research papers
A compact-rigid multi-analyser for energy and angle filtering of high-resolution X-ray experiments. Part 2. Efficiency of a single-crystal-comb
aInstitut Néel CNRS-UGA, 25 Avenue des Martyrs, 38042 Grenoble, France, bOSUG-FAME, CNRS-UGA-IRD-INRAe-MétéoFrance, 71 Avenue des Martyrs, 38000 Grenoble, France, cLAL, Univ. Paris-Sud XI, CNRS-IN2P3, Orsay, France, and dEuropean Synchrotron Radiation Facility, 71 Avenue des Martyrs, F-38000 Grenoble, France
*Correspondence e-mail: jean-louis.hodeau@neel.cnrs.fr
In memory of Alain Prat
Diffraction instruments using filtering by one or several analyser crystals exist since the 1980s and 1990s at synchrotron radiation sources, but, due to its low efficiency, this filtering is little used on laboratory sources. In order to overcome this limitation, the efficiency of a small diffraction filtering multi-analyzer block (MAD block) realized with a `single-crystal-comb' curved on a rigid support is demonstrated here. The geometry of this curved surface is logarithmic spiral and is optimized to allow multi-filtering over a relatively important diffraction angular range and to be also applicable over an X-ray spectral range. The efficiency of such a small rigid-compact MAD block consisting of this single-crystal-comb generating 20–50 Si(111) single-crystal blades, associated with a block of Soller collimators, is demonstrated. The angle between each crystal is 0.1°, so the measurement range of the comb is 2–5°. The geometry of this system has been optimized for operation with a synchrotron X-ray source over an energy range of 22 keV to 46 keV and could be used with laboratory X-ray sources (Ag Kα1, 22.1 keV). This MAD block complements and exploits the qualities of the `photon-counting' detectors which have very low intrinsic noise. Their joint efficacy is supported by powder pattern measurements of a LaB6 reference sample and of several heterogeneous samples of cultural heritage materials, carried out at 22 keV on the D2AM beamline at the ESRF. Their signal-to-noise ratio is excellent (1000/1) and allows the detection thresholds of the measurements (from 3–1% to 0.1%) to detect minor phases in the studies of `real' heterogeneous materials to be drastically improved.
Keywords: high-resolution powder diffraction; multi-analyser-diffraction filtering; heterogeneous materials; compact-rigid multi-analyser.
1. Introduction
A difficult problem to solve for structural experiments is the identification and quantification of phases in heterogeneous samples where amorphous and crystalline phases coexist and where the presence of minor phases is important. This is the case when the samples are materials for industrial application where the properties are not related to the support material but to the contribution of its few external materials/inclusions, and also when they are cultural heritage materials where the minor phases may be the signature of the origin of the raw materials or of an ancient know-how. This underlines the need to develop easily accessible high-angular-resolution experiments to perform accurate structural analyses of complex and heterogeneous materials in order to quantify their weak signals. Such studies often use the complementarity of diffraction/scattering and spectroscopic measurements; however, one difficulty of these two types of measurements is in obtaining a good energy and directional filtering of their data in order to select only the beams emitted or scattered by a dedicated part of the object. The performance of these analyses is only fully achieved if all components of the measurement chain (beam–optics–instrument–detection) have very good performances, both in terms of angular and energy resolution and detection threshold. For this purpose, the use of a crystal analyser as `diffraction filter' between the analysed object and the detector has long proved to be very effective both in spectroscopy (Zhong et al., 1999; Hazemann et al., 2009; Proux et al., 2017) and in diffraction (Cox et al., 1983; Cox, 1992; Hastings et al., 1984). These analyser crystals only detect `good' photons by acting as an energy bandpath filter and a `small-aperture slits/collimator' filter. However, the use of such a diffraction filter for accurate measurements leads to a decrease in the intensity of the signal received by the detector and therefore to long experimental times and generally to the impossibility of using laboratory X-ray sources. Consequently, these filtering optical geometries are mainly used for experiments and structural analyses using synchrotron radiation (SR) sources, whose specificities (high white beam, energy-tunable monochromatic beam, low divergence, energy resolution) make them powerful and efficient tools.
If such dedicated diffraction filters are developed to be combined with 1D or 2D detectors, several sets of data can be collected in parallel generating significant time-saving in data collection. Additionally, this combination of a `MAD filter' with a 1D or 2D `photon-counting' detector is very efficient to obtain a very good signal-to-noise ratio since these photon-counting detectors are sensitive to the individual measurement of each photon and provide measurements with almost zero residual noise over a fairly large area (Delpierre et al., 2001; Palancher et al., 2005; Basolo et al., 2007; Medjoubi et al., 2010; Bergamaschi et al., 2010; Dyadkin et al., 2016; Dejoie et al., 2018; Fitch & Dejoie, 2021). These two advances allow experiments to improve their detection thresholds of weak signals and may help to overcome some of the intensity losses of diffraction filtering. Most photon-counting detectors are multi-detectors (1D or 2D), whose small pixel size (100–200 µm) combined with a fairly large distance from the sample (200–900 mm) can complement the filtering device and allows fast and accurate angular measurement of several beams emitted by the sample over a large area. Therefore, in order to overcome the limitations of the low efficiency of diffraction filtering, it is important that these 1D–2D detectors can be combined with optics using multiple analyser crystals which allow diffraction filtering in parallel (MAD optics). This underlines the interest in developing MAD assemblies containing multiple analyser crystals for powder diffraction experiments but also for fluorescence spectroscopy and experiments, as has been done on several synchrotron beamlines (Toraya et al., 1996; Hodeau et al., 1996, 1998; Siddons et al., 1998; Zhong et al., 1999; Fitch, 2004; Gozzo et al., 2004; Patterson et al., 2005; Gozzo et al., 2006; Elkaïm et al., 2007; Lee et al., 2008; Wang et al., 2008; Tartoni et al., 2008; Thompson et al., 2009; Hazemann et al., 2009; Peral et al., 2011; Llorens et al., 2012, 2014; Sitaud et al., 2012; Fauth et al., 2013; Ketenoglu et al., 2015; Dejoie et al., 2018; Schökel et al., 2021; Fitch & Dejoie, 2021), and also for some laboratory experiments (Toraya, 2009; Bès et al., 2018; Honkanen et al., 2019).
