research papers
Beam heating from a fourth-generation synchrotron source
aEuropean Synchrotron Radiation Facility, 71 Avenue des Martyrs, 38040 Grenoble, France
*Correspondence e-mail: lawrence@esrf.fr
The high levels of in situ measurements of beam heating. Results showed significant increases in temperature for metal and ceria samples, which are compared with a lumped thermodynamic model, providing a tool for estimating beam heating effects. These temperature increases may have a drastic effect on samples and measurements, such as the rapid recrystallization of a copper wire shown here. These results demonstrate the importance of beam heating and provide information needed to consider, predict and mitigate these effects.
available at a fourth-generation synchrotron are shown to have significant beam heating effects for high-energy X-rays and radiation hard samples, leading to temperature increases of over 400 K with a monochromatic beam. These effects have been investigated at the ID11 beamline at the recently upgraded ESRF Extremely Brilliant Source, using thermal lattice expansion to performKeywords: beam heating; radiation damage.
1. Introduction
Fourth-generation synchrotron X-ray sources bring increasing levels of ; Garman, 2010; Garman & Weik, 2019); however, with increasing we have found that radiation effects become significant even for metal samples and high-energy X-rays through beam heating. With the Extremely Brilliant Source upgrade at the European Synchrotron Radiation Facility (ESRF-EBS), we show that a monochromatic 43.44 keV beam can cause temperature increases as large as 400 K. These significant increases in sample temperature may change sample properties such as lattice parameters and drive significant chemical or physical changes. Properties under investigation may be temperature dependent for many materials science experiments, such as mechanical response, chemical behaviour, microstructure etc., making it vital that any changes in temperature are known. If the materials being measured are used as standards, temperature can play a significant role in the accuracy of measurements. These factors make X-ray beam heating a potential issue for all samples.
and coherence, allowing unprecedented levels of resolution for a wide range of techniques, but with increasing risk of radiation damage. The high achievable at synchrotrons has been well known to cause damage in biological samples at around 5–20 keV (Ravelli & Garman, 2006While beam heating and radiation effects on biological samples are usually overcome with cryo-cooling, high-speed measurements, helical scans, or a combination (Holton, 2009; Stern et al., 2009; Borek et al., 2010), these measures may not be appropriate for many materials science experiments. Consequently, we need to be able to understand and model beam heating to predict the severity of the problem and determine possible methods of mitigation. Past studies have investigated beam heating effects on samples, modelling the effects of sample geometry, time, heat sinks and other parameters (Helliwell, 1984; Kuzay et al., 2001; Kriminski et al., 2003; Hopkins & Thorne, 2016; Wallander & Wallentin, 2017; Bonino et al., 2020). Some experimental investigations have shown increases in temperature in a third-generation synchrotron beam (Snell et al., 2007; Rosenthal et al., 2014; Warren et al., 2019); however, these studies mostly consider only the implications for soft, biological samples and lack a clear comparison between model and experiment.
Here, we provide experimental investigation into these effects at a fourth-generation synchrotron by measuring sample temperature changes via thermal lattice expansion using in situ X-ray diffraction performed at the ESRF-EBS. By designing samples to maximize effects and simplify the thermodynamics of the system, we set up a quantitative comparison to a model for beam heating to help to predict and understand the severity of the problem.
2. Experiment
Experiments were performed at the Materials Science Beamline (ID11) at the ESRF. The recent Extremely Brilliant Source (EBS) upgrade at the ESRF has increased the ). With this upgrade, ID11 can now achieve a of up to 2 × 1013 photons s−1 at an energy of 43.44 keV and bandpass of 10−3 using two in-vacuum undulators and a double bent crystal monochromator operating in horizontal Laue geometry (Wright et al., 2020).
of the X-ray beams produced by many of the beamlines compared with the previous source (Raimondi, 2016Measurements were performed at 96 m from the source, as shown in Fig. 1, with far-field diffraction and near-field imaging set-up for in situ measurements. FReLoN cameras were used for collecting imaging and diffraction data, with imaging used to determine sample dimensions and alignment relative to the beam. Diffraction data were used to determine sample temperature using lattice expansion. The experimental hutch was maintained at a stable temperature, keeping the sample environment at 294 K.
