research papers\(\def\hfill{\hskip 5em}\def\hfil{\hskip 3em}\def\eqno#1{\hfil {#1}}\)

Journal logoJOURNAL OF
APPLIED
CRYSTALLOGRAPHY
ISSN: 1600-5767

Angle-dependent X-ray absorption correction in small- and wide-angle X-ray scattering: accounting for constrained scattering volume, beam profiles and capillary wall attenuation

crossmark logo

aFaculty of Chemistry and Chemical Technology, University of Ljubljana, Večna pot 113, Ljubljana 1000, Slovenia, and bMTA-SZTE Momentum Biocolloids Research Group, Department of Physical Chemistry and Materials Science, Interdisciplinary Research Center, University of Szeged, Szeged H-6720, Hungary
*Correspondence e-mail: [email protected]

Edited by J. Ilavsky, Argonne National Laboratory, USA (Received 19 October 2025; accepted 29 June 2026; online 6 August 2026)

A GPU-accelerated numerical method has been developed to calculate angle-dependent transmission factors A(2θ) for small- and wide-angle X-ray scattering (SWAXS) data collected with a Kratky-type camera using a line-collimated beam and a horizontal cylindrical capillary with a vertically oriented detector. The procedure extends path-length-based absorption correction to the Kratky-type geometry by averaging Beer–Lambert attenuation over the constrained illuminated scattering volume and by including experimentally measured primary beam profiles and capillary-wall absorption. Sensitivity analysis shows that the vertical illumination geometry dominates the calculated angle-dependent transmission factor, whereas the horizontal beam-width effect is negligible. Using measured beam profiles makes the correction instrument specific and avoids introducing an effective uniform beam height as an approximation. For strongly absorbing nona­fluoro-tert-butanol, the conventional scalar primary beam transmission correction fails after subtraction of empty-capillary scattering, while the numerical correction enables physically meaningful absolute-scale data reduction. Comparisons between independently calculated theoretical SWAXS intensities and the two reduced experimental datasets obtained sequentially using two different detectors in the same experiment on the same Kratky-type SWAXS camera provide a direct experimental cross check for the data-reduction workflow based on calculated transmission factors. They indicate that the observed limitation is associated with transmission measurement of a spectrally modified primary beam, rather than with the Kratky-type geometry itself. Importantly, the results show that the limitations of conventional scalar transmission corrections are not confined to extreme cases but may also affect routine water-based absolute-scale SWAXS calibration. A simple Beer–Lambert-type numerical scalar correction can still provide a useful qualitative approximation, whereas reliable quantitative analysis requires the full angle-dependent A(2θ) correction.

1. Introduction

When performing X-ray scattering measurements on a series of samples, reliable absorption correction is essential for basic data reduction, because subsequent background subtraction and data interpretation critically depend on it. We focus our attention on small- and wide-angle X-ray scattering (SWAXS, including both SAXS and WAXS) measurements of liquid and soft-matter samples in cylindrical quartz capillaries performed using line-collimated Kratky-type cameras (Fritz & Bergmann, 2006View full citation), which remain established laboratory instruments in colloid and soft-matter research. The broader motivation for this study is summarized in Appendix 1 of the supporting information (Cerar et al., 2020View full citation; Tomšič, Jamnik et al., 2007View full citation; Lajovic et al., 2010View full citation; Cerar et al., 2021View full citation; Cerar et al., 2019View full citation; Tomšič et al., 2019View full citation; Tomšič et al., 2018View full citation; Cerar et al., 2017View full citation; Lajovic et al., 2012View full citation; Lajovic et al., 2009View full citation; Tomšič, Vlček et al., 2007View full citation). Note that the supporting information is a substantial document that forms an integral part of this paper, providing extensive details on all aspects of the discussion, and is intended to be read alongside the main text. As such, frequent references to it are made in the following.

In a conventional SAXS/SWAXS workflow, each raw curve Ij,raw(2θ) is first corrected using an experimental scalar transmission factor Tj,

Mathematical equation

where Ij,T(2θ) represents the absorption-corrected curve, and j = s, sol, w or ecap denotes the sample, solvent, water or empty-capillary measurement, respectively. In Kratky-type cameras with a semi-transparent beamstop, the factor Tj is determined from the primary beam transmission measured for the corresponding measurement, i.e. derived from the measured primary beam intensity at zero scattering angle, Ij,raw(0). The corresponding background scattering (solvent or empty capillary) is then subtracted, yielding the background-corrected SWAXS data for the sample Is,bc(2θ) and water Iw,bc(2θ),

Mathematical equation

The latter expression is important if the SWAXS data need to be put on an absolute scale, as the absolute scaling factor FT can be calculated using water as a secondary standard (Orthaber et al., 2000View full citation) as

Mathematical equation

Here, (dΣ/dΩ)w,th(0) is the known theoretical scattering cross section of water at zero scattering angle (1.632 × 10−2 cm−1 at 293 K) (Orthaber et al., 2000View full citation) and Iw,bc(0) is obtained by extrapolating the experimental Iw,bc(2θ) data to the zero scattering angle. Absolute scaling is then performed as

Mathematical equation

where (dΣ/dΩ)s(2θ) is the absolute-scale differential scattering cross section of the sample. For interested readers, additional comments are provided in Appendix 2 of the supporting information (Orthaber et al., 2000View full citation; Pauw, 2013View full citation; Bergmann et al., 2000View full citation).

