research papers
Turbo-XAS: dispersive
using sequential acquisitionaESRF, BP 220, 38043 Grenoble, France
*Correspondence e-mail: sakura@esrf.fr
A new experimental technique for time-resolved X-ray absorption studies in the sub-second range has been successfully tested on the dispersive i.e. the absence of movement of the optics and the extremely small and stable horizontal focal spot. In addition, it features simultaneous recording of I0 and I1 and the possibility of performing fluorescence and electron detection.
beamline (ID24) at the ESRF. It consists of a sequential acquisition of energy points using the dispersive optics scheme installed on the beamline. Turbo-XAS takes full advantage of the properties of third-generation radiation sources, overcoming many of the problems encountered in the classical dispersive mode, based on position-sensitive detectors. The new technique benefits from the basic assets of the dispersive set-up,1. Introduction
Time-resolved X-ray absorption spectroscopy using energy-dispersive optics has shown in the past 15 years to be an invaluable tool in the investigation of the electronic and local structure in condensed matter during dynamical processes. Moreover, thanks to the absence of movement of the optics and to the small and stable horizontal focal spot, the dispersive set-up has proved to have unique characteristics for the study of small samples in restrictive environmental cells as well as for the detection of very small signals, like those encountered in X-ray magnetic circular dichroism.
Starting from the pioneering work of Matsushita et al. (Matsushita & Kaminaga, 1980), the parallel energy acquisition mode (dispersive mode, or D-XAS) has been further developed at the different synchrotron radiation facilities (Phizackerley et al., 1983; Flank et al., 1983; Allinson et al., 1988; Hagelstein et al., 1989; D'Acapito et al., 1992; Allen et al., 1992; Lee et al., 1994) and extensively exploited in different scientific domains, from chemistry to biology to solid-state physics (for a review see Fontaine et al., 1989, and references therein).
The dispersive et al., 1997) has been specifically designed to benefit from the high of its undulator source, in order to reach a time resolution in the sub-millisecond range. A position-sensitive detector (PSD), consisting of a scintillator screen coupled to a CCD camera with a fast analog buffer, has been designed in order to reach such a time resolution (Koch et al., 1996). Thanks to the high available, both Bragg and Laue geometries (Hagelstein et al., 1995; San Miguel et al., 1998) for the polychromator crystal can be used, in order to cover the entire energy range from 5 to 27 keV with an optimized energy resolution.
beamline at the ESRF (HagelsteinRecently, a new design for the polychromator crystal has lead to an aberration-free horizontal focal spot (Pellicer Porres, San Miguel & Fontaine, 1998) of ∼20 µm FWHM in a large energy range, opening unique possibilities for performing X-ray absorption studies under extreme conditions of temperature and pressure (Itié et al., 1997; San Miguel et al., 1998; Pellicer Porres, Segura et al., 1998). Moreover, the implementation of quarter-wave plates on dispersive set-ups (Giles et al., 1994) has allowed the recording of extremely accurate XMCD spectra (Pizzini et al., 1998). Very recently the single-bunch filling mode of the machine at the ESRF has been combined with a microcoil to generate a pulsed magnetic field pump phased with respect to the probe to study nanosecond-resolved dispersive XMCD (Bonfim et al., 1998).
These examples show that the advent of third-generation sources has certainly given a new perspective to the field of time-resolved X-ray absorption in the dispersive mode. However, the reduction of source size, the lengthening of beamlines (due to the increase in machine diameters), as well as the increased power load on the optics, have introduced new problems related to the much higher sensitivity to beam instabilities (Pascarelli et al., 1999). With respect to the conventional scanning mode, the dispersive set-up is intrinsically more sensitive to such instabilities. In fact, up to now, no successful attempt to record simultaneously the intensity before (I0) and after (I1) the sample using a parallel acquisition mode has been made, leading to normalization problems in the case of beam instabilities. Moreover, as the spectra are obtained by conversion of a one-dimensional image, they suffer from distortions due to phase contrast phenomena associated with imperfections in all elements in the beam path, and to small-angle scattering from the sample, which strongly affect the quality of the data, particularly on powder samples (Hagelstein et al., 1997, 1998).