et al., 2011). The profile of the focused beam is displayed in Fig. 1, where it can be seen that the beam has an FWHM of 80 µm × 140 µm. Time-resolved instantaneous heating measurements were performed by acquiring data every 0.01 s when the focusing lenses were inserted into the beam. Flux-dependent measurements were performed by opening upstream slits that were close to the lenses while collecting diffraction data. The incident values were determined using a silicon diode placed in the beam before the sample and conversion to photons s−1 via the method provided by Owen et al. (2009).
on the samples was maximized by using a beryllium in-vacuum compound refractive lens transfocator that is 31.5 m from the source to give 1:2 focusing of the beam onto the sample (VaughanBeam heating measurements were performed on samples with different absorption coefficients, sample geometry, and orientation with respect to the beam. Aluminium wire samples with a diameter of 0.5 mm were chosen as an example with a low μ, while 0.19 and 0.025 mm-diameter copper wires provide a higher-μ example. As the 0.19 mm-diameter wire is similar to the focused beam size, these copper samples were designed to give large increases in temperature. A 0.5 mm capillary of CeO2 powder, commonly used for calibration in X-ray diffraction measurements, was also investigated. Calibration of sample–detector distance was done using CeO2 at low This CeO2 sample was then also measured at high flux.
Wire samples were mounted with their length oriented along either x, the direction of the X-ray beam, or y, the horizontal direction of X-ray polarization. To approximate a lumped model system, samples should be thermally isolated. To achieve this, wire samples were mounted end-on to a small strip of Kapton tape. This is not as stable as mechanical mounting; when the sample is in the orientation for achieving the highest increase in temperature, any movement from this position will decrease the temperature.
3. Results from example samples
Before describing the detailed modelling and experimental results, we highlight some key observations that motivate our approach. Fig. 2 shows the increase in temperature versus for all samples using temperatures calculated from the change in lattice parameter from X-ray diffraction.
A 0.5 mm capillary of CeO2 powder shows large increases in temperature under high up to 210 K at 1.75 × 1013 photons s−1, with even small increases possibly having a significant effect on calibration and leading to inaccuracies in results (e.g. 200 K in temperature gives a strain of 2.2 × 10−3). The change in temperature was calculated from the change in lattice parameter, with a coefficient of 11 × 10−6 K−1 (Sameshima et al., 2002). As CeO2 has a low and was in powder form, it is difficult to model the thermal effects. Imaging of the CeO2 powder before and after exposure to the focused beam showed movement of the particles, further adding to the difficulty of modelling such a system. Consequently, having experimental data for such a system is paramount, and the results here show that beam heating may be significant.
A Cu wire sample with diameter 0.19 mm and length 0.85 mm showed the largest beam heating effect when oriented parallel to the beam, increasing by over 330 K. This beam heating was enough to cause recrystallization of the drawn wire sample, seen by the diffraction pattern changing from powder rings to intense spots in Fig. 5(a) later in the paper. This `thick' Cu wire shows a non-linear relation between incident and temperature increase due to radiative cooling. The non-linear was accounted for in the calculation of temperature from the lattice parameter, as shown in Table 1 (Hahn, 1970).
Cu wire of the same length but 0.025 mm diameter showed a smaller temperature increase of 44 K. The average
on this `thin' wire is much higher than for the thick Cu wire as it is at the centre of the beam, with the beam size larger than the wire diameter. The lower temperature increases seen for the thin wire show that there are other factors, such as surface area to volume ratio, that influence beam heating effects.For both thin and thick Cu wire samples, temperature increases are lower when side-on to the beam rather than end-on. As the same samples were used for both side-on and end-on measurements, only the path length of the beam, and therefore the absorbed power, will be different in the two scenarios, showing the significance of absorbed power on beam heating effects.
4. Testing the lumped thermodynamic model
4.1. Model
In order to better understand these beam heating effects for samples and experimental set-ups that may be expected in a typical materials science synchrotron experiment, we introduce a simple beam heating model. By experimentally testing this model, we can investigate the severity of the problem, and test a method of predicting such effects.
