

beamlines
The High Energy diffraction beamline at the Canadian Light Source
aCanadian Light Source, 44 Innovation Blvd, Saskatoon, SK, S7N 2V3, Canada, and bUniversity of Guelph, 50 Stone Road E, Guelph, Ontario, Canada
*Correspondence e-mail: graham.king@lightsource.ca
The design, performance, and capabilities of the High Energy beamline at the Brockhouse Sector of the Canadian Light Source are described. The beamline uses a single bent silicon wafer as a side-bounce Laue monochromator, using the (111), (422), or (533) hkl reflections to access energies ranging from 25 to 90 keV. The cryogenically cooled crystal serves as the only optical element in the beamline providing a simple, convenient, and reliable configuration. The bending provides a vertical focus as small as 20 µm. The ranges from 1 × 1010 to 1 × 1013 photons s−1, depending on the energy, with typical pre-monochromator slit settings. A large translation table in the hutch moves to follow the beam as the energy is changed. Data are collected using large area detectors. Common uses include rapid collection of powder diffraction data, penetration of thick samples and devices, high pressure diffraction, and pair distribution function measurements.
Keywords: pair distribution function; high energy diffraction; high pressure diffraction; Laue monochromator; hard X-ray beamline.
1. Introduction
High energy X-ray beams produced by synchrotrons enable many types of experiments that are not possible using a laboratory source. The penetration power of these beams coupled with their high operando devices. Access to high Q data also allows the short and medium range structure of materials to be probed by means of the pair distribution function. Several synchrotron facilities currently operate successful user programs on high energy diffraction beamlines (Bernasconi et al., 2015; Chupas et al., 2007
; Connolley et al., 2020
; Dippel et al., 2015
; Drakopoulos et al., 2015
; Isshiki et al., 2001
; Kohara et al., 2007
; Martínez-Criado et al., 2012
, Martínez-Criado et al., 2016
; Mezouar et al., 2005
; Ohara et al., 2018
; Vaughan et al., 2020
; Wright et al., 2020
). This paper describes a new high energy diffraction and scattering beamline located at the Canadian Light Source (CLS).
The Brockhouse X-ray Diffraction and Scattering (BXDS) sector at the CLS is a set of three beamlines powered by two insertion devices that enable a variety of X-ray scattering techniques over a broad range of energies (5–90 keV) to service the scientific communities in Canada and abroad. The High Energy (HE) wiggler beamline is one of the beamlines at BXDS and is the focus of this publication. The HE beamline produces monochromatic X-rays between 25 and 90 keV.
2. Beamline design
The CLS is a 2.9 GeV storage ring consisting of 12 straight sections and typically operates in top up mode with a beam current of 220 mA. BXDS occupies Section 4 of the CLS storage ring. An in-vacuum undulator and in-vacuum wiggler share the straight section and are chicaned by 4 mrad. The wiggler beam first passes through a series of masks and filters. The wiggler beam is then split using a long toroidal mirror which reflects the inboard off-axis section of the beam (Fig. 1). The deflected beam then encounters a monochromator which further deflects the beam outboard to be used at the neighboring Low Energy (LE) wiggler beamline. The selection and design of the wiggler, optics common to both wiggler beamlines, the mirror, as well as the design and performance of the LE beamline have been described previously (Leontowich et al., 2021
).
![]() | Figure 1 Bird's eye layout of the BXDS-HE beamline. Optic 1 is a white beam mirror that splits a side portion of the wiggler beam to Optic 2, which is the BXDS-LE monochromator. The central portion of the wiggler beam passes through to Component 3, white beam slits, followed by a removable diamond screen for white beam viewing (Component 4). Optic 5 is the double bent Laue monochromator, which can switch between two crystals. The monochromatic beam is diffracted to the translation table (Component 6), which holds additional slits, sample stage, detector, etc. |
The central portion of the wiggler radiation fan, which passes by the mirror, then encounters a second filter assembly. This assembly has three filters (from upstream to downstream): 1.8 mm of annealed et al., 2014). This design allows three independently operable beamlines from two insertion devices in a single straight section.
