Practical measurement of the energy resolution for meV-resolved inelastic X-ray scattering

crossmark logo

aMaterials Dynamics Laboratory, RIKEN SPring-8 Center, 1-1-1 Kouto, Sayo, Hyogo 679-5148, Japan
*Correspondence e-mail: disikawa@spring8.or.jp, baron@spring8.or.jp

Edited by Y. Amemiya, University of Tokyo, Japan (Received 18 January 2021; accepted 26 March 2021; online 9 April 2021)

Several different ways of measuring the energy resolution for meV-resolved inelastic X-ray scattering (IXS) are compared: using scattering from poly(methyl methacrylate), PMMA, using scattering from borosilicate glass (Tempax), and using powder diffraction from aluminium. All of these methods provide a reasonable first approximation to the energy resolution, but, also, in all cases, inelastic contributions appear over some range of energy transfers. Over a range of ±15 meV energy transfer there is good agreement between the measurements of PMMA and Tempax at low temperature, and room-temperature powder diffraction from aluminium, so we consider this to be a good indication of the true resolution of our ∼1.3 meV spectrometer. The resolution over a wider energy range is self-consistently determined using the temperature, momentum and sample dependence of the measured response. The inelastic contributions from the PMMA and Tempax, and their dependence on momentum transfer and temperature, are then quantitatively investigated. The resulting data allow us to determine the resolution of our multi-analyzer array efficiently using a single scan. The importance of this procedure is demonstrated by showing that the results of the analysis of a spectrum from a glass are changed by using the properly deconvolved resolution function. The impact of radiation damage on the scattering from PMMA and Tempax is also discussed.

1. Introduction

Non-resonant inelastic X-ray scattering (IXS) is an accepted method of investigating atomic dynamics at ∼nm−1 momentum transfers with ∼meV resolution, as is useful for studying phonons in crystals, and excitations in liquids and glasses (Dorner et al., 1986[Dorner, B., Burkel, E. & Peisl, J. (1986). Nucl. Instrum. Methods Phys. Res. A, 246, 450-451.]; Burkel et al., 1987[Burkel, E., Peisl, J. & Dorner, B. (1987). Europhys. Lett. 3, 957-961.]; Sette et al., 1995[Sette, F., Ruocco, G., Krisch, M., Bergmann, U., Masciovecchio, C., Mazzacurati, V., Signorelli, G. & Verbeni, R. (1995). Phys. Rev. Lett. 75, 850-853.]). IXS provides relatively clean data with some advantages over other techniques [see discussion by Baron (2016[Baron, A. Q. R. (2016). Synchrotron Light Sources and Free-Electron Lasers: Accelerator Physics, Instrumentation and Science Applications, edited by E. Jaeschke, S. Khan, J. R. Schneider & J. B. Hastings, pp. 1643-1757. Cham: Springer International Publishing.])] for measuring small (mm to µm scale) samples, and for measuring liquids and glasses. Due to the relatively high incident X-ray energy (∼20000 eV) compared with the energy transfers of interest (typically <0.1 eV), the momentum and energy resolution for IXS are decoupled, so the influence of the finite energy resolution of a spectrometer can, in principle, be simulated by a straightforward one-dimensional convolution of the energy resolution with a model. However, that requires an accurate measurement of the energy resolution. Various methods of measuring the resolution function have been mentioned (Masciovecchio et al., 1996a[Masciovecchio, C., Bergmann, U., Krisch, M., Ruocco, G., Sette, F. & Verbeni, R. (1996a). Nucl. Instrum. Methods Phys. Res. B, 117, 339-340.]; Sinn et al., 2001[Sinn, H., Alp, E. E., Alatas, A., Barraza, J., Bortel, G., Burkel, E., Shu, D., Sturhahn, W., Sutter, J. P., Toellner, T. S. & Zhao, J. (2001). Nucl. Instrum. Methods Phys. Res. A, 467-468, 1545-1548.]; Baron et al., 2001[Baron, A. Q. R., Tanaka, Y., Miwa, D., Ishikawa, D., Mochizuki, T., Takeshita, K., Goto, S., Matsushita, T., Kimura, H., Yamamoto, F. & Ishikawa, T. (2001). Nucl. Instrum. Methods Phys. Res. A, 467-468, 627-630.]; Verbeni et al., 2005[Verbeni, R., Kocsis, M., Huotari, S., Krisch, M., Monaco, G., Sette, F. & Vanko, G. (2005). J. Phys. Chem. Solids, 66, 2299-2305.]; Said et al., 2011[Said, A. H., Sinn, H. & Divan, R. (2011). J. Synchrotron Rad. 18, 492-496.]; Shinohara et al., 2019[Shinohara, Y., Dmowski, W., Iwashita, T., Ishikawa, D., Baron, A. Q. R. & Egami, T. (2019). Phys. Rev. Mater. 3, 065604.]), but they usually involve the assumption that the scattering from some material, often a plastic, is a purely elastic, delta-function-like, line at zero energy transfer. This assumption, while approximately true, and sufficient for data analysis in some cases, is wrong in principle, as all materials have some dynamical response at finite energy transfer as is guaranteed on the meV scale, in a Born approximation limit, by the first-moment sum rule of Placzek (1952[Placzek, G. (1952). Phys. Rev. 86, 377-388.]). As the quality of IXS measurements then improves, there is a greater need for understanding and carefully determining the instrumental energy resolution for meV IXS. This is especially important for the analysis of data from samples where inelastic excitations appear near a large (quasi-)elastic line (such as in glasses, or imperfect crystals, or crystals with longer range density modulations), but is also expected to be more generally useful, to, for example, investigate weak excitations that may appear on the tails of stronger excitations.