This conventional procedure usually works well in SAXS/SWAXS on Kratky-type cameras when the attenuation of all measurements in a series is similar, for example for dilute samples and their corresponding solvent. However, it may become unreliable when the X-ray absorption differs substantially between the measured samples. Thus, the issue is not the conventional data-reduction workflow itself but rather the practical limitation of experimentally determining reliable scalar transmission factors Tj. We attribute this limitation primarily to differences in the degree of primary beam radiation hardening between samples with substantially different attenuation properties, an effect further enhanced by the absorption of the semi-transparent beamstop in Kratky-type instruments. SAXS/SWAXS position-sensitive detectors, including modern ones, generally provide no or only limited photon energy discrimination for the detected beam, so the measured primary beam intensities can become strongly biased in such cases (Baur et al., 2019View full citation; Hövelmann et al., 2019View full citation). We discuss primary beam radiation hardening in more detail in Appendix 3 (Bergmann et al., 2000View full citation; Oberta et al., 2012View full citation; Kirschbaum et al., 1997View full citation; Shimizu & Omote, 2008View full citation; Baur et al., 2019View full citation; Brooks & Di Chiro, 1976View full citation; Van de Casteele et al., 2002View full citation; Hövelmann et al., 2019View full citation; Wu & Li, 2024View full citation; Polikarpov et al., 2014View full citation; Chen et al., 2018View full citation). Appendix 4 presents a practical example of the failure of the conventional procedure for the strongly absorbing liquid nona­fluoro-tert-butanol (NFTB). NFTB is therefore used as a representative test case in this study.

A way of overcoming such experimental problems is to resort to calculated transmission factors from numerical absorption correction procedures, which can also include more sophisticated treatments, e.g. angle-dependent self-absorption correction. Numerous numerical procedures have been developed and tested over many decades to correct scattering data for X-ray absorption. These methods place particular emphasis on sample geometry and the X-ray path through the material in different experimental setups, as reviewed in Appendix 5 (Claassen, 1930View full citation; Bradley, 1935View full citation; Bond, 1959View full citation; Dwiggins, 1975View full citation; 1972View full citation; Ritter et al., 1951View full citation; Møller & Jensen, 1952View full citation; Schmitt & Ouladdiaf, 1998View full citation; Bowden & Ryan, 2010View full citation; Pauw, 2013View full citation; Pauw et al., 2017View full citation; Bendert et al., 2013View full citation; Van Hoesen et al., 2019View full citation; Jeffries et al., 2021View full citation; Sulyanov et al., 2012View full citation). Here we focus on the procedure of Bowden & Ryan (2010View full citation), who treated absorption in cylindrical and annular specimens and their containers or supports within a powder diffraction framework based on cross-sectional path-length integration. This makes their work the closest reference point for the cylindrical capillary geometry relevant to line-collimated Kratky-type SAXS and SWAXS instruments.

The novelty of our work is the extension of this type of path-length-based correction towards the actual geometry of line-collimated SWAXS instruments using routinely available instrument-specific information, namely the experimentally measured non-uniform primary beam profiles. In this way, the finite constrained illuminated scattering volume is retained, but its contribution is weighted by the measured primary beam intensity distribution rather than being treated as uniformly illuminated. In addition, we do not treat the capillary wall only as an absorbing medium along the photon path but also consider its scattering contribution separately in the empty-capillary measurement. These elements are then incorporated into the usual data reduction sequence, in which the experimental scalar transmission factor is replaced by calculated angle-dependent transmission factors before background subtraction and absolute scaling.

The objective of the present work is therefore to formulate, implement and evaluate a numerical angle-dependent absorption correction procedure directly applicable to line-collimated SWAXS measurements of liquid samples in thin-walled cylindrical capillaries. The procedure is designed to address the limitations of scalar transmission-based data reduction and absolute scaling that we encountered in our SWAXS studies of strongly absorbing samples, as discussed in detail in Appendix 4. We evaluate the procedure by testing the numerical convergence of the calculated transmission factors and by analysing their sensitivity to beam geometry, measured beam profiles and sample attenuation in Appendix 8. We further apply it to experimental SWAXS datasets from chemically diverse systems in Appendix 10 (Pi et al., 2009View full citation; Cerar et al., 2021View full citation). These applications also provide an important model-based consistency check by comparing corrected experimental SWAXS data with independently calculated theoretical SWAXS intensities for the same systems, obtained using the `complemented system approach' (Cerar et al., 2020View full citation; Tomšič, Jamnik et al., 2007View full citation; Lajovic et al., 2010View full citation). These theoretical intensities are calculated from molecular simulation results and are not fitted, scaled or otherwise adjusted to match the experimental data, thereby representing a fully independent reference dataset within the limitations of the chosen molecular force field. In addition, measurements performed on the same sample in the same capillary and on the same Kratky-type camera, without changing the sample/capillary arrangement but with a Dectris Mythen 1K detector, are used as a direct experimental cross check of the proposed absorption correction workflow, as discussed in Appendix 11.

