research papers
The Materials Science beamline upgrade at the Swiss Light Source
aSwiss Light Source, Paul Scherrer Institut, CH-5232 Villigen, Switzerland, bLaboratory for Micro- and Nanotechnology, Paul Scherrer Institut, CH-5232 Villigen, Switzerland, cExcelsus Structural Solutions SPRL, B-1150 Brussels, Belgium, and dDepartment of Machine Engineering, Paul Scherrer Institut, CH-5232 Villigen, Switzerland
*Correspondence e-mail: philip.willmott@psi.ch
The Materials Science beamline at the Swiss Light Source has been operational since 2001. In late 2010, the original wiggler source was replaced with a novel insertion device, which allows unprecedented access to high photon energies from an undulator installed in a medium-energy storage ring. In order to best exploit the increased
of this new source, the entire front-end and optics had to be redesigned. In this work, the upgrade of the beamline is described in detail. The tone is didactic, from which it is hoped the reader can adapt the concepts and ideas to his or her needs.Keywords: beamline upgrade; undulator radiation; X-ray optics; powder diffraction; surface diffraction; coherent X-ray diffraction imaging.
1. Introduction
The Materials Science (MS) beamline at the Swiss Light Source (SLS) was originally conceived to provide hard X-rays in the energy range 5–40 keV (Patterson et al., 2005) and serve experiments in powder diffraction (PD), surface diffraction (SD) and computed tomography. As such, it was the only insertion-device beamline at the SLS to exceed a photon energy of 20 keV.
In 2000, it was only possible to access such high photon energies at the intermediate storage-ring electron energies of = 2.4 GeV of the SLS by employing a so-called minigap wiggler (called `W61' to indicate the magnets' period of 60.5 mm), with a K = 8.6 at a gap size of 8 mm (Patterson et al., 2005). The transmitted at 12 keV from the wiggler of approximately 1013 photons s−1 after monochromatization through a Si(111) double-crystal monochromator (DCM) was at best 30% efficient. This was due primarily to the large heat load on the first DCM crystal, which caused a thermal bump and consequent reduction in despite the use of the dynamical correction system `TORII' (Schulte-Schrepping et al., 1995). Another consequence of the high output power of W61 was the need to use a so-called rotating carbon filter to remove the strongly absorbed soft X-ray part of the wiggler spectrum below approximately 5 keV, amounting to 1.6 kW, a power too high to be able to use a cooled diamond filter. This front-end component, as well as two beryllium windows, increased the effective source size to approximately 450 µm (h) × 250 µm (v).
of 7 keV and a deviation parameterSince 2000, however, undulator technology has matured, in particular with regards to magnet materials and in-vacuum technology. As described in detail in the next section, high photon energies can begin to be accessed at medium-energy storage rings, once radiation-hard small-period magnets with high maximum B0 become available. Such magnets were developed around 2003, and their implementation in undulator technology has allowed a considerable reduction of undulator-magnet period lengths from 19 mm to 14 mm.
With the advent of this technological advance, it was decided in 2007 to replace W61 with an in-vacuum cryogenically cooled permanent-magnet undulator [CPMU (Hara et al., 2004)], having a magnet period λu = 14 mm, resulting in the CPMU being called `U14'.
Because the beam divergence of U14, and hence the beam ). At the same time, the new design also took into account the possibility of introducing new techniques at the beamline, such as coherent X-ray diffraction imaging and micro Laue diffraction.
at the optics hutch, is so much smaller than that of W61, new optics had to be designed (see §6The paper is organized as follows: after a description of the specifications of the SLS storage ring, the U14 undulator and Materials Science beamline given in §2 and §3, details of the beamline layout, the front-end components and optics design are described in §4, §5 and §6, respectively. The performance of the beamline after the upgrade is covered in §7, and the impact of this on the two endstations, plus first results, are described in §8 and §9.
2. Specifications for the SLS and the Materials Science beamline
Since the last publication describing the MS beamline (Patterson et al., 2005), the storage-ring specifications have improved considerably in terms of vertical emittance and electron-beam size. This is due mostly to an exceedingly low coupling of the horizontal to the vertical emittance of 5 × 10−4. Such coupling is caused primarily by slight misalignments of the focusing quadrupole magnets. The most important parameters are listed in Table 1.
|
The SLS runs in `top-up' operation, in which the storage current is maintained to within 1% of the nominal value of 400 mA. This feature is important in providing a constant X-ray
and stable beam position.The specifications of the MS beamline are summarized in Table 2. The improved coherent fraction promised by undulator radiation should also be exploited. This feature sets stringent specifications on both the optics and on the vacuum quality; the gradual build-up of carbon deposits on the surface of the first crystal in the DCM, caused by the photo-induced cracking of carbon-containing gaseous species in the residual gas, can rapidly and adversely affect the quality of the wavefront. The residual gas pressure in the DCM was therefore specified as being less than 10−8 mbar.
|
3. The U14 undulator
3.1. Theoretical design considerations
The on-axis peak intensity In of the nth harmonic of undulator radiation is given by
whereby α = 1/137 is the fine-structure constant, γ is the Lorentz factor , Δν/ν = 1/nN is the relative spectral bandwidth of the nth harmonic peak, I is the current, e is the N is the number of magnet periods and Fn(K) is the so-called tuning function, given by
whereby
and J are Bessel functions of the first kind. Fn(K) is plotted as a function of K in Fig. 1(a). The deviation parameter K is defined as
whereby is the maximum angular excursion off-axis of the electrons passing through the undulator magnet array, m is the and c is the speed of light in a vacuum (Willmott, 2011).
One sees immediately from Fig. 1(a) that, even for relatively high values of K ≃ 1.5, Fn(K), and hence also In, is three orders of magnitude smaller for n ≃ 21 than for the first few harmonics. Fig. 1(b) is a plot of Fn(K) versus harmonic number n for three values of K.
The condition for constructive interference on-axis in an undulator is given by
or, in practical units,
Hence, for a given storage-ring energy , undulator magnet period λu and deviation parameter K, nλn is a constant, which, according to Fig. 1(a), we would like to be as small as possible, in order to access those tuning curves with as low as possible value of n. On the other hand, we require K to be as large as possible (Fig. 1b). The only possibility to achieve both these requirements is, therefore, to reduce λu [equation (6)] and increase B0 [equation (4)].