In order to measure the energy resolution, one would, ideally, like to uniformly illuminate the full analyzer solid angle using elastic scattering in a geometry very similar to that used to measure a sample (see Fig. 1[link]): this gives the best indication of the energy resolution of the spectrometer for a real measurement. However, the two requirements, pure elastic scattering and illumination over the full analyzer solid angle, are not easily satisfied simultaneously. Bragg scattering from a single crystal can provide an extremely favorable ratio of elastic to inelastic scattering, and effectively remove the inelastic components over a small solid angle. However, the solid angle is badly mis-matched to typical IXS analyzer acceptance: Bragg reflections have angular widths of ∼0.01 mrad, or less, and the analyzer solid angle is often 1 to 10 mrad. Therefore, using Bragg reflection from a single crystal is not a practical way to measure the resolution of a mrad-acceptance analyzer. More commonly one uses a dis­ordered material measured at the momentum transfer corresponding to the structure factor (SF) maximum, also sometimes called the first sharp diffraction peak. At this position one expects the scattering to be approximately elastic. Further, the angular width of the SF maximum is usually large enough so that the scattering results in relatively uniform illumination over an analyzer.1 Previous work has mentioned different samples used to measure the resolution including `plastic' (Masciovecchio et al., 1996a[Masciovecchio, C., Bergmann, U., Krisch, M., Ruocco, G., Sette, F. & Verbeni, R. (1996a). Nucl. Instrum. Methods Phys. Res. B, 117, 339-340.]), more specifically poly(methyl methacrylate) or PMMA/plexiglass (Sinn et al., 2001[Sinn, H., Alp, E. E., Alatas, A., Barraza, J., Bortel, G., Burkel, E., Shu, D., Sturhahn, W., Sutter, J. P., Toellner, T. S. & Zhao, J. (2001). Nucl. Instrum. Methods Phys. Res. A, 467-468, 1545-1548.]; Baron et al., 2001[Baron, A. Q. R., Tanaka, Y., Miwa, D., Ishikawa, D., Mochizuki, T., Takeshita, K., Goto, S., Matsushita, T., Kimura, H., Yamamoto, F. & Ishikawa, T. (2001). Nucl. Instrum. Methods Phys. Res. A, 467-468, 627-630.]; Verbeni et al., 2005[Verbeni, R., Kocsis, M., Huotari, S., Krisch, M., Monaco, G., Sette, F. & Vanko, G. (2005). J. Phys. Chem. Solids, 66, 2299-2305.]) or Kr gas (Sette et al., 1995[Sette, F., Ruocco, G., Krisch, M., Bergmann, U., Masciovecchio, C., Mazzacurati, V., Signorelli, G. & Verbeni, R. (1995). Phys. Rev. Lett. 75, 850-853.]), or borosilicate glass (Shinohara et al., 2019[Shinohara, Y., Dmowski, W., Iwashita, T., Ishikawa, D., Baron, A. Q. R. & Egami, T. (2019). Phys. Rev. Mater. 3, 065604.]) usually at room temperature, and, in all cases, without discussion of the inelastic contributions.

[Figure 1]
Figure 1
Schematic showing the setup/geometry with incident beam focused onto the sample position, and scattering into a single analyzer. The analyzer is usually at an angle of 2θ ≃ 5° out of the vertical plane containing the incident beam. (Not to scale.)

