research papers
Montel mirror based collimating analyzer system for high-pressure resonant inelastic X-ray scattering experiments
aCenter for Artificial Low Dimensional Electronic Systems, Institute for Basic Science (IBS), Pohang 790-784, Republic of Korea, bDepartment of Physics, Pohang University of Science and Technology, Pohang 790-784, Republic of Korea, and cAdvanced Photon Source, Argonne National Laboratory, Lemont, IL 60439, USA
*Correspondence e-mail: jhkim@anl.gov
Resonant inelastic X-ray scattering (RIXS) is increasingly playing a significant role in studying highly correlated systems, especially since it was proven capable of measuring low-energy magnetic excitations. However, despite high expectations for experimental evidence of novel magnetic phases at high pressure, unequivocal low-energy spectral signatures remain obscured by extrinsic scattering from material surrounding the sample in a diamond anvil cell (DAC): pressure media, Be gasket and the diamond anvils themselves. A scattered X-ray collimation based medium-energy resolution (∼100 meV) analyzer system for a RIXS spectrometer at the Ir L3-absorption edge has been designed and built to remediate these difficulties. Due to the confocal nature of the analyzer system, the majority of extrinsic scattering is rejected, yielding a clean low-energy of an iridate Sr2IrO4 sample in a DAC cell. Furthermore, the energy resolution of different configurations of the collimating and analyzing optics are discussed.
Keywords: resonant inelastic X-ray scattering; high pressure; scattered X-ray collimation; Montel mirror.
1. Introduction
Resonant inelastic X-ray scattering (RIXS) is a photon-in/photon-out spectroscopic technique that provides invaluable information on the electronic and magnetic properties of complex materials (Ament et al., 2011). After the momentous observation that RIXS can also detect single spin-flip excitations in cuprates (Ament et al., 2009; Braicovich et al., 2010), RIXS is considered as a unique alternative technique to inelastic neutron scattering for accessing elementary magnetic excitations and has been widely utilized to study various correlated systems (Chun et al., 2015; Kim, Casa et al., 2012; Kim, Said et al., 2012).
Applying pressure is an effective way to tune interactions in a correlated system so that novel phases can be realized. In particular, superconducting and quantum spin liquid phases are expected to emerge upon applying pressure to certain iridate compounds (Haskel et al., 2012, 2020; Majumder et al., 2018; Yadav et al., 2018). As such, novel phases can be identified by their particular spectral features corresponding to unique magnetic excitations; therefore, RIXS measurements at high pressure have been attempted on Sr2IrO4 and Sr3Ir2O7 (Kim, 2016; Rossi et al., 2019). However, it was instantly revealed that low-energy spectra are contaminated by extrinsic scattering from the high-pressure sample environment.
Indeed, when using a conventional RIXS spectrometer [using a diced spherical analyzer and a strip detector on a Rowland circle (Shvyd'ko et al., 2013)], this extrinsic scattering contamination is fundamentally inevitable. Although an attempt to use a post-sample clean-up slit alleviated the problem to some degree at the cost of signal reduction (Rossi et al., 2019), it is still difficult to achieve a sufficiently small field of view (FOV) with a slit and, therefore, to avoid extrinsic scattering from a Be gasket at high pressures as its inner edge is squeezed increasingly closer to the sample. The slit needs to be located closer to the sample position for a smaller FOV, but a bulky diamond anvil cell (DAC) restricts the slit to a minimum distance from the sample.
To circumvent this predicament, a scattered X-ray collimation (SXC) RIXS analyzer system was designed and built. Actual measurements in a DAC cell show that the confocal nature of the SXC optics can effectively reject extrinsic scattering and produce much cleaner spectra than a conventional setup based on a diced spherical analyzer, even in the extreme case where the pressure is increased to the point where the Be gasket actually touches the sample. Additionally, three different configurations of the SXC-RIXS spectrometer are tested and compared in terms of energy resolution.