The monochromator has a bent Laue design which serves the triple purposes of deflecting the beam inboard away from the undulator beam, selecting a particular wavelength, and focusing the beam (Gomez et al., 2018). Each crystal is precisely bent against a machined block to achieve a specific radius of curvature, optimizing the desired focus. The monochromator carriage is cooled with liquid nitrogen and contains two rotatable Si single crystal wafers oriented for different Bragg reflections to be accessed on the same rotation plane (Fig. 2
). One crystal is oriented to make use of the (111) reflection while the other is oriented so that the (422) or (533) reflections can be accessed. In general, the (111) crystal is intended to be used for relatively lower energy X-rays (25–37 keV) while the (422)/(533) crystal is meant for higher energy X-rays (39–90 keV). A beryllium window on the downstream end of the monochromator chamber is where the ultra-high vacuum environment of the beamline ends. The distance between the center of the wiggler and the monochromator is 24.54 m.
![]() | Figure 2 CAD model of the liquid N2 cooled crystal carriage of the HE monochromator. There are two carriages that house the Si(111) and the Si(422)/(533) crystals. The carriage is shown without the crystals so that the beam openings can be seen. The red line shows the white beam passing through the (111) opening and hitting the beam stop. The light blue lines show the diffracted monochromatic beam over the angular range it is possible to access by rotation of the stage. The stage can be translated perpendicular to the white beam to switch between crystals. |
This monochromator was built in-house, using Huber motorized stages for the crystal alignment on the beam. The silicon wafers were cut to a rectangular shape, etched to relieve et al., 2018). We also found that the horizontal curvature was skewed and uneven at different sections of the crystals, which could be the reason for the larger experimental horizontal focus when compared with the simulations.
The single crystal monochromator means the path of the beam is now energy dependent. To enable energy selection, all components downstream of the monochromator, including the experimental hutch components, endstations and detectors must follow the 2θ angle of the monochromator. The HE hutch has a large translation table with two translation rails enabling its motion to match the horizontal 2θ radial direction centered on the monochromator. The translation table can access an effective 2θ range of 6–14°, which defines the accessible energies of each monochromator, leaving only a narrow energy blind spot between 37 and 39 keV. The fixed curvature on the Laue crystals causes the optimum focal position to shift upstream as photon energy increases. The HE beamline table and sample motors are controlled through EPICS (EPICS, 2022) with direct user operation through SPEC (Swislow, 1998
) and Bluesky (Allan et al., 2019
).
3. Beamline performance
The et al., 2024) at the CLS using a radiometer. The diode readings were characterized using the absolute beam power readings of a radiometer over several energies and filters to accurately characterize the performance at each energy. The diode was then used at the HE beamline to measure the over the accessible energy range. The tungsten slits were set at 2 mm × 4 mm (horizontal × vertical) as this is the most commonly used setting for actual experiments. If the slits are fully open, the beamline will be roughly double the values described below.
The measured as a function of energy and monochromator reflection. A high of around 1 × 1013 photons s−1 is achieved off the (111) reflection for the lower energy range of this beamline. This compares very closely with simulated values calculated through ray tracing, and slightly higher is observed in the lower energy range using the (111) hkl reflection. It is possible there is variance in the thickness of the Al filters, which would dampen lower energy photons significantly more than higher energies. The Al filters cause the decrease in at energies below 30 keV.
![]() | Figure 3 Measured and simulated flux at maximum beam acceptance and 220 mA ring current across all three monochromator hkl options. |
The (422) and (533) reflections have a large overlap in accessible energies. The (422) reflection provides a higher et al., 2018). In practice, the (422) is generally used for energies between 55 and 70 keV and the (533) for energies between 70 and 90 keV, since for these energies the focal positions occur at convenient positions on the translation table.