So experimental powder diffraction beamlines using MAD arrays have been developed in SR facilities, because the ; Wittry et al., 1993; Zhong et al., 1999; Toraya, 2009). The characteristics of such a LogSpiral curve are that any analyser crystal located on it always sees its source/sample at the same ΘA, as shown for Si(111) analyser crystals in Fig. 1 and in Part 1 of this contribution (Prat & Hodeau, 2023; hereafter `Part 1'). So it would be best to build a rigid MAD system where, for a different set of experiments, the only parameters that can vary are the analyser angle ΘA and the detector angle 2ΘA. As the ΘA is imposed by the lattice distance dhkl of the (hkl) crystal analyser and by the energy of the X-ray beam used for the experiment, the positioning of these analysers + detectors are determined by the choice of an optimal energy range. These parameters also determine the length of the analyser crystals and their minimum separation, as shown in Figs. 1(a)–1(c) for three energies with the same distance D between the sample and the MAD filter. With low-energy X-rays [Fig. 1(a)], the space between the analyser crystals must be large enough to allow the beams to pass. On the contrary, with high-energy X-rays [Fig. 1(c)], it is the length of the analyser crystals that must be large enough to allow the filtering of a beam of reasonable width. Indeed, with a distance D = 460 mm and 14 mm-long analyser crystals, the width of the beams (and therefore their intensity), filtered by such a MAD system with a pitch of 1°, goes from 4.6 mm at 5 keV to 2.6 mm at 9.5 keV and it is much thinner at 40 keV. If we increase the length of the crystals (e.g. to 26 mm), the maximum width of the filtered beams is 1.6 mm at 40 keV, which increases to 4.6 mm at 9.5 keV [Fig. 1(b)], but at lower energies the experiment is impossible because the beams are partially cut by the adjacent analyser crystals. Variations in beam widths/paths are more important when the MAD filtering system is designed to be operated at low energies. From these constraints, it is preferable to choose a high-energy range of the X-ray source to develop a multi-analyser diffraction system and to define its angular pitch between each analyser, its geometry, its size and finally its number of filtered channels.
of these sources is intense, and also because their energy can be chosen and their optical system optimized. To easily obtain impacts of the incident beam on the middle of several analysing crystals, a MAD filtering system has a filtering surface which corresponds to a logarithmic spiral (`LogSpiral') mathematical curve (Sakayanagi, 1982As it is important not to limit ourselves to high-resolution measurements on SR sources, in this Part 2 we have investigated and tested a configuration that allows a small rigid MAD block to be used also for high-resolution powder diffraction analyses with laboratory X-rays sources. For this purpose, we have developed a configuration that seems to us to be the most optimal: use of MAD optics containing numerous Si(111) crystals working in reflection mode according to a geometry optimized for X-rays of energies close to that emitted by a laboratory Ag Kα1 X-ray source. Indeed, such a high energy (22 keV) makes it possible to benefit from the advantages of the low diffraction angle ΘA of the Si(111) planes. Our choice of Si(111) single crystals is guided by their good reflectivity which is associated with good angular resolution. The association of this MAD system with a Soller-collimator array is also fundamental to increase the signal-to-noise ratio of the measurements and thus to take advantage of the corresponding qualities of the 1D–2D photon-counting detectors. As reported in Part 1, such a small rigid MAD block can be realized by creating an array of 20–50 Si(111) crystals inserted in a rigid `comb-sandwich' block, but also by a specific design/fabrication of a `single-crystal-comb' block curved along a logarithmic spiral surface.
2. Specifications of the compact rigid multi-analyser
As the number of beams analysed is proportional to the crystal density of such a rigid system, it is important to have a relatively small angular pitch between them. The lower limit of this pitch is given by the manufacturing precision of the system: (i) a simple sandwich-comb of crystals, easily made by 3D printing, allows a pitch of 0.2° to be reached but not a pitch of 0.1°, (ii) a single-crystal-comb block having a rigid LogSpiral curvature, more difficult to build, enables a pitch of 0.1° to be achieved, as it offers higher precision and regularity of the orientations of each single-crystal plate of the block (Prat & Hodeau, 2023).
The results presented below relate to experiments carried out with such a rigidly curved single-crystal-comb block allowing a 0.1° pitch between 20–50 Si(111) crystals. This single-crystal MAD block is associated with a Soller-collimator block containing long collimators (60 mm) with the same angular pitch (0.1°) [Fig. 2(a)]. These two blocks were made using a multi-jet printing apparatus – ProJet MJP 3600: (i) the tungsten blades of the Soller block are placed in the slots of a 3D polymer holder, and (ii) the MAD block defines the LogSpiral curvature of its single-crystal-comb which is obtained by setting it on the corresponding curved surfaces of a second 3D polymer holder. Thanks to this process and to an ad hoc adjustment, a very good (±0.01°) accuracy is obtained for the alignment of the Si(111) crystals and allows a comb of perfect single crystals having a pitch of 0.1° to be produced, and this at a non-negligible but reasonable cost (about EUR 10000 in silicon).
The design of such a fixed MAD system requires that each beam cannot be cut by one of the adjacent analyser crystals. Therefore, to meet this requirement, we have designed a MAD system that can be used over a wide high-energy range (22 keV to 46 keV). Its distance to the sample under study was chosen to be large (D = 973 mm) to allow a small angular pitch (0.1°) between Si(111) crystals, corresponding to a distance between them of 1.7 mm. If the operating range allowed by the MAD system had been between 5 and 20 keV, the large variation in the impact of the beam on the analysers would have imposed a large interval distance between each analyser and a tenfold increase in angular pitch. For the wide high-energy range (22–46 keV) chosen for our system, the ΘA angle of the beam on the Si(111) crystals is relatively small and grazing (5.156°–2.463°), which imposes a relatively long analyser crystal length (12.6 mm) [Figs. 2(b) and 2(c)].