Here we investigate the inelastic contributions for scattering from two disordered materials that are interesting for resolution measurements, PMMA and Tempax glass. We compare these measurements with the use of powder diffraction from a polycrystalline aluminium sample which illuminates a line on the analyzer: by moving the analyzer so that different portions are illuminated, we use the powder scattering to estimate the resolution. We find good agreement of diffraction, and measurements of PMMA and Tempax at low temperature, leading us to believe our resolution determination is good. We then quantitatively determine the inelastic contributions to the PMMA and Tempax response. Using those data we can efficiently determine the energy resolution for our array of analyzers using a single scan of a sample at room temperature. The latter is important as IXS spectrometers employ an array of multiple (4 to 28, depending on the instrument) analyzers so that it is highly advantageous if the resolution measurement can be made in parallel over multiple analyzers at one time. We show how our procedure improves the fit (and changes the best fit parameters) for a low-Q spectrum from a glass.

2. Experimental setup, samples and methods

2.1. Spectrometer

All measurements were carried out at the high-resolution spectrometer at the RIKEN Quantum NanoDynamics beamline, BL43LXU (Baron, 2010[Baron, A. Q. R. (2010). SPring-8 Inf. Newsl. 15, 14-19.], 2016[Baron, A. Q. R. (2016). Synchrotron Light Sources and Free-Electron Lasers: Accelerator Physics, Instrumentation and Science Applications, edited by E. Jaeschke, S. Khan, J. R. Schneider & J. B. Hastings, pp. 1643-1757. Cham: Springer International Publishing.]), of the RIKEN SPring-8 Center. The beam was focused onto the sample by a bent cylindrical mirror to a spot size of about 50 µm and divergence of ∼0.2 mrad (V) × 0.5 mrad (H) [unless otherwise specified, sizes and spectral widths will be given as the full width at half maximum (FWHM)]. The BL43LXU spectrometer is designed for an array of up to 42 analyzers, but, in the present work, only the central 28 were active (see Fig. 2[link]) in a 4 × 7 array at a 9.8 m radius from the sample. The analyzer center-to-center spacing is 120 mm and each analyzer is pixelated (1 mm pitch) and operates in a slightly off-Roland geometry with a temperature gradient applied to optimize the resolution (Ishikawa et al., 2015[Ishikawa, D., Ellis, D. S., Uchiyama, H. & Baron, A. Q. R. (2015). J. Synchrotron Rad. 22, 3-9.]). By deliberate arrangement, not all analyzers have the same deviation from backscattering so there are systematic changes in the FWHM of the energy resolution [see Fig. 2[link] and Ishikawa et al. (2015[Ishikawa, D., Ellis, D. S., Uchiyama, H. & Baron, A. Q. R. (2015). J. Synchrotron Rad. 22, 3-9.])]. In addition, analyzer fabrication has process variations so that not all analyzers are of the same quality. Thus, different analyzers have different responses so it is desirable to measure the resolution of each analyzer for each experimental run, and for each setting of the slits used to define the analyzer acceptance/momentum resolution. It is then important that such measurements be efficient.

[Figure 2]
Figure 2
Schematic of the analyzer array setup at BL43LXU. (Not to scale.)

2.2. Energy scans

Energy scans are performed by holding the analyzer temperatures stable (typical variation < 0.12 mK r.m.s.) and continuously scanning the temperature of the backscattering monochromator. The energy transfer is then determined from the measured temperatures using the thermal expansion of silicon (approximately 17 mK meV−1 near room temperature at 21.7 keV) as calibrated by a known phonon line [see discussion by Fukui et al. (2008[Fukui, H., Katsura, T., Kuribayashi, T., Matsuzaki, T., Yoneda, A., Ito, E., Kudoh, Y., Tsutsui, S. & Baron, A. Q. R. (2008). J. Synchrotron Rad. 15, 618-623.])]. Typical scan speeds vary from 10 to 30 mK min−1, with sampling times of 10 to 3 s. For the resolution scans presented here, the data are sorted into 0.15 meV-wide bins near the elastic energy, though for experimental data, or sometimes, at larger energy transfers, larger bins (0.2 to 2 meV) are used to improve the statistical quality (for fitting of spectra, of course, the resolution should be binned with the same bin size as the data to avoid introducing systematic errors in fitted linewidths). The data processing (the `icscan' program) includes corrections to compensate for small changes in the incident power on the backscattering crystal2 and for scan direction3. For the present work, the beam was focused to a ∼50 µm spot using a bent cylindrical mirror 5 m upstream of the sample position. The position of the beam at the sample is held stable to better than 15 µm using feedback from a quadrant beam position monitor [see Baron et al. (2019[Baron, A. Q. R., Ishikawa, D., Fukui, H. & Nakajima, Y. (2019). AIP Conf. Proc. 2054, 020002.])]. We note, for completeness, that the bandwidth of the incident beam using the Si(11 11 11) back-reflection monochromator at 21.7 keV is calculated to be ∼0.8 meV, though we only measure the resolution function of the combined system of the analyzer and the monochromator.