The previous shortest-period undulator at the SLS, U19, had λu = 19 mm. In order for the insertion device to remain an undulator, K should not exceed 2.0 or thereabouts. Hence, from equation (4), the maximum must be equal to or less than approximately 1.1 T. For the maximum K of 2.0, nλn for U19 was equal to 12.92 Å at the SLS. Accessing photon energies above 30 keV thus required working on the 31st harmonic, for which the intensity is unacceptably low [see Fig. 1(a)]. Indeed, U19 at the protein crystallography beamline at the SLS operates only up to 19 keV photon energy.
Because the U14 undulator is an in-vacuum insertion device, the gap size can in principle be closed to as little as 3.5 mm and the maximum B0 be as large as 1.5 T (Hara et al., 2004). This was measured as a function of gap size g using a Hall probe (see Fig. 2). The resulting K values [equation (4)] were then fit, yielding
In practice, the minimum gap size has been limited to 3.8 mm.
From equation (4), the maximum value of K is approximately 1.65, and nλn = 7.49. Accessing 30 keV photons requires one to operate at the 17th harmonic, for which the is an order of magnitude higher than that for U19 [see Fig. 1(a)].
3.2. The U14 CPMU design
A detailed description of the U14 CPMU design and performance is given elsewhere (Calvi et al., 2013). Here, the most important features are highlighted.
The undulator design parameters compared with those for the W61 wiggler and U19 undulator are summarized in Table 3. The magnet material chosen for U14 was Neomax S45SH, which has the alloy composition Nd2Fe14B. The remanent field of this material maximizes at approximately 140 K (Hara et al., 2004) and is higher than related commercial magnets. Importantly, demagnetization caused by exposure to high-energy electrons is almost completely suppressed at this temperature. While at room temperature the drops by almost 30% after exposure to 1.5 × 1015 2 GeV electrons, the same exposure at 140 K produces a drop in field strength of approximately 1%. U14 is hence cryogenically cooled and maintained at 138 ± 0.1 K, by a combination of liquid-N2 cooling and heating elements. The temperature is measured at three positions (upstream, centre, downstream) on both the upper and lower undulator support structures. A second important aspect of this material is that, although the only increases modestly when cooled to 140 K, its coercivity increases by over a factor of two, allowing one to use it at small gap sizes.
|
The liquid-N2 pump system is also used to cool the first crystal of the DCM (see §6.2). This element, including the internal metallic bellows for the nitrogen flow, causes its thermally insulated goniometer stage to cool over approximately 100 h from room temperature to 290 K. This has the sadly unavoidable consequence that experiments at high energy, for which the DCM crystals' Darwin widths are narrow, are relatively unstable during this cool-down period, as the goniometer slowly drifts thermally.
The short undulator period allows more magnet poles to be installed (N = 120, compared with that of U19 at the SLS, for which N = 96; see Table 3). From equation (1), we see that the of U14 is also enhanced by a factor of approximately 1.5 because of this aspect. Despite the complication of cooling the CPMU, it was possible to maintain a phase error of approximately 2.5° between the useful gap sizes of 3.8–6 mm. Plots of the variation of the gap size with photon energy for the third to 19th harmonics, plus the theoretical taking into account phase errors, are shown in Fig. 3, calculated using the SRW code (Chubar et al., 1998; see also https://www.esrf.eu/Accelerators/Groups/InsertionDevices/Software/SRW ). Also included in the latter are experimentally determined fluxes, recorded at the SD station using the photon-counting Pilatus 100k detector, corrected for the various kapton, diamond, mylar and Si3N4 components, the detector efficiency [Fig. 4(a)], the mirrors' (a factor which begins to become important above approximately 20 keV) and the bandwidth of the Si(111) crystals in the DCM (see below). At lower energies, the experimentally obtained is 70% or better compared with theory, dropping to approximately 50% above 25 keV.
Of more direct interest to the user, however, is the actual number of photons per second that he/she can expect after the optics hutch, as a function of energy. This is shown in Fig. 5. For purposes of comparison, the total of the old wiggler insertion device W61 is also plotted. At the lowest photon energies, the from U14 is orders of magnitude higher, dropping to about 50–75% of the W61 at the highest photon energies. Given that the unfocused beam of U14 is over 200 times smaller than that of W61, the areal of U14 is thus over two orders of magnitude higher.
At the SD station, the best focus is approximately 130 µm (h) × 40 µm (v), or about 20 times smaller than previously obtainable (see the inset of Fig. 5) recorded at 9 keV.
4. Beamline layout
The positions of the most important beamline components are listed in Table 4, and shown schematically as a block diagram in Fig. 6. Most of these, in particular the optics between 20 and 31 m, are described below in detail. It is briefly mentioned here that, because the tomography station was removed from the beamline in 2007, an aspect of the upgrade program was to move the powder diffractometer 3.5 m upstream, while the focusing optics were moved approximately 70 cm downstream, compared with the former configuration. This means that it is now possible to demagnify the photon beam at PD to 0.6, which is beneficial for certain experimental set-ups such as those using diamond-anvil cells, or in micro-Laue experiments.
The surface diffractometer has also been moved 70 cm upstream in the second hutch. Although this has only a marginal impact on the demagnification of the focal spot size, it does allow the possibility of increasing the sample–detector distance to over 2 m. As described in §9.2.1, this becomes important for future possible experiments exploiting coherence, such as coherent X-ray diffraction imaging (CXDI).
5. The front-end
The most important front-end components are listed in Table 4, from the mask at 8.245 m to the CVD diamond window at 17.851 m. The two-blade X-ray beam-position monitors (XBPMs) allow one to monitor both the height and angle of the white X-ray beam produced by U14. Downstream from these are two water-cooled diaphragms which intercept the soft X-ray outer cone of the synchrotron beam. The transmitted power through the second (narrower) diaphragm is 334 W.
