Time-resolved pair distribution function analysis of disordered materials on beamlines BL04B2 and BL08W at SPring-8

CROSSMARK_Color_square_no_text.svg

aResearch and Utilization Division, Japan Synchrotron Radiation Research Institute, 1-1-1 Kouto, Sayo, Hyogo 679-5198, Japan, bSynchrotron X-ray Station at SPring-8, National Institute for Materials Science (NIMS), 1-1-1 Kouto, Sayo, Hyogo 679-5148, Japan, cInternational Center for Materials Nanoarchitectonics (WPI-MANA), National Institute for Materials Science (NIMS), 1-1 Namiki, Tsukuba, Ibaraki 305-0044, Japan, dDepartment of Chemical System Engineering, University of Tokyo, 7-3-1 Hongo, Bunkyo-ku, Tokyo 113-8656, Japan, eDepartment of Interdisciplinary Environment, Graduate School of Human and Environmental Studies, Kyoto University, Yoshida-nihonmatsu, Sakyo, Kyoto 606-8501, Japan, and fAdvanced Technology Research Laboratories, Idemitsu Kosan Co. Ltd, 1280 Kamiizumi, Sodegaura, Chiba 299-0293, Japan
*Correspondence e-mail: ohara@spring8.or.jp, futoshi.utsuno@idemitsu.com

Edited by I. Lindau, SLAC/Stanford University, USA (Received 25 April 2018; accepted 6 August 2018; online 26 September 2018)

A dedicated apparatus has been developed for studying structural changes in amorphous and disordered crystalline materials substantially in real time. The apparatus, which can be set up on beamlines BL04B2 and BL08W at SPring-8, mainly consists of a large two-dimensional flat-panel detector and high-energy X-rays, enabling total scattering measurements to be carried out for time-resolved pair distribution function (PDF) analysis in the temperature range from room temperature to 873 K at pressures of up to 20 bar. For successful time-resolved analysis, a newly developed program was used that can monitor and process two-dimensional image data simultaneously with the data collection. The use of time-resolved hardware and software is of great importance for obtaining a detailed understanding of the structural changes in disordered materials, as exemplified by the results of commissioned measurements carried out on both beamlines. Benchmark results obtained using amorphous silica and demonstration results for the observation of sulfide glass crystallization upon annealing are introduced.

1. Introduction

Pair distribution function (PDF) analysis of total scattering data covering a wide range of momentum transfer (Q), particularly up to a sufficiently large momentum transfer measured using high-energy X-rays available from synchrotron radiation, is widely utilized in the structural study of amorphous materials and disordered crystalline materials (Billinge & Levin, 2007[Billinge, S. J. L. & Levin, I. (2007). Science, 316, 561-565.]; Young & Goodwin, 2011[Young, C. A. & Goodwin, A. L. (2011). J. Mater. Chem. 21, 6464-6476.]; Keen & Goodwin, 2015[Keen, D. A. & Goodwin, A. L. (2015). Nature (London), 521, 303-309.]). From total scattering measurements, quantitative information on atomic arrangements, namely the relationship between interatomic distances, r, and atomic pair densities, i.e. the PDF, g(r), can be calculated from the experimentally accessible function of the reduced PDF, G(r), as follows:

[g(r) = {{G(r)} \over {4\pi r \rho }} + 1, \eqno(1)]

where ρ is the average number density of atoms. G(r) is calculated as follows by Fourier transformation of the total structure factor, F(Q), which is a normalized form of the total scattering data and represents the oscillation associated with the density distribution in a material:

[\eqalignno{G(r) = & \, {2 \over \pi} \int \limits_{Q_{\min}}^{Q_{\max}} F(Q) \sin (Qr) \, {\rm d}Q \cr = & \, {2 \over \pi} \int \limits_{Q_{\min}}^{Q_{\max}} Q[S(Q) - 1] \sin (Qr) \, {\rm d}Q. &(2)}]

Because this Fourier transform results in an r spacing of π/Qmax, S(Q) data measured up to a higher Q region result in PDFs with a higher real-space resolution. For example, high-energy X-rays of over 40 keV (or a shorter photon wavelength of λ < 0.31 Å) enable us to obtain high momentum transfers, Q > 20 Å−1, which also depend on the scattering angle, 2θ (Qλ = 4πsinθ).

Because the physical meaning of G(r) might not be straightforward, a total correlation function, T(r), is often used to discuss the structural information (Keen, 2001[Keen, D. (2001). J. Appl. Cryst. 34, 172-177.]),

[\eqalignno{T(r) = & \, G(r) + 4\pi r \rho \cr = & \, 4\pi r \rho \, g(r). & (3)}]

A peak in T(r) indicates the existence of atoms with a density exceeding the average number density at a distance r, whereas a valley suggests the absence of atoms. PDF analysis with high-energy X-rays has been reviewed in detail by Benmore (2012[Benmore, C. J. (2012). ISRN Mater. Sci. 2012, 852905.]).

