research papers
Diamond channel-cut crystals for high-heat-load beam-multiplexing narrow-band X-ray monochromators
aAdvanced Photon Source, Argonne National Laboratory, Argonne, IL 60439, USA, bTechnological Institute for Superhard and Novel Carbon Materials, 142190 Troitsk, Russian Federation, and cNational Synchrotron Radiation Centre SOLARIS, 30-392 Krakow, Poland
*Correspondence e-mail: shvydko@anl.gov
Next-generation high-brilliance X-ray photon sources call for new X-ray optics. Here we demonstrate the possibility of using monolithic diamond channel-cut crystals as high-heat-load beam-multiplexing narrow-band mechanically stable X-ray monochromators with high-power X-ray beams at cutting-edge high-repetition-rate X-ray free-electron laser (XFEL) facilities. The diamond channel-cut crystals fabricated and characterized in these studies are designed as two-bounce Bragg reflection monochromators directing 14.4 or 12.4 keV X-rays within a 15 meV bandwidth to 57Fe or 45Sc nuclear experiments, respectively. The crystal design allows out-of-band X-rays transmitted with minimal losses to alternative simultaneous experiments. Only ≲2% of the incident ∼100 W X-ray beam is absorbed in the 50 µm-thick first diamond crystal reflector, ensuring that the monochromator crystal is highly stable. Other X-ray optics applications of diamond channel-cut crystals are anticipated.
Keywords: X-rays; X-ray optics; Bragg diffraction; diamond; X-ray free-electron lasers.
1. Introduction
Hard X-ray free-electron lasers (XFELs) in the self-seeding mode (Saldin et al., 2001; Geloni et al., 2011) generate brilliant fully coherent X-ray beams with a well defined and narrow spectrum, typically with a bandwidth of ∼0.2–1 eV (Amann et al., 2012; Min et al., 2019; Inoue et al., 2019; Nam et al., 2021) and a of a few tens of femtoseconds.
When operated in a high-repetition-rate (of the order of megahertz) pulse sequence mode with properly tapered undulators, self-seeded XFELs may deliver X-rays with a pulse energy of ∼10 mJ per pulse (∼7 × 1012 photons per pulse of ∼10 keV photons) and a time-averaged power of more than 100 W, corresponding to a time-averaged spectral of ∼1017 photons s−1 in a bandwidth of ∼1 eV (Chubar et al., 2016). This is about three orders of magnitude larger than the average spectral currently possible with storage-ring-based synchrotron radiation sources [see also Yang & Shvyd'ko (2013) and Geloni et al. (2015)].
An X-ray source with an increase of three orders of magnitude in the spectral ), X-ray photon correlation spectroscopy (Shpyrko, 2014) and nuclear (Rüffer & Chumakov, 2016) in particular.
opens up exciting new opportunities for hard X-ray spectroscopic techniques, such as inelastic X-ray scattering (Baron, 2016Such opportunities can be realized today, because high-repetition-rate hard XFELs are now a reality. The European XFEL in Hamburg (Germany) (Decking et al., 2020) is the first operational high-repetition-rate XFEL facility. The LCLS-II-HE in Stanford (USA) (Raubenheimer, 2018) is coming soon.
However, these new hard X-ray sources with high average spectral
present not only opportunities but also challenges. One challenge is how to monochromate XFEL beams of very high average and peak power.To address these challenges, we have designed, manufactured and tested diamond channel-cut crystals for use as high-heat-load beam-multiplexing narrow-band (15 meV bandwidth) mechanically stable monochromators.
A schematic of our channel-cut crystal design, together with relevant X-ray beams, is shown in Fig. 1(a). We choose a channel-cut monolithic monocrystalline design for two reasons. First, this design enables multiple successive Bragg reflections (two, four etc.) by perfectly aligned parallel mechanically stable reflecting atomic planes in the reflecting crystal plates P1 and P2 of the monolithic system. Second, it ensures parallel propagation of the incident and multiply reflected beams. The design includes a drumhead structure shown in Fig. 1(b) to permit one crystal face to be an ultrathin membrane, as discussed later.
We choose diamond because diamond crystals have a unique combination of outstanding physical properties that are perfect for many X-ray optics applications for which traditional materials, such as silicon, fail to perform (Shvyd'ko et al., 2017). Single crystals of silicon are attractive because they have an almost perfect are commercially available in large quantities and are commonly used in X-ray crystal optics. However, many applications, including those considered here, require better performance than silicon can deliver, specifically, better transparency to X-rays, resilience to radiation damage, mechanical stiffness, X-ray Bragg reflectivity and Diamond is the material of choice in such cases, as it is superior to silicon because of its unrivaled radiation hardness, an order of magnitude higher transparency to X-rays, an order of magnitude higher stiffness (Field, 1992; Prelas et al., 1998), orders of magnitude higher (Wei et al., 1993; Inyushkin et al., 2018), small (Stoupin & Shvyd'ko, 2010, 2011) and almost 100% reflectivity in Bragg diffraction even in backscattering (Shvyd'ko et al., 2010, 2011). This unique combination of outstanding properties makes diamond the most promising material for the transparent resilient high-resolution wavefront-preserving X-ray optics components that are essential for many applications at third-generation light sources. These characteristics are especially important at next-generation high-brilliance high-coherence light sources, such as diffraction-limited storage rings (DLSRs) and XFELs.
