XTIP – the world's first beamline dedicated to the synchrotron X-ray scanning tunneling microscopy technique
aAdvanced Photon Source, Argonne National Laboratory, 9700 S Cass Ave, Lemont, IL 60439, USA, and bCenter for Nanoscale Materials, Argonne National Laboratory, 9700 S Cass Ave, Lemont, IL 60439, USA
*Correspondence e-mail: vrose@anl.gov
In recent years, there have been numerous efforts worldwide to develop the synchrotron X-ray scanning tunneling microscopy (SX-STM) technique. Here, the inauguration of XTIP, the world's first beamline fully dedicated to SX-STM, is reported. The XTIP beamline is located at Sector 4 of the Advanced Photon Source at Argonne National Laboratory. It features an insertion device that can provide left- or right-circular as well as horizontal- and vertical-linear polarization. XTIP delivers monochromatic soft X-rays of between 400 and 1900 eV focused into an environmental enclosure that houses the endstation instrument. This article discusses the beamline system design and its performance.
Keywords: synchrotron X-ray scanning tunneling microscopy; circularly polarizing undulator beamlines; soft X-rays.
1. Introduction
For more than three decades, scanning probe microscopes have been an indispensable tool for the direct high-resolution study of surfaces. Scanning tunneling microscopy (STM) can resolve surfaces with atomic resolution but direct elemental determination of materials is not easily accomplished. X-ray microscopies, on the other hand, provide elemental and magnetic selectivity but currently have a limited spatial resolution. Nevertheless, the strength of X-rays resides in their ability to excite core electrons of a specific level by tuning the incident photon energy to their binding energy. Furthermore, quantitative information about magnetic moments can be obtained because X-ray absorption by a magnetic material depends on the helicity of the incident X-rays. The emerging synchrotron X-ray scanning tunneling microscopy (SX-STM) technique combines the high spatial resolution of STM with the information obtained through X-ray/matter interaction. The technique takes advantage of the fact that X-ray excited electrons can modulate the tunneling current in STM leading to structural, elemental, electronic, and magnetic contrast with high spatial resolution (Saito et al., 2006; Chiu et al., 2008; Okuda et al., 2009; Cummings et al., 2012; Onderwaater et al., 2016). A specialized smart tip in close proximity to the sample surface serves as a detector for the X-ray enhanced tunneling current (Rose et al., 2011; Cummings et al., 2017). The tip height of typically less than 1 nm is controlled by an advanced feedback loop (Wang et al., 2013). Elemental sensitivity down to the limit of single atomic height (Shirato et al., 2014; Kersell et al., 2017) as well as magnetic contrast (Rose et al., 2012; DiLullo et al., 2016) have been demonstrated.
The XTIP beamline is the first beamline worldwide fully dedicated to SX-STM. Previously, substantial effort had been dedicated toward the development of the technique at other synchrotron facilities such as the ESRF (Comin, 2007). So far, only one related, but conceptually different, beamline combining X-ray absorption and scanning force microscopy currently operates at the Swiss Light Source (Pilet et al., 2012). The XTIP beamline uses a circularly polarizing undulator (CPU) that can provide left- or right-circular as well as horizontal- and vertical-linear polarization states. The beamline is capable of operating at soft X-ray energies from 400 to 1900 eV and can provide a focused beam down to ∼10 µm × 10 µm into the SX-STM endstation instrument. The spatial resolution in SX-STM is foremost governed by the properties of the tip detector and not the X-ray beam spot size on the sample. Therefore, the goal of XTIP is to provide high with variable (λ/Δλ) at a given wavelength λ into an X-ray spot that is large enough to illuminate the entire surface area that is usually imaged in SX-STM (typically in the nanometer square to few-micrometer square range). A temperature-controlled and soundproofed enclosure has been constructed in order to provide a stable and low-vibration environment for the endstation instrument.
