Performance of the BL03U beamline at SSRF
aCenter for Excellence in Superconducting Electronics, State Key Laboratory of Functional Materials for Informatics, Shanghai Institute of Microsystem and Information Technology, Chinese Academy of Sciences, Shanghai 200050, People's Republic of China, bCenter of Materials Science and Optoelectronics Engineering, University of Chinese Academy of Sciences, Beijing 100049, People's Republic of China, and cSchool of Physical Science and Technology, ShanghaiTech University, Shanghai 201210, People's Republic of China
*Correspondence e-mail: yemao@mail.sim.ac.cn, qiaoshan@mail.sim.ac.cn
The vacuum ultraviolet beamline BL03U with a photon energy range from 7 eV upwards has been constructed at the 3.5 GeV Shanghai Synchrotron Radiation Facility. Equipped with an APPLE-Knot undulator, this beamline is dedicated to angle-resolved photoemission spectroscopy. An energy-resolving power of higher than 4.6 × 104 has been achieved in the photon energy range 21.6–48 eV, which is almost the same as the theoretical estimation.
Keywords: vacuum ultraviolet beamline; APPLE-Knot undulator; high energy resolution; ARPES; plane-grating monochromator.
1. Introduction
Material science has developed tremendously in recent years due to rapidly improved experimental techniques based on synchrotron radiation light sources. In order to develop novel materials, it is necessary to have a fundamental understanding of how the functions of materials are related to their electronic structures as an origin of the physical properties of solids. The specific regime of very high energy resolution angle-resolved et al., 2003). However, owing to its surface sensitivity, ARPES could only be used to observe the electronic structure near the surface, due to the relatively short of photoelectrons (Brundle, 1974). Low energy photons with energy around 7 eV are capable of determining the bulk electronic state with higher energy resolution. On the other hand, higher energy photons with energies up to several hundred electron volts (eV) could also be used to determine the bulk electronic state. Furthermore, the polarizations of the incident photons provides a new degree of freedom in ARPES measurements, enabling us to selectively probe the electrons from orbitals with different symmetries. Thus, a beamline which can cover low photon energy down to 7 eV and up to several hundred electron volts with variable polarizations is long-awaited in the study of condensed matter physics by means of ARPES.
(ARPES) has proven to be particularly successful for the measurement of electronic states at low binding energies in correlated electron systems (DamascelliA vacuum ultraviolet (VUV) beamline (BL03U) has been built at the Shanghai Synchrotron Radiation Facility (SSRF), and has been designed for ARPES experiments with a photon energy range of 7–798 eV. It is rather challenging to generate 7 eV photons at SSRF, because the energy of the storage ring is as high as 3.5 GeV. In this case a deflection parameter K of greater than 10 is required and the heat load would be a serious problem (Hofmann, 2004). To solve this problem, an APPLE-Knot undulator was adopted, which can generate photons with variable polarizations and low on-axis heat load (Ji et al., 2015). In this paper, we report the design and performance test of this beamline.
2. Layout of the beamline
A schematic layout of the beamline is shown in Fig. 1 and the parameters selected are given in Table 1. In Fig. 1, the of the coordinates is set as the center of the undulator. The photon beam generated from the APPLE-Knot undulator is first focused horizontally onto the exit slit by an elliptical cylindrical mirror M1 (19.1 m) and deflected by 3.2° in the horizontal direction. Apertures set before (18.33 m) and after (21.13 m) M1 are used to control the acceptance angle of the photon beam. A plane mirror M2 is used to reflect the beam upwards toward the gratings (22 m). After the varied-line-space plane grating (VLPG) (Harada et al., 1984; Itou et al., 1989; Yan et al., 1998), the photon beam is focused onto the exit slit (29 m) vertically. The plane mirror M5 (23 m) set after the monochromator is used to switch the photon beam to pass the main or branch line. After the exit slit, the photon beam is refocused in both the horizontal and vertical directions by an ellipsoidal mirror (M3 at 38.6 m or M4 at 34.8 m for the main or branch line, respectively). The total length of the beamline from source to sample position is 40.3 m or 36.2 m for the main or branch line, respectively.
|
For better M1, M2, gratings, M5) are made of silicon. Also, a conventional side-cooling system was adopted for M1, M2 and the grating. The ellipsoidal mirrors M3 and M4 are made of silicon dioxide as they are at the end of the beamline where the heat load is negligible. All of the optical elements are coated with Au. Table 2 shows the parameters of the ellipsoidal mirrors. Since M1 is the first mirror of the beamline with long entrance arm (source to M1, 19.1 m) and exit arm (M1 to exit slit, 9.9 m), it has higher requirements regarding slope error and roughness.
most of the optical elements (
|
2.1. Undulator
The beamline is served from Section 03U of the SSRF. As a photon source, a 4.2 m-long APPLE-Knot undulator was put into practical use for the first time. Unlike conventional undulators, APPLE-Knot undulators can provide photons with arbitrary polarizations and a much lower on-axis heat load. According to simulations, the heat loads of the APPLE-Knot undulator are 59.4, 43.9 and 314 W for horizontal, circular and vertical modes, respectively, inside the 0.6 mrad × 0.6 mrad acceptance solid angle, which is satisfactory compared with 1106 W for a pure linear undulator (Ji et al., 2015). First-order light covers the photon energy range 7–165 eV, and higher harmonics could be used to provide higher-energy photons.