The paper is organized as follows. Section 2 covers the function and optical design of the diamond channel-cut crystals. Details of the fabrication are presented in Section 3. The results of characterization of the crystals by X-ray rocking curve imaging are presented in Section 4.1, and the results of X-ray performance tests are given in Section 4.2. Conclusions and outlook are discussed in Section 5.
2. Function and optical design
High-repetition-rate (HRR) hard X-ray self-seeded (HXRSS) XFELs will provide X-ray photon beams with an unprecedented spectral
advantageous for many applications, in particular for high-resolution spectroscopies, such as nuclear (NRS), and others. However, dealing with an X-ray beam of such high power is a challenge. Diamond channel-cut crystals can help solve this problem. Our recipe for the solution has four ingredients: a high-indexed Bragg reflection (for a narrow band reflection), diamond crystals (for X-ray transparency and excellent thermomechanical properties), a drumhead design (for low absorbed power and beam sharing) and a monolithic channel-cut design (for in-line geometry and stability).An important attribute of high-resolution spectroscopies is the use of X-ray beams with millielectronvolt or even sub-millielectronvolt bandwidths. As a result, such techniques require X-ray sources with a high spectral density, and ultra-high-resolution X-ray crystal monochromators with a bandpass of 1–0.1 meV and submicroradian angular stability; see Shvyd'ko (2004) for a review. Such optics usually cannot perform under high-heat-load conditions because of the resulting substantial angular instabilities. At storage-ring synchrotron radiation facilities, additional primary or high-heat-load crystal monochromators are used (Chumakov et al., 2014). This approach reduces the bandwidth of the undulator X-rays to about 1–2 eV and thus reduces substantially (by a factor of ∼100) the heat load on the ultra-high-resolution X-ray crystal monochromators.
However, the standard high-heat-load monochromators are not helpful to deal with X-ray beams generated by HRR–HXRSS XFELs, as these sources typically already have a relatively narrow bandwidth of ∼1 eV. Rather, such beams require a high-resolution monochromator with a bandpass of ∼10 meV that is stable under high heat loads. This smaller bandpass would reduce the heat load on the ultra-high-resolution optic by a factor of about 100. As an example, 10 keV X-rays can be monochromated to ΔE ≃ 10 meV with a single Bragg reflection1. In this case, about 1% of the XFEL radiation power in the small bandwidth can be directed to the ultra-high-resolution monochromator and associated spectroscopic experiments. Most of the remaining incident is transmitted, rather than being deposited in the crystal, thus avoiding crystal heating, the resulting angular instabilities and, therefore, angular instabilities in the reflected beam. If the monochromator is suitably designed, this remaining can be directed to other experiments.
Multiplexing beams from one X-ray source to perform simultaneous experiments is a common approch at synchrotron radiation (Als-Nielsen et al., 1994; Mattenet et al., 1998) and free-electron laser facilities (Zhu et al., 2014; Stoupin et al., 2014; Feng et al., 2015). This multiplexing of the beam requires an X-ray-transparent Bragg-reflecting crystal. Such a crystal would allow most of the incident X-rays, with the exception of the Bragg reflections in the narrow bandwidth, to pass through without absorption. Silicon cannot perform in this way because the ten extinction lengths required for maximum Bragg reflectivity are comparable to the photoabsorption length in silicon. In contrast, for diamond crystals, ten extinction lengths equal about one tenth of the photoabsorption length. Exactly because the extinction length in diamond is much smaller than the absorption length, diamond crystals experience close to 100% Bragg reflectivity, even in backscattering, and simultaneously offer high X-ray transparency (Shvyd'ko et al., 2011). Therefore, we chose to fabricate a diamond crystal membrane with a thickness of about five extinction lengths for high X-ray reflectivity and transparency.
However, this dimension of five extinction lengths measures only about 20 to 100 µm, depending on the chosen Bragg reflection. It is challenging to fabricate and handle such ultrathin crystal components without introducing crystal damage and strain. A solution is to use a drumhead crystal: a monolithic (b). This design ensures mechanically stable strain-free mounting of the membrane while still allowing efficient thermal transport (Kolodziej et al., 2016).
comprised of a thin membrane furnished with a surrounding solid collar, shown schematically in Fig. 1Finally, to ensure a fixed direction of the reflected X-rays independent of the . To ensure mechanical stability, both crystal plates are a part of a single monolithic channel-cut Channel-cut crystals have been popular in X-ray optics applications since the 1960s (Bonse & Hart, 1965a,b, 1966; Deslattes, 1968). Typically they are fabricated from silicon or germanium crystals. Manufacturing a channel-cut crystal from diamond is a challenge, but becomes possible using modern advanced techniques, as discussed in Section 3.
we add one more equivalent Bragg reflection from another crystal plate in a non-dispersive setting, as shown in Fig. 1In the present discussion, we focus on a diamond monolithic channel-cut crystal for use with 14.4125 keV X-rays for 57Fe nuclear resonant experiments (Gerdau & de Waard, 1999, 2000; Röhlsberger, 2004; Rüffer & Chumakov, 2016). We also explored a similar design for use with 12.4 keV X-rays for 45Sc experiments (Shvyd'ko & Smirnov, 1990).
The 733 Bragg reflection in diamond is the highest indexed reflection that is accessible with 14.4 keV X-rays; it has the largest θ = 80.75226°) and the smallest Bragg reflection bandwidth (13 meV). However, because of practical considerations related to channel-cut manufacturing (discussed in Section 3), we opted for the reflection with the next highest index, the 800 Bragg reflection, with a smaller of θ = 74.7207° and a larger bandwidth of 20 meV.
