including a comparison of different scintillator screens, the overall efficiencies observed for the two detector technologies tested here are comparable. This is especially true when considering that an event-mode system optimized for radiography was used for this demonstration of diffraction data collection, without any optimizations of the setup for diffraction whatsoever.

To verify the accuracy of resolutions determined with McStas as a function of 2ϑ, results are compared with nominal data from Wenk et al. (2003[Wenk, H. R., Lutterotti, L. & Vogel, S. (2003). Nucl. Instrum. Methods Phys. Res. A, 515, 575-588.]) in Fig. 5[link]. Simulation and nominal data are in good agreement, validating simulated resolutions. Furthermore, a resolution estimation made using the analytical function

[{{\Delta}}\vartheta = \arcsin \left({{r \over {{L_2}}}} \right)\eqno (3)]

to approximate Δϑ using only L2 and the sample radius r is depicted in orange. The resolution calculated by inserting equation (3)[link] into equation (1)[link] and using the same parameters for L, ΔL, t and Δt as described in Section 2.4[link], agrees well with nominal and simulated data, illustrating that for approximate resolution assessments time-consuming simulations can be omitted.

[Figure 5]
Figure 5
Resolution of a diffraction detector at 83 cm distance to the sample simulated with McStas (blue) and nominal data from Wenk et al. (2003[Wenk, H. R., Lutterotti, L. & Vogel, S. (2003). Nucl. Instrum. Methods Phys. Res. A, 515, 575-588.]) (in black). The orange curve depicts the resolution estimated calculating Δϑ with equation (3)[link]. The agreement between simulation and nominal data points validates our McStas model, thus allowing us to compare the predicted and observed resolutions for the Timepix3 setup.

Fig. 6[link] shows the results of the data reduction with MAUD for 8 × 32 [(a) and (c)] and 1 × 8 [(b) and (d)] superpixel sizes. As for the GSAS analysis described above, the silicon lattice parameter and atomic displacement parameter were kept constant during the refinement. Besides three Chebyshev polynomial parameters for the background and the aforementioned detector calibration parameters, only a velocity absorption correction to accommodate the deviations from the estimated incident intensity spectrum and the peak broadening parameters were varied. The raw data (lower panel) in Figs. 6[link](a) and 6[link](c) show the presence of spurious diffraction peaks from the aluminium holders of the imaging detector outside the calibrated sample position as diagonal stripes. The much smaller 1 × 8 superpixel size leads to 56 times more superpixels at the cost of reduced counts in each superpixel. However, the resulting peak width in the integrated spectrum in Fig. 6[link](d) is virtually identical to that obtained with the much coarser 8 × 32 superpixels in Fig. 6[link](b). This indicates that the geometric resolution of the setup (moderator and sample broadening, short L2) and not the resolution of the imaging detector is the limiting factor for the Δd/d resolution, which is in agreement with the McStas findings described above. The main deviations between fit and data in Figs. 6[link](b) and 6[link](d) originate from the contributions of the aluminium to the background, which are almost impossible to model and will be mitigated in future versions of this camera by minimizing material contributing to the diffraction signal.

[Figure 6]
Figure 6
(a) Contour plot produced by MAUD of diffracted intensity as a function of d spacing for 48 superpixels of 8 × 32 size, each shown as horizontal stripes, as raw data (bottom panel) and after the fit (top panel). (b) Integrated histogram of all 48 superpixels of 8 × 32 size after the Rietveld fit (red curve) to silicon powder data (black dots) of all individual histograms, with the difference curve below and silicon reflection positions marked by tick marks. (c) Contour plot produced by MAUD for 2688 superpixels of 1 × 8 size. (d) Integrated histogram of all 2688 superpixels of 1 × 8 size. See text for more information.

