research papers
A multipurpose instrument for time-resolved ultra-small-angle and coherent X-ray scattering
aESRF – The European Synchrotron, 38043 Grenoble, France
*Correspondence e-mail: narayan@esrf.fr
This article presents the main technical features and performance of the upgraded beamline ID02 at the ESRF. The beamline combines different small-angle X-ray scattering techniques in one unique instrument, enabling static and kinetic investigations from ångström to micrometre size scales and time resolution down to the sub-millisecond range. The main component of the instrument is an evacuated detector tube of length 34 m and diameter 2 m. Several different detectors are housed inside a motorized wagon that travels along a rail system, allowing an automated change of the sample–detector distance from about 1 to 31 m as well as selection of the desired detector. For optional combined wide-angle scattering measurements, a wide-angle detector is installed at the entrance cone of the tube. A scattering vector (of magnitude q) range of 0.002 ≤ q ≤ 50 nm−1 is covered with two sample–detector distances and a single-beam setting for an X-ray wavelength of 1 Å. In the high-resolution mode, two-dimensional ultra-small-angle X-ray scattering patterns down to q < 0.001 nm−1 can be recorded, and the resulting one-dimensional profiles have superior quality as compared to those measured with an optimized Bonse–Hart instrument. In the highest-resolution mode, the beam is nearly coherent, thereby permitting multispeckle ultra-small-angle X-ray photon correlation spectroscopy measurements. The main applications of the instrument include the elucidation of static and transient hierarchical structures, and nonequilibrium dynamics in soft matter and biophysical systems.
Keywords: small-angle X-ray scattering; SAXS; ultra-small-angle X-ray scattering; USAXS; X-ray photon correlation spectroscopy; XPCS; time-resolved X-ray scattering.
1. Introduction
Over the past several decades, high-brilliance small-angle X-ray scattering (SAXS) and related techniques have been widely used in the investigation of soft matter and noncrystalline biological systems (Narayanan et al., 2017; Tuukkanen et al., 2017). A unique feature of these scattering methods is that they offer simultaneously high spatial and temporal resolution, albeit in Moreover, scattering experiments can be combined with a variety of thermophysical, rheological and biophysical techniques, thereby further enhancing the structural and dynamical information (Narayanan et al., 2017; Sakurai, 2017). Traditionally, SAXS and wide-angle X-ray scattering (WAXS) experiments and ultra-small-angle X-ray scattering (USAXS) experiments have been performed using two separate instruments in tandem (Narayanan et al., 2001; Ilavsky et al., 2018). In this case, the USAXS setup involves a Bonse–Hart multi-crystal instrument (Sztucki & Narayanan, 2007; Ilavsky et al., 2009). However, pinhole two-dimensional USAXS has some advantages, especially for investigating highly oriented specimens and in time-resolved studies (Yagi & Inoue, 2003; Shinohara et al., 2007; Kishimoto et al., 2014). In addition, high angular resolution and two-dimensional diffraction patterns are essential for performing crystallography at the colloidal scale (Petukhov et al., 2015).
This article presents the main technical features and performance of the upgraded beamline ID02 at the ESRF. The primary motivation for this beamline project, named time-resolved ultra-small-angle X-ray scattering (TRUSAXS), was to enlarge the applicability of SAXS to relatively unexplored nonequilibrium processes of fundamental and practical interest. Examples range from complexities of hierarchically self-assembled biomimetic systems (Valéry et al., 2011) to pathways of cellular regulation (Piazzesi et al., 2018). To make SAXS techniques useful in the investigation of such complex systems, significant enhancements of the range of accessible scattering vector, angular resolution, scattering detectivity and time resolution are required. While improving the performance of high-brilliance SAXS and WAXS techniques, the TRUSAXS instrument offers a unique feature of two-dimensional USAXS and ultra-small-angle X-ray photon correlation spectroscopy (USA-XPCS), allowing one to probe microstructure and dynamics in the same experiment (Möller, Chushkin et al., 2016). Although USA-XPCS can be performed with a Bonse–Hart instrument (Sztucki & Narayanan, 2007; Zhang et al., 2011), multispeckle analysis (Möller, Chushkin et al., 2016) and time-resolved measurements are required for the investigation of nonequilibrium systems (Dattani et al., 2017; Möller & Narayanan, 2017). Having SAXS, WAXS and USAXS in the same setup greatly enhances the exploitation of USAXS in soft-matter and biophysical studies.
The main technical feature of the TRUSAXS instrument is an evacuated detector flight tube (Van Vaerenbergh et al., 2016) that permits a continuous variation of the sample–detector distance from 0.8 to 30.8 m. Using several sample–detector distances and an X-ray wavelength (λ) of 1 Å, measurements over a q range of 10−3 ≤ q ≤ 8 nm−1 can be performed. Here, q is the magnitude of the scattering vector given by q = (4π/λ)sin(θ/2), with θ the scattering angle. When SAXS and USAXS are combined with WAXS, the setup covers a q range of 10−3 ≤ q ≤ 50 nm−1, corresponding to more than four orders of magnitude in dimensions. This broad q range together with sub-millisecond time resolution enables a wide range of dynamical studies from the molecular level to the upper limit of the colloidal scale. The use of long collimation distances diminishes the background of the instrument over the conventional SAXS range, thereby allowing studies of extremely dilute systems. To cover a high in q and scattered intensity, I(q), several configurations are implemented. In the high-brilliance mode, the (in photons s−1) is maximized for the highest time resolution and detection capability. In the high-resolution mode, the high angular resolution is obtained with moderate In the coherence mode, the beam is tightly collimated at the expense of flux.
2. Technical features
The schematic layout of the beamline is depicted in Fig. 1. The beamline consists of five sections: the source, principal optics, collimation slits, experimental station and detector tube. The relatively long separation between the principal optical components and the sample permits better collimation of the beam, thereby significantly curtailing the background, which is essential for performing USAXS with a two-dimensional detector. The SAXS detectors are enclosed in the wagon within the detector tube and the WAXS detector is mounted outside in air above the entrance cone of the tube.
