research papers
Evaluating scintillator performance in time-resolved hard X-ray studies at synchrotron light sources
aInstitute of Shock Physics, Blackett Laboratory, Imperial College London, London, UK, bDiamond Light Source, I12 Joint Engineering, Environmental, Processing (JEEP) Beamline, Didcot, Oxfordshire, UK, and cEuropean Synchrotron Radiation Facility, Grenoble, France
*Correspondence e-mail: d.eakins@imperial.ac.uk
The short
small effective source size and high of synchrotron radiation is ideally suited for probing a wide range of transient deformation processes in materials under extreme conditions. In this paper, the challenges of high-resolution time-resolved indirect X-ray detection are reviewed in the context of dynamic synchrotron experiments. In particular, the discussion is targeted at two-dimensional integrating detector methods, such as those focused on dynamic radiography and diffraction experiments. The response of a scintillator to periodic synchrotron X-ray excitation is modelled and validated against experimental data collected at the Diamond Light Source (DLS) and European Synchrotron Radiation Facility (ESRF). An upper bound on the accessible in a time-resolved experiment for a given bunch separation is calculated for a range of scintillators. New bunch structures are suggested for DLS and ESRF using the highest-performing commercially available crystal LYSO:Ce, allowing time-resolved experiments with an interframe time of 189 ns and a maximum of 98 (6.6 bits).Keywords: dynamic imaging; shock compression; time-resolved; X-ray; scintillator.
1. Introduction
The
of synchrotron radiation enables the study of phenomena across a range of spatial and temporal scales, from diffraction experiments able to probe the dynamics of atomic structure to radiography techniques capable of resolving deformation at the microscopic level. By combining new mesoscopic experimental measurements of material deformation with leading numerical models, better-performing next-generation materials may be designed from the ground up.In recent years, a number of studies (e.g. Luo et al., 2012; Hu et al., 2013; Chen et al., 2014; Eakins & Chapman, 2014; Rack et al., 2014; Kantor et al., 2014; Kareh et al., 2014; Karagadde et al., 2015; Jensen et al., 2015; and references therein) have began to exploit synchrotron radiation to provide some of the first observations of important damage mechanisms at the mesoscale. Radiographic studies report observations of the early stages of buried pore collapse (Eakins & Chapman, 2014), crack pattern formation and instability growth (Luo et al., 2012; Jensen et al., 2015), while diffraction experiments have identified remarkably complex phase behaviour in bismuth (Hu et al., 2013).1 However, several authors highlight the challenges associated with bringing dynamic experiments to synchrotron environments, noting trade-offs between signal-to-noise ratio (SNR), time resolution, interframe time and image ghosting (Luo et al., 2012; Eakins & Chapman, 2014).
As the required temporal resolution of dynamic X-ray experiments approaches microseconds and below, increasingly severe demands are placed on detector technology, often resulting in X-ray detection, not the X-ray source, limiting the temporal and spatial resolution realisable in experiments. The challenges of dynamic time-resolved X-ray detection are exacerbated at higher X-ray energies (20 keV) where indirect detection via scintillating materials is in most cases a necessity (Gruner et al., 2002). It is these higher energies, however, that permit the study of appreciable sample volumes ( several mm3), which in turn allow longer-timescale processes, such as the of new phases, to evolve and be stroboscopically examined. Accordingly, the choice of scintillator is crucially important to the temporal resolution accessible by experiment, and can be regarded as a significant bottleneck in the development of dynamic X-ray techniques.
In this paper, the challenges of time-resolved, hard, indirect X-ray detection on the sub-µs timescale are reviewed in the context of dynamic synchrotron experiments for the first time. A succinct review of scintillators suitable for X-ray detection in these experiments is then presented. As this discussion is tailored towards experiments recording sub-µs two-dimensional datasets (e.g. radiographs or diffraction patterns), it compliments recent reviews focused on less-rapid medical imaging scenarios (Nikl, 2006; Rack et al., 2008; van Loef & Shah, 2014), and does not include photon-counting systems, which perform poorly in response to the large instantaneous fluxes expected in radiography experiments (Hatsui & Graafsma, 2015). Using the reported scintillation decay modes, the scintillator response at synchrotron light sources is modelled and validated against experimental data. The scintillator emission is then modelled for a range of bunch separations, allowing the usefulness of these scintillators to be evaluated for a number of existing bunch modes. New bunch modes are then proposed for dynamic experiments at Diamond Light Source (DLS) and European Synchrotron Radiation Facility (ESRF), optimized for existing detector and scintillator technologies. The X-ray energies of focus here (20 keV) make the conclusions of this paper most applicable to existing third-generation synchrotron light sources. However, it is intended that the discussions presented will be of use for detector development at the soon-to-be-online European XFEL (Roling et al., 2014), LCLS II and MaRIE facilities (Barnes et al., 2014).