2. Formulation of a realistic SWAXS absorption correction model

We focus on Kratky-type cameras and similar modern line-collimated SWAXS geometries (Glatter, 2018View full citation) in which a thin-walled horizontal quartz capillary is illuminated by a horizontally collimated primary beam passing through its central axis. In such instruments, the capillary is consistently aligned to ensure it is always centrally illuminated. The detector is positioned perpendicular to the capillary axis and oriented vertically, with the scattering signal recorded in the upward direction, as depicted schematically in Fig. 1[link]. For this geometry, we extend the existing path-length-based absorption correction framework (Bowden & Ryan, 2010View full citation) by incorporating weighting functions based on experimentally determined primary beam profiles, constraining the scattering volume vertically by the beam's `height' and horizontally by its `width' profile, both of them perpendicular to the beam propagation direction. Here, beam height and beam width denote the beam height in the vertical (zy) plane and the beam width in the horizontal (xy) plane, respectively. The introduced weighting functions provide a direct way to connect idealized path-length calculations with the actual illumination conditions in a line-collimated Kratky-type SWAXS instrument. In contrast to uniform illumination or small-beam approximations, the transmission factor is averaged over the finite illuminated region and weighted by the experimentally measured primary beam profiles. Additionally, we explicitly account for attenuation in the quartz capillary wall and use a full three-dimensional correction model. These features become particularly important for highly absorbing samples. To the best of our knowledge, this specific combination of constrained illuminated volume, capillary wall attenuation and experimentally measured beam profile weighting has not previously been implemented for absolute-scale line-collimated SWAXS data reduction in thin-walled capillaries.

[Figure 1]
Figure 1
Schematic representation of the considered experimental geometry.

The experimental geometry is defined in a Cartesian coordinate system with the primary beam propagating along the y axis, the horizontal direction along the x axis and the vertical direction along the z axis. A schematic representation of this geometry, including the sample, the capillary wall and the detector arrangement, is shown in Fig. 2[link](a).

[Figure 2]
Figure 2
Visualization of the sample and detector geometry considered in our derivation of the path lengths, (a) without considering the capillary wall absorption and scattering, (b) considering only the scattering and absorption in the capillary wall, and (c) considering scattering inside the sample and absorption in the sample and capillary wall.

The scattering-angle-dependent absorption correction is applied directly to the measured raw scattering intensity Iraw(2θ), analogously to equation (1)[link], using the relation

Mathematical equation

where 2θ is the scattering angle, Icorr(2θ) is the corrected scattering intensity, and A(2θ) is the angle-dependent transmission factor calculated for the relevant absorbing material and experimental geometry. The geometric meaning of this correction is clarified by the definition of the transmission factor A(2θ), introduced in the following. In a general 3D geometry, the transmission factor can be expressed as a volume integral over the illuminated part of the sample (and its container, if included among attenuation contributions), accounting for attenuation along the incident path to the scattering point and along the scattered path from that point to the detector:

Mathematical equation

where μ is the linear attenuation coefficient of the material, L(r, 2θ) is the X-ray path length through the material at a scattering angle 2θ and V is the illuminated volume. Equation (6)[link] is essentially the Beer–Lambert attenuation law averaged over all the corresponding incident and exit X-ray paths relevant for the detected scattering at each 2θ. Accordingly, A(2θ) represents the average transmission coefficient defined solely for photons contributing to the detected scattering intensity at a given 2θ. For SAXS data, it is often even more natural to express the angular dependence in terms of the length of the scattering vector q, defined as

Mathematical equation

where λ is the wavelength of the X-rays. To facilitate comparison with previous absorption correction studies, we compute and present the correction primarily as a function of the scattering angle 2θ, but also provide scattering-angle-dependent transmission correction factors A(q) where relevant (see Table S3 in Appendix 7).

Note that the wavelength of the X-rays does not enter this absorption correction formulation explicitly. Its influence on the absorption correction coefficients is fully accounted for through the wavelength dependence of the linear attenuation coefficient μ, which is treated as an input material parameter (Hubbell & Seltzer, 2004View full citation).

In our SWAXS experimental configuration, the incident beam exhibits an experimentally determined non-uniform intensity distribution, which is measured as two orthogonal profiles: (i) a `width' profile Pwid(x) along the horizontal axis x, and (ii) a `height' profile Phei(z) along the vertical axis z, as shown in Fig. 3[link]. These two profiles are incorporated into the absorption correction scheme through the corresponding dimensionless weighting function W(x, z). The latter is independent of the beam-propagation coordinate y and is defined for the primary beam path lengths through the capillary with the sample as

Mathematical equation

where the normalization factor is given by

Mathematical equation

ensuring that

Mathematical equation

[Figure 3]
Figure 3
The experimentally obtained (a) height Phei(z) and (b) width Pwid(x) profiles of the horizontal line-collimated primary X-ray beam in our SWAXS setup.

The transmission coefficient A(2θ), which accounts for attenuation in both the capillary wall and the sample, is then calculated as

Mathematical equation

where V is the constrained illuminated volume defined by the limits of integration, μcw and μs are the attenuation coefficients of the capillary wall and the sample, respectively, and Lcw(x, y, z, 2θ) and Ls(x, y, z, 2θ) denote the corresponding path lengths contributing to the scattering at an angle 2θ. In these path-length terms, the coordinate y denotes that the scattering event may occur at any position along the y axis, including within the capillary wall, as depicted schematically in Fig. 2[link](b).