The alignment and angle between the analyser crystals and the downstream Soller collimators varies according to the diffraction angle ΘA of the analyser. This requires the Soller collimator to be rotated so that its position can be adjusted/optimized when there is a change in energy. The size of the fan aperture of the beams entering the Soller collimator also varies with the distance of this collimator from the sample. The MAD block is mounted on the ΘA axis of the diffractometer arm so as to have its Si(111) diffraction planes parallel to this axis; this MAD system must be rigid because its alignment must be maintained regardless of the ΘA angle. As it is essential that all analyser crystals and Sollers blades are correctly oriented, the Soller-collimator block is mounted with a goniometer head just before the 2D detector on the 2ΘA axis of the diffractometer, with its central channel aligned with the central filtered beam path of the MAD block and with its input face perpendicular to this central beam path [Figs. 2(b′) and 2(c′)]. Small vertical/horizontal translations/rotations allow this Soller block to be perfectly adjusted to associate the path of each filtered beam to a collimator. As reported in Part 1, the images of the beam directly seen on the 2D detector greatly facilitate the optimization of the positions/orientations of the MAD block and the Soller-collimator block before stiffening their alignment [Fig. 7(d) of Part 1]. This adjustment should only be made when the energy of the X-ray source changes. For the two extreme energies 22 keV and 46 keV of the chosen usable range, these two blocks were designed to be identical and to be placed at the same distance from the sample; only their position/rotation corresponding to ΘA and 2ΘA angles varies (ΘA = 5.1556° at 22 keV and ΘA = 2.4632° at 46 keV) [Figs. 2(b′) and 2(c′)]. Thanks to this simple adjustment process, a very good alignment accuracy can be achieved, allowing the use of this comb of perfect single crystals with a pitch of 0.1°.
This single-crystal-comb was cut directly from a single-crystal block of silicon, its size (110 mm) allowing the design of 50 strictly parallel Si(111) single-crystal blades spaced 1.7 mm apart [Figs. 3(a) and 3(a′)]. This comb is then bent to the desired logarithmic spiral curvature. As this bending could induce crystal stresses and some additional anticlastic curvature (Sparks et al., 1982; Batterman & Berman, 1983; Ferrer et al., 1992; Hazemann et al., 1995), the single-crystal-comb design contains a thin bond and a large hole under each crystal [Fig. 3(a′)] which allows the stresses on the crystal blades to be relaxed as the comb is bent by its supporting block [Fig. 3(a)]. The supporting MAD block contains a small mechanical fixing/stiffness system that allows the curvature of the single-crystal-comb to be optimally adjusted to the LogSpiral curvature of the support [Fig. 3(a′′)]. The efficiency of this device was tested by measurements made on a LaB6 reference sample using the goniometer of the ESRF D2AM beamline [Fig. 3(b)] (Ferrer et al., 1998; Basolo et al., 2007; Chahine et al., 2021).
3. Signal-to-noise ratio, intensity and resolution gains for powder diffraction experiments using a single-crystal-comb
The measurements and results presented below were collected with a quasi-parallel beam of 0.4 mrad using this rigid curved single-crystal-comb block associated with the Soller-collimator block, both having a 0.1° pitch. The experiment energy was 22 keV in order to easily extrapolate its potential gains to an experiment that could be performed (i) on a conventional source with an Ag Kα1 X-ray tube in order to perform measurements in a laboratory or directly at an investigation site, or (ii) on a compact inverse Compton scattering X-ray source based on the power of lasers and the sharpness of optical cavities, which can provide an intense and thin beam accessible to laboratories at a cost one or two orders of magnitude lower than that of SR sources (Variola et al., 2009; Jacquet, 2016; Eggl et al., 2016; Hornberger et al., 2019; Dupraz et al., 2020).
The single-crystal-comb is mounted on the ΘA axis of the D2AM beamline diffractometer at a distance D = 973 mm from the LaB6 sample and the Soller-collimator block is mounted with the 2D detector on its 2ΘA arm. Fig. 3(b) highlights the small size of this compact MAD system. The 2D detector is an XPAD S70 hybrid pixel photon-counting 2D detector which consists of a succession of seven chips each containing 80 × 120 pixels; the pixel size is 125 µm × 125 µm and the used distance D > 1200 mm corresponds to an angular precision of 0.006° per pixel (Basolo et al., 2007). In order to take full advantage of the background reduction achieved by the diffraction filtering, an additional shielding with a lead foil was placed between the single-crystal-comb, the Soller-collimator block and the 2D detector [Fig. 3(b)].
The 2D images of the beams received by the 2D detector visualize 20 or so beams filtered by the single-crystal-comb [Figs. 4(a), 4(a′) and 4(a′′)]; these images also allow us to more easily optimize the alignment of the position/orientation of the Soller-collimator block before its stiffening. As shown in this figure, we have a very good parallelism and regularity of the angular alignment of the Si(111) blades of the single-crystal-comb, which makes it possible to precisely define very thin detection windows on the 2D detector and therefore to have a very low background noise and a very good separation of the 20 beams filtered by each analyser crystal with an angular pitch of 0.1°.
Comparison of Figs. 4(a), 4(a′) and 4(a′′) and Figs. 4(b), 4(b′) and 4(b′′) illustrates the almost identical distribution of intensities and beam positions obtained between a first measurement of the (110) reflection of LaB6 and a second measurement made a few days later on the same reflection after assembly/disassembly of the LaB6 sample. This second measurement was collected after several sample changes and after data collections made on several cultural heritage samples at the same energy; these data illustrate the alignment stability of this MAD filter that uses a single-crystal-comb. This stability is important because it allows the measurement of about 20 diffractograms in parallel with a constant contribution from each analyser crystal: even if they show some deviations in intensity, these constant measurement ratios allow all spectra/samples to be measured under the same conditions and thus to be corrected by an ad hoc data processing, as we had observed in 1995 when the first multiple analyser system was set up at ESRF (Hodeau et al., 1996).
Figs. 4(a)–4(d) show profiles of the 20 single-crystal-comb contributions for several LaB6 reflections at low and high angles. In the outer parts of the comb, some lower intensities and asymmetric broadenings are observed and correspond to a partial measurement of the diffraction cone. These constant distortions can be related to a possible tilt of some of the Si(111) crystals or, more probably, to an imperfect residual adjustment of (i) the single-crystal-comb on the MAD block, (ii) the distances between the MAD-comb/Soller blocks and the sample, or (iii) their relative orientations on the ΘA/2ΘA circles. As these residual beam intensity variations, observed in the thin detection windows defined on the 2D detector, are constant for the different experiments at a given energy, such effects can also be taken into account and corrected by data processing of the kind recently carried out on the ESRF ID22 beamline for analysis of diagrams collected with a 2° pitch MAD system associated with a 2D photon-counting detector (Dejoie et al., 2018; Fitch & Dejoie, 2021).
