Fly-scan-oriented motion analyses and upgraded beamline integration architecture for the High-Dynamic Double-Crystal Monochromator at Sirius/LNLS
aBrazilian Synchrotron Light Laboratory (LNLS), Brazilian Center for Research in Energy and Materials (CNPEM), Rua Giuseppe Máximo Scolfaro, 10000 Campinas, São Paulo, Brazil, bEindhoven University of Technology, Department of Mechanical Engineering, Control Systems Technology Group, De Zaale, 5612 AJ Eindhoven, The Netherlands, and cTNO, Optomechatronics Department, Stieltjesweg 1, 2628 CK Delft, The Netherlands
*Correspondence e-mail: renan.geraldes@lnls.br
The High-Dynamic Double-Crystal Monochromator (HD-DCM) is a mechatronic system with unique control-based architecture and deep paradigm changes as compared with traditional beamline monochromators. Aiming at unprecedented inter-crystal positioning stability in vertical-bounce double-crystal monochromators (DCMs) of the order of 10 nrad RMS (1 Hz to 2.5 kHz), and not only in fixed-energy but also in fly-scan operation, it has been developed according to a `first-time right' predictive design approach for hard X-ray beamlines at Sirius, the fourth-generation light source at the Brazilian Synchrotron Light Laboratory (LNLS/CNPEM). This work explores some of the challenges that emerge with this new technology and presents the latest commissioning results that demonstrate the unparallel performances of the HD-DCM at the undulator-based EMA (Extreme Methods of Analysis) beamline at Sirius. With the enabled fast spectroscopy fly-scan possibilities, a new energy-tuning evaluation method, based on wave-propagation simulations, becomes part of a motion-oriented analysis that is carried out to derive the multi-axis non-linear positioning problem, covering not only energy selection and fixed exit in the HD-DCM but also the
of an adjustable-phase undulator (APU). The HD-DCM control scheme and its flexible operation modes are described in detail as well. Furthermore, a new integration topology between the HD-DCM and EMA's APU, coming already close to ultimate motion levels, is described and validated.Keywords: vertical double-crystal monochromator; synchrotron instrumentation; high-performance mechatronics; continuous energy scanning; beamline integration.
1. Introduction
At synchrotron light sources, hard X-ray monochromatic beams are often provided by monochromators that are implemented as a non-dispersive parallel arrangement of two identical monocrystalline structures. Constructively, this arrangement can be implemented in two basic ways, namely: (1) a single crystal with a groove, a design concept known as channel-cut; and (2) two independent crystals, generally referred to as a double-crystal monochromator (DCM). This work is focused on DCMs, whose key conceptual advantage over channel-cuts is the ability to keep a fixed-exit beam independently of the rotation of the crystals for energy selection, which can be important for beamline alignment purposes and relative motion of the beam with respect to the sample under analysis.
The first DCMs were developed in the late 1970s and early 1980s (Hastings, 1977; Lemonnier et al., 1978; Golovchenko et al., 1981; Hussain et al., 1982; Cowan et al., 1983), since when incremental upgrades have been gradually implemented to adapt to ever more stringent requirements over time (Kelly et al., 2013; Kristiansen et al., 2015; Waterstradt et al., 2018; Baker et al., 2018; Dolbnya et al., 2018; Fan et al., 2020). This evolutionary approach has generally persisted even after the workshop on X-ray DCMs, held in 2014 by the European Synchrotron Radiation Facility in Grenoble, France (https://www.esrf.eu/home/events/conferences/2014/double-crystal-monochromator-workshop.html), in which it was recognized by a great number of worldwide experts in the field that no existing DCM was able to meet some of the requirements of the new-generation beamlines. Indeed, starting in the 2010s, the new fourth-generation storage rings push the X-ray and coherence fraction to unprecedented levels, opening unique science opportunities in terms of temporal and spatial resolutions (Eriksson et al., 2014; Hettel, 2014), but require that the beamline instrumentation performs accordingly. Still, after so many years of development, the main recurrent challenge in the DCMs is found to be keeping the inter-crystal parallelism, so that variations in and beam position are kept within acceptable levels, while handling multiple moving axes, high power loads, radiation and in-vacuum operation.
Aiming at improved fixed-energy stability and unlocking high-performance fly-scan perspectives with the inter-crystal parallelism target of 10 nrad RMS (1 Hz to 2.5 kHz), the Brazilian Synchrotron Light Laboratory (LNLS/CNPEM) has developed the High-Dynamic Double-Crystal Monochromator (HD-DCM) for the fourth-generation light source Sirius (Liu et al., 2021), the first DCM to implement a high-performance control-based isolated mechatronic architecture. Herein, fixed-energy or stand-still is understood as all motion axes in position during data acquisition, which can also be related to traditional step-scan operation, whereas fly-scan refers to continuous synchronous motion of axes with simultaneous data acquisition. The first HD-DCM unit has been operational at the MANACÁ (MAcromolecular micro and NAno CrystAllography) beamline since 2020, mostly working on fixed-energy experiments, whereas the second unit at the EMA (Extreme Methods of Analysis) beamline has been operational since 2021, where the scanning perspectives started to be explored.
To advocate in favor of its powerful technology and the applied development methodologies, and allay insecurities in the community, a lot has been shared over the years at topical conferences: the basic conceptual design, mechatronic principles and thermal management were described by Geraldes et al. (2017a,b) and Saveri Silva et al. (2017); the first results of in-air validation of the core, together with system identification and control techniques in the prototyping hardware, were shown by Caliari et al. (2018) and Moreno et al. (2018); the offline performance of the full in-vacuum cryocooled system, including energy fly-scan, was demonstrated by Geraldes et al. (2018); the dynamic modeling work, updated control design and NI LabVIEW FPGA implementation in the final NI's CompactRIO (cRIO) were addressed by Geraldes et al. (2020), Caliari et al. (2020) and Moraes et al. (2020); calibration and commissioning procedures, together with the first experimental results with beam, were shown by Geraldes et al. (2021a); and, finally, integration aspects with undulator sources started to be discussed by Geraldes et al. (2021b). More recently, the detailed mechanical design and mechatronic architecture of the HD-DCM has been thoroughly described by Geraldes et al. (2022a,b); whereas, together with updated commissioning results, a didactic discussion on the fundamental conceptual differences between the HD-DCM and standard DCM mechanical architectures has been provided by Geraldes et al. (2022c).
This work mainly focuses on detailing and experimentally validating the positioning problem formulations for both the HD-DCM and its current adjustable-phase undulator (APU) source from an energy fly-scan (or spectroscopy) perspective. Indeed, although basic geometrical and physical equations can be found elsewhere (Hidas et al., 2022), to the best of the authors' knowledge a mechatronic approach concerning ranges, velocities, acceleration levels, forces and torques has not been covered in the literature yet — perhaps due to step-based actuation technologies and dominant fly-scan limitations related to motion errors in standard DCMs. With the HD-DCM, however, this understanding is essential to realize the full potential of the instrument, allowing for high-stability fixed-exit spectroscopy experiments to reach the order of 1 s, i.e. a time reduction of up to two or three orders of magnitude with respect to standard step-scans — which might drastically increase experimental throughput and/or even create new scientific perspectives. To that end, a partly tutorial approach is chosen here, such that the well known geometrical and physical properties within a DCM scope can be revisited from the necessary mechatronic approach, from where practical limits, tuning tolerances, calibration strategies and advanced beamline integration needs can be derived.
This paper is organized as follows. Firstly, in Section 2, the energy selection and fixed-exit concepts of a DCM, both in nominal geometry and including crystal miscut, are used to derive the non-linear characteristics of a DCM and its implications in fly-scans. Then, in Section 3, the relevant aspects of APUs regarding their integration with the HD-DCM, particularly focusing on tuning tolerances and fly-scan positioning demands, are developed. Next, Section 4 describes the generic and flexible beamline integration architecture developed for the HD-DCM, as well as its operation modes, and the latest communication topology with its APU source, which finally enabled practical high-performance fly-scan spectroscopy. After that, Section 5 summarizes some experimental results at EMA, including: (1) stand-still inter-crystal stability measurements via rocking-curve edges and knife-edge; (2) fly-scan motion performance evaluation according to the new integration architecture between the HD-DCM and APU; and (3) a long-term scientific high-pressure commissioning experiment via step and fly-scan spectroscopy. Conclusions and further considerations are given in Section 6. Also, a list of abbreviations is provided for general reference in Table 1.
|
2. DCM positioning problem formulation
The main geometrical concepts of a DCM are illustrated in Fig. 1, which is thoroughly explored in the following subsections to first describe a nominal geometry, then elaborate on crystal asymmetry effects, and finally derive the main time-dependent fly-scan aspects, with highly non-linear motion characteristics that pose challenging requirements in terms of actuation, metrology, control and integration.
2.1. Nominal case
The energy selection in a DCM results from Bragg's law of diffraction (Ashcroft, 2003), which describes that only photons of a given fundamental energy E and its harmonics are diffracted for a given incidence θB, i.e.
where d is the parameter, also known as d-spacing, n a positive integer that defines the diffraction order, h is Planck's constant and c is the speed of light. Thus, different energies can be selected by adjusting θB. Yet, in reality, the and the energy are not discrete values, but narrow distributions defined by an intrinsic crystallographic property, the so-called Darwin width (Koningsberger & Prins, 1988), such that the monochromatic beam results from a convolution of the diffraction in both crystals.
Hence, for two crystals with identical d, an ideal monochromatic beam, i.e. parallel to the incoming beam and with fixed exit, is nominally achieved by having parallel crystals and adjusting the gap g between them, such that a constant offset H is obtained according to
This is illustrated for two arbitrary angles in the side-view schematic of Fig. 1(a), which also follows the specific geometrical architecture chosen for the HD-DCM, namely: with the center of rotation of the crystals set on the surface of the first crystal; and with a long second crystal to comply with the beam walk for the different energies — i.e. preventing a complementary motion stage that otherwise would be needed to longitudinally move a short second crystal according to the different downstream positions of the beam after the diffraction in the first crystal.
Partially differentiating (1) with respect to E and d individually, and applying simple manipulation, it can be shown that
from where a few aspects can be highlighted. Firstly, it can be seen that the sensitivity to variations in θB in the energy selection starts from zero at the upper limit of θB = π/2 and tends to infinity as the moves towards θB = 0. Secondly, a given percentage of change in d-spacing is one-to-one related to the percentage of change in energy. Thus, if different d-spacings are found in the two crystals, an ideal energy matching for maximum would be related to slightly different θB in the crystals, such that the monochromatic beam would no longer be exactly parallel to the incoming beam, and the magnitude of the deviation would be variable over the operational energy range.
