research papers
High-resolution
spectroscopy setup at the X-ray absorption spectroscopy beamline P65 of PETRA IIIaFelix-Bloch-Institut für Festkörperphysik, Universität Leipzig, Linnéstraße 5, 04103 Leipzig, Germany, bDeutsches Elektronen-Synchrotron DESY, Notkestr. 85, 22607 Hamburg, Germany, and cLaboratory for Photovoltaics, Department of Physics and Materials Science, University of Luxembourg, Rue du Brill 41, 4422 Belvaux, Luxembourg
*Correspondence e-mail: sergiu.levcenco@uni-leipzig.de
A newly designed setup to perform steady-state X-ray excited optical luminescence (XEOL) spectroscopy and simultaneous 2 thin film, single-crystalline GaN thin film and single-crystalline ZnO bulk semiconductor samples are performed.
and X-ray absorption spectroscopy characterization at beamline P65 of PETRA III is described. The setup is equipped with a He-flow cryostat and state-of-the-art optical detection system, which covers a wide wavelength range of 300–1700 nm with a high spectral resolution of 0.4 nm. To demonstrate the setup functioning, low-temperature studies on polycrystalline CuInSe1. Introduction
X-ray excited optical luminescence (XEOL) spectroscopy is a useful technique to investigate radiative recombination properties using a synchrotron radiation source (Liu & Sun, 2018; Sham, 2014; Rogalev & Goulon, 2002). The method has been increasingly used to access the optical properties of a wide range of materials and to explore the nature and structural environment of their luminescent centers, including studies of metal clusters (Grandjean et al., 2018), rare-earth and rare-earth-doped oxides (Soderholm et al., 1998; Ishii et al., 2013; Fortes et al., 2014), and various semiconductor bulk samples (Goulon et al., 1983; Pettifer & Bourdillon, 1987), thin films (Sham et al., 1993; Ishii et al., 2001) and nanostructures (Larcheri et al., 2006; Wang et al., 2014; Murphy et al., 2014). Initially, was suggested as an alternative optical detection method for X-ray absorption fine-structure (XAFS) measurements (Bianconi et al., 1978). Later on, it was discovered that in certain cases offers element- and site-selective information, especially when combined with transmission (Grandjean et al., 2018; Goulon et al., 1983; Pettifer & Bourdillon, 1987), fluorescence (Ishii et al., 2013; Wang et al., 2014; Murphy et al., 2014) and total-electron-yield (Sham et al., 1993; Wang et al., 2014; Murphy et al., 2014) measurements. However, in other cases, migration of electronic excitations and various energy-transfer processes preclude any element or site selectivity of the signal (Rogalev & Goulon, 2002; Soderholm et al., 1998). In this context and in view of elucidating the possible origin of the luminescence in the material, it becomes clear that conventional spectroscopy must support measurements whenever aiming for site-selective analysis (Boscherini, 2008). In recent years, has also been used as an alternative detection channel for mapping and imaging nanostructures with high spatial resolution by applying a synchrotron nanobeam (Steinmann et al., 2020; Hageraats et al., 2021; Lin et al., 2020).
Herein, we discuss the capabilities of a new setup designed at beamline P65 (Welter et al., 2019) of PETRA III to perform high-spectral-resolution measurements in wide spectral and temperature ranges. The main advantage of the elaborated setup is that it does not require modification of either the sample alignment or the basic instrument to perform simultaneous and measurements (2D-XEOL-XAFS). To demonstrate the setup applicability in the near-infrared (NIR), visible (VIS) and ultraviolet (UV) wavelength ranges, we performed low-temperature hard X-ray investigations on polycrystalline CuInSe2 thin film, single-crystalline GaN thin film and single-crystalline ZnO bulk semiconductor samples, respectively, at fixed X-ray energies. In addition, the simultaneous and characterization of a crystalline GaN thin film is also demonstrated.