Using this single-crystal-comb + Soller-collimator + photon-counting 2D detector setup we have collected the complete LaB6 diffractogram in ∼9 h with a 0.002°/1.5 sec step size/time (a continuous scanning mode can also be used). Figs. 5(a) and 5(a′) show the complete data and enlarge the residual noise of this pattern of a LaB6 reference powder collected on the D2AM beamline with a 0.4 mrad quasi-parallel beam, using the filtering of 20 analyser crystals of the single-crystal-comb. The of the intensity profiles, performed with the FullProf program (Rietveld, 1969; Rodríguez-Carvajal, 1993; Roisnel & Rodríquez-Carvajal, 2001), can be carried out with the Pearson VII pseudo-Voigt line profiles as variables (Hall et al., 1977; Thompson et al., 1987); it gives RBragg and Rf factors of 1.498 and 1.164, respectively. The of LaB6 reflections are 0.008°, their shape is a convolution of Gaussian and Lorentzian curves (η = 0.19) [Fig. 5(b)]. The intensity of the strongest LaB6 reflection is 60000 counts s−1 while the residual background is 60 counts s−1 [i.e. 3.0 counts s−1 per Si(111) channel!], which illustrates the very good signal-to-noise ratio provided by this small compact system.
Using the same alignment and adjustment of the single-crystal-comb + Soller-collimator blocks of this MAD system, we collected powder diffractograms of several complex heterogeneous samples from cultural heritage studies of the `Patrimalp' project (Patrimalp, https://patrimalp.univ-grenoble-alpes.fr; Martinetto et al., 2021; Bordet et al., 2021; Autran, 2021; Autran et al., 2021). The analysis of these cultural heritage materials is a real challenge for crystallography and, in the following experiment, illustrates the power of our MAD filter system, as they are very heterogeneous materials containing mixtures of mineral and/or organic compounds, corresponding to several crystallized and/or amorphous phases.
For these experiments, we first used a sample from Rocher du Château, a site in the Alps in Savoy-France with schematic rock paintings attributed to the Neolithic period and located on a high serpentinite cliff. This site shows rock paintings composed of a group of eight red deer and of some schematic figures (grids, dots,…); a first report of all the paintings was made in 1975 (Nelh, 1976) and they were studied in detail by Defrasne et al. (2019). In this last study, the surface information of the archaeological drawings was evaluated in detail by non-invasive in situ methods (digital microscopy and Raman spectroscopy), and some micro-samples were analysed by energy-dispersive (SEM-EDX) and X-ray diffraction (XRD) methods. We used one of these micro-sampling units to test the sensitivity of our MAD filter by performing an XRD experiment on it, with a quasi-parallel beam of 0.4 mrad at an energy of 22 keV. The corresponding powder diffractogram (Fig. 6), collected using Debye–Scherrer geometry in ∼4 h with a step size/time of 0.004°/1.5 sec, illustrates the richness/complexity of such samples that contain the signatures of various materials. The green curve in Fig. 6 corresponds to a direct data collection using a conventional 2D detector, the orange curve corresponds to the data collection using the MAD filtering using the 20 analyser crystals of the single-crystal-comb and the Soller-collimator block, and the blue curve corresponds to the signal received by a single analyser. It is obvious that the diffraction filtering has a very weak intensity response compared with that of the 2D detector, but with a single 2D detector the background signal is high and the line widths are so wide that a peak may contain one, two, three reflections…, so we cannot differentiate the diffracted lines and the crystallinity of these different phases [Figs. 6(a′) and 6(a′′)]. This powder pattern reveals about ten phases, the majority of which are phyllosilicates, already observed by Defrasne et al. (2019), which could be derived from the serpentine bedrock substrate, the most important being: chamosite (Mg5Fe5)Al2.7(Si5.7Al2.3)O20(OH)16, paragonite NaAl2(AlSi3)O10(OH)2, alurgite (K1–xNax)[Al1–y(Mg,Fe,Mn)y]2(AlSi3)O10(OH)2, lizardite Mg3(Si2O5)(OH)4, albite (NaAlSi3O8), quartz (SiO2)…. In contrast to the diagram resulting from the sum of the 20 beams filtered by the single-crystal-comb (orange curves), which allow the collection of weak refections, the diagram measured and filtered by only a single analyser crystal (blue curves) is really too weak to distinguish the weak signals that contain rich information on the minor phases.
The second cultural heritage sample we used to test the single-crystal-comb filtering system is a black powder from vessels found in houses in Pompeii. These black powders were chosen for their mention in ancient texts (Littre, 1855; Discorides, 2000) and for their possible artistic uses in antiquity; they were therefore studied to specify such uses (Tomasini et al., 2012; Cersoy et al., 2016). These samples contain pigments that can correspond to various uses: inks, cosmetics, paints. Fig. 7(b) shows the full diffraction pattern of a powder contained in a bronze container, collected with a single-crystal analyser and then using the 20 signals from the single-crystal-comb with a step size/time of 0.004°/1.5 sec. The two enlargements [Figs. 7(b′) and 7(b′′)] of this figure highlight the importance of the multi-filtering (×20) of the single-crystal-comb to generate gains in intensity and resolution that allow the visualization/quantification of different minor phase reflections. These powder diagrams show no diffuse scattering and there is a very good separation of the crystalline phases which allows their crystallinity to be estimated. Previous photographs and microanalysis of an identical sample coming from the same container showed its heterogeneity and revealed the presence of a mixture of coloured grains (dark grey, white, black and green) and a variety of compositions, notably in Ca, Cu and Si (Welcomme, 2007; Cersoy et al., 2016). Figs. 7(a) and 7(b) compare two powder diffractograms collected using Debye–Scherrer geometry on two samples coming from this same vessel (b) with the MAD single-crystal-comb on the D2AM beamline, and (a) with the ID22 high-resolution powder diffraction beamline at ESRF (Autran, 2021; Autran et al., 2021). They show a very good agreement, giving the proportions and crystallinities of a mixture of a dozen phases including gypsum (CaSO4.2H2O), calcite (CaCO3), malachite [Cu2CO3(OH)2] and quartz (SiO2). The malachite observed could be the result of degradation of the bronze vessel, the presence of gypsum and calcite having already been mentioned for the preparation of such inks and used in particular as a mineral filler to accelerate the drying process (Welcomme, 2007; Gamberini et al., 2008; Canevali et al., 2011). Comparison of the measurements made (b) with the MAD filtering setup on the D2AM beamline and (a) with the ID22 reference beamline for high-resolution powder diffraction highlights the excellent signal-to-noise ratio, the good resolution and the intensities obtained by using the filtering with the proposed combination of small rigid single-crystal-comb block + Soller-collimator block + 2D photon-counting detector. It underlines the interest of this system to carry out clean measurements of very heterogeneous ordered and/or disordered materials. Although the instrumental resolution of the MAD single-crystal-comb (FWHM ≃ 0.008°) is much larger than that of the ESRF ID22 powder diffraction reference beamline (FWHM ≃ 0.0023°) (Dejoie et al., 2018; Fitch & Dejoie, 2021), the width of the diffraction peaks of each sample is related to the crystallinity of its phases, the fluorescence is suppressed and the experimental noise is very low; therefore, its use allows the detection and quantification of minor phases of such complex heterogeneous samples.
