The use of a Hamamatsu X-ray image intensifier with a cooled CCD as a solution X-ray scattering detector
aStructural Biophysics Laboratory, RIKEN Harima Institute/SPring-8, SPring-8, 1-1-1 Kouto, Mikazuki, Sayo Hyogo 679-5148, Japan, bDivision of Biophysical Engineering, Graduate School of Engineering Science, Osaka University, Machikaneyama, Toyonaka 560-8531, Japan, and cExperimental Research Division, JASRI (Japan Synchrotron Radiation Research Institute), SPring-8, 1-1-1 Kouto, Mikazuki, Sayo, Hyogo 679-5198, Japan
*Correspondence e-mail: fujisawa@sp8sun.spring8.or.jp
CCD detectors are now widely used in many synchrotron small-angle X-ray scattering beamlines. The use of an X-ray image intensifier with cooled CCD (XR-II + CCD) was studied, especially for use in synchrotron solution X-ray scattering. Two samples, polystyrene latex and apoferritin, were used. These two samples have fine structure in the solution scattering profile due to symmetry and narrow size distribution. The recorded scattering profile, in comparison with that obtained by a position-sensitive ). Rev. Sci. Instrum. 66, 2290–2294], have very little effect on the SAXS results. This paper presents a practical protocol for obtaining a reliable solution scattering profile given the limitations of XR-II + CCD for synchrotron solution X-ray scattering.
(PSPC), showed that XR-II + CCD has a much smaller practical (100:1) than that of a pixel well (7500:1). This limited was overcome by placing various-size masks on the detection plane, thereby eliminating the high-intensity region. The images recorded with various masks were combined, and the reconstituted solution scattering profile was submitted to various analyses, including Guinier analysis, power-law analysis, size distribution analysis and calculation of radial density distributions. The results were the same as those obtained with the PSPC. This indicates that spatial distortion as well as shading, a decrease in sensitivity from the centre to the edge of the detecting region [Amemiya, Ito, Yagi, Asano, Wakabayashi, Ueki & Endo (1995Keywords: SPring-8; solution X-ray scattering; X-ray image intensifier with cooled CCD; apoferritin.
1. Introduction
Solution X-ray scattering is the scattering from particles in a solution, which includes polymers,
micelles, proteins and Because of its wide application, solution X-ray scattering therefore dominates the field of small-angle scattering. In order to clarify the problems in the use of a new detector, the characteristics of solution X-ray scattering are as follows.(i) High intensity in the inner scattering region, especially around the beam stop, with a gradual decay in intensity (1/100–1/10000) toward the edge.
(ii) A weak signal in the region near the edge comparable with that of natural radiation.
(iii) Stability in the time scale (days or hours) in order to ensure identical conditions between sample and buffer profiles.
(iv) Uniformity in the detection area, since solution scattering analysis often uses the tangent to the intensity rather than the position of a peak.
(v) High resolution (∼200 µm) in the centre area, while near the edge low resolution (<1 mm) is sufficient because no sharp peaks are seen in the profile.
(vi) Capability of performing time-resolved experiments.
In the early days of solution scattering the detector was a ; Fowler et al., 1983; Boulin et al., 1982). The integration-type detector represented by an imaging plate (IP) has not been used for quantitative measurements. The main problem with using an IP for solution scattering is that time-resolved experiments cannot be conducted. Furthermore, there is no systematic evaluation for solution X-ray scattering. One of the problems is the time-dependent feature of phosphorus decay, which makes it difficult to satisfy the third requirement in the above list.
and this was later replaced by a position-sensitive (PSPC). The successive derivatives of PSPCs, such as the quadrant detector and two-dimensional PSPC, used for solution scattering were all gas detectors (Gabriel, 1977The gas-type detector represented by the PSPC has been widely used in the second-generation synchrotron sources because it has long given satisfactory results. However, the modern synchrotron facilities with insertion devices such as wigglers or undulators changed the situation: standard PSPCs cannot tolerate the intense scattering from the sample. Besides, the CCD detector is more efficient than the standard PSPC system, which reduces both collection time and radiation damage on protein solutions. As a result many SAXS beamlines at second- or third-generation sources have started to employ the CCD-based detector as the main detector (Amenitsch et al., 1999; Bras et al., 1999; Fujisawa et al., 1997; Narayanan et al., 1999; Seifert et al., 1999). There are growing interests from SAXS users, especially solution scattering synchrotron users, who want to know how the familiar results of second-generation sources with the standard PSPC differ from those of third-generation sources with CCDs.