Before continuing the discussion on indirect X-ray detectors, advances in time-resolved direct detector technology (Hatsui & Graafsma, 2015) should be noted. The development of detectors such as the Keck-PAD (Koerner & Gruner, 2011), which may resolve X-ray bunches of spacing on the order of ∼150 ns, and projects dedicated to resolving the 4.5 MHz pulse train at the European XFEL (LPD, DSSC, AGIPD) promise to see application in the experiments of interest here. However, their large pixel size (150 µm) and low efficiency for X-rays above ∼20 keV will deliver a limited spatial resolution and SNR, respectively, making these detectors less applicable to mesoscale problems such as spatially resolving grain-scale deformation or local phase transitions.
2. Time-resolved hard X-ray detection: challenges
In this section, the challenges of time-resolved hard X-ray detection in dynamic synchrotron experiments are outlined and justified. Firstly, it is important to clarify the timescales of interest to a growing number of dynamic mechanical experiments at synchrotron light sources. Understanding the response of materials to the transient conditions experienced in the aerospace, advanced manufacturing, nuclear and defence industries, as well as astrophysical environments, requires subjecting samples to severe loading conditions, which may only be briefly supported. In these extreme experiments, samples are under loading for nanoseconds (laser-compression), hundreds of nanoseconds (pulsed power loading), <10 µs (gas-gun loading) and tens of microseconds (quasi-static loading) timescales. During such experiments, several important physical processes such as crack propagation and sub-surface defect generation evolve on timescales governed by material sound speeds, which are of the order of km s−1 (µm ns−1). For processes evolving at 1 km s−1 and an assumed system resolution of 50 µm (a value easily achievable with modern optically coupled systems), exposure times of no more than 50 ns are required if data are to be recorded without detectable motion blur. These short exposure times must be combined with interframe times appropriate to the loading conditions in order to reliably resolve material behaviour. For example, an interframe time of <1 µs is desirable in a gas-gun experiment lasting 5 µs.
Several factors contribute to the exposure time and temporal resolution achievable in dynamic synchrotron experiments:
(i) Synchrotron
and bunch structure.(ii) Detector sensitivity and gating capability.
(iii) Choice of scintillator material.
Across the many synchrotron facilities in operation there is a significant variation in storage ring energy, current and filling patterns, making some facilities far better suited to probing certain sample conditions over others. For example, the standard operating mode at the Advanced Photon Source (APS) (153 ns bunch separation) and the 4- and 16-bunch modes at ESRF (704 ns and 175 ns bunch separation, respectively), in combination with their high storage ring energies (7 GeV and 6 GeV, respectively), are already well suited to probing dynamic experiments with sub-µs temporal resolution. More continuous bunch structures can be modified to be better suited to dynamic experiments through bunch structure shuttering or chopping. A three-wheel chopper design has been demonstrated on ID09B, ESRF, which is able to isolate a single bunch from any of the ESRF bunch structures (Cammarata et al., 2009), and, more recently, a chopper capable of delivering 50 ps, 8 keV pulses with a 1.25 MHz repetition rate has been developed at BESSY II (Förster et al., 2015). More complex shuttering techniques, such as phonon Bragg switching (Bucksbaum & Merlin, 1999), continue to be developed as a method to isolate single X-ray bunches (Gaal et al., 2014). Currently, the engineering and technical challenges, particularly the complications due to a high heat load, associated with bunch structure chopping means they are not yet commonplace on synchrotron beamlines, and thus have not been included in this review. In contrast with dedicated bunch modes or chopping, the BESSY-VSR project (HZB, 2015) intends to provide the option of a high average or 500 MHz pulse trains via new accelerator technology. Due to the amount of ongoing research in this area, a similar review of scintillator performance in the near future when these new techniques are more widely available would be of benefit to the dynamic loading community.
The sensitivity and exposure capabilities of several commercial cameras [e.g. Princeton Instruments PI-MAX4 and PCO Dicam Pro: ∼50% peak (QE)2, ≥2.81 ns exposure time] have been demonstrated to provide sufficient SNR in dynamic synchrotron experiments (Wang et al., 2008; Luo et al., 2012; Eakins & Chapman, 2014). Thus, with a suitable bunch structure and detector, the choice of scintillator material is key to the temporal resolution and signal-to-noise levels accessible in a time-resolved experiment. In this paper, the discussion focuses on the influence of the scintillator and does not analyse the required SNR or in an experiment, as the latter may vary significantly from high-resolution quantitative studies to more forgiving qualitative analyses.
3. Scintillator materials
For data collection on the sub-µs timescale at synchrotron light sources, the scintillator material must meet the following four conditions:3
(1) A high stopping power to absorb an appreciable fraction of the incident X-rays. The experiments of interest here demand high spatial resolution (e.g. 1–100 µm) as well as high temporal resolution. The optical systems required to achieve this level of spatial resolution have depths of fields spanning micrometres to hundreds of micrometres. To retain this spatial resolution with the use of a scintillator the crystal must be no thicker than the system's depth of field, assuming a single-crystal scintillator is used. These crystal thicknesses, however, come at the expense of X-ray absorption thus requiring a high to absorb and convert as many of the X-rays as possible. An attenuation length of <200 µm at 25 keV is taken as a lower bound in this review.