2.3. Analyzer slits

The analyzer acceptance, or solid angle, is determined by a set of venetian-blind slits located about 0.9 m upstream of the analyzer crystals (or about 8.9 m from the sample position). The maximum aperture is 85 mm (V) × 80 mm (H), while the smallest gap is ∼2.5 mm × 2.5 mm, though in practical experiments they have never been used with gaps smaller than 4 mm × 4 mm, and more typical values are 20 mm × 20 mm to fully open. These slits have one motor to control all the gaps in the vertical and one for the horizontal, and in the usual operating configuration all analyzers have the same slit sizes [it is also possible to mount masks independently to set each analyzer aperture (Baron et al., 2019[Baron, A. Q. R., Ishikawa, D., Fukui, H. & Nakajima, Y. (2019). AIP Conf. Proc. 2054, 020002.]), but this is not done in the present work]. All measurements presented here were made with slit sizes of 80 mm (V) × 40 mm (H), unless stated otherwise.

2.4. Samples

The samples used included commercially available poly(methyl methacrylate), PMMA, borosilicate glass (Tempax), and a high-purity polycrystalline aluminium plate. Radiation damage of the PMMA was an issue, with noticeable changes in the structure factor occurring on ∼hour time scales in a beam of 4 × 1010 s−1 into a 50 µm spot. Fig. 3[link] shows the progression of a series of I(Q) measurements made one night. [I(Q) is used to indicate that it is the measured intensity without correction for atomic form factors, or for polarization.] The change in response is clear. After X-ray exposure, visual inspection showed structure (looking, sometimes, like a trail of bubbles) in the PMMA along the beam path. Therefore, all resolution measurements using PMMA are made while scanning the sample through the incident beam, typically at a rate of 0.05 mm/5 minutes. This preserves the structure factor. No equivalent issue was found for the Tempax despite >24 h of illumination: color centers did form, leading to brown dis­coloration, but the structure factor was not noticeably affected as is clear in Fig. 3[link](b). Fig. 3[link](c) shows the result of 2θ scans of one analyzer held at zero energy transfer: similar peak rates of ∼6 kHz are obtained at the SF maximum for 2 mm PMMA and for 1.5 mm Tempax. For pristine PMMA the SF maximum occurs at 9.6 nm−1 and the peak width is about 5.9 nm−1, while for the Tempax glass the SF maximum is at 15.4 nm−1 and the peak width is about 7.2 nm−1.

[Figure 3]
Figure 3
I(Q) for (a) PMMA and (b) Tempax, both at room temperature. The PMMA (Tempax) measurements were made with an incident beam intensity of 4 × 1010 (2 × 1010) 21.747 keV photons s−1 into a 50 µm-diameter spot. (c) Measured intensity from one analyzer set at zero energy transfer from a 2 mm-thick sample of PMMA and a 1.5 mm-thick sample of Tempax. [2 × 1010 s−1, 1.3 meV resolution at 21.747 keV, analyzer acceptance set to 4.5 mrad (H) × 9 mrad (V).] The notation `0 h', `24 h', etc. indicate the time (in hours) of integrated illumination.

3. Resolution measurement

In order to determine the resolution of our system, we measured the analyzer response under the following conditions:

(i) Using the (111) powder diffraction ring from a sample of pure aluminium, 1.2 mm thick. This was measured at five different 2θ angles, effectively illuminating the analyzer surface.

(ii) Using a 2 mm-thick PMMA sample at 11 and 300 K measured at the SF maximum.

(iii) Using a 1.5 mm-thick Tempax sample at 20, 300 and 500 K measured at the SF maximum.

The excellent agreement, see Fig. 4[link], of the low-T measurements of Tempax and PMMA and the 300 K Al powder diffraction over the range of ±15 meV leads us to conclude that this accurately represents the resolution function. However, beyond this range, various inelastic contributions begin to appear in the spectra, and are different for each material.

[Figure 4]
Figure 4
Comparison of the direct measurement of the resolution using PMMA at low temperature, Tempax at low temperature and powder diffraction at the Al (111) diffraction ring, shown on (a) linear and (b) log scales.