Time-resolved X-ray absorption in the sequential mode, using a fast-scanning double-crystal monochromator (Frahm, 1988, 1989; Als-Nielsen et al., 1995) or a piezo-driven vibrating double-crystal monochromator (Bornebusch et al., 1999), leads to a set-up with an intrinsic lower sensitivity to beam instabilities, due to the simultaneous recording of I0 and I1. Also, sequential acquisition enables detection of decay channels, thereby opening the way to time-resolved investigations of the changes in the structure or valence states of a wide variety of samples which are inaccessible for transmission spectroscopy. The time resolution, the position stability of the focal spot and the stability of the energy scale in such methods are, however, limited by the mechanical movement of the two crystals.
We have recently successfully tested on ID24 a new experimental set-up which allows a very good time resolution to be maintained, while offering a correct normalization due to the simultaneous recording of I0 and I1. In turbo-XAS (T-XAS), absorption spectra are recorded using two gas detectors or photodiodes, positioned before and after the focal point, respectively, which measure the signal generated by a monochromatic beam, obtained by scanning a narrow slit through the energy-dispersed fan of radiation immediately after the polychromator. This technique is successful at the ESRF thanks to the small source size, which allows energy resolution to be controlled by the width of the slit.
The simultaneous recording of I0 and I1 is the essential new feature with respect to D-XAS, leading to an experimental set-up which is intrinsically less sensitive to beam instabilities. A good time resolution is maintained, as preliminary spectra have been recorded on timescales of 0.05 (XANES) to 0.5 s (EXAFS). The experimental set-up also profits from the small and stable horizontal focal point and high energy scale stability and reproducibility of the dispersive optics thanks to the decoupling of the monochromatization and of the focusing action. Finally, T-XAS enables detection of decay channels, opening the way to time-resolved studies of surfaces and dilute samples using dispersive optics.
The main applications of the new technique envisaged on ID24 are in the field of dynamical processes on the second timescale, mainly in the transmission mode but also using the reflection mode or fluorescence/electron yield detection. Due to the possibility of obtaining a correct normalization of the data, T-XAS will also be widely employed for the large number of static et al., 1997; Pellicer Porres, Segura et al., 1998) and micro-XANES (Mosbah et al., 1998), or which need a high stability of the energy scale, such as XMCD measurements (Pizzini et al., 1998; Bonfim et al., 1998).
experiments which need dispersive optics because of the small and stable focal spot, such as high-pressure experiments (Itié2. Experimental
Fig. 1(a) shows a schematic diagram of the optics layout of beamline ID24, as well as the experimental set-up required for T-XAS. The beamline is installed on an undulator source on a high-beta section of the ring, and operates in the energy range 5–27 keV. The first mirror (M1) focuses the beam in the vertical plane at the sample position while the second mirror (M2) has the double action of demagnifying the horizontal source and of opening a ∼1 mrad horizontally diverging fan of radiation which impinges on the curved crystal. The crystal is positioned on the θ axis of a θ–2θ spectrometer, with the PSD lying at the extreme of the 2θ axis.
The radius of curvature of the crystal is obtained by the well known relationship of cylindrical optics,
where R is the crystal radius, θ is the and p and q are defined in Fig. 1(a). The change of the along the surface of the crystal, Δθ, opens an energy bandpass, ΔE, given by
where l is the illuminated length of the crystal and P is defined as
The crystal yields a polychromatic focus point, I, at a distance q from the crystal (typically 1–2 m), which is the image of the effective source, Se, at a distance from the crystal of p ≃ 30 m. I is the optimal position for the sample. In the standard dispersive set-up (D-XAS), the PSD converts the `energy–direction' correlation into `energy–position' correlation, and allows for parallel recording of the entire in the energy interval ΔE, in the form of a one-dimensional image.