2. X-ray optical layout
The XTIP beamline is located at 4-ID of the Advanced Photon Source. Besides a single optical element positioned after the exit slit, the beamline uses horizontally deflecting grazing-incidence elements, which allow a monochromator to accept the entire vertically coherent portion of the beam. The XTIP branch shares the CPU source with the 4-ID-C branch. The CPU is 2.4 m long and features a period of 12.8 cm with 35 vertical and 36 horizontal poles. Additionally, Sector 4 includes a hard X-ray branch (4-ID-D) with the standard Advanced Photon Source undulator A (Freeland et al., 2002). The canted orientation of the hard X-ray undulator and the CPU provides a 400 µrad splitting of the two beams. Fig. 1 shows a schematic of the optical path of the XTIP beamline. The first two mirrors in the optical path, M0 (plane) and M1 (spherical), are used by both the C branch and XTIP. Also located in the first optical enclosure (FOE) is M2, a horizontally deflecting plane mirror, which when inserted into the beam serves to deflect the soft X-ray beam into the XTIP branch. When the mirror is retracted, the beam proceeds unaltered down the C-branch optical path. A horizontally focusing cylindrical mirror, M3, focuses the beam onto the entrance slit of the spherical grating monochromator (SGM), which in turn focuses the diffracted beam onto a moveable exit slit. An optical chopper located between the SGM and the exit slit allows modulation of the X-ray beam. A pair of mirrors is used to focus the beam vertically and horizontally (M4 and M5, respectively) at the sample position to a 10 µm × 10 µm-sized spot. The M5 mirror and the endstation instrument are located inside of the environmental enclosure.
3. Beamline physical layout
The physical layout of the XTIP beamline is summarized in Table 1. The shared mirrors M0 [distance from source (dfs): 30.23 m] and M1 (dfs: 32.65 m) are described elsewhere (Freeland et al., 2002). Fig. 2(a) shows the side-bounce mirror assembly of the M2 mirror (dfs: 33.65 m), which deflects the beam horizontally, when inserted into the beam, away from the 4-ID-C branch and into XTIP. The assembly provides a vertical travel of ±25 mm to allow for the selection of one of two optical mirror coatings, carbon or platinum. The carbon coating can be used for harmonic rejection of the undulator beam and has higher reflectivity at lower photon energies of between 400 and 1000 eV. Fig. 2(b) displays the M3 assembly (dfs: 39.15 m) that focuses the beam horizontally onto the SGM entrance slit. The M3 mirror is a platinum-coated meridional cylinder with a radius of 311.2 m. Both M2 and M3 utilize a modular mirror holder [Fig. 2(c)]. The mirror mount design features six fixed contact points with six corresponding spring-loaded contacts. The cooling blocks clamp around the water tubing and are spring loaded against the side of the mirror. The mirror mount also acts as a baffle to shield the mirror surface from an ion pump that is mounted at the bottom of the vacuum chamber. In addition, a glow discharge screen is situated at the opening to the to protect the mirror from ion-pump discharges. On both M2 and M3 there are upstream masks to protect the mirror surfaces and electrically isolated downstream masks that are used as a diagnostic when aligning the beam.
|
Fig. 3 depicts a schematic of the M2 assembly. When M2 is inserted, it deflects the beam into the XTIP branch [Fig. 3(b)]. When M2 is retracted, the beam passes into the C branch [Fig. 3(c)]. The hard X-ray beam of the D branch passes through a separate beam pipe outside of the M2 assembly.
The monochromator is a horizontal diffracting SGM (Chen, 1987). It has an included angle of 175.5°, a 3.5 m entrance arm and a nominal 5.5 m exit arm (McNulty et al., 1996). The water-cooled entrance slit (dfs: 42.65 m) is located at a fixed position upstream of the SGM (dfs: 46.15 m). The SGM is operated in the first inside order. The diffracted beam is focused on a movable exit slit adjustable over a 660 mm range (dfs: 52.0 ± 0.33 m). Each of the slits has a phosphorus-coated mask upstream of the slit blades which can be used for steering the beam. The SGM focus location as a function of energy is shown in Fig. 4. The grating chamber houses two holographically recorded ion-beam etched gratings with 114.6 m curvature radius and line densities of 400 and 600 lines mm−1. The 400 lines mm−1 grating with 25 nm-thick Pt coating is optimized at ∼800 eV. The 600 lines mm−1 grating, previously used at beamline 2-ID-B of the Advanced Photon Source (McNulty et al., 1997), is coated with 20 nm Rh and exhibits the highest at ∼1100 eV allowing extension of the operation to higher energies. The gratings are contact cooled with copper foils with each optic being supported in kinematic holders with six adjustment axes. The SX-STM technique requires modulation of the X-ray beam. Therefore, an ultra-high-vacuum compatible optical chopper (dfs: 50.5 m) operating at frequencies of up to 6 kHz is added to the beam path (Chang et al., 2016).