XR-II + CCD, developed by Amemiya's group (Amemiya et al., 1995), has two major merits: (a) a large (b) capability of performing time-resolved experiments. The drawbacks, on the other hand, are spatial distortion and shading effect, i.e. loss of especially in the region near the edge. These features are common to CCD-based detectors. Many studies on the application of XR-II + CCD to diffraction studies have been carried out, and a calibration program has been developed (Hammersley et al., 1994; Ito, 1998). On the other hand, there have been few studies on the application of this detector to solution scattering. The results of detector correction can be seen directly in the image in the case of diffraction, while solution scattering requires further processing and the detector correction is only apparent in the final stage of analysis. The CCD-based detector apparently compromises extra difficulties compared with standard PSPCs. However, there are no alternative detectors available for third-generation sources. In this paper we describe how to use the XR-II + CCD properly for the use of solution scattering, and the effects on their final results.
2. Materials and methods
2.1. Detector
The detector is composed of (i) an X-ray image intensifier, which has a beryllium entrance window; (ii) an optical lens; (iii) a CCD used as an image sensor; and (iv) a data-acquisition system. The X-ray image intensifier (Hamamatsu, V5445P-MOD) has an aperture of 150 mm diameter with a 1 mm-thick beryllium window. Incident X-rays are converted into visible photons by the phosphor (CsI:Na) at the beryllium window, and they are then converted into photoelectrons by a photocathode. An output phosphor [Gd2O2S:Tb (P43), 13 mm-diameter] is used to convert photoelectrons into intensified visible light. Optical coupling between the image intensifier and the image sensor is performed by an optical lens. An iris (hereafter called aperture) of the CCD regulates the amount of light in the optical coupling. Two CCD sensors are used: one (Hamamatsu, C4880-80-14A) for time-resolved measurements (hereafter called `CCDfast') and the other (Hamamatsu, C4880-10-14A) for static measurements (hereafter called `CCDslow'). The signal from CCDslow is digitized by a 14-bit A/D converter and the signal from CCDfast is digitized by a 14-bit or 10-bit A/D converter in static and real modes (28 Hz), respectively (Amemiya et al., 1995).
2.2. Sample preparations
Polystyrene latex (PS) was purchased from Sekisui Chemical Co. PS is very stable against X-ray radiation and has a spherical structure. The mean diameter of N-100 (Sekisui Chemical Co.) is 1200 Å with 2.5% standard deviation. The solid content of PS was measured by weight after evaporation of water. In order to enhance contrast as well as reduce the fluctuation contribution to total scattering, solid PS was dissolved in glycerol. The final concentrations of latex and glycerol were 5.1 wt% and 52.1 wt%, respectively. Apoferritin was prepared from horse spleen ferritin (Sigma) by stepwise dialysis against 1 to 0% thioglycollic acids (TGC) in 0.1 M acetic acid/NaOH, 0.1 M NaCl, pH 5.5. The ferritin solution (10 mg ml−1) was dialyzed first against 1% TGC overnight and then repeatedly by 0.1% TGC containing 1 mM 2,2′-bipyridyl until the pink colour of the solution disappeared. The solution was then dialyzed against 0% TGC, followed by a change of dialysis buffer to 0.1 M HEPES, pH 7.0. The apoferritin concentration was determined from its absorption at 280 nm (e = 19500 M−1 cm−1 per subunit; Heusterpreute & Crichton, 1981).
2.3. Solution scattering measurement by XR-II + CCD
Measurement of the solution scattering pattern was performed at RIKEN structural biology beamline I (BL45XU). A detailed description of the SAXS optics will be presented elsewhere (Fujisawa et al., 1999). The half size of the focus at BL45XU is 0.2 mm (vertical) × 0.4 mm (horizontal). The camera length was 2.2 m for PS, and 0.5 m and 1.8 m for apoferritin. The aperture of the CCD was determined so as to set the maximum charge of one pixel to less than 40000 e−, which is ∼60–70% of a pixel well. Incident intensity was scaled by an current. A dark image was collected after each data collection with the same collection time but without X-ray irradiation, by which the transfer noise from the CCD camera to the interface board as well as that from the after-glow of the phosphor were recorded. Typical output of a dark image is within 200 A/D counts, most of which are offset counts for an A/D converter. The X-ray wavelength was 1.0 Å.