Kuzay et al. provide a clear analysis of beam heating effects, distinguishing between a `lumped' model, where internal thermal conduction in the sample is negligible, and a `distributed' model, where it is not (Kuzay et al., 2001). As the focus of this study is on beam heating effects, rather than heat conduction models, we use the simpler lumped model.
For the lumped model to be a reasonable approximation, the et al. (2001) for a quantitative description]. Consequently, beam heating effects should be dependent on radiation dose rate, i.e. the incident power, P, on the sample, rather than the total dose. Similarly, beam heating effects should be dependent on total incident and not This contrasts with macromolecular crystallography, where total dose is significant for both beam heating and radiation damage effects (Owen et al., 2006; Sliz et al., 2003; Leiros et al., 2006).
of the sample must be high enough or the sample size small enough that temperature is uniform across the sample on the time scales to be investigated [see KuzayThe power absorbed by the sample, Pin, is determined by the beam energy, Eb, and Φ, multiplied by the proportion of the beam absorbed by the sample:
The portion of the total beam power, P = ΦEb, absorbed by the sample is calculated from the attenuation, μ, along the beam path through the sample, x, at a point y, z in the sample, multiplied by the at point y, z and integrated over all y and z. Previous works in the 10–15 keV range have used the total recognizing that this is likely to be an overestimate (Rosenthal et al., 2014; Wallander & Wallentin, 2017). At 10–15 keV, the generation of photoelectrons is the dominant mechanism of X-ray attenuation; however, at the higher energy used in this study, the contribution from Compton scattering is significant. Consequently, we can also consider the photoelectron contribution to the determined by including the factor μpe/μ, where μpe is the photoelectron with values obtained from the XCOM database (Berger et al., 2010).
For a sample larger than the beam size, and equal path lengths through the sample for all points in the beam (such as a wire aligned with its length along the beam), we can simplify equation (1) to
With a lumped model system, we can calculate the equilibrium temperature, Teqm, at which the rate of thermal energy deposition in the sample by the beam (Pin) is equal to the rate of thermal energy loss (Maruyama & Moriya, 2021). With an environment temperature of T0, we find that
where the right-hand terms show the convective and radiative heat loss terms, respectively. The convective heat loss rate in equation (3) is dependent on surface area, A, and the heat transfer coefficient, h, while the radiative heat loss rate is dependent on surface area, emissivity, ε, and the Stefan–Boltzmann constant, σ.
Using a fourth-generation synchrotron source and the set-up typically expected for a material science experiment, it is possible to reach temperatures at which radiative effects are not negligible and a radiative cooling term should be included. This is shown by the relationship between ΔT = Teqm − T0 and incident for the 0.19 mm-thick Cu wire shown in Fig. 2. Assuming h and ε to be independent of temperature, fitting this data with equation (3) gives an emissivity value of ε = 0.015 ± 0.004, matching literature values (Window & Harding, 1981; Estalote & Ramanathan, 1977).
Though solving equation (3) is straightforward, determining the appropriate value of h is not. Consequently, this has been a focal point for research on beam heating effects (Kriminski et al., 2003). This heat transfer coefficient will depend on sample material, surface roughness, shape, orientation, and the surrounding temperature and airflow, making it challenging to determine.
By integrating equation (3), neglecting the radiative cooling term, the sample temperature at a time t after the X-ray beam is incident on the sample is found to be
The characteristic time (hA/CVρ) is dependent on surface area, A, volume, V, specific heat, C, and density, ρ (Kuzay et al., 2001). This equation can be used to analyse time-resolved temperature measurements, allowing h to be experimentally determined for a specific sample and experimental set-up.
We can also consider the instantaneous sample heating that occurs before heat loss mechanisms become significant, i.e. adiabatic heating. In this case, the heating rate will be
This simple model can be used to predict the changes in temperature for samples and experimental set-ups that may be expected in a materials science synchrotron experiment. By investigating the time dependence of temperature changes during beam heating, we will be able to determine the heat transfer coefficient for each sample. This value, along with the
dependence of temperature changes, can then be used to calculate the thermal output of the system. Comparing this with the attenuated beam power will then show the effective `efficiency' of beam heating.We focus on using this model to predict scenarios with the highest increase in temperature in order to investigate the severity of the problem. Additionally, larger changes in temperature will allow for a more precise comparison with the model and improved sensitivity to the effects of radiative heat losses.