(2) Large light yields to provide acceptable signal-to-noise levels. The required short exposure times (e.g. 50 ns) have the potential to introduce significant noise, meaning a high conversion efficiency is required to maximize SNR. Based on recent leading dynamic synchrotron experiments which recorded data with a single bunch exposure (Luo et al., 2012; Eakins & Chapman, 2014; Rack et al., 2014), an efficiency of 20 photons per absorbed keV is taken as a lower bound on efficiency.
(3) Rapid scintillation decay modes to avoid the accumulation of afterglow due to periodic excitation. The time available to collect the scintillator emission in a single-bunch experiment is limited by the employed bunch structure. For example, a collection time of 704 ns is permissible at ESRF in the 4-bunch mode. Suitable scintillator materials must in general exhibit a primary decay mode with a characteristic time of <100 ns to avoid the build-up of deleterious background intensity between bunches. Otherwise, data will be compromised by motion blur and ghosting artefacts.
(4) Visible, rather than ultraviolet, emission for efficiency and flexibility in optical coupling. The QE of the required detectors drops off significantly for wavelengths <400 nm (∼20% at 350 nm for the Princeton Instruments PI-MAX and PCO Dicam Pro), which poses a large issue with respect to SNR. Furthermore, the transmission and imaging performance of optical relays, required to couple the scintillator emission to the detector, are, in general, severely reduced in the UV due to absorption and limitations, respectively. Therefore, visible scintillator emission ensures superior signal-to-noise levels and imaging performance.
Criteria (1), (2) and (3) combine to allow a sufficient SNR with a single-bunch X-ray exposure, which is typically of the order of 100 ps. This exposure time is, crucially, fast enough to permit high spatial resolution data collection without motion blur in extreme deformation scenarios such as imaging 10 km s−1 processes at 1 µm resolution.
Table 1 lists scintillator materials meeting the criteria above, and the parameters of relevance to time-resolved experiments. Of these materials, only three, YAG:Ce (Ce-doped Y3Al5O12), LuAG:Ce (Ce-doped Lu3Al5O12:Ce) and LYSO:Ce (Ce-doped Lu2–xYxSi2O5:Ce), are available commercially, with the others still under development. LYSO:Ce is the best-performing readily available crystal with a high density and large light yield and rapid (τ = 41 ns) single term.
|
The most promising scintillator crystals under development are the metal-iodides (e.g. YI3, GdI3, LuI3). These crystals exhibit a good (attenuation length of <200 µm for 50 keV), green emission, high efficiency (90–115 photons keV−1) and fast decays (primary decay modes <50 ns). However, it is important to acknowledge the influence of the growth process and scintillator form (ceramic, single crystal, film, columnar) on the performance. Micro-columnar scintillators offer an improved effective efficiency as they channel the isotropically emitted light towards the scintillator rear surface, allowing the possibility of using a slightly thicker crystal while preserving spatial resolution. Single-crystal films allow access to doping concentrations not possible in the Bridgman and Czochralski methods and often display fewer long-lived decay components due to the absence of antisite defects (Martin et al., 2009). However, the robustness of existing single-crystal production methods means that single crystals are still preferred for the hard X-ray studies of interest here where the required thicknesses on the order of 100 µm can be easily achieved via polishing.
4. Model
The temporal evolution of the scintillator emission in response to repetitive X-ray excitation was modelled numerically. The train of X-rays emitted by a synchrotron was modelled in time according to the chosen synchrotron period and bunch mode. Every bunch was assigned to a time, tbunch, assumed to have unit intensity, and given a Gaussian temporal profile, which was chosen to approximate the measured temporal profiles (Wulff et al., 1997, 2007). The emission of intensity from a given scintillator was modelled as a series of processes. For a given scintillator, the number of terms, the decay constants and their relative weighting were obtained from the literature; the constants used are summarized in Table 2. At a time t, the intensity emitted by the scintillator was calculated as the sum of intensity generated by all preceding bunches. Mathematically, this is given by equation (1):
where the sum is over the N () scintillation modes in a crystal, each of which having a weighting constant, Ci (<1), and (ms). The scintillator decay is convolved with Fpulse(t), which accounts for the temporal shape of the synchrotron pulses, and Fexposure(t), which accounts for the finite exposure time used to measure the decay response.
By summing the contributions of previous bunches in this way, the effect of repetitive X-ray excitation was captured; scintillation modes with decay constants significantly longer than the bunch separation have little chance to decay, creating an increasing background. In all cases, the model was run for a sufficiently long time (typically up to 100 synchrotron periods) so that the background reached a steady-state value and, thus, an equilibrium state to compare with the experimental data. Notably, 100 synchrotron periods is of the order of hundreds of microseconds to 1 ms, much less than the opening time of typical fast X-ray shutters (∼50–200 ms), meaning the influence of long-lived decay modes will be significant in shuttered synchrotron experiments. It was assumed that the referenced decay modes represent the scintillation process in the materials independent of supplier, and that the X-ray excitation process studied in this work did not change the overall scintillation modes.