Approximately half of this transmitted power lies in the soft X-ray regime below approximately 5 keV, which interacts very strongly with condensed matter and must therefore be removed to protect the X-ray optical elements. This is achieved with a combination of a 120 µm-thick water-cooled diamond filter, and an 80 µm-thick water-cooled diamond window, after which the transmitted power is 157 W. The transmission spectrum of 236 µm of diamond is shown in Fig. 4(b). The additional 36 µm of diamond is included in order to account for the three quadrant CVD-diamond beam-position monitors (qBPMs) permanently positioned in the beam in the optics hutch (Schulze-Briese et al., 2001).
6. Optics design concept
6.1. Choice of configuration
Two mirrors are necessary to sufficiently suppress harmonic contamination (Patterson et al., 2005). The previous optics configuration with W61 was collimating-mirror–DCM–focusing-mirror. As such, the first mirror (1 m long) absorbed as much as 1200 W of power from the wiggler source and therefore required water cooling (Patterson et al., 2005). There are some important advantages in replacing this set-up with one in which the two mirrors are downstream of the DCM (i.e. DCM–mirror–mirror, see Fig. 7). These include not needing to cool the first mirror, plus the option of using only one of the mirrors for experiments which profit from using as few optical elements as possible, such as CXDI. Lastly, in such a configuration, both mirrors can be housed in the same vacuum vessel, making the design more compact and cheaper.
In reaching a decision as to whether such a change of configuration should be adopted, the relative sizes of the natural vertical divergence (FWHM) of the undulator radiation and the Darwin width of the DCM Si(111) crystals had to be compared. The standard deviation of the undulator radiation perpendicular to the orbital plane is given by
and the FWHM of the beam in this direction is therefore
This is plotted in Fig. 8 along with the theoretical Darwin widths of Si(111) for selected energies. If one makes the simplification that the Darwin curve approximates a top-hat profile of width 2WD, the fraction fD of the beam diffracted by the Si(111) crystal is given by
This is also shown in Fig. 8. Where the two angles overlap (approximately 16 keV), one FWHM of the undulator radiation is intercepted by the Darwin curve, that is, 76% of the At the highest desired energy of 40 keV, almost 50% of the beam intensity is lost. Nonetheless, because the large majority of experiments are conducted at photon energies below 25 keV, the advantages of this new configuration were deemed to outweigh the loss of at higher energies.
6.2. Detailed geometry
The fixed-exit DCM design is similar to those already used at the protein crystallography beamline PX1 (Schulze et al., 1998) and the computed tomography beamline TOMCAT (Stampanoni et al., 2007), both at the SLS. The DCM was designed and built by CINEL s.r.l., Padova, Italy (https://www.cinel.com ). Both Si(111) crystals are mounted on their own goniometer. The second crystal can be bent sagittally to provide horizontal focusing (Schulze et al., 1998) (Fig. 9). A crucial difference, however, between the new optics design at MS and those at PX1 and TOMCAT is the beam offset (i.e. the difference in height between the incoming white beam from the undulator and the outgoing monochromatic beam). At PX1, this is +50 mm. Because at the MS beamline, we wish to access photon energies as high as 40 keV; having a fixed-exit beam with this offset would require the second crystal (X2) to move by approximately 600 mm in the beam-propagation direction (z), setting unreasonable specifications on the linear-motion stage for X2 with regards to parasitic motions, and also increase the volume of the monochromator vacuum chamber by a significant fraction. Radiation dosage studies of the Bremsstrahlung radiation at different offsets were performed, from which an offset of +20 mm was deemed to be well within the safety limits and meant that the maximum z-translation of X2 is reduced to approximately 200 mm.
This smaller offset compared with that of the PX1 beamline meant, however, that we could not take a carbon copy of the two crystal designs used in the PX1 monochromator (Schulze et al., 1998). The reason for this is that at low photon energies the crystals and their holders would overlap vertically, both occluding the beam and increasing the danger of crashing the crystals into each other when changing the photon energy. As calculated in this section, the vertical separation of X1 and X2 at 5 keV is only approximately 25 mm (see Fig. 10 below), too small to be able to use the same model of the PX1 beamline. The design concept, however, could be preserved, requiring only that the dimensions of X1 and X2 be scaled down. The most important design parameter was that the ratio of the length to width of the thinned bendable section of X2 (0.4 mm thick, used for sagittal focusing) be approximately 6 or greater to avoid anticlastic distortions [see Fig. 9(b)].
Another novel feature is shown in Figs. 9(c) and 9(d). It is well known that the coherent wavefront of quasimonochromatic synchrotron radiation can degrade over time because of the build-up of carbonaceous deposits on X-ray optics surfaces (especially on those exposed to high-intensity polychromatic beams), due to the cracking of carbon-containing species in the residual gas within the vacuum chamber (Chauvet et al., 2011). Considerable effort was invested to optimize the DCM vacuum without incurring unacceptable costs associated with ultrahigh-vacuum (UHV) motors and motion stages. An ultimate pressure better than 10−8 mbar was specified. This was met by CINEL.
However, the effective incident c)] by installing a copper `cap' which only opens up to an angle of approximately 25° perpendicular to the crystal surface (Fig. 9d). The opening solid angle of the cap is approximately 1 sr, or 16% of an unprotected surface. The first Si(111) crystal X1 and thus also the cap are maintained at liquid-nitrogen temperatures, and the latter therefore acts as a sacrificial cryopump, protecting X1. After more than one year of operation, no sign of a burn spot could be seen.
of residual-gas species can be further reduced by an order of magnitude by limiting the solid angle `seen' by the optics. This was achieved on X1 [on which the high-power white beam is incident, Fig. 9(A schematic diagram of the DCM and mirrors' configuration is shown in Fig. 7. The second Si crystal X2 is at room temperature. Because of the temperature difference of the two crystals (ΔT ≃ 200 K), they also have marginally different Bragg angles, due to the thermal contraction of silicon in X1.
In order to maintain a constant height for the exit beam after M2 (+20 mm relative to the incoming beam) for all photon energies, the second crystal must be translated both vertically (y) and horizontally (z), while the first mirror M1 must be translated vertically. In addition, the mirror angles are adjusted with photon energy, in order to minimize harmonic content. Both mirrors have three regions: bare silicon, Rh-coating and Pt-coating. The first mirror is kept flat, while the second mirror M2 can be bent to provide vertical (meridional) focusing.
coefficient of silicon changes almost linearly between 77 and 300 K (Slack & Bartram, 1975as one cools from 300 to 77 K.