The final focus of ∼10 µm × 10 µm is achieved with a Kirkpatrick–Baez mirror pair, M4 and M5 (Fig. 5). Both mirrors are uncooled and have a platinum coating. The vertical focusing mirror, M4 (dfs: 54.0 m), is a meridional elliptical cylinder and is the only vertically deflecting element in the beamline. It images the source, originating from the center of the CPU, onto the sample position. Since the exit slit has 660 mm of total travel, M5 (dfs: 58.0 m) is a bendable mirror focusing the horizontal exit slit onto the sample. The final focus is located 60.0 m away from the source, at the position of the endstation instrument. The environmental enclosure enhances the temperature stability and reduces noise at the final-focus position. This is achieved through wall panels with acoustical-type solid fill as well as by means of double-pane glass windows. A supply air diffuser is used to minimize air-flow disturbances. The experiments are carried out remotely from inside the adjacent XTIP control room.
An effort was made to place a diagnostic before each component in the optical path. As such, there are diagnostic chambers before M3, the SGM, M4 and M5. These diagnostic chambers include a gold-coated tungsten mesh with 12.5 µm wire diameter on a 64 µm grid, a Ce-doped YAG crystal and, after the monochromator, a photodiode. The M5 diagnostic further houses standard foils for energy calibration. In addition to the diagnostics mentioned above, horizontal and vertical electrically isolated tungsten wire probes are located between M2 and M3. The wire probes can be independently scanned through the beam. This is used to characterize the position, size and intensity of the beam. The vertical wire is also used to determine the relative position of the straight through beam (C branch) and the deflected beam (XTIP) to allow fast and easy switching between the two branches. In addition to knowing the beam profile at the wire location, the wire can be used to cast a shadow down the optical path aiding in the beamline alignment. An image of the wire probes projected on the YAG screen of the M3 diagnostic is shown in Fig. 2(d).
4. Beamline performance
The goal of XTIP is to deliver a monochromatic X-ray beam of ∼10 µm × 10 µm with selectable polarization and high positional stability to the final-focus location. Since its horizontal source is the moving exit slit, special emphasis was devoted to the accuracy and reproducibility of the exit-slit movement. Additionally, in order to guarantee ultimate positional stability of the final-focus position, calculations were carried out to evaluate potential focal-spot broadening for small-range energy scans (30 eV), in which the exit-slit position remains fixed. This energy range covers, for example, spectroscopy scans over the L2 and L3 absorption edges of transition metals. Fig. 6 shows the calculated beam profile at the final focus for photon energies of 500, 530, 770 and 800 eV. These shadow ray-tracing calculations (SHADOW hybrid) (Shi et al., 2014) include diffraction effects from optics apertures and slope errors as listed in Table 1. In the case of 500 and 530 eV the FWHM beam size changes from 8.7 µm × 8.5 µm to 8.5 µm × 8.5 µm. Likewise, for 770 and 800 eV the beam profile changes from 8.1 µm × 8.2 µm to 8.1 µm × 8.1 µm, respectively. Obviously, the spot size is preserved for small-range energy scans demonstrating that the exit-slit location can be fixed if the highest positional stability of the final focal is desired.
The calculated beam stability has been verified experimentally. Fig. 7 shows the beam profile measured at the final focus using a YAG crystal and a high-resolution CCD camera. The final-focus position is very well preserved for small-range energy scans as demonstrated for energy ramps from 500 to 530 eV and 770 to 800 eV using fixed exit-slit positions at the 530 eV and 800 eV SGM focus, respectively. Similar to the final-focus position, the beam spot size is also well preserved. The final focus has been fit to a Gaussian distribution. With an exit-slit size of 25 µm, the measured spot sizes are 12 µm × 10 µm (500 eV), 11 µm × 9 µm (530 eV), 11 µm × 10 µm (770 eV) and 10 µm × 9 µm (800 eV) FWHM. Hence, the experimental spot sizes agree well with the theoretical prediction.