5. Validation of the model
5.1. Experimental method
To validate the modelled scintillator response, a number of experiments were performed on beamline I12, DLS (Drakopoulos et al., 2015), and beamline ID19, ESRF (Weitkamp et al., 2015). At DLS, experiments were performed with the monochromatic beam (55 keV, 0.05% bandwidth). At the time of the experiments, the Diamond Light Source was operating with a reduced bunch current of 234 mA due to a preceding RF-cavity failure. A 686-fill bunch mode was used for the measurements. Per revolution, this bunch mode delivers a train of 686 adjacent bunches of 0.34 mA, separated by 2.0 ns. The remaining 250 buckets are unfilled meaning a 500 ns gap follows the pulse train. The scintillators and thicknesses examined on I12, DLS, are listed in Table 3.
|
The scintillator emission was relayed to a PI-MAX4:1024i ICCD camera (Princeton Instruments) via a pair of back-to-back achromatic doublets (AC508-075-A-ML, Thorlabs, fb = 61.7 mm) operating at numerical aperture (NA) = 0.38. The camera was equipped with a Gen III filmless `HBf' photocathode with QE of 35% at 400 nm, 50% at 500 nm and 35% at 700 nm.
The camera was synchronized with the DLS bunch structure via the RF bunch clock, in a similar manner to that described previously (Eakins & Chapman, 2014). Following receipt of the RF bunch clock as a trigger, a decay scan was performed by capturing a series of images (e.g. 1000) with a short (e.g. 5 ns) exposure time and an increasing RF-to-exposure delay. The intensifer gain was set to 100 in order to maximize signal during the short exposures.
A series of decay scans were also performed at ESRF. Scans were performed on beamline ID19, ESRF, using the 4-bunch mode (4BM) and 16-bunch mode (16BM). Per revolution, the 4BM delivers four 140 ps bunches of up to 10 mA each separated by 704 ns, and the 16BM delivers sixteen 140 ps bunches of up to 5.6 mA each separated by 175 ns. The scintillators and their thicknesses examined on ID19, ESRF, are listed in Table 4. The scintillator emission was relayed to the same PI-MAX camera, operating with a gain of 100, using back-to-back Hasselblad HC 2.2:100 mm lenses (NA = 0.15).
|
Decay scans were synchronized similarly to those at DLS. The ID19 beam (U17-6c undulator, filtering: 1.4 mm diamond, 2.8 mm aluminium) was used. The single-harmonic undulator, U17-6c, was used to minimize heat load delivered to the scintillators. Heat load was further controlled in the 16BM scans, which employed a 200 ms X-ray shutter. The spectral
delivered by this beam configuration was dominated by a peak at 17.8 keV. All the decay scans in this work used a small beam size, ∼2.8 mm × 2.8 mm, to further reduce heat load to the scintillators.6. Results
Figs. 1 and 2 show, respectively, decay curves collected for the studied scintillators on DLS and ESRF. Details of the individual scans (number of frames, averaging) are noted in the figures. In these figures the intensity values in each scan have been normalized to the maximum value in each scan so that the traces may be compared on the same scale. The performance of each scintillator is summarized by the measured defined as the ratio of the peak intensity to the minimum intensity in the measured response. The measured maxima, minima and values (before normalization) are shown in Appendix A. Because these values are influenced by the experimental conditions (incident choice of lens, etc.) they should not be used as an absolute guide of scintillator performance.4
Early work showed that on the timescales of interest here (tens to hundreds of nanoseconds) the K and D in equation (2):
did not affect the modelled decay response. Furthermore, in some cases, the equilibrium background intensity produced by the model could not reproduce the background observed in experiment. Therefore, the modelled scintillator response was fitted to the experimental decay scan curves by varyingwhere K is a time-independent scaling factor accounting for the efficiency (e.g. incident detector sensitivity) of the experimental system, and D is an additive constant matching the apparent background.
Overlaid on each curve in Figs. 1 and 2 is the scintillator response modelled using equation (2). The magnitude of the experimentally measured background and the inclusion of the constant D is discussed further in §7. The constants K and D were fitted in a least-squares manner with C and τ fixed to the literature values.
7. Discussion
It was expected that the experimentally observed background offset was a result of the repetitive excitation of the scintillator on a timescale quicker than the dominant decay modes. Therefore, crystals such as LYSO:Ce should accrue almost no background between pulses whereas crystals such as LuAG:Ce should develop a large background offset, characterized by a lower −10 this is unlikely to be a dominant factor (Rack et al., 2014). Thirdly, could have contributed to the measured discrepancy. To assess this further, the highest heat load scenario was considered. This was the experiment performed on LYSO:Ce in the 4-bunch mode at ESRF due to the large delivered by ID19 and the absence of an X-ray shutter during these scans. The temperature increase in LYSO:Ce during this scan, accounting for periodic X-ray heating and cooling via scintillator emission and blackbody emission, is calculated to be less than 10 K. At ∼310 K the glow curves of LuAG:Ce (Nikl et al., 2014), LYSO:Ce (Blahuta et al., 2011) and YAG:Ce (Zych et al., 2000) are non-zero suggesting that thermoluminesence could have been a small source of spurious background counts. However, the reduction of due to dopant materials, which are understood to be included in the commercial crystals used here, is also reported (Blahuta et al., 2013). Thus, quantifying the effect of here is challenging. Finally, the significance of slowly decaying scintillation modes has been shown to increase with increased X-ray exposure time (Koch et al., 1999; Martin et al., 2009). As the exposure times in this study ranged from 200 ms at ESRF in the 16BM to minutes at ESRF in the 4BM and DLS in the 686BM, an increased weighting of long-lived components may have contributed to the background seen in experiment but not in the model. Because the relative contributions of these effects could not be retrieved from the experimental data collected to date, curves were fitted with a simple additive constant, rather than any additional time-dependent functions. As such, it is important to note that the modelled data represent a lower/upper bound on the expected background/dynamic range between bunches.