Energies, bandwidth, and vertical divergence were determined by analyzing rocking curves of the (333) and (777) reflections of a Si wafer, measured using a two-circle Huber 414 stage stack as described by Leontowich et al. (2021). The results are shown in Figs. 4
and 5
and agree reasonably well with theoretical values.
![]() | Figure 4 Experimentally determined vertical beam divergence measured using the rocking curves of Si (333) and (777) reflections, over-plotted with the simulated divergences for all three crystals hkl reflections. |
![]() | Figure 5 Experimentally measured energy bandwidths using rocking curves of Si (333) and (777) reflections. These values are over-plotted with simulated values from ray tracing calculations. |
Fig. 6 shows the horizontal and vertical beam profile at the smallest vertically focused position at 59.19 keV using the (422) reflection. The profile is created through the first derivative of a knife edge scan. The beam focuses with a full width at half-maximum (FWHM) of 20 µm vertically, but has a large horizontal FWHM of 2.23 mm. Ray tracing simulations predict a similar minimal vertical focus of 15 µm and a horizontal beam size of 1.6 mm. Similar measurements were taken at 30 keV using the (111) reflection and 80 keV using the (533) reflection. The results are shown in Table 1
. Both vertical and horizontal experimental beam are about 25% larger than calculated. Discrepancies in the actual bending radii in comparison with the designed dimensions could account for these differences.
|
![]() | Figure 6 Experimentally determined focused beam profile at the vertical focus position for 60 keV X-rays using the Si (422) monochromator. |
Simulated values of beam xrt (Klementiev & Chernokov, 2014). The ray tracing simulations were initiated by first reconstructing the beamline layout and defining the wiggler source using the targeted storage ring and wiggler magnet parameters including the electron beam energy and current, the wiggler deflection parameter, and an appropriate energy range to simulate (Cutler et al., 2007
; Gomez et al., 2018
). A minimum of a million rays are generated from the wiggler source to ensure a realistic representation of the beamline. Great care was taken when modeling the monochromator crystals as it is often very difficult to accurately model anticlastic crystals. The xrt software implements the Takagi–Taupin model to include effects of crystal strain and volumetric diffraction (Klementiev & Chernokov, 2023
). The monochromator crystals were built within the simulations according to the targeted radii of curvature and asymmetry angle. No crystal mosaicity is considered in these calculations, and a Poisson ratio of 0.26 has been adopted (Kim et al., 2001
; Hopcroft et al., 2010
). This model accurately predicts the sagittal focal points expected for these crystals at various energies. Beam size was subsequently determined from the FWHM of the simulated beam profiles when the beam was at sagittal focus. Beam divergence was determined geometrically from the beam size at sagittal focus and the beam size at the monochromator crystal. X-ray beam and energy bandpass are determined automatically by xrt during ray generation. Ray generation in xrt is performed through pulses where each photon is assigned an energy within the energy range of the source. The beam is then determined by monitoring the number of photons that successfully reach the detector position over the duration of the pulse. Energy bandpass is subsequently calculated from the FWHM of the energy distribution of photons that have reached the detector position.
4. Beamline capabilities and uses
The two detectors currently available are a Varex XRD 4343CT and a Dectris Eiger2 4M. As nearly the entire Debye-Scherrer ring is collected and integrated (except where the shadow of the rod holding the beamstop is located), very high statistic data can be collected very rapidly. Complete powder diffraction patterns with low noise can usually be collected in just a few seconds and often in just tenths of a second. This makes the beamline well suited for rapidly acquiring data on large numbers of samples, monitoring fast chemical reactions, or collecting temperature dependent data using a fine temperature increment (King, 2020). The two-dimensional information given by the area detector can also be used for texture analysis. Examples of a typical two dimensional diffraction pattern and a of the radially integrated data are shown Fig. 7
.