Beamline BL04B2 at SPring-8 was designed for high-energy X-ray diffraction measurements to perform high-resolution X-ray PDF analysis (Isshiki et al., 2001[Isshiki, M., Ohishi, Y., Goto, S., Takeshita, K. & Ishikawa, T. (2001). Nucl. Instrum. Methods Phys. Res. A, 467-468, 663-666.]; Kohara et al., 2007[Kohara, S., Itou, M., Suzuya, K., Inamura, Y., Sakurai, Y., Ohishi, Y. & Takata, M. (2007). J. Phys. Condens. Matter, 19, 506101.]), contributing to scientific research on the structural study of disordered materials (Matsunaga et al., 2011[Matsunaga, T., Akola, J., Kohara, S., Honma, T., Kobayashi, K., Ikenaga, E., Jones, R. O., Yamada, N., Takata, M. & Kojima, R. (2011). Nat. Mater. 10, 129-134.]; Ohara et al., 2012[Ohara, K., Temleitner, L., Sugimoto, K., Kohara, S., Matsunaga, T., Pusztai, L., Itou, M., Ohsumi, H., Kojima, R., Yamada, N., Usuki, T., Fujiwara, A. & Takata, M. (2012). Adv. Funct. Mater. 22, 2251-2257.]; Akola et al., 2013[Akola, J., Kohara, S., Ohara, K., Fujiwara, A., Watanabe, Y., Masuno, A., Usuki, T., Kubo, T., Nakahira, A., Nitta, K., Uruga, T., Weber, J. K. R. & Benmore, C. J. (2013). Proc. Natl Acad. Sci. USA, 110, 10129-10134.]; Kohara et al., 2014[Kohara, S., Akola, J., Patrikeev, L., Ropo, M., Ohara, K., Itou, M., Fujiwara, A., Yahiro, J., Okada, J. T., Ishikawa, T., Mizuno, A., Masuno, A., Watanabe, Y. & Usuki, T. (2014). Nat. Commun. 5, 5892.]; Liu et al., 2016[Liu, Z., Okabe, K., Anand, C., Yonezawa, Y., Zhu, J., Yamada, H., Endo, A., Yanaba, Y., Yoshikawa, T., Ohara, K., Okubo, T. & Wakihara, T. (2016). Proc. Natl Acad. Sci. USA, 113, 14267-14271.]; Umeda et al., 2017[Umeda, T., Yamada, H., Ohara, K., Yoshida, K., Sasaki, Y., Takano, M., Inagaki, S., Kubota, Y., Takewaki, T., Okubo, T. & Wakihara, T. (2017). J. Phys. Chem. C, 121, 24324-24334.]; Onodera et al., 2017[Onodera, Y., Kohara, S., Masai, H., Koreeda, A., Okamura, S. & Ohkubo, T. (2017). Nat. Commun. 8, 15449.]). One measurement for PDF analysis on BL04B2 takes at least 2–3 h, owing to the necessity of scanning a point detector with an energy window such as a Ge-SDD to eliminate the higher harmonic contamination of X-rays. This long measurement due to the use of scanning point detectors has been typical at SPring-8, hampering observations of rapid structural changes in dis­ordered materials. For example, observations of structural changes upon the growth or ageing of disordered materials, which may occur within a few minutes, are required to understand the roles of such processes. Understanding the structural relaxation mechanism in amorphous materials, including organic materials such as pharmaceutical molecules (Tominaka et al., 2017[Tominaka, S., Kawakami, K., Fukushima, M. & Miyazaki, A. (2017). Mol. Pharm. 14, 264-273.]), and the crystallization mechanism of glasses during annealing treatment are of great interest. Nowadays, such time-resolved PDF analysis is being utilized at other synchrotron facilities, such as the APS (Chupas et al., 2003[Chupas, P. J., Qiu, X., Hanson, J. C., Lee, P. L., Grey, C. P. & Billinge, S. J. L. (2003). J. Appl. Cryst. 36, 1342-1347.], 2004[Chupas, P. J., Chaudhuri, S., Hanson, J. C., Qiu, X., Lee, P. L., Shastri, S. D., Billinge, S. J. L. & Grey, C. P. (2004). J. Am. Chem. Soc. 126, 4756-4757.], 2007[Chupas, P. J., Chapman, K. W. & Lee, P. L. (2007). J. Appl. Cryst. 40, 463-470.]; Kramer, 2007[Kramer, M. (2007). J. Appl. Cryst. 40, 77-86.]; Jensen et al., 2012[Jensen, K. M. Ø., Christensen, M., Juhas, P., Tyrsted, C., Bøjesen, E. D., Lock, N., Billinge, S. J. L. & Iversen, B. B. (2012). J. Am. Chem. Soc. 134, 6785-6792.]), the ESRF (Abdala et al., 2014[Abdala, P. M., Mauroy, H. & van Beek, W. (2014). J. Appl. Cryst. 47, 449-457.]; Terban et al., 2015[Terban, M. W., Johnson, M., Di Michiel, M. & Billinge, S. J. L. (2015). Nanoscale, 7, 5480-5487.]) and the DLS (Keen et al., 2018[Keen, D. A., Keeble, D. S. & Bennett, T. D. (2018). Phys. Chem. Miner. 45, 333-342.]), using two-dimensional amorphous silicon (a-Si) digital X-ray flat-panel detectors (FPDs). This is because FPDs possess characteristics suitable for time-resolved PDF analysis, such as high efficiency for the detection of high-energy X-rays, a high dynamic range and contrast, and a high throughput with a frame rate of 15 frames s−1.