We let the θ for the room-temperature crystal X2, and that for the same photon energy of X1 be θ + δ. Starting from Bragg's law, it is trivial to demonstrate that
at 300 K bewhere δ 1 is given in radians.
The consequence of this is that the beam is tilted upwards by an angle 2δ after X2 (see Fig. 7). This is compensated by tilting the first mirror to an angle of α + δ/2 relative to the beam and the second mirror to α − δ/2, so that the exiting beam is again horizontal. Note that δ/2α ≃ 0.01 and hence this readjustment has no significant impact on the reflectivity of the mirrors.
The two mirrors were fabricated by WinlightX, Pertuis, France (https://www.winlightx.com ) from single-crystal silicon instead of the more conventional material, fused silica. The main reason for this choice was the fact that the of silicon is 100 times greater than that for silica, thereby minimizing the formation of thermal bumps when operating in pink-beam mode (i.e. bypassing the DCM) for possible future micro-Laue experiments. The mirrors' reflecting surfaces consisted of 7 mm-wide stripes of Rh coating, bare Si, and Pt coating (coating thickness of 55 nm and 37.5 nm, respectively), separated from each other by 4.5 mm. The useful mirror length was 400 mm and the mirror 40 mm × 40 mm. The tangential slope error was specified to be better than 0.30 µrad (RMS). Each stripe is accessed by translation of the entire mirror chamber in the x-direction by ±11.5 mm (see Fig. 7).
The critical angle for total external reflection, αc, is inversely proportional to the photon energy, and proportional to the square-root of the electron density. As is often observed, the actual densities of the sputtered Rh- and Pt-coating stripes are marginally less than for bulk material. These were accurately determined by recording the reflectivity curves of test pieces prepared in the same sputter-deposition run used to make the mirror coatings themselves, and at three different photon energies. The Rh- and Pt-coatings were found to be 12.0 g cm−3 (96.8% dense, 3.142 e Å−3) and 20.3 g cm−3 (94.9% dense, 4.86 e Å−3), respectively. The atomic roughness could also be fit, and yielded 0.25 nm for the Rh surface and 0.50 nm for Pt.
The mirrors should be tilted to close to the critical angle, in order to suppress harmonic contamination as much as possible. However, one should avoid getting too close to αc, as the drop in reflectivity at the critical angle is not infinitely abrupt. We therefore chose a value of 85% of αc. The reflectivity at the lower and upper limits of the energy ranges used for each stripe is given in Table 5.
|
We express this set angle in convenient units as
where αset is in degrees, ρ is the electron density in e Å−3, and the photon energy E is in keV. This yields
Referring to Fig. 7, we see there are some constants to the optics geometry. These are:
(i) The vertical offset of the monochromatic beam emerging from the optics compared with that of the incoming pink beam, Δy = y1 + y2 − y3 = +20 mm.
(ii) The horizontal distance between the centres of M1 and M2, z3 = 700 mm.
(iii) The horizontal distance between the centres of X1 and M1, z0 = z1 + z2 = 1372 mm.
We now determine expressions for the positions and angles of the four optical components. First, the
of X2 is simplyRemembering that Δy = y1 + y2 − y3 leads to
From this, we directly obtain the horizontal distance between the centres of X2 and M1, z2 = z0 − z1, from which y1, y2 and y3 directly follow.
The crystal and mirror positions as a function of photon energy for the three mirror coatings are provided in Fig. 10, with the coating densities given above. The limits of the positions at the lowest (5 keV, using Si reflection on the mirrors) and highest energies (40 keV, using Pt) are shown in Fig. 11.
7. Performance
7.1. DCM energy resolution
Initial calibration of the DCM Bragg angles, plus the DCM K-edge, which has a well studied pre-edge feature at 8.9813 keV. The raw spectrum, only corrected for the small calibration offset in the is shown in Fig. 12(a), whereby the pre-edge feature is highlighted with an asterisk. The separation in the valley and peak energies of the pre-edge is 1.7 eV, or 1.9 × 10−4 of the edge energy. This high visibility demonstrates that the DCM energy resolution ΔE/E is at the theoretical value of approximately 1.4 × 10−4.
were tested using an unfocused beam by recording the CuThe calibration offsets for the Bragg angles of X1 and X2 determined from the Cu-edge XANES spectrum are only constant if there is insignificant parasitic pitch associated with the relatively large horizontal movement of X2 between low and high photon energies (see Figs. 10 and 11). A second Pd-edge XANES spectrum was recorded at 24.35 keV (Fig. 12b), from which it was observed that θ was accurate to within 5 × 10−5 degrees, and that therefore the absolute accuracy of the set energy lay within the Darwin width of the crystals.
Once the DCM has thermally stabilized after initial cooling of the first crystal to 77 K, the photon energy remains constant to better than 0.3 eV (one part in 105), as was demonstrated by recording at 25 keV a silicon powder pattern every 45 s over 10 h.
7.2. Focusing elements
7.2.1. Primary optics
As mentioned earlier, vertical focusing is achieved using a flexural hinge-based mirror bender, similar to systems described previously (Rossetti et al., 2002). The X-ray beam is focused in the horizontal plane by sagittally bending the second DCM crystal X2 (Schulze et al., 1998).
From the lensmaker equation
whereby p is the source–lens distance and q is the lens–image distance, we can establish the required focal lengths for focusing at the PD station, at the SD station and for parallel beam (Willmott, 2011). For horizontal focusing, ph ≃ 20.268 m (the midpoint of the horizontal translation of X2; see Table 4), while, in the vertical plane, pv = 22.214 m, the centre of mirror 2. For focusing at the powder diffractometer, qh,PD = 12.529 m and qv,PD = 10.583 m, while, at the surface diffractometer, qh,SD = 19.819 m and qv,SD = 17.873 m. Table 6 lists in the second and third columns the focal lengths in the vertical and horizontal planes for focusing at the powder diffractometer, the surface diffractometer and for parallel beam.
|
The bending radii Rs and Rm for sagittal focusing using the second DCM crystal and meridional focusing using the second mirror, respectively, are given by
and
whereby θi is the incident angle of the X-ray beam on the optical element and is equal to the θ in the case of sagittal focusing using X2; and equal to the mirror incident angle α for vertical focusing using M2. Both θ and α are energy dependent. The limiting values for 5 and 40 keV are listed in Table 6. The change in the bending radii as the beam focus is adjusted between 30 and 43 m from the centre of U14 is shown in Fig. 13.