![]() | Figure 7 Top: 2D diffraction pattern of LaB6 collected using 55.32 keV X-rays with the Varex detector placed 923 mm away from the sample. Bottom: a of the integrated LaB6 data. Black circles are the data points, the red line is the fit, and the difference is shown beneath. The Rwp for the fit is 1.91. |
Many experiments have used the higher energies of the beamline to collect powder diffraction data on thick samples or devices. Examples include monitoring phase changes in high strength steels during rapid heating and cooling (Ghosh et al., 2022; Singh et al., 2022
; Wang et al., 2021
) and tensile measurements on pipes or sheets of metal (Kong et al., 2023
). Another popular use is monitoring of phase changes within batteries. Either prototype or fully functional commercial battery pouch cells or coin cells can be placed on the beamline and the composition can be studied at various locations in the battery as a function of cycling (Bond et al., 2024
).
The beamline is also capable of performing high pressure experiments using a diamond anvil cell (DAC) (Ukoji et al., 2024; Guan et al., 2024
). Typically, an energy around 35 keV is used for such experiments, as it has the proper penetration to transmit through the cell and provides data in a useful d-spacing range that matches the exit angle of the cell. A setup containing a collimator and pinhole is used to reduce the beam size to a diameter as small as 20 µm. A motorized stage stack is used to hold and center the DAC. Alignment is achieved through an active beamstop which has a diode embedded in it to monitor the transmitting through the components.
High energy X-rays also allow access to high Q data, which is essential for producing pair distribution functions (PDFs) with good real space resolution. One of the main uses of this beamline is the collection of high quality PDF data. The PDF gives the distribution of interatomic distances in a material and is an excellent probe of the short and medium range structural features. It can be applied to almost any area of research, including poorly crystalline minerals (King et al., 2020), single atom catalyst (Chen et al., 2022
; Xia et al., 2021
), magnetic materials with site disorder (Lozano-Gorrín et al., 2021
), glasses (Singh et al., 2021
; Wang et al., 2022
), and nanoparticles (Sulaiman et al., 2022
), to name a few. On this beamline, PDF data are most commonly collected using 60 keV X-rays, which allows a Qmax around 27 Å−1 with the large 43 cm × 43 cm Varex detector placed close to the sample. Using higher energies, it is possible to collect data well above 30 Å−1. A new methodology for PDF data collection has recently been developed on this beamline which uses the area detector in an inclined geometry to cover a 2θ range 0–90°. This geometry has several advantages such as extending the accessible Q range and an improved signal to noise ratio at higher scattering angles (Burns et al., 2023
). Some examples of data collected in this configuration are shown in Fig. 8
and a photograph of the setup is shown in Fig. 9
.
![]() | Figure 8 (a) A one-dimensional reduced total scattering structure function F(Q) and a two-dimensional QF(Q) of amorphous selenium where white/red indicates positive intensity and black indicates negative intensity. (b) A PDF of amorphous selenium generated using a Qmax of 50 Å−1. (c) A PDF of Ni metal generated using a Qmax of 42.5 Å−1. Data were collected on a Perkin Elmer detector using (a, b) 72 keV and (c) 60 keV radiation. |
![]() | Figure 9 Photograph of inside the BXDS-HE experimental hutch with the detector arranged in tilted geometry. The detector is inclined approximately 60° from normal incidence geometry. |
A number of different sample environments are available. For variable temperature experiments a cryostream can precisely control the temperature between 80 and 500 K and a capillary furnace can heat as high as 1473 K while also flowing gas. A gas loading system is available for high pressure experiments. Special battery holders have been designed to use with battery cyclers. The beamline is equipped with a 10 kN stress rig integrated with an infra-red furnace to accommodate various types of strain experiments.