In practice, even for nominally identical crystals, d-spacing variations may result from crystal imperfections or clamping distortions, but they are most noticeably related to thermal effects. Indeed, d-spacing variations due to together with local curvature in the are well known issues in DCMs (and other types of monochromators) due to the local heat loads deposited in the first crystal, which may reach hundreds of W mm−2 (Zhang et al., 2013). This effect tends to be highly non-linear in energy, since it may depend on: (1) the power variation as a function of energy for undulators; (2) the particular crystal heat extraction capacity; and (3) the power load distribution over the beam footprint b, which can be written as a function of θB and the beam size a as
Clearly, this is in contrast with the concept of an ideal DCM, requiring some kind of compromise or compensation strategy regarding
beam parallelism and/or beam position at a given point of interest, as elaborated next.Figure 1(b) illustrates how a small angle ΔRx between crystals is related to a shift δyΔRx of the virtual source according to
where L is the distance between the DCM and the source. Then, considering that variations of the virtual source are often proportionally related to shifts of the beam at the sample through the beamline optics, a common requirement is having them small compared with the source size. With X-ray source sizes of about 5 µm and L commonly of the order of 30 m for modern beamlines, a typical budget of 10% pushes ΔRx to the range of 10 nrad only. This immediately reveals the stringent dynamical angular stability requirements for state-of-the-art DCMs, as well as suggests that within an experiment there would be virtually no margin for intentional variations of ΔRx, for thermal effects compensation, for instance.
Yet, recalling (2), it can be seen that displacements in the virtual source related to ΔRx may be at least partly compensated by energy-dependent beam offset corrections via gap adjustments. Again using partial derivatives, δyH can be written as
Moreover, although not practically useful for intentional offset compensation because of its correlation with energy, it should be noted that the offset is also sensitive to variations in θB according to
Indeed, in addition to compensations with , these equations can be directly used for motion error specifications [see also Geraldes et al. (2022a)]. It can be seen (refer also to Fig. 2 in Section 2.2) that at low energies (high angles) the offset is sensitive to changes in θB but quite insensitive to the gap g, and vice-versa. This is unfortunate, since the higher chances of corrections related to thermal effects occur precisely at low energies, where higher power densities occur due to smaller footprints [see (4)]. Thus, acceptable boundaries may need to be identified for each experiment individually. Ultimately, the target would be to have the effective virtual source variation δy = in an experiment within a fraction of the source size, i.e. typically about 0.5 µm, or the corresponding behavior at the sample position.
The angular boundaries for ΔRx around an ideal energy tuning, which might be already out or perfect parallelism due to d-spacing variations, can be derived as a fraction of the angular bandwidth ΔθDW of the Darwin width of the crystals. This can be used, for example, to evaluate acceptable losses in trying to keep the incoming and outgoing beam parallel despite thermal effects. It turns out that ΔθDW can be described to a good approximation by a `rearrangement' of (3),
where ΔE/E becomes the intrinsic energy resolution of the crystal, a constant dimensionless physical quantity that is typically in the range between 10−3 and 10−5. With typical orientations of silicon crystals, such as Si(111) and Si(311), which are used in the HD-DCM, ΔθDW varies from hundreds of µrad at large θB to sub-µrad at small θB (see also Section 5.2.1). As a side note, detuning of crystals also often finds practical use in harmonic rejection strategies.
2.2. Asymmetric crystals
One last factual aspect to be considered is the inevitable existence of deviations between the actual crystal surface and its diffraction planes. Although many times asymmetric cut crystals reaching several degrees are intentionally designed for different purposes, including beam compression or expansion [see, for example, Hastings (1977)], here ideally symmetric-cut crystals with only manufacturing miscut limitations are considered. This is depicted in Fig. 1(c), which is adapted from the discussion recently carried out by Sterbinsky & Heald (2021) on how miscuts affect the gap in DCMs, where it can be seen that, for parallel diffraction planes in the first and second crystals, the distance between their surfaces varies over the propagation of the beam as a function of the small miscut angles β and α, respectively.
Considering here the gap value to be taken perpendicularly to the diffraction plane of the crystals and aligned with the rotation axis, a derivation equivalent to that described by Sterbinsky & Heald (2021) shows that (2) becomes
from where it can be seen that, in addition to the gain , which is very close to unity for small α, the larger deviations from (2) occur for small θB. Then, updating (6) and (7) yields
One may notice that β does not appear explicitly in any of these equations. This is because no beam walk is assumed in the first crystal. In practice, with imperfect alignment of the crystal surface together with the incidence beam on the axis of rotation, parasitic beam walk will occur, resulting in some contribution from β to the gap. Nonetheless, in this geometry, this contribution will be negligible compared with that of α.
Figure 2 shows plots comparing these quantities for α = 0, i.e. the nominal case with no miscut, and two miscut cases, with α = ±0.1° to represent reasonable manufacturing limitations. The angular range from 0.01 rad ≃ 0.57° to π/2 rad = 90° is shown in logarithmic scale for clarity, and the angular limits of the HD-DCM are represented by the vertical dash-dotted lines at 3° and 60° for reference. The dimensionless quantities are shown on the left side, whereas the relative ratio with respect to the nominal case on the right. In the upper plots, it can be seen that the gap no longer tends to an asymptotic value as θB becomes smaller for higher energies. Similarly, the remaining plots demonstrate the increased sensitivity in the offset position to both the gap and θB, the latter in particular increasing by several orders of magnitude, when miscuts are considered.
This indicates that in reality the gap motion range may need to be larger by a few percent than nominally expected, that sensitivities at low angles may vary by more than one order of magnitude, that calibrations for fixed exit may require more than the simpler trigonometric relation of (2) [see also Sterbinsky & Heald (2021)], and that the required velocities and accelerations related to the fly scan are in practice different from nominal ones.
2.3. Fly-scans
To conclude, the previous equations can be further discussed in terms of their implications in fly-scan spectroscopy. Firstly, for θB, E from (1) can be substituted into (3), and the result taken with respect to time, such that
Then, for the gap, (9) can be derived for a constant offset with respect to time, yielding
which, in terms of the energy rate, using (11), becomes
Naturally, here θB in the trigonometric arguments are actually related to E via (1), but replacing it with the inverse trigonometric function does not add to clarity.
The functions η0, η1 and η2 are introduced in (11), (12) and (13) to simplify the notation of these equations by capturing the complex trigonometric dependences on θB. Indeed, from Fig. 3, the strong non-linear relationships in following the desired energy variation rates in fly-scans become evident. From η0, the exponential increase in the required Bragg velocity for higher angles (lower energies) is clearly visible, changing by nearly three orders of magnitude within the operational range of the HD-DCM. From η1 and η2, although with different slopes, similar exponential behavior occurs for the gap with the nominal geometry with α = 0. With miscuts, however, deviations occur in the low-angle (high-energy) range, as also seen in Fig. 2. In particular, because η2 — representing the variation of g with respect to E — is the product of η0 and η1, it actually asymptotically converges to a small constant value, either for positive or negative α. Hence, differences of three to four orders of magnitude occur for the gap velocity for a given energy variation rate within the operational range of the HD-DCM.
For a control-based instrument like the HD-DCM, these aspects place demanding specifications on metrology and acquisition hardware, since very high resolution and low noise are required for the lower angular (higher energy) range, whereas high rates are necessary at the opposite limit. For example, while the angular resolution in the it can be found that, for an energy scan of 1 keV s−1, the Si(111) crystal angular speed requirements would be around 0.1° s−1 at the high energy range and as much as 40° s−1 at the low energy limit. In the latter case, the counting rates would have to be higher than the current electronics capacity of 10 MHz. Similarly for the gap, with a resolution of 0.1 nm from the quadrature laser interferometers for the nanometre-level control performance, an equivalent energy rate scan speed with Si(111) crystals without a miscut would translate via (13) to about 0.8 µm s−1 and 20 mm s−1 at the high and low energy limits, respectively. In the latter case, counting rates would need to reach 200 MHz.
quadrature encoder is 50 nrad (or about 3 µ°) for high angular resolution and small control errors, from (11)Moreover, reaching such high speeds in a reasonable traveling range may be associated with significant acceleration levels, since the main rotation mass 2 and 6 kg [see also Geraldes et al. (2022b)]. Indeed, taking the derivatives of (11), (12) and (13) with respect to time, the acceleration levels might span over even higher ranges than those of velocity, stressing the demands on the actuators, amplifiers and converters, again with very high resolution and low noise for the lower angular range, but higher forces and torques at the opposite limit. Furthermore, even considering the control-based and disturbance-oriented design of the HD-DCM, the higher speeds and accelerations at higher angles inevitably increase mechanical and setpoint-related disturbances.
and the gap stage mass are close to, respectively, 14 kg mThe performance of the HD-DCM under these challenges and the ongoing research are discussed via results in Section 5, but, before that, the following section elaborates a corresponding analysis for an APU source, showing how it impacts energy fly-scans with the HD-DCM.
3. Undulator properties and positioning analyses
So far, Sirius beamlines have been designed with two types of sources; namely, bending magnets, from the 3.2 T permanent-magnet dipoles that are part of its five-bend achromat lattice (5BA) (Liu et al., 2021), and undulators. The first provides a continuous and broad-band spectrum (with at 19 keV), that can be used in a number of beamlines, for different characterization methods, and is transparent with respect to a DCM operation [aside from power density variations due to the footprint dependence on θB, as seen in (4)]. The latter, on the other hand, has an adjustable photon justifying the specification of a particular device for each beamline individually, depending on the scientific research area, and requiring energy tuning with the monochromator. Therefore, key undulator characteristics and positioning tolerances for energy fly-scans with the HD-DCM at Sirius are investigated next. The analyses will be restricted to the case of APUs, which is the first type to be integrated with the HD-DCMs at Sirius beamlines, but, naturally, equivalent analyses can be made for different types of undulators.