The cylindrical capillary has an inner radius R and a wall thickness t. It is illuminated non-uniformly, with the illumination represented by W(x, z) over the constrained cross-section region z ∈ [−a, a] and x ∈ [−b, b], while the primary beam propagates along the y axis. The limits a and b define the constrained illuminated scattering volume cross section used in the numerical integration and were chosen from the measured extent of the experimental beam profiles, i.e. from the regions where their intensities remain significant. A scattering event occurs at a point C(x, y, z) within the constrained scattering volume – either within the capillary wall or within the sample [Fig. 2[link](c)]. Due to the non-uniform illumination of the scattering volume, the corresponding path lengths must be weighted by the actual local intensity of the primary beam to obtain the weighted average transmission factor in equation (11)[link]. To keep the geometric description consistent with the actual experimental setup used in this work, the detector is positioned vertically at a distance d from the central capillary axis and is curved such that the detector points lie on the cylindrical surface:

Mathematical equation

The path length through the sample, Ls(x, y, z, 2θ), is [Fig. 2[link](c)]

Mathematical equation

and the total path through the capillary wall, Lew(x, y, z, 2θ), is

Mathematical equation

Given the general coordinates of point Mathematical equation, the magnitude of the incident segment inside the sample is

Mathematical equation

For the scattered beam path length towards the detector, |CD|, we consider the ray from point C in the direction towards the detector point at the scattering angle of 2θ,

Mathematical equation

which intersects the inner cylinder y2 + z2 = R2 of the sample. This yields the quadratic equation for parameter p,

Mathematical equation

Of the two roots, the larger, p+, corresponds to the sample exit point, giving the magnitude

Mathematical equation

The magnitudes of |AC| and |CE| are computed analogously using the outer radius (R + t).

For scattering events occurring within the wall,

Mathematical equation

the beam segments through the inner sample cylinder must be treated conditionally: (i) if y > 0 and |z| < R, the incident beam intersects the inner surface, yielding

Mathematical equation

but (ii) if two positive roots p+,inner > p−,inner exist for the scattered ray intersecting the inner sample cylinder, then

Mathematical equation

In summary, we note that, in the present context, the transmission factor A(2θ) is not a general material property but rather a numerically determined geometry-specific quantity. It is used here as a calculated, angle-dependent and geometry-specific Beer–Lambert transmission factor for the detected scattering intensity, not as an experimentally measured direct-beam transmission factor. It effectively encapsulates information about the specific experimental configuration of the instrument – derived from the primary beam profiles – as well as information on the sample material through its linear attenuation coefficient. In our case, this corresponds to a Kratky-type line-collimated SWAXS setup equipped with a focusing Göbel mirror. In the calculations, only two absorption contributions are considered: (i) attenuation of the X-ray beam within the capillary up to the position of the scattering event, and (ii) attenuation of the scattered ray along its exit path through the remaining capillary material towards the detector. Consequently, the transmission factor A(2θ) accounts exclusively for these two contributions and does not include additional instrumental effects such as detector response, multiple scattering or spectral changes in the incident beam. In a practical sense, A(2θ) represents a weighted average over all individual path-length contributions arising from the different possible scattering positions within the illuminated scattering volume.

3. Experimental and computational methods

Full details of the materials, SWAXS measurements, mol­ecular dynamics simulations and calculation of the angle-dependent transmission factor A(2θ) using the developed Python software are provided in Appendix 6 (Glatter, 2018View full citation; Jorgensen et al., 1996View full citation; Watkins & Jorgensen, 2001View full citation; Dodda et al., 2017View full citation; Jorgensen et al., 1983View full citation; Abascal & Vega, 2005View full citation; Mahoney & Jorgensen, 2000View full citation; Berendsen et al., 1987View full citation; Abraham et al., 2024View full citation).

4. Results and discussion

As a starting point for treating absorption correction in our Kratky-type SAXS/SWAXS instrument, we used the work of Bowden & Ryan (2010View full citation), since their approach is conceptually closest to the capillary geometry considered here. In their terminology, the relevant case corresponds to a cylindrical specimen inside an annular support, treated in powder-diffraction capillary geometries. We adapt this geometry to the one shown in Fig. 1[link], which is commonly used in SAXS and SWAXS experiments on liquid systems contained in thin-walled cylindrical capillaries, as is typical in colloidal and soft-matter research. Although Bowden and Ryan considered an incident beam smaller than the specimen diameter and offset from the cylindrical axis, they discuss this aspect only in relation to 2θ-dependent offsets in experimental peak positions in their specific powder-diffraction geometry (Bowden & Ryan, 2010View full citation). However, we explicitly define and analyse the constrained illuminated scattering volume for the symmetric capillary alignment typical of the Kratky-type geometry to study its effect on the transmission correction factor in a systematic way. This is introduced through the integration limits a and b in equation (11)[link], where a defines the vertical half-height of the constrained illuminated volume and b defines the horizontal half-width along the capillary axis.

To gain a better understanding of the practical consequences of this adaptation, we first examined how the calculated transmission coefficient A(2θ) changes when the main geometric and material parameters of the model are varied separately. The effect of varying the beam half-height a in the vertical z direction is analysed systematically in Fig. 4[link](a) for a constant beam half-width b = 12 mm and uniform central illumination of the capillary. Overall, it is evident that increasing the vertical beam height results in a higher transmission coefficient at a given scattering angle. This is because illuminating a larger proportion of the capillary in the z direction increases the fraction of shorter path lengths, which contribute less to the total attenuation. These results highlight that the beam height has a significant influence on the transmission coefficient. Therefore, it is crucial to determine it carefully and include it as realistically as possible in the calculation of the coefficient A(2θ).

[Figure 4]
Figure 4
(a) The effect of increasing the primary beam half-height a at constant beam half-width b = 12 mm, capillary-wall thickness t = 0.01 mm and linear attenuation coefficient μ = 1.0 mm−1. (b) The effect of changing the beam half-width b and including the experimental beam profiles, all at constant t = 0.01 mm and μ = 1.0 mm−1. (c) The effect of increasing the linear attenuation coefficient μ on the angular dependence of the transmission coefficient A(2θ) at constant a = 0.1 mm, b = 12 mm and t = 0.01 mm.