For wider energy scans, the position of the exit slit has to be adjusted to track the focus of the SGM (c.f. Fig. 4). Fig. 8 shows the shift of the center of mass at the final-focus position as a function of energy using the 400 lines mm−1 grating. The curvature of the horizontal-focusing mirror is adjusted to maintain the smallest spot size for each energy. However, the positional of the M4 and M5 focusing mirrors are not utilized. In the horizontal and vertical directions, the final focus continuously shifts over ∼15 µm upwards and 15 µm outwards for the scanned energy range 400–1300 eV. The slightly higher noise observed in the vertical direction can be attributed to insufficient stiffness of the mount supporting the high-resolution camera imaging the beam on the YAG screen. Because it is important in SX-STM to always illuminate the junction between the sample and the tip, a shift larger than half of the final-focus size needs to be corrected. The positional of the M4 and M5 mirrors can be utilized for that purpose. Alternatively, the microscope inside of the SX-STM endstation instrument can be adjusted to follow the final-focus location.
The calculated λ/Δλ and the of the 400 lines mm−1 grating at the final-focus position as a function of photon energy are shown in Fig. 9. In the calculations, the emitted by the insertion device, the reflectivity of all mirrors and the grating efficiency are taken into account. The entrance- and exit-slit sizes are set to the same value and calculations for 10, 20, 30, 50 and 100 µm are presented. In the case of 10 µm entrance and exit slits, a of up to ∼7200 at ∼520 eV can be achieved [Fig. 9(a)]. The then decreases with increasing slit size. The expected at the final-focus position as a function of photon energy is shown in Fig. 9(b). As can been seen in the figure, the covers multiple orders of magnitude, typically from 1010 to 1013 photons s−1, depending on the entrance- and exit-slit settings. In contrast to the the increases with increasing slit size. The is experimentally evaluated using a calibrated photodiode located in the M5 diagnostic assembly [Fig. 9(c)]. The measured is in good agreement with the expected at the M5 diagnostics as demonstrated for the example of 50 µm entrance and exit slits. The experimental reaches ∼2 × 1012 photons s−1 at 800 eV.
Fig. 10 shows the calculated and the of the 600 lines mm−1 grating at the final-focus position. Generally, in comparison with the 400 lines mm−1 grating, higher groove densities, i.e. more lines per millimetre, deliver higher reciprocal dispersion and therefore higher resolution for the same slit settings. However, the coma error starts to dominate the resolution at energies below ∼600 eV. Therefore, a maximal of 7020, similar to the 400 lines mm−1 grating, is obtained, but it is shifted to a higher photon energy of ∼670 eV [Fig. 10(a)]. Likewise, the highest of ∼1011 to 1013 photons s−1 is shifted to ∼1100 eV [Fig. 10(b)]. The 600 lines mm−1 grating slightly underperforms in peak compared with the expectation, as shown in Fig. 10(c). In the case of 50 µm slits, at the M5 diagnostic chamber a of ∼2 × 1012 photons s−1 is observed experimentally at 1100 eV. At the same time, the maximum is shifted to a lower energy than expected. The measured peaks with 2.3 × 1012 photons s−1 at ∼940 eV indicating that the grating does not entirely conform with the specifications listed in the work of McNulty et al. (1997). Therefore, to enhance the at higher energies, a new high-quality 600 lines mm−1 grating is being prepared using new substrates polished together with the current high-quality 400 lines mm−1 grating and will be installed in the near future.
The influence of selecting a specific i.e. energy resolution, can be seen in Fig. 11. It shows spectra of an oxidized Mn metal calibration standard with an area density of 59.6 µg cm−2 ± 5% located in the M5 diagnostics. The spectra were taken with 10 µm and 30 µm slit sizes using the 400 lines mm−1 grating. Besides the main Mn L3 peak at 638.7 eV, there are additional peaks observed at 637.4, 639.8, 640.4 and 642.3 eV. These additional peaks can be assigned to the three oxidation states of manganese, namely M2+, M3+, and M4+ (Anjum et al., 2011; Khan et al., 2014). Obviously, the oxidation-state peaks are much more clearly resolved in the case of a 10 µm exit slit (λ/Δλ ≃ 6400) compared with a 30 µm exit slit (λ/Δλ ≃ 2900). The entrance slit is fixed at 20 µm. Likewise, the FWHM of the main Mn L3 peak decreases from 2.0 to 1.3 eV when using the smaller exit-slit size. It should be noted that the natural width of absorption peaks in solid-state samples is normally larger than the calculated energy resolution of the beamline. A direct experimental evaluation of the energy resolution would therefore require noble-gas resonances. Generally, the increases with reduced slit sizes. However, at the same time the decreases. In practice, by selecting one of the two gratings, the energy, and can be tailored for a specific experiment. Usually, imaging experiments (Shirato et al., 2014) at XTIP require high while spectroscopy measurements (Chang et al., 2018) benefit from high energy resolution.