3.1. Undulator emission properties
Indeed, as illustrated in Fig. 4(a), the in an undulator spectrum is characterized by relatively narrow emission bands, which are harmonics of a fundamental energy and depend on physical constants and parameters both of the storage ring and of the particular device. Then, by online adjustment of the so-called undulator deflection parameter Ku — that is a function of its magnetic period λu and magnetic field Bu, which may have components in both axes transversal to the propagation of the electron beam — the energy spectrum and possibly the polarization of the photon beam can be selected accordingly by the user. The plot covers the first nine harmonics of the APU22 undulator that is currently installed at both EMA and MANACÁ beamlines, in an arbitrary configuration with the third harmonic tuned to 9 keV for a given Ku. It clearly illustrates the gradual reduction in over several orders of magnitude as the harmonic order increases towards higher energies, and also that the peak energy width and the depend on the angular acceptance window at the beamline.
In APUs, Ku can be varied from a maximum value to virtually zero by relatively shifting one of its magnetic cassettes with respect to the other, defining a variable generally referred to as the undulator phase pu, which has a typical stroke of the order of 10 mm. Without diving into the development of the equations and into the physical parameters, as given in detail by Onuki & Elleaume (2002), it can be shown that an peak for a given energy Eu can be nominally written in terms of pu as
where nu is the emission harmonic and ci, with i = 1,…5, are constants resulting from physical constants and parameters of the storage ring and of the device. Thus, the peak emission energy dependence on pu for such an APU has a generic shape as depicted for normalized phase values in Fig. 4(b), i.e. ranging from a minimum to a maximum value for each harmonic. In most cases, the harmonics overlap, meaning that the same energy may be reached in two or more harmonics/phases, only with different fluxes (see also Fig. 5 in Section 3.2). This is especially useful for energy scans realized around the overlapping regions, such that full scans can be executed without commanding the undulator over discontinuous trajectories, which would be actually impractical in fly-scans. In other cases, energy gaps may exist, meaning that peak emissions do not occur for a given energy range with the particular combination of parameters and would not be available to the user, being a design choice.
Consequently, for DCMs operating at undulator beamlines, θB must be tuned to pu for the desired output energy. This aspect is illustrated in Fig. 4(c). Firstly, a zoomed spectral plot of the emission around the third harmonic at 9 keV is shown to illustrate in more details in linear scale (and with the smaller acceptance in the right axis for readability) the broadening of the peak and its shift towards lower energies as the angular acceptance is increased for higher fluxes. Then, wave-propagation simulations using Synchrotron Radiation Workshop [SRW (Chubar & Elleaume, 1998)] show the beam profile for perfectly monochromatic energies at and around the 9 keV resonances without any slit acceptance restrictions, from where it can be seen that, if the DCM and the APU are detuned beyond given energy-dependent margins, either the downstream is rapidly reduced to one side or the ideal Gaussian beam profile is distorted to ring-like structures to the other.
Therefore, for energy fly-scans at undulator beamlines the traditional stand-still tuning between θB and pu becomes dependent on the motion capabilities of the undulator and synchronism as well. As done in the previous sections, (14) can also be derived as a function of time, resulting in
which can be used in (11) and (13) to find the required motion correlation between all the axes of interest for an energy scan with an APU and a DCM. This correlation between the variation of Eu and pu is also qualitatively shown in Fig. 4(b), from where both the sensitivity in the due to motion errors in the undulator, for instance, as well as the effectiveness in realizing scans can be observed.
For a desired energy variation rate, large speeds would be required from the undulator motion at the phase limits. In addition, as for the DCM, linear energy scans require non-linear undulator trajectories, via an exotic function as compared with industrial standards, which may not be directly available for undulator controllers. So, it becomes clear that ideal and `simple' user settings for constant energy scanning rates, in practice, must be translated according to highly non-linear functions, often tending either to zero, or infinity, as shown next for the APU22 at Sirius.
3.2. The APU22 source
The two operational units of the HD-DCM are at the EMA and the MANACÁ beamlines at Sirius, which currently operate in a commissioning configuration with an APU22 undulator source by Kyma: λu = 22 mm, length = 1.12 m and Ku,max = 1.44. By 2024, final configurations are expected, with two of the APU22 units in series for MANACÁ, and a new IVU18.5 (in-vacuum undulator) for EMA: λu = 18.5 mm, length = 2 m and Ku,max = 2.1.
Hence, in this early stage both beamlines are working with lower fluxes, not only because the storage ring current is still limited to 100 mA, while expected to reach 350 mA, but also because of the shorter undulator. Besides that, EMA is working with a more limited energy range, due to an existing energy gap between 3.7 and 5.7 keV, occurring between h1 and h3, i.e. the first and third harmonics (even harmonics do not provide adequate sources) of the APU22. Moreover, due to an initial perspective of short-time replacement (or upgrade) of the sources, only basic features have been originally specified for the APU22 Beckhoff's control system. This has eventually posed integration limitations with respect to the HD-DCM in the early operation phase, motivating the development of alternative solutions, as discussed in the next sections.
Following the generic shape anticipated in Fig. 4(b), Fig. 5(a) shows the harmonic emission peaks between h3 and h13 as a function of phase, thus covering the energy range that has been explored so far at EMA and MANACÁ. Indeed, the useful range of h1 by the HD-DCM, i.e. between 2.3 and 3.7 keV, has a sharp reduction in this range (not shown) and would drastically suffer from attenuation from air and windows in the setups currently available at both beamlines. Higher energies, in turn, have prohibitively low fluxes with the APU22 and anyhow face reflectivity limitations from downstream mirrors. In the plot, the harmonics are truncated after an arbitrary overlap of 1 keV due to gains.
Then, Fig. 5(b) illustrates the variation correlation between the energy and the phase, but as a function of the energy for the various harmonics, instead of as a function of the phase as in Fig. 4(b). This can be directly used as a guide to evaluate how desired rates in keV s−1 are translated to mm s−1 in the undulator. For example, energy rates of 1 keV s−1 can be achieved with reasonable phase speeds in the range of 1 mm s−1 depending on the desired energy range, whereas it would require `infinite' speed around energies close to pu = 0. Before fast orbit feedback (FOFB) and feedforward systems, that should be capable of compensating possibly higher orbit distortions caused by the fast motion of the undulators, become available at Sirius' storage ring, an upper limit velocity of 3 mm s−1 has been specified to the APU22. Yet, due to the current limitations, at EMA the APU22 has been operated in practice well below 1 mm s−1 in fly-scans, reaching only a few tens of eV s−1 in this initial stage. Hence, ultimate fly-scan spectroscopy velocities are expected to be achieved later, with the final sources and with the storage ring systems fully implemented.
Next, to define synchronism and motion performance requirements for the APU22, a tuning tolerance method using wave-propagation simulations like those in Fig. 4(c) has been developed. It consists of analyzing a finite number of energies in each harmonic, and, for each of them, mapping an energy tuning tolerance ΓE corresponding to a variation of 10% of the full width at half-maximum (FWHM) of the beam with respect to the ideal `Gaussian-like' profile. The ratio between the tolerances and the central (nominal) energies are shown in Fig. 5(c), in which it can be seen that there is a higher forgiveness, around 0.075%, in h3 (and even higher for h1, around 0.18%, not shown), but similar ratios for h5 to h13, around 0.045% to 0.05%. This is because h1 and h3 are related to lower energies in which the X-ray source size and divergence are larger due to diffraction limits (not shown), whereas beyond h5 these parameters reach asymptotic values. Occasional residual inaccuracies in the estimation of ΓE seem insufficient to hinder any conclusions.
Finally, the dependence from Fig. 5(b) can be combined with the tolerances from Fig. 5(c) to determine phase tolerances for the undulators. This is shown in Fig. 5(d), from where it can be seen that at the lower phase limit (<2 mm), related to higher velocities for given energy rates, the error tolerances exceed 10 µm, whereas only 4 to 7 µm are required as the most stringent values. At this point, it is worth noting that, using (3), the same energy tolerances from Fig. 5(c) can be taken to derive angular tolerances for the with respect to an ideal energy setpoint. It turns out that, within this energy range, these calculated tolerances remain above 30 and 60 µrad for Si(111) and Si(311), respectively, which are much larger than the typical motion errors observed in the HD-DCM, as shown in Section 5.3.1. Indeed, the obtained energy tolerances varied between 4 and 16 eV, whereas spectroscopy resolutions in this energy range generally aim at sub-eV targets, such that the performance should be truly at least one order of magnitude better than that.
Consequently, in practice the mapped tolerance budget does remain entirely for the undulator, but must include metrology accuracy, motion control errors, and delay effects, which might possibly lead to integration-related limitations beyond the performance of each instrument individually. This has been the case with the APU22 until very recently, as elaborated by Geraldes et al. (2021b, 2022c). And it highlights the need of a holistic approach in beamline instrumentation for increased performances and efficiency, as discussed in the following sections.
4. HD-DCM integration and operation schemes
Once the essential positioning characteristics for the HD-DCM and the APU22 have been explained — including higher-level motion parameters for fly-scan spectroscopy that are now for the first time enabled with the HD-DCM — this section summarizes how the HD-DCM can be effectively used at Sirius beamlines, including its integration schemes and operation modes. Sufficient details, also updated with respect to Geraldes et al. (2021b), are given to clarify the several operational possibilities within more complex integration demands, leading to different hardware and software complexity requirements.
The HD-DCM application within a flexible integration scheme with a generic APU source is illustrated in the simplified control implementation diagram of Fig. 6. Integration with the well known EPICS control system (https://epics-controls.org/), via the so-called process variables (PVs) for high-level user operation, is represented in the diagram by the abstracted multidimensional structures PVUND and PVDCM, for variables related to the undulator and the HD-DCM, respectively.
The APU is represented by a generic single control loop UND for the phase (see Fig. 6), typically consisting of a servo-motor and a linear encoder in a third-party controller, with sample rates in the order of 1 kHz. Evidently, APUs with different embodiment or different types of undulators can be correspondingly adapted. The HD-DCM, in turn, can be defined by four main control loops (see Fig. 6), namely: Bragg (BRG), gap (GAP), pitch (PTC) and roll (RLL), running on NI's CompactRIO (cRIO) hardware at 20 kHz sample rate [see also Geraldes et al. (2021b, 2022b)], as detailed next.