The effective electron beam size σr0 and divergence of the storage ring of SSRF in the vertical (horizontal) direction are 8.36 µm (143.2 µm) and 4.18 µrad (34 µrad), respectively. Then, from consideration of the spatial coherence (diffraction limit), the r.m.s. of size σr and divergence angle of the source are (Onuki & Elleaume, 2003)
where λ and L are the wavelength of the photon and the total length of the undulator, respectively. The above equations give us the σr and values at the harmonic energy. However, for certain acceptance angles, the peak energy is lower than the harmonic energy and the r.m.s. of the source size at peak energy (σv) is smaller than that calculated by equation (1) (σv0) (Onuki & Elleaume, 2003). In order to solve this problem we use the SPECTRA program (Tanaka & Kitamura, 2001) to calculate the source sizes at peak energies. Table 3 shows a comparison of the source sizes obtained by the two methods at three different photon energies corresponding to the gas absorption edges. One can clearly see that the source sizes at peak energy σv are significantly smaller than those at resonance energies σv0 by 30–40%.
|
2.2. Monochromator
In order to achieve good efficiency of the gratings, i.e. as high as possible for different photon energies and a small beam size at the sample position, we choose a variable-groove-depth (VGD) and a variable-line-space plane-grating in the design of the monochromator (VLPGM). The design principles of the VLPGM, as well as its advantages compared with other types of monochromators, have been described in detail earlier (Reininger & Castro, 2005; Wang et al., 2015). Therefore only a brief description of the geometry and optical elements used in this design is given below. Three Au-coated gratings with line densities of 188, 620 and 2000 lines mm−1 are used, covering the energy range 7–798 eV. Further details of these gratings are listed in Table 4. To obtain high low-line-density gratings (188 lines mm−1) are used, and consequently the diffraction limitation takes an important role in the energy resolution. To decrease the contribution from the diffraction limitation, the monochromator parameters are carefully selected, and almost the same illuminated length, which is also the effective grating length, on the grating can be achieved for different photon energies, which also means that the maximum incident angles or, say, the maximum reflectivity has been achieved for the whole energy range. Suppose the line density of the VLPGM is
where the positive direction of w is along the photon propagation direction, then the grating diffraction and the focus condition for the −1 order (inner order) diffraction are
where r, are distances between the grating and source and exit slit, and λ is the photon wavelength. According to equations (4) and (5), each photon wavelength corresponds to a set of incident angle (α) and diffraction angle (β). Therefore, photons of different energies can be delivered to the exit slit through the coordinated rotation of M2 and the grating. When the values of r and are different from the theoretical ones, the focus condition can be remedied by the change of α and β achieved by the rotations of the pre-mirror (M2) and grating, which is another advantage of the VLPGM. The vertical magnification of the grating is
To keep the fixed direction of the exit photons, the M2 mirror needs both translational and rotational movements. This complex movement can be achieved by an off-axis rotation of M2. As shown in Fig. 2, the distance between the entrance light and the rotation center of the grating is set as 28 mm. The best performance is achieved when we set the rotation axis of M2 at (−0.02 mm, 14.41 mm) referring to the grating center, and the movement of the beam center on the grating surface is shorter than 20.6 µm during the energy scan.
|
3. Theoretical performance from beamline design
Firstly, we estimate the energy resolution of the beamline by considering the contribution from the source size, diffraction limitation, slope error of M2, slope error of the grating and coma focus, which could be calculated using the following equations,
where σv, σs, l, , σm and σg are the vertical source size, vertical width of the exit slit, illuminated length, vertical source divergence, mirror slope error and grating slope error, and
Here, F30 is set to zero at the midpoint of the energy range covered by each grating by choosing a suitable b3. In the calculations, the vertical source sizes (σv) are obtained using the SPECTRA program as listed in Table 3. The slope errors of the gratings and M2 are selected as the measured values from real optics of this beamline, which are 0.12 µrad, 0.13 µrad and 0.14 µrad for G1, G2 and M2, respectively. Fig. 3 shows the total energy resolution calculated by the above equations for a 10 µm vertical width of the exit slit. One can see that grating 2 has a better energy resolution and its energy resolution reaches below 0.5 meV of the full width at half-maximum (FWHM) at the photon energy 21.6 eV, in which most contributions are from the diffraction limitation (0.31 meV), the vertical width of the exit slit (0.30 meV) and the source size (0.24 meV), and the others can be ignored.