Although this trend was observed, the modelled scintillator response did not reliably match the experimental background, indicating there are additional contributions to the observed background in experiment. Several experimental factors could have contributed to the anomalous background offset recorded in this work. Parasitic light signals or camera errors (thermal sensor drift, electronic offsets) could have contributed to the unexplained offset. Secondly, undesired electrons in adjacent bunches could have been a small source of persistent excitation, although with operating contrast levels of towards 10The discussion here emphasizes the significant impact of long-lived decay modes on decreasing the contrast in scintillator emission at synchrotron light sources. To refine the scintillator recommendations made in Table 1, the response of the recommended crystals was modelled for a range of bunch separations using the constants reported in Table 2. Four crystals (K2LaI5:Ce, Cs2LiLaBrCl:Ce, SrHfO3:Ce and BaHfO3:Ce) were not modelled as the relative contributions of the decay modes are not reported. Fig. 3 shows a plot of % background versus bunch separation for the ten modelled crystals.5 Here, % background is defined as the ratio of the minimum intensity observed in the scan to the maximum intensity. An increase of the background signal translates into a decreased available Fig. 4 shows the modelled as a function of bunch separation.
As suggested by the experimental data in Figs. 1 and 2, the modelled curves show that the absence of dominant long-lived decay constants allows the bunch separation to be brought closer together. The calculations show that LuAG:Ce should be used with caution in time-resolved experiments on the sub-µs timescale as the does not exceed 2 bits until the bunch separation is increased beyond 1000 ns. Dynamic experiments with moving targets using LuAG:Ce are thus expected to encounter significant difficulty in extracting quantitative contrast information (Eakins & Chapman, 2014; Jensen et al., 2015). In contrast, micro-columnar LuI3:Ce and single-crystal LYSO:Ce deliver a of over 16 bits (the limit of the previously identified intensified CCD cameras) at a bunch separation of 456 ns. Overlaid on Fig. 3 are dashed grey lines indicating the bunch separation in several synchrotron bunch modes suited to dynamic experiments. These suggest that to fully exploit the capabilities of the fastest time-resolved synchrotron bunch modes (e.g. the APS standard mode, ESRF 16BM, and PETRA III time-resolved mode) only a very small subset of crystals may be used.
The validated model may be further used to direct bunch structure development. Using the commercially available LYSO:Ce, synchrotrons could reliably employ a bunch separation of 189 ns for 1% background between bunches. At this bunch separation, experiments with LYSO would deliver a maximum et al., 2012; Eakins & Chapman, 2014; Jensen et al., 2015).
of 98 (6.6 bits). With periods of 1873 ns and 2816 ns, this ideal separation may be practically implemented as a `8 × 25-bunch mode' (184 ns separation) and `15-bunch mode' (188 ns separation) at DLS and ESRF, respectively. To maximize in a dynamic experiment without significant motion blur, the `8 × 25-bunch mode' at DLS includes pulse trains each with 25 bunches and duration 50 ns. Notably, the 16-bunch mode (175 ns separation) already implemented at ESRF is close to the proposed separation and should see <2% background between bunches with LYSO:Ce. Furthermore, by modelling the scintillator response in this way the extent of image ghosting can be approximated in a growing number of dynamic radiography experiments (LuoThe inclusion of the offset parameter, D, in the modelled scintillator response suggests additional care should be taken to calibrate sources of background intensity in future experiments. In addition to the elimination of parasitic light signals via improved shielding, and more careful monitoring of sensor fluctuations, can be further suppressed by faster X-ray shuttering, and care should be taken to understand the influence of X-ray exposure on the relative weighting of short- and long-lived decay modes.
While 14 promising crystals have been highlighted in the literature, it is clear that there is still scope for the further via co-doping or alternative growth methods, or the development of entirely new crystals (Derenzo et al., 2011). The enormous light yield of single-crystal LuI3 (van Loef et al., 2008) is compromised by long decay mechanisms arising from the growth process. In contrast, LuI3 grown in the micro-columnar form (Marton et al., 2014) is identified as the most promising crystal in this work, with a single scintillation mode of 28 ns.
of crystal performanceFinally, the suitability of a 189 ns bunch structure separation with commercially available LYSO:Ce suggests that synchrotrons must push to tailor their bunch structures to user needs, and continue to lead the development of bunch chopping options, thereby further increasing user flexibility. Lastly, were camera quantum efficiencies significantly improved in the 300–400 nm range then many more rapidly decaying scintillator materials would become a viable choice, assuming appropriate optical relays are available.