4.1. The HD-DCM control architecture
BRG is responsible for controlling the θB in the operational range from 3° to 60°, which is done via an in-vacuum direct-drive rotary stage with a rotary encoder, in a bandwidth of 20 Hz. Then, GAP, PTC and RLL, in turn, are part of the so-called crystal cage (CCG), responsible for controlling the position of the second crystal with respect to the first crystal with nanometre-level performance, which is done via voice-coil actuators and laser interferometers, in bandwidths between 150 and 250 Hz. GAP controls the parameter g from Section 2, whereas PTC and RLL, referring to the pitch and roll angles, control the inter-crystal parallelism. PTC is related to the crystals DW tuning and to vertical shifts in the virtual source via ΔRx in (5), with the strict target of 10 nrad RMS (1 Hz to 2.5 kHz). RLL is related to lateral shifts in the virtual source, but with dependence on θB and much lower sensitivity than PTC, leading, for instance, to a positioning stability target of 90 nrad RMS. The complete set of specifications of the HD-DCM has been given by Geraldes et al. (2022a).
Due to a careful mechatronic architecture, in practice the originally multiple-input–multiple-output (MIMO) system can be statically decoupled, so that the loops may be addressed independently and simpler single-input–single-output (SISO) controller design techniques can be used [see also Geraldes et al. (2022b)]. Thus, in Fig. 6 the individual plants and controllers are represented as Pi and Ci, respectively, where i refers to UND, BRG, GAP, PTC and RLL. Each loop follows a reference ri and provides a measurement yi, being subject to actuator and plant disturbances, as well as measurement noise from sensors (encoders and laser interferometers). Measured error signals are defined as ei = ri − yi.
The reference signals for the CCG loops are defined as functions fGAP, fPTC and fRLL of θB, which can follow nominal correlations, according to (2) or (9) for the gap, for instance, or result from beam-based calibrations [see Geraldes et al. (2021a)], that can be updated by the user via PVDCM. Regarding the control-based architecture and fly-scan demands, it is useful to have these relations as continuous smooth functions, rather than arbitrary discrete look-up tables that might be used in step-scan applications, such that discontinuities are prevented and high-frequency content is minimized in the reference signals. So far, eighth-order polynomials have been used as a standard.
Furthermore, because the measurement signal yBRG includes the disturbances in the BRG loop and electronic noise contamination that would deteriorate the individual performances of the CCG loops, it has been found that under normal conditions the control error eBRG is sufficiently small that the reference signal rBRG may be preferable as an indicator of θB. Although perhaps not a usual control strategy in leading–following systems, it benefits from favorable tolerances in the HD-DCM and its reliability. Indeed, PTC and RLL have only (at most) a small and smooth dependence on BRG, resulting essentially from fine adjustment for parallelism and fixed-exit calibrations [see Geraldes et al. (2021a)], with virtually no effect within small θB mismatching.
The GAP, in turn, as specified by Geraldes et al. (2022a), can afford as much as 300 nm of mismatch in the worst-case scenario before its influence on the beam offset exceeds 10% of the beam size [see also (6) and (10)]. Then, using (2) — or, equivalently, (9) for miscuts — the derivative of the gap with respect to θB can be written as dg/dθB = . Hence, with H = 18 mm and assuming an ideal gap motion, the corresponding BRG tracking mismatch, again for 10% of the beam sizes, should be bound to |eBRG| << 10 µrad in the 3° to 60° working range. As in the case of the energy tuning tolerances, this is one order of magnitude larger than the tracking errors that have been observed so far in fly-scans with the HD-DCM (see Section 5.3.1). This means that the loops can be connected only to a necessary extent, allowing for better individual performances, particularly of the PTC — which is the most sensitive degree of freedom, operating close to the specification boundaries.
A more elaborate alternative — that, however, has not proven practical until now — to prevent sensor noise, but still take the system response into account, consists of using an estimator χBRG of the closed-loop response via the complementary sensitivity function (T) of the BRG loop with χBRG = TBRGrBRG. Whichever the case, the desired reference selection for the the CCG loops may be done in the operation mode block OPNBRG via EPICS with PVDCM, which is also used to define the origin of the reference rBRG itself.
Indeed, it is then the definition of how rBRG is updated in the HD-DCM control loop that eventually defines how the instrument can be used at the beamline. In that sense, within the architecture in Fig. 6, three operation modes can be identified such that: (1) a trajectory is asynchronously calculated for rBRG internally in the cRIO just before motion, being independent of other instruments; (2) a real-time conversion for rBRG is made from a leading signal; and (3) a known trajectory for rBRG is previously stored in a file in the cRIO and executed after a trigger for synchronism. As these modes are related to different experimental capabilities and implementation complexity concerning other instruments at the beamline, they are separately discussed in the following subsection [see also Geraldes et al. (2021b) for a preliminary discussion].
4.2. Operation modes
The first operation mode is the stand-alone mode, in which the HD-DCM can be operated by the user in a simple way at high level via PVDCM. In this case, a trajectory generator block TRGBRG in cRIO uses user-defined parameters and default trajectory options to generate a suitable trajectory on demand and, then, triggers rBRG to be updated accordingly at the 20 kHz sample rate. This is a basic mode for fixed-energy or step-scan experiments, and the undulator phase must be independently adjusted by the user via PVUND for the appropriate energy tuning. Although not optimum in terms of timing, this can often be handled via standard step-scan beamline pieces of software. Moreover, this would also be a straightforward option for energy fly-scans at bending-magnet beamlines, in which the source spectrum is continuous.
Next, in the follower mode, rBRG is updated in real-time as a function of a leading input signal, which might be any external reference, but here, naturally, consists of the undulator phase. Thus, fBRG is defined as a function of the undulator phase and the desired harmonic (HAR), which is selected by the user as part of PVDCM. It can be nominally defined via (1) and (14), but, in practice, due to the relatively small tuning tolerances (see Section 3.2), requires beam-based calibration, as discussed by Geraldes et al. (2021a). Hence, by controlling the undulator alone via PVUND with reasonable motion setpoints, the user should be able to obtain a well tuned fixed-exit monochromatic beam over the whole energy range, allowing the fly-scan spectroscopy potential to be explored at undulator beamlines. Integrating the HD-DCM with different types of undulators would basically require only adapting fBRG to different parameters.
Still about the follower mode — in a similar fashion to the reference choice of the CCG loops — the phase signal can be also considered from three options that are managed via the selection block SELUND, according to user settings in EPICS as part of PVDCM. They are: the measurement signal yUND; the reference rUND from the undulator controller; or a position estimator χUND of the closed-loop response via the complementary sensitivity function with χUND = TUNDrUND (as mentioned for the BRG loop). In the first, which was the original option available for the APU22, the encoder signal yUND is directly derived to the cRIO — with the reading rates typically in the order of MHz for a quadrature signal, for example, being then internally downsampled to 20 kHz. However, as in the BRG case and the CCG loops, this signal includes disturbances from the UND loop and electronic noise that are propagated to rBRG. Thus, some filtering inside SELUND was found to be essential already at the early implementation phase. Nevertheless, as described by Geraldes et al. (2021a, 2022c), even with the additional filters the performance degradation to the the HD-DCM loops ended up being unacceptably large for the desired fly-scan spectroscopy quality (with pitch stability exceeding 30 nrad RMS and correspondingly large Bragg control errors, for instance), such that further alternatives needed to be developed, as explained below.
Indeed, the two remaining options for the follower mode are based on the reference signal as the undulator phase indicator. Here, again connecting the loops only to the necessary extent, the tuning tolerances in the range of a few micrometres between the HD-DCM and the APU [see Fig. 5(d)] can be explored to leave out disturbances and noise from the undulator system, which is typically only a micrometre-level-performance machine. Naturally, both cases require a minimum level of reliability/repeatability in the undulator, which must be proven by each machine, and is demonstrated for the APU22 in Section 5.3. With rUND, the input may come via a digital signal from the undulator controller, with the advantage of being a noiseless signal, but numerical representation limitations and lower control rates may still exist, requiring additional filtering or interpolation functionalities in SELUND to prevent significant setpoint discontinuities in rBRG. This is the case for the current APU22 solution, discussed in the following subsection. For the χUND option, there is the additional challenge of knowing enough about the control of the undulator, which is not always the case, particularly with third-party instruments.
The last operation mode is the triggered mode, in which trajectories for rBRG can be directly read from a file TRFBRG that is pre-generated by the user. At bending-magnet beamlines, this may open possibilities in terms of fly-scans with non-standard trajectories. At undulator beamlines, in turn, it separates the undulator and the HD-DCM from the control perspective, preventing the issues of numerical representation, sample rates or noise that were just described in the follower mode. In this case, the undulator controller is also required to be able to switch between modes, according to an equivalent operation block OPNUND, and to store and run the setpoints from a file TRFUND, which must match those of the HD-DCM for energy tuning. Then, trigger and synchronization must be handled at the hardware level or via sufficiently fast/deterministic software implementation. Here, as well, sufficient reliability is needed in the undulator. This mode is unfortunately not yet available with the APU22, since an additional board is needed in the Beckhoff's controller for external trigger capabilities. Yet, this feature is expected to be implemented and tested in the near future.
Finally, complementary performance estimators in Fig. 6 can be optionally computed in real-time to indicate compliance in the experiments. Firstly, the sensitive virtual source vertical shift χVSV, as functions fVSV of BRG, GAP and PTC, i.e. θB, g and ΔRx, can be calculated using (5) and (9), or their calibrated equivalent, and evaluated against the source size. Then, the mismatch χENG in the energy tuning between the source and the HD-DCM, as functions fENG of BRG, UND and HAR, i.e. θB, pu and nu, can be computed using (1) and (14), or their calibrated equivalent, and compared with ΓE [see Fig. 5(c)]. This should become common practice in the future to increase the maturity and the robustness of the experimental procedures.
4.3. Upgraded APU22 HD-DCM integration topology
Due to the inherent fly-scan capabilities of the HD-DCM, the known significant benefits related to fly-scan spectroscopy (elaborated in Section 5.3), and the inconvenient limitations found in the originally available follower operation that were mentioned in the previous subsection, integration alternatives started to be investigated. Then, while the absence of the Beckhoff's triggering board prevented fly-scan spectroscopy operation in the triggered mode, a solution complementing the follower mode operation has been developed in-house. Now, differently from the previous condition, the new architecture allows the HD-DCM to work close to ideal scanning conditions, as shown in the next section.
The upgraded topology is depicted in Fig. 7. As before, for standard communication with the external world, the Beckhoff's controller, running a Windows Embedded CE operating system (WinEmbCE OS), exchanges variables via a Python layer (Variables Layer) and an ethernet port (Eth0 driver) with an external virtual machine (VM-Debian), which runs an EPICS IOC server for exchanging the EPICS PVs with the network. The cRIO controller, running a linux real-time operating system (NI Linux RTOS), in turn, has an embedded EPICS IOC layer [Nheengatu (Alnajjar et al., 2019)], directly communicating with the network via an ethernet port (Eth0 driver).
