ROBL-II at ESRF: a synchrotron toolbox for actinide research
aThe Rossendorf Beamline (BM20), European Synchrotron Radiation Facility, 71 Avenue des Martyrs, 38043 Grenoble, France, bInstitute of Resource Ecology, Helmholtz Zentrum Dresden Rossendorf, Bautzner Landstrasse 400, 01328 Dresden, Germany, cDepartment of Research Technology, Helmholtz Zentrum Dresden Rossendorf, Bautzner Landstrasse 400, 01328 Dresden, Germany, dInstitut für Experimentelle Physik, TU Bergakademie Freiberg, 09596 Freiberg, Germany, eEuropean Synchrotron Radiation Facility, 71 Avenue des Martyrs, 38043 Grenoble, France, and fSwiss Norwegian Beamlines, European Synchrotron Radiation Facility, 71 Avenue des Martyrs, 38043 Grenoble, France
*Correspondence e-mail: scheinost@esrf.fr, a.scheinost@hzdr.de
This paper is dedicated to the memory of Heino Nitsche (1949–2014), who pioneered with ROBL the world's first beamline dedicated to actinide sciences.
ROBL-II provides four different experimental stations to investigate actinide and other alpha- and beta-emitting radionuclides at the new EBS storage ring of ESRF within an energy range of 3 to 35 keV. The L and M edges together with other elements accessible in the 3 to 20 keV energy range. The six-circle heavy duty goniometer of XRD-1 is equipped for both high-resolution powder diffraction as well as surface-sensitive CTR and RAXR techniques. Single crystal diffraction, powder diffraction with high temporal resolution, as well as X-ray tomography experiments can be performed at a Pilatus 2M detector stage (XRD-2). Elaborate radioprotection features enable a safe and easy exchange of samples between the four different stations to allow the combination of several methods for an unprecedented level of information on radioactive samples for both fundamental and applied actinide and environmental research.
station consists of a highly automatized, high sample throughput installation in a glovebox, to measure and conventional XANES of samples routinely at temperatures down to 10 K, and with a in the sub-p.p.m. range. The station with its five bent-crystal analyzer, Johann-type setup with Rowland circles of 1.0 and 0.5 m radii provides high-energy resolution fluorescence detection (HERFD) for XANES, and RIXS measurements, covering both actinideKeywords: actinides; EXAFS; XANES; HERFD-XANES; XAS; XES; RIXS; XRD; CTR; RAXR; surface diffraction.
1. Introduction
The Rossendorf beamline has been operated at the ESRF since 1996 by Helmholtz-Zentrum Dresden-Rossendorf (HZDR) (Nitsche, 1995; Matz et al., 1999; Reich et al., 2000) and has been redesigned in 2016–2020 (ROBL-II). Embedded within the Helmholtz Association's program on nuclear safety, ROBL-II is dedicated to research on actinides and other elements with no stable isotopes (Tc, Po, Ra). The beamline hosts four major experimental stations for synchrotron X-ray experiments in two hutches (RCH-1 and RCH-2) (Fig. 1):
(1)
station with fluorescence and transmission detection for X-ray absorption fine-structure (XAFS) spectroscopy, including (conventional) X-ray absorption near-edge structure (XANES) and extended X-ray absorption fine-structure (EXAFS) spectroscopies.(2)
station with a five-crystal Johann-type spectrometer for high-energy-resolution fluorescence-detection X-ray absorption near-edge spectroscopy (HERFD-XANES), and resonant inelastic X-ray scattering (RIXS) measurements.(3) XRD-1 station with a heavy-duty, Eulerian cradle, six-circle goniometer for (high-resolution) powder X-ray diffraction (PXRD), surface-sensitive crystal truncation rod (CTR) and resonant anomalous X-ray reflectivity (RAXR) measurements.
(4) XRD-2 station with a Pilatus3 x2M detector stage for single crystal X-ray diffraction (SCXRD) and in situ/in operando PXRD measurements.