8. Conclusion
In summary, dynamic experiments at synchrotron light sources promise to address numerous outstanding problems in the materials science community. However, the scope of physics accessible to novel radiographic, diffraction and spectroscopic methods is limited by detector technology and, in particular, the choice of scintillator material. Fourteen crystals well suited to the dynamic experiments of interest here were identified, but only three (YAG:Ce, LuAG:Ce and LYSO:Ce) are available commercially.6 Further analysis of the performance of these crystals in experiment and via modelling highlights that LYSO:Ce is the best commercially available choice for time-resolved experiments on the sub-µs timescale, and was the only experimentally studied crystal to effectively resolve the DLS and ESRF bunch structures. Assuming the use of LYSO:Ce, new bunch modes achieving the best balance between and interframe-time were proposed for DLS and ESRF. The modeling framework developed here can be used to direct future dynamic experiments seeking to build on those reported previously (Luo et al., 2012; Chen et al., 2014; Eakins & Chapman, 2014) although an experiment-specific decay scan remains a necessary step.
APPENDIX A
Measured K and D
and fitted values ofMeasured maxima, minima and K and D, are shown in Table 5 (see §6).
values (before normalization), and fitted values of
|
Footnotes
1Due to the short exposure times of these experiments, X-ray induced radiation damage is not significant in comparison with the damage initiated by mechanical loading methods. For example, an upper bound on X-ray heating in 1 cm3 aluminium is calculated, using XOP (del Rio & Dejus, 2011), to be less than 1 K s−1 in response to the white beam on beamline ID19, ESRF. This temperature rise is negligible in comparison with temperature changes of tens to hundreds of Kelvin expected during dynamic loading. X-ray probing can thus be considered a non-destructive tool for time-resolved experiments.
2Quantum efficiency is defined as the ratio of the number of electrons produced in the photocathode for one X-ray photon incident on the photocathode.
3A more general discussion of ideal scintillator properties, which also appeals to manufacturing concerns and higher energy experiments, is presented by Derenzo et al. (2003).
4The experimental data presented in this article are available open access at https://dx.doi.org/10.5281/zenodo.48170 .
5Note that the decay response is heavily dependent on the dopant, not the crystal alone.
6The proprietary nature of CRY-019 and it's similar performance to LYSO:Ce means discussions have focused on the latter.
Acknowledgements
The authors acknowledge the continued support of Imperial College London, EPSRC, STFC and AWE. The authors also extend their thanks to the I12 and ID19 beamline teams for facilitating the beamtime, and the Diamond Light Source for delivering the custom bunch structure.
References
Barnes, C. W., Funk, D. J., Hockaday, M. P., Sarrao, J. L. & Stevens, M. F. (2014). J. Phys. Conf. Ser. 500, 092001. CrossRef Google Scholar
Blahuta, S., Bessiere, A., Viana, B., Dorenbos, P. & Ouspenski, V. (2013). IEEE Trans. Nucl. Sci. 60, 3134–3141. Web of Science CrossRef CAS Google Scholar
Blahuta, S., Bessière, A., Viana, B., Ouspenski, V., Mattmann, E., Lejay, J. & Gourier, D. (2011). Materials, 4, 1224–1237. Web of Science CrossRef CAS Google Scholar
Bucksbaum, P. H. & Merlin, R. (1999). Solid State Commun. 111, 535–539. Web of Science CrossRef CAS Google Scholar
Cammarata, M., Eybert, L., Ewald, F., Reichenbach, W., Wulff, M., Anfinrud, P., Schotte, F., Plech, A., Kong, Q., Lorenc, M., Lindenau, B., Räbiger, J. & Polachowski, S. (2009). Rev. Sci. Instrum. 80, 015101. Web of Science CrossRef PubMed Google Scholar
Chen, W. W., Hudspeth, M. C., Claus, B., Parab, N. D., Black, J. T., Fezzaa, K. & Luo, S. N. (2014). Philos. Trans. R. Soc. A, 372, 20130191. Web of Science CrossRef Google Scholar