2. Experimental
The key elements of the experimental setup for the and several photographs are given in the supporting information. The design of this setup is constrained by the vacuum requirements for the incoming monochromatic X-ray beam, the transmitted X-ray beam and the to access the full range of X-ray energies provided at beamline P65 of PETRA III at DESY in Hamburg, Germany. The main technical characteristics of this undulator beamline are an X-ray energy range of 4–44 keV, a monochromatic of up to 1012 s−1, an energy resolution ΔE/E of 1.4 × 10−4 for Si(111) and of 0.6 × 10−4 for Si(311) crystals, and a beam spot size at the sample of 1 mm × 0.5 mm in the horizontal and vertical directions, respectively (Welter et al., 2019).
measurement are shown schematically in Fig. 1To simultaneously measure x-axis direction, which coincides with the propagation of the X-ray beam (Fig. 1). The optical window made of quartz is located at 140° (XEOL, 25° opening angle) to the x-axis direction. A motorized positioner stage with rotation (360°) around the z-axis and translation (60 mm travel distance) along the z-axis enables loading several specimens on the sample holder and their adjustment with respect to the incident X-ray beam. In all measurements reported herein, samples were aligned at the incidence angle of 45°. A temperature controller (Lake Shore 335) maintains a fixed sample temperature in the range 5–300 K with an accuracy of ±0.1 K via a feedback loop with a silicon diode temperature sensor. A six-way cross-type vacuum chamber hosts the cryostat, integrates it into the end-station of the beamline, and provides access to all X-ray and optical viewports. The positioning of the chamber in the y-axis direction perpendicular to the X-ray beam is facilitated by a translation stage with 30 mm travel distance in order to align entrance window and sample with the incoming X-ray beam.
spectra and signals, a modified He-flow cryostat (Janis STVP-FTIR) was manufactured with three X-ray windows made of 25 µm Kapton and located at 180° (incoming X-ray beam, 4 mm diameter), at 90° (X-ray fluorescence, 25° opening angle) and at 0° (transmitted X-ray beam, 4 mm diameter) to theThe xyz stage with additional rotational freedom around the z-axis.
signal is collected by a 90° off-axis parabolic mirror (Thorlabs, 10 cm focal length) and coupled into an optical fiber with a reflective collimator. Aluminium or silver coating of the mirror and collimator can be selected for optimal signal collection in the UV and NIR-VIS ranges, respectively. These optical elements are the main components of the so-called optical head positioned in front of the optical viewport. In order to simplify the alignment of the optical head, a laser beam is directed at the sample through a center hole in the parabolic mirror. This laser beam can also serve as an excitation source to probe (PL) from the sample. A 633 nm was used in the measurements reported herein, while a 405 nm is also available. The alignment of the optical head is optimized with anThe optical detection system further includes a Czerny-Turner spectrograph (Shamrock 500i) with a 0.5 m focal length, a thermally cooled (Peltier type) linear InGaAs (IGA) array detector (Andor DU490A-1.7, 512 pixels) and a charge coupled device (CCD) camera (Andor DV420A-OE, 1024 × 256 pixels). The triple grating turret of this spectrograph provides the desired band-pass performance and spectral resolution in the region of interest. Two 150 lines mm−1 gratings blazed at 1250 and 500 nm and a 300 lines mm−1 grating blazed at 500 nm were used in measurements reported herein. Additionally, the spectrograph has been recently updated with a second turret hosting a 1200 lines mm−1 grating blazed at 300 nm, a 600 lines mm−1 grating blazed at 500 nm and a 300 lines mm−1 grating blazed at 1200 nm, which can be mounted upon user request. A motorized entrance slit (10 µm to 2.5 mm) of the spectrograph allows the intensity and the spectral resolution of the signal sent through the optical fiber to be adjusted, while a set of suitable long-pass filters mounted on the filter wheel of the spectrograph rejects second-order effects in the acquired spectra. An Hg-Ar light source is used to calibrate the spectrograph and to determine its spectral resolution, defined as the full width at half-maximum (FWHM) of a particular Hg or Ar line. The raw data are corrected for the spectral response of the detection system using a calibrated halogen lamp as a broadband source.