2.1. Undulator source
The source is located at a high β section of the present storage ring with a full width at half-maximum (FWHM) size and divergence of, respectively, 9 µm and 14 µrad in the vertical and 900 µm and 28 µrad in the horizontal. With the Extremely Brilliant Source (EBS) upgrade, the horizontal beam size and divergence will be reduced to about 56 µm and 17 µrad, respectively, while the vertical parameters remain similar. The X-ray beam is delivered by two phased U21.4 or a single U35 undulator of length 1.6 m each and minimum gap of 11 mm. While U21.4 undulators are primarily used for fixed-energy (E) operation (at 12.4 keV and maximum up to 15 keV), the U35 device covers the full energy range of the beamline, 8−20 keV. The front end has a polished diamond window (thickness ∼300 µm) that separates the vacuum of the storage ring from the beamline. The phasing of the two U21.4 undulators increases the as compared to two separate devices. The exact gain in depends on the opening of the primary slits and it increases with decreasing slit gap. In the standard operation, the primary slits are closed to 0.8 × 0.8 mm and the is about 2.3 times larger when two U21.4s are used in tandem as compared to a single U21.4, while this factor increases to 3 for a primary slit opening of 0.3 × 0.3 mm. The combination of two U21.4 undulators and a primary slit aperture of 0.8 × 0.8 mm provides a monochromatic (ΔE/E ≃ 2 × 10−4) of about 6 × 1013 photons s−1 at the sample position. Between 12.4 and 15.2 keV, the U21.4 device delivers more than the U35 undulator.
2.2. Primary optics
The main optical elements consist of the primary and secondary slits, the cryogenically cooled monochromator, and the focusing optics. The liquid-nitrogen-cooled monochromator is a generic design using a channel-cut Si 111 crystal (Crystal Scientific Ltd) with a channel width of 5 mm. The channel-cut is similar to a Z-cut, which allowed fine polishing and etching of the diffracting surfaces. The channel-cut configuration is not a fixed exit monochromator, and the 5 mm gap results in an increase of beam height by 9.87 mm at 12.4 keV and a height variation of about 50 µm when the energy is increased from 12.4 to 16 keV. The monochromator can withstand a heat load up to 200 W and a power density of the order of 100 W mm−2 without significant loss in intensity and broadening of the rocking curve (Zhang et al., 2013). In addition, an independent tuning of the second diffracting surface is facilitated by a motorized pusher arrangement. The reproducibility of the energy scan is about 0.3 eV.
The choice of focusing optics was guided by several factors. First of all, they should allow observation of the smallest possible scattering angle for a given sample–detector distance and maintain very high temporal stability. Secondly, to preserve the high Δβ, the sagittal slope error) since the deviation from ideal focusing (δv) is given by δv = 2sin(αi)Δβ, with αi the incident grazing angle. In other words, the sin(αi) multiplicative factor diminishes the effect of slope error and enables an ideal source size to be obtained in the vertical direction. However, the horizontal beam size becomes more sensitive to the meridional slope error (Δα) as the deviation (δh) is given by δh = 2Δα. With the present horizontal source parameters, no difference can be observed for a slope error Δα ≃ 0.6 µrad. As a result the double-mirror configuration with the best possible toroidal mirror followed by a fine planar mirror preserves the source properties and maintains high temporal stability. The main optical parameters of the mirrors (Thales SESO SAS) are provided by Van Vaerenbergh et al. (2016).
the optics should be able to accept the full beam after the monochromator. Thirdly, the change of the sample–detector distance should be relatively easy without needing to refocus the beam. Additionally, the constraints imposed by the building require that the reflected beam is maintained parallel to the incident beam all along the beamline. Finally, the focusing optics should also reject the harmonics in the undulator spectrum which, in order to fulfill the radiation protection regulations, must be below 0.01%. All these conditions are met with a double-mirror scheme consisting of a toroidal mirror and a planar mirror in tandem, which reflects the beam in the horizontal direction. In this scheme, the vertical focusing (sagittal) is less sensitive to the slope error of the mirror (The focusing distance is varied from the sample position to the farthest detector position by changing the inclination of the double-mirror assembly from about 2.6 to 3.5 mrad while maintaining the reflected beam along the same path. When changing the inclination of a toroidal mirror, the horizontal and vertical focal lengths change in opposite directions and consequently the horizontal and vertical beam sizes cannot be simultaneously optimized at a given position. This is not a severe constraint since the beam at the mirror is only 1.5 times larger than at the source horizontally, while it is 50 times larger vertically. Therefore, changing the focal length by inclining the mirror has much less influence horizontally than vertically. The minimum FWHM vertical beam size varies from approximately 30 to 60 µm from the sample to the farthest detector position and the corresponding horizontal beam size changes from 600 to 1200 µm (for a primary slit size of 0.8 × 0.8 mm). Fig. 2(a) shows the beam size measured along the detector tube using a high-resolution camera (beam viewer). The variation of beam size along the detector tube for a fixed focusing is depicted in Fig. 2(b). The measured values are consistent with the expected sizes from the geometrical optics and deviations can be described by ray-tracing calculation in XOP (Sánchez del Río & Dejus, 2011). The enlargement of the vertical beam size from the ideal source size is attributed to a cumulative effect of residual optical aberrations and mechanical vibrations. Further reduction of the beam size is achieved by slit collimation as shown in §3.3.
2.3. Secondary optics
Apart from the main optical elements described above, high-power primary slits (Marion & Zhang, 2004) are used to reduce the heat load on the monochromator. In addition, these slits are used together with the secondary slits for beam collimation. In the standard configuration, the last two pairs of slits serve as guard slits to curtail the parasitic background from the upstream optics and slits. The last guard slits are scatterless slits (Li et al., 2008) with hybrid metal/silicon crystal blades (Xenocs SA). When a smaller size than the focused beam is required, the first guard slits are also employed for collimation and only the last slits serve as guard slits. However, the slit collimation is effective only up to a certain aperture size (ca 50 and 20 µm along vertical and horizontal directions, respectively), below which the diffraction effect broadens the beam size. The differences are due to different vertical and horizontal transverse coherence lengths. By optimizing primary, secondary and guard slits, a beam of 20 × 20 µm and divergence of ∼1 µrad can be obtained near the sample position, but at the expense of (∼1010 photons s−1). For these small sizes, the beam is highly coherent, producing speckles in the scattered field (Sutton, 2008).
Alternatively, the horizontal beam divergence could be reduced by using multiple-bounce pseudo-channel-cut crystals (Sztucki et al., 2008), but again at the expense of As compared to extreme slit collimation, this scheme provides a lower background when relatively weak scattering needs to be measured. For variable beam sizes and coherence applications, the slit collimation is preferred.
2.4. Detector tube
The detector flight tube, which has a length of 34 m and a diameter of 2 m, is made of 8 mm thick stainless steel (Added Value Solutions, AVS). Fig. 3(a) shows the top view of the detector tube and the experiment station. Three different SAXS detectors are housed inside a sealed motorized wagon that travels along a rail system within the vacuum tube as depicted in Fig. 3(b). For SAXS/USAXS experiments, this arrangement enables automated change of the sample–detector distance between 0.8 and 30.8 m in a precise and reproducible way (Van Vaerenbergh et al., 2016). The parasitic lateral movement of the wagon is within ±300 µm over the full stroke of 30 m. The wagon assembly comprises a cabin, a carriage and a translation table which permits further repositioning of the cabin within 100 µm over the 30 m stroke. For normal operation, the interior of the cabin is at atmospheric pressure with vacuum outside. The 100 m3 tube volume is evacuated to 10−2 mbar (1 Pa) in less than 2 h by an industrial dry screw pump (Edwards GX450). An X-ray-transparent window made of sheet (thickness ∼400 µm and diameter 240 mm) is glued and bolted onto the front flange of the cabin, and it separates the vacuum in the tube and the atmosphere inside the wagon. The temperature inside the cabin is regulated to 298 K by a heat exchanger system.
