short communications
Synchrotron radiation X-ray microtomography for the visualization of intra-cochlear anatomy in human temporal bones implanted with a perimodiolar cochlear implant electrode array
aEuropean Institute for ORL-HNS, GZA Hospitals Antwerp, Oosterveldlaan 24, Wilrijk, Antwerp 2610, Belgium, bElettra Sincrotrone Trieste S.C.p.A., S.S. 14 Area Science Park, 34149 Basovizza, Trieste Italy, cDepartment Biology, Lab Functional Morphology, University of Antwerp, Universiteitsplein 1, Wilrijk, Antwerp 2610, Belgium, dLaboratory of Biomedical Physics, University of Antwerp, Groenenborgerlaan 171, Antwerp 2020, Belgium, and eDepartment of Radiology, GZA Hospitals Antwerp, Oosterveldlaan 24, Wilrijk, Antwerp 2610, Belgium
*Correspondence e-mail: fergio.sismono@gza.be
Recently, synchrotron radiation computed microtomography (SRµCT) has emerged as a promising tool for non-destructive, in situ visualization of cochlear implant electrode arrays inserted into a human cochlea. Histological techniques have been the `gold standard' technique for accurate localization of cochlear implant electrodes but are suboptimal for precise three-dimensional measurements. Here, an SRµCT experimental setup is proposed that offers the benefit of a high spatial and contrast resolution (isotropic voxel size = 4.95 µm and propagation-based phase-contrast imaging), while visualizing the soft-tissue structures and electrode array of the cochlear implant simultaneously. In this work, perimodiolar electrode arrays have been tested, which incorporate thick and closely spaced platinum–iridium contacts and wiring. These data can assist cochlear implant and hearing research, can be used to verify electrode segmentation techniques for clinical computed tomography or could be utilized to evaluate cochlear implant electrode array designs.
Keywords: perimodiolar electrodes; cochlear implants; synchrotron X-ray microtomography; phase-contrast imaging.
1. Introduction
Cochlear implants (CIs) are medical devices that restore hearing and speech understanding in moderate to profoundly deaf patients who are unable to receive adequate benefit from conventional hearing aids. CIs directly stimulate the auditory nerve through an electrode array inserted in the cochlea. Using a microphone and an external speech processor, sounds are captured from the environment and transformed into electrical signals that are distributed over the electrode contacts of the implanted electrode array. Each contact carries the information corresponding to a specific acoustic frequency band and electrically stimulates the target auditory neurons within the cochlea. In this way, the whole speech-related frequency spectrum of captured sound can be conveyed to the auditory system.
The position of the CI electrodes with respect to the anatomical structures in the cochlea varies as a result of the surgical procedure and is of importance for the final hearing outcome of the CI recipient. For instance, the final performance on speech recognition tests is significantly better when the electrodes are implanted into the scala tympani, the inferior partition of the cochlear duct, rather than the scala vestibuli, the superior partition (Holden et al., 2013; Shaul et al., 2018). Also, various studies suggest that CI patients have improved audiological outcomes when the electrode contacts are positioned in close proximity to the auditory nerves in the cochlear medial wall, compared with when the contacts are positioned further away (Holden et al., 2013; Chakravorti et al., 2019).
A detailed investigation of the spatial relation between the electrode contacts and the neural target structures and how this affects hearing outcomes would be of great scientific and clinical interest. However, the accurate localization of the CI-electrode remains a significant challenge that has not yet been overcome. In clinical computed tomography (CT), cochlear microstructures, such as the basilar membrane, the osseous spiral lamina and Rosenthal's canal – the structure that contains the auditory nerve cell somas and is assumed to be the primary neural target region for electrical stimulation (Kawano et al., 1998) – are invisible. This is due to the small differences in X-ray attenuation coefficients between soft-tissue and fluids within the cochlea combined with the high structural density of the surrounding temporal bone. Moreover, the electrode contacts, which are made out of a platinum–iridium alloy, absorb a great portion of the incident X-ray beam, which results in streaking artefacts in the final reconstructions (Saeed et al., 2014).
The current `gold standard' for accurate localization of the CI-electrode is ex vivo histological sectioning, but this requires extensive and complicated sample preparations, introduces processing artefacts, is destructive, time consuming and costly (Rau et al., 2013; Johnson et al., 2014; Iso-Mustajärvi et al., 2017). Another critical limitation of histological techniques is that the spatial resolution in the longitudinal plane tends to be relatively low compared with that of the transversal plane, due to the technical constraints imposed by the serial grinding of the sample (Sun et al., 2002; Rau et al., 2013). This limits the maximally achievable accuracy when performing three-dimensional measurements.