(Fig. 2 shows the predicted X-ray photon energy dependencies of the double 800 Bragg reflection profiles from a diamond channel-cut crystal, calculated numerically using equations of the of X-ray diffraction in crystals (Authier, 2001; Shvyd'ko, 2004). The first and second reflections are from 50 and 500 µm-thick crystal plates, respectively. The peak Bragg reflectivity of the 50 µm-thick crystal reduces to 91%, from 99% for the 500 µm-thick crystal (assuming σ polarization), resulting in a double-reflection peak reflectivity of 89%. The energy bandwidth ΔE is 17 meV for the σ-polarization and 14.5 meV for the π-polarization components of the beam. An angular divergence of the incident X-rays of 2.5 µrad (FWHM) is used in the calculations. The calculations show that only 2% of the incident X-ray beam is absorbed in the first plate (the 50 µm-thick diamond drumhead crystal), thus making possible high stability of the monochromator crystal.
Such a diamond monolithic channel-cut crystal can function as a two-bounce 800 Bragg reflection monochromator, deflecting 14.4 keV X-rays within a ∼15 meV bandwidth to 57Fe nuclear experiments, while the out-band X-rays are transmitted through the thin drumhead crystal and could be transported to a simultaneous experiment2.
Similarly, 620 Bragg reflections are suitable for a diamond channel-cut crystal narrow-band monochromator for 12.4 keV X-rays, corresponding to the nuclear resonance in 45Sc (Shvyd'ko & Smirnov, 1990); see Fig. 3.
3. Manufacturing
Two conditions are key to successful realization of a properly functioning diamond channel-cut crystal X-ray monochromator: (i) the availability of flawless single-crystal diamond material in sufficiently large workable sizes, and (ii) the ability to machine diamond without introducing crystal defects.
The characteristic linear dimensions of presently available high-quality synthetic diamond crystals are in the range of a few millimetres [see Shvyd'ko et al. (2017) for a recent review]. The crystals are relatively small but sufficient for use with XFEL beams, which are typically less than 1 mm2 in Fig. 4 shows a schematic diagram of the diamond channel-cut crystals produced in the present studies.
The channel-cut crystals were cut from synthetically grown raw diamond stones of type IIa (with nitrogen content below 1 p.p.m.). The diamond stones were grown using a temperature-gradient method under high-pressure and high-temperature (HPHT) conditions. The stones used in these studies were grown and machined at the Technological Institute for Superhard and Novel Carbon Materials (TISNCM) in Troitsk, Russia (Blank et al., 2007; Polyakov et al., 2011).
Raw HPHT-grown diamond stones typically have a cubo-octahedral form. It is well known that the top part of the (001) cubic growth sector (the area furthest from the 001 crystal seed) features the highest crystal quality with the lowest defect density (Burns et al., 2009; Sumiya & Tamasaku, 2012). Therefore, in these studies, the channel-cut crystals are extracted from the top part of the stone, as indicated in Fig. 4(b). With this cut, the crystal plate surfaces parallel to the atomic planes (hk0) and to the [001] direction take the largest volume of highest quality. For this reason, we chose to make an (800) monochromator for use with 14.4 keV X-rays, instead of the narrower band (733) design. Additional reasons for the (800) choice are related to laser machining, as discussed later. Similarly, we chose to make a (620) channel-cut monochromator for 12.4 keV X-rays in the 45Sc case.
Production of the monochromator crystals begins with selection by visual screening of high-quality as-grown diamond crystal stones for machining. To ensure the fewest defects, crystals with well developed defect-free cubic facets are selected because, as noted previously, the (100) cubic growth sector contains the smallest amount of contamination and lowest number of structural defects. The diamond stones are machined with the newest laser machining technologies.
The monochromator atomic planes [(800) or (620)] of the selected diamond stones are oriented to an accuracy of better than 0.5°, and the channel-cut crystals are cut to the designed shape using a pulsed Nd:YAG laser with the following parameters: 532 nm wavelength (second harmonic), 15 W average power, 10 kHz repetition rate, ∼1.5 mJ pulse energy, 40 ns −1.
and 15 µm focused spot size (FWHM). The laser beam is positioned and steered on the workpiece with a two-coordinate optical mirror scanning system equipped with an F-Theta scanning lens. The laser beam is moved with a linear speed of 100 mm sThe drumhead structure is formed by layer-by-layer ablation with picosecond pulses from a Nd:YAG laser operated in the third harmonic (355 nm wavelength) with the following parameters: 1 W average power, 10 ps −1 with a similar two-coordinate scanning system with an F-Theta lens. The removal rate is about 0.3 µm per layer. The surface roughness of the drumhead surface after laser ablation of a 0.5 mm-thick crystal is ≲1 µm (r.m.s.). Similar laser parameters and procedures have been used previously to manufacture diamond X-ray lenses (Terentyev et al., 2015) and drumhead crystals (Kolodziej et al., 2016).