4.2. Aluminium wire results
A range of lengths of 0.5 mm-diameter Al wire were used to test the beam heating model with a low μ = 1.2897 cm−1 and μpe = 0.7258 cm−1 (Berger et al., 2010). Radiographs used to determine sample dimensions and align each sample to the beam are shown in Fig. 3. Fig. 4(a) shows the change in temperature of these wire samples as a function of time exposed to the focused X-ray beam. Temperature values were determined from the change in lattice parameter using a linear coefficient of 22 × 10−6 K−1 (Wilson, 1942).
ofUpon exposure to the focused X-ray beam, the initial rate of temperature increase (i.e. adiabatic heating rate) is the same for all Al wire lengths, showing that X-ray attenuation is approximately linear over the range of wire lengths measured. Because these measurements required fast acquisition of data, measurements could not be taken simultaneously and the values of ΔT could not be normalized and are therefore not comparable. Consequently, the heating efficiency cannot be calculated from adiabatic heating rate here.
As the Al wire samples have a diameter greater than the beam size, it is possible that there are internal thermal gradients. An examination of the change in strain calculated from each of the measured Bragg reflections gives further insight into this possibility. The values plotted in Figs. 4(b) to 4(e) show the difference between measured strain at a time t and strain at the final equilibrium temperature, Teqm, calculated from the (113), (222), (331) and (024) reflections, respectively, with the values for each length offset for clarity. Assuming the strain induced in the Al samples upon exposure to the focused beam is caused by these strain values give an indication of the change in thermal gradient across the sample.
The longest samples measured clearly showed some changes, with the (222) reflections showing the greatest changes in strain over time for most samples. These changes occur over a time of under 2 s for the longest samples, and shorter times for shorter samples, showing that thermal equilibrium within the wire is reached quickly. As the changes in strain are faster than the time taken to reach thermal equilibrium, it is likely that the thermal impedance of the wire surface is much greater than that of the bulk of the wire. Consequently, we assume that the radial thermal gradient in the wire is likely to be negligible and so equation (4) should be a good approximation.
Each data set in Fig. 4(a) has been fitted with the function described in equation (4), shown with a solid line. When fitting, only values of the heat transfer coefficient, h, were allowed to vary. Values of C and ρ (see Table 1) and surface area and volume were kept constant with geometry calculated from radiographs, e.g. Fig. 3, assuming the wires to be cylindrical. All wire lengths show a good fit to this function, with the longer wire lengths clearly showing greater characteristic times. The results plotted in Fig. 4(f) show a decrease in h with length, fitted with an h ∝ 1/length trend.
Measured changes in temperature with increasing incident (g), fitted with a linear trend, showing that radiative cooling was not significant. Some discrepancies can be seen, such as the dip seen for the 3.73 mm sample after 1.2 × 1013 photons s−1, which we believe to be due to movement of the sample. In such cases, only the data before the discrepancy were used for analysis.
are plotted in Fig. 4The results from Fig. 4(g) are compared with the modelled temperature increase at 1013 photons s−1 in Fig. 4(h). These values were calculated by solving equation (3) for Teqm, using the values given in Table 1 and the calculated h values (blue circles), and the fitted h (length) function (orange line). Experimental values of Teqm (green diamonds) show a similar trend to the modelled values with calculated and fitted h values, but with experimental values lower than modelled. Experimental values are found to be around 50% of the predicted temperature increases, with the exception of values for the 1.17 and 1.42 mm wires, which show a larger discrepancy. We find that Teqm increases with length as absorption is approximately linear over this length range.
The calculated thermal output at equilibrium calculated for an incident 13 photons s−1 (69.6 mW at 43.44 keV) is plotted against attenuated power calculated from the photoelectric in Fig. 4(i). The fitted gradient of 86 ± 9% shows the effective heating efficiency relative to the photoelectric attenuation. If only the total is used, a lower total heating efficiency of 48 ± 5% is obtained. The 1.0 ± 0.9 mW x intercept shows that photoelectron losses are not significant on the length scales being considered here.
of 104.3. Copper wire results
Similar measurements and analysis were performed on 0.19 mm-diameter Cu wire samples with a range of lengths, with results shown in Fig. 5. Cu has μ = 34.6 cm−1 and μpe = 31.8 cm−1 at 43.44 keV, greater than the values for Al, and the wire is comparable with the beam size, leading to significant beam heating effects. This is displayed in Fig. 5(a), where the diffraction rings from a 0.86 mm-length sample are shown for a range of times, with focusing lenses inserted into the beam at 0 s. The powder rings initially increase in intensity and shift to lower Q before separating into intense spots around the ring, showing the heating and subsequent recrystallization of the drawn wire.