The 2 and the incoming X-ray density was 1.4 × 1012 s−1 mm−2 at 9 keV. The time per point was constant over the entire scan range and determined by the counting statistics of the signal.
detected spectra were measured in step-wise scan mode (∼355 points) starting 200 eV below the respective edge and ending 700 eV above the edge using a silicon drift detector (SDD). The irradiated sample area was 0.7 mm3. Results and discussion
3.1. measurements
The standard procedure of the e.g. below and above an of interest. In parallel, the I0 signal is measured at the first which ensures both the monitoring of the incident X-ray beam and the normalization of the intensity probed at each X-ray energy. The signal can also be measured in the absence of the X-ray beam or even the sample to investigate the presence of various artifacts in the spectrum due to the scattering light effects in the experimental hutch (Ossig et al., 2021) or possible emission from the cryostat windows, holders, substrates, etc. Once these measurements are completed and typical exposure times are determined, the and spectra are detected simultaneously during an X-ray energy scan across the specific X-ray energy region of interest.
measurement is as follows. The signal is first measured in the spectral region of interest with the CCD or IGA detector at selected excitation X-ray energies,3.2. NIR-XEOL: a case study on CuInSe2
For NIR applications, the 150 lines mm−1 grating blazed at 1250 nm and the image area of the IGA detector yield a band-pass, i.e. a spectral range or wavelength interval at fixed angle position of the grating, of about 160 nm and a spectral resolution of up to 0.6 nm (0.9 meV at λ = 922.45 nm) for a slit width of 25 µm (Table 1). This setup configuration has been utilized to study the low-temperature near the Cu K-edge on a polycrystalline CuInSe2 thin film sample as shown in Fig. 2. The CuInSe2 layer with a thickness of ∼2 µm was deposited on a glass/Mo substrate using the physical vapor deposition process (Babbe et al., 2019). To cover a wide spectral range of 0.79–1.10 eV (1127–1569 nm) for the spectrum, three center wavelengths, i.e. three different angle positions of the grating, were necessary. A spectral resolution of ∼2.4 nm (3.6 meV at λ = 922.45 nm) obtained with a slit width of 200 µm and an exposure time of 60 s were used for each center wavelength. This long exposure time is due to the small film thickness (∼2 µm) compared with the much larger absorption length (approximately 12–15 µm) at the Cu K-edge in CuInSe2. This means that less than 20% of the X-ray flux [an X-ray density of 1.4 × 1012 s−1 mm−2 (1.4 × 1014 s−1 cm−2)] is absorbed in the sample and, consequently, the excitation density for these spectra is rather low. The latter is verified with 633 nm laser excitation, showing an intense PL signal with an exposure time of only ∼1 s at moderate illumination (<1 W cm−2 or a density of 3 × 1018 s−1 cm−2). Nevertheless, the characteristic features of Cu-rich CuInSe2 PL (Babbe et al., 2019; Spindler et al., 2019) are clearly visible in the spectra presented in Fig. 2. In particular, the donor–acceptor pair type emissions DA1 at 1.00 eV, DA2 at 0.97 eV and DA3 at 0.90 eV and their longitudinal optical (LO) phonon replicas with a phonon energy of 28 meV as well as the free excitonic emission Ex at 1.04 eV are observed. In addition, there are two weak features near 1.02 eV in the lower-energy region of the Ex transition, which are usually ascribed to the bound transitions (Spindler et al., 2019). It has been found that the intensity of the DA and the excitonic emission becomes stronger by about 30% when the excitation X-ray energy varies from 8929 eV (EK − 50 eV) to 9029 eV (EK + 50 eV) across the Cu K-edge. Although intensity maps with nanoscale spatial resolution measured at 15.25 keV (just above the Rb K-edge) have been reported for Cu(In,Ga)Se2 solar cells (Ossig et al., 2021), to our knowledge, we present here for the first time energy-resolved spectra of chalcopyrite CuInSe2 material measured at different excitation X-ray energies.