We now consider the necessary range of motion and resolution of the actuators used to bend X2 and M2. We begin by assuming cylindrical focusing. The pertinent geometries are sketched in Fig. 14. We begin with the simpler configuration of the mirror bender (Fig. 14b). From Pythagoras' theorem,
from which one obtains
under the valid assumption that ym/Rm 1. Both mirrors in the new optics set-up have a usable length 2Lm = 400 mm. The maximum value for ym is thus at the minimum bending radius, which is required for focusing 5 keV photons at the PD station (Rm = 2.723 km), and therefore ymmax = 7.34 µm.
Differentiation of (26) with respect to Rm leads to
This tells us the change in ym (dym) for a small change in the meridional bending radius dRm. What interests us more, however, is the amount we need to move the actuator for a given small change in q, the position of the focal spot (i.e. dq). We differentiate by parts equation (24) with respect to q to obtain
Multiplying (27) with (28), we obtain
Substituting for Rm using (24) leads to
We want to control the focal position with an accuracy of 10 cm. From (30), the smallest change in ym, dym, is required for large q (i.e. at the surface diffractometer) and for high photon energies (small incident angle α). Under these conditions, control of q with an accuracy dq = 10 cm therefore requires dym = 5.4 nm, approximately 1400 times smaller than ymmax.
Bending mirror 2 was achieved using the flexural hinge system described by Rossetti et al. (2002). Silicon blades, 10 mm-thick, are attached to a rigid clamp system holding the ends of the Si mirror. The bending depth ym is proportional to the amount the Si blades are pushed outwards (P1 and P2 in Fig. 14). It is hence clear from equation (26) that the meridional radius of the second mirror, Rm, is inversely proportional to the positions of the Si-blade actuators. This is demonstrated in Fig. 15 using a long-trace profiler. Driving both motors to push out the blades by P1 = P2 = 0.5 mm was shown to induce the minimum bending radius of approximately 2 km.
X2 is sagittally bent by applying equal forces on four actuators (two pairs, each pair being separated by 36 mm). From Fig. 14(a), we see that the height of the four actuators, ys, above that in the flat configuration, is composed of two parts: ys1, due to the bending of the thinned central part of width 2Ls = 7 mm, and ys2, produced by the tilting of the rigid blocks. ys1 is derived in the same manner as ym in equation (26), so that
ys2 can be simply derived by using the principle of similar triangles,
Combining (31) and (32) we obtain
The maximum travel of the sagittal actuators is therefore approximately 75 µm.
Using the same approach as above for meridional focusing, and referring to (33) and (23), leads to
Because of the larger divergence in the horizontal plane compared with that in the vertical plane, it is unnecessary to control the focal point to better than approximately 1 m. Control of q to this amount thus requires (at low photon energies at the surface station) a minimum step size in ys of 180 nm. In the present set-up, the pitch of the sagittal actuators is 250 µm and the motors have a resolution of 1000 steps per turn, sufficient for all conditions except the lowest-energy experiments at the SD station.
Both mirrors can be dynamically tilted (`pitch' motion), a requirement set by the change in incident angle with photon energy [Fig. 16(a) and equations (14)–(16)]. We specified the tilt accuracy to be equal to or better than 10 µm in the vertical position at the SD station. This imposes an accuracy of better than 80 nm for the linear positions of the 3 × 2 linear drives used to tilt the mirrors. These drives, adapted in-house from an earlier design from DESY, Hamburg, were found to have an accuracy of better than 50 nm when operated in open loop. Under normal operating conditions, the absolute positions were controlled using encoders with 50 nm resolution. A sketch of the linear drive design is shown in Fig. 16(b).
The best focal spots obtained to date using the primary optics have been 80 µm × 30 µm at the PD station and 130 µm × 40 µm at the SD station.
7.2.2. FZP microfocusing
Fresnel zone plates (FZPs) fabricated in-house by the Laboratory of Micro- and Nanotechnology can be employed to demagnify the source by approximately a factor of 50. The first FZP, made of 0.82 µm-thick gold Fresnel zones on a Si3N4 membrane (outer zone width of 100 nm and diameter of 1 mm) was tested at the SD station using 9 keV radiation. The FZP nominal focal length is 725.9 mm. The source–sample distance at the SD station is 40.087 m (see Table 4); hence, using the lensmaker's equation, we would expect the tightest focus at a distance of 739.5 mm downstream from the FZP. The experimentally determined smallest focus was found to be at this position to within 0.5 mm, had a FWHM of 1.32 µm (Fig. 17) and transmitted a of 5 × 1010 photons s−1, equating to a of 3.7 × 1016 photons s−1 mm−2, approximately 12 times higher than the best focus at SD using only the primary optics. The FWHM beam divergence using this FZP is 1.35 mrad (0.077°).
8. Endstations
8.1. Powder diffraction
8.1.1. Mythen II microstrip detector
The Mythen II microstrip detector has become the workhorse of the PD station (Bergamaschi et al., 2010) since its installation in 2007. The Mythen detector has recently undergone a significant upgrade, in which a second layer of detector modules was installed. Each layer covers an angle of 120° and consists of 24 modules. The geometry of this new set-up is shown in Fig. 18. The radii of the circles tangential to the layer surfaces are 761.5 mm (inner layer, Si-module thickness 320 µm) and 784.45 mm (outer layer, thickness 450 µm). The modules of both layers are 1280 channels wide (channel separation of 50 µm) and are repeated every 5°. The outer layer is offset relative to the inner layer by approximately 1.25°, in order to provide a gapless angular response, thereby obviating the need to move the detector to cover the small gaps between modules (0.179° and 0.320° for the inner and outer layer, respectively). The minimum readout time (i.e. dead-time when operating in the non-symmetric mode; see below) is 90 µs for 4-bit low-dynamic-range data.