To achieve such time-resolved PDF analysis at SPring-8, we have developed a new apparatus for total scattering measurements with a variable camera length using a 16-inch two-dimensional FPD (XRD1621 CN3), which we tested on beamlines BL04B2 and BL08W. The sample environment is controlled using a variety of sample stages. For example, (i) solvothermal/hydrothermal conditions are realized using a microwave reactor (Initiator from Biotage) in the temperature range from room temperature to 573 K at a pressure of less than 20 bar (1 bar = 100 000 Pa; Tominaka et al., 2018[Tominaka, S., Hiroki, Y., Satoshi, H., Saori , I. K. & Koji, O. (2018). ACS Omega, 3, 8874-8881.]), and (ii) high-temperature conditions up to 873 K are realized by a heating cell. In addition to being designed for in situ measurements, the apparatus used for time-resolved PDF analysis is compatible with the existing facilities on BL04B2, such as an aerodynamic levitation furnace, an acoustic levitation furnace and a diamond anvil cell. In this paper, we present the design and experimental verification of the new apparatus for time-resolved PDF analysis.

2. PDF analysis on high-energy X-ray total scattering beamlines at SPring-8

Beamlines BL04B2 and BL08W have been widely used in structural studies of disordered materials over the past 20 years (Ohno et al., 2001[Ohno, H., Kohara, S., Umesaki, N. & Suzuya, K. (2001). J. Non-Cryst. Solids, 293-295, 125-135.]; Kohara et al., 2004[Kohara, S., Suzuya, K., Takeuchi, K., Loong, C. K., Grimsditch, M., Weber, J. K. R., Tangeman, J. A. & Key, T. S. (2004). Science, 303, 1649-1652.], 2014[Kohara, S., Akola, J., Patrikeev, L., Ropo, M., Ohara, K., Itou, M., Fujiwara, A., Yahiro, J., Okada, J. T., Ishikawa, T., Mizuno, A., Masuno, A., Watanabe, Y. & Usuki, T. (2014). Nat. Commun. 5, 5892.]; Inui et al., 2007[Inui, M., Matsuda, K., Ishikawa, D., Tamura, K. & Ohishi, Y. (2007). Phys. Rev. Lett. 98, 185504.]; Matsunaga et al., 2011[Matsunaga, T., Akola, J., Kohara, S., Honma, T., Kobayashi, K., Ikenaga, E., Jones, R. O., Yamada, N., Takata, M. & Kojima, R. (2011). Nat. Mater. 10, 129-134.]; Li et al., 2018[Li, B., Wang, H., Kawakita, Y., Zhang, Q., Feygenson, M., Yu, H. L., Wu, D., Ohara, K., Kikuchi, T., Shibata, K., Yamada, T., Ning, X. K., Chen, Y., He, J. Q., Vaknin, D., Wu, R. Q., Nakajima, K. & Kanatzidis, M. G. (2018). Nat. Mater. 17, 226-230.]). Total scattering measurement (or PDF analysis) using high-energy X-rays is useful for quantitatively determining the local structure of disordered materials at low scattering angles with a wide Q range. BL04B2 is equipped with two Si crystals as a monochromator, which provides fixed-energy X-rays of 37.7 keV from Si(111), 61.4 keV from Si(220) and 113.1 keV from Si(111) (third-harmonic generation). The energy of 61.4 keV is mainly used in the PDF analysis. On the other hand, BL08W can provide fixed-energy X-rays of 115 keV using Si(440). The energy resolutions ΔE/E of BL04B2 and BL08W are approximately 5 × 10−3 and 1 × 10−3, respectively. The beam dimensions at the sample position are 1.0 mm (vertical) × 1.0 mm (horizontal). The photon flux on BL08W is approximately 100 times higher than that on BL04B2, at 1.0 × 1013 photons s−1. Further details of these beamlines are available in the literature (Kohara et al., 2007[Kohara, S., Itou, M., Suzuya, K., Inamura, Y., Sakurai, Y., Ohishi, Y. & Takata, M. (2007). J. Phys. Condens. Matter, 19, 506101.]; Marechal et al., 1998[Maréchal, X.-M., Hara, T., Tanabe, T., Tanaka, T. & Kitamura, H. (1998). J. Synchrotron Rad. 5, 431-433.]). On BL04B2, a total scattering apparatus consisting of point detectors installed along the flight path to eliminate the harmonic contamination of X-rays is continuously available. This system is mainly used for ex situ study of disordered materials. On the other hand, on BL08W, a temporary apparatus has been installed to enable the use of X-rays with higher energy or flux to obtain total scattering data from samples containing a heavy element (Yoneda et al., 2013[Yoneda, Y., Kohara, S., Ito, M., Abe, H., Takeuchi, M., Uchida, H. & Matsumura, Y. (2013). Trans. Mater. Res. Soc. Jpn, 38, 109-112.]; Kohara et al., 2014[Kohara, S., Akola, J., Patrikeev, L., Ropo, M., Ohara, K., Itou, M., Fujiwara, A., Yahiro, J., Okada, J. T., Ishikawa, T., Mizuno, A., Masuno, A., Watanabe, Y. & Usuki, T. (2014). Nat. Commun. 5, 5892.]) or from samples in an unstable environment subjected to aerodynamic (Kohara et al., 2014[Kohara, S., Akola, J., Patrikeev, L., Ropo, M., Ohara, K., Itou, M., Fujiwara, A., Yahiro, J., Okada, J. T., Ishikawa, T., Mizuno, A., Masuno, A., Watanabe, Y. & Usuki, T. (2014). Nat. Commun. 5, 5892.]) or acoustic levitation.