Recently published findings suggest that synchrotron radiation computed microtomography (SRµCT) may be a suitable alternative to conventional histological techniques. While SRµCT has been shown to be capable of visualizing intra-cochlear soft-tissue structures with near-histological quality, it is able to generate a 3D, high isotropic spatial resolution view of the specimen, which makes it particularly suitable for precise measurements. In addition, it is non-destructive, greatly more time-efficient and requires only minimal sample processing (Elfarnawany et al., 2017; Iyer et al., 2018; Schart-Morén et al., 2019). In recent work, the first SRµCT data on human temporal bones implanted with a clinically representative CI electrode were presented, which showed promising results in terms of visualizing intra-cochlear structures (Rohani et al., 2018; Li et al., 2019). However, further investigation is still necessary to assess the benefits SRµCT could provide over conventional methods.
Up to now, only cochleas implanted with lateral wall CI-electrode arrays, which incorporate relatively thin electrodes, have been investigated. Here, we present data from our recent experiment where we tested perimodiolar electrode arrays. These types of electrode arrays are commonly used in CI surgery, as they are pre-curved in the fabrication process to take a position closer to the auditory neurons in the cochlear medial wall, which is, in turn, assumed to offer the benefit of improved audiological outcome. However, such electrodes also typically incorporate relatively thick platinum contacts that are more closely spaced together, which poses a major challenge even for SRµCT imaging. The aim of this study was to develop an experimental protocol for visualizing perimodiolar electrode arrays in situ with SRµCT and to assess whether it would be possible to measure the distance from the surface of the electrode contacts to neural structures.
2. Materials and methods
2.1. Sample preparation
A total of eight human temporal bones were used in this study. All samples were excised from cadaveric head specimens that were donated to the European Institute for Otorhinolaryngology in Antwerp (Belgium) for scientific and educational use. The temporal bones were fixated in a 4% formaldehyde-buffer solution. To accommodate the size restrictions imposed by the experimental setup employed, samples were trimmed manually with a drill bit to include only the cochleovestibular system, leaving a few millimetres of surrounding bone. This process resulted in a set of specimens that were each approximately 20–30 mm in length and 20–25 mm in diameter (Fig. 1). Prior to tomographic scanning, the samples were dehydrated for approximately 12 h in air. Draining of the liquid inside the cochlea was achieved by carefully placing the samples on a piece of paper tissue with the round window oriented downward. No additional pre-processing steps such as decalcification or sample staining were performed.
2.2. Cochlear implantation
Seven out of eight temporal bone samples were implanted with a CI-electrode array and one specimen was left unimplanted. The implantations were performed by an otorhinolaryngologist with over 20 years of experience with cochlear implantation. Two different types of CI-electrodes were used: (i) a relatively rigid perimodiolar electrode array (Contour Advance, CI512, Cochlear Corporation); and (ii) its successor, a slim, flexible perimodiolar electrode array (Slim Modiolar, CI532, Cochlear Corporation). The semi-ring-shaped electrode contacts are approximately 0.3 mm in width and 0.4–0.7 mm in length, depending on the location along the array and the type. A metal alloy with a ratio of 90% platinum to 10% iridium is commonly used for CI-electrodes (Zeng et al., 2008). For one sample, the surgeon had intentionally misplaced the electrode, simulating a scala transition from the scala tympani into the scala vestibuli. To ensure minimal movement of the electrode during transportation, the electrode was cut off at the tail and mechanically fixed within a receptacle drilled in the sinus tympani.
2.3. Synchrotron radiation computed microtomography scans
The SRµCT experiments were performed at the SYRMEP beamline of the Elettra synchrotron facility (Basovizza, Trieste, Italy). This beamline provides, with a bending magnet source at a distance of about 24 m from the sample stage, a laminar-section X-ray beam with a maximum area of approximately 160 mm (h) × 5 mm (v) (Tromba et al., 2010). For the present experiment, the electron storage ring was operating at 2.0 GeV, and a filtered polychromatic X-ray beam (filters: 1.5 mm Si + 1.0 mm Al, mean X-ray energy ca. 27 keV) was used to obtain X-rays with a high allowing us to reduce the scan duration of the µCT experiments when working at high spatial resolution (pixel size <5 µm) and for X-ray energies >20 keV (Goyens et al., 2018). In order to capture the entire region of interest, two scans per sample were acquired at different vertical positions and then the images were vertically stitched.