Chewpraditkul, W., Swiderski, L., Moszynski, M., Szczesniak, T., Syntfeld-Kazuch, A., Wanarak, C. & Limsuwan, P. (2009). Phys. Status Solidi A, 206, 2599–2605. Web of Science CrossRef CAS Google Scholar
Derenzo, S., Bizarri, G., Borade, R., Bourret-Courchesne, E., Boutchko, R., Canning, A., Chaudhry, A., Eagleman, Y., Gundiah, G., Hanrahan, S., Janecek, M. & Weber, M. (2011). Nucl. Instrum. Methods Phys. Res. A, 652, 247–250. Web of Science CrossRef CAS Google Scholar
Derenzo, S. E., Weber, M. J., Bourret-Courchesne, E. & Klintenberg, M. K. (2003). Nucl. Instrum. Methods Phys. Res. A, 505, 111–117. Web of Science CrossRef CAS Google Scholar
Drakopoulos, M., Connolley, T., Reinhard, C., Atwood, R., Magdysyuk, O., Vo, N., Hart, M., Connor, L., Humphreys, B., Howell, G., Davies, S., Hill, T., Wilkin, G., Pedersen, U., Foster, A., De Maio, N., Basham, M., Yuan, F. & Wanelik, K. (2015). J. Synchrotron Rad. 22, 828–838. Web of Science CrossRef CAS IUCr Journals Google Scholar
Eakins, D. E. & Chapman, D. J. (2014). Rev. Sci. Instrum. 85, 123708. Web of Science CrossRef PubMed Google Scholar
Förster, D. F., Lindenau, B., Leyendecker, M., Janssen, F., Winkler, C., Schumann, F. O., Kirschner, J., Holldack, K. & Föhlisch, A. (2015). Opt. Lett. 40, 2265–2268. Web of Science PubMed Google Scholar
Gaal, P., Schick, D., Herzog, M., Bojahr, A., Shayduk, R., Goldshteyn, J., Leitenberger, W., Vrejoiu, I., Khakhulin, D., Wulff, M. & Bargheer, M. (2014). J. Synchrotron Rad. 21, 380–385. Web of Science CrossRef IUCr Journals Google Scholar
Grezer, A., Zych, E. & Kepinski, L. (2010). Radiat. Meas. 45, 386–388. Web of Science CrossRef CAS Google Scholar
Gruner, S. M., Tate, M. W. & Eikenberry, E. F. (2002). Rev. Sci. Instrum. 73, 2815. Web of Science CrossRef Google Scholar
Hatsui, T. & Graafsma, H. (2015). IUCrJ, 2, 371–383. Web of Science CrossRef CAS PubMed IUCr Journals Google Scholar
Hu, J., Ichiyanagi, K., Doki, T., Goto, A., Eda, T., Norimatsu, K., Harada, S., Horiuchi, D., Kabasawa, Y., Hayashi, S., Uozumi, S., Kawai, N., Nozawa, S., Sato, T., Adachi, S. & Nakamura, K. G. (2013). Appl. Phys. Lett. 103, 161904. Web of Science CrossRef Google Scholar
HZB (2015). BESSY VSR – Variable pulse length Storage Ring – Upgrade for BESSY II. Technical Design Study, June 2015. Helmholtz Zentrum Berlin, Berlin, Germany. Google Scholar
Glodo, J., Higgins, W. M., van Loef, E. & Shah, K. S. (2006). Nucl. Sci. Symp. Conf. Rec. 3, 1574–1577. Google Scholar
Jensen, B. J., Cherne, F. J., Prime, M. B., Fezzaa, K., Iverson, A. J., Carlson, C. A., Yeager, J. D., Ramos, K. J., Hooks, D. E., Cooley, J. C. & Dimonte, G. (2015). J. Appl. Phys. 118, 195903. Web of Science CrossRef Google Scholar
Kantor, I., Labiche, J.-C., Collet, E., Siron, L., Thevenin, J.-J., Ponchut, C., Borrel, J., Mairs, T., Marini, C., Strohm, C., Mathon, O. & Pascarelli, P. (2014). J. Synchrotron Rad. 21, 1240–1246. Web of Science CrossRef CAS IUCr Journals Google Scholar
Karagadde, S., Lee, P. D., Cai, B., Fife, J. L., Azeem, M. A., Kareh, K. M., Puncreobutr, C., Tsivoulas, D., Connolley, T. & Atwood, R. C. (2015). Nat. Commun. 6, 8300. Web of Science CrossRef PubMed Google Scholar
Kareh, K. M., Lee, P. D., Atwood, R. C., Connolley, T. & Gourlay, C. M. (2014). Nat. Commun. 5, 4464. Web of Science CrossRef PubMed Google Scholar
Koch, A., Peyrin, F., Heurtier, P., Ferrand, B., Chambaz, B., Ludwig, W. & Couchaud, M. (1999). Proc. SPIE, 3659, 170–179. CrossRef Google Scholar
Koerner, L. J. & Gruner, S. M. (2011). J. Synchrotron Rad. 18, 157–164. Web of Science CrossRef IUCr Journals Google Scholar
Loef, E. V. D. van, Dorenbos, P., van Eijk, C. W. E., Krämer, K. W. & Güdel, H. U. (2003). Phys. Rev. B, 68, 045108. Google Scholar
Loef, E. V. van, Higgins, W. M., Glodo, J., Brecher, C., Lempicki, A., Venkataramani, V., Moses, W. W., Derenzo, S. E. & Shah, K. S. (2007). IEEE Trans. Nucl. Sci. 54, 741–743. Google Scholar