We have installed a large flat-panel area detector on both beamlines, as illustrated in Fig. 1[link]. The X-ray diffraction measurements on these beamlines have different features, particularly for the photon energy and flux as described above, which in turn result in different resolutions in real/reciprocal space, as summarized in Table 1[link]. For example, the camera length can be changed from 372 mm up to 866 mm on BL08W, which corresponds to Q = 50–25 Å−1 (dQ = 0.04–0.02 Å−1). This difference in the specifications does not result in an obvious difference in the PDF data, at least in the standard data for amorphous silica, as shown in Fig. 2[link]. The figure shows that the oscillations in F(Q) do not decay up to the high-Q region, as is always the case with Si—O covalent bonds. Thus, it is suitable for verifying measurements. Regarding F(Q), compared with the data obtained using six point detectors on BL04B2, the total scattering data collected using the FPD are identical, at least in the observed range of Q (up to 25 Å−1).

Table 1
Comparison of two beamlines for high-energy X-ray diffraction at SPring-8

Beamline BL04B2 BL08W
Energy (keV) 61.4 115.6
Energy resolution 5 × 10−3 1.5 × 10−3
Photon flux (photons s−1 mm−2) 9.1 × 1010 1.0 × 1013
Qmax, dQ−1) (camera length 300 mm) 25, 0.02 50, 0.04
Qmax, dQ−1) (camera length 800 mm) 12, 0.01 25, 0.02
[Figure 1]
Figure 1
Photographs and an illustration of the experimental setups of the flat-panel area-detector apparatus for the analysis of disordered materials. (a) High-resolution mode on BL04B2 with an acoustic levitator. (b) High-Q measurement mode on BL08W with a furnace. (c) A schematic drawing of the apparatus used for in situ PDF analysis.
[Figure 2]
Figure 2
(a) Total structure factors F(Q) and (b) total correlation function T(r) of amorphous silica obtained on BL04B2 and BL08W using the FPD (1 min), compared with those obtained on BL04B2 using point detectors (3 h). Fourier transforms were carried out for the data for a constant Qmax = 26 Å−1 using the Lorch function (Lorch, 1969[Lorch, E. (1969). J. Phys. C Solid State Phys. 2, 229-237.]). A fit to the first Si—O peak in the X-ray correlation function (Mozzi & Warren, 1969[Mozzi, R. L. & Warren, B. E. (1969). J. Appl. Cryst. 2, 164-172.]) is also shown by the red dashed line.

The total correlation functions, T(r), for the amorphous silica were obtained from F(Q) at intervals of 0.05 Å−1 in the Q range up to 25 Å−1 and are shown in Fig. 2[link](b). The data obtained using the FPD are similar, have a sufficient real-space resolution and exhibit very small unphysical peaks in the range r < 1 Å. Moreover, they are in good agreement with the data obtained on BL04B2 using point detectors, for which a 180 times longer exposure time is required to obtain sufficient statistics. A fit to the first peak in the amorphous silica (Mozzi & Warren, 1969[Mozzi, R. L. & Warren, B. E. (1969). J. Appl. Cryst. 2, 164-172.]) is shown by the dashed line in Fig. 2[link](b). The fit parameter is an Si—O distance of 1.62 ± 0.01 Å with a coordination number of 3.89 ± 0.02, which is consistent with the earlier result (Wright, 1990[Wright, A. C. (1990). J. Non-Cryst. Solids, 123, 129-148.]). Thus, we have confirmed that an FPD with a short exposure time of 1 min provides sufficient data for PDF analysis. Furthermore, an exposure time of 10 s still provides sufficient data for the PDF analysis of amorphous bulk silica, as shown in Fig. 3[link]. Thus, in situ PDF analysis within 10 s is expected to extend the study of structural changes in disordered materials at SPring-8 by reducing the measurement time from hours to seconds.