5. Summary
A new high energy wiggler beamline at the Canadian Light Source is available for conducting high energy diffraction, high pressure diffraction, and pair distribution function studies. The beamline uses a single bent Si crystal monochromator and a translation table in the hutch to follow the beam. The in situ/operando experiments under extreme conditions with large samples and environments for new alloys, catalysis, batteries, among others. Since opening for users in 2020, the user's community has been rapidly growing, taking advantage of these new beamline capabilities in numerous research areas.
bandwidth, divergence, and focus of the beam have been measured and compare well with theoretical values determined from ray tracing calculations. The high energies are suitable forAcknowledgements
We thank the entire CLS staff for their support during the construction and commissioning of this beamline. We specifically thank Johannes Vogt and Omar Marinos for assistance in using the radiometer. All of the research described in this paper was performed at the Canadian Light Source, a national research facility of the University of Saskatchewan, which is supported by the Canada Foundation for Innovation (CFI), the Natural Sciences and Engineering Research Council (NSERC), the National Research Council (NRC), the Canadian Institutes of Health Research (CIHR), the Government of Saskatchewan, and the University of Saskatchewan.
Funding information
The following funding is acknowledged: Canada Foundation for Innovation; Natural Sciences and Engineering Research Council of Canada; National Research Council Canada; Canadian Institutes of Health Research; University of Saskatchewan.
References
Allan, D., Caswell, T., Campbell, S. & Rakitin, M. (2019). Synchrotron Radiat. News, 32(3), 19–22. CrossRef Google Scholar
Bernasconi, A., Wright, J. & Harker, N. (2015). Powder Diffr. 30, S2–S8. CrossRef CAS Google Scholar
Bond, T., Gauthier, R., King, G., Dressler, R., Abraham, J. J. & Dahn, J. R. (2024). J. Electrochem. Soc. 171, 110514. CrossRef Google Scholar
Burns, N., Rahemtulla, A., Annett, S., Moreno, B. & Kycia, S. (2023). J. Appl. Cryst. 56, 510–518. Web of Science CrossRef CAS IUCr Journals Google Scholar
Chen, F.-Y., Wu, Z.-Y., Gupta, S., Rivera, D. J., Lambeets, S. V., Pecaut, S., Kim, J. Y. T., Zhu, P., Finfrock, Y. Z., Meira, D. M., King, G., Gao, G., Xu, W., Cullen, D. A., Zhou, H., Han, Y., Perea, D. E., Muhich, C. L. & Wang, H. (2022). Nat. Nanotechnol. 17, 759–767. CrossRef CAS PubMed Google Scholar
Chupas, P. J., Chapman, K. W. & Lee, P. L. (2007). J. Appl. Cryst. 40, 463–470. Web of Science CrossRef CAS IUCr Journals Google Scholar
Connolley, T., Beavers, C. M. & Chater, P. (2020). Synchrotron Radiat. News 33(6), 31–36. CrossRef Google Scholar
Cutler, J., Hallin, E., de Jong, M., Thomlinson, W. & Ellis, T. (2007). Nucl. Instrum. Methods Phys. Res. A, 582, 11–13. Web of Science CrossRef CAS Google Scholar