At room temperature, both the PMMA and the Tempax show inelastic contributions, albeit small, at the scale of a few meV energy transfer. This can be seen in Fig. 5[link] (also Fig. 8) and numerical values can be found in Table 1[link]. However, noting that there are not significant structural changes in these materials with temperature, we fit the results at 11 K and 300 K with a common set of inelastic excitations scaled by the single phonon Bose factor. This allows us to separate the inelastic contributions from the sample from the elastic scattering, and then to self-consistently determine the resolution. Table 1[link] indicates the contribution of inelastic excitations integrated over the indicated energy ranges and sample conditions at structure factor maximum.

Table 1
Fraction of the total scattering from the sample that is inelastic at the position of the SF maximum for PMMA and TEMPAX at the indicated temperatures

  PMMA   TEMPAX
Energy range [min, max] (meV) 11 K (%) 300 K (%)   20 K (%) 300 K (%) 500 K (%)
Total 0.48 4.86   0.24 1.01 1.80
[−10, +10] 0.30 4.41   0.07 0.66 1.24
[−20, −10], [+10, +20] 0.04 0.25   0.04 0.16 0.30
[+20, +40] 0.04 0.07   0.03 0.05 0.09
[+40, +60] 0.04 0.05   0.03 0.04 0.05
[+60, +120] 0.04 0.05   0.04 0.05 0.06
[Figure 5]
Figure 5
Measured scattering from (a) PMMA and (b) TEMPAX at the indicated temperatures (points) compared with the derived resolution function described in the text (green line). The inelastic contribution (convolved with the resolution function) is given by the blue lines.

4. Comparison with ray-tracing

We compare our measured resolution for this particular analyzer with a ray-tracing calculation assuming dynamical diffraction from perfect thick silicon crystals [the same code as used by Ishikawa et al. (2015[Ishikawa, D., Ellis, D. S., Uchiyama, H. & Baron, A. Q. R. (2015). J. Synchrotron Rad. 22, 3-9.])]. Examining Fig. 6[link], it is clear that the measured resolution is slightly worse than calculated, indicating that imperfections are introduced in the fabrication process (previous tests of flat crystals suggest that the intrinsic silicon quality is not an issue at this level). This is unsurprising given long experience with analyzer-to-analyzer variations. It also is consistent with the work of Said et al. (2011[Said, A. H., Sinn, H. & Divan, R. (2011). J. Synchrotron Rad. 18, 492-496.]) where, for example, some analyzers were seen to have better resolution than others, and systematic improvements to resolution tails are possible by increasing the thickness of the silicon used. At the level of ∼1.3 meV resolution, spherical analyzer crystals have process variations in their fabrication that lead to a generally good, but still imperfect, response.

[Figure 6]
Figure 6
Comparison of the measured resolution with a ray-tracing calculation based on Ishikawa et al. (2015[Ishikawa, D., Ellis, D. S., Uchiyama, H. & Baron, A. Q. R. (2015). J. Synchrotron Rad. 22, 3-9.]). The FWHMs are nearly the same, but the tail is worse for the calculation. Insets show expanded views of the region near zero energy transfer on (a) log and (b) linear scales.

5. Inelastic excitations away from the SF maximum

The inelastic excitations in PMMA and Tempax were investigated by a series of measurements at different values of 2θ with the same analyzer crystal. The analyzer resolution is then constant, allowing the inelastic contribution from the sample to be extracted. Fig. 7[link] shows the fraction of the total scattering from −20 to +120 meV that is from inelastic excitations as a function of Q for the samples at different temperature: low temperatures, have, unsurprisingly, rather lower inelastic contributions, especially above the SF maximum. Meanwhile, Fig. 8[link] shows the energy-resolved plots with a low-energy excitation, and then optical bands that appear most strongly for PMMA at ∼47, 75 and 100 meV (380, 605 and 807 cm−1), reasonably consistent with peaks observed in IR spectroscopy.

[Figure 7]
Figure 7
Momentum transfer (Q) dependence of the inelastic contribution to the total scattering for the indicates samples and temperatures, expressed as a percentage of the integrated total intensity over an energy range of [−20, 120] meV for (a) PMMA and (b) Tempax. Note the highest momentum transfer in each case corresponds to the first minimum of the structure factor and in included for reference only: all practical resolution measurements are at momentum transfers close the SF maximum. See also Fig. 8[link] where the I(Q, ω) plane is shown.
[Figure 8]
Figure 8
I(Q, ω) at room temperature for (a) PMMA and (b) Tempax after subtraction of the elastic contribution. Expressed as a fraction of the intensity at zero energy transfer. (Note that the intensity has been corrected for the incident X-ray polarization so this plot is for σσ scattering.)