|
3.3. UV-XEOL: a case study on ZnO
For measurements in the UV range, the 300 lines mm−1 grating blazed at 500 nm and the image area of the CCD detector provide a band-pass of about 170 nm with a spectral resolution of up to 0.4 nm (3.4 meV at λ = 365.01 nm) as indicated in Table 1. To demonstrate their feasibility, we performed UV-XEOL measurements on a single-crystalline ZnO- bulk sample at 10 K near the Zn K-edge as shown in Fig. 3. Here we focus only on the near-band-edge (NBE) region of ZnO, even though the measured spectral range extended from 2.66 to 4.23 eV (293 to 466 nm) and the spectra contained some contribution from the broad defect band centered at ∼2.0 eV. The investigation of this ZnO defect emission has already been reported in the literature (Sham, 2014; Goulon et al., 1983; Larcheri et al., 2006) and is beyond the scope of this work. To enhance the signal-to-noise ratio of some weak luminescence in the NBE region, an exposure time of 20 s is used. Moreover, the ZnO sample is aligned in E ⊥ c (perpendicular) polarization mode geometry, for which all A, B and C excitonic type transitions originating from the different states of the are allowed (Teke et al., 2004). The spectra are composed of the most intense bound to neutral donor (DBE) line at 3.360 eV and its LO phonon replicas with a phonon energy of 72 meV denoted as DBE-LO, DBE-2LO and DBE-3LO lines (von Wenckstern et al., 2007a,b; Wagner et al., 2011). The two-electron satellite (TES) transitions of the bound at 3.32 eV are also detected (Teke et al., 2004; Wagner et al., 2011). When the X-ray energy increases from 9609 eV (EK − 50 eV) to 9709 eV (EK + 50 eV), the DBE intensity increases by almost a factor of 6, while the intensity of the TES and DBE-LO lines increases by only a factor of 1.5.
3.4. UV-VIS a case study on GaN
The combination of the 150 lines mm−1 grating blazed at 500 nm and the image area of the CCD detector defines the largest band-pass of about 350 nm for the spectrograph (Table 1). As an illustrative example, this setup configuration enables the monitoring of both the NBE emission and the defect emission of the wide-band-gap semiconductor GaN in the extended photon energy range 1.82–3.69 eV (336–682 nm). The low-temperature measurements near the Ga K-edge (EK = 10367 eV) on a 4 µm-thick single-crystalline GaN-(0001) thin film grown on c-Al2O3 are shown in Fig. 4. An exposure time of just 1 s per spectrum was obtained with a slit width of 200 µm, however, at the cost of lower spectral resolution of ∼2.4 nm (22 meV at λ = 365.01 nm), which causes an obvious broadening of the UV lines. A spectral resolution of 0.6 nm (6 meV at λ = 365.01 nm) may be achieved by carrying out measurements with 25 µm slit width, which approaches the size of the individual pixel of the CCD camera, however, at the cost of a significantly increased exposure time. The DBE emission at 3.48 eV, the ultraviolet emission presented as a donor–acceptor pair (DAP) band at 3.28 eV and its LO phonon (91.5 meV) replicas, and a yellow defect (YL1) luminescence at 2.20 eV are clearly identified in the spectra (Reshchikov, 2021; Reshchikov et al., 2018). A note of caution is that the strain effects in GaN thin layers grown on sapphire substrates increase the band gap of GaN by 5–20 meV and as a result the spectral positions of luminescence lines are different from those of strain-free GaN samples (Reshchikov, 2021; Reshchikov et al., 2021). The small oscillation features with separation of ∼50 meV between the maxima on the band shape of the YL1 emission are most likely caused by the interference effect in the GaN thin layer (Reshchikov, 2021). Although the DBE emission is relatively weak, it exhibits the strongest increase in intensity by almost a factor of 5 when the X-ray is tuned across the Ga K-edge, compared with values of 3.5 for the YL1 band and 3 for the DAP bands.