There are further advantages in having the second layer. Firstly, the thicker outer modules are more sensitive to high-energy X-rays. Secondly, rate corrections for intense (e.g. Bragg-peak) signal can be avoided or at least reduced, since the data on the inner layer can be discarded, because the statistics for the signal from the outer layer are already high.
The intrinsic resolution of Mythen II in 2θ is arctan(0.05/761.5) = 0.00376°. The effective resolution, however, is determined by the sample or illumination size, such that, for example, the resolution of data acquired from a homogeneously illuminated capillary of 200 µm diameter is 0.0188° (Gozzo et al., 2004, 2010). In the large majority of cases, this is more than sufficient for PD applications, be they structural solutions, profile/full-pattern analysis or total scattering. Only in rare cases in which the sample intrinsic line broadening is close to line-profile standards is the resolution using Mythen II inadequate. In particular, Mythen II has been shown to be perfectly sufficient for full structural determination (Bruni et al., 2011). Moreover, the parallel data acquisition over large angular ranges means that (a) time-resolved studies on a millisecond timescale or shorter and (b) the complete exclusion of radiation damage, are possible with Mythen II, which would be impossible using analyzer detectors (Bergamaschi et al., 2010; Fadenberger et al., 2010; Willmott, 2011).
The upgrade of Mythen II has also included operational improvements. It can now be used in a symmetrical configuration (±60°), whereby data acquisition from one half is delayed with respect to the other, for dead-time-free measurements. When operating continuously and with the maximum −1. This scales inversely with the number of modules used and the Lastly, the on-board memory allows for storage of data up to several thousand frames per burst before data must be transferred to file.
of 24 bits (16 Mcounts), the maximum data acquisition rate for the full 120° is 20 frames s8.1.2. SAXS/WAXS facility
A new combined small-angle/wide-angle X-ray scattering (SAXS/WAXS) set-up has been designed at the MS beamline, with a view to investigating phase changes and catalytic processes on micrometre- and submicrometre-sized powder systems. Fig. 19 shows the possible configurations. WAXS data are recorded using the large Mythen II microstrip detector. A crystalline transparent window at one edge of the WAXS detector housing allows the SAXS signal to propagate with zero dead region on-axis (in reality, the dead region is determined by a beam stop directly in front of the SAXS detector). The sample–SAXS detector distance can be varied, according to the length of the flight tube in between, which consists of modular 1 m sections. The choice of the SAXS detector depends on the needs of the experiment; the three-module Mythen II detector provides better resolution in but only in one dimension, and has a smaller maximum scattering vector Qmax. Conversely, the Pilatus 2M detector is 50% larger and can therefore in principle detect features that are correspondingly smaller, although the largest objects that can be resolved are three times smaller than those studied with the Mythen modules.
The largest scattering vector Qmax and best resolution ΔQ are listed for both WAXS and SAXS at a photon energy of 10 keV in Table 7.
‡Limited by the opening angle (2θ = 0.50°) of aperture between PD and SD hutch 5.6 m downstream from the centre of the powder diffractometer (COPD). |
8.1.3. Other equipment
Several sample environments are available for users at the PD station, listed in Table 8. In addition, a robotic sample changer (Stäubli, model TX60L) allows rapid and remote sample exchange, particularly useful for experiments in which many samples need to be investigated (Fig. 20a). Most of such experiments are performed using the Mythen detector (and not the crystal analyzer), hence it is imperative that the sample capillary is mounted as coaxially as possible with the diffractometer axis, as the sample holders used by the robot system allow no adjustment of the sample orientation (Fig. 20b). A system is presently under development to simply and rapidly mount capillaries of all diameters concentrically on the magnetic cone sample holders.
|
8.2. Surface diffraction
8.2.1. Detectors
The two-dimensional photon-counting Pilatus 100k detector has been permanently installed at the SD station since 2005 (Schlepütz et al., 2005, 2011). The original graphite-crystal analyzer has been removed. In the present set-up, the sample–detector distance of 1140.8 mm and pixel size of 172 µm × 172 µm provides an angular resolution of 0.00862° (ΔQ = 7.6 × 10−4 Å at 10 keV).
At the time of writing, steps are being taken to replace the Pilatus 100k detector with the next-generation `Eiger 500k' detector (Dinapoli et al., 2010), consisting of 1024 (h) × 512 (v) pixels, each pixel being 75 µm × 75 µm in size. At the same sample–detector distance as presently used, this will provide a resolution of ΔQ = 3.3 × 10−4 Å at 10 keV. As discussed in §9.2.1, extending the sample–detector distance to 2 m will enhance the resolution to approximately 1.9 × 10−4 Å.
The `standard' Eiger module is water-cooled. Although this poses no problem for experiments in which the detector is stationary, such as in SAXS, water-cooling is difficult to implement at the surface diffractometer, as the detector module is mounted on the rotatable detector arm. The Eiger module being developed for the SD station is therefore air-cooled.
8.2.2. Environmental chambers
For robust sample surfaces, in particular those of many perovskites and related complex metal-oxides, it is only necessary to maintain an oxygen/water via a Teflon tube with a turbomolecular pump to approximately 10−2 mbar. The alternative to this is an 80 mm-diameter 25 µm-thick thermoformable polyimide dome filled with helium gas. This system has the important advantage over the Be dome that there is essentially no diffuse scattering of the incident and exit beam by the dome walls, and, because the plastic is amorphous, polycrystalline Debye–Scherrer cone signal from the dome is entirely absent.
below approximately 1 mbar in order to avoid radiation damage. For such samples, two very simple and easy-to-use systems are available. The first is a 50 mm-diameter 0.5 mm-thick beryllium dome, pumpedRecently, a new UHV cryostat has been designed and constructed for SD experiments. The lowest achievable temperature is 13 K. This replaces the previous system, which was designed for bulk single-crystal diffraction experiments and was thus far from optimal for SD experiments. The sample movements in the new design (i.e. rotation of the sample about its surface normal, and fine adjustment of the surface orientation using the hexapod) are decoupled from the vacuum vessel (which is fixed to the optical table on the surface diffractometer) using bellows and a ferrofluid rotary feedthrough (see Fig. 21a). A rod is attached to the vacuum side of this feedthrough, which is in turn connected via a ceramic spacer to the cooled sample holder, made from copper. The spacer thereby thermally isolates the sample mount from the rest of the chamber, allowing for rapid thermal equilibration and mechanically stable conditions. The sample holder is connected via copper braids to the Janis ST-400-1 UHV continuous-flow cryostat. As well as cooling, the cryostat can be heated to 500 K. All vacuum seals are UHV-compatible conflat flanges. The ultimate pressure lies in the mid 10−10 mbar range.