The effect of varying the beam half-width b from 2 to 12 mm for a constant beam half-height a = 0.1 mm is presented in Fig. 4[link](b). The two curves (black open circles and red crosses) practically coincide, indicating that the horizontal width of the primary beam has almost no detectable effect on the transmission coefficient. Note that, although variations in the primary beam profile in the horizontal plane strongly affect the projection of scattering intensity onto the detector, leading to considerable instrumental smearing effects in the SAXS regime (Glatter, 2018View full citation), they evidently have only a negligible influence on the absorption correction. This is because, for fixed material attenuation coefficients, the absorption depends directly on the path length of the X-ray beam within the absorbing material, i.e. the sample and the capillary wall. Since the incident beam propagates along the y axis, displacements of the scattering event along the capillary axis (x) do not change the incident path length through this material. They affect only the exit path, and even this effect is small because the sample-to-detector distance (267 mm in our setup) is much larger than the horizontal displacement within the illuminated beam. The corresponding change in the total path length within the absorbing material is therefore small and has a barely noticeable effect on the absorption correction, as demonstrated in Fig. 4[link](b). To test this point further, we also considered the limiting case where the primary beam width is reduced to values comparable to the beam height, as may occur in experimental setups with strongly reduced slit width; an example with a = b = 0.1 mm is provided in Section 8.1 of Appendix 8.

The green curve with open squares in Fig. 4[link](b) represents a key result of the presented extension of the absorption correction procedure. It shows the A(2θ) data obtained by applying both experimentally measured primary beam profiles as weights within the constrained illuminated scattering volume, while also accounting for the capillary-wall attenuation. The fact that this curve does not deviate substantially from the other two curves in Fig. 4[link](b) may be somewhat misleading when assessing the importance of this extension. However, this only means that the selected value a = 0.1 mm is rather close to the appropriate effective primary beam half-height in our Kratky-type camera.

The message of Fig. 4[link](b) is one of the most important messages of this study and is twofold. First, properly accounting for the vertical illumination geometry of the primary beam is crucial for absorption correction, and including the non-uniform experimentally measured height profile of the primary beam in the correction procedure is the most reliable way to do this. Second, even when only a uniform primary beam is considered, there exists an effective beam half-height value that can closely approximate the effect of the primary beam height on A(2θ). However, this effective value is not known a priori and would have to be estimated correctly. The most realistic solution is therefore to incorporate the experimentally obtained incident beam profiles directly into the procedure, as implemented through the weighting function W(x, z) in equation (11)[link].

Having established this key geometrical point, we next consider the material parameter entering the correction. Therefore, Fig. 4[link](c) illustrates the effects of increasing the linear attenuation coefficient of the sample. Higher attenuation not only reduces the overall transmission but also enhances the angular dependence of the transmission coefficient. The latter occurs because increasing μ amplifies the differences between the contributions from different path lengths within the illuminated scattering volume. At small scattering angles, the average total path length through the absorbing material remains similar to the corresponding forward-transmission path length, whereas at larger angles the distribution of total path lengths broadens. Correspondingly, over the scattering angle range up to about 50° in Fig. 4[link](c) – roughly matching the SWAXS range of our Kratky-type instrument – this effect on the angular dependence is still moderate but becomes increasingly pronounced at higher angles.

The numerical reliability of the numerical integration in equation (11)[link] and geometrical robustness of the procedure are examined further in Appendix 8. The root-mean-squared error of A(2θ) falls below 0.1% for all tested attenuation coefficients, while the influence of capillary conicity was found to be negligible even for unrealistically large conicities.

In practice, the purpose of the proposed procedure is to correct measured SWAXS curves only over the measured angular range from 2θmin to 2θmax. Therefore, only the calculated A(2θ) values within this range are required in equation (5)[link], thereby avoiding the practical problems associated with primary beam transmission measurements discussed in Appendix 3. Nevertheless, it is useful to relate the calculated A(2θ) to the scalar factor Tj from equation (1)[link]. In equation (11)[link], the Beer–Lambert attenuation is averaged over the incident and exit paths of the photons that contribute to the detected elastically scattered intensity at a given 2θ. The forward-scattering limit A(0) in equation (11)[link] follows from the same reasoning. Nonetheless, the photon paths in the forward-scattering limit are the same as those of the directly transmitted photons recorded in primary beam transmission measurements. Therefore, A(0) could also be applied consistently to correct Ij,raw(0) from equation (1)[link], provided that the latter can be measured reliably (Appendix 3).

Consequently, for weakly absorbing samples with negligible angular dependence of the absorption correction, A(0) can be used as an approximate scalar numerical correction factor, analogously to Tj in equation (1)[link]. This is a useful qualitative alternative when Tj cannot be measured reliably. It also raises the question of whether A(0) could be approximated even more simply and still provide a satisfactory qualitative approximation. A straightforward option would be to use a deliberately simplified numerical correction based on the Beer–Lambert attenuation law, which considers only the sample and capillary-wall thicknesses and a single effective path length for all photons (Appendix 9).

We evaluate this simplified scalar correction strategy against the full A(2θ) transmission coefficients in Fig. 5[link]. It shows the A(2θ) values (full lines with symbols) together with the corresponding simplified Beer–Lambert scalar estimates (horizontal dotted lines) for the empty capillary, water and NFTB, for several capillary radii. The simplified Beer–Lambert scalar estimate is not physically equivalent to the full angle-dependent correction because it neglects the finite illuminated volume, beam-profile weighting and angle-dependent exit paths. Nevertheless, its values remain very close to the corresponding A(0) values for all tested cases, indicating that this similarity is not an artefact of the selected capillary radius. At larger scattering angles, however, the full A(2θ) curves increasingly deviate from this scalar value.