4. Conclusion

We demonstrated the first proof-of-concept for an event-mode imaging based scattering detector for neutron scattering data acquisition. This demonstration may pave the way to replacing conventional neutron detectors such as scintillator/photomultiplier or 3He detector tubes requiring custom-made electronics and software. The ability to operate imaging, scattering (diffraction, total scattering/pair distribution function analysis, small-angle scattering, reflectometry) and neutron resonance spectroscopy beamlines all with the same data acquisition software and the same simple hardware can greatly reduce investment and operational cost for neutron facilities. The components of a detector system as demonstrated here – TPX3Cam, image intensifier, optical lens, scintillator screen, light-tight box – are all available commercially off the shelf or can be readily fabricated, thus keeping the cost low, especially compared with hardware and software for custom-made readout electronics. While this work demonstrates the feasibility of applying a Timepix3-based system for diffraction detection at least for low- and medium-resolution applications, the applicability to other neutron scattering techniques such as high-resolution single-crystal diffraction, total scattering, small-angle scattering etc. with different demands on temperature stability, count rates etc. requires further evaluation. Demonstration of the technique with a larger scintillator to cover a larger fraction of 4π is planned for the near future by our team. The detection efficiency of the imaging-based system depends on multiple factors like readout software, image intensifier gain, optical lenses and the scintillator screen, and a detailed analysis of the absolute detection efficiency will be a topic of future publications. However, it is safe to say that detection efficiencies comparable to those of conventional scattering detector technology such as 3He tubes can be achieved. The inclusion of an optical lens as well as the ability to design tiled versions of this detector system with multiple TPX3Cams promises scalability to larger detector area coverage with the field of view per camera optimized to handle expected event rates. While scintillator areas up to 625 cm2 have been tested successfully in other experiments, we plan to demonstrate scintillator areas exceeding 625 cm2 per camera in the near future.

The widely used Rietveld code MAUD was adapted to calibrate and process the 2D TOF data generated by the camera-based setup. Building on experience from data analysis for 2D X-ray experiments at synchrotron or X-ray free-electron lase facilities, MAUD allows refinement of parameters related to the data reduction, such as sample/detector distances, during the Rietveld analysis rather than separating data reduction and data analysis into two independent steps.

We found that the Δd/d resolution obtained with the camera-based setup is as expected for the short distance between sample and detector used in this demonstration: at a sample-to-detector distance of 11.3 cm the resolution from the GSAS analysis was 2.1%, in good agreement with simulations carried out with McStas yielding Δd/d = 2.2% at [2\vartheta = {90^ \circ }]. The validity of resolutions determined by McStas simulations was established by comparing them with the nominal resolution of HIPPO as published by Wenk et al. (2003[Wenk, H. R., Lutterotti, L. & Vogel, S. (2003). Nucl. Instrum. Methods Phys. Res. A, 515, 575-588.]). Simulations indicate that at suitable distances, reducing sample broadening, similar resolutions can be achieved with the novel imaging detector system and with the current HIPPO 3He-based detector system. Furthermore, more solid angle can be covered by the camera-based system compared with conventional detector panels, thereby detecting more neutrons and reducing measurement time. The camera-based setup will additionally be able to record energy-resolved radiography data of the sample in the transmitted beam direction, therefore enabling simultaneous neutron diffraction and radiography characterization including Bragg-edge (Vogel, 2000[Vogel, S. (2000). PhD thesis, Christian-Albrechts Universität Kiel, Germany.]; Woracek et al., 2018[Woracek, R., Santisteban, J., Fedrigo, A. & Strobl, M. (2018). Nucl. Instrum. Methods Phys. Res. A, 878, 141-158.]) and neutron absorption resonance imaging (Losko & Vogel, 2022[Losko, A. S. & Vogel, S. C. (2022). Sci. Rep. 12, 6648.]). Ongoing efforts to further develop this system include the identification of optimal scintillator screens (material and thickness).

Funding information

This work has benefitted from the use of the Los Alamos Neutron Science Center (LANSCE) at Los Alamos National Laboratory (LANL). LANL is operated by Triad National Security, LLC, for the National Nuclear Security Administration of the US Department of Energy under contract No. 89233218NCA000001. We acknowledge support for an extended visit by Dr Luca Lutterotti at Los Alamos in the summer of 2023 through a grant by LANL's Technology Evaluation and Demonstration (TED) program. We gratefully acknowledge support from LANL's Laboratory Directed Research and Development (LDRD) program through grant No. 20230592ER. Furthermore, we thank the Danish Agency for Science, Technology and Innovation for funding the instrument center (DanScatt, grant No. 7129-00006B). We gratefully acknowledge support from the BMBF in the framework of the research project 05K22WO5 and the ZARA-LAN project 050306/50001339.