2.4. Solution scattering measurement by PSPC
The reference data of XR-II + CCD were recorded by a SAXS camera at BL-10C, Photon Factory, Tsukuba (Ueki et al., 1985). The effective length of the PSPC was 200 mm with a delay line of 400 nS. The signal from the PSPC was converted to time–distance analogue information by a time-to-digital converter, then the time difference was converted to digital information of the position by a time-to-digital converter. The data were stored in histogram memory. The camera length was 200 cm and the X-ray wavelength was 1.488 Å.
A direct comparison of PSPC and XR-II + CCD was made at BL45XU at a wavelength of 1.0 Å. The effective length of the PSPC was 50 mm. The process of the output signal was basically the same as described above. The small area of the CCD image, which was identical to the detection area of the PSPC, was used for direct comparison.
2.5. Preliminary data processing of the solution scattering pattern
Subtraction of the buffer image was performed by using the following equation,
where S = 2 sinθ/λ, 2θ is the scattering angle and λ is the wavelength. I(S), I(S)sample, I(S)buffer, I(S)dark-sample and I(S)dark-buffer are the circular-averaged intensity of subtracted intensity, intensity of the sample, intensity of the buffer, intensity of dark collection after the sample, and intensity of dark collection of the buffer, respectively. ν is the of the solute and Asample and Abuffer are scaling factors of each intensity. The subtraction procedure was performed by using the Fortran program IISGNAPR (Fujisawa et al., 1997).
3. Results and discussion
3.1. Effects of spatial distortion and shading by image intensifier curvature
The effects of spatial distortion and shading are discussed in this section. The distortion correction program is based on the difference in position and integrated intensity of spots of a grid-patterned mask between XR-II + CCD and IP. The distortion is approximated as a three-dimensional function, such as a cubic spline function, by which the redistribution of intensity and the position of each pixel was performed (Hammersley et al., 1994). When a direct beam is radiated into the centre of the detection area, the distortion of XR-II + CCD at a given radial pixel is independent of the azimuthal position of XR-II + CCD and only a function of the pixel distance from the centre in Fig. 1. The correction factor for distortion and shading was approximated by a polynomial function by least-squares fitting. When the correction curves of CCDslow and CCDfast were scaled by pixel size, they coincided. There are several possible reasons for these effects; for example, the beryllium window curvature, inhomogeneity of XR-II, and optical lens response. These effects are also dependent on sample-to-detector distance, which affects the incident angle of the X-rays on the detection plane.
3.2. Direct comparison between standard PSPC and XR-II + CCD
The performance of XR-II + CCD was shown in solution scatterings from standard proteins with various molecular weights (Fig. 2). The scattering curves are identical although the quality of data with XR-II + CCD was superior to that of PSPC. The quality of PSPC data will be improved when Xe gas is used instead of Ar. Both data were identical; however, it should be noted that the incident beam should be attenuated (by ∼1/10) in order to keep the linearity of the PSPC.
3.3. Effect on the determination of radius of gyration
Based on these calibration features, the detector effect on the Rg, is discussed. Rg determination by {ln[I(S)] versus S2 plot} uses the following formula,
The above formula indicates the following. (i) In the case of the shading effect of XR-II + CCD, the detected I(S) becomes smaller than the ideal decrease in I(S) as S increases, and therefore the tangent of the becomes steeper and, consequently, Rg becomes larger. (ii) On the other hand, the distortion effect of XR-II + CCD enlarges the horizontal direction in the which reduces the tangent, i.e. Rg. The more pixels used, the larger the error in the determination of Rg. When two detector effects are compared, the distortion effect is negligible, since the uses the inner part of the detection area. The limit of the number of pixels for the determination of Rg is therefore determined by the shading effect which is the most important one. For example, Rg from CCDslow without correction lies within 1% error when 55 pixels from the centre of the detection plane were used. The determination of Rg was normally performed in the S region where 2πS [Å−1] Rg [Å] < 1.3. One pixel corresponds to (pixel size)/L, and therefore the camera length used for a given Rg [Å] should be less than L [mm] < Rg [Å] × 2π × 55 × (pixel size) [mm]/1.3, i.e. L [mm] < 40Rg [Å] for CCDslow. This equation is valid for CCDslow. In conclusion, the circular average of XR-II + CCD data can be used for Rg determination without a shading and deviation correction, if an appropriate camera length is chosen.