2. Low-energy contamination and scattered X-ray collimation
A conventional RIXS spectrometer is shown schematically in Fig. 1(a). It consists of a diced spherical analyzer (A) and a strip detector (D) in a Rowland circle geometry that includes the sample (S) (Shvyd'ko et al., 2013). The flat crystal dices of the analyzer disperse the scattered radiation from the sample, and the corresponding intensities are then assigned to different energies based on their position along the detector around the reference point F. An extrinsic source of S′ near S will disperse around a point F′ along the detector, therefore adding spectral intensity at different energies due to the displacement.
A bent analyzer, on the contrary, is rather insensitive to the presence of extrinsic scatterers because it focuses X-rays scattered from a single scatterer into a single focal point, as shown in Fig. 1(a). If an extrinsic scatterer (S′) were introduced, X-rays from the scatterer would be focused on another focal point (F′), which is newly defined by an extrinsic scatterer (S′) and analyzer (A) following the Rowland circle geometry. Thus, if an area detector has enough resolution to resolve these two focal points (F and F′), then a RIXS spectrum of the target sample (S) would not be spoiled by extrinsic signals. This spatial resolution can be applied to direct tomography (Huotari et al., 2011). However, the rather coarse resolution (few hundreds of meV) due to bending strain on the crystal makes it unsuitable for the measurements discussed here.
For a successful RIXS measurement under high pressure, here we explore a RIXS spectrometer based on flat-crystal optics that offers a high energy resolution and an important advantage regarding suppressing extrinsic signals, compared with the diced spherical analyzer system (Kim, 2016). The confocal property comes from its first optical element, a collimating laterally graded multilayer Montel mirror, shown schematically in Fig. 2(a). The Montel mirrors for RIXS and spectrometers were designed and their performances were confirmed earlier (Kim et al., 2016; Mundboth et al., 2014). The relatively narrow reflectivity curve of the mirror [Fig. 2(b)] fundamentally rejects X-rays scattered from an extrinsic scattering source (S′). This concept can also be understood as a limited FOV of the Montel mirror at the sample position, which is the angular acceptance multiplied by the working distance, in this case 0.028° × 200 mm = 98 µm. Since the distance between the sample center and the Be gasket center, 110 µm, is larger than half of the Montel mirror's FOV, extrinsic signals from the Be gasket can be effectively rejected. We note that a collimating polycapillary optics and a flat analyzer crystal have been used for a few-eV-resolution high-pressure emission spectroscopy to collect the signal only from a small volume at its focus (Heald et al., 2012) and also for a RIXS or spectrometer for exceptionally high resolutions (Kim et al., 2018; Shvyd'ko et al., 2014).
3. Experimental
Fig. 3 introduces two configurations of the SXC-RIXS spectrometers for the experiments. In the configuration in Fig. 3(a), a crystal (C), asymmetrically (b = −0.0642) cut Si(111), is used to further collimate the X-rays exiting the Montel mirror (M). Detailed information on the Montel mirror can be found in a previous study (Kim et al., 2016). Then, the symmetric flat analyzer crystal (A) selects the energies of the collimated X-rays. Fig. 3(b) describes a one-step collimating configuration of SXC-RIXS consisting of the Montel mirror (M) and a flat analyzer crystal (A). These configurations of SXC-RIXS spectrometers in Figs. 3(a) and 3(b) are referred to as the `MCA' and `MA' configurations, respectively.
The experiments were organized into two steps. First, RIXS spectra of an Sr2IrO4 single crystal in a DAC were obtained by a conventional-type spectrometer, such as shown in Fig. 1 which uses a diced spherical Si(844) analyzer on a 2 m arm. These were compared with the spectra measured by both MCA and MA setups with a symmetric Si(844) flat crystal analyzer to demonstrate and quantify the extrinsic scattering rejection capabilities of the SXC system. Afterwards, the confocal performance of the MA configuration was once more tested in the extreme case where the inner edge of the Be gasket touches the sample at high pressures (6.8 and 15.6 GPa).