6. Efficient determination of the resolution

It is desirable to measure the resolution for each experiment.4 This provides the resolution in the conditions used in the experiment, confirms that the spectrometer is performing as expected, and gives a reference point for the evolution of spectrometer performance. However, it is also highly desirable that such measurements do not take a large amount of beam time: typical experimental runs are 3 to 8 days, so spending a few hours measuring the resolution function is reasonable, but to spend ∼1 day is too much. Therefore, after collecting data to determine a self-consistent resolution function at the SF maximum for one analyzer, we then went on to measure the response of the PMMA and Tempax over a wide range of momentum transfers, as is discussed in Section 5[link]. This allows us to determine the inelastic excitations in each material as a function of momentum transfer. Once this is known, then we can replace carrying out a longer series of measurements with each analyzer near the SF maximum with a single scan followed by a deconvolution to remove the inelastic excitations. This approach is demonstrated in Fig. 9[link] where we show the scattering from Tempax glass for all analyzers as measured from a single scan. The analyzer resolution is then extracted by fitting the measured response including the previously determined inelastic excitations in Tempax where the free parameters for the fit are the width, amplitude and center, of one Gaussian and a set of 6 to 10 Lorentzians used to simulate the resolution function, giving a smooth approximation to the resolution function (sarf). This allows the resolution to be determined for each analyzer, as shown by the black lines in the figure. One notes that there is some analyzer-to-analyzer fabrication process variation as is clearly visible in the right-hand column, where all resolutions for a given row of analyzers are collected.5

[Figure 9]
Figure 9
Scattering for all analyzers as measured using a single scan from Tempax glass. The data points are in blue, the red line shows the total fit, and the black lines the derived resolution. Green numbers indicate momentum transfer in nm−1 while other numbers give the full width at 1/2, 1/10 and 1/100 maximum. All the resolutions from each row are collected in the far right column. (A software error during this scan prevented a41 from being counted so a40 is presented to fill the graph layout.) The detector/analyzer alignment is deliberately chosen so that the analyzers in the bottom row have the smallest deviation from backscattering and therefore the best resolution (see Fig. 2[link]).

7. Comparison: PMMA versus Tempax for measuring the resolution

PMMA and Tempax may both be used to measure the resolution function. At room temperature the phonon contribution from Tempax is about 1% of the total scattered intensity while that of PMMA is about 5% near the SF maximum, which suggests Tempax may be preferable. In addition, the Tempax is less radiation sensitive than the PMMA. However, PMMA is easier to cut/machine into specific shapes, and attenuates the beam less for a given thickness. Given these considerations, we are now transitioning to using Tempax, except in cases where a specific shape may be needed to, for example, measure the resolution with the PMMA placed inside a specific sample cell.

8. Impact on analysis of a spectrum from a glass

As noted above, the impact of the corrections considered here are most significant when observing weak modes on the tails of strong modes, or near a strong elastic peak. This can be particularly significant for the IXS spectra of glasses at small momentum transfers, where it is not uncommon that the acoustic mode intensity is only a few percent of the elastic scattering. In Fig. 10[link] we show the measured spectrum of a silica-based glass at low momentum transfer (Q ≃ 3 nm−1). From general considerations, this spectrum is expected to be well approximated as the sum of a delta-function elastic peak and a damped harmonic oscillator (DHO) function for the acoustic mode, scaled by a one-phonon Bose factor, and convolved with the resolution (e.g. Masciovecchio et al., 1996[Masciovecchio, C., Ruocco, G., Sette, F., Krisch, M., Verbeni, R., Bergmann, U. & Soltwisch, M. (1996b). Phys. Rev. Lett. 76, 3356-3359.]b). We then fit the measured spectrum using that model, but taking the resolution function as either (a) the measured spectrum from Tempax at room temperature, or (b) the deconvolved (sarf) result from the process described in Section 6[link]. The results from minimizing (James & Roos, 1975[James, F. & Roos, M. (1975). Comput. Phys. Commun. 10, 343-367.]) χ2 are shown in the figure. As noted in the table inset in Fig. 10[link], the deconvolved (sarf) resolution gives a smaller χ2 and significantly different values for the acoustic mode width6 and intensity. Thus, the correction discussed here is important for analysis of the spectra of glasses, even when using the relatively good (see Sections 5[link] and 7[link]) response of Tempax to estimate the resolution.