500 kHz repetition rate and 2 µJ pulse energy. The use of the third harmonic enables reduction of the focused beam spot size and, most importantly, reduction of the diameter of the crater in the diamond produced by a single laser pulse. The crater diameter is 8 µm, provided the beam is focused on the diamond surface. During ablation, the beam is steered over the workpiece at a speed of 150 mm sFig. 5(a) shows photographs of the first machined (800) channel-cut crystal. The crystals appear black after machining. Apparently, laser machining is accompanied by graphitization of the diamond surface. To erase the blackening, we anneal the crystals in air at a temperature of 630–650°C for 3 h, as in our previous and recent studies (Kolodziej et al., 2016; Pradhan et al., 2020). After annealing, the blackening has vanished, as seen on the photographs in Figs. 5(c)–5(d). Most importantly, annealing erases any crystal strain induced in the process of laser cutting or ablation, which is critical for the proper functioning of the channel-cut crystal as a narrow-band X-ray monochromator. The annealing temperature is chosen such that all residuals of graphite and other carbon compounds are burned in air, while keeping diamond intact (but note that increasing the annealing temperature or time may result in etching of the crystal surface). We will refer to this procedure as medium-temperature in-air annealing (MTA).
The first reflecting crystal plate of the channel-cut crystal (P1 in Fig. 5) features a 50 µm-thick drumhead, designed to achieve high reflectivity (≳90%) of X-rays in the Bragg reflection band and 98% transmissivity of the out-of-band X-rays. X-rays in the 15 meV band reflected from both crystal plates (P1 and P2) of the channel-cut crystal (see Fig. 2) are directed to a high-spectral-resolution X-ray instrument. The major part, 98% of the incident X-rays in a 1 eV bandwidth (Chubar et al., 2016; Geloni et al., 2015), is transmitted and can be used in simultaneous parallel experiments. The 2 mm-thick base allows the crystal to be mounted mechanically tight and strain-free and to be water-cooled efficiently, ensuring stable operation in high-power XFEL beams.
Fig. 6 shows an example of a (620) diamond channel-cut crystal designed and manufactured for 45Sc nuclear resonant experiments with 12.4 keV X-rays. The (620) channel-cut crystal also features a 50 µm drumhead on the first plate P1. Making drumhead crystals in orientations different from (100) is a challenge (Kolodziej et al., 2016).
Anisotropy of the physical properties of diamond crystals leads to different results in laser ablation of surfaces with different crystal orientations. Surfaces with the (100) orientation do not present significant problems for laser ablation. In contrast, surfaces with the (111) crystallographic orientation are most difficult, both for laser ablation and for polishing. Laser ablation of the (111) surface creates microcracks.
The quality of the (620) surfaces after laser ablation is typically worse than that of the (100) surfaces. We found that increasing the crystal temperature improves the surface quality of the machined surfaces. During machining of the drumhead with the (620) surface orientation, the crystal is heated to 600°C by a resistive heater installed in the crystal holder.
Altogether, three (800) and two (620) channel-cut monochromator crystals were manufactured and characterized. In the following section, we present X-ray characterization results of the best (800) channel-cut crystal monochromator (third iteration), shown in Fig. 7.
4. Characterization of channel-cut crystals
4.1. Crystal quality test: X-ray rocking-curve imaging
The quality of the diamond channel-cut crystals was characterized using X-ray Bragg diffraction rocking curve imaging (RCI), also known as sequential topography (Lübbert et al., 2000). In this technique, Bragg reflection images of the crystal under study are measured with a pixel X-ray detector, with images being taken sequentially at different incidence angles to the reflecting atomic planes of the well collimated X-rays; Bragg reflection maps are calculated for the reflection peak intensities, angular widths and angular positions. We used an RCI setup (Stoupin et al., 2016; Pradhan et al., 2020) on the X-ray optics testing beamline 1BM (Macrander et al., 2016) at the Advanced Photon Source (APS), shown schematically in Fig. 8.
The RCI setup employs a nearly non-dispersive double-crystal C1–C2 arrangement. The first conditioning crystal C1 is an asymmetrically cut high-quality silicon crystal in the 531 Bragg reflection; the second crystal C2 is the structure under study, the (800) diamond channel-cut crystal, in the 400 Bragg reflection. We used the 400 Bragg reflection rather than the 800 reflection to characterize the (800) channel-cut crystal for the following reason. We want both reflecting plates P1 and P2 of the channel-cut crystal to be illuminated and imaged simultaneously so that we can evaluate whether there is any angular misorientation of the plates due to crystal defects. To meet this requirement, the must be chosen close to 45°, which in turn would require high-energy (19.66 keV) X-ray photons if the 800 reflection were used. Such a choice would result in time-consuming measurements, as the is relatively low at this high photon energy and the angular reflection width (Δθ800 = 1.2 µrad) is very narrow (compared with Δθ800 = 5.2 µrad for 14.4 keV X-rays). Therefore, we chose instead the 400 Bragg reflection and 9.831 keV X-rays, for which the θ400 = 45°, with angular width Δθ400 = 9.1 µrad if the crystal is 0.5 mm thick (reflecting plate P1) and Δθ400 = 9.8 µrad if the crystal is 0.05 mm thick (drumhead on reflecting plate P2).
The upstream double-crystal Si(111) monochromator is therefore tuned to select 9.831 keV X-rays. The monochromator bandwidth is ∼1.5 eV. For the conditioning crystal C1, the is θ531 = 43.385°. The asymmetry angle is chosen as η531 = 41.4°, resulting in the asymmetry parameter b = −28.7. This choice ensures ∼1 µrad angular collimation and an X-ray beam size greater than 20 mm illuminating the channel-cut crystal C2. Under these conditions, both plates of the channel-cut crystal are illuminated and can be imaged simultaneously in Bragg diffraction. The setup enables RCI mapping with submicroradian angular and 2.5 µm spatial resolution.