Loef, E. V. van, Higgins, W. M., Glodo, J., Churilov, A. V. & Shah, K. S. (2008). J. Cryst. Growth, 310, 2090–2093. Google Scholar
Loef, E. V. van & Shah, K. S. (2014). Proc. SPIE, 9214, 92140A. Google Scholar
Luo, S. N., Jensen, B. J., Hooks, D. E., Fezzaa, K., Ramos, K. J., Yeager, J. D., Kwiatkowski, K. & Shimada, T. (2012). Rev. Sci. Instrum. 83, 073903. Web of Science CrossRef PubMed Google Scholar
Mares, J. A., Nikl, M., Beitlerova, A., Horodysky, P., Blazek, K., Bartos, K. & D'Ambrosio, C. (2012). IEEE Trans. Nucl. Sci. 59, 2120–2125. Web of Science CrossRef CAS Google Scholar
Martin, T., Douissard, P.-A., Couchaud, M., Cecilia, A., Baumbach, T., Dupre, K. & Rack, A. (2009). IEEE Trans. Nucl. Sci. 56, 1412–1418. Web of Science CrossRef CAS Google Scholar
Marton, Z., Nagarkar, V. V., Miller, S. R., Brecher, C., Bhandari, H. B., Kenesei, P., Ross, S. K., Almer, J. D. & Singh, B. (2014). J. Phys. Conf. Ser. 493, 012017. CrossRef Google Scholar
Nikl, M. (2006). Meas. Sci. Technol. 17, R37–R54. Web of Science CrossRef CAS Google Scholar
Nikl, M., Kamada, K., Babin, V., Pejchal, J., Pilarova, K., Mihokova, E., Beitlerova, A., Bartosiewicz, K., Kurosawa, S. & Yoshikawa, A. (2014). Cryst. Growth Des. 14, 4827–4833. Web of Science CrossRef CAS Google Scholar
Pidol, L., Kahn-Harari, A., Viana, B., Virey, E., Ferrand, B., Dorenbos, P., de Haas, J. & van Eijk, C. (2004). IEEE Trans. Nucl. Sci. 51, 1084–1087. Web of Science CrossRef CAS Google Scholar
Rack, A., Scheel, M., Hardy, L., Curfs, C., Bonnin, A. & Reichert, H. (2014). J. Synchrotron Rad. 21, 815–818. Web of Science CrossRef CAS IUCr Journals Google Scholar
Rack, A., Zabler, S., Müller, B. R., Riesemeier, H., Weidemann, G., Lange, A., Goebbels, J., Hentschel, M. & Görner, W. (2008). Nucl. Instrum. Methods Phys. Res. A, 586, 327–344. Web of Science CrossRef CAS Google Scholar
Rio, M. S. del & Dejus, R. J. (2011). Proc. SPIE, 8141, 814115. Google Scholar
Roling, S., Appel, K., Braun, S., Buzmakov, A., Chubar, O., Gawlitza, P., Samoylova, L., Siemer, B., Schneidmiller, E., Sinn, H., Siewert, F., Tschentscher, T., Wahlert, F., Wöstmann, M., Yurkov, M. & Zacharias, H. (2014). Proc. SPIE, 9210, 92100B. Google Scholar
Shah, K. S., Cirignano, L., Grazioso, R., Klugerman, M., Bennett, P. R., Gupta, T. K., Moses, W. W., Weber, M. J. & Derenzo, S. E. (2002). IEEE Trans. Nucl. Sci. 49, 1655–1660. Web of Science CrossRef CAS Google Scholar
Shirwadkar, U., Glodo, J., van Loef, E. V., Hawrami, R., Mukhopadhyay, S., Churilov, A., Higgins, W. M. & Shah, K. S. (2011). Nucl. Instrum. Methods Phys. Res. A, 652, 268–270. Web of Science CrossRef CAS Google Scholar
Tyagi, M., Desai, V. V., Singh, A. K., Singh, S. G., Sen, S., Nayak, B. K. & Gadkari, S. C. (2015). Phys. Status Solidi A, 212, 2213–2218. Web of Science CrossRef CAS Google Scholar
Wang, Y., Liu, X., Im, K.-S., Lee, W.-K., Wang, J., Fezzaa, K., Hung, D. L. S. & Winkelman, J. R. (2008). Nat. Phys. 4, 305–309. Web of Science CrossRef CAS Google Scholar
Wei, H., Du, M.-H., Stand, L., Zhao, Z., Shi, H., Zhuravleva, M. & Melcher, C. L. (2015). arXiv:1509.07822. Google Scholar
Weitkamp, T., Tafforeau, P., Boller, E., Cloetens, P., Valade, J.-P., Bernard, P., Peyrin, F., Ludwig, W., Helfen, L., Baruchel, J., Denecke, M. & Walker, C. T. (2015). AIP Conf. Proc. 1221, 33–38. Google Scholar
Wulff, M., Kong, Q., Cammarata, M., Lo Russo, M., Anfinrud, P., Schotte, F., Lorenc, M., Ihee, H., Kim, T. K. & Plech, A. (2007). AIP Conf. Proc. 879, 1187–1194. CrossRef CAS Google Scholar
Wulff, M., Schotte, F., Naylor, G., Bourgeois, D., Moffat, K. & Mourou, G. (1997). Nucl. Instrum. Methods. Phys. Res. A, 398, 69–84. CrossRef CAS Web of Science Google Scholar
Zych, E., Brecher, C. & Glodo, J. (2000). J. Phys. Condens. Matter, 12, 1947–1958. Web of Science CrossRef CAS Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.