Figure 2
(a) Integrated overview of the SIBIPIRUNA beamline, showing photon delivery and the two experimental stations located at SIRIUS and Orion, including their respective sample environments and transfer systems. (b) Optical layout based on dipole radiation, plane grating monochromator, and reflective optics for beam conditioning and energy selection in the soft X-ray regime. (c) Full-field transmission X-ray microscope configuration, highlighting the condenser optics, sample stage, Fresnel zone plate objective, and detector system used for cryogenic imaging. (d) Sample preparation and correlative workflow, illustrating cryogenic preparation, transfer, and integration with complementary imaging modalities. |