4. Conclusion
We have developed a multiple analyser system using diffraction designed to be combined with 1D or 2D detectors for accurate structural analyses. This MAD filter has the shape of a single-crystal-comb block curved on a support surface following a logarithmic spiral curve. This system consists of a rigid single-crystal analyser array which diffracts in reflection mode and contains 20–50 single Si(111) crystal blades accurately aligned (±0.01°), each filtering the sample signal with a 0.1° pitch when it is mounted on a long arm allowing a large distance D between the sample and the MAD multi-analyser + Soller + detector setup (of the order of 800–1000 mm). The most difficult part of the realization of this MAD filter is the cutting and polishing of the single-crystal-comb itself, which has a non-negligible cost. To simplify the manufacture of the MAD filter and reduce its cost, it is possible to replace the single-crystal-comb of this MAD system by a sandwich-comb (Part 1), but the lower precision in the orientations of the successive Si(111) crystals (±0.05°) will require the design of a system with only 10–25 single-crystal blades, filtering with a pitch of 0.2° for a device with the same large distance D between the sample and the MAD filter system.
The geometry of this rigid MAD filter was optimized to be used over a wide range of synchrotron X-ray high energies with an accurate analyser crystal alignment and was tested by data collection of complete diffractograms of a LaB6 reference powder and of a few complex cultural heritage samples. For these data collections, our rigid MAD filter associates a single-crystal-comb, a Soller-collimator and a 2D photon-counting detector, which provide excellent signal-to-noise ratios, reasonable intensity gains and narrow reflection profiles. These excellent results, obtained on the D2AM beamline at an energy of 22 keV using the ESRF source, could partly be extrapolated to laboratory experiments with an Ag Kα1 tube X-ray source. The reflection profiles will probably be wider due to less efficient pre-sample parallel optics of the source/beam, and the intensities will be much lower due to the intensity of tube-type sources. In this case, the possible use of a comb composed of crystals with a larger mosaic/stresses would increase the filtered intensity but would decrease the resolution. The use of a 1D photon-counting detector instead of a 2D detector is possible, if it is chosen to filter only a narrow band of beams in the direction perpendicular to the Θ diffraction plane. In all the cases the gains obtained in intensity and mainly in signal-to-noise ratios are important and are strongly determined by the post-sample optics and therefore by the qualities of the combination of single-crystal-comb + Soller-collimator + 2D photon-counting detector.
The small size of the single-crystal-comb and Soller-collimator blocks also suggests that such a MAD filtering system could also be used to increase the resolution of some experiments using laboratory sources. The complete MAD-filtering system is small (less than 150 mm × 250 mm × 150 mm, with the addition of the 2D detector) and lightweight, allowing it to be mounted on laboratory equipment having a long 2Θ arm. Indeed, the main constraint to obtain a respectable intensity gain is that this MAD system must be rigid and able to insert 20–50 single-crystal blades, each filtering the sample signal with a 0.1° pitch, which imposes a large distance D between the sample and the multi-analyser + detector system (of the order of 800–1000 mm). If the geometry of the goniometer only allows a sample-to-detector distance of 400–500 mm, the MAD system will only be able to hold 10 to 25 single-crystal blades.
If the energy of the X-ray source is unique, as in the case for a laboratory device using only an Ag Kα1 X-ray source, it is possible to reduce the variability of the settings by making a rigid system consisting of a single compact-rigid block that incorporates both the MAD comb and the Soller-collimator array. The final alignment of such a system would first need to be optimized and then rigidly fixed to the instrument during its assembly, prior to its experimental uses in the laboratory. In this case it might also be possible to use mosaic crystals instead of perfect Si(111) crystals to match the resolution offered by the other components of a laboratory source and to use a sandwich-comb block allowing filtering with a 0.2° pitch.
The benefits of such an improvement in the resolution of powder diffraction instruments/experiments and in particular in their signal-to-noise ratios are also essential for studies of mixed samples containing both crystalline and amorphous components. Although, after diffraction filtering, the measured intensities can be relatively low, the very good signal-to-noise ratio given by the MAD single-crystal-comb allows minority phases to be extracted and also the diffusion of amorphous phases to be quantified when they are present. This gain is essential for studies of complex heterogeneous materials such as `real' materials of industrial type, cultural heritage, etc., as these materials generally present difficult problems in identifying and quantifying their many minor phases.
Acknowledgements
We thank the X-ray Optics ESRF and particularly R. Verbeni, and C. Lapras of the Crystal Analyser Laboratory, for their support in cutting the single-crystal-comb. We thank the ESRF for provision of experimental time on beamlines BM02-D2AM and ID22. We also thank the referees for their relevant comments and suggestions.