References

First citationArnold, O., Bilheux, J. C., Borreguero, J. M., Buts, A., Campbell, S. I., Chapon, L., Doucet, M., Draper, N., Ferraz Leal, R., Gigg, M. A., Lynch, V. E., Markvardsen, A., Mikkelson, D. J., Mikkelson, R. L., Miller, R., Palmen, K., Parker, P., Passos, G., Perring, T. G., Peterson, P. F., Ren, S., Reuter, M. A., Savici, A. T., Taylor, J. W., Taylor, R. J., Tolchenov, R., Zhou, W. & Zikovsky, J. (2014). Nucl. Instrum. Methods Phys. Res. A, 764, 156–166.  Web of Science CrossRef CAS Google Scholar
First citationBertelsen, M. (2017). PhD thesis, The Niels Bohr Institute, Faculty of Science, University of Copenhagen, Denmark.  Google Scholar
First citationBertelsen, M. (2022). The Union Components, https://mads-bertelsen.github.io/tutorial/Union_tutorial_1_processes_and_materials.htmlGoogle Scholar
First citationCai, X. X. & Kittelmann, T. (2020). Comput. Phys. Commun. 246, 106851.  Web of Science CrossRef Google Scholar
First citationDiawara, Y. (2023). Neutron Detectors for Scattering Applications. Geneva: Springer Nature.  Google Scholar
First citationExosens Products (2024). Cricket Advanced Image Intensifier Adapter, https://exosens.com/products/crickettmGoogle Scholar
First citationFenske, J., Rouijaa, M., Šaroun, J., Kampmann, R., Staron, P., Nowak, G., Pilch, J., Beran, P., Šittner, P., Strunz, P., Brokmeier, H. G., Ryukhtin, V., Kadeřávek, L., Strobl, M., Müller, M., Lukáš, P. & Schreyer, A. (2016). J. Phys. Conf. Ser. 746, 012009.  CrossRef Google Scholar
First citationIno, T., Ooi, M., Kiyanagi, Y., Kasugai, Y., Maekawa, F., Takada, H., Muhrer, G., J. Pitcher, E. & J. Russell, G. (2004). Nucl. Instrum. Methods Phys. Res. A, 525, 496–510.  Google Scholar
First citationJohnson, M. S. & Losko, A. S. (2023). IEEE Trans. Nucl. Sci. 70, 1413–1415.  Web of Science CrossRef CAS Google Scholar
First citationKirstein, O., Hall-Wilton, R., Stefanescu, I., Etxegarai, M., Anastasopoulos, M., Fissum, K., Gulyachkina, A., Höglund, C., Imam, M., Kanaki, K. & Khaplanov, A. (2014). arXiv:1411.6194.  Google Scholar
First citationLarson, A. C. & Von Dreele, R. B. (2004). General Structure Analysis System (GSAS). Report LAUR 86-748. Los Alamos National Laboratory, New Mexico, USA.   Google Scholar
First citationLefmann, K. & Nielsen, K. (1999). Neutron News, 10(3), 20–23.  CrossRef Google Scholar
First citationLisowski, P. W. & Schoenberg, K. F. (2006). Nucl. Instrum. Methods Phys. Res. A, 562, 910–914.  Web of Science CrossRef CAS Google Scholar
First citationLiu, Y., Cao, H., Rosenkranz, S., Frost, M., Huegle, T., Lin, J. Y., Torres, P., Stoica, A. & Chakoumakos, B. C. (2022). Rev. Sci. Instrum. 93, 073901.  Web of Science CrossRef PubMed Google Scholar
First citationLosko, A. S., Han, Y., Schillinger, B., Tartaglione, A., Morgano, M., Strobl, M., Long, J., Tremsin, A. S. & Schulz, M. (2021). Sci. Rep. 11, 21360.  Web of Science CrossRef PubMed Google Scholar
First citationLosko, A. S. & Vogel, S. C. (2022). Sci. Rep. 12, 6648.  Web of Science CrossRef PubMed Google Scholar
First citationLutterotti, L., Bortolotti, M., Ischia, G., Lonardelli, I. & Wenk, H. R. (2007). Z. Kristallogr. Suppl. 2007, 125–130.  CrossRef Google Scholar
First citationLutterotti, L., Wenk, H. & Matthies, S. (1999). Proceedings of the Twelfth International Conference on Textures of Materials (ICOTOM-12), Vol. 2, pp. 1599–1604. NRC Research Press.  Google Scholar
First citationNavitar Lenses (2024). (Navitar) High Speed Lenses, https://www.navitar.com/products/imaging-solutions/high-speedGoogle Scholar
First citationNeuwirth, T., Walfort, B., Sebold, S. & Schulz, M. (2020). J. Imaging, 6, 134.  Web of Science CrossRef PubMed Google Scholar