Inside the cabin, three detectors, workhorse SAXS/USAXS (Rayonix MX 170 HS), high-resolution USAXS (FReLoN 4M) and time-resolved SAXS (Pilatus 300K, Dectris) detectors, and a high-resolution camera (beam viewer) are installed on a translation table. This arrangement allows positioning of the desired detector behind the X-ray window while shielding the unused detectors. The WAXS detector (Rayonix LX 170 HS) is mounted outside the tube around the entrance cone as depicted in Fig. 3(c), and it can be lifted up by 0.5 m when not in use. A flexible beamstop system is installed in front of the X-ray window in vacuum, which blocks the direct beam (Van Vaerenbergh et al., 2016). The primary beamstop is made of a lead block (height 2 mm, width 5 mm and thickness 2 mm) with an embedded silicon p–i–n photodiode. The signal from the photodiode is fed to a variable-gain low-noise current amplifier (DLPCA-200, FEMTO) that enables the measurement of transmitted beam intensity (IT) with high precision. In addition, the beamstop system is equipped with six independent frames with auxiliary beamstops (Van Vaerenbergh et al., 2016) of circular or rectangular shape with sizes ranging from 1 to 12 mm. They provide a high degree of flexibility in different applications, including fiber diffraction, XPCS and grazing-incidence SAXS. The detector wagon has an absolute position encoder which enables the measurement of translation along the rail with a precision better than 0.1 mm. In addition, the distances from a reference flange in front of the detector tube to the different detectors are precisely calibrated using silver behenate at multiple detector positions. Therefore, the SAXS/USAXS detector distance calibration involves only an accurate measurement of the separation between the sample and the reference flange outside the tube. The WAXS detector distance is usually calibrated with a reference sample, e.g. para-bromobenzoic acid (PBBA).
2.5. Experiment station
The experiment station is equipped with essential infrastructure (power outlets, Ethernet sockets, compressed air or nitrogen, chilled water etc.) for multipurpose in situ experiments. Fig. 3(c) is a sketch of the sample area inside the lead hutch. The main sample table has motorized movements of 600 mm along the horizontal and 350 mm along the vertical with 1 µm precision. This table provides a very stable platform, and relatively large sample environments (maximum 300 kg) can be placed on it. When required, additional fast translation and precision rotation stages are installed on the main sample table. These auxiliary stages facilitate fast vertical movement, rotation along the vertical and tilt along two horizontal axes. The sample environment is placed in air. The vacuum upstream and in the detector tube is isolated by thin high-quality mica windows of thickness ∼15–25 µm, and diameter 10 and 12–16 mm, respectively. The last guard slits are mounted on a sturdy retractable telescopic tube arrangement which enables minimization of the air gap without breaking the vacuum. The experiment control cabin is located just outside the radiation-protected lead hutch. Adequate coaxial cables and serial lines are provided to record signals from sample environments and interface user-supplied instruments with the data acquisition.
2.6. Detectors
In general, SAXS, WAXS and USAXS detectors need to have low intrinsic noise and sufficient sensitivity to provide single-photon detectivity together with high e.g. MX170 HS) meets some of these specifications in terms of overall size (170 × 170 mm), spatial resolution (∼85 µm), sensitivity (the average single-photon level is above the readout noise at 193 K for typical integration times) and frame rate (100 frames per second in 10 × 10 binning with significantly reduced spatial resolution).
and spatial resolution. A high frame rate is essential for time-resolved experiments. In addition, the SAXS detector should have a relatively large active area, preferably without gaps. It is not always possible to meet these specifications with one detector. Therefore, several detectors are utilized to obtain the required characteristics in combination. To meet the high-count-rate, spatial resolution and radiation hardness requirements, fiber-optically coupled CCD-based detectors are employed. For SAXS, a Rayonix detector using split frame transfer CCDs (For USAXS, a high spatial resolution and et al., 2007) based on a large-area Kodak CCD sensor (KAF-4320) bonded to a thick fiber-optic faceplate (50 × 50 mm) with 1:1 coupling (FReLoN 4M). This detector provides a higher spatial resolution (∼44 µm), high sensitivity (single-photon signal about 5 times above the noise level) and reasonable (107 photons mm−2). For compatibility and software integration reasons, a Rayonix LX170 HS (85 × 170 mm) was chosen as the WAXS detector. Moreover, this detector has a groove cut at the middle of the long edge that permits the SAXS intensity to pass unobstructed. To complement the CCD detectors, a photon-counting PILATUS 300K detector is installed for highly time-resolved SAXS and XPCS experiments. This detector delivers a maximum frame rate of 400 Hz with a minimum 0.2 ms exposure. Table 1 provides a summary of the observed technical specifications of the detectors. The Rayonix detectors are operated in the low-noise mode with slightly reduced frame rate compared with the standard mode. The spatial resolution refers to the FWHM point spread function (PSF), which is measured by means of an attenuated beam of 20 µm centered on the pixel and by applying the corresponding Gaussian deconvolution. The pixel detector has a sharp PSF, nearly a box profile with one pixel width, which when approximated by a Gaussian function has an FWHM roughly half the pixel size, as listed in Table 1. Furthermore, an Eiger 500k detector module (Dinapoli et al., 2011), which has a pixel size of 75 µm, active area 80 × 40 mm and frame rate up to 22 kHz can be used for XPCS measurements. In the future, it is foreseen that the Rayonix MX170 HS and Pilatus 300K detectors will be replaced by a larger Eiger 4M (Dectris) pixel detector.
are critical. Therefore, a specific development was undertaken within the ESRF detector group to assemble a fast readout low-noise (FReLoN) detector (Labiche
|
Besides the three SAXS detectors, the cabin also houses a high-resolution camera primarily for recording the direct beam profile. This beam viewer is based on a high-resolution scintillator lens coupled to a Basler Gigabit Ethernet CCD camera with a frame rate of about 30 s−1. The pixel size is 4.8 µm, corresponding to a Nyquist resolution of 9.6 µm (Martin & Koch, 2006). In addition to the beamstop photodiode, several beam monitors based on Kapton foil (25 µm) scatterers and p–i–n photodiodes fed to FEMTO current amplifiers (AMP) are installed along the beamline for monitoring the incident beam intensity (I0) and alignment purposes. These intensity monitors are calibrated using a reference photodiode where the absolute photocurrent and absorption are directly measured.