References
Autran, P. O. (2021). Thesis, Université Grenoble Alpes, France (https://www.theses.fr/2021GRALY048). Google Scholar
Autran, P. O., Dejoie, C., Bordet, P., Hodeau, J. L., Dugand, C., Gervason, M., Anne, M. & Martinetto, P. (2021). Anal. Chem. 93, 1135–1142. CrossRef CAS PubMed Google Scholar
Basolo, S., Bérar, J.-F., Boudet, N., Breugnon, P., Caillot, B., Clemens, J.-C., Delpierre, P., Dinkespiler, B., Hustache, S., Koudobine, I., Meessen, C., Menouni, M., Mouget, C., Palancher, H., Pangaud, P., Potheau, R. & Vigeolas, E. (2007). J. Synchrotron Rad. 14, 151–157. Web of Science CrossRef CAS IUCr Journals Google Scholar
Batterman, B. W. & Berman, L. (1983). Nucl. Instrum. Methods Phys. Res. 208, 327–331. CrossRef Web of Science Google Scholar
Bergamaschi, A., Cervellino, A., Dinapoli, R., Gozzo, F., Henrich, B., Johnson, I., Kraft, P., Mozzanica, A., Schmitt, B. & Shi, X. (2010). J. Synchrotron Rad. 17, 653–668. Web of Science CrossRef CAS IUCr Journals Google Scholar
Bès, R., Ahopelto, T., Honkanen, A. P., Huotari, S., Leinders, G., Pakarinen, J. & Kvashnina, K. (2018). J. Nucl. Mater. 507, 50–53. Google Scholar
Bordet, P., Kergourlay, F., Pinto, A., Blanc, N. & Martinetto, P. (2021). J. Anal. At. Spectrom. 36, 1724–1734. CrossRef CAS Google Scholar
Canevali, C., Gentile, P., Orlandi, M., Modugno, F., Lucejko, J. J., Colombini, M. P., Brambilla, L., Goidanich, S., Riedo, C., Chiantore, O., Baraldi, P., Baraldi, C. & Gamberini, M. C. (2011). Anal. Bioanal. Chem. 401, 1801–1814. Web of Science CrossRef CAS PubMed Google Scholar
Cersoy, S., Martinetto, P., Bordet, P., Hodeau, J. L., Van Elslande, E. & Walter, P. (2016). J. Appl. Cryst. 49, 585–593. CrossRef CAS IUCr Journals Google Scholar
Chahine, G. A., Blanc, N., Arnaud, S., De Geuser, F., Guinebretière, R. & Boudet, N. (2021). Metals, 9, 352–364. CrossRef Google Scholar
Cox, D. E. (1992). Synchrotron Radiation Crystallography, pp. 186–254. New York: Academic Press. Google Scholar
Cox, D. E., Hastings, J. B., Thomlinson, W. & Prewitt, C. T. (1983). Nucl. Instrum. Methods Phys. Res. 208, 573–578. CrossRef CAS Web of Science Google Scholar
Defrasne, C., Chalmin, E., Bellot-Gurlet, L., Thirault, E. & Andre, G. (2019). Archaeol. Anthropol. Sci. 11, 6065–6091. CrossRef Google Scholar
Dejoie, C., Coduri, M., Petitdemange, S., Giacobbe, C., Covacci, E., Grimaldi, O., Autran, P.-O., Mogodi, M. W., Šišak Jung, D. & Fitch, A. N. (2018). J. Appl. Cryst. 51, 1721–1733. Web of Science CrossRef CAS IUCr Journals Google Scholar
Delpierre, P., Berar, J. F., Blanquart, L., Caillot, B., Clemens, J. C. & Mouget, C. (2001). IEEE Trans. Nucl. Sci. 48, 987–991. Web of Science CrossRef Google Scholar
Dioscorides (2000). De Materia Medica, Five Books in One Volume: A new English translation by T. A. Osbaldeston. Johannesburg: IBIDIS Press. Google Scholar
Dupraz, K., Alkadi, M., Alves, M., Amoudry, L., Auguste, D., Babigeon, J., Baltazar, M., Benoit, A., Bonis, J., Bonenfant, J., Bruni, C., Cassou, K., Cayla, J., Chabaud, T., Chaikovska, I., Chance, S., Chaumat, V., Chiche, R., Cobessi, A., Cornebise, P., Dalifard, O., Delerue, N., Dorkel, R., Douillet, D., Dugal, J., El Kamchi, N., El Khaldi, M., Ergenlik, E., Favier, P., Fernandez, M., Gamelin, A., Garaut, J., Garolfi, L., Gauron, P., Gauthier, F., Gonnin, A., Grasset, D., Guerard, E., Guler, H., Haissinski, J., Herry, E., Iaquaniello, G., Jacquet, M., Jules, E., Kubytskyi, V., Langlet, M., Le Barillec, T., Ledu, J., Leguidec, D., Leluan, B., Lepercq, P., Letellier-Cohen, F., Marie, R., Marrucho, J., Martens, A., Mageur, C., Mercadier, G., Mercier, B., Mistretta, E., Monard, H., Moutardier, A., Neveu, O., Nutarelli, D., Omeich, M., Peinaud, Y., Petrilli, Y., Pichet, M., Plaige, E., Prévost, C., Rudnicky, P., Soskov, V., Taurigna-Quéré, M., Trochet, S., Vallerand, C., Vitez, O., Wicek, F., Wurth, S., Zomer, F., Alexandre, P., Ben El Fekih, R., Berteaud, P., Bouvet, F., Cuoq, R., Diaz, A., Dietrich, Y., Diop, M., Pedeau, D., Dupuy, E., Marteau, F., Bouvet, F., Gamelin, A., Helder, D., Hubert, N., Veteran, J., Labat, M., Lestrade, A., Letrésor, A., Lopes, R., Loulergue, A., Louvet, M., Louvet, M., Marchand, P., El Ajjouri, M., Muller, D., Nadji, A., Nadolski, L., Nagaoka, R., Petit, S., Pollina, J., Ribeiro, F., Ros, M., Salvia, J., Bobault, S., Sebdaoui, M., Sreedharan, R., Bouanai, Y., Hazemann, J., Hodeau, J., Roy, E., Jeantet, P., Lacipière, J., Robert, P., Horodynski, J., Bzyl, H., Chapelle, C., Biagini, M., Walter, P., Bravin, A., Del Net, W., Lahéra, E., Proux, O., Elleaume, H. & Cormier, E. (2020). Phys. Open, 5, 100051. CrossRef Google Scholar