[Figure 3]
Figure 3
Total structure factors F(Q) of amorphous silica on BL04B2. The data were collected with integration periods of 0.5, 10, 30 and 300 s.

3. Processing of two-dimensional diffraction data (PIXIA)

Because time-resolved PDF measurements provide many two-dimensional diffraction images, rapid and accurate processing of this huge quantity of information is necessary to extract information on structural changes. Thus, we use the PIXIA program for data conversion from two-dimensional images into one-dimensional total scattering patterns. PIXIA stands for pixel-based image analyser, and it is dedicated to such data conversion for total scattering data. Similar to other programs such as Fit2D (Hammersley, 1997[Hammersley, A. P. (1997). ESRF Internal Report ESRF97HA02T. ESRF, Grenoble, France.]) used for diffractometry, PIXIA uses equations applicable over a wide Q range suitable for PDF studies, and moreover adopts noise filtering based on an image-processing algorithm. Roughly speaking, PIXIA deals with all the pixels of a two-dimensional detector as an array of point detectors to determine a total scattering pattern that corresponds to millions of point detectors (cf. 2000 × 2000 pixels). In contrast with conventional programs that perform simple integrations to increase the signal-to-noise ratio, PIXIA adopts a mathematical noise-filtration algorithm in different data dimensions.

The PIXIA program was written in Python and uses the SciPy and NumPy packages, in which all the mathematical treatments are dealt with in matrix calculations. This program is a part of the Orochi program suite, which is an ensemble of Python-based programs for X-ray diffraction studies including image processing, PDF conversion and data fitting. In this study, we used PIXIA (β version 2.x), but the latest version, β version 3.x, can work with the MaterialsPDF PDF conversion program in the Orochi suite, and can monitor PDFs in real time during experiments. Such a demonstration will be reported in the future. The program will be available free of charge and is currently under testing as a β version available from the website of the National Institute for Materials Science (Tominaka, https://samurai.nims.go.jp/profiles/tominaka_satoshi).

The Orochi program suite has a graphical user interface written in wxPython and data are illustrated using Matplotlib. Examples of plots of data treated using PIXIA are shown in Fig. 4[link]: panel (a) illustrates the raw two-dimensional image data; panel (b) illustrates the time dependence of the total scattering data as an I(θ, t) plot; panel (c) compares initial and latest total scattering patterns; panel (d) illustrates the latest relative PDF calculated based on the initial PDF; and panel (e) illustrates the time dependence of the relative PDFs. These data were analysed and plotted within a short time simultaneously with the measurements, i.e. substantially in real time. This enables us to discuss the data in real time and optimize the measurements taken in a period of beam time.

[Figure 4]
Figure 4
The main graphical user interface of the data-analysis software (PIXIA), showing results for the Li7P3S11 sulfide glass. (a) The raw two-dimensional image data. (b) The time dependence of the total scattering data as an I(θ, t) plot. (c) A comparison of initial and latest total scattering patterns. (d) The latest relative PDF calculated based on the initial PDF. (e) The time dependence of the relative PDFs.

4. Observation of crystallization of sulfide glass during annealing

The sulfide glass Li7P3S11 is a very important material for all-solid-state Li-ion batteries, in which it is used as a solid electrolyte. Its Li ionic conductivity increases upon thermal crystallization (Mizuno et al., 2005[Mizuno, F., Hayashi, A., Tadanaga, K. & Tatsumisago, M. (2005). Adv. Mater. 17, 918-921.]) and, intriguingly, the conductivity depends on the annealing conditions such as the annealing rate. Thus, we investigated its structural changes upon thermal crystallization at two different annealing rates (2 and 10 K min−1). The time-dependent data obtained at the annealing rate of 2 K min−1 are shown in Fig. 4[link]. As can be seen in Fig. 4[link](e), the relative PDF indicates a clear structural change after 6000 s, which corresponds to a sample temperature of 523 K (Fig. S1 in the supporting information). This is the typical crystallization temperature of the sulfide.

To determine the effect of such annealing on the structure of the sulfide glass at the atomic and molecular levels, time-resolved PDFs, collected in the temperature range from 473 to 573 K at 5 K intervals as shown in Fig. 5[link], were analysed. The first peak around 2.0 Å is assigned to the P—S bonds in the PS4 tetrahedral anions. This peak does not depend on the annealing treatment, meaning that the PS4 tetrahedral anions remain in the structure regardless of the crystallization temperature. On the other hand, the relative intensities of the peaks found at longer r (3.4, 4.0, 6.8, 9.6, 10.6, 12.5, 15.5, 16.2 and 19.6 Å) gradually change upon crystallization, as shown by dashed lines in Fig. 5[link].