2.7. Fast beam shutter
The X-ray exposure of the sample is controlled by a fast beam shutter (FBS) that opens only for the duration of the specified acquisition time, and prevents the sample and detectors from continuous exposure during the idle time. The FBS is constructed by a tandem arrangement of two laser shutters (Uniblitz LS2, Vincent Associates) with an aperture of 2 mm. This shutter control has an electronic delay of 0.7 ms after receiving the gate signal and a total shutter transit time of 0.3 ms, corresponding to a combined delay of 1 ms. The exposure time is precisely adjusted by the delay between the gate signals (TTL) to the two shutter controllers. The shutter has a maximum operating frequency of 400 Hz, and sub-millisecond exposure time is obtained by tuning the delay between the two gate signals. This scheme provides a nearly rectangular intensity profile down to 300 µs and even shorter exposure times (triangular) in stroboscopic time-resolved experiments. However, the laser shutter (LS2) does not completely attenuate the X-ray beam at 12.4 keV. Therefore, a lead spot of size ∼2 mm and thickness 0.2 mm is affixed at the position of the X-ray window on the shutter blade without introducing significant additional mechanical inertia.
2.8. Data acquisition scheme
All detectors are integrated into the beamline control software SPEC (Certified Scientific Software), via LIMA (Library for Image Acquisition) device servers. The LIMA is an ESRF development for the unified control of two-dimensional detectors with the same acquisition, file saving and image processing features (Petitdemange et al., 2018). Acquired data are directly saved by the LIMA running on the detector control Linux workstations to the central data storage. At any given time, a desired combination of the configured detectors can be selected for image acquisition. The synchronization of the detectors and the fast beam shutter and other accessories is achieved by a time-frame-generator (TFG) unit, which is embedded in a compact PCI module, C216, developed by the Digital Electronics laboratory at the ESRF. The C216 offers a high level of flexibility by combining a TFG, multi-channel scaler (MCS) and voltage-to-frequency converter (VFC) in a single module. The MCS allows simultaneous registering of essential parameters such as incident and transmitted sample temperature, pressure, etc. with each two-dimensional image. The TFG can program 4096 frames or 2048 frame pairs with eight independent outputs (TTL). The minimum length of a frame is 100 µs and the maximum is 214 s. The acquisition can be triggered either internally or externally by a TTL signal from the sample environment or user-supplied electronics. The MCS features 16 channels with 32 bit counter depth and the VFC has eight channels. The C216 is used not only for timing and synchronization of detectors and sample environments but also as a counter for normal scans during beamline alignment and sample positioning.
Fig. 4 schematically depicts the data acquisition scheme based on the C216. In an acquisition sequence, the frame pairs can be equally spaced or varied in a geometric progression or divided into subgroups with different exposure times and waiting times. By varying the length of frames in a geometric progression both the exposure time and the between exposures can be adjusted to match the evolution of scattered intensity and the kinetics under investigation. However, the minimum is determined by the readout rate of a given detector. For CCD detectors, multiple dark images of the same exposure time are taken prior to each acquisition sequence. The delay between the two fast beam shutter controllers is adjusted using a delay generator based on a NIM module (OPIOM). The voltage outputs from the FEMTO current amplifiers of different photodiodes are converted to frequency and counted by the MCS. The offset due to is subtracted from the measured counts and then converted to absolute number of photons.
2.9. Data reduction
Two pipelines are available for online data reduction, which can be configured by the same graphical user interface that is accessible from SPEC. Different levels of corrections can be chosen depending on the type of detector (dark image subtraction, flat field division, spatial distortion correction etc.). For better performance, detector data and metadata are independently saved to the central data storage by the LIMA servers and SPEC, respectively. Subsequently, two-dimensional images are normalized to an absolute intensity scale, azimuthally regrouped and averaged to obtain one-dimensional scattering profiles, while the data acquisition can be continued in parallel. The first data reduction pipeline is based on the online SAXS/WAXS data reduction package (SPD) that was developed for the former beamline (Boesecke, 2007). It offers a high level of flexibility and stores two-dimensional data in the ESRF data format (EDF, with binary data and a comprehensive ASCII header) and one-dimensional data directly in ASCII format. The one-dimensional data can be processed easily by users and read by many data analysis packages. The second pipeline is based on Python fast azimuthal integration (pyFAI; Kieffer & Karkoulis, 2013). This approach makes use of parallel GPU computation, which is a significant improvement for the data reduction in measurements involving high time resolution and throughput. Both two-dimensional and one-dimensional data files are saved in HDF5 containers following the NeXus (Könnecke et al., 2015) standard.
The HDF5 data as well as EDF image data and ASCII one-dimensional data can be visualized using the advanced version of the MATLAB-based SAXSutilities package (https://www.saxsutilities.eu). In addition, conversion routines from HDF5 to EDF (for two-dimensional) and ASCII (for one-dimensional data) are included. SAXSutilities enables the display of multiple data sets after background subtraction in different plots used in small-angle scattering. Other features include averaging, rebinning, merging and background subtraction of selected data sets and offline data reduction including error estimation. In addition, routines for masking unusable regions of scattering patterns, determination of the beam center, calibration of the WAXS detector position using PBBA etc. are included in this software.
3. Technical performance
The instrument can be operated in different resolution modes. Fig. 5 illustrates the different features observed depending on the collimation for the same sample in the USAXS range. The sample consisted of a dilute suspension of silica particles (size ∼450 nm and polydispersity ∼0.05) in water. The intensity modulations arise from the form factor of spherical particles. At medium resolution (Fig. 5a), one can observe the oscillations in the form factor over a very large q range, which is important for the analysis of the morphology of the particles (Narayanan, 2014). Alternatively, the low-q region can be measured with high resolution (Fig. 5b), which is important to avoid instrumental smearing effects and to accurately determine the of interactions. Finally, in the coherence mode, one can observe speckles in the scattering pattern (Fig. 5c) and the analysis of the temporal fluctuations of these intensities yields the equilibrium dynamics of the suspension (Sutton, 2008).