Second, three different SXC-RIXS configurations are compared in terms of energy resolution using 3M Magic Scotch tape as an elastic reference scatterer. MCA/MA configurations with a flat Si(844) analyzer and an MA configuration with a flat sapphire(078) analyzer were tested. Experimentally obtained energy resolutions are then compared with the calculated ones.
The sample environment is shown schematically in Figs. 2(a) and 3(c). A pair of 0.25-carat brilliant-cut diamond anvils with 400 µm culets were fixed at the seats of a Mao-type symmetric pressure cell, which has four windows with ∼30° opening (Li et al., 2018). A piece of Sr2IrO4 single crystal with c-axis oriented in the lateral plane was prepared by a focused ion beam (FIB) technique to a 100 µm × 100 µm lateral size with a thickness of 20 µm and was placed inside a 220 µm-diameter hole of a Be gasket along with ruby chips as a pressure reference. After closing the cell, Ne gas was loaded into the cell with a base pressure of 1.2 GPa. A c-axis Bragg peak was measured but the sample orientation was not fully determined in this study. The pressure was measured and controlled by an in situ membrane-driven and ruby-measurement system. During the experiment, two other pressures (6.8 GPa, 15.6 GPa) were reached besides the base pressure.
Measurements were performed at the 27-ID beamline at the Advanced Photon Source, where a diamond (111) high-heat-load monochromator and a four-bounce asymmetric Si(400) medium-resolution monochromator produced an X-ray beam at 11.215 keV with 70 meV bandpass (Shvyd'ko et al., 2013). The beam was focused on 54 µm (H) × 16 µm (V) onto the sample using a KB mirror pair.
4. Result and discussion
4.1. Performance of Montel mirror and flat crystals
Fig. 4(a) shows RIXS spectra measured by three different spectrometers, demonstrating that SXC-RIXS spectrometers effectively reduce extrinsic signals from the sample environment, which contaminate the low-energy part of the spectrum measured using a diced spherical analyzer. It can be more clearly seen by fitting the spectra using known peaks at low-energy RIXS spectra of Sr2IrO4 (Kim, Casa et al., 2012), as exhibited in Figs. 4(c) and 4(d). The RIXS spectrum taken by the MCA spectrometer in Fig. 4(c) is well fitted by three Gaussian-line-shaped peaks, which correspond to an elastic line, magnon, and dd-excitation from low energy loss and, also, magnon and dd-excitations are located at the correct energies with proper widths. However, the RIXS spectrum taken by the conventional RIXS spectrometer with a spherical analyzer cannot be fully reproduced with those three peaks, showing a discrepancy between the line made by the three peaks (solid black line) and the actual spectrum (black dots), especially around the zero energy-loss.
Note that the count rate of the MA configuration is comparable with that of the conventional spectrometer using a diced spherical analyzer; the intensities around 600 meV (dd-excitation) in Fig. 4(a) are similar for both setups. Since the solid angle acceptance of scattered X-rays was kept constant by adjusting the mask size of the diced spherical analyzer, the intensities shown in Fig. 4(a) directly correspond to the efficiencies of the spectrometers.
The inset of Fig. 4(b) shows the sample environment at 15.6 GPa where the Be gasket collapses to contact the sample. As a collapsed Be gasket invades the FOV of the mirror at higher pressure, elastic signals increase upon applying higher pressures. Nevertheless, the SXC-RIXS spectrometer still works well and gives a clean low-energy even in this extreme sample environment.