Tomographic scans were acquired in propagation-based phase-contrast mode (Cloetens et al., 1997). A total of 1800 projections were acquired over a total scan angle of 360°, setting the sample off-centre in the field of view (FOV) of the detector (extended horizontal FOV) with an exposure time per projection of 1.5 s. The detector was a water-cooled, 16 bit, sCMOS camera (Hamamatsu C11440-22 C) with a 2048 × 2048 pixel chip coupled with a LSO:Tb scintillator screen (45 µm-thick on top of a 170 µm-thick LSO substrate) through high numerical aperture optics. The effective pixel size was set at 4.95 µm, yielding a maximum FOV of about 20 mm (h) × 4 mm (v). The sample-to-detector distance was set to 0.91 m.
The microtomographic axial slices were reconstructed using the SYRMEP Tomo Project (STP) software suite, custom-developed at Elettra (Brun et al., 2015), applying the filtered back-projection algorithm (Herman, 1980) and then were converted to 16-bit tiff format. The virtual slices were visualized through the Fiji software (Schindelin et al., 2012).
Prior to tomographic reconstruction, a single-distance phase-retrieval algorithm, based on the transport of intensity equation (Paganin et al., 2002) was applied to the projection images. We set a value of 200 for the δ/β = γ parameter (ratio between the real and imaginary parts of the complex refractive index) after a manual optimization to enhance the contrast of fine anatomical microstructures while minimizing the image blurring. In the supporting information we report some examples of axial slices reconstructed without phase-retrieval and after phase-retrieval using Paganin's algorithm with different γ ratios. These images show how the γ parameter used in this study was optimized and how this affected the image quality. Fig. S7 of the supporting information shows that after application of phase-retrieval, a fivefold improvement could be observed in the contrast-to-noise ratio (Goyens et al., 2018). Application of the algorithm also did not affect the observed thickness of the structures of interest (Figs. S3 and S5).
2.4. Comparator imaging techniques
To evaluate the improvements of SRµCT over clinical imaging as well as over conventional ex vivo imaging, cone-beam computed tomography (CBCT) and conventional X-ray microfocus tomography (XµCT) were performed on the same specimens. Clinical CBCT scans were acquired with a NewTom 5G XL CBCT scanner (NewTom, Verona, Italy) and XµCT scans were captured using a SkyScan 1172 scanner (Bruker micro-CT, Kontich, Belgium). The scan parameters used for the CBCT and XµCT experiments are summarized in Tables 1 and 2, respectively.
|
|
3. Results and discussion
For all eight temporal bones, SRµCT successfully visualized the intra-cochlear anatomical structures of interest (Fig. 2). The locations of the cochlear scalae, the osseous spiral lamina, the basilar membrane and Rosenthal's canal were evident in the final reconstructions. In addition, the Reissner's membrane (RM), the thinnest membrane of interest which consists of two epithelial cell layers could be visualized. The spatial resolution of the current experimental setup was evaluated directly from the tomographic slices by considering the thinnest region of RM whose thickness could reliably be measured, and was estimated at approximately 10 µm. The thickness of the RM was measured on a tomographic slice at roughly 2–4 pixels, or about 10–20 µm. Previous studies that employed either histological sectioning (Shibata et al., 2009) or (Fraissinette et al., 1993) to evaluate the structural properties of the RM reported a thickness in the approximate range 4–12 µm. Based on the values in the literature, we conclude that the 10 µm spatial resolution is insufficient to determine the thickness of the membrane. The RM was only found to be present in two of the specimens. Because the membrane is a delicate structure, it was possibly damaged during the processing and fixation of the other samples. In future studies, shortening of the fixation time and employing fresh temporal bones is expected to allow for better preservation of this structure.
Streak artefacts due to the highly absorbing electrode contacts were also observed in the reconstructed images of the implanted samples. However, despite the erroneous voxel intensities in close proximity to the electrode array, it remained possible to identify the location of the intra-cochlear structures with respect to the electrode contacts. Although automatic threshold-based segmentation cannot be achieved with the present data, the image quality allows for precise manual measurements of the electrode locations, which was the main objective of the current study. This is a distinct advantage over conventional XµCT, where in our case the severity of the streaking artefacts was found to be substantially higher (Fig. 3).