Yet, now, the Beckhoff's and the cRIO controllers are also directly connected via a unidirectional ethernet access (Eth1 drivers), which can stream the APU22 position setpoint rUND to the HD-DCM (see Fig. 6). In the same motion control TwinCAT RT thread in the Beckhoff, motion variables (e.g. position and velocity) are calculated, sampled and streamed via UDP (user datagram protocol) through the dedicated ethernet link. At the cRIO side, LabVIEW RT code (as part of SELUND in Fig. 6) is responsible for decoding and treating the data.
Indeed, in spite of the peer-to-peer ethernet link and both implementations being on real-time threads, there are inherent delays, jitters and latencies in the system, which can be in principle characterized and partly compensated for in the LabVIEW RT code. One of these was the mechanical delay in the undulator response, thus allowing the HD-DCM to be closer to the real position rather than simply following an unrealistic reference setpoint. Moreover, given that the undulator control loop runs at 1 kHz only and the HD-DCM at 20 kHz, a LabVIEW FPGA code was implemented (also as part of the abstract SELUND block in Fig. 6) to generate an upsampled position reference to the HD-DCM control algorithm.
In this configuration, the Beckhoff's controller is the one responsible for defining a suitable higher-order trajectory profile for smooth performances. Hence, obtaining deeper access and suitably defining Beckhoff's internal variables for acceleration and jerk has also been an upgrade action, as compared with the early operation phase. Yet, as typical in high-performance mechatronic systems [see Butler et al. (2020)], a lot more can be expected to be investigated and optimized in terms of trajectories in the future, once other performance/operation bottlenecks at the beamline are eliminated. The following section demonstrates the latest progress and discusses the current operational limitations via concrete experimental examples.
5. Experimental results
Before becoming available to the general beamline user, the HD-DCM goes through a primary commissioning procedure that has been iteratively developed at the MANACÁ and the EMA beamlines at Sirius, according to the particularities of this new instrument and its integration system. As elaborated by Geraldes et al. (2021a), this includes: (1) energy calibration for the angular metrology of the Bragg rotary stage with respect to reference absorption edges; (2) DCM-undulator tuning between the and the undulator phase for the multiple harmonics (see also Section 3.2); and (3) fixed-exit calibration concerning the HD-DCM crystal cage (gap, pitch and roll). Then, part of these routines may be executed by the beamline crew or users via automated Python scripts on a regular basis, as sanity checks or re-calibration for improved accuracy. The first commissioning results of the HD-DCMs with X-rays at MANACÁ and EMA were described by Geraldes et al. (2021a, 2022c). In the following subsections, updated results for stand-still and scanning performances at EMA, as well as methods developed for data analysis, are disclosed.
5.1. EMA beamline setup
The experimental setup layout is depicted in Fig. 8. The APU22 source, with its Beckhoff's motion controller, is at the origin; electron beam position monitors (eBPMs) positioned 3 m upstream and downstream can provide transversal (x and y) and steering (pitch and yaw) data with sampling rates up to 572 kHz via dedicated electronics in the storage ring (SR); a set of high-heat-load slits (Slit1) is used to define the acceptance to the HD-DCM at 29 m; a set of bendable Kirkpatrick–Baez (KB) mirrors at 44 m is used to collimate or focus the downstream X-ray beam; two sample stages are present at 45.5 and 46.2 m; a set of pneumatic metal filters can be used at 52 m; a set of monochromatic slits is positioned at 53.3 m; finally, ionization chambers (IC1, IC2, IC3 and IC4) are used both before and after the samples for absorption experiments, as well as after the monochromatic slits (Slit2) for stability measurements.
The ICs were connected to Stanford Research Instruments' SR570 low-noise current preamplifiers, whose voltages were then digitized at cRIO units (cRIO1 and cRIO2) using NI's 16-bit NI-9215 analog input boards with 100 kHz sampling rate. Indeed, the cRIO has been chosen as one of the standard data acquisition hardware/controllers for Sirius beamlines, thus not only hosting the entire control of the HD-DCM but also being used to handle digital and analog signals of a variety of devices at rates up to 10 MHz. At the higher-level user operation, the so-called Nheengatu solution has been developed by the Beamline Software Group group (Alnajjar et al., 2019) to integrate cRIO variables in LabVIEW into the EPICS framework. Moreover, a special software module, known as the time and trigger unit (TATU) (Piton et al., 2021), has been developed in-house by the Beamline Software Group for the NI-9401 board to work as a synchronization unit in the microsecond range. This way, data from the eBPMs, the HD-DCM and the ICs can be already synchronized in the sub-millisecond domain using TTL (transistor–transitor logic) trigger signals. The APU22 is expected to become part of the complete triggering framework as soon as the complementary Beckhoff's triggering board is commissioned.
5.2. Stand-still performance
Stand-still performances of the HD-DCM with pitch levels around 10 nrad RMS up to 2.5 kHz according to its embedded metrology have been reported by Geraldes et al. (2022a) and Geraldes et al. (2022c) for the units at MANACÁ and EMA, respectively. With X-rays, however, only a preliminary stability evaluation attempt at MANACÁ has been reported so far by Geraldes et al. (2021a). Here, further analyses of the angular vibrations are carried out at EMA using both the rising edge of rocking curves and knife-edge methods, as described by Yamazaki et al. (2013), Chumakov et al. (2014) and Sergueev et al. (2016).
5.2.1. Rocking curves
Starting with the rocking curves, the experiments were performed at 25.5 keV, i.e. close to the edge of Ag. This has been about the highest energy used so far at the beamline mainly due to the weak emission of the APU22 [see Fig. 4(a)] at higher energies, with the 13th harmonic being required for 25.5 keV. The selection of the highest possible energy was an attempt to maximize the sensitivity provided by narrower rocking curves [see Sergueev et al. (2016)]. The high-heat-load slits were set to 430 µm × 600 µm (v × h), providing an acceptance close to 15 µrad × 20 µrad, which offers a reasonable compromise between spectral purity for different diffraction orders and (see Fig. 4). Both Si(333) and Si(555) reflections were used, hence with Bragg angles around 13.45°and 22.81°. The measurements were taken with the IC4, having the monochromatic slits completely open, and the sample stages and the IC2 and IC3 removed from the beam. The IC1 was left in place to work as a filter for the lower-order reflections, being sufficient for the Si(333) measurements. For the Si(555) measurements, in turn, Al filters from the filter box were added to reduce the tails in the rocking curve, as shown next. The IC1 was filled with Ar95%/Kr5% and IC4 with N240%/Ar60% for efficiency at 25.5 keV. Finally, the amplifier gains for the IC4 were set to 1 nA V−1 and 100 pA V−1 for the Si(333) and the Si(555) measurements, respectively.
The step-scan rocking curve measurements, with integration time of 0.1 s per point and total measurement times around 1 min, are shown in Fig. 9(a). In spite of the residual tails from lower-order reflections, particularly in the Si(555) case, the agreement with the nominal simulated values (Sanchez del Rio et al., 2011) is remarkable, with FWHMs of about 3 µrad and 1.12 µrad. Furthermore, even though the total power in the HD-DCM in these conditions is only of the order of 1 W, i.e. two orders of magnitude lower than the expected final load when the storage ring is at full current and the final undulator is installed, these results prove the high quality of the crystals [manufactured at the Advanced Photon Source (APS) by Elina Kazman's team] and of the clamping and cryogenic cooling solutions [see Geraldes et al. (2022a)]. Next to that, as a side evaluation of fly-scan options in pitch, repeatability and long-term stability, Fig. 9(b) shows a series of thirteen 10 s-long fly-scan measurements for the Si(333), taken every 12 minutes in a total of about 2.5 h. The results find good agreement with the initial step-scan, prove the internal metrology repeatability, and demonstrate stability over a few hours, apparently with only a small drift of about 0.1 µrad h−1.
Then, with the rocking curves, conversions between intensity and inter-crystal parallelism can be made for stability analyses. Fig. 9(c) shows the intensity sensitivity obtained for both step-scan measurements by taking the derivative of fitted peak functions (not shown) of the rocking curves. It is worth noting that, perhaps counter-intuitively, although the Si(555) has a steeper rising edge, the larger available with Si(333) made its sensitivity larger, even with a tenfold smaller amplifier gain. Besides that, the storage ring is not yet operating in top-up mode and its current during these measurements varied from 92 to 58 mA, respectively, thus reducing the potential sensitivity with Si(555) by 42% (with respect to the 100 mA injection level).
The actual stability experiments consisted of running staircase-like rocking curves, as proposed by Yamazaki et al. (2013). Hence, while making continuous acquisition at sampling rates of 1 kHz for the eBPMs and the IC4, and 20 kHz for the HD-DCM, the second crystal of the HD-DCM would start from a detuned condition and move over the rocking curves in a given number of steps. At every step, a dwell time of 5 s was taken for further statistical analyses. Figs. 9(d) and 9(e) show the time data for these scans. Together with the IC4 signal, estimated intensity levels for the HD-DCM are calculated using the high-quality internal metrology — which was consistently around 11 nrad RMS at all times — and the intensity sensitivity curves.
Next, the RMS values obtained in each step are summarized in Fig. 9(f). While the HD-DCM estimates replicate the sensitivity shapes, the IC4 signals show different results. For the Si(333) measurement, it leaves a base noise floor and starts to build up close to what might be interpreted as the HD-DCM contribution, but, instead of decreasing with the HD-DCM after the sensitivity peak, it continues to rise, in a closer approximation with the absolute intensity shape. For the Si(555) measurement, in turn, the RMS value simply remains at a constant noise floor, regardless of the absolute intensity value.
To further investigate these results, Figs. 9(g) and 9(h) provide the power spectral density (PSD) for each step in color gradients: the IC4, in the front and between 1 and 500 Hz, goes from light to dark (`viridis' colormap) as the pitch angle increases, whereas the HD-DCM, at the back and between 1 Hz and 10 kHz, goes from dark to light (`plasma' colormap). In neither case would it be expected that the HD-DCM estimate would overtake the IC4 spectra because, by definition, the HD-DCM is only one of the contributions that might affect the intensity measurements. Moreover, the most prominent peaks in the HD-DCM estimate do not appear as the most relevant ones in the IC4 spectra.