The parameters chosen for each grating are shown in Table 4, and the photon fluxes at the sample position of the three gratings are shown in Fig. 4, calculated using the code REFLECT (Schäfers et al., 2002). As exhibited in Fig. 4, all three gratings could reach fluxes of more than 1013 photons s−1 (0.1% bandwidth)−1 in the photon energy range 7–100 eV. In the calculations, the ring current is set as 300 mA, which is that of SSRF at full state.
4. Beam delivery performance
The energy resolution of the monochromator was tested by performing gas-phase photoabsorption measurements on He and Ne. The gas cell employed in these studies was mounted behind the post-focusing mirror (M3). A perforated plate valve with thin film was used to isolate the gas from the ultrahigh vacuum (UHV), and the thickness of the Al film was 150 nm, and is transparent to photons in the photon energy range 20–70 eV. The pressure of the gas was set to approximately 5 × 10−2 mbar as measured using a Pirani vacuum gauge. A photodiode of type AXUV100G (OptoDiode Corporation, California, USA) inside the gas cell was used to measure the photon intensity I. The photocurrent generated on M3 was used as the initial current I0. In the gas scan, the minimum energy step was set at 0.05 meV, which was close to the driving limit of the stepper motor of M2 and the grating, which is 75 nrad step−1, corresponding to 0.01 meV. The experiments were carried out at room temperature with horizontally polarized light. The results of the photoabsorption (I/I0) for different gases are displayed in Fig. 5. Here only the photoabsorption spectra of G2 with line density 620 lines mm−1 are illustrated. More test results of other gratings are listed in Table 5.
|
The photoabsorption spectra with different photon energies are shown in Figs. 5(a)–5(c) with the vertical width of the exit slit being 10 µm. The experimental resolution was estimated by a multi-peak fitting method [equation (14)], and the standard deviation σ of the energy distribution of the incident monochromatic light could be obtained,
where xn is the position of the nth peak. An energy resolution (FWMH = 2.35σ) of 0.47 meV was achieved at 21.6 eV, corresponding to a of E/ΔE = 4.6 × 104. Similarly, the resolving powers for photon energies at 48.6 eV and 65.5 eV are 4.7 × 104 and 3.1 × 104, respectively [Figs. 5(b)–5(c)]. Fig. 5(d) shows a typical series of Ne spectra at a photon energy of 21.6 eV with different vertical widths of the exit slit. The multi-peak fittings are shown as blue curves in Fig. 5(d) and the results are presented as blue squares in Fig. 5(e). One can see that the energy resolution changes linearly with the vertical width of the exit slit.
The M3. The diode response was calibrated using a standard curve from the diode manufacturer, Opto Diode Corporation. The resulting calibrated photon fluxes for different vertical widths of the exit slit are also shown on the right-hand axis of Fig. 5(e), with 240 mA ring current. As the vertical width of the exit slit becomes larger, the energy resolution becomes linearly worse and the received photons increase linearly, which leads to basically an unchanged that remains above 2.2 × 1013 photons s−1 (0.1% bandwidth)−1 when the vertical width of the exit slit is above 30 µm. The drops by half when the vertical width of the exit slit is adjusted to 10 µm because the energy resolution is no longer linearly proportional to the slit width and also other factors (such as the diffraction limitation) become major contributions to the energy resolution as discussed before.
on the samples is measured using an AXUV100G photodiode downstream of mirrorMore experimental results of the . It is worth noting that the of G2 remains above 4.6 × 104 with a flux of over 1013 photons s−1 (0.1% bandwidth)−1 in the photon energy range 21–48 eV. The resolving powers of G1 and G2 are slightly decreased when the photon energy is above 48 eV, which is consistent with the previous theoretical calculations. The measured is approximately (1.6–2.9) × 1013 photons s−1 (0.1% bandwidth)−1 for the 188 lines mm−1 grating (G1) and (2.1–2.6) × 1013 photons s−1 (0.1% bandwidth)−1 for the 620 lines mm−1 grating (G2) in the energy range 21–65 eV. Considering that the ring current is 240 mA during the test, the would increase accordingly when the ring current is in a 300 mA full state. Overall, the measured results show good agreement with the estimated values.
and for different gratings are exhibited in Table 5Next, we tested the performance of the beamline in delivering low-energy photons. The higher harmonic contamination of the photon beam becomes a serious problem when very low energy photons are used. Fig. 6(a) shows the photoemission spectrum with 7.2 eV photon energy. One can see that the higher harmonics give a huge background and that the first-order light of 7.2 eV is submerged in it. To solve this problem, we use a 200 µm-thick CaF2 filter at the end of the beamline. As shown in Fig. 6(b), the harmonics with photon energy above 10 eV are rejected and the first-order light at 7.2 eV shows a sharp fine shape. After the CaF2 filter, the from the photodiode at 7.2 eV photon energy is reduced to 1.65 × 10−8 A and the corresponding is 1.54 × 1011 photons s−1. The theoretical energy-resolving power is about 8000 when the photon energy is 7.2 eV (Wang et al., 2015), thus the normalized is 1.23 × 1012 photons s−1 (0.1% bandwidth)−1. The photon flux of 7.2 eV is one order of magnitude lower than that of 21.6 eV. Because the of the increases dramatically with the decrease of photon energy, the ARPES experiments of low-energy photons still have a high enough efficiency.