3.4. The of XR-II + CCD for an actual scattering image
The et al., 1995). The practical solution scattering image, however, is continuous and therefore may introduce another problem for the the intensity of one spot may influence a remote area, while the of the spot is assured. We used polystyrene latex in reverse contrast media as a standard sample for evaluation, since PS is a solid sphere and its scattering can be accurately approximated by the form factor of a solid sphere. In comparison with the PSPC, XR-II + CCD deviated in the S region where the scattering is less than 1/100 of the highest intensity (Fig. 3a). When the high-intensity regions were masked, XR-II + CCD recorded the same solution scattering as that recorded by the PSPC (Fig. 3b). This indicates that the practical is 100:1, far below the pixel depth. The situation is much worse when CCDfast is used; the actual was 40:1 (data not shown here). The most important point is that this effect cannot be corrected geometrically. One possible explanation for this effect is the contrast decrease in XR-II, for example, stray light in output window (Ito, 1998). Another possible explanation is that it comes from the lens optics. High intensity causes interference in a lens like a `cataract'.
of a detector is obtained by measuring the integrated intensities of small spots while changing the incident beam intensity. The of XR-II + CCD is ∼6000:1 (AmemiyaWhen the
is within limits, circular integration of the detected A/D counts greatly reduces the apparent signal-to-noise ratio. However, it should be noted that a sufficient quantum of A/D counts for one pixel is necessary to avoid digitization errors. A/D counts less than 2 after background subtraction often resulted in deviations from the ideal solution scattering profile.3.5. Power law analysis
Here we deal with power law analysis, which uses a wider scattering region than that of the . The scattering intensity of a solid sphere should theoretically follow the S−4 law, and the PS scattering from the PSPC as well as the reconstituted curve from XR-II + CCD does follow this law. The upper limit of fitting, 120 pixels from the centre of the detector, can be considered as the limit of this analysis without detector corrections.
Introducing various masks onto the detection plane in order to maintain a of 1:100 as well as collecting sufficient A/D counts gave the reconstituted scattering curve shown in Fig. 43.6. Determination of size distribution
In the following section the effects of shading and distortion on the final results of solution scattering analysis, which includes inverse Fourier transformation, are discussed. The mathematical transformation often results in an exaggeration of detector distortions and is more conspicuous in the final result than in the analysis in N(D), which estimates the volume-weighted size distribution of particles with diameter D. This is performed by solving the following equations,
One of the most sensitive methods is the regularization method for particle size distributionIn these equations function V(D) relates D to the particle volume. The solution of the equations is usually not stable and is greatly affected by the scattering curve itself. The volume-weighted size distribution N(D)V2(D) was calculated by using the software package SIZES based on the regularization method (Jemian, 1995). The final fit of the experimental data as well as the volume-weighted size distribution N(D)V2(D) are shown in Fig. 5. The peak diameter was 1250 Å and the half width of the distribution was 3.8%, which are consistent with those characterized by It should be noted that the scattering curves from XR-II + CCD are not corrected for distortion and shading effects. Detector correction for both shading and distortion has very little effect on the final data.
3.7. Box analysis on apoferritin
If a particle has spherical symmetry its radial electron density can be determined by using the box et al., 1989). The protein complex incorporates Fe+++ ions inside its shell and transports them across the cell membrane. Scattering from a spherical shell has a steep trough, which is easily smeared by detector effects. Fig. 6 shows the detector effect on the solution scattering profile of apoferritin. By applying various masks, smearing of the first trough was overcome, but there are still discrepancies for the second and third troughs. In dealing with a protein system the possibility of radiation damage cannot be ruled out. The smearing of troughs by changing different sample preparations was observed even with a PSPC, although they were much smaller than these with the XR-II + CCD. The limit to the could not be compensated by using masks in the case of a solution scattering profile with steep troughs (1/100 of peak maximum).