X-ray images of the (800) channel-cut crystal taken at the maximum of the sample-integrated Bragg reflection curves are shown in Figs. 9(a) and 10(a). In Fig. 9, X-rays illuminate the front (working) face of plate P1 and the rear face of plate P2, respectively, as in Fig. 7(a). In Fig. 10, the sample is exposed to X-rays from the reverse side, as in Fig. 7(b).
The angular dependencies of Bragg reflectivity (rocking curves) measured with each detector pixel are used to calculate Bragg reflection maps. Figs. 9(b) and 10(b) show color maps of the angular widths (FWHM) of the rocking curves. The color maps of the rocking curves' mid-points are shown in Figs. 9(c) and 10(c). The maps are calculated using a dedicated program (Stoupin, 2015). Cross sections through the FWHM and mid-point maps along the blue lines in Figs. 9(a) and 10(a) are presented in Figs. 9(b′), 9(c′), 10(b′) and 10(c′), respectively.
The microscopic defects can be derived from the Bragg reflection FWHM maps. The mesoscopic and macroscopic crystal strain and Bragg plane slope errors can best be evaluated from the mid-point maps. In particular, if the plates of the channel-cut crystal are misoriented due to crystal defects, the mid-point maps will provide this information, because both plates of the channel-cut crystal are imaged simultaneously.
The X-ray images and RCI maps in Figs. 9 and 10 reveal the good crystal quality of the manufactured (800) channel-cut crystal, especially in its upper part. Nevertheless, some defects can be detected. Dislocations and stacking faults are typical defects in HPHT-grown diamond. These defects result in increased angular widths and varying angular positions of the rocking curves. In the center of the plates (in the working area), the angular reflection widths are increased by a few microradians from the expected 9 µrad angular width of the 400 Bragg reflection. The angular position of the rocking curves also varies within a range of a few microradians, which indicates angular misalignment of the plates. These distortions are caused chiefly by the crystal defects in the as-grown crystal stone. Machining of the diamond has a secondary effect.
4.2. X-ray performance test: two-bounce Bragg reflection from the diamond channel-cut crystal
The performance of the manufactured (800) diamond channel-cut crystal with a 50 µm drumhead on the first plate P1 was tested by measuring the angular width Δθ of two successive Bragg reflections of X-rays from its crystal plates P1 and P2. The angular width is measured using a non-dispersive double-crystal arrangement, with the first reference crystal chosen to have an angular reflection width much smaller than the angular width of the channel-cut crystal. The spectral width of the channel-cut double reflection can be estimated as ΔE/E = Δθ/tanθ using the differential presentation of Bragg's law and the measured Δθ value.
We used the 777 Bragg reflection of 14.412 keV X-rays from the reference Si crystal with a θ777 = 73.7848° and an angular reflection width of Δθ777 = 1.2 µrad. It matches well with the 800 Bragg reflection from diamond, with a of θ800 = 74.7206° and an angular width of Δθ800 = 5.2 µrad.
ofIn the first step, the setup is tested using two successive 777 Bragg reflections from Si crystals, Si(777) × Si(777), as shown schematically in Fig. 11(a). The angular dependence of the Bragg reflection is measured by rocking the angular orientation of the second crystal and thus changing the angle of incidence of the X-rays to the diffracting atomic planes. The angular reflection profile shown in Fig. 11(a′) features an angular width of Δθa = 2.7 µrad, which is close to the expected ∼2 µrad. The additional broadening is attributed to the large horizontal spread of X-rays on a bending magnet beamline. The appropriate spectral bandwidth is calculated to be ΔEa = EΔθa/tan(θ777) ≃ 11 meV. The angular width Δθa and the spectral width ΔEa represent the angular and spectral resolution of the setup with the reference Si(777) crystal.
In the next step, we used a similar nearly non-dispersive double-crystal Bragg reflection arrangement, Si(777) × C(800), as shown schematically in Fig. 11(b), to measure the angular width of the single 800 Bragg reflection from one crystal plate of the channel-cut crystal. The reflection profile shown in Fig. 11(b′) on a logarithmic scale features an angular width of Δθb = 7.8 µrad, which is about 2.5 µrad broader than the expected value. The broadening is attributed to imperfections in the diamond crystal. The appropriate Bragg reflection spectral width calculated from the differential Bragg law is ΔEb ≃ 31 meV, which should be compared with a 20 meV single-reflection spectral width.
In the final step, we used two successive Bragg reflections from both plates of the (800) channel-cut crystal, as shown schematically in Fig. 11(c), to measure the combined angular Bragg reflection width of the channel-cut crystal. Compared with the single-reflection profile in Fig. 11(b′), the double-reflection profile of the channel-cut crystal, shown in Fig. 11(c′) on a logarithmic scale, features steeper tails in the angular profile and a narrower angular width of Δθc = 7.1 µrad. The appropriate Bragg reflection spectral width calculated from the differential Bragg law is ΔEc ≃ 28 meV, which is broader than the expected 17 meV (see Fig. 2). The double-reflection peak reflectivity drops by a factor of ∼1.8 compared with the single-reflection reflectivity in Fig. 11(b′). We attribute the decrease in the double-reflection reflectivity to the angular misorientation of the crystal plates.
Despite the deviations from the theoretical predictions, the manufactured diamond channel-cut crystal is performing within expectations. The experimental data demonstrate the feasibility of using channel-cut diamond crystals to monochromate X-ray beams at cutting-edge XFELs.