First citationPeng, L.-M., Ren, G., Dudarev, S. L. & Whelan, M. J. (1996). Acta Cryst. A52, 456–470.  CrossRef CAS Web of Science IUCr Journals Google Scholar
First citationPietropaolo, A., Angelone, M., Bedogni, R., Colonna, N., Hurd, A. J., Khaplanov, A., Murtas, F., Pillon, M., Piscitelli, F., Schooneveld, E. M. & Zeitelhack, K. (2020). Phys. Rep. 875, 1–65.  Web of Science CrossRef CAS Google Scholar
First citationRC Tritec (2004). 6LiF-Based Scintillators for Neutron Imaging, https://www.rctritec.com/en/isotopes/li-f-based-scintillators-2Google Scholar
First citationRietveld, H. M. (1969). J. Appl. Cryst. 2, 65–71.  CrossRef CAS IUCr Journals Web of Science Google Scholar
First citationRussell, G. J., Bowman, C. D., Whitaker, E. R., Robinson, H. & Meier, M. M. (1985). ICANS-VIII: Proceedings of the Eighth Meeting of the International Collaboration on Advanced Neutron Sources, pp. 272–293. Report RAL-85-110, Vol. 2. Rutherford Appleton Laboratory, Didcot, Oxford, UK.   Google Scholar
First citationStefanescu, I., Christensen, M., Fenske, J., Hall-Wilton, R., Henry, P. F., Kirstein, O., Müller, M., Nowak, G., Pooley, D., Raspino, D., Rhodes, N., Šaroun, J., Schefer, J., Schooneveld, E., Sykora, J. & Schweika, W. (2017). J. Instrum. 12, P01019.  Google Scholar
First citationSykora, G. J., Schooneveld, E. M., Rhodes, N. J. & Van Eijck, L. (2012). 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC), pp. 1567–1571. IEEE.  Google Scholar
First citationTakajo, S. & Vogel, S. C. (2018). J. Appl. Cryst. 51, 895–900.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationTPX3Cam Product (2024). TPX3Cam – Single Photon Counting Camera, https://quarkphotonics.au/product/tpx3cam-single-photon-counting-camera/Google Scholar
First citationTremsin, A. S., Losko, A. S., Vogel, S. C., Byler, D. D., McClellan, K. J., Bourke, M. A. M. & Vallerga, J. V. (2017). AIP Adv. 7, 015315.  Google Scholar
First citationVogel, S. (2000). PhD thesis, Christian-Albrechts Universität Kiel, Germany.  Google Scholar
First citationVogel, S., Williams, D., Zhao, Y., Bennett, K., Von Dreele, B. & Wenk, H. R. (2004). APS March Meeting Abstracts, pp. R1–225.  Google Scholar
First citationVogel, S. C. (2011). J. Appl. Cryst. 44, 873–877.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationVon Dreele, R. B., Jorgensen, J. D. & Windsor, C. G. (1982). J. Appl. Cryst. 15, 581–589.  CrossRef CAS Web of Science IUCr Journals Google Scholar
First citationWenk, H. R., Lutterotti, L., Kaercher, P., Kanitpanyacharoen, W., Miyagi, L. & Vasin, R. (2014). Powder Diffr. 29, 220–232.  Web of Science CrossRef CAS Google Scholar
First citationWenk, H. R., Lutterotti, L. & Vogel, S. (2003). Nucl. Instrum. Methods Phys. Res. A, 515, 575–588.  Web of Science CrossRef CAS Google Scholar
First citationWenk, H. R., Lutterotti, L. & Vogel, S. C. (2010). Powder Diffr. 25, 283–296.  Web of Science CrossRef CAS Google Scholar
First citationWoracek, R., Santisteban, J., Fedrigo, A. & Strobl, M. (2018). Nucl. Instrum. Methods Phys. Res. A, 878, 141–158.  Web of Science CrossRef CAS Google Scholar
First citationZeitelhack, K. (2012). Neutron News, 23(4), 10–13.  CrossRef Google Scholar
First citationZhao, A., van Beuzekom, M., Bouwens, B., Byelov, D., Chakaberia, I., Cheng, C., Maddox, E., Nomerotski, A., Svihra, P., Visser, J., Vrba, V. & Weinacht, T. (2017). Rev. Sci. Instrum. 88, 113104.  Web of Science CrossRef PubMed Google Scholar

This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.

Journal logoJOURNAL OF
APPLIED
CRYSTALLOGRAPHY
ISSN: 1600-5767
Follow J. Appl. Cryst.
Sign up for e-alerts
Follow J. Appl. Cryst. on Twitter
Follow us on facebook
Sign up for RSS feeds