Radiation damage is a limiting factor in many soft-matter and biophysical investigations. Since samples have to be studied under stringent thermodynamic or physiological conditions, specific solutions are adapted for each system. To avoid damage, the beam size on the sample, the exposure time and the delay between successive exposures are assessed in each case. The focusing and collimation slits are adjusted such that the beam size is maximized on the sample and minimized on the detector. When a small beam spot is required on the sample, the
exposure time and delay between exposures are optimized to remain below the damage threshold. The specimen is usually displaced between successive exposures, which in the case of fluid samples is effected by pushing the column using a flow-through capillary cell.The performance of the detectors depends on how well they are calibrated. The pixel size and spatial distortion of each detector are calibrated using a copper grid with a regular array of holes precisely separated by 5 mm. The calibration of spatial inhomogeneities (flat-field response) for all detectors is performed using the fluorescence from a 10% HBr solution in a 3 mm capillary excited by an incident energy of 13.5 keV. The corresponding fluorescence emission is at 11.92 keV. To obtain good statistics, typically several hundred thousand counts per pixel are accumulated by multiple exposures. The efficiencies of detectors are estimated using SAXS intensity from deionized water (MilliQ) in the compressibility limit (1.6 × 10−3 mm−1 at 298 K) (Narayanan, 2014).
The spatial resolution of the detector is another important parameter which ultimately determines the resolution of the instrument. Fig. 6 displays the azimuthally averaged USAXS profiles of the silica particle suspension measured with different detectors. These profiles can be modeled by a polydisperse sphere scattering function with mean radius (RMean) 226 ± 2 nm and polydispersity ∼5%. The oscillations in the intensities have the same amplitude, indicating that the detector spatial resolution is not affecting the measurement at this sample–detector distance (30.8 m) for this size and polydispersity of particles. Slight smearing of the first minimum in this case is due to multiple scattering (Semeraro, Möller & Narayanan, 2018) and not a resolution effect. The vertical shifts in intensities manifest the differences in the detection sensitivity, i.e. the number of detector counts in analog–digital units (ADU) produced per photon. The average sensitivity factors at 12.46 keV are 11.1, 2.7 and 0.56 for the FReLoN, Rayonix and Pilatus detectors, respectively. Once divided by these factors, all profiles superimpose as shown in the inset of Fig. 6. These factors indicate that for both CCD detectors the average single-photon level is above the mean pixel noise level, which is about 1.6–1.8 ADU. The Pilatus detector is a true photon-counting detector and the single-photon signal is discriminated from the noise.
3.1. High-brilliance SAXS and WAXS
The main applications of high-brilliance SAXS and WAXS are either probing kinetic processes or detecting weak structural signals. In radiation-sensitive samples, long exposure times cannot be afforded. The illustrates the absolute which is given by the residual dark counts in normalized intensity units, for the Rayonix detector at two extremes of the detector position. These difference dark patterns are measured by blocking all scattered photons on the detector while the intensity monitors are still receiving the indicated The two-dimensional patterns were subsequently normalized and azimuthally averaged using the same procedure as in a SAXS measurement. The scales down with the solid angle subtended by the pixels since each pixel collects more photons for the same noise level. Similarly, it is inversely proportional to the total number of incident photons. The improves with hardware binning as each binned pixel collects more scattered photons for a noise level similar to that of an unbinned pixel, as evident in Fig. 7. Therefore, for the detection of weak signals and radiation-sensitive samples, maximum affordable hardware binning should be used. The corresponding curves of the WAXS detector are not shown since they lie roughly two orders of magnitude lower owing to the larger solid angle subtended. The FReLoN detector with a comparable pixel area (binned) has a lower as a result of the higher sensitivity factor. The traces in Fig. 7 are recorded without a sample and absorption in the sample scales up the proportionally. The scattering by water lies below the at 31 m sample–detector distance. Evidently, for weakly scattering samples where the count rate is not an issue, the Pilatus detector is better than the CCD-based detectors.
is often set by the noise of the detector, and in addition the frame rate in kinetic studies and parasitic background at ultra-low angles. Fig. 73.2. High-resolution SAXS and USAXS
Fig. 6 demonstrated that the effect of detector resolution is not significant for 0.5 µm particles (with 5% polydispersity) at 31 m sample–detector distance. The resolution effect becomes more pronounced at shorter distances with larger particles and lower polydispersities. This is illustrated in Fig. 8 using 2 µm size particles with 2.5% polydispersity. The normalized and background-subtracted I(q) cannot be fully described by a polydisperse sphere scattering function (mean radius = 1010 nm), and convolution by a Gaussian resolution function with FWHM 4 × 10−4 nm−1 is necessary to fit the data. In this way, the effective resolution of the setup can be determined in the absence of multiple scattering.
A comparison of the data acquired with the other two detectors is shown in Fig. 9. Different degrees of smearing can be noticed for the three detectors, which can be described by the sphere scattering function convoluted by a Gaussian resolution function with the FWHM values indicated in the legend. The Pilatus detector has a better resolution than the Rayonix detector despite its much larger pixel sampling size (172 versus 44 µm). The smearing is primarily contributed by the detector PSF, which is nearly a box profile in the case of the Pilatus detector while the PSF of the Rayonix detector has a slowly decaying tail due to the phosphor response and imperfections in the fiber-optic taper (scattering and spatial distortion).
3.3. High resolution and coherence
In order to further improve the resolution, a higher degree of collimation and smaller beam presents the minimum (diffraction-limited) beam sizes measured using the beam viewer along the detector tube for the mirror inclination adjusted to focusing at the shortest and longest distances. In this configuration, the beam is nearly coherent (i.e. the beam is comparable to the transverse coherence length) with a reduced of about 5 × 1010 photons s−1. Scattering patterns recorded with this beam display speckles as depicted in Fig. 5(c).
are required. Fig. 10Further reduction of the beam size can be obtained by offsetting the last secondary and first guard slits in opposite directions while maintaining the aperture above the diffraction limit. A nearly symmetric beam 10 photons s−1. Fig. 11 displays the beam profile recorded at 30.8 m, which has near equal contributions from the detector PSF and the beam size. The FWHM limiting resolution is about 1.5 × 10−4 nm−1, that is a factor of 6 better than an Si 220 crystal-based Bonse–Hart instrument (Sztucki & Narayanan, 2007). In terms of angular resolution, this roughly corresponds to 2 µrad and the corresponding real-space resolution is about 50 µm (Petukhov et al., 2015). Therefore, the actual spatial resolution probed in the scattering volume is now limited by the beam rather than the angular resolution. The beam is highly coherent in this configuration, which is useful for XPCS measurements with strongly scattering samples. The optimum condition for high-resolution USAXS measurements is a beam of 25 µm and angular resolution of 4 µrad, which is comparable to the resolution obtained with an Si 333 crystal.