The fact is that even though the acquisition rate for the IC4 was 1 kHz, the large gains (≤1 nA V−1) required in these measurements were associated with low bandwidths of only 15 Hz for the Si(333) and 10 Hz for the Si(555) with the SR570 amplifier (as specified by the supplier). As a matter of fact, bandwidth limitations are often the case for most current amplifiers at large gains, as a way to handle spurious noise. Yet, with the Si(333) an increase in the IC4 spectra can be seen, with the intensity being dominated by the low-frequency contribution and by peaks at 60, 180 and 300 Hz. The Si(555), in turn, barely changes with the absolute intensity, being again dominated by peaks at 60, 180 and 300 Hz. This suggests that the first is sensitive to contributions from the source, whereas the latter is already limited by the noise in the detecting elements.
This can be further confirmed by including the noise levels (floors) that had been previously measured for the ICs in the setup without the beam, and by qualitatively comparing the IC4 spectra with those of eBPM signals, as shown in Figs. 9(i) and 9(j). Here, the multiple steps are represented by the same color that is associated with the different elements, namely: IC4 signal, electron beam horizontal position, electron beam vertical, electron beam yaw, electron beam pitch, and horizontal position of a dispersive eBPM (qualitatively related to the oscillation of the energy of the electrons in the storage ring, not shown in Fig. 8). In Fig. 9(i), the IC4 PSD is mostly above the noise floor, whereas the low-frequency behavior, the peaks at 6 and 12 Hz, and the `bump' between 30 and 100 Hz do correlate with features in the eBPM signals, such that it becomes clear that the beam instabilities in the storage ring, with RMS values up to 2 µm and 1.5 µrad, are still truly affecting the intensity stability at the beamline. In Fig. 9(j), only the very lowest range of the Si(555) signal is above the noise floor, endorsing the discussion above.
Consequently, being close to 10 nrad RMS according to its internal metrology, the pitch stability of the HD-DCM seems indeed too small to be measurable via et al. (2016) and more than 50 times lower than measured by Yamazaki et al. (2013). Still, the data analysis methodology proposed here allows for a systematic evaluation of the problem, demonstrating that the current practical measurement limitations span from instabilities in the source itself to the detecting elements, due not only to low signals (related to low fluxes with the APU22) but also to spurious electronic noise. Then, it can be used as a validation tool in following the required improvements in experimental conditions over time.
over rocking-curve edges under the current conditions of the EMA beamline and Sirius storage ring. It should be emphasized, however, that the stability levels investigated with the HD-DCM are about 20 times lower than those measured by Sergueev5.2.2. Knife edge
Next, taking advantage of the experimental setup with the Si(333) configuration, an equivalent stability analysis was carried out with a knife-edge method. Firstly, the beam was focused with the KB set to about 7 µm × 10 µm (v × h) at the monochromatic slits. Then, with the crystals tuned at the peak of the rocking curve, the bottom blade of the slit was scanned in front of the beam and finally centered halfway for maximum sensitivity. Next, the second crystal of the HD-DCM was step-scanned in pitch to obtain the intensity sensitivity curve shown in Fig. 10(a). Lastly, the second crystal was repositioned at the tuned condition (maximum sensitivity) and continuous data were collected for 10 s.
Fig. 10(b) shows the time data, with the IC4 signal, the estimated intensity level for the HD-DCM (derived from its internal metrology signal and the sensitivity curve) and also the IC1, which was used as a reference signal with a gain of 5 nA V−1. At the top of the rocking curve, the HD-DCM should have nearly zero impact in the IC1, which is before the knife and also takes the Si(111) background Besides that, equivalently to the staircase measurements, under ideal experimental conditions the distribution of the IC4 could not possibly be smaller than that of the HD-DCM.
Turning once again to the frequency domain, Figs. 10(c) and 10(d) show the PSDs comparing all signals and including the noise floors from dark conditions. In Fig. 10(c), it can be seen that the ICs are well above the noise level for most of the frequencies. Yet, they surprisingly show very similar levels and behavior between them, from which any contributions from the HD-DCM can be hardly recognized. Only the largest peaks from the HD-DCM at 120, 204 and 255 Hz seem to find counterparts in the IC4, but with 1000-fold reduced amplitudes. Then, from Fig. 10(d), both ICs take the structures from the eBPMs spectra, as also discussed for Fig. 9(i), which again limited the investigation of the inter-crystal pitch stability of the HD-DCM with X-rays. Still, the sensitivity peak here reached 2 V µrad−1 (not shown), which is 2.5 times larger than what was achieved with the Si(333) rocking edge [see Fig. 9(c)], making this the most sensitive and promising method available at EMA at the moment.
Unfortunately, the beam time was over before a knife-edge measurement with Si(111) could be performed. Indeed, since this method does not rely on the sharpness of the rocking curve, working with the first diffraction order would allow fluxes one or two orders of magnitude larger (with a broader Darwin width), which could use lower gains (≫1 nA V−1) and reach bandwidths of hundreds of Hz to a few kHz, while improving the signal-to-noise ratio. In that sense, lower energies at lower harmonics should also be explored, as more than two extra orders of magnitude might be gained in [see Fig. 4(a)]. Nevertheless, as discussed by Sergueev et al. (2016), one of the drawbacks of the knife-edge concept is its sensitivity to the mechanical stability of other elements in the beamline, including the mirrors and the knife itself. For example, in Figs. 10(c) and 10(d) the sharp peak at 150 Hz in IC4 is likely to be related to an unmapped mechanical resonance in the optical chain. These contributions may become more relevant once the other dominant effects are eliminated.
To conclude, although the replacement date of the official undulator source — which will improve the conditions for pitch stability measurements based on rocking curve edges — remains unclear, significant improvements in overall stability are expected already in the short term. These include: (1) the implementation of the FOFB system, which will extend the electron beam control bandwidth well beyond the 1 Hz that is currently provided by the slow orbit feedback (SOFB) system (Marques et al., 2022); (2) the top-up operation mode at the storage ring, which should minimize current-related transients; and (3) a refined grounding work to reduce electronic noise at EMA. Other than that, this commissioning round fostered the development of this systematic data analysis pipeline and a in the real-time integration between several components of the storage ring and the beamline, which proved to be decisive in interpreting the experimental data and will be essential for fast experiments with sufficient statistics. A new round of beam time for stand-still stability assessment of the HD-DCM will be realized after the identified dominant sources of beam variation and noise are solved.
5.3. Scanning performances
In spite of the electron beam instabilities and electronic noise levels that limited the evaluation of the HD-DCM performance in the stand-still experiments with X-rays, EMA has been perfectly able to host users from the scientific community and carry out relevant spectroscopy experiments on a regular basis. The following subsections provide results showing the remarkable motion performance and spectroscopy capabilities currently available at EMA.
5.3.1. Fly-scan motion evaluation
The current possibilities in energy fly-scans, with the HD-DCM in the follower mode and the upgraded integration topology with the APU22 (see Section 4.3), are exemplified in Fig. 11. Performance variables in 1 keV scans taking about 15 s with Si(111) are shown for a selection of energies for elements of interest covering the currently useful range of energies at EMA. The undulator reference signal rUND entering the HD-DCM control loops is shown to cover 1 to 2 mm over different harmonics for the different energies. The undulator velocity vUND, obtained by differentiating rUND, is bound generally below 1 mm s−1, except for larger random peaks found in most of the scans. Their cause seems to be related to discontinuities in the trajectory generated by the Beckhoff's controller, but this issue is still under investigation for a definitive solution. Yet, although they partly degrade the HD-DCM loops, as discussed below, this is to a far less extent than the degradation observed with the previous architecture [see Geraldes et al. (2021b, 2022c)]. Regarding the undulator tracking error eUND, it shows good performance, with deviations of ≤3 µm, thus obeying the phase error tolerances specified in Fig. 5(d).
The amplitude of the rBRG decreases with energy, going from a stroke of 0.05 rad (2.85°) to 0.002 rad (0.1°) only. The angular velocity vBRG is generally bound to ≤30 mrad s−1 (1.7° s−1), but the reaction to the peaks in the undulator velocity is clear. The control error eBRG is also significantly reduced as the photon energy increases, typically going from about 1 µrad (in the cruising range outside the acceleration/deceleration regions) to less than 0.2 µrad at high energies. Naturally, the influence of the undulator velocity bursts is again obvious. These angular errors can be converted to energy variations according to (3), here falling <0.1 eV. Their particular impact on the experiment depend on the contributions within the integration times. More details about this will be provided in another publication under preparation.
reference signalRegarding crystal cage loops, the amplitude of the gap reference signal rGAP also decreases with energy, going from a stroke of 0.15 mm to 0.001 mm only. The gap velocity vGAP is generally bound to ≤0.1 mm s−1, mostly to dynamically correct the gap error (∣eGAP∣ ≤ 10 nm, not shown), since the gap stroke for these energies are low, except during the reactions to undulator peaks. Finally, the control error for the pitch between crystal ePTC is always within 15 nrad RMS, even at the lower energies and with the disturbances of the undulator signal. In these experiments, the destination position, velocity and motion trigger were sent to the Beckhoff's controller via PVs, and the trajectories had constant speed in the undulator coordinates. For trajectories linear in energy, pre-calculated vectors must be passed to the APU22 controller, a functionality not yet validated, but that is expected to become available soon.
Still, to remove any doubts about the undulator signal being responsible for the residual performance degradation in Fig. 11, it also shows the Bragg and crystal cage variables for similar trajectories with the HD-DCM in the triggered mode, i.e. following a pre-planned trajectory and completely decoupled from the undulator. Small differences in the reference signals are because the follower mode used polynomials (fBRG in Fig. 6) calibrated at the beamline, whereas the triggered mode tests used nominal polynomials for this undulator. It can be seen that eBRG and ePTC are close to ideal performance, at ≤0.35 µrad and ≤15 nrad RMS (at 20 kHz), respectively.
The undulator could not make part of this last proof of concept because, as mentioned above, a functional board is still missing in the Beckhoff's controller to allow for external triggers (and integration with the TATU framework). As soon as it is available, it will be possible to compare the results from the current streaming implementation with fully triggered performance for a definitive beamline operation choice. It should be noted that this high-performance synchronization effort is in agreement with what has been developed in other synchrotrons (Hidas et al., 2022), but here with additional subtleties related to the large sensitivity of the HD-DCM as a high-performance mechatronic system.