3.5. Simultaneous and measurements: a case study on GaN
The most distinctive advantage of our (a) shows an example of the low-temperature 2D-XEOL-XAFS mapping of the YL1 defect band at 10 K on a single-crystalline GaN-(0001) thin film. Details on the X-ray energy range are given in Section 2. The inset of Fig. 5(b) shows the experiment geometry. Note that the intensity of the YL1 defect band is almost independent of temperature in the low-temperature region (Reshchikov, 2021; Reshchikov et al., 2018) and the results obtained at 15 K (discussed in Section 3.4) are also valid here. The normalized intensity is plotted in a color-coded scheme as a function of both the photon energy (x-axis) and the exciting X-ray energy (y-axis) around the Ga K-edge. The technical details of this measurement include a spectral resolution of 0.6 nm (3 meV at λ = 546.08 nm), an exposure time of 3 s and a dwell time of 0.1 s per spectrum. Note that at this exposure time the CCD detector signal (∼6 × 105 counts) almost approached its saturation point (∼6.5 × 105 counts) and, hence, the exposure time can be safely reduced by factor of five or more, still providing good quality data. A horizontal cut through the 2D map in Fig. 5(a) yields spectra at fixed X-ray energy similar to those shown in Fig. 4. A strong increase of the intensity for the YL1 band at the X-ray energy of the Ga K-edge (EK = 10367 eV) is clearly observed in excellent agreement with the spectra plotted in Fig. 4 for X-ray energies below or above the edge. In contrast, a vertical cut through the 2D map in Fig. 5(a) yields the intensity at fixed luminescence photon energy as a function of the exciting X-ray energy. As an example, we extracted and normalized the integrated intensity of the YL1 defect band in the 1.9–2.6 eV region (477–653 nm) and plotted it together with the simultaneously measured fluorescence spectrum in Fig. 5(b). The X-ray energy dependent spectrum shows not only the edge step at the Ga K-edge but also resembles all fine-structure features of the fluorescence spectrum. Interestingly, our results of a positive edge step for the YL1 band and NBE emission (presented in the previous section, see Fig. 4) are consistent with the room-temperature polarization-dependent study of the NBE emission on a 500 µm GaN- wafer (Lin et al., 2019), but they are different from earlier measurements at 80 K on a 100 µm free-standing GaN sample, which exhibited a negative edge step and inverted fine-structure features for the spectra from the blue defect emission at 2.8 eV and the green defect emission at 2.5 eV (Martínez-Criado et al., 2006). Obviously, further work is needed to resolve these somewhat controversial results on GaN. While this example on GaN was primarily given for the demonstration of the simultaneous and measurements with our setup, it should be pointed out that experiments on silver clusters in Ag-Linde Type A (LTA) zeolites (Grandjean et al., 2018), porous Si (Sham et al., 1993), dopants in TiO2:Sm (Ishii et al., 2013) and Si:Er (Ishii et al., 2001), ZnO and zinc mesotetraphenylporphyrin (Goulon et al., 1983), ZnS and ZnSe mixtures (Pettifer & Bourdillon, 1987) mixtures, and ZnO/CdS core-shells (Wang et al., 2014) yielded different XEOL-detected spectra as compared with conventional measurements performed in fluorescence or transmission modes, thus providing unique information about the nature and local structure of the luminescent centers in these materials.
cryostat is its ease to simultaneously record and spectra in the fluorescence geometry, being applicable to dilute samples or thin films, especially those grown on X-ray opaque substrates. Fig. 54. Conclusions
The performance of the new high-resolution 2 with spectroscopy. These experiments clearly show that the new setup facilitates the optical characterization of materials over wide spectral and temperature ranges, namely UV-VIS and NIR ranges and 5–300 K, respectively. Furthermore, it easily enables hard X-ray energy-dependent simultaneous XEOL-XAFS measurements. The setup is available for the user experiments and we expect that future studies will provide valuable information about the radiative recombination properties for a broad range of materials and will enable a more detailed insight into the complex relaxation and energy-transfer processes following the absorption of X-ray photons.
setup at beamline P65 of PETRA III has been demonstrated on different semiconductor materials including a single-crystalline GaN thin film and a ZnO bulk sample. We also present the first measurements on the chalcopyrite semiconductor CuInSeSupporting information
Photographs of the https://doi.org/10.1107/S1600577522007287/ok5078sup1.pdf
setup. DOI:Acknowledgements
We acknowledge DESY (Hamburg, Germany), a member of the Helmholtz Association HGF, for the provision of experimental facilities. This research was carried out at beamline P65 of PETRA III under the proposal II-20200005, and we would like to thank Dr Morgane Desmau for numerous valuable discussions and for her assistance during the measurements. The authors are grateful to the Semiconductor Physics Group at the Felix Bloch Institute for Solid State Physics at Leipzig University, in particular to Professor Michael Lorenz and Gabriele Benndorf, for the provision and PL characterization of the GaN and ZnO samples. Open Access funding enabled and organized by Projekt DEAL.