For experiments requiring only the best vacuum conditions, a second `mini chamber' is available, based on the design of Lee & Zegenhagen (2006). Samples can be inserted and removed by attaching a vacuum load lock onto a side port and are held using screws that are manipulated using a wobble stick (see Fig. 21b). The ultimate pressure lies in the low 10−10 mbar range.
Lastly, an in situ thin-film growth chamber using pulsed laser deposition is also available, as described in detail elsewhere (Willmott et al., 2005).
9. Initial results
9.1. Powder diffraction
Standard Si-PD patterns were recorded at 20 keV using the Mythen II detector, under comparable conditions [with regards to the capillary diameter (300 µm), filling fraction, length of capillary exposed, and vertical beam divergence], and using parallelized radiation from the W61 wiggler in 2010 and naturally divergent radiation from the U14 undulator in 2012. In the former the capillary was exposed for 160 s, while in the latter the exposure time was 40 s. The results are compared in Fig. 22. The count rate is approximately eight times larger when using undulator radiation. This corresponds very well to the theoretically expected ratio of the beam cross-sections at the powder station, assuming similar total from U14 and W61 at this energy.
9.2. Surface diffraction
A comparison of the (00l) crystal-truncation rod (CTR) of the same LaNiO3 thin-film sample using wiggler and undulator radiation at 15.5 keV is shown in Fig. 23. A similar increase in signal intensity (after background subtraction) of approximately a factor of ten was obtained. Marginal differences in the details of the CTR profiles can be explained by the footprint of the undulator radiation being approximately a factor of five smaller than that of the wiggler source, plus the fact that the LaNiO3 sample was two years older when the U14-CTR was recorded.
9.2.1. Coherent X-ray diffraction imaging
Coherent X-ray diffraction imaging (CXDI) in the Bragg geometry is an emerging quantitative technique for imaging the internal defect structure of nano- and micro-crystalline objects and materials (Robinson & Harder, 2009) and extended samples (Robinson et al., 2005; Le Bolloc'h et al., 2005). Phase retrieval and inversion of the oversampled intensity distribution around a Bragg peak (Sayre, 1952) allows one to obtain the electron density and relative displacements (i.e. strain fields) with sub-ångström resolution of the atoms from their ideal bulk-like positions (Pfeifer et al., 2006).
At present the coherence volume available at the MS beamline is of the order of ξT,h = 10 µm, ξT,v = 100 µm, ξL = 1 µm. The 2 + 3 circle surface diffractometer allows sufficient to access and map in three dimensions individual Bragg reflections. The secondary optic FZP described in §7.2.2 allows micrometre focusing.
At the end of November 2012, the first CXDI experiments were performed at the MS beamline. Two modes of operation were tested: one in which the transmitted coherent
was selected from the unfocused beam using a coherence defining slit [20 µm (h) × 100 µm (v)], the second with the FZP optics described above.In Fig. 24, the source coherence was confirmed in the diffraction pattern of a 500 nm gold crystal by the high visibility of the fringes. The example shown was collected at 9 keV using the FZP optics; 100 min total exposure yielded 1.45 × 107 photons in the diffraction pattern, with a peak intensity of 6.8 × 104 photons.
At present the detector–sample distance and pixel size determine the upper limit on the crystallite size to approximately 500 nm, while the lower limit is set by the
in the unfocused mode, 300 nm is approximately the lower limit, given the increased exposure time required to count statistically significant scatter. Below this size, focusing optics are mandatory. Implementation of the Eiger detector and an increased detector-arm length will relax the upper limit considerably to closer to two micrometres, at which point the longitudinal coherence length will become the limiting factor.In the long term, multiple Bragg reflections from the same crystal would provide the full three-dimensional strain tensor (Newton et al., 2010) and could be exploited to characterize the internal structure of devices in situ, during synthesis, and operation in a working environment.
10. Concluding remarks
The Materials Science beamline has been upgraded with an undulator X-ray source having the smallest ever reported magnet-pole periodicity of 14 mm to be routinely used in synchrotron storage rings. This allows photon energies up to nearly 40 keV to be accessible from the medium-energy storage ring of the SLS. New X-ray optics suited to the power density and beam
of the source were designed and installed. The beamline performs close to theoretical values up to approximately 25 keV, dropping to approximately 50% thereafter.The improved
has resulted in an order-of-magnitude improvement in diffraction intensity at both the PD and SD stations. The increased transverse coherence lengths of the undulator radiation has opened up the possibility of performing coherent diffraction imaging in the Bragg geometry.Footnotes
‡Present address: MAX IV Laboratory, Ole Römers väg 1, SE-223 63 Lund, Sweden.
Acknowledgements
The authors thank Professor Franz Pfeiffer, Dr Clemens Schulze-Briese and Professor Friso van der Veen for valuable discussions.