4.2. Comparing performances of the three different configurations
Figs. 5(a) and 5(b) show the measured and calculated incident-energy-dependence curves of three different SXC-RIXS spectrometers, and their resolutions and throughputs are tabulated in Table 1 (the calculation method is described in Appendix A). In the experimental results, the MCA setup with a Si(844) analyzer and the MA setup with a sapphire(078) analyzer have better energy resolutions than that of the MA setup with the Si(844) analyzer. Although the same kind of calculation greatly reproduced the measured resolutions of a SXC-RIXS spectrometer in the previous study (Kim et al., 2018), in our current case there is one major difference with the calculation: energy resolutions of the MA configurations are notably sharper than that calculated. This discrepancy could come from the multi-bounce monochromator used in this experiment, which consists of asymmetric crystals.
|
A multi-bounce monochromator consisting of asymmetric crystals is known to produce a vertically energy-dispersed beam spot at the sample position (Huang et al., 2012), while the calculation here assumes no spatial energy dispersion. Therefore, when using a multi-bounce asymmetric monochromator, the Montel mirror emittance is correspondingly dispersed in the energy–angle space due to the finite beam size, which in turn results in a reduced width when rocking an analyzer crystal, which has a narrower acceptance than the divergence of the Montel mirror (the detailed explanation is given in Appendix B). These dispersion effects are not significant when using a C crystal since it collimates the mirror emittance to fit the analyzer crystal acceptance by design or trivially when using a symmetric channel-cut monochromator. Additionally, this dispersion effect also produces Lorentzian-like line shapes of the resolution curves for MA setups.
Throughputs are estimated by the integrated intensities under the curves, and used to compare the efficiency of each setup. In Table 1 it is seen that the relative throughput of two MA setups is reasonably explained by the calculations. On the other hand, the MCA setup shows a noticeably smaller value in the measured throughput than in the calculated throughput. It seems that the spatial energy dispersion and resulting complicated beam profile not only affect the spectrometer resolution but also the spectrometer throughputs.
5. Summary and outlook
A collimation-based analyzer system for a RIXS spectrometer, dubbed SXC-RIXS, was designed and implemented to study low-energy excitation spectra of samples under high pressure. In contrast to conventional spectrometers with diced spherical analyzers, this new setup can effectively remove extrinsic signals from a high-pressure sample environment and provide clean low-energy excitation spectra, thanks to the confocal nature of its first collimating element, a multilayer Montel mirror. Even in the extreme case when the applied pressure on the diamond anvil cell collapses the Be gasket into the sample, the SXC-RIXS spectrometer still provides a clean low-energy excitation, albeit with an increased elastic line. Experimental comparison of the energy resolutions of three different SCX-RIXS reveals that the MA and MCA setups give similar resolutions when using a medium-energy bandpass monochromator (70 meV), although the calculation claims that the MA setup would have worse resolutions. We attribute this discrepancy to the spatially energy-dispersed beam known to result from a monochromator based on asymmetrical crystals. Finally, an eventual SXC-RIXS type spectrometer is ideally suited to fulfill increasing demands for high-resolution RIXS measurements at high pressures, especially after the planned APS upgrade (APS-U).
APPENDIX A
Calculating A-crystal rocking curves and incident energy curves
The ). Multiplying the reflectivity of an analyzer and a double-Gaussian-shaped, which is assumed, exit beam from the Montel [Fig. 6(a)] gave the intensity curvature of the reflected beam from an analyzer [Fig. 6(b)]. One more identical step of multiplying reflectivity is added when a C-crystal is included. The volume under the curvature corresponds to the total intensity of the reflected beam after the analyzer. Incident energy dependence curves in Fig. 5(b) were acquired by changing an energy-centered position of the incident beam.