Diaz, B., Gomez, A., Meyer, B., Duffy, A., Hallin, E. & Kycia, S. (2014). Rev. Sci. Instrum. 85, 085104. Web of Science CrossRef PubMed Google Scholar
Dippel, A.-C., Liermann, H.-P., Delitz, J. T., Walter, P., Schulte-Schrepping, H., Seeck, O. H. & Franz, H. (2015). J. Synchrotron Rad. 22, 675–687. Web of Science CrossRef CAS IUCr Journals Google Scholar
Drakopoulos, M., Connolley, T., Reinhard, C., Atwood, R., Magdysyuk, O., Vo, N., Hart, M., Connor, L., Humphreys, B., Howell, G., Davies, S., Hill, T., Wilkin, G., Pedersen, U., Foster, A., De Maio, N., Basham, M., Yuan, F. & Wanelik, K. (2015). J. Synchrotron Rad. 22, 828–838. Web of Science CrossRef CAS IUCr Journals Google Scholar
EPICS (2022). The Experimental Physics Control System, https://epics-controls.org/. Google Scholar
Gasilov, S., Webb, M. A., Panahifar, A., Zhu, N., Marinos, O., Bond, T., Cooper, D. M. L. & Chapman, D. (2024). J. Synchrotron Rad. 31, 1346–1357. CrossRef CAS IUCr Journals Google Scholar
Ghosh, S., Wang, S., Singh, H., King, G., Xiong, Y., Zhou, T., Huttula, M., Kömi, J. & Cao, W. (2022). J. Mater. Res. Technol. 20, 485–495. CrossRef CAS Google Scholar
Gomez, A., Dina, G. & Kycia, S. (2018). Rev. Sci. Instrum. 89, 063301. Web of Science CrossRef PubMed Google Scholar
Guan, R., Liu, J., Kutty, A., Yu, Z., Ji, Y., Li, Y. & Song, Y. (2024). J. Mater. Chem. C. 12, 14833–14843. CrossRef CAS Google Scholar
Hopcroft, M. A., Nix, W. D. & Kenny, T. W. (2010). J. Microelectromech. Syst. 19, 229–238. Web of Science CrossRef CAS Google Scholar
Isshiki, M., Ohishi, Y., Goto, S., Takeshita, K. & Ishikawa, T. (2001). Nucl. Instrum. Methods Phys. Res. A, 467–468, 663–666. Web of Science CrossRef CAS Google Scholar
Kim, J., Cho, D. & Muller, R. S. (2001). Transducers '01 Eurosensors XV – The 11th International Conference on Solid-State Sensors and Actuators, 10–14 June 2001, Munich, Germany. Google Scholar
King, G. (2020). Acta Cryst. B76, 789–794. CrossRef IUCr Journals Google Scholar
King, G., Celikin, M., Gomez, M. A., Becze, L., Petkov, V. & Ventura, G. D. (2020). Environ. Sci. Nano, 7, 3735. CrossRef Google Scholar
Klementiev, K. & Chernokov, R. (2014). Proc. SPIE, 9209, 92090A. Google Scholar
Klementiev, K. & Chernokov, R. (2023). Synchrotron Radiat. News, 35(5), 23–27. CrossRef Google Scholar
Kohara, S., Itou, M., Suzuya, K., Inamura, Y., Sakurai, Y., Ohishi, Y. & Takata, M. (2007). J. Phys. Condens. Matter, 19, 506101. Web of Science CrossRef Google Scholar
Kong, J., Li, T. J. H., Haché, M. J. R., Tam, J., McCrea, J. L., Howe, J. Y. & Erb, U. (2023). Acta Mater. 246, 118711. CrossRef Google Scholar
Leontowich, A. F. G., Gomez, A., Diaz Moreno, B., Muir, D., Spasyuk, D., King, G., Reid, J. W., Kim, C.-Y. & Kycia, S. (2021). J. Synchrotron Rad. 28, 961–969. Web of Science CrossRef CAS IUCr Journals Google Scholar
Lozano-Gorrín, A. D., Wright, B., Dube, P. A., Marjerrison, C. A., Yuan, F., King, G., Ryan, D. H., Gonzalez-Silgo, C., Cranswick, L. M. D., Grosvenor, A. P. & Greedan, J. E. (2021). ACS Omega, 6, 6017–6029. PubMed Google Scholar