Fig. 5(b) shows the change in temperature as a function of time exposed to the focused X-ray beam. Temperature transients of these Cu samples occur over less than 2 s, faster and to much higher temperatures than the Al samples. As fast measurements could not be taken simultaneously, the magnitude of these temperature changes cannot be normalized and compared. Fits to equation (4) are shown with a solid line, while the adiabatic heating rate is shown with a dotted line. The noise in the data for samples that reached higher temperatures is due to high-intensity reflections appearing as a result of recrystallization. This noise impeded strain analysis; however, as the wire diameter size is close to the beam size and attenuation through the sample length is approximately linear, it is reasonable to assume that internal conduction did not have a significant effect on characteristic time.
The rapid beam heating caused some movement of the samples, as seen in Fig. 5(d) for the two longest samples. Linear fits were used to analyse these data, taking the steepest gradient for each data set, which removes contributions from sample movement and radiative cooling. With these gradients it is possible to compare the measured temperature changes with the model at 1013 photons s−1, as shown in Fig. 5(e). The data and model using calculated and fitted h values have similar shape, with a peak in Teqm at around 0.25 mm length. However, there are large differences between the model and experiment, with predicted values increasing over 860 K, while experimental values do not go above 265 K. The discrepancy between model and experiment is evident when the thermal output is plotted against the calculated photoelectric attenuation in Fig. 5(f). Fitted with a linear trend, the large errors from the calculated h values and surface area give an x intercept of 12 ± 10 mW. The fitted gradient gives a heating efficiency of 55 ± 10% relative to the calculated photoelectric attenuation and 50 ± 10% relative to the calculated total attenuation.
4.4. Comparison of Al and Cu
Comparing the results from the 0.19 mm Cu and 0.50 mm Al wires, we find some differences, both with their fit to the lumped model and measured efficiency. The heating rate in the adiabatic regime (before heat loss becomes significant) is the same for all Al wires but decreases with increasing wire length for Cu due to the high attenuation of X-rays in Cu. This difference in attenuation between Al and Cu is also evident when comparing Fig. 4(h) and Fig. 5(e), where the experimental and modelled temperature changes decrease after only 0.25 mm length, while for Al they continue to increase over the measured range of sample lengths.
Al and Cu wires also showed differences in the measured photoelectron heating efficiency, with values of 86 ± 9% and 55 ± 10%, respectively. These differences and the values themselves may be understood by considering unaccounted energy loss mechanisms. Heat lost through the Kapton holder may contribute to energy losses. The effect of wire length on this conductive cooling would be more significant for shorter wire lengths, which is the pattern seen for Cu samples, and may explain the non-zero x intercept seen in Fig. 5(f). We would expect the rate of heat lost through conduction to be proportional to the temperature difference ΔT, which correlates with the lower thermal efficiency and greater x intercept seen for the Cu samples compared with Al. Additionally, we expect the rate of heat lost through conduction to be proportional to the contact area between the Kapton and the wire, i.e. the wire Heat loss through the Kapton holder should be more significant for the thicker Al wire, but this does not seem to be the case. Perhaps the temperature gradient from the Al wire to the Kapton is not large enough to drive heat loss via conduction.