at the farthest detector position can be obtained in this way. However, the is correspondingly diminished to about 103.4. Improvements with EBS upgrade
Here the expected performance with the ESRF EBS upgrade is briefly summarized. The main improvement with the upgraded storage ring will be in the horizontal direction where the beam size and divergence are expected to decrease by about a factor of 16 and 1.6, respectively. The change in the vertical beam parameters will not be significant. With the reduction of horizontal source size, the depicts the calculated beam sizes for the present source and optics, for the EBS with the current optics, and with the mirror specifications optimized for the EBS properties. The horizontal beam improvement will be complemented by installing a high-performance pixel detector. Therefore, both USAXS and XPCS will greatly benefit from the EBS and the optimized optics. Further improvement of the beam in the future will be possible using a focusing and collimation scheme based on compound refractive lenses (Vaughan et al., 2011).
will be higher by a factor of 25−30 and correspondingly the coherent fraction of the will be roughly increased by a factor of 30. Nevertheless, the power density on the monochromator will remain the same. Fig. 124. Scientific applications
This section describes some representative examples to demonstrate the potential of the instrument. More in-depth studies are reported in the corresponding references cited. One of the applications of a combined USAXS/SAXS instrument is to probe in situ the structural features from micrometre scale to nanometres (Beaucage, 2012). This aspect is illustrated in recent work on the ultrastructure of Escherichia coli bacteria under in vivo conditions (Semeraro et al., 2017). USAXS and SAXS elucidated the structure from the whole cell to the ultrastructure of the cell membrane, providing complementary information to both optical and cryo-electron microscopies. Another notable example is an investigation of the hierarchical structure of microtubes formed by the self-assembly of an archetypical sodium dodecyl sulfate (SDS), and a naturally abundant polysaccharide, β-cyclodextrin (Ouhajji et al., 2017). In this case, the combination of SAXS/USAXS revealed the multiscale structure from their molecular capsids to folded microtubes. Here the molecular capsids organize to create rhombic two-dimensional crystalline bilayers which stack further to form multilamellar structures, and they subsequently fold to microtubes.
The scientific applications also critically depend on the availability of sample environments, and data analysis and modeling capabilities. Therefore, an array of sample environments is provided, which include rapid stopped-flow mixing (Narayanan et al., 2014), a high-resolution rheometer (Panine et al., 2003), and temperature and pressure jumps (Möller, Léonardon et al., 2016). For data evaluation, relatively simple analysis and visualization features are implemented in the SAXSutilities package. More advanced modeling is performed using specialized program suits such as SASfit (Breßler et al., 2015), SasView (Alina et al., 2017) and Irena (Ilavsky et al., 2018).
4.1. Broad-q-range time-resolved studies
One of the advantages of having access to a wide q range is the ability to follow the structural development at multiple length scales. A classic example is the crystallization of a polymer from the melt to form the semi-crystalline mesostructure, which further evolves to microscopic spherulite morphology. Fig. 13 displays the typical evolution of the SAXS and WAXS intensities during the isothermal crystallization of isotactic polypropylene. In this case, SAXS/WAXS and USAXS/WAXS data were acquired from two separate experiments under identical conditions. The crystallinity (deduced from WAXS) and the invariant, which is a measure of the mesostructure (derived from SAXS), follow the behavior as found in previous work (Panine et al., 2008). In addition, the USAXS region shows the formation of a large-scale structure which could be the precursor for the spherulite morphology. Development of this microscopic structure at relatively early stages of crystallization was not evident in the previous study using a Bonse–Hart instrument, presumably because of the higher background and time required to perform a USAXS scan, which was much longer than the times shown in Fig. 13. Qualitatively, these USAXS data may be interpreted in terms of the long-range density fluctuations at the initial stages, which subsequently develop into the well defined surface-fractal-like morphology of spherulites.
A more striking illustration is the recent study of the β-cyclodextrin mentioned above, following a temperature quench from 348 to 298 K (Landman et al., 2018). Time-resolved SAXS/USAXS experiments deciphered the structure development at different size scales and the underlying kinetics. The results allowed the identification of an inward growth mechanism of microtubes, which may be applicable to a variety of similar classes of systems.
of microtubes of SDS and4.2. Time-resolved USAXS
Examples described in the previous section have already illustrated the utility of time-resolved USAXS for probing hierarchical structure development. A key strength of this method is in the investigation of optically opaque systems which cannot be easily observed under an optical microscope or by et al., 2016). For example, the kinetics of in concentrated bovine serum albumin solutions undergoing liquid–liquid induced by trivalent ions was studied. The results revealed an arrested for deep temperature quenches into the liquid–liquid coexistence region (Da Vela et al., 2016).
methods. In many synthetic systems, it may be possible to match the optical by choosing an appropriate solvent, thereby making the system less turbid. This approach is not feasible with protein solutions, most of which are opaque at high concentrations. Therefore, time-resolved USAXS has become a powerful tool for probing the early stages of liquid–liquid in concentrated protein solutions (Da VelaAnother prototypical case is the et al., 2017). Nearly charge neutral percolated clusters formed in the millisecond scale, which coalesced with time to form large agglomerates. At the same time, the morphology of the initially formed clusters became more compact by chain relaxation. In other words, the complexation is a multiscale process and time-resolved USAXS enabled deciphering of the kinetics at different structural levels.
of polyelectrolytes, which occurs instantaneously upon mixing oppositely charged polyelectrolytes, and the solution becomes very turbid. Time-resolved USAXS allowed probing of the millisecond scale assembly processes upon mixing anionic sodium polyacrylate and cationic polyallylamine hydrochloride at different salt (NaCl) concentrations (Takahashi4.3. High-resolution fiber diffraction
The small beam et al., 2018). Fig. 14 shows the background-subtracted ultra-low-angle diffraction pattern from rabbit skeletal muscle (extensor digitorum longus) in the resting state. The diffraction diagram clearly displays the lower-order peaks from the sarcomere which can be used to follow the changes at the cellular level during the contraction. This approach turned out to be a valuable tool for the investigation of the length-dependent activation in cardiac muscle (Reconditi et al., 2017). The results provided compelling evidence for the mechanosensing role of the myosin filament as a second permissive step along the pathway of muscle regulation, supplementing the well established actin-based mechanism. The mechanosensing function facilitates high metabolic efficiency of striated muscle and explains the molecular basis for the Frank–Starling law of the heart (Piazzesi et al., 2018).
high degree of collimation and high-resolution detector greatly benefit fiber diffraction experiments. A well known case is structure–function studies in striated muscle (skeletal and cardiac), which until now have mainly focused on the structural dynamics at the level of filamental proteins myosin and actin. The long fiber–detector distance of 30.8 m permits accessing the structural changes at the sarcomere (muscle cell) level (PiazzesiOne of the important assets for this type of experiments is the ability to investigate the structural changes at the level of filamental proteins as well as the sarcomere in the same preparation. The complementary information thereby derived significantly aids in interpreting the diffraction diagrams. Furthermore, these investigations undoubtedly will benefit from the upcoming EBS and having a high-resolution pixel array detector.