method. This method is also greatly affected by the shape of the solution scattering profile and therefore could be a good test for detector usage. Apoferritin is a protein complex consisting of a hollow spherical shell made of 24 protein chains arranged in a 432-point symmetry (SmithIf the smearing of troughs cannot be avoided, the next question is whether or not smearing affects the final results. For particles with spherical symmetry it is theoretically possible to calculate the radial electron density ρ(r) from the scattering intensity since, as a consequence of symmetry, no information is lost in the process of spherical averaging. The scattering amplitude A(S) = ±[I(S)]1/2 relates to ρ(r) by the following equation,
If the signs of the scattering amplitude are determined, ρ(r) can be restored. This solution to the sign (phase) problems is obtained under the condition where ρ(r) is finite, i.e. Router > r > Rinner, ρ(r) = 0, where Router and Rinner are values of the outer and inner radii of the shell, respectively. First, the phase is constructed by an arbitrary value of ρ(r). Then using that phase, ρ*(r), ρ(r) of the next iteration, is calculated but it is not finite. By imposing the condition Router > r > Rinner, ρ*(r) = 0, ρ(r) of the next iteration is constructed. Based on that ρ(r), the phases are recalculated and used for producing ρ(r) at the next iteration. The continuation of the iteration converges to a true ρ(r) (Svergun et al., 1984). Since this analysis is affected not only by the peak position but also by the slope of each peak, it could be a good detector test.
Fig. 7 shows radial electron density distributions of apoferritin taken from PSPC and XR-II + CCD and their fitting to experimental data. In the case of a solid spherical shell, ρ(r) becomes a step function, ρ(r) ≠ 0 in Router > r > Rinner. Since the X-ray crystallographic structure of apoferritin complex has been solved (Hempstead et al., 1997), from the atomic coordinates one can also calculate the radial electron density distribution. The difference between PSPC and XR-II + CCD was very small compared with the resolution from the algorithm itself: the peak position of radial electron density and Router and Rinner basically coincide with each other. The value of ρ(r) from XR-II + CCD is similar to that from crystal coordinates.
Fig. 8 demonstrates the power in the of XR-II + CCD. By using two-dimensional data, a fine structure in a medium-angle region can easily be obtained. The ripples are supposed to derive from the form factor of the shell.
4. Conclusions
Table 1 summarizes the comparison among standard SAXS detectors. XR-II + CCD has detector distortion and shading, as well as a low practical It is not an appropriate detector for standard X-ray generators. Despite these drawbacks, this detector has very attractive features; high efficiency, short readout time, capability of time-resolved experiment (>1 ms), tolerance of high dose and easy maintenance. These advantages are essential to modern synchrotron beamlines, therefore this detector will be used in various synchrotron facilities for the time being. This limit of can be overcome by using appropriate masks. The practical protocol for obtaining reliable solution scattering is as follows.
|
(i) Set the time and aperture so that the maximum intensity becomes 60 or 70% of the pixel depth.
(ii) Record the dark image after every sample and buffer measurement.
(iii) The net signal should be large enough to avoid digitization error. Parasitic scattering should therefore be minimized so as to maximize the net count in the limited data-collection time.
(iv) Practical dynamic ranges for CCDfast and CCDslow are 40:1 and 100:1, respectively. The high-intensity region should be masked.
(v) In the case of Guinier analysis, a shorter camera length is better; for example, L [mm] < 40Rg [Å], if there are no distortion and shading effects.
As long as the centre of the detection plane is used for analysis, the following samples can be measured without correction of distortion and shading:
(i) Protein solutions (Guinier analysis, distance distribution function and Kratky plot);
(ii) Polymer solutions (Fractal analysis);
(iii) Protein with high symmetry, such as virus shell protein complex.
In the case of a protein solution, the peak height of the Rg and I(S) are identical to those of the PSPC, most indirect methods give the same results.
is only relevant, for which the centre of the detector is used. When the slope or oscillation of the is relevant, it uses the edge region and the image needs special care. The calculation of the distance distribution function is more stable than that of the radial electron density distribution. When the values ofAcknowledgements
Dr Hidenori Toyokawa kindly helped with PSPC measurements and analysis at SPring-8. The authors thank Professor Yoshiyuki Amemiya and Dr Kazuki Ito for helpful discussions and comments in contrast effect of XR-II. Dr Tatzuo Ueki directed the beamline construction (BL45XU) at SPring-8. Dr Katsumi Kobayashi at the Photon Factory played a key role in maintenance of BL-10C at the Photon Factory. The synchrotron radiation experiments were performed at SPring-8 with the approval of the Japan Synchrotron Radiation Research Institute (JASRI) (proposal No. 1998A0102-CL-np). Part of this work (experiments at the Photon Factory) has been performed under the approval of the Photon Factory Advisory Committee (proposal No. 97G123). This work was supported by both the special coordination funds for promoting science and technology and SPring-8 Joint Research Promotion Scheme of Japan Science and Technology Corporation.
