Figure 9
Scattering for all analyzers as measured using a single scan from Tempax glass. The data points are in blue, the red line shows the total fit, and the black lines the derived resolution. Green numbers indicate momentum transfer in nm−1 while other numbers give the full width at 1/2, 1/10 and 1/100 maximum. All the resolutions from each row are collected in the far right column. (A software error during this scan prevented a41 from being counted so a40 is presented to fill the graph layout.) The detector/analyzer alignment is deliberately chosen so that the analyzers in the bottom row have the smallest deviation from backscattering and therefore the best resolution (see Fig. 2[link]).