In our model, we have only used the photoelectron et al., 2018; Nave & Hill, 2005), meaning that a much higher level of precision would be required to see such effects here. It is challenging to compute photoelectron energy loss mechanisms such as fluorescence and generation; however, we can make some simple assessments. The is dominated by Kα emission for both Al and Cu (Krause et al., 1978). For Al the Kα energy is 1.5 keV with an attenuation length of 0.009 mm, so very few fluorescence photons will escape the 0.5 mm-diameter Al wire. Cu Kα emission at 8.0 keV has an attenuation length of 0.022 mm which is about 10% of the wire thickness. Fluorescence photons emitted from the centre of the wire will be reabsorbed but those emitted from nearer the surface may escape. Energy loss via fluorescence emission may partially explain the difference in the measured photoelectron heating efficiency between Al and Cu samples.
of the material under investigation to determine the deposited energy, and have not considered the various energy loss mechanisms of the photoelectrons, not all of which will be converted to thermal energy. Electron escape is expected to be on a scale much smaller than our samples, with photoelectrons spreading on the order of 200 nm from the absorption event (TorselloBonino et al. used a more complex theoretical model using Monte Carlo (MC) simulations considering both photo-atomic and electron interactions to calculate the proportion of the incident X-ray beam converted into thermal energy in the sample (Bonino et al., 2020). Comparing explicit (using MC) to implicit modelling (using Beer–Lambert) of the experimental results of Snell et al. (2007), Bonino et al. find around 1 K difference between explicit and implicit models for a 5 to 10 K increase in the temperature of a glass bead in a 15 keV X-ray beam. This gives a heating efficiency similar to that seen for the Al wire samples. Though their experimental and modelled values are similar, the significant differences in their time dependence show the limits of the model, making it difficult to draw any conclusions about the thermal efficiency and therefore the accuracy of the model using MC simulations.
5. Discussion
Our results show significant beam heating effects for materials science samples in a monochromatic hard X-ray beam. Using high-energy X-rays allows the use of transmission geometry and bulk measurements, favouring the use of samples such as capillaries of powders, plates, wires or small single crystals. Radiation damage effects have not previously been a concern for these radiation-resistant materials, but we have seen that beam heating effects can become significant for certain sample sizes even at high X-ray energy with high
Consequently, when performing synchrotron experiments, beam heating effects should be considered, especially when using a high-flux focused beam and absorbing samples which match the beam size.Knowing the potential effect of beam heating, we can now consider how to predict and mitigate these effects. Predicting the exact temperature increases due to beam heating can be challenging, but by considering samples as thermally isolated and assuming maximum heating efficiency, we can use a lumped model to give an upper limit of temperature increase. In this case, the main challenge is determining h, which will be dependent on sample shape, orientation and environment. There are many formulas for calculating h using the Nusselt, Grasof and Prandtl dimensionless parameters, but few seem suited to small samples. [Calculated values for the samples used here based on the work of Incropera et al. (2007) gave h values around 50 to 200 Wm−2 K−1 lower than the experimentally determined values.] Consequently, calculated values of h may not give accurate predictions, but may be useful as a conservative low estimate of h for getting an upper limit of predicted temperature increase. In low-pressure or vacuum environments, samples will be better thermally isolated and h decreases significantly so that beam heating is a much more severe problem.
An alternative to predicting beam heating effects is to check if they are present during an experiment by measuring with reduced beam
Results may be compared to quickly see evidence of beam heating effects.Beam heating effects may be reduced in several ways. Using equation (3), we can see that heating is reduced by increasing the ratio of the surface area to length along the beam. Where possible, it may help to maximize the heat transfer coefficient, h, through sample orientation and exterior cooling. Providing thermal connection to a heat sink is also likely to be an effective way to increase the heat loss rate, reducing sample heating. If a sample has a thermal connection to a heat sink with a length l, k and cross-sectional area D, assuming the convective cooling of the heat sink connection is negligible, a further term can be added to the equation for equilibrium temperature, giving
Beam properties will also have significant effects on beam heating. We have shown that at a given energy Teqm is proportional to when radiative cooling is not significant. Beam heating effects can be reduced by lowering however, this will slow down measurements and prevent fast processes from being studied.
We may also increase the speed of measurements, in order to complete them before Teqm is reached. In the instantaneous heating regime, we expect adiabatic heating with ΔT proportional to total dose before the heat loss becomes significant. This diffraction before destruction approach has been extensively discussed for macromolecular crystallography. The ultimate realization is perhaps serial crystallography, where the sample is completely destroyed by the X-ray beam.