4.4. Ultra-small-angle XPCS
Using the high degree of coherence obtained with tight collimation, XPCS experiments can be performed on samples with sufficient scattering power. USA-XPCS provides an easier access to the equilibrium dynamics in low-viscous media owing to the intrinsic slowing down at low q. Therefore, faster diffusive and advective dynamics in aqueous media can be probed by multispeckle XPCS without the need to manipulate the solvent viscosity (Möller, Chushkin et al., 2016). This feature has already been exploited for the investigation of velocity fluctuations at the early stages of colloidal sedimentation (Möller & Narayanan, 2017), phoretic dynamics of colloids during the of the solvent (Dattani et al., 2017) and active dynamics of Janus colloids (Semeraro, Dattani & Narayanan, 2018). The larger size of speckles at ultra-low angles allowed the use of a Pilatus detector for multispeckle XPCS measurements.
Recently, the performance of multispeckle USA-XPCS was significantly improved by using an Eiger 500k detector, which has a small pixel size (75 µm) and high frame rate (up to 22 kHz) (Zinn et al., 2018). This allows a much broader range of experiments involving sub-millisecond dynamics. When this detector is used in combination with the EBS, it is expected that dynamics in many functional biological systems will become accessible.
5. Summary
Following the upgrade, the ID02 beamline at the ESRF has evolved to a unique multipurpose time-resolved USAXS/SAXS/WAXS (TRUSAXS) instrument. The combined q range covered using 1 Å X-ray wavelength is about 0.001 ≤ q ≤ 50 nm−1, and q resolution down to 1.5 × 10−4 nm−1 can be obtained. Two-dimensional USAXS patterns from q < 0.001 nm−1 can be acquired, and the resulting one-dimensional profiles have superior resolution and statistics as compared to the same data measured with an optimized Bonse–Hart instrument. The overall size range covered by these techniques is from micrometre to ångström scales, and time resolution down to the sub-millisecond range can be achieved. Fig. 15 summarizes the length and time scales accessed by these techniques. In addition, the improved coherence properties of the beam can be exploited for performing USA-XPCS, which probes sub-millisecond dynamics at large length scales.
Despite the advances in different microscopy techniques, scattering methods remain essential for deriving quantitative structural and kinetic information in complex soft matter and biological systems. This impact is more significantly felt in the investigation of transient processes over millisecond time range and at large length scales in optically opaque systems. Further improvement of these methods is expected with the upcoming EBS upgrade, especially time-resolved SAXS and USAXS, and USA-XPCS will greatly benefit from the new source. The main applications of the instrument are in soft-matter and biophysical sciences, but it is also used for a broad range of industrial research and development.
Acknowledgements
The authors profusely express their gratitude to various service groups at the ESRF (Survey and Alignment, Vacuum, Optics, Advanced Modeling and Analysis, Mechanical Engineering, Safety, Building and Infrastructure, Detectors and Electronics, Software, and Purchasing), and in particular to D. Martin, C. Herve, J.-C. Labiche, R. Barrett and P. Mackrill, as well as colleagues from the Institut Laue–Langevin (Grenoble) and the AVS/CADINOX companies (Spain) for the collaboration on the detector tube project. G. Lotze, J. Möller and S. Prevost are thanked for their active participation in the late stage of commissioning.
Funding information
ESRF is acknowledged for the financial support of the project and provision of synchrotron beam time.
References
Alina, G., Butler, P., Cho, J., Doucet, M. & Kienzle, P. (2017). SasView For Small Angle Scattering Analysis, https://www.sasview.org/. Google Scholar
Beaucage, G. (2012). Polymer Science: A Comprehensive Reference, pp. 399–409. Amsterdam: Elsevier. Google Scholar
Boesecke, P. (2007). J. Appl. Cryst. 40, s423–s427. Web of Science CrossRef CAS IUCr Journals Google Scholar
Breßler, I., Kohlbrecher, J. & Thünemann, A. F. (2015). J. Appl. Cryst. 48, 1587–1598. Web of Science CrossRef IUCr Journals Google Scholar
Dattani, R., Semeraro, E. F. & Narayanan, T. (2017). Soft Matter, 13, 2817–2822. Web of Science CrossRef PubMed Google Scholar
Da Vela, S., Braun, M. K., Dörr, A., Greco, A., Möller, J., Fu, Z., Zhang, F. & Schreiber, F. (2016). Soft Matter, 12, 9334–9341. Web of Science CrossRef PubMed Google Scholar
Dinapoli, R., Bergamaschi, A., Henrich, B., Horisberger, R., Johnson, I., Mozzanica, A., Schmid, E., Schmitt, B., Schreiber, A., Shi, X. & Theidel, G. (2011). Nucl. Instrum. Methods Phys. Res. A, 650, 79–83. Web of Science CrossRef CAS Google Scholar
Ilavsky, J., Jemian, P. R., Allen, A. J., Zhang, F., Levine, L. E. & Long, G. G. (2009). J. Appl. Cryst. 42, 469–479. Web of Science CrossRef CAS IUCr Journals Google Scholar
Ilavsky, J., Zhang, F., Andrews, R. N., Kuzmenko, I., Jemian, P. R., Levine, L. E. & Allen, A. J. (2018). J. Appl. Cryst. 51, 867–882. Web of Science CrossRef IUCr Journals Google Scholar
Kieffer, J. & Karkoulis, D. (2013). J. Phys. Conf. Ser. 425, 202012. CrossRef Google Scholar
Kishimoto, H., Shinohara, Y., Suzuki, Y., Takeuchi, A., Yagi, N. & Amemiya, Y. (2014). J. Synchrotron Rad. 21, 1–4. Web of Science CrossRef IUCr Journals Google Scholar
Könnecke, M. et al. (2015). J. Appl. Cryst. 48, 301–305. Web of Science CrossRef IUCr Journals Google Scholar