[Figure 5]
Figure 5
Reduced PDFs G(r) of Li7P3S11 used to investigate its thermal crystallization. The data were collected in the temperature range from 473 to 573 K at annealing rates of 2 and 10 K min−1.

Although different structural changes were expected in the samples annealed at different rates, no clear difference was observed, except for the crystallization temperature: the structural changes commenced at a lower temperature in the sulfide annealed at 2 K min−1 than in the sample annealed at 10 K min−1. This difference is clear in PDFs obtained at longer r, as can be seen from the peaks at r = 15.5, 16.2 and 19.6 Å. Moreover, a peak around 6 Å, which is attributed to the formation of Li4P2S6 (Sadowski et al., 2018[Sadowski, M., Sicolo, S. & Albe, K. (2018). Solid State Ion. 319, 53-60.]), is observed at 546 K only in the case of annealing at 2 K min−1. Such decomposition of Li7P3S11 into Li4P2S6 was also reported by Busche et al. (2016[Busche, M. R., Weber, D. A., Schneider, Y., Dietrich, C., Wenzel, S., Leichtweiss, T., Schröder, D., Zhang, W., Weigand, H., Walter, D., Sedlmaier, S. J., Houtarde, D., Nazar, L. F. & Janek, J. (2016). Chem. Mater. 28, 6152-6165.]) and Minami et al. (2010[Minami, K., Hayashi, A. & Tatsumisago, M. (2010). J. Ceram. Soc. Jpn, 118, 305-308.]). ΔG(r, Δt) was calculated as the difference from G(r) at 373 K in the r range of 15–17 Å (Fig. 6[link]a). To discuss the kinetics, we plot the relationship between dG/dt and temperature in Fig. 6[link](b). dG/dt was calculated by differentiating ΔG(r, Δt). As can be seen in this figure, the crystallization of Li7P3S11 was facilitated at around 546 K in the 10 K min−1 case. This suggests that crystallization proceeds through the formation of sulfur-deficient fragments of P2S6 anions as an intermediate between the amorphous phase and the crystalline phase, both of which consist of PS4 tetrahedral anions. Further detailed investigations will be reported in the future.

[Figure 6]
Figure 6
Rate law of structural changes in Li7P3S11 in the temperature range from 473 to 573 K at annealing rates of 2 and 10 K min−1. (a) ΔG(r, Δt) was calculated as the difference from G(r) at 373 K in the r range of 15–17 Å. (b) dG/dt was calculated by differentiating ΔG(r, Δt).

These results demonstrate that the new apparatus for time-resolved PDF analysis on BL08W and BL04B2 can collect PDF data for amorphous and disordered crystalline materials in real time. The acquisition times of 10–60 s per frame enable the investigation of structural changes in disordered materials in real time with high accuracy.

5. Conclusions

The dedicated setups on beamlines BL04B2 and BL08W, in which a large two-dimensional flat-panel detector is used for time-resolved PDF analysis, have been described in this paper. The data for amorphous silica collected on the two-dimensional detector in a few seconds were consistent with the data collected on a point detector scanned for 2–3 h in the existing PDF apparatus on BL04B2. In addition, the apparatus used for time-resolved PDF analysis is compatible with existing instruments for different sample environments, such as an aerodynamic levitation furnace, an acoustic levitation furnace and a diamond anvil cell. To obtain a better understanding of the annealing behaviour of Li7P3S11 sulfide glass, time-resolved PDFs were observed at intervals of 5 K. The sulfide glass annealed at 10 K min−1 crystallized faster than that annealed at 2 K min−1. This might be associated with the formation of sulfur defects, which were observed only in the crystallization at 2 K min−1. Further detailed investigations will be reported in the future.

It has been demonstrated that the new apparatus for time-resolved PDF analysis on BL08W and BL04B2 can collect PDF data for amorphous and disordered crystalline materials in real time. We believe that many structural studies on disordered materials by time-resolved PDF analysis will be performed at SPring-8 using the new apparatus.

Supporting information


Acknowledgements

This development of the time-resolved pair distribution function analysis using the flat-panel area detector was supported by RIKEN, Japan. We wish to thank the Information-Technology Promotion Division of JASRI for experimental assistance. In particular, an information-processing control system (Matsumoto et al., 2018[Matsumoto, T., Abe, T., Furukawa, Y., Masunaga, H., Matsushita, T. & Nakada, K. (2018). Proceedings of the 16th International Conference on Accelerator and Large Experimental Control Systems (ICALEPCS2017), 8-13 October 2017, Barcelona, Spain, pp. 1281-1285. THMPL07.]) contributed greatly to this work. Moreover, discussions with Shinji Kohara, Yasuo Ohishi, Shigeru Kimura, Tomoya Uruga and Yoshiharu Sakurai are appreciated.