5. Conclusions and outlook
We have demonstrated the feasibility of diamond channel-cut crystals designed to function as high-heat-load beam-multiplexing narrow-band mechanically stable X-ray monochromators for high-power X-ray beams at cutting-edge high-repetition-rate XFEL facilities. Laser machining techniques were used to manufacture complex 3D structures in the channel-cut crystals. The crystals were designed for use as X-ray monochromators for 14.4 keV X-rays, corresponding to the 57Fe nuclear resonance, and for 12.4 keV X-rays, corresponding to the 45Sc nuclear resonance.
The channel-cut crystals were characterized by X-ray rocking curve imaging (topography) and by measurement of the angular and energy widths of double Bragg reflections. The studied channel-cut crystals perform close to theoretical expectations. However, the reflection curves are somewhat broadened and the channel-cut reflecting plates are misoriented by a few microradians because of crystal defects, which further reduces the double-reflection crystal reflectivity.
High crystal quality in the initial diamond stone used in fabricating the channel-cut crystal is essential for the proper functioning of this diamond monochromator.
Along with applications as monochromators, other applications of the diamond channel-cut crystals are anticipated. In particular, because diamond crystals feature very high X-ray reflectivity, they can be used as efficient four-bounce or six-bounce channel-cut polarizers and analyzers to achieve a very high degree of X-ray polarization purity (Bernhardt et al., 2016).
Footnotes
1XFEL beams typically have a small angular divergence of Δθ ≃ 2 µrad (Saldin et al., 2000; Huang & Kim, 2007). Because of this property, monochromation of X-rays to a ∼10 meV bandwidth can be achieved with a single high-indexed Bragg reflection, because the spectral broadening due to the angular divergence can be kept small (δE/E ≃ Δθ/tanθ ≲ 10−6), provided the θ is large and tanθ ≫ 1. The use of high-indexed Bragg reflections with θ close to 90° is therefore important.
2To ensure further stability of such a multiplexed setup, it would be advantageous to choose the for the diamond channel-cut crystal to be perpendicular to the of the downstream ultra-high-resolution optic, as discussed by Chubar et al. (2016).
Acknowledgements
Diamond crystal growth, quality checks and laser machining at FSBI TISNCM were performed on a Shared-Equipment User Facility `Research of Nanostructured, Carbon and Superhard Materials'.
Funding information
Work at Argonne National Laboratory was supported by the US Department of Energy, Office of Science, Office of Basic Energy Sciences, under contract DE-AC02- 06CH11357.
References
Als-Nielsen, J., Freund, A., Grübel, G., Linderholm, J., Nielsen, M., del Rio, M. & Sellschop, J. (1994). Nucl. Instrum. Methods Phys. Res. B, 94, 306–318. CAS Google Scholar
Amann, J., Berg, W., Blank, V., Decker, F.-J., Ding, Y., Emma, P., Feng, Y., Frisch, J., Fritz, D., Hastings, J., Huang, Z., Krzywinski, J., Lindberg, R., Loos, H., Lutman, A., Nuhn, H.-D., Ratner, D., Rzepiela, J., Shu, D., Shvyd'ko, Yu., Spampinati, S., Stoupin, S., Terentiev, S., Trakhtenberg, E., Walz, D., Welch, J., Wu, J., Zholents, A. & Zhu, D. (2012). Nat. Photon. 6, 693–698. Web of Science CrossRef CAS Google Scholar
Authier, A. (2001). Dynamical Theory of X-ray Diffraction. Vol. 11 of IUCr Monographs on Crystallography. Oxford University Press. Google Scholar
Baron, A. Q. R. (2016). Synchrotron Light Sources and Free-Electron Lasers, pp. 1643–1757. Cham: Springer International Publishing. Google Scholar
Bernhardt, H., Marx-Glowna, B., Schulze, K., Grabiger, B., Haber, J., Detlefs, C., Loetzsch, R., Kämpfer, T., Röhlsberger, R., Förster, E., Stöhlker, T., Uschmann, I. & Paulus, G. G. (2016). Appl. Phys. Lett. 109, 121106. Web of Science CrossRef Google Scholar
Blank, V., Kuznetsov, M., Nosukhin, S., Terentiev, S. & Denisov, V. (2007). Diamond Relat. Mater. 16, 800–804. Web of Science CrossRef CAS Google Scholar
Bonse, U. & Hart, M. (1965a). Z. Phys. 188, 154–164. CrossRef CAS Web of Science Google Scholar