[Figure 5]
Figure 5
The angular dependence of the transmission coefficient A(2θ) at constant a = 0.1 mm, b = 12 mm, t = 0.01 mm, μquartz = 8.099 mm−1, μwater = 1.01 mm−1 and μNFTB = 2.20 mm−1 for the empty capillary (green solid line), water (blue solid line) and NFTB (black solid line) compared with the corresponding simplistic model values (dotted lines) for capillary radii of (a) 0.24 mm, (b) 0.34 mm, (c) 0.49 mm and (d) 0.74 mm.

Fig. 5[link] therefore supports two very important practical conclusions. First, the simplified Beer–Lambert-type scalar transmission estimate is a qualitatively useful approximation when the angular dependence of the transmission function is negligible, but it cannot replace the full angle-dependent correction when quantitative SWAXS data reduction is required. Second, the transmission functions for water and the empty capillary differ substantially, both in their absolute transmission level and in their angular dependence. This indicates that the water/empty-capillary correction step must be executed with care. If unreliable experimental transmission factors are used, this step may introduce notable errors into absolute-scale calibration using water as a secondary standard.

The practical consequences of these correction strategies are briefly summarized here and documented in detail in the supporting information. Appendix 4 first shows the failure of the conventional scalar primary beam transmission correction for the strongly absorbing NFTB sample. Section 10.1 of Appendix 10 then shows a successful SWAXS data reduction for this sample using the proposed numerical correction procedure. The water/empty-capillary correction relevant to absolute-scale calibration is presented and discussed in Section 10.2 of Appendix 10, a diverse range of Brij 35/alcohol/water systems in Section 10.3, and the simplified Beer–Lambert-type scalar approximation in Section 10.4. Appendix 11 provides further direct experimental cross checks of the proposed correction procedure for the water and NFTB datasets. These checks use independent Mythen 1K SAXS detector measurements performed on the same sample, in the same capillary and on the same Kratky-type camera, without changing the sample/capillary arrangement. Together, these examples and cross checks support the practical applicability of the calculated transmission workflow for background subtraction and absolute scaling in the experimental cases considered here.

5. Conclusions

The central result of this study is that, when experimentally determined scalar primary beam transmission factors become unreliable, the corresponding transmission correction problem can be bypassed using calculated geometry-specific transmission factors A(2θ) based only on the experimental geometry, measured primary beam profiles and tabulated attenuation coefficients. For the line-collimated Kratky-type SWAXS geometry considered here, the vertical illumination profile is the key geometric parameter controlling the absorption correction, whereas the horizontal beam width effect is negligible. Using the measured primary beam profiles removes the need to choose an effective uniform beam height as an additional approximation. The performance of the proposed angle-dependent transmission correction procedure has been demonstrated for the strongly absorbing NFTB sample. For this sample, the conventional scalar transmission correction led to a physically inconsistent result after background subtraction, while the numerical correction resolved this problem and enabled physically meaningful absolute-scale SWAXS data reduction.

The results also show that the bias in experimentally determined scalar transmission factors associated with primary beam radiation hardening is not limited to strongly absorbing samples. Such bias can become important whenever a measurement series contains samples with substantially different X-ray attenuation. One example is the water/empty-capillary measurement pair used in routine absolute scaling with water as a secondary standard. The proposed correction procedure successfully bypasses the corresponding transmission correction problem in this absolute-scaling step as well. Comparisons of independent model-based theoretical SWAXS curves (Lajovic et al., 2010View full citation) with the reduced experimental data, and additional measurements with the Mythen 1K detector, provide consistency checks of the corrected data-reduction procedure. Importantly, neither the correction factors nor the corrected experimental SWAXS curves were fitted, tuned or adjusted to improve the agreement between the compared data.

Taken together, the presented results suggest the following practical hierarchy of correction strategies: (i) conventional primary beam transmission correction is adequate when the measured transmission factors Tj are reliable and absorption differences between measurements are small; (ii) a simplified Beer–Lambert-based scalar correction can be useful as a reasonably fast qualitative or approximate correction when the angular dependence is weak; and (iii) the full A(2θ) correction is required for physically consistent quantitative SWAXS analysis when absorption differences between measurements are significant. This practical distinction is not specific to the Kratky-type geometry considered here and is relevant more generally whenever SAXS/SWAXS data reduction relies on experimentally determined scalar transmission factors whose reliability cannot be ensured.

6. Related literature

For further literature related to the supporting information, see Merck Life Science (2025View full citation), Nürnberg et al. (2016View full citation), Prince (2004View full citation), Rodriguez et al. (2001View full citation) and Tanaka et al. (2001View full citation).

Supporting information


Acknowledgements

We are most grateful to Professor Otto Glatter for his generous contribution to the light-scattering instrumentation of our laboratory.

Conflict of interest

We have no conflicts of interest to declare.

Data availability

The experimental data reported in this article are provided as a zip archive in the supporting information.

Funding information

We acknowledge support from the Slovenian Research and Innovation Agency (research core funding No. P1-0201, project No. N1-0308 `Nanoplastics in Aqueous Environments: Structure, Migration, Transport and Remediation', and project No. N1-0483 `Ionic Liquids and Nanoclays in Aqueous PFAS Removal') and from the National Research, Development and Innovation Office (project Nos. SNN142258 and SNN152145).