Martínez-Criado, G., Tucoulou, R., Cloetens, P., Bleuet, P., Bohic, S., Cauzid, J., Kieffer, I., Kosior, E., Labouré, S., Petitgirard, S., Rack, A., Sans, J. A., Segura-Ruiz, J., Suhonen, H., Susini, J. & Villanova, J. (2012). J. Synchrotron Rad. 19, 10–18. Web of Science CrossRef IUCr Journals Google Scholar
Martínez-Criado, G., Villanova, J., Tucoulou, R., Salomon, D., Suuronen, J.-P., Labouré, S., Guilloud, C., Valls, V., Barrett, R., Gagliardini, E., Dabin, Y., Baker, R., Bohic, S., Cohen, C. & Morse, J. (2016). J. Synchrotron Rad. 23, 344–352. Web of Science CrossRef IUCr Journals Google Scholar
Mezouar, M., Crichton, W. A., Bauchau, S., Thurel, F., Witsch, H., Torrecillas, F., Blattmann, G., Marion, P., Dabin, Y., Chavanne, J., Hignette, O., Morawe, C. & Borel, C. (2005). J. Synchrotron Rad. 12, 659–664. Web of Science CrossRef CAS IUCr Journals Google Scholar
Ohara, K., Tominaka, S., Yamada, H., Takahashi, M., Yamaguchi, H., Utsuno, F., Umeki, T., Yao, A., Nakada, K., Takemoto, M., Hiroi, S., Tsuji, N. & Wakihara, T. (2018). J. Synchrotron Rad. 25, 1627–1633. CrossRef IUCr Journals Google Scholar
Singh, H., Shu, Q., King, G., Liang, Z., Wang, Z., Cao, W., Huttula, M. & Fabritius, T. (2021). J. Am. Ceram. Soc. 104, 4505–4517. CrossRef CAS Google Scholar
Singh, H., Xiong, Y., Rani, E., Wang, S., Kharbach, M., Zhou, T., Yao, H., Niu, Y., Zakharov, A., King, G., de Groot, F. M. F., Kömi, J., Huttula, M. & Cao, W. (2022). NPJ Mater. Degrad. 6, 54. Google Scholar
Sulaiman, K. O., Zubair, M., King, G., Bedford, N. M. & Scott, R. W. J. (2022). Phys. Chem. Chem. Phys. 24, 24834–24844. CrossRef CAS PubMed Google Scholar
Swislow, G. (1998). SPEC X-ray Diffraction Software, https://www.certif.com. Google Scholar
Ukoji, N., Rodriguez, D., Kuang, H., Desgreniers, S. & Tse, J. S. (2024). Commun. Chem. 7, 162. CrossRef PubMed Google Scholar
Vaughan, G. B. M., Baker, R., Barret, R., Bonnefoy, J., Buslaps, T., Checchia, S., Duran, D., Fihman, F., Got, P., Kieffer, J., Kimber, S. A. J., Martel, K., Morawe, C., Mottin, D., Papillon, E., Petitdemange, S., Vamvakeros, A., Vieux, J.-P. & Di Michiel, M. (2020). J. Synchrotron Rad. 27, 515–528. Web of Science CrossRef CAS IUCr Journals Google Scholar
Wang, S., Kistanov, A. A., King, G., Ghosh, S., Singh, H., Pallaspuro, S., Rahemtulla, A., Somani, M., Kömi, J., Cao, W. & Huttula, M. (2021). Acta Mater. 221, 117361. CrossRef Google Scholar
Wang, S., Rani, E., Gyakwaa, F., Singh, H., King, G., Shu, Q., Cao, W., Huttula, M. & Fabritius, T. (2022). Inorg. Chem. 61, 7017–7025. CrossRef CAS PubMed Google Scholar
Wright, J., Giacobbe, C. & Majkut, M. (2020). Curr. Opin. Solid State Mater. Sci. 24, 100818. Web of Science CrossRef Google Scholar
Xia, C., Qiu, Y., Xia, Y., Zhu, P., King, G., Zhang, X., Wu, Z., Kim, J. Y., Cullen, D. A., Zheng, D., Li, P., Shakouri, M., Heredia, E., Cui, P., Alshareef, H. N., Hu, Y. & Wang, H. (2021). Nat. Chem. 13, 887–894. CrossRef CAS PubMed Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.