Funding information
The following funding is acknowledged: German Federal Ministry of Education and Research (grant No. 05K19OL1).
References
Babbe, F., Elanzeery, H., Wolter, M. H., Santhosh, K. & Siebentritt, S. (2019). J. Phys. Condens. Matter, 31, 425702. CrossRef PubMed Google Scholar
Bianconi, A., Jackson, D. & Monahan, K. (1978). Phys. Rev. B, 17, 2021–2024. CrossRef CAS Web of Science Google Scholar
Boscherini, F. (2008). X-ray Absorption Fine Structure in the Study of Semiconductor Heterostructures and Nanostructures, in Characterization of Semiconductor Heterostructures and Nanostructures, edited by C. Lamberti, pp. 289–330. Amsterdam: Elsevier. Google Scholar
Fortes, L. M., Gonçalves, M. C., Pereira, J. C. & D'Acapito, F. (2014). J. Non-Cryst. Solids, 402, 244–251. CrossRef CAS Google Scholar
Goulon, J., Tola, P., Lemonnier, M. & Dexpert-Ghys, J. (1983). Chem. Phys. 78, 347–356. CrossRef CAS Google Scholar
Grandjean, D., Coutiño-Gonzalez, E., Cuong, N. T., Fron, E., Baekelant, W., Aghakhani, S., Schlexer, P., D'Acapito, F., Banerjee, D., Roeffaers, M. B. J., Nguyen, M. T., Hofkens, J. & Lievens, P. (2018). Science, 361, 686–690. CrossRef CAS PubMed Google Scholar
Hageraats, S., Keune, K., Stanescu, S., Laurent, J.-M., Fresquet, W. & Thoury, M. (2021). J. Synchrotron Rad. 28, 1858–1864. Web of Science CrossRef CAS IUCr Journals Google Scholar
Ishii, M., Crowe, I. F., Halsall, M. P., Hamilton, B., Hu, Y. F., Sham, T. K., Harako, S., Zhao, X. W. & Komuro, S. (2013). J. Appl. Phys. 114, 133505. CrossRef Google Scholar
Ishii, M., Tanaka, Y., Ishikawa, T., Komuro, S., Morikawa, T. & Aoyagi, Y. (2001). Appl. Phys. Lett. 78, 183–185. CrossRef CAS Google Scholar
Larcheri, S., Armellini, C., Rocca, F., Kuzmin, A., Kalendarev, R., Dalba, G., Graziola, R., Purans, J., Pailharey, D. & Jandard, F. (2006). Superlattice Microstruct. 39, 267–274. CrossRef CAS Google Scholar
Lin, B. H., Wu, Y. C., Lee, J. F., Tang, M. T. & Hsieh, W. F. (2019). Appl. Phys. Lett. 114, 091102. CrossRef Google Scholar
Lin, B.-H., Wu, Y.-H., Li, X.-Y., Hsu, H.-C., Chiu, Y.-C., Lee, C.-Y., Chen, B.-Y., Yin, G.-C., Tseng, S.-C., Chang, S.-H., Tang, M.-T. & Hsieh, W.-F. (2020). J. Synchrotron Rad. 27, 217–221. Web of Science CrossRef CAS IUCr Journals Google Scholar
Liu, L. & Sun, X. (2018). X-ray Excited Optical Luminescence and its Applications, in Synchrotron Radiation Applications, edited by X. Zhang, pp. 493–534. Singapore: World Scientific Publishing. Google Scholar
Martínez-Criado, G., Alen, B., Homs, A., Somogyi, A., Miskys, C., Susini, J., Pereira-Lachataignerais, J. & Martínez-Pastor, J. (2006). Appl. Phys. Lett. 89, 221913. Google Scholar
Murphy, M. W., Yiu, Y. M., Ward, M. J., Liu, L., Hu, Y., Zapien, J. A., Liu, Y. K. & Sham, T. K. (2014). J. Appl. Phys. 116, 193709. CrossRef Google Scholar