To conclude, the choice of scans within 15 s was arbitrary and tests at least twice as fast have achieved comparable results (not shown). Naturally, at some point, higher accelerations and faster motions on bearings, together with larger strokes required for lower energies, start to increase disturbances. These will require case-by-case analyses, particularly as the lowest energy range of the HD-DCM starts to be explored after the replacement of the APU22.
5.3.2. Spectroscopy results
The d transition metals, whose energy levels involved at the L2,3 edges are above 10 keV, therefore compatible with diamond anvil cells (DACs) (Itié et al., 2016). This kind of study is among the central scientific goals of the EMA beamline at Sirius.
technique is a powerful tool for studying the electronic behaviors of materials under extreme thermodynamical conditions, such as high pressure and low temperatures. In particular, high-pressure studies are especially interesting for the 5Here, to evaluate the spectroscopy capabilities with the HD-DCM during the EMA scientific commissioning, a high-pressure L2 and L3 edges (2p → 5d transition) to probe the pressure evolution of spin–orbit coupling of Pt metal. The measurements, at room temperature, were performed in transmission geometry, as illustrated in Fig. 12, using a membrane-driven copper–beryllium DAC prepared with two full anvils, with a culet size of 350 µm to reach pressures as high as 60 GPa. A 5 µm-thick platinum foil was loaded in a 60 µm hole of a rhenium gasket, together with a 5 µm ruby sphere for in situ pressure calibration, and neon gas was used as a pressure-transmitting medium. Referring to Fig. 8, the DAC was mounted at the Sample1 position, where the beam was focused to 1 µm × 2 µm (v × h), whereas a Pt standard foil (Exafs Materials) was placed at Sample2, while IC1, IC2 and IC3 were used for intensity measurements.
experiment was carried out with the Si(111) set of crystals at the PtThe experiment consisted of incrementally increasing the sample pressure in steps of roughly 5 GPa, and, in each pressure step, measuring three spectra in step-scan mode followed by ten spectra in fly-scan mode, for both the L2 and L3 edges. The total energy range was 490 eV in all cases. For the step mode, energy step sizes of 0.5 eV were used around the edge (30 eV before and 30 eV after it) and 7 eV for the pre- and post-edge regions. The step-scan integration time was 0.5 s, and the total time for each spectra was about 6 min, with most of it consumed by overheads from start/stop actions, as also mentioned by Hidas et al. (2022). The fly-scans, in turn, were performed with 0.1 mm s−1 phase speed in the APU22 and 20 kHz sampling frequency for the HD-DCM and the ICs cRIOs (see Fig. 8), with a measurement time of 27 s per fly-scan spectrum.
One representative set of data, for the L2 edge at 42 GPa, is depicted in Fig. 13. The colorful lines in the back (`viridis' colormap) are the fly-scan measurements, for which the effective sampling rate was averaged down to 40 Hz to improve the signal-to-noise ratio while achieving an energy resolution of 0.5 eV; the black dashed line is their average; the diamond markers show the points of three step-scan measurements, which overlap in most cases. The accurate correspondence between the two measurement methods demonstrates the reliability of the HD-DCM, the effectiveness of the upgraded follower mode operation, and the potential of the fly-scan method, with maximum practical speeds currently limited mostly by the APU22 and spurious electronic noise in the detection electronics (which is currently under improvement). Furthermore, it was found that the dispersion in the intensity signal of the fly-scan measurements is, in this case, dominated by a 12 Hz contribution (not shown in detail for conciseness), which was also observed in the stability measurements in Figs. 9 and 10, being likely caused by the instabilities in the electron beam. Again, these should be greatly improved with the implementation of the FOFB system.
It should be noted that the high-pressure compares a step-scan spectrum of the standard Pt foil in Sample2 (collected simultaneously with one of the step-scan measurements in the main plot) with a standard Pt L2 spectrum that had been measured at the APS 4-ID-D beamline in 2012, where the same features can be identified.
experiments using diamond anvil cells are particularly challenging because of the presence of diamond Bragg peaks in the spectra, such as those around 13.34 keV and above 13.5 keV. Although these Bragg peaks, inherent to the high-pressure experiments, do bring special complications to the data analysis, for our purposes they reinforce the accuracy of the fly-scan as being identical to the step-scan along the entire energy range. Moreover, in addition to the drastic time efficiency increase, it becomes clear that the fly-scans have benefits related to continuous measurements, such as the wiggle feature close to 13.52 keV, which might be misinterpreted in a step-scan with a poorer resolution. As a further validation check, the inset in Fig. 13Finally, evaluating the robustness and repeatability of the experimental system, Fig. 14 shows five sets of measurements for the L2 edge from 10 to 53 GPa, with the step-scans now represented by solid black dots and vertical offset between the sets for readability. The changes in the signature of the spectra as a function of pressure can be promptly identified, but the same aspects emphasized for Fig. 13 hold true. The whole measurement time was approximately 36 h, including injection windows, since the storage ring is still not operating in top-up mode. Without any further compensations, the variation of the peak of the L2 derivatives for all spectra (not shown), i.e. for more than 300 scans and including the current noise and electron beam stability limitations, was within ±0.75 eV. Also, considering that for each pressure step the HD-DCM would travel to the L3 edge (around 11.5 keV) and back, this proves the high degree of reliability achieved at EMA.
6. Conclusions
The efforts in designing and implementing the High-Dynamic Double-Crystal Monochromator from scratch, as a high-end mechatronic machine for Sirius new-generation beamlines, have been honored with unmatched fixed-exit and fly-scan spectroscopy capabilities. Indeed, the well known inter-crystal pitch stability performance bottleneck in standard DCMs is in the HD-DCM solved thanks to an improvement level varying from about 5- to more than 100-fold, since 10 nrad RMS (1 Hz to 2.5 kHz) can be achieved even during fast fly-scans, which, in addition, can also reduce measurement times by at least one or two orders of magnitude.
This article has summarized, from a new perspective, engineering and operational aspects that acquire critical relevance in enabling next-level experimental possibilities with the HD-DCM. The multi-axis and highly non-linear motion control problem is stated and an energy-tuning evaluation method, based on wave-propagation simulations, is also proposed to define positioning tolerances between the HD-DCM and the APU source. Then, limitations in the originally available integration architecture with the APU — which could have been actually predicted and prevented if a dedicated analysis had been carried out during the procurement phase of the commissioning undulators — have been overcome with a new control topology, that has been already able to provide close-to-ideal motion performance results. Furthermore, it was demonstrated how sensitive fast-acquisition experiments are to the whole beamline system, from the stability of the electron beam in the storage ring to high-bandwidth noise levels on detectors. Hence, the essential role of a holistic approach in integrating the HD-DCM to the beamlines, so as to achieve ultimate performances, higher efficiency and maximum throughput in fourth-generation light-source beamlines, cannot be emphasized enough. Still, scientific commissioning results demonstrating operational reliability and already tenfold faster fly-scan
measurements are discussed.Technical commissioning of the HD-DCM will proceed in parallel with beamline operation, as the parameters and functional systems of the storage ring gradually reach final values, and the definitive undulators are installed at the beamlines. These will provide higher electron beam stability and higher fluxes, eventually enabling practical experiments with faster acquisition rates than today, but also affecting the HD-DCM with higher power loads. Moreover, as a next step, spectroscopy performance in triggered mode will be compared with the current solution in follower mode also with X-rays. Finally, a new HD-DCM model for even faster scans, the so-called HD-DCM-Lite, is now under assembly phase for two new beamlines, and first results are expected in early 2023.
Acknowledgements
The authors would like to acknowledge the fundamental contribution of the LNLS staff, including scientists, leaders, engineers and designers of different groups, and of the team at MI-Partners, in the development of this collaborative project. Special thanks to Marlon Saveri Silva, Ricardo Caliari and Marcelo Moraes, who worked directly as the HD-DCM application engineers, and to Marcos Eleotério, José Claudio Corsaletti and Mauricio Donatti, who made the experimental setup at EMA possible. This work is also part of a PhD project carried out with the CST group at TU/e.
Funding information
This work is supported by the Brazilian Center for Research in Energy and Materials (CNPEM) under contract with the Brazilian Ministry of Science and Technology and Innovation (MCTI).