References
Bergamaschi, A., Cervellino, A., Dinapoli, R., Gozzo, F., Henrich, B., Johnson, I., Kraft, P., Mozzanica, A., Schmitt, B. & Shi, X. (2010). J. Synchrotron Rad. 17, 653–668. Web of Science CrossRef CAS IUCr Journals Google Scholar
Bruni, G., Gozzo, F., Capsoni, D., Bini, M., Macchi, P., Simoncic, P., Berbenni, V., Milanese, C., Girella, A., Ferrari, S. & Marini, A. (2011). J. Pharm. Sci. 100, 2321–2332. Web of Science CSD CrossRef CAS PubMed Google Scholar
Calvi, M., Schmidt, T., Anghel, A., Cervellino, A., Leake, S. J., Willmott, P. R. & Tanaka, T. (2013). J. Phys. Conf. Ser. 425, 032017. CrossRef Google Scholar
Chauvet, C., Polack, F., Silly, M. G., Lagarde, B., Thomasset, M., Kubsky, S., Duval, J. P., Risterucci, P., Pilette, B., Yao, I., Bergeard, N. & Sirotti, F. (2011). J. Synchrotron Rad. 18, 761–764. Web of Science CrossRef CAS IUCr Journals Google Scholar
Chubar, O., Elleaume, P. & Chavanne, J. (1998). J. Synchrotron Rad. 5, 481–484. Web of Science CrossRef CAS IUCr Journals Google Scholar
Dinapoli, R., Bergamaschi, A., Henrich, B., Horisberger, R., Johnson, I., Kraft, P., Mozzanica, A., Schmitt, B., Shi, X. & Suter, D. (2010). Nucl. Instrum. Methods Phys. Res. A, 617, 384–386. Web of Science CrossRef CAS Google Scholar
Fadenberger, K., Gunduz, I. E., Tsotsos, C., Kokonou, M., Gravini, S., Brandstetter, S., Bergamaschi, A., Schmitt, B., Mayrhofer, P. H., Doumanidis, C. C. & Rebholz, C. (2010). Appl. Phys. Lett. 97, 144101. Web of Science CrossRef Google Scholar
Gozzo, F., Cervellino, A., Leoni, M., Scardi, P., Bergamaschi, A. & Schmitt, B. (2010). Z. Kristallogr. 225, 616–624. CrossRef CAS Google Scholar
Gozzo, F., Schmitt, B., Bortolamedi, T., Giannini, C., Guagliadi, A., Lange, M., Meister, D., Maden, D., Willmott, P. & Patterson, B. D. (2004). J. Alloys Compd. 362, 206–217. Web of Science CrossRef CAS Google Scholar
Hara, T., Tanaka, T., Kitamura, H., Bizen, T., Maréchal, X. M., Seike, T., Kohda, T. & Matsuura, Y. (2004). Phys. Rev. ST Accel. Beams, 7, 050702. CrossRef Google Scholar
Le Bolloc'h, D., Ravy, S., Dumas, J., Marcus, J., Livet, F., Detlefs, C., Yakhou, F. & Paolasini, L. (2005). Phys. Rev. Lett. 95, 116401. Web of Science PubMed Google Scholar
Lee, T.-L. & Zegenhagen, J. (2006). Private communication. Google Scholar
Newton, M. C., Leake, S. J., Harder, R. & Robinson, I. K. (2010). Nat. Mater. 9, 120–124. Web of Science CrossRef CAS PubMed Google Scholar
Patterson, B. D., Abela, R., Auderset, H., Chen, Q., Fauth, F., Gozzo, F., Ingold, G., Kühne, H., Lange, M., Maden, D., Meister, D., Pattison, P., Schmidt, T., Schmitt, B., Schulze-Briese, C., Shi, M., Stampanoni, M. & Willmott, P. R. (2005). Nucl. Instrum. Methods Phys. Res. A, 540, 42–67. Web of Science CrossRef CAS Google Scholar
Pfeifer, M. A., Williams, G. J., Vartanyants, I. A., Harder, R. & Robinson, I. K. (2006). Nature (London), 442, 63–66. Web of Science CrossRef PubMed CAS Google Scholar
Robinson, I. K., Da, Y., Spila, T. & Greene, J. (2005). J. Phys. D, 38, A7–A10. CrossRef CAS Google Scholar
Robinson, I. K. & Harder, R. (2009). Nat. Mater. 8, 291–298. Web of Science CrossRef PubMed CAS Google Scholar
Rossetti, D., Lienert, U., Pradervand, C., Schneider, R., Shi, M., Zelenika, S., Rossat, M., Hignette, O., Rommeveaux, A. & Schulze-Briese, C. (2002). Proc. SPIE, 4782, 86–93. CrossRef Google Scholar
Sayre, D. (1952). Acta Cryst. 5, 843. CrossRef IUCr Journals Web of Science Google Scholar
Schlepütz, C. M., Herger, R., Willmott, P. R., Patterson, B. D., Bunk, O., Brönnimann, Ch., Henrich, B., Hülsen, G. & Eikenberry, E. F. (2005). Acta Cryst. A61, 418–425. Web of Science CrossRef CAS IUCr Journals Google Scholar
Schlepütz, C. M., Mariager, S. O., Pauli, S. A., Feidenhans'l, R. & Willmott, P. R. (2011). J. Appl. Cryst. 44, 73–83. Web of Science CrossRef IUCr Journals Google Scholar
Schulte-Schrepping, H., Materlik, G., Heuer, J. & Teichmann, T. (1995). Rev. Sci. Instrum. 66, 2217–2219. CrossRef CAS Web of Science Google Scholar
Schulze, C., Heidenreich, G., Auderset, H., Vermeulen, D. & Freund, A. K. (1998). Proc. SPIE, 3448, 156–165. Google Scholar
Schulze-Briese, C., Ketterer, B., Pradervand, C., Brönnimann, C., David, C., Horisberger, R., Puig-Molina, A. & Graafsma, H. (2001). Nucl. Instrum. Methods Phys. Res. A, 467, 230–234. Google Scholar
Slack, G. A. & Bartram, S. F. (1975). J. Appl. Phys. 46, 89–98. CrossRef CAS Web of Science Google Scholar
Stampanoni, M., Groso, A., Isenegger, A., Mikuljan, G., Chen, Q., Meister, D., Lange, M., Betemps, R., Henein, S. & Abela, R. (2007). AIP Conf. Proc. 879, 848–851. CrossRef CAS Google Scholar
Willmott, P. R. (2011). An Introduction to Synchrotron Radiation: Techniques and Applications. New York: John Wiley and Sons. Google Scholar
Willmott, P. R., Herger, R., Schlepütz, C. M., Martoccia, D. & Patterson, B. D. (2005). Synchrotron Radiat. News, 18, 37–42. CrossRef Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.