[Figure 10]
Figure 10
Impact of different resolution functions on the fit to the spectrum from a glass sample. The points are the measured spectra for Tempax (red circles) and the glass sample (back squares) while the solid lines show the deconvolved resolution (sarf) and the fits and the contribution of the DHO acoustic modes to the fits. The inset gives the χ2 in 139 degrees of freedom and the optimized fit parameters with errors. See the text for discussion.

9. Conclusion

We have measured the resolution of our meV spectrometer over a range [−15, +15] meV and find essentially identical results for low-temperature PMMA, Tempax and powder diffraction from aluminium, giving us confidence that we know the real spectrometer resolution over this range. Meanwhile, the consistency of results at 20 to 500 K with a simple model suggests that the resolution is reasonably determined out to ∼50 meV, and comparison between different materials, and the assumption that the resolution function is smooth, then allows us some confidence that we know the resolution function out to 120 meV energy transfer, or more. The measured response is similar to ray-tracing calculation in the FWHM, but is somewhat worse in the tails. Scanning an analyzer over a wide Q range then allowed us to measure the relative contribution of inelastic excitations for PMMA and Tempax. With that information, we then can make an efficient and practical measurement of the resolution for a multi-analyzer system using a single scan of Tempax or PMMA at room temperature – deconvolution of the inelastic contribution gives a smooth approximation to the resolution function (sarf). The impact of this approach is seen to be significant for analysis of a spectrum from a glass at low Q, and is generally expected to be important whenever a weak mode is measured on the tail of stronger modes. We also have shown that PMMA, if used incautiously, shows radiation damage on ∼1 hour time scales even in our meV beam, while Tempax glass is relatively radiation hard.

Footnotes

1Sometimes small-angle scattering (SAXS) has also been used. However, using spherical analyzers at small scattering angles is tricky in a typical setup as it is easy to see scattering from a large volume around the sample, so that window and air scattering can cause significant background. Also, SAXS patterns usually change strongly with momentum transfer (e.g. they are often displayed on a log scale), so it may be difficult to achieve uniform illumination of a single analyzer and to measure more than one analyzer at a time.

2The temperature sensor attached to the backscattering crystal is located about 20 mm from the X-ray spot location and the power of the X-rays incident onto the backscattering crystal creates a temperature gradient between the X-ray beam and the sensor location. The temperature difference is then dependent on incident power load, so that we apply a linear correction that is typically good for incident intensity variations of about 20%, as may occur, for example, if upstream optics drift, or if the top-up of the storage ring is temporarily suspended (the typical stability is much better).

3This is put in as a `lag time' – assuming the temperature readout lags behind that at the X-ray spot by a small delay – typically 0.5 to 1 s. This correction then allows scanning in both directions: when this correction is included, peaks from scans upward or downward in energy transfer coincide to better than 0.02 meV.

4We sometimes make one measurement under standard conditions at the start of an experiment and then an additional measurement at the end, where the latter measurements are made with a reduced analyzer acceptance as may have been used during the experiment to improve momentum resolution.

5The resolution of a09 shows a `bump' at about 9 meV energy transfer. This can appear if the analyzer diffraction plane (Psi-rotation about the Bragg vector) is chosen such that one excites a multibeam condition, as can occasionally happen after the spectrometer and detector alignment is changed. This bump is consistent with 24-beam calculations of the full diffraction from the Si (11 11 11) back-reflection and integration over a range of Psi-angles consistent with the experiment (Sutter & Baron, unpublished[Sutter, J. P. & Baron, A. Q. R. (unpublished). Multibeam Calculations Near Backscattering at the Si(11 11 11) Reflection of Silicon.]).

6The reported width in the table is the relevant parameter from the DHO model which is approximately the half width at half-maximum (HWHM). See Masciovecchio et al. (1996b[Masciovecchio, C., Ruocco, G., Sette, F., Krisch, M., Verbeni, R., Bergmann, U. & Soltwisch, M. (1996b). Phys. Rev. Lett. 76, 3356-3359.]) and Baron (2016[Baron, A. Q. R. (2016). Synchrotron Light Sources and Free-Electron Lasers: Accelerator Physics, Instrumentation and Science Applications, edited by E. Jaeschke, S. Khan, J. R. Schneider & J. B. Hastings, pp. 1643-1757. Cham: Springer International Publishing.]).

Acknowledgements

These experiments were carried out at BL43LXU, The RIKEN Quantum NanoDynamics Beamline of the RIKEN SPring-8 Center. While some of this work began in 2014, the bulk of the work became possible due to beam time that became available during the disruption caused by the COVID-19 pandemic in 2020.