Finally, we tested the size of the focused beam at the sample position, since a sufficiently small spot size is critical to guarantee excellent angular resolution of photoelectrons in state-of-the-art ARPES measurements. Fig. 7 exhibits the experimental vertical and horizontal spot sizes at the sample position when the vertical width of the exit slit is set as 30 µm with 21.6 eV photon energy. The experimental data were collected by measuring the first-order derivation of the photocurrent from a square aperture scanning vertically and horizontally, respectively, yielding a spot size of 7.5 µm (V) × 67 µm (H) FWHM. The vertical size of the spot at the sample position is determined by the width of the exit slit and M3. Since M3 remains unchanged, the vertical size of the spot can be changed through the change of the width of the exit slit, while the horizontal size of the spot at the sample position is determined by the original source size of the undulator, M1 and M3. According to equation (1), the source size decreases as the photon energy increases, and theoretically the horizontal size of the spot at the sample position would decrease accordingly. However, we found that the horizontal size of the spot at the sample position is almost unchanged as the photon energy increases, which is due to the slope error of M3 playing a major role in the influence of the spot size at the sample position.
5. Summary
The vacuum ultraviolet beamline BL03U has been constructed and commissioned at SSRF for ARPES experiments. The beamline is equipped with an APPLE-Knot undulator and three VGD-VLP gratings, covering the photon energy from 7 eV upwards with variable polarizations and much lower heat load. The beamline delivers photons with a resolving power exceeding 4.6 × 104 and up to 1013 photons s−1 (0.1% bandwidth)−1 in the range 21–48 eV, when the ring current is 240 mA. The beam sizes at the sample position are 7.5 µm FWHM and 67 µm FWHM in the vertical and horizontal directions, respectively. The energy resolution and beam size at the sample both meet the design expectations, and the ARPES endstation of BL03U has been opened to users.
Acknowledgements
The authors thank Bo Zhao and Tao Xu for help and advice during the performance test of BL03U at SSRF.
Funding information
Funding for this research was provided by: National Natural Science Foundation of China (grant No. U1632266; grant No. 11927807).
References
Brundle, C. R. (1974). J. Vac. Sci. Technol. 11, 212–224. CrossRef CAS Web of Science Google Scholar
Damascelli, A., Hussain, Z. & Shen, Z. (2003). Rev. Mod. Phys. 75, 473–541. Web of Science CrossRef CAS Google Scholar
Harada, T., Itou, M. & Kita, T. (1984). Proc. SPIE, 0503, 114–118. CrossRef Google Scholar
Hofmann, A. (2004). The Physics of Synchrotron Radiation. Cambridge University Press. Google Scholar
Itou, M., Harada, T. & Kita, T. (1989). Appl. Opt. 28, 146–153. CrossRef CAS PubMed Web of Science Google Scholar
Ji, F., Chang, R., Zhou, Q., Zhang, W., Ye, M., Sasaki, S. & Qiao, S. (2015). J. Synchrotron Rad. 22, 901–907. Web of Science CrossRef CAS IUCr Journals Google Scholar
Onuki, Hideo & Elleaume, Pascal. (2003). Wigglers, Undulators and Their Applications. London: Taylor and Francis. Google Scholar
Reininger, R. & de Castro, A. R. B. (2005). Nucl. Instrum. Methods Phys. Res. A, 538, 760–770. Web of Science CrossRef CAS Google Scholar
Schfers, F., Abramsohn, D. & Krumrey, M. (2002). REFLEC, a Program to Calculate VUV/X-ray Optical Elements and Synchrotron Radiation Beamline. BESSY Press. Google Scholar
Tanaka, T. & Kitamura, H. (2001). J. Synchrotron Rad. 8, 1221–1228. Web of Science CrossRef CAS IUCr Journals Google Scholar
Wang, J., Ye, M., Shi, T., Chang, R. & Qiao, S. (2015). Chin. Phys. C. 39, 048001. Web of Science CrossRef Google Scholar
Yan, Y., Shigemasa, E. & Yagishita, A. (1998). J. Synchrotron Rad. 5, 246–251. Web of Science CrossRef CAS IUCr Journals Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.