Dyadkin, V., Pattison, P., Dmitriev, V. & Chernyshov, D. (2016). J. Synchrotron Rad. 23, 825–829. Web of Science CrossRef CAS IUCr Journals Google Scholar
Eggl, E., Dierolf, M., Achterhold, K., Jud, C., Günther, B., Braig, E., Gleich, B. & Pfeiffer, F. (2016). J. Synchrotron Rad. 23, 1137–1142. Web of Science CrossRef IUCr Journals Google Scholar
Elkaïm, E., Legrand, F., Guillet, G., Mezière, H. & Ravy, S. (2007). Acta Cryst. A63, s285. CrossRef IUCr Journals Google Scholar
Fauth, F., Peral, I., Popescu, C. & Knapp, M. (2013). Powder Diffr. 28, S360–S370. Web of Science CrossRef CAS Google Scholar
Ferrer, J.-L., Simon, J.-P., Bérar, J.-F., Caillot, B., Fanchon, E., Kaïkati, O., Arnaud, S., Guidotti, M., Pirocchi, M. & Roth, M. (1998). J. Synchrotron Rad. 5, 1346–1356. Web of Science CrossRef CAS IUCr Journals Google Scholar
Ferrer, S., Krisch, M., de Bergevin, F. & Zontone, F. (1992). Nucl. Instrum. Methods Phys. Res. A, 311, 444–447. CrossRef Google Scholar
Fitch, A. & Dejoie, C. (2021). J. Appl. Cryst. 54, 1088–1099. CrossRef CAS IUCr Journals Google Scholar
Fitch, A. N. (2004). J. Res. Natl Inst. Stand. Technol. 109, 133–142. Web of Science CrossRef CAS PubMed Google Scholar
Gamberini, M. C., Baraldi, C., Palazzoli, F., Ribechini, E. & Baraldi, P. (2008). Vib. Spectrosc. 47, 82–90. Web of Science CrossRef CAS Google Scholar
Gozzo, F., De Caro, L., Giannini, C., Guagliardi, A., Schmitt, B. & Prodi, A. (2006). J. Appl. Cryst. 39, 347–357. Web of Science CrossRef CAS IUCr Journals Google Scholar
Gozzo, F., Schmitt, B., Bortolamedi, T., Giannini, C., Guagliardi, A., Lange, M., Meister, D., Maden, D., Willmott, P. & Patterson, B. D. (2004). J. Alloys Compd. 362, 206–217. Web of Science CrossRef CAS Google Scholar
Hall, M. M., Veeraraghavan, V. G., Rubin, H. & Winchell, P. G. (1977). J. Appl. Cryst. 10, 66–68. CrossRef IUCr Journals Web of Science Google Scholar
Hastings, J. B., Thomlinson, W. & Cox, D. E. (1984). J. Appl. Cryst. 17, 85–95. CrossRef CAS Web of Science IUCr Journals Google Scholar
Hazemann, J. L., Nayouf, K. & de Bergevin, F. (1995). Nucl. Instrum. Methods Phys. Res. B, 97, 547–550. CrossRef CAS Web of Science Google Scholar
Hazemann, J.-L., Proux, O., Nassif, V., Palancher, H., Lahera, E., Da Silva, C., Braillard, A., Testemale, D., Diot, M.-A., Alliot, I., Del Net, W., Manceau, A., Gélébart, F., Morand, M., Dermigny, Q. & Shukla, A. (2009). J. Synchrotron Rad. 16, 283–292. Web of Science CrossRef CAS IUCr Journals Google Scholar
Hodeau, J. L., Bordet, P., Anne, M., Prat, A., Fitch, A., Dooryhee, E. & Vaughan, G. (1996). 45th Denver X-ray Conference & Powder Satellite Meeting of the XVII Congress of the IUCr, 3–8 August 1996, Denver, USA. Google Scholar
Hodeau, J.-L., Bordet, P., Anne, M., Prat, A., Fitch, A. N., Dooryhee, E., Vaughan, G. & Freund, A. K. (1998). Proc. SPIE, 3448, 353–361. CrossRef Google Scholar
Honkanen, A. P., Ollikkala, S., Ahopelto, T., Kallio, A. J., Blomberg, M. & Huotari, S. (2019). Rev. Sci. Instrum. 90, 033107. CrossRef PubMed Google Scholar
Hornberger, B., Kasahara, J., Gifford, M., Ruth, R. & Loewen, R. (2019). Proc. SPIE, 11110, 1111003. Google Scholar
Jacquet, M. (2016). Phys. Med. 32, 1790–1794. CrossRef PubMed Google Scholar
Ketenoglu, D., Harder, M., Klementiev, K., Upton, M., Taherkhani, M., Spiwek, M., Dill, F.-U., Wille, H.-C. & Yavaş, H. (2015). J. Synchrotron Rad. 22, 961–967. Web of Science CrossRef CAS IUCr Journals Google Scholar
Lee, P. L., Shu, D., Ramanathan, M., Preissner, C., Wang, J., Beno, M. A., Von Dreele, R. B., Ribaud, L., Kurtz, C., Antao, S. M., Jiao, X. & Toby, B. H. (2008). J. Synchrotron Rad. 15, 427–432. Web of Science CrossRef CAS IUCr Journals Google Scholar
Littré, E. (1855). Translator of Histoire Naturelle, Livre XXXV by Pline l'Ancien, https://www.mediterranees.net/geographie/pline/livre34.html. Google Scholar
Llorens, I., Lahera, E., Delnet, W., Proux, O., Braillard, A., Hazemann, J. L., Prat, A., Testemale, D., Dermigny, Q., Gelebart, F., Morand, M., Shukla, A., Bardou, N., Ulrich, O., Arnaud, S., Berar, J. F., Boudet, N., Caillot, B., Chaurand, P., Rose, J., Doelsch, E., Martin, P. & Solari, P. L. (2012). Rev. Sci. Instrum. 83, 063104. Web of Science CrossRef PubMed Google Scholar
Llorens, I., Solari, P. L., Sitaud, B., Bes, R., Cammelli, S., Hermange, H., Othmane, G., Safi, S., Moisy, P., Wahu, S., Bresson, C., Schlegel, M. L., Menut, D., Bechade, J. L., Martin, P., Hazemann, J. L., Proux, O. & Den Auwer, C. (2014). Radiochim. Acta, 102, 957–972. CrossRef CAS Google Scholar