The volumetric SRµCT image data were rigidly co-registered to the data acquired from the clinical CBCT system with proprietary image-processing software (MATLAB image-processing toolbox, Mathworks, NA). The aligned slices are shown in Fig. 4. Fig. 4(a) demonstrates a typical atraumatic insertion of the CI-electrode into the scala tympani. Clinical CBCT was unable to visualize the internal structures of the cochlea, whereas synchrotron-based imaging could clearly reveal the separation between the cochlear scalae, thus confirming that the electrode array was positioned fully in the scala tympani. Fig. 4(b) instead shows an example of the traumatic dislocation of the CI-electrode in the sample, which had been intentionally induced by the surgeon by misorienting the electrode array during the implantation. SRµCT images showed that the electrode array had translocated into the scala vestibuli, by both rupturing and displacing the basilar membrane.
4. Conclusions and outlook
The results presented in this report demonstrate that, by using propagation-based phase-contrast synchrotron X-ray microtomography, it is possible to visualize both the electrode contacts and the intra-cochlear anatomical structures simultaneously, even when imaging cochleas implanted with a perimodiolar electrode array. However, the presence of the highly absorbing, closely spaced platinum–iridium contacts still resulted in significant streak artefacts that could interfere with the accurate analysis of the CI-electrode positions. Therefore, our future studies will focus on further reducing the contribution of the CI-related metal artefact. In particular, we will investigate the capabilities of synchrotron-based imaging in monochromatic beam mode and at higher beam energies, ca. 40 keV, in order to improve the X-ray transmission while maximally preserving the soft tissue contrast within the specimen. We will also focus on the development of an application-specific tomographic reconstruction algorithm to reduce the metal artefact.
Supporting information
Examples of images obtained by phase-contrast synchrotron-based computed microtomography reconstructed without and after the application of the single-distance phase-retrieval Paganin's algorithm. DOI: https://doi.org/10.1107/S1600577520014952/rv5138sup1.pdf
3D animation (refer to supporting PDF file). DOI: https://doi.org/10.1107/S1600577520014952/rv5138sup2.avi
Acknowledgements
We acknowledge Elettra Sincrotrone Trieste for providing access to its synchrotron radiation facilities and for financial support. We also thank Franco Zanini and Alessio Veneziano for their assistance in using the SYRMEP beamline. Finally, we thank the two anonymous referees for their helpful comments and suggestions.
Funding information
The following funding is acknowledged: Fonds Wetenschappelijk Onderzoek (grant No. 12R5118N to JG); Hercules Foundation (grant No. UABR/11/004).
References
Brun, F, Pacilè, S, Accardo, A., Kourousias, G,, Dreossi, D., Mancini, L., Tromba, G. & Pugliese, R. (2015). Fund. Inform. 141, 233–243. Google Scholar
Chakravorti, S., Noble, J. H., Gifford, R. H., Dawant, B. M., O'Connell, B. P., Wang, J. & Labadie, R. F. (2019). Otol. Neurotol. 40, 617–624. CrossRef PubMed Google Scholar
Cloetens, P., Pateyron-Salomé, M., Buffière, J. Y., Peix, G., Baruchel, J., Peyrin, F. & Schlenker, M. (1997). J. Appl. Phys. 81, 5878–5886. CrossRef CAS Web of Science Google Scholar
De Chaumont, F., Dallongeville, S., Chenouard, N., Hervé, N., Pop, S., Provoost, T., Meas-Yedid, V., Pankajakshan, P., Lecomte, T., Le Montagner, Y., Lagache, T., Dufour, A. & Olivo-Marin, J. C. (2012). Nat. Methods, 9, 690–696. CrossRef CAS PubMed Google Scholar