References
Amemiya, Y., Ito, K., Yagi, N., Asano, Y., Wakabayashi, K., Ueki, T. & Endo, T. (1995). Rev. Sci. Instrum. 66, 2290–2294. CrossRef CAS Web of Science
Amenitsch, H., Bernstorff, S., Dubcek, P., Kriechbaum, M., Menk, R., Mio, H., Pabst, G., Rappolt, M., Steinhart, M. & Laggner, P. (1999). Abstract in XIth International Conference on Small-Angle Scattering, 239. BNL, NY 11973-5000, USA.
Boulin, C., Dainton, D., Dorrington, E., Elsner, G., Gabriel, A., Bordas, J. & Koch, M. H. J. (1982). Nucl. Instrum. Methods, 201, 209–220. CrossRef CAS Web of Science
Bras, W., Oversluizen, M., Vlieg, E. & Goedtkindt, P. (1999). Abstract in XIth International Conference on Small-Angle Scattering, 272. BNL, NY 11973-5000, USA.
Fowler, A. G., Foote, A. M., Moody, M. F., Vachette, P., Provencher, S. W., Gabriel, A., Bordas, J. & Koch, M. H. J. (1983). J. Biochem. Biophys. Methods, 7, 387–392. CrossRef
Fujisawa, T., Inoue, K., Oka, T., Iwamoto, H., Uruga, T., Kumasaka, T., Inoko, Y., Yagi, N., Yamamoto, M. & Ueki, T. (1999). Abstract in XIth International Conference on Small-Angle Scattering, 276. BNL, NY 11973-5000, USA.
Fujisawa, T., Yagi, N., Inoue, K., Oka, T., Iwamoto, H. & Ueki, T. (1997). SPring-8 Annual Report 1997, p. 238. SPring-8, Hyogo 679-5198, Japan.
Gabriel, A. (1977). Rev. Sci. Instrum. 48, 1303–1305. CrossRef Web of Science
Hammersley, A. P., Svensson, S. O. & Thompson, A. (1994). Nucl. Instrum. Methods, A346, 312–321. CrossRef Web of Science
Hempstead, P. D., Yewdall, S. J., Fernie, A. R., Lawson, D. M., Artymiuk, P. J., Rice, D. W., Ford, G. C. & Harrison, P. M. (1997). J. Mol. Biol. 268, 424–448. CrossRef CAS PubMed Web of Science
Heusterpreute, M. & Crichton, R. R. (1981). FEBS Lett. 129, 322–327. PubMed Web of Science
Ito, K. (1998). PhD thesis, The Graduate University for Advanced Studies, Japan.
Jemian, P. R. (1995). Late-Nite Software. UNICAT, APS, IL 60439, USA (http://www.uni.aps.anl.gov/~jemian/pete/sas.html) .
Narayanan, T., Bosecke, P. & Diat, O. (1999). Abstract in XIth International Conference on Small-Angle Scattering, 256. BNL, NY 11973-5000, USA.
Seifert, S., Winans, R., Tiede, D. & Thiyagarajan, P. (1999). Abstract in XIth International Conference on Small-Angle Scattering, 257. BNL, NY 11973-5000, USA.
Smith, J. M., Ford, G. C., Harrison, P. M., Yariv, J. & Kalb, A. J. (1989). J. Mol. Biol. 205, 465–467. CrossRef CAS PubMed Web of Science
Svergun, D. I., Feigin, L. A. & Schedrin, B. M. (1984). Acta Cryst. A40, 137–142. CrossRef CAS Web of Science IUCr Journals
Ueki, T., Hiragi, Y., Kataoka, M., Inoko, Y., Amemiya, Y., Izumi, Y., Tagawa, H. & Muroga, Y. (1985). Biophys. Chem. 23, 115–124. CrossRef CAS PubMed Web of Science
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.