Funding information

This work was supported by JSPS KAKENHI Grants-in-Aid for Scientific Research (A, B and C) and for Young Scientists (B) (grant Nos. 16H0390311, 17H0103411, 18K05192 and 17K18444). The synchrotron radiation experiment was approved by the Japan Synchrotron Radiation Research Institute (proposal Nos. 2016A1841, 2017A1217, 2017A1855 and 2017B1847).

References

First citationAbdala, P. M., Mauroy, H. & van Beek, W. (2014). J. Appl. Cryst. 47, 449–457.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationAkola, J., Kohara, S., Ohara, K., Fujiwara, A., Watanabe, Y., Masuno, A., Usuki, T., Kubo, T., Nakahira, A., Nitta, K., Uruga, T., Weber, J. K. R. & Benmore, C. J. (2013). Proc. Natl Acad. Sci. USA, 110, 10129–10134.  Web of Science CrossRef PubMed Google Scholar
First citationBenmore, C. J. (2012). ISRN Mater. Sci. 2012, 852905.  Google Scholar
First citationBillinge, S. J. L. & Levin, I. (2007). Science, 316, 561–565.  Web of Science CrossRef PubMed CAS Google Scholar
First citationBusche, M. R., Weber, D. A., Schneider, Y., Dietrich, C., Wenzel, S., Leichtweiss, T., Schröder, D., Zhang, W., Weigand, H., Walter, D., Sedlmaier, S. J., Houtarde, D., Nazar, L. F. & Janek, J. (2016). Chem. Mater. 28, 6152–6165.  Web of Science CrossRef Google Scholar
First citationChupas, P. J., Chapman, K. W. & Lee, P. L. (2007). J. Appl. Cryst. 40, 463–470.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationChupas, P. J., Chaudhuri, S., Hanson, J. C., Qiu, X., Lee, P. L., Shastri, S. D., Billinge, S. J. L. & Grey, C. P. (2004). J. Am. Chem. Soc. 126, 4756–4757.  Web of Science CrossRef PubMed CAS Google Scholar
First citationChupas, P. J., Qiu, X., Hanson, J. C., Lee, P. L., Grey, C. P. & Billinge, S. J. L. (2003). J. Appl. Cryst. 36, 1342–1347.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationHammersley, A. P. (1997). ESRF Internal Report ESRF97HA02T. ESRF, Grenoble, France.  Google Scholar
First citationInui, M., Matsuda, K., Ishikawa, D., Tamura, K. & Ohishi, Y. (2007). Phys. Rev. Lett. 98, 185504.  Web of Science CrossRef PubMed Google Scholar
First citationIsshiki, M., Ohishi, Y., Goto, S., Takeshita, K. & Ishikawa, T. (2001). Nucl. Instrum. Methods Phys. Res. A, 467–468, 663–666.  Web of Science CrossRef Google Scholar
First citationJensen, K. M. Ø., Christensen, M., Juhas, P., Tyrsted, C., Bøjesen, E. D., Lock, N., Billinge, S. J. L. & Iversen, B. B. (2012). J. Am. Chem. Soc. 134, 6785–6792.  Web of Science CrossRef CAS PubMed Google Scholar
First citationKeen, D. (2001). J. Appl. Cryst. 34, 172–177.  Web of Science CrossRef IUCr Journals Google Scholar
First citationKeen, D. A. & Goodwin, A. L. (2015). Nature (London), 521, 303–309.  Web of Science CrossRef CAS PubMed Google Scholar
First citationKeen, D. A., Keeble, D. S. & Bennett, T. D. (2018). Phys. Chem. Miner. 45, 333–342.  Web of Science CrossRef Google Scholar
First citationKohara, S., Akola, J., Patrikeev, L., Ropo, M., Ohara, K., Itou, M., Fujiwara, A., Yahiro, J., Okada, J. T., Ishikawa, T., Mizuno, A., Masuno, A., Watanabe, Y. & Usuki, T. (2014). Nat. Commun. 5, 5892.  Web of Science CrossRef PubMed Google Scholar
First citationKohara, S., Itou, M., Suzuya, K., Inamura, Y., Sakurai, Y., Ohishi, Y. & Takata, M. (2007). J. Phys. Condens. Matter, 19, 506101.  Web of Science CrossRef Google Scholar
First citationKohara, S., Suzuya, K., Takeuchi, K., Loong, C. K., Grimsditch, M., Weber, J. K. R., Tangeman, J. A. & Key, T. S. (2004). Science, 303, 1649–1652.  Web of Science CrossRef PubMed Google Scholar
First citationKramer, M. (2007). J. Appl. Cryst. 40, 77–86.  Web of Science CrossRef IUCr Journals Google Scholar