of diffraction was used to calculate incident energy curves (Authier, 2006APPENDIX B
Effects of the spatial energy dispersion
As mentioned in the discussion section, the spatial energy dispersion generated by the asymmetrical monochromator could lead to better energy resolutions of the MA than that expected from calculations. We will explain how this could lead to better energy resolutions more specifically. Fig. 7(a) shows that the energy-dispersed beam was focused on the sample position with a finite beam size (∼16 µm). To simplify the case, we assume that energies are uniformly distributed in the vertical direction. The energy of the top ray (red) is 35 meV higher than that of the center ray (dashed black) and 70 meV higher than that of the bottom ray (blue). Then, the angle difference between the top (or bottom) ray and the center ray is roughly = = ≃ (s/2)/WD, where s is the vertical beam size, and WD is the distance between the sample position and the Montel mirror. With s = 16 µm and WD = 200 mm, the angle difference is about 40 µrad. Due to the large acceptance of the Montel mirror, all the scattered rays from the sample will be reflected to the analyzer. However, in the MA setup used in the experiments, the top (bottom) ray is incident on the analyzer with a larger (smaller) angle than the reference angle whereas it has higher energy (lower energy) or smaller (larger) than that of the reference ray. This effect can be described by a diagonally shrunk profile of the incident beam in a DuMond diagram as seen in Fig. 7(b). As the profile of the incident beam is shrunk along the reflectivity curve of the analyzer crystal, the energy resolution curve will be narrower than the non-dispersive case (dashed black circle). However, the energy resolutions are not easy to estimate without a real calculation because they are evaluated by a convolution of two complex and anisotropic two-dimensional curves. Still, we could see that this effect is more drastic when using Si(844) as an analyzer (34% reduction in resolution) than when using Sapph(078) (28% reduction), since it has smaller acceptance than that of Sapph(078) with a similar energy window at the Ir L3-edge (Table 2), and is consequently more susceptible to this effect.
|
Funding information
J.-K. Kim was supported by the Global Ph.D. Fellowship Program from National Research Foundation of Korea (Grant No. 2018H1A2A1059958) and by IBS-R014-A2. This material is based upon work supported by Laboratory Directed Research and Development (LDRD) funding from Argonne National Laboratory, provided by the Director, Office of Science, of the U.S. Department of Energy under Contract No. DE-AC02-06CH11357
References
Ament, L. J., Ghiringhelli, G., Sala, M. M., Braicovich, L. & van den Brink, J. (2009). Phys. Rev. Lett. 103, 117003. Web of Science CrossRef PubMed Google Scholar
Ament, L. J. P., van Veenendaal, M., Devereaux, T. P., Hill, J. P. & van den Brink, J. (2011). Rev. Mod. Phys. 83, 705–767. Web of Science CrossRef CAS Google Scholar
Authier, A. (2006). International Tables for Crystallography, Vol. B, Reciprocal Space, edited by U. Shmuelipp, pp. 626–646. International Union of Crystallography. Google Scholar
Braicovich, L., van den Brink, J., Bisogni, V., Sala, M. M., Ament, L. J., Brookes, N. B., De Luca, G. M., Salluzzo, M., Schmitt, T., Strocov, V. N. & Ghiringhelli, G. (2010). Phys. Rev. Lett. 104, 077002. Web of Science CrossRef PubMed Google Scholar
Haskel, D., Fabbris, G., Kim, J. H., Veiga, L. S. I., Mardegan, J. R. L., Escanhoela, C. A., Chikara, S., Struzhkin, V., Senthil, T., Kim, B. J., Cao, G. & Kim, J. W. (2020). Phys. Rev. Lett. 124, 067201. Web of Science CrossRef PubMed Google Scholar