In T-XAS, a narrow (∼30–50 µm) slit is scanned through the polychromatic fan of radiation downstream of the crystal, selecting a monochromatic beam, the intensity of which is simultaneously recorded before and after the sample by two ionization chambers, as shown in Fig. 1(a). The slit, of width σ, is located at a distance from the sample dslit = xq (0 < x < 1). Energy resolution is determined, as in D-XAS, by four factors: the crystal diffracting planes (which determine the intrinsic energy resolution), the dimensions of the slit (or of the PSD pixels in D-XAS), the source horizontal dimensions, and the crystal penetration depth.
Preliminary tests were directed at first in establishing the adequacy of such a concept, the main concern being the effects of source size on energy resolution.
It is well known (Tolentino et al., 1988) that in D-XAS the contribution to energy resolution due to a finite source size can be minimized by placing the PSD on or close to the Rowland circle of energy E at a distance from the crystal (see Fig. 1b). Note that b ≃ P−1 whenever p >> q, as is the case on ID24 [see equation (3)].
In T-XAS the source size contribution cannot be minimized using this method and can effectively be the limiting factor to energy resolution. In fact, the total angular spread in δθ, relative to the portion of the fan selected by the slit is given by
where δθslit and δθsource are the angular spreads in related to the width of the slit and to the source size, respectively (see Appendix A),
and
α is the fan of rays emitted by the source of size Se and impinging on the crystal at the same Bragg angle,
On ID24, α ≃ 7 µrad and the contribution δθsource is always smaller than the intrinsic angular acceptance (Ω111) of the Si(111) crystal1 thanks to the small size (Se ≃ 200 µm) of the effective source.
In order to have δθ ≃ Ω, the following condition on the ratio σ/x must be respected,
3. Results
Preliminary tests have been performed at the K-edges of Fe, Cu, Ga, Ge and Zr and at the LIII-edges of Pt and Au using Si(111) and Si(311) crystals in the Bragg geometry. The measured on the sample through the slit was ∼109 photons s−1. The current generated in the N2- or Ar-filled ion chambers was of the order of 10−7–10−9 A. Different signal-processing methods were tested. In all cases the current generated by the ion chambers was amplified by fast current amplifiers. The output signal was then either converted by a 1 MHz voltage-to-frequency converter or by a 16 bit ADC. The acquisition of the I0 and I1 signals is performed simultaneously at constant time intervals Δt while the slit is continuously scanned through the beam. Using the ADC it was possible to use Δt = 400 µs and to record good quality XANES and spectra in ∼50 and 500 ms, respectively.
3.1. Energy resolution
The first tests were directed at investigating the achievable energy resolution using this new concept. We have observed that energy resolution can be effectively controlled by varying the width, σ, of the slit. Fig. 2(a) shows a comparison of the Pt LIII recorded with an Si(111) crystal using T-XAS and D-XAS (inset). The slit width was σ = 30 µm while x = 0.5. Energy resolution is indeed comparable in the two set-ups. Fig. 2(b) compares Pt LIII XANES recorded with σ = 30, 50, 70, 100 and 120 µm. Degradation in energy resolution is visible for σ > 50 µm, in agreement with equation (4), where Ω ≃ 25 µrad and P−1 ≃ b ≃ 2.85 m.
3.2. Time resolution
Fig. 3 illustrates results on one of the first applications of T-XAS. This experiment was aimed at providing insight into the evolution of a supercooled liquid as a function of temperature. Taking advantage of the sub-second time resolution, snapshots of the electronic and local atomic structure of undercooled Ge melt have been taken before and during the appearance of the crystallites. Such an investigation is of fundamental interest due to the interplay between the covalent and metallic character of the bonding which characterize the solid and liquid phases of Ge, respectively. Pure Ge was chosen as a model sample; however, this experiment opens the way to the study of the electronic and local structure in metastable phases in semiconductor and metallic alloys. Polycrystalline Ge in powder form was heated to the melting point and then cooled down at different rates in order to observe the supercooled phase. The XANES spectra (Fig. 3a) clearly show a gradual shift towards higher energies and a strong attenuation of the white line, indicating the onset of the liquid metal–semiconductor transition, while the modifications in the oscillations (Fig. 3b) mark the onset of the reconstruction of the tetrahedral network with the formation of crystallites. Such an experiment has also been carried out using D-XAS; however, due to the strong SAXS from the powder particles, which degrades energy resolution (see Fig. 4), all the fine structure features are broadened.