The XTIP beamline utilizes a CPU with full polarization control providing the ability to study magnetic properties of materials. Fig. 12 shows absorption spectra of an iron thin film measured with left-circular polarized (LCP) and right-circular polarized (RCP) X-rays. The Fe film with a thickness of ∼50 monolayers was deposited on a Au(111) substrate and post-annealed at 740 K for 90 min. The sample preparation and measurements were carried out under ultra-high vacuum conditions. For the measurement, the sample was kept at 90 K and illuminated by a focused soft X-ray beam at a grazing angle of ∼5°. The spectra measured using a specialized smart tip exhibit peaks at 706.8 and 719.9 eV, which are associated with the Fe L3 and Fe L2 absorption edges, respectively. The relative peak intensities change with the X-ray polarization because the X-ray absorption of a magnetic sample is spin dependent. In this case, the X-ray absorption proceeds from the 2p core shell, involving the excitation of electrons from the spin-orbit split 3p3/2 (Fe L3) and 3p1/2 (Fe L2) levels into the 3d shell, whose incomplete occupation is the origin of a The magnetic moments of the Fe film can then in principle be obtained by recording an X-ray magnetic dichroism (XMCD) spectrum, derived from the differences in the absorption intensities of X-rays with right- and left-circular polarization (Stöhr, 1995).
5. Conclusions
A new beamline, XTIP, has been constructed at the Advanced Photon Source to deliver monochromatic soft X-rays between 400 and 1900 eV for the emerging SX-STM technique. The source for this beamline is a CPU, providing full polarization control. A pair of focusing mirrors delivers a focused beam of 10 µm × 10 µm with a variable 12 to 1013 photons s−1 with a of up to ∼7200. The beamline is the world's first dedicated branch for the SX-STM technique. It will offer the scientific user community unique ways to explore nanoscale materials.
in the range of ∼10Acknowledgements
We thank the members of the XTIP advisory committee Thomas Jung, George Srajer, Maria Iavarone and John Freeland for their leadership. Ian McNulty is acknowledged for his help during the design phase of the beamline. We thank David Gagliano and Don Jensen for their generous support during the construction phase. The students Ryan Tumbleson, Shaoze Wang and Keara Ginell are acknowledged for their help with data analysis. We thank Daniel Trainer for the preparation of the iron thin film. Kurt Goetze and Tim Mooney are acknowledged for their help with the beamline control system.
Funding information
This work was performed at the Advanced Photon Source and the Center for Nanoscale Materials, a US Department of Energy Office of Science User Facility under Contract No. DE-AC02-06CH11357.
References
Anjum, G., Kumar, R., Mollah, S., Thakur, P., Gautam, S. & Chae, K. H. (2011). J. Phys. D Appl. Phys. 44, 075403. Web of Science CrossRef Google Scholar
Chang, H., Cummings, M., Shirato, N., Stripe, B., Rosenmann, D., Preissner, C., Freeland, J. W., Kersell, H., Hla, S.-W. & Rose, V. (2016). AIP Conf. Proc. 1696, 020001. Google Scholar
Chang, H., Shirato, N., Zhang, Y., Hoffman, J., Rosenmann, D., Freeland, J. W., Bhattacharya, A., Rose, V. & Hla, S.-W. (2018). Appl. Phys. Lett. 113, 061602. Web of Science CrossRef Google Scholar
Chen, C. T. (1987). Nucl. Instrum. Methods Phys. Res. A, 256, 595–604. CrossRef Web of Science Google Scholar
Chiu, C.-Y., Chan, Y.-L., Hsu, Y. J. & Wei, D. H. (2008). Appl. Phys. Lett. 92, 103101. Web of Science CrossRef Google Scholar