Ossig, C., Strelow, C., Flügge, J., Kolditz, A., Siebels, J., Garrevoet, J., Spiers, K., Seyrich, M., Brückner, D., Pyrlik, N., Hagemann, J., Seiboth, F., Schropp, A., Carron, R., Falkenberg, G., Mews, A., Schroer, C. G., Kipp, T. & Stuckelberger, M. E. (2021). Materials, 14, 228. CrossRef Google Scholar
Pettifer, R. F. & Bourdillon, A. J. (1987). J. Phys. C.: Solid State Phys. 20, 329–335. CrossRef CAS Google Scholar
Reshchikov, M. A. (2021). J. Appl. Phys. 129, 121101. CrossRef Google Scholar
Reshchikov, M. A., Andrieiev, O., Vorobiov, M., McEwen, B., Shahedipour-Sandvik, S., Ye, D. & Demchenko, D. O. (2021). Phys. Status Solidi B, 258, 2100392. CrossRef Google Scholar
Reshchikov, M. A., Ghimire, P. & Demchenko, D. O. (2018). Phys. Rev. B, 97, 205204. CrossRef Google Scholar
Rogalev, A. & Goulon, J. (2002). X-ray Excited Optical Luminescence Spectroscopies, in Chemical Applications of Synchrotron Radiation, Part II: X-ray Applications, Vol. 12B, edited by T. K. Sham, pp. 707–760. River Edge: World Scientific. Google Scholar
Sham, T. K. (2014). Adv. Mater. 26, 7896–7901. Web of Science CrossRef CAS PubMed Google Scholar
Sham, T. K., Jiang, D. T., Coulthard, I., Lorimer, J. W., Feng, X. H., Tan, K. H., Frigo, S. P., Rosenberg, R. A., Houghton, D. C. & Bryskiewicz, B. (1993). Nature, 363, 331–334. CrossRef CAS Web of Science Google Scholar
Soderholm, L., Liu, G. K., Antonio, M. R. & Lytle, F. W. (1998). J. Chem. Phys. 109, 6745–6752. Web of Science CrossRef CAS Google Scholar
Spindler, C., Babbe, F., Wolter, M. H., Ehré, F., Santhosh, K., Hilgert, P., Werner, F. & Siebentritt, S. (2019). Phys. Rev. Mater. 3, 090302. CrossRef Google Scholar
Steinmann, R. G., Martinez-Criado, G., Salomon, D., Vitoux, H., Tucoulou, R., Villanova, J., Laboure, S., Eymery, J. & Segura-Ruiz, J. (2020). J. Synchrotron Rad. 27, 1074–1079. Web of Science CrossRef CAS IUCr Journals Google Scholar
Teke, A., Özgür, Ü., Doğan, S., Gu, X., Morkoç, H., Nemeth, B., Nause, J. & Everitt, H. O. (2004). Phys. Rev. B, 70, 195207. CrossRef Google Scholar
Wagner, M. R., Callsen, G., Reparaz, J. S., Schulze, J. H., Kirste, R., Cobet, M., Ostapenko, I., Rodt, S., Nenstiel, C., Kaiser, M., Hoffmann, A., Rodina, A. V., Phillips, M. R., Lautenschläger, S., Eisermann, S. & Meyer, B. K. (2011). Phys. Rev. B, 84, 035313. CrossRef Google Scholar
Wang, Z. Q., Wang, W., Sham, T. K. & Yang, S. G. (2014). Nanoscale, 6, 9783–9790. CrossRef CAS PubMed Google Scholar
Welter, E., Chernikov, R., Herrmann, M. & Nemausat, R. (2019). AIP Conf. Proc. 2054, 040002. Google Scholar
Wenckstern, H. von, Schmidt, H., Grundmann, M., Allen, M. W., Miller, P., Reeves, R. J. & Durbin, S. M. (2007a). Appl. Phys. Lett. 91, 022913. Google Scholar
Wenckstern, H. von, Schmidt, H., Hanisch, C., Brandt, M., Czekalla, C., Benndorf, G., Biehne, G., Rahm, A., Hochmuth, H., Lorenz, M. & Grundmann, M. (2007b). Phys. Status Solidi (RRL), 1, 129–131. Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.