References

Return to citationAbascal, J. L. & Vega, C. (2005). J. Chem. Phys. 123, 234505.   Google Scholar
Return to citationAbraham, M., Alekseenko, A., Bergh, C., Blau, C., Briand, E., Doijade, M., Fleischmann, S., Gapsys, V., Garg, G., Gorelov, S., Gouaillardet, G., Gray, A., Irrgang, M. E., Jalalypour, F., Jordan, J., Junghans, C., Kanduri, P., Keller, S., Kutzner, C., Lemkul, J. A., Lundborg, M., Merz, P., Miletić, V., Morozov, D., Páll, S., Schulz, R., Shirts, M., Shvetsov, A., Soproni, B., van der Spoel, D., Turner, P., Uphoff, C., Villa, A., Wingbermühle, S., Zhmurov, A., Bauer, P., Hess, B. & Lindahl, E. (2024). GROMACS Archive , https://doi.org/10.5281/zenodo.11104865Google Scholar
Return to citationBaur, M., Uhlmann, N., Pöschel, T. & Schröter, M. (2019). Rev. Sci. Instrum. 90, 025108.   Google Scholar
Return to citationBendert, J. C., Blodgett, M. E. & Kelton, K. F. (2013). Acta Cryst. A69, 131–139.  Web of Science CrossRef CAS IUCr Journals Google Scholar
Return to citationBerendsen, H. J., Grigera, J.-R. & Straatsma, T. P. (1987). J. Phys. Chem. 91, 6269–6271.  CrossRef CAS Google Scholar
Return to citationBergmann, A., Orthaber, D., Scherf, G. & Glatter, O. (2000). J. Appl. Cryst. 33, 869–875.  Web of Science CrossRef CAS IUCr Journals Google Scholar
Return to citationBond, W. L. (1959). Acta Cryst. 12, 375–381.  CrossRef CAS IUCr Journals Web of Science Google Scholar
Return to citationBowden, M. & Ryan, M. (2010). J. Appl. Cryst. 43, 693–698.  Web of Science CrossRef CAS IUCr Journals Google Scholar
Return to citationBradley, A. (1935). Proc. Phys. Soc. 47, 879–899.  CrossRef CAS Google Scholar
Return to citationBrooks, R. A. & Di Chiro, G. (1976). Phys. Med. Biol. 21, 390–398.  CAS PubMed Web of Science Google Scholar
Return to citationCerar, J., Jamnik, A., Pethes, I., Temleitner, L., Pusztai, L. & Tomšič, M. (2020). J. Colloid Interface Sci. 560, 730–742.  CrossRef CAS PubMed Google Scholar
Return to citationCerar, J., Jamnik, A., Szilágyi, I. & Tomšič, M. (2021). J. Colloid Interface Sci. 594, 150–159.  CrossRef CAS PubMed Google Scholar
Return to citationCerar, J., Jamnik, A. & Tomšič, M. (2019). J. Mol. Liq. 276, 307–317.  CrossRef CAS Google Scholar
Return to citationCerar, J., Lajovic, A., Jamnik, A. & Tomšič, M. (2017). J. Mol. Liq. 229, 346–357.  CrossRef CAS Google Scholar
Return to citationChen, Y., Wei, L., Qian, F., Yang, Z., Wang, S., Wu, Y., Zhang, Q., Fan, Q. & Cao, L. (2018). Chin. Phys. B 27, 024101.  CrossRef Google Scholar
Return to citationClaassen, A. (1930). London Edinb. Dubl. Philos. Mag. J. Sci. 9, 57–65.  CrossRef Google Scholar
Return to citationDodda, L. S., Vilseck, J. Z., Tirado-Rives, J. & Jorgensen, W. L. (2017). J. Phys. Chem. B 121, 3864–3870.  CrossRef CAS PubMed Google Scholar
Return to citationDwiggins, C. W. (1972). Acta Cryst. A28, 219–220.  CrossRef IUCr Journals Web of Science Google Scholar
Return to citationDwiggins, C. W. (1975). Acta Cryst. A31, 146–148.  CrossRef IUCr Journals Web of Science Google Scholar
Return to citationFritz, G. & Bergmann, A. (2006). J. Appl. Cryst. 39, 64–71.  Web of Science CrossRef CAS IUCr Journals Google Scholar
Return to citationGlatter, O. (2018). Scattering Methods and Their Application in Colloid and Interface Science. Amsterdam: Elsevier.  Google Scholar
Return to citationHövelmann, J., Stawski, T. M., Besselink, R., Freeman, H. M., Dietmann, K. M., Mayanna, S., Pauw, B. R. & Benning, L. G. (2019). Nanoscale 11, 6939–6951.  PubMed Google Scholar
Return to citationHubbell, J. & Seltzer, S. (2004). X-ray Mass Attenuation Coefficients, NIST Standard Reference Database 126, National Institute of Standards and Technology, Gaithersburg, Maryland, USA. https://doi.org/10.18434/T4D01FGoogle Scholar
Return to citationJeffries, C. M., Ilavsky, J., Martel, A., Hinrichs, S., Meyer, A., Pedersen, J. S., Sokolova, A. V. & Svergun, D. I. (2021). Nat. Rev. Methods Primers 1, 70.  Web of Science CrossRef Google Scholar
Return to citationJorgensen, W. L., Chandrasekhar, J., Madura, J. D., Impey, R. W. & Klein, M. L. (1983). J. Chem. Phys. 79, 926–935.  CrossRef CAS Web of Science Google Scholar
Return to citationJorgensen, W. L., Maxwell, D. S. & Tirado-Rives, J. (1996). J. Am. Chem. Soc. 118, 11225–11236.  CrossRef CAS Web of Science Google Scholar