References
Alnajjar, D. A. D., Fedel, G. S. & Piton, J. R. (2019). Proceedings of the 17th International Conference on Accelerator and Large Experimental Control Systems (ICALEPCS2019), New York, NY, USA, pp. 997–1000. WEMPL002. Google Scholar
Ashcroft, N. W. (2003). Solid State Physics. Thomson Press. Google Scholar
Baker, R., Baboulin, D., Barrett, R., Bernard, P., Berruyer, G., Bonnefoy, J., Brendike, M., Brumund, P. M., Dabin, Y., Ducotté, L., Gonzalez, H., Malandrino, G., Marion, P., Mathon, O., Roth, T. & Tucoulou, R. (2018). Proceedings of the 10th International Conference on Mechanical Engineering Design of Synchrotron Radiation Equipment and Instrumentation (MEDSI 2018), 25–29 June 2018, Paris, France, pp. 440–444. FROAMA07. Google Scholar
Butler, H., Heertjes, M. F., Dirkx, N., van der Meulen, S., Ahlawat, R., O'Brien, K., Simonelli, J., Teng, K. & Zhao, Y. (2020). 2020 American Control Conference, 1–3 July 2020, Denver, CO, USA, pp. 3686–3703. IEEE. Google Scholar
Caliari, R. M., Geraldes, R. R., Moraes, M. A. L., Moreno, G. B. Z. L., Faassen, R., Ruijl, T. A. M. & Schneider, R. M. (2018). Proceedings of the 16th International Conference on Accelerator and Large Experimental Control Systems (ICALEPCS'17), 12–16 October 2017, Barcelona, Spain, pp. 997–1002. TUSH203. Google Scholar
Caliari, R. M., Geraldes, R. R., de Moraes, M. A. L. & Witvoet, G. (2020). ASPE 2020 Spring – Design and Control of Precision Mechatronic Systems, 6–8 May 2020, pp. 125–130. American Society for Precision Engineering. Google Scholar
Chubar, O. & Elleaume, P. (1998). Proceedings of the Sixth European Particle Accelerator Conference (EPAC98), 22–26 June 1998, Stockholm, Sweden, pp. 1177–1179. Google Scholar
Chumakov, A. I., Sergeev, I., Celse, J.-P., Rüffer, R., Lesourd, M., Zhang, L. & Sánchez del Río, M. (2014). J. Synchrotron Rad. 21, 315–324. Web of Science CrossRef CAS IUCr Journals Google Scholar
Cowan, P. L., Hastings, J. B., Jach, T. & Kirkland, J. P. (1983). Nucl. Instrum. Methods Phys. Res. 208, 349–353. CrossRef Google Scholar
Dolbnya, I. P., Sawhney, K. J. S., Scott, S. M., Dent, A. J., Cibin, G., Preece, G. M., Pedersen, U. K., Kelly, J. & Murray, P. (2019). J. Synchrotron Rad. 26, 253–262. Web of Science CrossRef CAS IUCr Journals Google Scholar
Eriksson, M., van der Veen, J. F. & Quitmann, C. (2014). J. Synchrotron Rad. 21, 837–842. Web of Science CrossRef CAS IUCr Journals Google Scholar
Fan, Y., Qin, H., Zhu, W., Jia, W., Liu, Y., Wang, J. & Li, Z. (2020). Nucl. Instrum. Methods Phys. Res. A, 983, 164636. CrossRef Google Scholar
Geraldes, R. R., Caliari, R. M., Moreno, G. B. Z. L., Ronde, M. J. C., Ruijl, T. A. M. & Schneider, R. M. (2017b). Proceedings of the 9th Mechanical Engineering Design of Synchrotron Radiation Equipment and Instrumentation (MEDSI 2016), 11–16 September 2016, Barcelona, Spain, pp. 44–47. MOPE19. Google Scholar
Geraldes, R. R., Caliari, R. M., Moreno, G. B. Z. L., Sanfelici, L., Saveri Silva, M., Souza Neto, N. M., Tolentino, H. C. N., Westfahl Jr, H., Ruijl, T. A. M. & Schneider, R. M. (2017a). Proceedings of the 9th Mechanical Engineering Design of Synchrotron Radiation Equipment and Instrumentation (MEDSI 2016), 11–16 September 2016, Barcelona, Spain, pp. 141–146. TUCA05. Google Scholar
Geraldes, R. R., Caliari, R. M., Moreno, G. B. Z. L., Sanfelici, L., Saveri Silva, M., Tolentino, H. C. N., Westfahl Jr, H., Ruijl, T. A. M. & Schneider, R. M. (2018). Proceedings of the 10th International Conference on Mechanical Engineering Design of Synchrotron Radiation Equipment and Instrumentation (MEDSI 2018), 25–29 June 2018, Paris, France, pp. 147–152. WEOAMA01. Google Scholar
Geraldes, R. R., de Brito Neto, J. L., Caliari, R. M., Eleoterio, M. A. S., Luiz, S. A. L., Moraes, M. A. L., Perna, A. V., Silva, M. S. & de Albuquerque, G. S. (2021a). Proceedings of the 11th International Conference on Mechanical Engineering Design of Synchrotron Radiation Equipment and Instrumentation (MEDSI 2020), 26–29 July 2021, Chicago, IL, USA, pp. 25–28. MOPB03. Google Scholar
Geraldes, R. R., de Brito Neto, J. L., Coelho, E. P., Do Carmo, L. P., Horita, A. Y., Luiz, S. A. L. & Moraes, M. A. L. (2021b). Proceedings of the 18th International Conference on Accelerators and Large Experimental Physics Control Systems (ICALEPCS2021), 14–22 October 2021, Shanghai, China, pp. 370–375. TUPV004. Google Scholar
Geraldes, R. R., de Brito Neto, J. L., Saveri Silva, M., de Albuquerque, G. S., Luiz, S. A. L., Pinto, A. C., Rigamonti Jr, H. & Eleotério, M. A. S. (2022c). Presented at 14th International Conference on Synchrotron Radiation Instrumentation (SRI 2021), 28 March–1 April 2021, Hamburg, Germany. Google Scholar
Geraldes, R. R., Moraes, M. A. L., Caliari, R. M. & Witvoet, G. (2020). ASPE 2020 Spring – Design and Control of Precision Mechatronic Systems, 6–8 May 2020, pp. 119–124. American Society for Precision Engineering. Google Scholar
Geraldes, R. R., Moraes, M. A. L., Witvoet, G. & Vermeulen, J. P. M. B. (2022b). Precis. Eng. 77, 90–103. CrossRef Google Scholar
Geraldes, R. R., Witvoet, G. & Vermeulen, J. P. M. B. (2022a). Precis. Eng. 77, 110–126. CrossRef Google Scholar
Golovchenko, J. A., Levesque, R. A. & Cowan, P. L. (1981). Rev. Sci. Instrum. 52, 509–516. CrossRef CAS Web of Science Google Scholar
Hastings, J. B. (1977). J. Appl. Phys. 48, 1576–1584. CrossRef Web of Science Google Scholar
Hettel, R. (2014). J. Synchrotron Rad. 21, 843–855. Web of Science CrossRef IUCr Journals Google Scholar
Hidas, D., Kiss, A. M., Rakitin, M., Sinsheimer, J., Tanabe, T. & Musardo, M. (2022). Nucl. Instrum. Methods Phys. Res. A, 1031, 166505. CrossRef Google Scholar
Hussain, Z., Umbach, E., Shirley, D. A., Stöhr, J. & Feldhaus, J. (1982). Nucl. Instrum. Methods Phys. Res. 195, 115–131. CrossRef CAS Google Scholar
Itié, J. P., Baudelet, F. & Rueff, J.-P. (2016). High Pressure XAS, XMCD and IXS, ch. 14, edited by J. A. Van Bokhoven & C. Lamberti, pp. 385–705. Google Scholar
Kelly, J., Lee, T., Alcock, S. & Patel, H. (2013). J. Phys. Conf. Ser. 425, 052009. CrossRef Google Scholar
Koningsberger, D. & Prins, R. (1988). X-ray Absorption: Principles, Applications, Techniques of EXAFS, SEXAFS and XANES. Wiley-Interscience. Google Scholar
Kristiansen, P., Horbach, J., Döhrmann, R. & Heuer, J. (2015). J. Synchrotron Rad. 22, 879–885. Web of Science CrossRef CAS IUCr Journals Google Scholar
Lemonnier, M., Collet, O., Depautex, C., Esteva, J. & Raoux, D. (1978). Nucl. Instrum. Methods, 152, 109–111. CrossRef CAS Web of Science Google Scholar
Liu, L., Alves, M. B., Oliveira, A. C. S., Resende, X. R. & de Sá, F. H. (2021). Proceedings of the 12th International Particle Accelerator Conference (IPAC2021), 24–28 May 2021, Campinas, Brazil, pp. 13–18. MOXA03. Google Scholar
Marques, S. R., Alves, M. B., Arroyo, F. C., Calcanha, M. P., Canova, H. F., Limeira, B. E., Liu, L., Neuenschwander, R. T., Pereira, A. G. C., Tavares, D. O., de Sá F. H., Brunheira, G. O., Cardoso, A. C. T., Cardoso, R. B., Junqueira Leão, R., Leão, L. R., Martins, P. H. S., Moreira, S. S., Oliveira Neto, R. & Siqueira, M. G. (2022). Proceedings of the 13th International Particle Accelerator Conference (IPAC2022), 12–17 June 2022, Bangkok, Thailand, pp. 226–229. MOPOPT002. Google Scholar
Moraes, M. A. L., Caliari, R. M. & Geraldes, R. R. (2020). ASPE 2020 Spring – Design and Control of Precision Mechatronic Systems, 6–8 May 2020, pp. 131–136. American Society for Precision Engineering. Google Scholar
Moreno, G. B. Z. L., Caliari, R. M., Geraldes, R. R. & Moraes, M. A. L. (2018). Proceedings of the 16th International Conference on Accelerator and Large Experimental Control Systems (ICALEPCS2017), 12–16 October 2017, Barcelona, Spain, pp. 1941–1946. THPHA214. Google Scholar
Onuki, H. & Elleaume, P. (2002). Undulators, Wigglers and their Applications. CRC Press. Google Scholar
Piton, J. R., Alnajjar, D. A. D., Araujo, D. H. C., Brito Neto, J. L., do Carmo, L. P., Guedes, L. C. & Moraes, M. A. L. (2021). Proceedings of the 18th International Conference on Accelerators and Large Experimental Physics Control Systems (ICALEPCS2021), 14–22 October 2021, Shanghai, China, pp. 908–911. THPV021. Google Scholar
Sanchez del Rio, M., Canestrari, N., Jiang, F. & Cerrina, F. (2011). J. Synchrotron Rad. 18, 708–716. Web of Science CrossRef CAS IUCr Journals Google Scholar
Saveri Silva, M., Geraldes, R. R., Gilmour, A., Ruijl, T. A. M. & Schneider R. M. (2017). Proceedings of the 9th Mechanical Engineering Design of Synchrotron Radiation Equipment and Instrumentation (MEDSI 2016), 11–16 September 2016, Barcelona, Spain, pp. 194–197. TUPE15. Google Scholar
Sergueev, I., Döhrmann, R., Horbach, J. & Heuer, J. (2016). J. Synchrotron Rad. 23, 1097–1103. Web of Science CrossRef CAS IUCr Journals Google Scholar
Sterbinsky, G. E. & Heald, S. M. (2021). J. Synchrotron Rad. 28, 1737–1746. CrossRef CAS IUCr Journals Google Scholar
Waterstradt, T., Diete, W., Schacht, A., Engblom, C., Scheel, M. & Weitkamp, T. (2018). Presented at 10th International Conference on Mechanical Engineering Design of Synchrotron Radiation Equipment and Instrumentation (MEDSI 2018), 25–29 June 2018, Paris, France. Poster WEPH42. Google Scholar
Yamazaki, H., Ohashi, H., Senba, Y., Takeuchi, T., Shimizu, Y., Tanaka, M., Matsuzaki, Y., Kishimoto, H., Miura, T., Terada, Y., Suzuki, M., Tajiri, H., Goto, S., Yamamoto, M., Takata, M. & Ishikawa, T. (2013). J. Phys. Conf. Ser. 425, 052001. CrossRef Google Scholar
Zhang, L., Sánchez del Río, M., Monaco, G., Detlefs, C., Roth, T., Chumakov, A. I. & Glatzel, P. (2013). J. Synchrotron Rad. 20, 567–580. Web of Science CrossRef CAS IUCr Journals Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.