Bonse, U. & Hart, M. (1965b). Appl. Phys. Lett. 6, 155–156. CrossRef Web of Science Google Scholar
Bonse, U. & Hart, M. (1966). Z. Phys. 189, 151–162. CrossRef CAS Web of Science Google Scholar
Burns, R. C., Chumakov, A. I., Connell, S. H., Dube, D., Godfried, H. P., Hansen, J. O., Härtwig, J., Hoszowska, J., Masiello, F., Mkhonza, L., Rebak, M., Rommevaux, A., Setshedi, R. & Van Vaerenbergh, P. (2009). J. Phys. Condens. Matter, 21, 364224. Web of Science CrossRef PubMed Google Scholar
Chubar, O., Geloni, G., Kocharyan, V., Madsen, A., Saldin, E., Serkez, S., Shvyd'ko, Y. & Sutter, J. (2016). J. Synchrotron Rad. 23, 410–424. Web of Science CrossRef CAS IUCr Journals Google Scholar
Chumakov, A. I., Sergeev, I., Celse, J.-P., Rüffer, R., Lesourd, M., Zhang, L. & Sánchez del Río, M. (2014). J. Synchrotron Rad. 21, 315–324. Web of Science CrossRef CAS IUCr Journals Google Scholar
Decking, W. et al. (2020). Nat. Photon. 14, 391–397. Web of Science CrossRef CAS Google Scholar
Deslattes, R. D. (1968). Appl. Phys. Lett. 12, 133–135. CrossRef CAS Web of Science Google Scholar
Feng, Y., Alonso-Mori, R., Barends, T. R. M., Blank, V. D., Botha, S., Chollet, M., Damiani, D. S., Doak, R. B., Glownia, J. M., Koglin, J. M., Lemke, H. T., Messerschmidt, M., Nass, K., Nelson, S., Schlichting, I., Shoeman, R. L., Shvyd'ko, Y. V., Sikorski, M., Song, S., Stoupin, S., Terentyev, S., Williams, G. J., Zhu, D., Robert, A. & Boutet, S. (2015). J. Synchrotron Rad. 22, 626–633. Web of Science CrossRef CAS IUCr Journals Google Scholar
Field, J. E. (1992). Editors. The Properties of Natural and Synthetic Diamond. Cambridge, Massachusetts, USA: Academic Press. Google Scholar
Geloni, G., Kocharyan, V. & Saldin, E. (2011). J. Mod. Opt. 58, 1391–1403. Web of Science CrossRef Google Scholar
Geloni, G., Kocharyan, V. & Saldin, E. (2015). arXiv:1508.04339. Google Scholar
Gerdau, E. & de Waard, H. (1999). Editors. Nuclear Resonant Scattering of Synchrotron Radiation. Hyperfine Interactions, Vol. 123, pp. 1–879. Springer. Google Scholar
Gerdau, E. & de Waard, H. (2000). Editors. Nuclear Resonant Scattering of Synchrotron Radiation. Hyperfine Interactions, Vol. 125, pp. 1–221. Springer. Google Scholar
Huang, Z. & Kim, K.-J. (2007). Phys. Rev. ST Accel. Beams, 10, 034801. Web of Science CrossRef Google Scholar
Inoue, I., Osaka, T., Hara, T., Tanaka, T., Inagaki, T., Fukui, T., Goto, S., Inubushi, Y., Kimura, H., Kinjo, R., Ohashi, H., Togawa, K., Tono, K., Yamaga, M., Tanaka, H. & Yabashi, T. I. M. (2019). Nat. Photon. 13, 1749–4893. Web of Science CrossRef Google Scholar
Inyushkin, A. V., Taldenkov, A. N., Ralchenko, V. G., Bolshakov, A. P., Koliadin, A. V. & Katrusha, A. N. (2018). Phys. Rev. B, 97, 144305. Web of Science CrossRef Google Scholar
Kolodziej, T., Vodnala, P., Terentyev, S., Blank, V. & Shvyd'ko, Y. (2016). J. Appl. Cryst. 49, 1240–1244. Web of Science CrossRef CAS IUCr Journals Google Scholar
Lübbert, D., Baumbach, T., Härtwig, J., Boller, E. & Pernot, E. (2000). Nucl. Instrum. Methods Phys. Res. B, 160, 521–527. Google Scholar
Macrander, A., Erdmann, M., Kujala, N., Stoupin, S., Marathe, S., Shi, X., Wojcik, M., Nocher, D., Conley, R., Sullivan, J., Goetze, K., Maser, J. & Assoufid, L. (2016). AIP Conf. Proc. 1741, 030030. Google Scholar
Mattenet, M., Schneider, T. & Grübel, G. (1998). J. Synchrotron Rad. 5, 651–653. Web of Science CrossRef CAS IUCr Journals Google Scholar
Min, C.-K., Nam, I., Yang, H., Kim, G., Shim, C. H., Ko, J. H., Cho, M.-H., Heo, H., Oh, B., Suh, Y. J., Kim, M. J., Na, D., Kim, C., Kim, Y., Chun, S. H., Lee, J. H., Kim, J., Kim, S., Eom, I., Kim, S. N., Koo, T.-Y., Rah, S., Shvyd'ko, Y., Shu, D., Kim, K.-J., Terentyev, S., Blank, V. & Kang, H.-S. (2019). J. Synchrotron Rad. 26, 1101–1109. Web of Science CrossRef CAS IUCr Journals Google Scholar
Nam, I., Min, C.-K., Oh, B., Kim, G., Na, D., Young Jin Suh, H. Y., Cho, M. H., Kim, C., Kim, M.-J., Shim, C. H., Jun Ho Ko, H. H., Park, J., Kim, J., Park, S., Park, G., Kim, S., Sae Hwan Chun, H. H., Lee, J. H., Kim, K. S., Eom, I., Rah, S., Deming Shu, K.-J. K., Terentyev, S., Blank, V., Shvyd'ko, Yu., Lee, S. J. & Kang, H.-S. (2021). Nat. Photon. 15, 435–441. Web of Science CrossRef CAS Google Scholar