3.3. between scans
Due to the nature of the movement of the slit, which has to return to the initial position in between scans, the
between spectra is approximately 800 ms. By recording scans also during the return trip (two-way scan acquisition), this can be reduced slightly, but not eliminated due to the fact that the slit must stop before inverting its direction. This drawback will be eliminated in the near future by using a rotating disc with an engraved spiral. One revolution of the disc swipes the slit once through the radiation fan. This solution also eliminates the need of backlash corrections.3.4. Reproducibility in slit position/energy
If two-way scans are recorded, a backlash correction of about 10 encoder units2 has to be performed on the encoder position of every second scan. After removing this quantity from every second scan, the overall position reproducibility is ±3 encoder units, which corresponds to δE ≃ ±0.07 eV at 11.1 keV (see Fig. 5). This δE is about one order of magnitude smaller than the global energy resolution.
3.5. Restored sensitivity to dilute samples
Tests were also devoted to assess the feasibility of fluorescence detection on dilute samples. Fig. 6 illustrates an example of the possibilities in the study of dilute samples. It shows a Ge K-edge spectra on a test sample containing 1 wt% of the same Ge powder in a BN matrix, recorded in 2 s. Clearly, self-absorption effects are present, due to the nature of the sample (pure Ge powder of average grain size ∼1 µm). The was measured using an unbiased Si photodiode in the current mode. Due to the relatively low concentration, the measurement in the transmission mode using the CCD camera is prohibitive on this sample.
4. Conclusions
A new simple method of recording time-resolved X-ray absorption spectra has been successfully tested. T-XAS features the basic assets of dispersive
namely the absence of movement of the optics during data acquisition, leading to a high stability in energy scale and focal spot size and position. It also allows for simultaneous acquisition of the incident and transmitted intensity, yielding correct normalization of data, while maintaining a sub-second time resolution. In no way can T-XAS compete with D-XAS concerning time resolution: T-XAS and D-XAS are therefore complementary. This is illustrated with different examples, and lies primarily in the possibility of T-XAS to detect decay channels and to investigate powder samples. T-XAS therefore opens the way to time-resolved investigations of the changes in the structure or valence states of a wide variety of samples which are inaccessible for energy-dispersive spectroscopy in the transmission mode.References
Allen, P. G., Conradson, S. D. & Penner-Hahn, J. E. (1992). Synchrotron Rad. News, 5, 16–24. CrossRef
Allinson, N. M., Baker, G., Greaves, G. N. & Nicoll, K. J. (1988). Nucl. Instrum. Methods, A266, 592–597. CrossRef CAS Web of Science
Als-Nielsen, J., Grubel, G. & Clausen, B. S. (1995). Nucl. Instrum. Methods Phys. Res. B, 97, 522–525. CAS
Bonfim, M., Mackay, K., Pizzini, S., San Miguel, A., Tolentino, H., Giles, C., Neisius, T., Hagelstein, M., Baudelet, F., Malgrange, C. & Fontaine, A. (1998). J. Synchrotron Rad. 5, 750–752. Web of Science CrossRef IUCr Journals
Bornebusch, H., Clausen, B. S., Steffensen, G., Lutzenkirchen-Hecht, D. & Frahm, R. (1999). J. Synchrotron Rad. 6, 209–211. Web of Science CrossRef CAS IUCr Journals
D'Acapito, F., Boscherini, F., Marcelli, A. & Mobilio, S. (1992). Rev. Sci. Instrum. 63, 899–901. CrossRef CAS Web of Science
Flank, A. M., Fontaine, A., Jucha, A., Lemonnier, M., Raoux, D. & Williams, C. (1983). Nucl. Instrum. Methods, 208, 651–654. CrossRef CAS Web of Science
Fontaine, A., Dartyge, E., Itie, J. P., Polian, A., Tolentino, H. & Tourillon, G. (1989). Topics in Current Chemistry 151, pp. 180–203. Berlin: Springer.