Return to citationKirschbaum, K., Martin, A. & Pinkerton, A. A. (1997). J. Appl. Cryst. 30, 514–516.  CrossRef CAS Web of Science IUCr Journals Google Scholar
Return to citationLajovic, A., Tomšič, M., Fritz-Popovski, G., Vlček, L. & Jamnik, A. (2009). J. Phys. Chem. B 113, 9429–9435.  CrossRef PubMed CAS Google Scholar
Return to citationLajovic, A., Tomšič, M. & Jamnik, A. (2010). J. Chem. Phys. 133, 174123.  CrossRef PubMed Google Scholar
Return to citationLajovic, A., Tomšič, M. & Jamnik, A. (2012). Acta Chim. Slov. 59, 520–527.  CAS PubMed Google Scholar
Return to citationMahoney, M. W. & Jorgensen, W. L. (2000). J. Chem. Phys. 112, 8910–8922.  Web of Science CrossRef CAS Google Scholar
Return to citationMerck Life Science (2025). Brij 35 Material Data Sheet. Sigma–Aldrich.  Google Scholar
Return to citationMøller, E. & Jensen, E. (1952). Acta Cryst. 5, 345–348.  CrossRef IUCr Journals Web of Science Google Scholar
Return to citationNürnberg, F., Kühn, B. & Rollmann, K. (2016). Laser-Induced Damage in Optical Materials 2016, pp. 42–54. SPIE.  Google Scholar
Return to citationOberta, P., Platonov, Y. & Flechsig, U. (2012). J. Synchrotron Rad. 19, 675–681.  Web of Science CrossRef CAS IUCr Journals Google Scholar
Return to citationOrthaber, D., Bergmann, A. & Glatter, O. (2000). J. Appl. Cryst. 33, 218–225.  Web of Science CrossRef CAS IUCr Journals Google Scholar
Return to citationPauw, B. R. (2013). J. Phys. Condens. Matter 25, 383201.  Web of Science CrossRef PubMed Google Scholar
Return to citationPauw, B. R., Smith, A. J., Snow, T., Terrill, N. J. & Thünemann, A. F. (2017). J. Appl. Cryst. 50, 1800–1811.  Web of Science CrossRef CAS IUCr Journals Google Scholar
Return to citationPi, H. L., Aragones, J. L., Vega, C., Noya, E. G., Abascal, J. L., Gonzalez, M. A. & McBride, C. (2009). Mol. Phys. 107, 365–374.  CrossRef CAS Google Scholar
Return to citationPolikarpov, M., Snigireva, I. & Snigirev, A. (2014). J. Synchrotron Rad. 21, 484–487.  Web of Science CrossRef CAS IUCr Journals Google Scholar
Return to citationPrince, E. (2004). International Tables for Crystallography, Vol. C, Mathematical, Physical and Chemical Tables, ch. 6.3, pp. 599–608. Heidelberg: Springer.  Google Scholar
Return to citationRitter, H., Harris, R. & Wood, R. (1951). J. Appl. Phys. 22, 169–176.  CrossRef CAS Google Scholar
Return to citationRodriguez, A., Canosa, J. & Tojo, J. (2001). J. Chem. Eng. Data 46, 1506–1515.  CAS Google Scholar
Return to citationSchmitt, D. & Ouladdiaf, B. (1998). J. Appl. Cryst. 31, 620–624.  Web of Science CrossRef CAS IUCr Journals Google Scholar
Return to citationShimizu, K. & Omote, K. (2008). Rigaku J. 24, 1–9.  CAS Google Scholar
Return to citationSulyanov, S., Gogin, A. & Boysen, H. (2012). J. Appl. Cryst. 45, 93–97.  Web of Science CrossRef CAS IUCr Journals Google Scholar
Return to citationTanaka, M., Girard, G., Davis, R., Peuto, A. & Bignell, N. (2001). Metrologia 38, 301–309.  CrossRef CAS Google Scholar
Return to citationTomšič, M., Cerar, J. & Jamnik, A. (2018). J. Mol. Liq. 259, 291–303.  Google Scholar
Return to citationTomšič, M., Cerar, J. & Jamnik, A. (2019). J. Colloid Interface Sci. 557, 328–335.  PubMed Google Scholar
Return to citationTomšič, M., Jamnik, A., Fritz-Popovski, G., Glatter, O. & Vlček, L. (2007). J. Phys. Chem. B 111, 1738–1751.  PubMed Google Scholar
Return to citationTomšič, M., Vlček, L., Jamnik, A. & Fritz-Popovski, G. (2007). Acta Chim. Slov. 54, 484–491.  Google Scholar
Return to citationVan de Casteele, E., Van Dyck, D., Sijbers, J. & Raman, E. (2002). Phys. Med. Biol. 47, 4181–4190.  Web of Science PubMed CAS Google Scholar
Return to citationVan Hoesen, D. C., Bendert, J. C. & Kelton, K. F. (2019). Acta Cryst. A75, 362–369.  CrossRef IUCr Journals Google Scholar
Return to citationWatkins, E. K. & Jorgensen, W. L. (2001). J. Phys. Chem. A 105, 4118–4125.  CrossRef CAS Google Scholar
Return to citationWu, H. & Li, Z. (2024). J. Synchrotron Rad. 31, 1197–1208.   CrossRef CAS IUCr Journals Google Scholar

This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.

Journal logoJOURNAL OF
APPLIED
CRYSTALLOGRAPHY
ISSN: 1600-5767
Follow J. Appl. Cryst.
Sign up for e-alerts
Follow J. Appl. Cryst. on Twitter
Follow us on facebook
Sign up for RSS feeds