Elfarnawany, M., Alam, S. R., Rohani, S. A., Zhu, N., Agrawal, S. K. & Ladak, H. M. (2017). J. Microsc. 265, 349–357. CrossRef CAS PubMed Google Scholar
Fraissinette, A. de, Felix, H., Hoffmann, V., Johnsson, L. G. & Gleeson, M. J. (1993). Orl, 55, 68–72. PubMed Google Scholar
Goyens, J., Vasilopoulou-Kampitsi, M., Claes, R., Sijbers, J. & Mancini, L. (2018). J. Anat. 233, 770–782. CrossRef CAS PubMed Google Scholar
Herman, G. T. (1980). Image Reconstruction from Projections: The Fundamentals of Computerized Tomography. New York: Academic Press. Google Scholar
Holden, L. K., Finley, C. C., Firszt, J. B., Holden, T. A., Brenner, C., Potts, L. G., Gotter, B. D., Vanderhoof, S. S., Mispagel, K., Heydebrand, G. & Skinner, M. W. (2013). Ear Hear. 34, 342–360. CrossRef PubMed Google Scholar
Iso-Mustajärvi, M., Matikka, H., Risi, F., Sipari, S., Koski, T., Willberg, T., Lehtimäki, A., Tervaniemi, J., Löppönen, H. & Dietz, A. (2017). Otol. Neurotol. 38, e327–e334. PubMed Google Scholar
Iyer, J. S., Zhu, N., Gasilov, S., Ladak, H. M., Agrawal, S. K. & Stankovic, K. M. (2018). Biomed. Opt. Expr. 9, 3757. CrossRef Google Scholar
Johnson, S. B., Cureoglu, S., O'Malley, J. T. & Santi, P. A. (2014). Otol. Neurotol. 135, 1. Google Scholar
Kawano, A., Seldon, H., Clark, G. M., Ramsden, R. T. & Raine, C. H. (1998). Acta Otolaryngol. 118, 313–326. CAS PubMed Google Scholar
Li, H., Schart-Morén, N., Rohani, S. A., Ladak, H. M., Rask-Andersen, H. & Agrawal, S. (2019). Ear Hear. 41, 173–181. CrossRef CAS Google Scholar
Paganin, D., Mayo, S. C., Gureyev, T. E., Miller, P. R. & Wilkins, S. W. (2002). J. Microsc. 206, 33–40. Web of Science CrossRef PubMed CAS Google Scholar
Rau, T. S., Würfel, W., Lenarz, T. & Majdani, O. (2013). Int. J. CARS, 8, 481–509. CrossRef Google Scholar
Rohani, S. A., Iyaniwura, J. E., Zhu, N., Agrawal, S. K. & Ladak, H. M. (2018). J. Microsc. 273, 127–134. CrossRef PubMed Google Scholar
Saeed, S. R., Selvadurai, D., Beale, T., Biggs, N., Murray, B., Gibson, P., Risi, F. & Boyd, P. (2014). Otol. Neurotol. 35, 1338–1344. CrossRef PubMed Google Scholar
Schart-Morén, N., Agrawal, S. K., Ladak, H. M., Li, H. & Rask-Andersen, H. (2019). Ear Hear. 40, 393–400. PubMed Google Scholar
Schindelin, J., Arganda-Carreras, I., Frise, E., Kaynig, V., Longair, M., Pietzsch, T., Preibisch, S., Rueden, C., Saalfeld, S., Schmid, B., Tinevez, J. Y., White, D. J., Hartenstein, V., Eliceiri, K., Tomancak, P. & Cardona, A. (2012). Nat. Methods, 9, 676–682. Web of Science CrossRef CAS PubMed Google Scholar
Shaul, C., Dragovic, A. S., Stringer, A. K., O'Leary, S. J. & Briggs, R. J. (2018). J. Laryngol. Otol. 132, 1000–1006. CrossRef CAS PubMed Google Scholar
Shibata, T., Matsumoto, S., Agishi, T. & Nagano, T. (2009). Am. J. Otolaryngol. 30, 112–120. CrossRef PubMed Google Scholar
Sun, Q., Chang, K. H., Dormer, K. J., Dyer, R. K. & Gan, R. Z. (2002). Med. Eng. Phys. 24, 595–606. CrossRef PubMed CAS Google Scholar
Tromba, G., Longo, R., Abrami, A., Arfelli, F., Astolfo, A., Bregant, P., Brun, F., Casarin, K., Chenda, V., Dreossi, D., Hola, M., Kaiser, J., Mancini, L., Menk, R. H., Quai, E., Quaia, E., Rigon, L., Rokvic, T., Sodini, N., Sanabor, D., Schultke, E., Tonutti, M., Vascotto, A., Zanconati, F., Cova, M., Castelli, E. & Siu, K. K. W. (2010). AIP Conf. Proc. 1266, 18–23. CrossRef Google Scholar
Zeng, F. G., Rebscher, S., Harrison, W. V., Sun, X. & Feng, H. (2008). Biomed. Eng. 1, 115–142. Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.