The fast measurement approach will be most effective for samples with larger characteristic times, i.e. samples with low or thermally isolated samples e.g. in vacuum. For thermal insulators, heat is not as effectively dissipated through the sample, and therefore decreasing the will also decrease beam heating effects, e.g. using an unfocused beam rather than a focused beam when measuring a CeO2 calibration sample and spending a few seconds longer to collect the data.
6. Conclusions
The high
levels achievable at fourth-generation synchrotron sources offer many advantages for measuring weak signals; however, it is important to be aware of possible effects of these high radiation levels. We have shown that X-ray beams can cause significant thermal, physical and structural changes to metal samples, clearly demonstrating that the effects of beam heating are of potential concern for a wide range of samples and can have a profound impact on experimental results. Consequently, it is vital to be able to understand and predict these effects.By performing measurements on a system designed to be realistic as an experimental set-up but to simplify modelling of the system, we have been able to test a model for beam heating predictions. Comparison between model and experiment showed these predictions are accurate at determining the relative effects of
and sample size. However, discrepancies were found between the results of Al and Cu wire samples, with Cu samples showing lower heating efficiency. This discrepancy may be because far higher temperatures are reached by Cu samples, or other unaccounted heat loss mechanisms. A better understanding of energy loss mechanisms will be needed to accurately predict the magnitude of temperature changes. Overall, these predictions provide an experimentally verified starting point for analysis of more complex samples and experiments, and have given useful insight into how to predict and mitigate beam heating effects.Acknowledgements
We thank the ESRF for provision of synchrotron beam time.
Funding information
Funding for this research was provided by: European Synchrotron Radiation Facility.
References
Berger, M., Hubbell, J., Seltzer, S., Chang, J., Coursey, J., Sukumar, R., Zucker, D. & Olsen, K. (2010). XCOM: Photon Cross Sections Database. National Institute of Standards and Technology, Gaithersburg, MD, USA (https://www.nist.gov/pml/xcom-photon-cross-sections-database). Google Scholar
Bonino, V., Torsello, D., Prestipino, C., Mino, L. & Truccato, M. (2020). J. Synchrotron Rad. 27, 1662–1673. Web of Science CrossRef IUCr Journals Google Scholar
Borek, D., Cymborowski, M., Machius, M., Minor, W. & Otwinowski, Z. (2010). Acta Cryst. D66, 426–436. Web of Science CrossRef CAS IUCr Journals Google Scholar
Estalote, E. A. & Ramanathan, K. G. (1977). J. Opt. Soc. Am. 67, 39. CrossRef Web of Science Google Scholar
Garman, E. F. (2010). Acta Cryst. D66, 339–351. Web of Science CrossRef CAS IUCr Journals Google Scholar
Garman, E. F. & Weik, M. (2019). J. Synchrotron Rad. 26, 907–911. Web of Science CrossRef CAS IUCr Journals Google Scholar
Hahn, T. A. (1970). J. Appl. Phys. 41, 5096–5101. CrossRef CAS Web of Science Google Scholar
Helliwell, J. R. (1984). Rep. Prog. Phys. 47, 1403–1497. CrossRef CAS Web of Science Google Scholar
Holton, J. M. (2009). J. Synchrotron Rad. 16, 133–142. Web of Science CrossRef CAS IUCr Journals Google Scholar
Hopkins, J. B. & Thorne, R. E. (2016). J. Appl. Cryst. 49, 880–890. Web of Science CrossRef CAS IUCr Journals Google Scholar