Frahm, R. (1988). Nucl. Instrum. Methods, A270, 578–581. CrossRef CAS Web of Science
Frahm, R. (1989). Rev. Sci. Instrum. 60, 2515–2518. CrossRef CAS Web of Science
Giles, C., Malgrange, C., Goulon, J., de Bergevin, F., Vettier, C., Dartyge, E., Fontaine, A., Giorgetti, C. & Pizzini, S. (1994). J. Appl. Cryst. 27, 232–240. CrossRef CAS Web of Science IUCr Journals
Hagelstein, M., Cunis, S., Frahm, R., Niemann, W. & Rabe, P. (1989). Physica B, 158, 324–325. CrossRef CAS Web of Science
Hagelstein, M., Ferrero, C., Hatje, U., Ressler, T. & Metz, W. (1995). J. Synchrotron Rad. 2, 174–180. CrossRef CAS Web of Science IUCr Journals
Hagelstein, M., Lienert, U., Ressler, T., San Miguel, A., Freund, A., Cunis, S., Schulze, C., Fontaine, A. & Hodeau, J. L. (1998). J. Synchrotron Rad. 5, 753–755. Web of Science CrossRef CAS IUCr Journals
Hagelstein, M., San Miguel, A., Fontaine, A. & Goulon, J. (1997). J. Phys. IV, 7, 303–308. CrossRef
Itié, J. P., Gauthier, M., Polian, A. & San Miguel, A. (1997). ESRF Highlights 1996/97. ESRF, Grenoble, France.
Koch, M., Hagelstein, M., San Miguel, A., Fontaine, A. & Ressler, T. (1996). Proc. SPIE, 2416, 85–93. CrossRef
Lee, P. L., Beno, M. A., Jennings, G., Ramanathan, M., Knapp, G. S., Huang, K., Bai, J. & Montano, P. A. (1994). Rev. Sci. Instrum. 65, 1–6. CrossRef CAS Web of Science
Matsushita, T. & Kaminaga, U. (1980). Jpn. J. Appl. Cryst. 13, 465–472. CrossRef CAS IUCr Journals Web of Science
Mosbah, M., Metrich, N., Duraud, J. P., San Miguel, A., Delaney, J. S. & McCammon, C. (1998). Exp. Mineral. Petrol. Geochem. VII Int., April 1998, Orleans, France.
Pascarelli, S., Neisius, T., De Panfilis, S., Bonfim, M., Pizzini, S., Mackay, K., David, S., Fontaine, A., San Miguel, A., Itié, J. P., Gauthier, M. & Polian, A. (1999). J. Synchrotron Rad. 6, 146–148. Web of Science CrossRef IUCr Journals
Pellicer Porres, J., San Miguel, A. & Fontaine, A. (1998). J. Synchrotron Rad. 5, 1250–1257. Web of Science CrossRef IUCr Journals
Pellicer Porres, J., Segura, A., San Miguel, A. & Munoz, V. (1998). Phys. Status Solidi, B211, 389–393.
Phizackerley, R. P., Rek, Z. U., Stephenson, G. B., Conradson, S. D., Hodgson, K. O., Matsushita, T. & Oyanagi, H. (1983). J. Appl. Cryst. 16, 220–232. CrossRef CAS Web of Science IUCr Journals
Pizzini, S., Bonfim, M., Baudelet, F., Tolentino, H., San Miguel, A., MacKay, K., Malgrange, C., Hagelstein, M. & Fontaine, A. (1998). J. Synchrotron Rad. 5, 1298–1303. Web of Science CrossRef IUCr Journals
San Miguel, A., Hagelstein, M., Borrel, J., Marot, G. & Renier, M. (1998). J. Synchrotron Rad. 5, 1396–1397. Web of Science CrossRef CAS IUCr Journals
Tolentino, H., Dartyge, E., Fontaine, A. & Tourillon, G. (1988). J. Appl. Cryst. 21, 15–21. CrossRef CAS Web of Science IUCr Journals