Polyakov, S. N., Denisov, V. N., Kuzmin, N. V., Kuznetsov, M. S., Martyushov, S. Y., Nosukhin, S. A., Terentiev, S. A. & Blank, V. D. (2011). Diamond Relat. Mater. 20, 726–728. Web of Science CrossRef CAS Google Scholar
Pradhan, P., Wojcik, M., Huang, X., Kasman, E., Assoufid, L., Anton, J., Shu, D., Terentyev, S., Blank, V., Kim, K.-J. & Shvyd'ko, Y. (2020). J. Synchrotron Rad. 27, 1553–1563. Web of Science CrossRef CAS IUCr Journals Google Scholar
Prelas, M. A., Popovici, G. & Bigelow, L. K. (1998). Editors. Handbook of Industrial Diamonds and Diamond Films. New York: Marcel Dekker Inc. Google Scholar
Raubenheimer, T. (2018). Proceedings of the 60th ICFA Advanced Beam Dynamics Workshop (FLS'18), 5–9 March 2018, Shanghai, China, pp. 6–11. MOP1WA02. Google Scholar
Röhlsberger, R. (2004). Nuclear Condensed Matter Physics with Synchrotron Radiation. Basic Principles, Methodology and Applications. Vol. 208 of Springer Tracts in Modern Physics. Berlin-Heidelberg: Springer Verlag. Google Scholar
Rüffer, R. & Chumakov, A. (2016). Synchrotron Light Sources and Free-Electron Lasers, pp. 1–32. Cham: Springer International Publishing. Google Scholar
Saldin, E. L., Schneidmiller, E. A., Shvyd'ko, Yu. V. & Yurkov, M. V. (2001). Nucl. Instrum. Methods Phys. Res. A, 475, 357–362. Web of Science CrossRef CAS Google Scholar
Saldin, E. L., Schneidmiller, E. A. & Yurkov, M. V. (2000). The Physics of Free Electron Lasers. Advanced Texts in Physics. Berlin Heidelberg: Springer. Google Scholar
Shpyrko, O. G. (2014). J. Synchrotron Rad. 21, 1057–1064. Web of Science CrossRef CAS IUCr Journals Google Scholar
Shvyd'ko, Yu. (2004). X-ray Optics – High-Energy-Resolution Applications, Vol. 98 of Optical Sciences. Berlin: Springer. Google Scholar
Shvyd'ko, Yu., Blank, V. & Terentyev, S. (2017). MRS Bull. 42, 437–444. Google Scholar
Shvyd'ko, Yu. V. & Smirnov, G. V. (1990). Nucl. Instrum. Methods Phys. Res. B, 51, 452–457. Google Scholar
Shvyd'ko, Yu. V., Stoupin, S., Blank, V. & Terentyev, S. (2011). Nat. Photon. 5, 539–542. CAS Google Scholar
Shvyd'ko, Yu. V., Stoupin, S., Cunsolo, A., Said, A. & Huang, X. (2010). Nat. Phys. 6, 196–199. CAS Google Scholar
Stoupin, S. (2015). Python-DTXRD, https://www.aps.anl.gov/Science/Scientific-Software/DTXRD. Google Scholar
Stoupin, S., Shvyd'ko, Yu., Trakhtenberg, E., Liu, Z., Lang, K., Huang, X., Wieczorek, M., Kasman, E., Hammonds, J., Macrander, A. & Assoufid, L. (2016). AIP Conf. Proc. 1741, 050020. Google Scholar
Stoupin, S. & Shvyd'ko, Yu. V. (2010). Phys. Rev. Lett. 104, 085901. Web of Science CrossRef PubMed Google Scholar
Stoupin, S. & Shvyd'ko, Yu. V. (2011). Phys. Rev. B, 83, 104102. Web of Science CrossRef Google Scholar
Stoupin, S., Terentyev, S. A., Blank, V. D., Shvyd'ko, Y. V., Goetze, K., Assoufid, L., Polyakov, S. N., Kuznetsov, M. S., Kornilov, N. V., Katsoudas, J., Alonso-Mori, R., Chollet, M., Feng, Y., Glownia, J. M., Lemke, H., Robert, A., Sikorski, M., Song, S. & Zhu, D. (2014). J. Appl. Cryst. 47, 1329–1336. Web of Science CrossRef CAS IUCr Journals Google Scholar
Sumiya, H. & Tamasaku, K. (2012). Jpn. J. Appl. Phys. 51, 090102. CrossRef Google Scholar
Terentyev, S., Blank, V., Polyakov, S., Zholudev, S., Snigirev, A., Polikarpov, M., Kolodziej, T., Qian, J., Zhou, H. & Shvyd'ko, Yu. (2015). Appl. Phys. Lett. 107, 111108. Web of Science CrossRef Google Scholar
Wei, L., Kuo, P. K., Thomas, R. L., Anthony, T. R. & Banholzer, W. F. (1993). Phys. Rev. Lett. 70, 3764–3767. CrossRef PubMed CAS Web of Science Google Scholar
Yang, X. & Shvyd'ko, Yu. (2013). Phys. Rev. ST Accel. Beams, 16, 120701. Web of Science CrossRef Google Scholar
Zhu, D., Feng, Y., Stoupin, S., Terentyev, S. A., Lemke, H. T., Fritz, D. M., Chollet, M., Glownia, J. M., Alonso-Mori, R., Sikorski, M., Song, S., van Driel, T. B., Williams, G. J., Messerschmidt, M., Boutet, S., Blank, V. D., Shvyd'ko, Yu. V. & Robert, A. (2014). Rev. Sci. Instrum. 85, 063106. Web of Science CrossRef PubMed Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.