Haskel, D., Fabbris, G., Zhernenkov, M., Kong, P. P., Jin, C. Q., Cao, G. & van Veenendaal, M. (2012). Phys. Rev. Lett. 109, 027204. Web of Science CrossRef PubMed Google Scholar
Heald, S., Seidler, G. T., Mortensen, D., Mattern, B., Bradley, J. A., Hess, N. & Bowden, M. (2012). Proc. SPIE, 8502, 85020I. CrossRef Google Scholar
Huang, X. R., Macrander, A. T., Honnicke, M. G., Cai, Y. Q. & Fernandez, P. (2012). J. Appl. Cryst. 45, 255–262. Web of Science CrossRef CAS IUCr Journals Google Scholar
Huotari, S., Pylkkänen, T., Verbeni, R., Monaco, G. & Hämäläinen, K. (2011). Nat. Mater. 10, 489–493. Web of Science CrossRef CAS PubMed Google Scholar
Hwan Chun, S., Kim, J.-W., Kim, J., Zheng, H., Stoumpos, C., Malliakas, C., Mitchell, J., Mehlawat, K., Singh, Y., Choi, Y., Gog, T., Al-Zein, A., Sala, M., Krisch, M., Chaloupka, J., Jackeli, G., Khaliullin, G. & Kim, B. J. (2015). Nat. Phys. 11, 462–466. Web of Science CrossRef Google Scholar
Kim, J. (2016). High. Press. Res. 36, 391–403. Web of Science CrossRef CAS Google Scholar
Kim, J., Casa, D., Said, A., Krakora, R., Kim, B. J., Kasman, E., Huang, X. & Gog, T. (2018). Sci. Rep. 8, 1958. Web of Science CrossRef PubMed Google Scholar
Kim, J., Casa, D., Upton, M. H., Gog, T., Kim, Y. J., Mitchell, J. F., van Veenendaal, M., Daghofer, M., van den Brink, J., Khaliullin, G. & Kim, B. J. (2012). Phys. Rev. Lett. 108, 177003. Web of Science CrossRef PubMed Google Scholar
Kim, J., Said, A. H., Casa, D., Upton, M. H., Gog, T., Daghofer, M., Jackeli, G., van den Brink, J., Khaliullin, G. & Kim, B. J. (2012). Phys. Rev. Lett. 109, 157402. Web of Science CrossRef PubMed Google Scholar
Kim, J., Shi, X., Casa, D., Qian, J., Huang, X. & Gog, T. (2016). J. Synchrotron Rad. 23, 880–886. Web of Science CrossRef CAS IUCr Journals Google Scholar
Li, B., Ji, C., Yang, W., Wang, J., Yang, K., Xu, R., Liu, W., Cai, Z., Chen, J. & Mao, H. K. (2018). Proc. Natl Acad. Sci. USA, 115, 1713–1717. Web of Science CrossRef CAS PubMed Google Scholar
Majumder, M., Manna, R. S., Simutis, G., Orain, J. C., Dey, T., Freund, F., Jesche, A., Khasanov, R., Biswas, P. K., Bykova, E., Dubrovinskaia, N., Dubrovinsky, L. S., Yadav, R., Hozoi, L., Nishimoto, S., Tsirlin, A. A. & Gegenwart, P. (2018). Phys. Rev. Lett. 120, 237202. Web of Science CrossRef PubMed Google Scholar
Mundboth, K., Sutter, J., Laundy, D., Collins, S., Stoupin, S. & Shvyd'ko, Y. (2014). J. Synchrotron Rad. 21, 16–23. Web of Science CrossRef CAS IUCr Journals Google Scholar
Rossi, M., Henriquet, C., Jacobs, J., Donnerer, C., Boseggia, S., Al-Zein, A., Fumagalli, R., Yao, Y., Vale, J. G., Hunter, E. C., Perry, R. S., Kantor, I., Garbarino, G., Crichton, W., Monaco, G., McMorrow, D. F., Krisch, M. & Moretti Sala, M. (2019). J. Synchrotron Rad. 26, 1725–1732. Web of Science CrossRef CAS IUCr Journals Google Scholar
Shvyd'ko, Y., Stoupin, S., Shu, D., Collins, S. P., Mundboth, K., Sutter, J. & Tolkiehn, M. (2014). Nat. Commun. 5, 4219. Web of Science PubMed Google Scholar
Shvyd'ko, Y. V., Hill, J. P., Burns, C. A., Coburn, D. S., Brajuskovic, B., Casa, D., Goetze, K., Gog, T., Khachatryan, R., Kim, J. H., Kodituwakku, C. N., Ramanathan, M., Roberts, T., Said, A., Sinn, H., Shu, D., Stoupin, S., Upton, M., Wieczorek, M. & Yavas, H. (2013). J. Electron Spectrosc. Relat. Phenom. 188, 140–149. CAS Google Scholar
Yadav, R., Rachel, S., Hozoi, L., van den Brink, J. & Jackeli, G. (2018). Phys. Rev. B, 98, 121107(R). Web of Science CrossRef Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.