Martinetto, P., Blanc, N., Bordet, P., Champdavoine, S., Fabre, F., Guiblain, T., Hodeau, J.-L., Lelong, F., Leynaud, O., Prat, A., Pouyet, E., Uher, E. & Walter, Ph. (2021). J. Cult. Herit. 47, 89–99. CrossRef Google Scholar
Medjoubi, K., Bucaille, T., Hustache, S., Bérar, J.-F., Boudet, N., Clemens, J.-C., Delpierre, P. & Dinkespiler, B. (2010). J. Synchrotron Rad. 17, 486–495. Web of Science CrossRef CAS IUCr Journals Google Scholar
Nelh, G. (1976). Bull. Info. GERSAR, 3, 71–76. Google Scholar
Palancher, H., Hodeau, J.-L., Pichon, C., Bérar, J.-F., Lynch, J., Rebours, B. & Rodriguez-Carvajal, J. (2005). Angew. Chem. Int. Ed. 44, 1725–1729. Web of Science CrossRef CAS Google Scholar
Patterson, B. D., Abela, R., Auderset, H., Chen, Q., Fauth, F., Gozzo, F., Ingold, G., Kühne, H., Lange, M., Maden, D., Meister, D., Pattison, P., Schmidt, Th., Schmitt, B., Schulze-Briese, C., Shi, M., Stampanoni, M. & Willmott, P. R. (2005). Nucl. Instrum. Methods Phys. Res. A, 540, 42–67. Web of Science CrossRef CAS Google Scholar
Peral, I., McKinlay, J., Knapp, M. & Ferrer, S. (2011). J. Synchrotron Rad. 18, 842–850. Web of Science CrossRef IUCr Journals Google Scholar
Prat, A. & Hodeau, J. L. (2023). J. Synchrotron Rad. 30, this issue. Google Scholar
Proux, O., Lahera, E., Del Net, W., Kieffer, I., Rovezzi, M., Testemale, D., Irar, M., Thomas, S., Aguilar-Tapia, A., Bazarkina, E. F., Prat, A., Tella, M., Auffan, M., Rose, J. & Hazemann, J.-L. (2017). J. Environ. Qual. 46, 1146–1157. Web of Science CrossRef CAS PubMed Google Scholar
Rietveld, H. M. (1969). J. Appl. Cryst. 2, 65–71. CrossRef CAS IUCr Journals Web of Science Google Scholar
Rodríguez-Carvajal, J. (1993). Physica B, 192, 55–69. CrossRef Web of Science Google Scholar
Roisnel, T. & Rodríquez-Carvajal, J. (2001). Mater. Sci. Forum, 378–381, 118–123. Web of Science CrossRef CAS Google Scholar
Sakayanagi, Y. (1982). Jpn. J. Appl. Phys. 21, L225–L226. CrossRef Web of Science Google Scholar
Schökel, A., Etter, M., Berghäuser, A., Horst, A., Lindackers, D., Whittle, T. A., Schmid, S., Acosta, M., Knapp, M., Ehrenberg, H. & Hinterstein, M. (2021). J. Synchrotron Rad. 28, 146–157. CrossRef IUCr Journals Google Scholar
Siddons, D. P., Yin, Z., Furenlid, L., Pietraski, P., Li, Z. & Harlow, R. (1998). Proc. SPIE, 3448, 120–131. CrossRef CAS Google Scholar
Sitaud, B., Solari, P. L., Schlutig, S., Llorens, I. & Hermange, H. (2012). J. Nucl. Mater. 425, 238–243. Web of Science CrossRef CAS Google Scholar
Sparks, C. J., Ice, G. E., Wong, J. & Batterman, B. W. (1982). Nucl. Instrum. Methods Phys. Res. 195, 73–78. CrossRef Google Scholar
Tartoni, N., Thompson, S. P., Tang, C. C., Willis, B. L., Derbyshire, G. E., Wright, A. G., Jaye, S. C., Homer, J. M., Pizzey, J. D. & Bell, A. M. T. (2008). J. Synchrotron Rad. 15, 43–49. Web of Science CrossRef IUCr Journals Google Scholar
Thompson, P., Cox, D. E. & Hastings, J. B. (1987). J. Appl. Cryst. 20, 79–83. CrossRef ICSD CAS Web of Science IUCr Journals Google Scholar
Thompson, S. P., Parker, J. E., Potter, J., Hill, T. P., Birt, A., Cobb, T. M., Yuan, F. & Tang, C. C. (2009). Rev. Sci. Instrum. 80, 075107. Web of Science CrossRef PubMed Google Scholar
Tomasini, E., Siracusano, G. & Maier, M. S. (2012). Microchem. J. 102, 28–37. Web of Science CrossRef CAS Google Scholar
Toraya, H. (2009). J. Appl. Cryst. 42, 485–489. Web of Science CrossRef CAS IUCr Journals Google Scholar
Toraya, H., Hibino, H. & Ohsumi, K. (1996). J. Synchrotron Rad. 3, 75–83. CrossRef CAS Web of Science IUCr Journals Google Scholar
Variola, A., Brasile, J. P., Bruni, C., Chehab, R., Chiche, R., Cizeron, R., Couchot, F., Fedala, Y., Haissinski, J., Jacquet, M., Jehanno, D., Lacroix, M., Lepercq, P., Mouton, B., Roux, R., Soskov, V., Vivoli, A. & Zomer, F. (2009). Nucl. Instrum. Methods Phys. Res. A, 608, S83–S86. CrossRef CAS Google Scholar
Wang, J., Toby, B. H., Lee, P. L., Ribaud, L., Antao, S. M., Kurtz, C., Ramanathan, M., Von Dreele, R. B. & Beno, M. A. (2008). Rev. Sci. Instrum. 79, 085105. Web of Science CrossRef PubMed Google Scholar
Welcomme, E. (2007). Thesis, Université Pierre et Marie Curie, Paris, France (https://www.theses.fr/2007PA066523). Google Scholar
Wittry, D. B., Chang, W. Z. & Li, R. Y. (1993). J. Appl. Phys. 74, 3534–3540. CrossRef CAS Google Scholar
Zhong, Z., Chapman, L. D., Bunker, B. A., Bunker, G., Fischetti, R. & Segre, C. U. (1999). J. Synchrotron Rad. 6, 212–214. Web of Science CrossRef CAS IUCr Journals Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.