Comin, F. (2007). X-TIP: Nano-scale chemical mapping and surface structural modification by joined use of X-ray microbeams and tip assisted local detection. European Commission CORDIS, STRP 505634-1 X-TIP. Google Scholar
Cummings, M., Shirato, N., Kersell, H., Chang, H., Rosenmann, D., Freeland, J. W., Miller, D., Hla, S. W. & Rose, V. (2017). J. Appl. Phys. 121, 015305. Web of Science CrossRef Google Scholar
Cummings, M. L., Chien, T. Y., Preissner, C., Madhavan, V., Diesing, D., Bode, M., Freeland, J. W. & Rose, V. (2012). Ultramicroscopy, 112, 22–31. Web of Science CrossRef CAS PubMed Google Scholar
DiLullo, A., Shirato, N., Cummings, M., Kersell, H., Chang, H., Rosenmann, D., Miller, D., Freeland, J. W., Hla, S.-W. & Rose, V. (2016). J. Synchrotron Rad. 23, 574–578. Web of Science CrossRef CAS IUCr Journals Google Scholar
Freeland, J. W., Lang, J. C., Srajer, G., Winarski, R., Shu, D. & Mills, D. M. (2002). Rev. Sci. Instrum. 73, 1408–1410. Web of Science CrossRef CAS Google Scholar
Kersell, H., Shirato, N., Cummings, M., Chang, H., Miller, D., Rosenmann, D., Hla, S.-W. & Rose, V. (2017). Appl. Phys. Lett. 111, 103102. Web of Science CrossRef Google Scholar
Khan, M., Suljoti, E., Singh, A., Bonke, S. A., Brandenburg, T., Atak, K., Golnak, R., Spiccia, L. & Aziz, E. F. (2014). J. Mater. Chem. A, 2, 18199–18203. Web of Science CrossRef CAS Google Scholar
McNulty, I., Feng, Y. P., Frigo, S. P. & Mooney, T. M. (1997). Proc. SPIE, 3150, 195–204. CrossRef CAS Google Scholar
McNulty, I., Khounsary, A., Feng, Y. P., Qian, Y., Barraza, J., Benson, C. & Shu, D. (1996). Rev. Sci. Instrum. 67, 3372. CrossRef Google Scholar
Okuda, T., Eguchi, T., Akiyama, K., Harasawa, A., Kinoshita, T., Hasegawa, Y., Kawamori, M., Haruyama, Y. & Matsui, S. (2009). Phys. Rev. Lett. 102, 105503. Web of Science CrossRef PubMed Google Scholar
Onderwaater, W. G., van der Tuijn, P. C., Mom, R. V., van Spronsen, M. A., Roobol, S. B., Saedi, A., Drnec, J., Isern, H., Carla, F., Dufrane, T., Koehler, R., Crama, B., Groot, I. M. N., Felici, R. & Frenken, J. W. M. (2016). Rev. Sci. Instrum. 87, 113705. Web of Science CrossRef PubMed Google Scholar
Pilet, N., Raabe, J., Stevenson, S. E., Romer, S., Bernard, L., McNeill, C. R., Fink, R. H., Hug, H. J. & Quitmann, C. (2012). Nanotechnology, 23, 475708. Web of Science CrossRef PubMed Google Scholar
Rose, V., Chien, T. Y., Freeland, J. W., Rosenmann, D., Hiller, J. & Metlushko, V. (2012). J. Appl. Phys. 111, 07E304. Web of Science CrossRef Google Scholar
Rose, V., Chien, T. Y., Hiller, J., Rosenmann, D. & Winarski, R. P. (2011). Appl. Phys. Lett. 99, 173102. Web of Science CrossRef Google Scholar
Saito, A., Maruyama, J., Manabe, K., Kitamoto, K., Takahashi, K., Takami, K., Yabashi, M., Tanaka, Y., Miwa, D., Ishii, M., Takagi, Y., Akai-Kasaya, M., Shin, S., Ishikawa, T., Kuwahara, Y. & Aono, M. (2006). J. Synchrotron Rad. 13, 216–220. Web of Science CrossRef CAS IUCr Journals Google Scholar
Shi, X., Reininger, R., Sanchez del Rio, M. & Assoufid, L. (2014). J. Synchrotron Rad. 21, 669–678. Web of Science CrossRef IUCr Journals Google Scholar
Shirato, N., Cummings, M., Kersell, H., Li, Y., Stripe, B., Rosenmann, D., Hla, S.-W. & Rose, V. (2014). Nano Lett. 14, 6499–6504. Web of Science CrossRef CAS PubMed Google Scholar
Stöhr, J. (1995). J. Electron Spectrosc. Relat. Phenom. 75, 253–272. Google Scholar
Wang, K., Rosenmann, D., Holt, M., Winarski, R., Hla, S.-W. & Rose, V. (2013). Rev. Sci. Instrum. 84, 063704. Web of Science CrossRef PubMed Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.