Funding information

The following funding is acknowledged: RIKEN SPring-8 Center.

References

First citationBaron, A. Q. R. (2010). SPring-8 Inf. Newsl. 15, 14–19.  Google Scholar
First citationBaron, A. Q. R. (2016). Synchrotron Light Sources and Free-Electron Lasers: Accelerator Physics, Instrumentation and Science Applications, edited by E. Jaeschke, S. Khan, J. R. Schneider & J. B. Hastings, pp. 1643–1757. Cham: Springer International Publishing.  Google Scholar
First citationBaron, A. Q. R., Ishikawa, D., Fukui, H. & Nakajima, Y. (2019). AIP Conf. Proc. 2054, 020002.  Google Scholar
First citationBaron, A. Q. R., Tanaka, Y., Miwa, D., Ishikawa, D., Mochizuki, T., Takeshita, K., Goto, S., Matsushita, T., Kimura, H., Yamamoto, F. & Ishikawa, T. (2001). Nucl. Instrum. Methods Phys. Res. A, 467–468, 627–630.  Web of Science CrossRef CAS Google Scholar
First citationBurkel, E., Peisl, J. & Dorner, B. (1987). Europhys. Lett. 3, 957–961.  CrossRef CAS Web of Science Google Scholar
First citationDorner, B., Burkel, E. & Peisl, J. (1986). Nucl. Instrum. Methods Phys. Res. A, 246, 450–451.  CrossRef Web of Science Google Scholar
First citationFukui, H., Katsura, T., Kuribayashi, T., Matsuzaki, T., Yoneda, A., Ito, E., Kudoh, Y., Tsutsui, S. & Baron, A. Q. R. (2008). J. Synchrotron Rad. 15, 618–623.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationIshikawa, D., Ellis, D. S., Uchiyama, H. & Baron, A. Q. R. (2015). J. Synchrotron Rad. 22, 3–9.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationJames, F. & Roos, M. (1975). Comput. Phys. Commun. 10, 343–367.  CrossRef Web of Science Google Scholar
First citationMasciovecchio, C., Bergmann, U., Krisch, M., Ruocco, G., Sette, F. & Verbeni, R. (1996a). Nucl. Instrum. Methods Phys. Res. B, 117, 339–340.  CrossRef CAS Web of Science Google Scholar
First citationMasciovecchio, C., Ruocco, G., Sette, F., Krisch, M., Verbeni, R., Bergmann, U. & Soltwisch, M. (1996b). Phys. Rev. Lett. 76, 3356–3359.  CrossRef PubMed CAS Web of Science Google Scholar
First citationPlaczek, G. (1952). Phys. Rev. 86, 377–388.  CrossRef CAS Web of Science Google Scholar
First citationSaid, A. H., Sinn, H. & Divan, R. (2011). J. Synchrotron Rad. 18, 492–496.  Web of Science CrossRef IUCr Journals Google Scholar
First citationSette, F., Ruocco, G., Krisch, M., Bergmann, U., Masciovecchio, C., Mazzacurati, V., Signorelli, G. & Verbeni, R. (1995). Phys. Rev. Lett. 75, 850–853.  CrossRef PubMed CAS Web of Science Google Scholar
First citationShinohara, Y., Dmowski, W., Iwashita, T., Ishikawa, D., Baron, A. Q. R. & Egami, T. (2019). Phys. Rev. Mater. 3, 065604.  Web of Science CrossRef Google Scholar
First citationSinn, H., Alp, E. E., Alatas, A., Barraza, J., Bortel, G., Burkel, E., Shu, D., Sturhahn, W., Sutter, J. P., Toellner, T. S. & Zhao, J. (2001). Nucl. Instrum. Methods Phys. Res. A, 467–468, 1545–1548.  Web of Science CrossRef CAS Google Scholar
First citationSutter, J. P. & Baron, A. Q. R. (unpublished). Multibeam Calculations Near Backscattering at the Si(11 11 11) Reflection of Silicon.  Google Scholar
First citationVerbeni, R., Kocsis, M., Huotari, S., Krisch, M., Monaco, G., Sette, F. & Vanko, G. (2005). J. Phys. Chem. Solids, 66, 2299–2305.  Web of Science CrossRef CAS Google Scholar

This article is published by the International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.

Journal logoJOURNAL OF
SYNCHROTRON
RADIATION
ISSN: 1600-5775
Follow J. Synchrotron Rad.
Sign up for e-alerts
Follow J. Synchrotron Rad. on Twitter
Follow us on facebook
Sign up for RSS feeds