Labiche, J.-C., Mathon, O., Pascarelli, S., Newton, M. A., Ferre, G. G., Curfs, C., Vaughan, G., Homs, A. & Carreiras, D. F. (2007). Rev. Sci. Instrum. 78, 091301. Web of Science CrossRef PubMed Google Scholar
Landman, J., Ouhajji, S., Prévost, S., Narayanan, T., Groenewold, J., Philipse, A. P., Kegel, W. K. & Petukhov, A. V. (2018). Sci. Adv. 4, eaat1817. Web of Science CrossRef PubMed Google Scholar
Li, Y., Beck, R., Huang, T., Choi, M. C. & Divinagracia, M. (2008). J. Appl. Cryst. 41, 1134–1139. Web of Science CrossRef CAS IUCr Journals Google Scholar
Marion, P. & Zhang, L. (2004). AIP Conf. Proc. 705, 320–323. CrossRef Google Scholar
Martin, T. & Koch, A. (2006). J. Synchrotron Rad. 13, 180–194. Web of Science CrossRef CAS IUCr Journals Google Scholar
Möller, J., Chushkin, Y., Prevost, S. & Narayanan, T. (2016). J. Synchrotron Rad. 23, 929–936. Web of Science CrossRef IUCr Journals Google Scholar
Möller, J., Léonardon, J., Gorini, J., Dattani, R. & Narayanan, T. (2016). Rev. Sci. Instrum. 87, 125116. Web of Science PubMed Google Scholar
Möller, J. & Narayanan, T. (2017). Phys. Rev. Lett. 118, 198001. Web of Science PubMed Google Scholar
Narayanan, T. (2014). Structure from Diffraction Methods, pp. 259–324. Chichester: Wiley. Google Scholar
Narayanan, T., Diat, O. & Bösecke, P. (2001). Nucl. Instrum. Methods Phys. Res. A, 467–468, 1005–1009. Web of Science CrossRef CAS Google Scholar
Narayanan, T., Gummel, J. & Gradzielski, M. (2014). Advances in Planar Lipid Bilayers and Liposomes, Vol. 20, pp. 171–196. Burlington: Elsevier. Google Scholar
Narayanan, T., Wacklin, H., Konovalov, O. & Lund, R. (2017). Crystallogr. Rev. 23, 160–226. Web of Science CrossRef CAS Google Scholar
Ouhajji, S., Landman, J., Prévost, S., Jiang, L., Philipse, A. P. & Petukhov, A. V. (2017). Soft Matter, 13, 2421–2425. Web of Science CrossRef PubMed Google Scholar
Panine, P., Di Cola, E., Sztucki, M. & Narayanan, T. (2008). Polymer, 49, 676–680. Web of Science CrossRef Google Scholar
Panine, P., Gradzielski, M. & Narayanan, T. (2003). Rev. Sci. Instrum. 74, 2451–2455. Web of Science CrossRef CAS Google Scholar
Petitdemange, S., Claustre, L., Homs, A., Papillon, E., Homs, R., Naudet, D., Henry, A., Noureddine, A., Langlois, F. & Mant, G. (2018). 16th International Conference on Accelerator and Large Experimental Control Systems (ICALEPCS'17), Barcelona, Spain, 8–13 October 2017, pp. 886–890. Geneva: JACOW. Google Scholar
Petukhov, A. V., Meijer, J.-M. & Vroege, G. J. (2015). Curr. Opin. Colloid Interface Sci. 20, 272–281. Web of Science CrossRef Google Scholar
Piazzesi, G., Caremani, M., Linari, M., Reconditi, M. & Lombardi, V. (2018). Front. Physiol. 9, 736. Web of Science CrossRef PubMed Google Scholar
Reconditi, M., Caremani, M., Pinzauti, F., Powers, J. D., Narayanan, T., Stienen, G. J., Linari, M., Lombardi, V. & Piazzesi, G. (2017). Proc. Natl Acad. Sci. USA, 114, 3240–3245. Web of Science CrossRef PubMed Google Scholar
Sakurai, S. (2017). Polym. Int. 66, 237–249. Web of Science CrossRef Google Scholar
Sánchez del Río, M. & Dejus, R. J. (2011). Proc. SPIE, 8141, 814115. Google Scholar
Semeraro, E. F., Dattani, R. & Narayanan, T. (2018). J. Chem. Phys. 148, 014904. Web of Science CrossRef PubMed Google Scholar
Semeraro, E. F., Devos, J. M., Porcar, L., Forsyth, V. T. & Narayanan, T. (2017). IUCrJ, 4, 751–757. Web of Science CrossRef PubMed IUCr Journals Google Scholar
Semeraro, E. F., Möller, J. & Narayanan, T. (2018). J. Appl. Cryst. 51, 706–713. Web of Science CrossRef IUCr Journals Google Scholar
Shinohara, Y., Kishimoto, H., Inoue, K., Suzuki, Y., Takeuchi, A., Uesugi, K., Yagi, N., Muraoka, K., Mizoguchi, T. & Amemiya, Y. (2007). J. Appl. Cryst. 40, s397–s401. Web of Science CrossRef CAS IUCr Journals Google Scholar
Sutton, M. (2008). C. R. Phys. 9, 657–667. Web of Science CrossRef CAS Google Scholar
Sztucki, M., Gorini, J., Vassalli, J.-P., Goirand, L., van Vaerenbergh, P. & Narayanan, T. (2008). J. Synchrotron Rad. 15, 341–349. Web of Science CrossRef CAS IUCr Journals Google Scholar
Sztucki, M. & Narayanan, T. (2007). J. Appl. Cryst. 40, s459–s462. Web of Science CrossRef CAS IUCr Journals Google Scholar
Takahashi, R., Narayanan, T. & Sato, T. (2017). J. Phys. Chem. Lett. 8, 737–741. Web of Science CrossRef CAS PubMed Google Scholar
Tuukkanen, A. T., Spilotros, A. & Svergun, D. I. (2017). IUCrJ, 4, 518–528. Web of Science CrossRef PubMed IUCr Journals Google Scholar
Valéry, C., Artzner, F. & Paternostre, M. (2011). Soft Matter, 7, 9583–9594. Google Scholar
Van Vaerenbergh, P., Léonardon, J., Sztucki, M., Boesecke, P., Gorini, J., Claustre, L., Sever, F., Morse, J. & Narayanan, T. (2016). AIP Conf. Proc. 1741, 030034. Google Scholar
Vaughan, G. B. M., Wright, J. P., Bytchkov, A., Rossat, M., Gleyzolle, H., Snigireva, I. & Snigirev, A. (2011). J. Synchrotron Rad. 18, 125–133. Web of Science CrossRef IUCr Journals Google Scholar
Yagi, N. & Inoue, K. (2003). J. Appl. Cryst. 36, 783–786. Web of Science CrossRef CAS IUCr Journals Google Scholar
Zhang, F., Allen, A. J., Levine, L. E., Ilavsky, J., Long, G. G. & Sandy, A. R. (2011). J. Appl. Cryst. 44, 200–212. Web of Science CrossRef CAS IUCr Journals Google Scholar
Zhang, L., del Rio, M. S. & Monaco, G. (2013). J. Phys. Conf. Ser. 425, 052008. CrossRef Google Scholar
Zinn, T., Homs, A., Sharpnack, L., Tinti, G., Fröjdh, E., Douissard, P. A., Kocsis, M., Möller, J., Chushkin, Y. & Narayanan, T. (2018). J. Synchrotron Rad. 25, https://doi.org/10.1107/S1600577518013899. CrossRef IUCr Journals Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.