Incropera, F. P., Lavine, A. S., Bergman, T. L. & DeWitt, D. P. (2007). Fundamentals of Heat and Mass Transfer, 6th ed. New York: Wiley. Google Scholar
Krause, M., Nestor, C. Jr, Sparks, C. Jr & Ricci, E. (1978). X-ray fluorescence cross sections for K and L X-rays of the elements. Technical Report. Oak Ridge National Laboratory, TN, USA. Google Scholar
Kriminski, S., Kazmierczak, M. & Thorne, R. E. (2003). Acta Cryst. D59, 697–708. Web of Science CrossRef CAS IUCr Journals Google Scholar
Kuzay, T. M., Kazmierczak, M. & Hsieh, B. J. (2001). Acta Cryst. D57, 69–81. Web of Science CrossRef CAS IUCr Journals Google Scholar
Leadbetter, A. J. (1968). J. Phys. C Solid State Phys. 1, 1481–1488. CrossRef Web of Science Google Scholar
Leiros, H.-K. S., Timmins, J., Ravelli, R. B. G. & McSweeney, S. M. (2006). Acta Cryst. D62, 125–132. Web of Science CrossRef CAS IUCr Journals Google Scholar
Maruyama, S. & Moriya, S. (2021). Int. J. Heat Mass Transfer, 164, 120544. Web of Science CrossRef Google Scholar
Nave, C. & Hill, M. A. (2005). J. Synchrotron Rad. 12, 299–303. Web of Science CrossRef CAS IUCr Journals Google Scholar
Owen, R. L., Holton, J. M., Schulze-Briese, C. & Garman, E. F. (2009). J. Synchrotron Rad. 16, 143–151. Web of Science CrossRef CAS IUCr Journals Google Scholar
Owen, R. L., Rudiño-Piñera, E. & Garman, E. F. (2006). Proc. Natl Acad. Sci. USA, 103, 4912–4917. Web of Science CrossRef PubMed CAS Google Scholar
Raimondi, P. (2016). Synchrotron Radiat. News, 29, 8–15. CrossRef Google Scholar
Ravelli, R. B. & Garman, E. F. (2006). Curr. Opin. Struct. Biol. 16, 624–629. Web of Science CrossRef PubMed CAS Google Scholar
Rosenthal, M., Doblas, D., Hernandez, J. J., Odarchenko, Y. I., Burghammer, M., Di Cola, E., Spitzer, D., Antipov, A. E., Aldoshin, L. S. & Ivanov, D. A. (2014). J. Synchrotron Rad. 21, 223–228. Web of Science CrossRef CAS IUCr Journals Google Scholar
Sameshima, S., Kawaminami, M. & Hirata, Y. (2002). J. Ceram. Soc. Jpn, 110, 597–600. Web of Science CrossRef CAS Google Scholar
Sliz, P., Harrison, S. C. & Rosenbaum, G. (2003). Structure, 11, 13–19. Web of Science CrossRef PubMed CAS Google Scholar
Snell, E. H., Bellamy, H. D., Rosenbaum, G. & van der Woerd, M. J. (2007). J. Synchrotron Rad. 14, 109–115. Web of Science CrossRef CAS IUCr Journals Google Scholar
Stern, E. A., Yacoby, Y., Seidler, G. T., Nagle, K. P., Prange, M. P., Sorini, A. P., Rehr, J. J. & Joachimiak, A. (2009). Acta Cryst. D65, 366–374. Web of Science CrossRef CAS IUCr Journals Google Scholar
Torsello, D., Mino, L., Bonino, V., Agostino, A., Operti, L., Borfecchia, E., Vittone, E., Lamberti, C. & Truccato, M. (2018). Phys. Rev. Mater. 2, 014801. Web of Science CrossRef Google Scholar
Vaughan, G. B. M., Wright, J. P., Bytchkov, A., Rossat, M., Gleyzolle, H., Snigireva, I. & Snigirev, A. (2011). J. Synchrotron Rad. 18, 125–133. Web of Science CrossRef IUCr Journals Google Scholar
Wallander, H. & Wallentin, J. (2017). J. Synchrotron Rad. 24, 925–933. Web of Science CrossRef IUCr Journals Google Scholar
Warren, A. J., Axford, D. & Owen, R. L. (2019). J. Synchrotron Rad. 26, 991–997. Web of Science CrossRef CAS IUCr Journals Google Scholar
White, G. K. & Collocott, S. J. (1984). J. Phys. Chem. Ref. Data, 13, 1251–1257. CrossRef CAS Web of Science Google Scholar
Wilson, A. J. (1942). Proc. Phys. Soc. 54, 487–491. CrossRef CAS Google Scholar
Window, B. & Harding, G. (1981). J. Opt. Soc. Am. 71, 354–357. CrossRef CAS Web of Science Google Scholar
Wright, J., Giacobbe, C. & Majkut, M. (2020). Curr. Opin. Solid State Mater. Sci. 24, 100818. Web of Science CrossRef Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.