First citationLi, B., Wang, H., Kawakita, Y., Zhang, Q., Feygenson, M., Yu, H. L., Wu, D., Ohara, K., Kikuchi, T., Shibata, K., Yamada, T., Ning, X. K., Chen, Y., He, J. Q., Vaknin, D., Wu, R. Q., Nakajima, K. & Kanatzidis, M. G. (2018). Nat. Mater. 17, 226–230.  Web of Science CrossRef PubMed Google Scholar
First citationLiu, Z., Okabe, K., Anand, C., Yonezawa, Y., Zhu, J., Yamada, H., Endo, A., Yanaba, Y., Yoshikawa, T., Ohara, K., Okubo, T. & Wakihara, T. (2016). Proc. Natl Acad. Sci. USA, 113, 14267–14271.  Web of Science CrossRef PubMed Google Scholar
First citationLorch, E. (1969). J. Phys. C Solid State Phys. 2, 229–237.  CrossRef Web of Science Google Scholar
First citationMaréchal, X.-M., Hara, T., Tanabe, T., Tanaka, T. & Kitamura, H. (1998). J. Synchrotron Rad. 5, 431–433.  Web of Science CrossRef IUCr Journals Google Scholar
First citationMatsumoto, T., Abe, T., Furukawa, Y., Masunaga, H., Matsushita, T. & Nakada, K. (2018). Proceedings of the 16th International Conference on Accelerator and Large Experimental Control Systems (ICALEPCS2017), 8–13 October 2017, Barcelona, Spain, pp. 1281–1285. THMPL07.  Google Scholar
First citationMatsunaga, T., Akola, J., Kohara, S., Honma, T., Kobayashi, K., Ikenaga, E., Jones, R. O., Yamada, N., Takata, M. & Kojima, R. (2011). Nat. Mater. 10, 129–134.  Web of Science CrossRef CAS PubMed Google Scholar
First citationMinami, K., Hayashi, A. & Tatsumisago, M. (2010). J. Ceram. Soc. Jpn, 118, 305–308.  Web of Science CrossRef Google Scholar
First citationMizuno, F., Hayashi, A., Tadanaga, K. & Tatsumisago, M. (2005). Adv. Mater. 17, 918–921.  Web of Science CrossRef Google Scholar
First citationMozzi, R. L. & Warren, B. E. (1969). J. Appl. Cryst. 2, 164–172.  CrossRef CAS IUCr Journals Web of Science Google Scholar
First citationOhara, K., Temleitner, L., Sugimoto, K., Kohara, S., Matsunaga, T., Pusztai, L., Itou, M., Ohsumi, H., Kojima, R., Yamada, N., Usuki, T., Fujiwara, A. & Takata, M. (2012). Adv. Funct. Mater. 22, 2251–2257.  Web of Science CrossRef Google Scholar
First citationOhno, H., Kohara, S., Umesaki, N. & Suzuya, K. (2001). J. Non-Cryst. Solids, 293–295, 125–135.  Web of Science CrossRef Google Scholar
First citationOnodera, Y., Kohara, S., Masai, H., Koreeda, A., Okamura, S. & Ohkubo, T. (2017). Nat. Commun. 8, 15449.  Web of Science CrossRef PubMed Google Scholar
First citationSadowski, M., Sicolo, S. & Albe, K. (2018). Solid State Ion. 319, 53–60.  Web of Science CrossRef Google Scholar
First citationTerban, M. W., Johnson, M., Di Michiel, M. & Billinge, S. J. L. (2015). Nanoscale, 7, 5480–5487.  Web of Science CrossRef CAS PubMed Google Scholar
First citationTominaka, S., Hiroki, Y., Satoshi, H., Saori , I. K. & Koji, O. (2018). ACS Omega, 3, 8874–8881.  Web of Science CrossRef Google Scholar
First citationTominaka, S., Kawakami, K., Fukushima, M. & Miyazaki, A. (2017). Mol. Pharm. 14, 264–273.  Web of Science CrossRef PubMed Google Scholar
First citationUmeda, T., Yamada, H., Ohara, K., Yoshida, K., Sasaki, Y., Takano, M., Inagaki, S., Kubota, Y., Takewaki, T., Okubo, T. & Wakihara, T. (2017). J. Phys. Chem. C, 121, 24324–24334.  Web of Science CrossRef Google Scholar
First citationWright, A. C. (1990). J. Non-Cryst. Solids, 123, 129–148.  CrossRef Web of Science Google Scholar
First citationYoneda, Y., Kohara, S., Ito, M., Abe, H., Takeuchi, M., Uchida, H. & Matsumura, Y. (2013). Trans. Mater. Res. Soc. Jpn, 38, 109–112.  CrossRef Google Scholar
First citationYoung, C. A. & Goodwin, A. L. (2011). J. Mater. Chem. 21, 6464–6476.  Web of Science CrossRef CAS Google Scholar

This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.

Journal logoJOURNAL OF
SYNCHROTRON
RADIATION
ISSN: 1600-5775
Follow J. Synchrotron Rad.
Sign up for e-alerts
Follow J. Synchrotron Rad. on Twitter
Follow us on facebook
Sign up for RSS feeds