short communications
A new image-plate reader for various sizes and shapes
aAdvanced Technology Research Laboratories at Futtsu, Nippon Steel Corporation, 20-1 Shintomi, Futtsu, Chiba 293, Japan, bAdvanced Technology Research Laboratories at Hiyoshi, Nippon Steel Corporation, 3-35-1 Ida, Nakahara-ku, Kawasaki 211, Japan, and cMAC Science Co. Ltd, 1-5-1 Shin-Yokohama, Kohoku-ku, Yokohama 222, Japan
*Correspondence e-mail: takagi@kaiseki.re.nsc.co.jp
A new image-plate reader which can read image plates (IPs) having various sizes and shapes has been developed. The machine consists of a rotating laser head to induce luminescence photons from IPs and a porous cylindrical holder made of a composite of cast iron and porous ceramic – so-called `breathnite'. Image plates can be fixed onto the inner surface of the cylindrical holder by evacuating the outer surface through a vacuum chamber with a mechanical pump. Image plates having various shapes and sizes up to 400 × 400 mm can be fixed without covering the open surface on the cylinder. In spite of such flexibility to various applications, the reader produces competitive performances compared with other commercial readers.
Keywords: image plates; readers; breathnite.
1. Introduction
Since Sonoda et al. (1983) invented image plates (IPs) as a radiological tool, they have become indispensable tools for various experimental techniques as highly sensitive two-dimensional radiation detectors. As various applications using IPs have been developed, IPs having various sizes and shapes have become commonly used. However, most IP readers can only read IPs having fixed sizes and rectangular shapes, such as 40 × 20 cm or 20 × 25 cm. Although some IP readers can read variable-size IPs, special attachments and settings depending on the sizes and shapes of the IPs are necessary to read them properly.
A new IP reader which can read IPs having various sizes and shapes without requiring any attachments was developed to alleviate such inconveniences.
2. Mechanism of the IP reader
As shown in Fig. 1, the main parts of the machine, SHIN-MAC ONE (made at MAC Science Co. Ltd, Yokohama, Japan), consists of a rotating laser [He–Ne (λ = 633 nm), 10 mW] head and a porous cylindrical holder made of a composite of cast iron and porous ceramic materials, called `breathnite' (Nabeya Corporation Ltd, Gifu, Japan) (Horie, 1988, 1989; New Materials Technology Update, 1987).
The IP is vacuum-sucked onto the inner surface of the cylinder by evacuating the outer surface through a vacuum chamber with a mechanical pump. The laser beam induces luminescence photons from the IP. By rotating the laser head (382 r.p.m.) and displacing the cylinder towards the vertical direction of the laser rotation, the entire surface of the IP can be scanned. A side view of the reader is shown in Fig. 2. The luminescence photons are detected by two parallel photomultipliers having different gains (250:1) to switch with each other, depending on the photon intensities, and send the data to a workstation (NSSUN SPARC STATION IPX) through ADC (16 bits). The mechanism ensures a wide and linearity, as shown in Fig. 3.
Image plates having various shapes and sizes (smaller than 400 × 400 mm) can be read without any attachments covering the open surface of the cylinder, since the air conductance of the breathnite, having controlled pore sizes (∼0.1 mm diameter) and ratios (∼60–90 vol.%), is large enough to evacuate the vacuum chamber and small enough to maintain a pressure difference between the inner and the outer surface of the cylinder holder to vacuum-suck IPs without covering the open surface. To ensure spatial .
of reading, the accuracy of the radius of the cylinder (200 mm diameter) was controlled carefully to less than 20 µm, which is the depth of the focus of the laser beam. The structure of breathnite is schematically described in Fig. 43. Main specifications and machine performances
The main specifications of the reader are summarized in Table 1.
|
The readout linearity was checked using an RI source (55Fe, 3 µCi) with changing exposure times. The source intensity was changed using Al foil attenuators. The results are plotted in Fig. 3. The readable intensity range of the luminescence photon is from 1 to 106 photons pixel−1, and five-digit linearity is ensured; the performance matches that of other commercial readers (see, for example, Miyahara et al., 1986).
To examine the readout uniformity, a 200 × 250 mm image plate was set parallel to the direct-beam direction inside the entrance door of the experimental hutch of PF-BL15C, and exposed for 8 s to an air-scattered white beam. The exposed plate was scanned by the laser beam towards either the longitudinal or traverse direction of the IP to check the effect of the reading direction on readout uniformity. The results are shown in Fig. 5. In the analysis of the data, the IP area was divided into nine sections having equivalent area, and 25 × 25 pixels were chosen from the centre part of each section for analysis of the total intensity and its standard deviation in the area. No systematic deviation from the centre of the IP, both in the sum and the standard deviation of the total intensities, was observed in both the longitudinal and traverse scans. The is as good as −0.05 in the summation, and ±0.1 in the standard deviation. The values are close to those of typical commercial readers (Amemiya et al., 1988).
4. Examples for application
A 100 nm-thick sputter-deposited α-Fe2O3 thin film was used for a test sample. The sample was prepared on an Si (111) wafer at 773 K by the magnetron sputtering deposition method with an oxide target having the same composition as the deposited film. A highly focused and high-flux synchrotron radiation beam was used as an incident beam with an incident angle of 0.5° under a total reflection condition at PF-BL3A. The IP was mounted on the spherical-type goniometer (Takagi et al., 1995) and set for the G-GIXD geometry (Takagi & Kimura, 1998). In Fig. 6 an IP image read by SHIN-MAC ONE is shown as an example. Both strong diffraction arcs from α-Fe2O3, indicating a of the film, and some weak spots from the single-crystal Si wafer are clearly shown in the image.
5. Conclusions
As was shown in the above sections, SHIN-MAC ONE, the first IP reader developed to read IPs having various sizes and shapes without any attachments, by using breathnite as the IP holder, showed competitive performances compared with other commercially available IP readers in spite of its flexibility for various applications. Although applications for various experiments must be investigated further in the future, we believe that the reader will become a powerful tool which may be applicable to various experiments in which special settings of IPs having various sizes and shapes are required.
Acknowledgements
The authors would like to thank Satoshi Matsumoto of MAC Science Co. Ltd for technical contributions to the design and assembly of the IP reader. Kazumi Obora of Nabeya Corporation Ltd contributed to the design and assembly of the breathnite cylinder. The contributions of Toshiyuki Mizutani of Nippon Steel Corporation to the early stages of this work were essentially important.
References
Amemiya, Y., Satow, Y., Matsushita, T., Chikawa, J., Wakabayashi, K. & Miyahara, J. (1988). Topics in Current Chemistry, Vol. 147, pp. 121–144. Berlin/Heidelberg: Springer-Verlag.
Horie, T. (1988). UK Patent Appl. 2196623A.
Horie, T. (1989). US Patent 4833106.
Miyahara, J., Takahashi, K., Amemiya, Y., Kamiya, N. & Satow, Y. (1986). Nucl. Instrum. Methods Phys. Res. A, 246, 572–578. CrossRef Web of Science
New Materials Technology Update (1987). Metalwork. Eng. Mark. September Issue, pp. 93–95.
Sonoda, M., Takano, M., Miyahara, J. & Kato, H. (1983). Radiology, 148, 833. CrossRef PubMed
Takagi, Y., Kikuchi, T., Mizutani, T., Imafuku, M., Sasaki, S. & Mori, T. (1995). Rev. Sci. Instrum. 66, 1802–1805. CrossRef CAS Web of Science
Takagi, Y. & Kimura, M. (1998). J. Synchrotron Rad. 5, 488–490. Web of Science CrossRef CAS IUCr Journals
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.