2. Optics and control system
ROBL-II receives synchrotron beam from a short bending magnet (SBM) with 0.86 T of the new ESRF storage ring, a 7BA lattice operating at 200 mA and 6 GeV. This SBM provides ROBL-II with a spectral 13 photons s−1 (0.1% bandwidth)−1 at 18 keV, i.e. with about the same as the 0.85 T bending magnet of the previous storage ring of the ESRF, but with almost an order of magnitude higher due to the smaller source size. The synchrotron beam is conditioned by the multipurpose optics (designed and delivered by FMB Oxford) shown in Fig. 2.
of 1.4 × 10This optics setup consists of a combined double-crystal monochromator (DCM), and a double-multilayer monochromator (DMM). The DCM produces monochromatic X-rays within 3 to 35 keV with energy resolution of a few eV (depending on the employed crystal pair and energy) for spectroscopy at the expense of e.g. using the five-crystal spectrometer of the station. Two mirrors before and after the monochromator collimate and focus the beam, respectively, and suppress higher harmonics. Diagnostic modules (DM) consisting of (removable) fluorescence screens and/or photon-sensitive movable blades and diodes are installed after the portend and after each optic element. Fig. 2 (bottom) shows also the final footprint of photons for the two groups of experimental stations, a large, unfocused beam for and a small, focused beam for the other three stations, as calculated by ray tracing (Shadow-Vui, Oasys Version 1.0) and verified by computationally more expensive wavefront propagation methods. A more detailed description of the optics components is given in the following.
The DMM is used to produce an about 100 times higher in the energy range 8 to 19 keV at the expense of energy resolution (∼100 eV) for scattering and selected spectroscopy applications, where energy resolution is not as critical or is improved by secondary monochromatization,2.1. Monochromator
This unique design (FMB Oxford) combines a double-crystal (DCM) setup with three laterally mounted sets of crystals (which can be exchanged by laterally shifting the UHV vessel on the granite support), with a double-multilayer (DMM) setup with two laterally mounted sets of multilayers (which can be again exchanged by laterally shifting the UHV vessel), which can be exchanged without breaking the vacuum. The upward-bouncing, liquid-notrogen-cooled first crystals, as well as the upward-bouncing, water-cooled first multilayer crystals, are mounted on a common Bragg axis. The exchange from the DCM to the DMM setup is simply effectuated by moving the monochromator vessel upward until the multilayer crystals are in the beam (Fig. 3, right). The DCM setup comprises two Si(111) pairs, with a 30° difference in the crystal orientation towards the incoming beam, for an improved management of so-called glitches, caused by Umweganregung, especially in the energy region of actinide LIII-edges (16 to 20 keV). A third crystal pair uses the improved energy resolution of the Si(311) plane. The DMM setup consists of two pairs of multilayers (ML1 and ML2, produced by AXO), with ML1 covering the energy range 8.5 to 12.6 keV (optimized for 10 keV), and ML2 from 12.6 to 18.7 keV (optimized for 15 keV). The diffracting optics consists of a silicon substrate coated with Mo/B4C multilayers of two different period thicknesses plus a ∼4 nm Si cover layer for protection, which can be rotated within an angle of 0.76 to 1.13° and provide an energy resolution ΔE/E < 1%. ML1 has 190 layers with a period thickness of 2.5 nm providing a reflectivity R of 0.9%; ML2 has 80 layers with a period thickness of 3.7 nm and an R of 2.3%.
2.2. Mirror 1 (collimating)
The first, 1.3-m long, water-cooled mirror (Thales SESO) is set at an upward reflecting angle of 2.5 mrad and longitudinally bent to ∼20 km to collimate the white beam vertically. The silicon body is coated with two 37 mm-wide stripes of Rh and Pt, with an uncoated bare Si surface of 32 mm width in between (Fig. 3, left). The three different reflecting surfaces have been selected to maximize the across the working energy range (3 to 35 keV), while minimizing the contribution of higher-order harmonics (Fig. 3, left). Surface metrology performed by the ESRF metrology laboratory showed a surface (micro-) roughness of 0.4 nm for the 700 Å Rh and Pt coatings, and 0.6 nm for the bare Si surface. Slope errors (r.m.s.) vary between 0.7 and 1.1 µrad depending on stripe and bending radius. The meridional bending acts with excellent linearity between radius and motor steps (R2 > 0.996) with no significant hysteresis between increasing and decreasing the radius.
2.3. Mirror 2 (focusing)
The 1.2 m long, downward-reflecting mirror of monocrystalline silicon (Thales SESO) has two toroidal grooves both with a radius of 61.8 mm, one coated with Rh and the other with Pt, to focus the beam both horizontally and vertically into the and 3). The 38 mm-wide flat section between both toroidal grooves is uncoated, i.e. the bare Si surface can be used to provide a collimated beam of <14 keV for diffraction experiments, where a beam with minimized divergence is required. Surface metrology provided surface (micro-) roughnesses of 0.3 nm for the 600 Å Rh coating and the bare Si surface, while the 1.3 nm for the 600 Å Pt coating is less favorable. Slope errors (r.m.s.) are 0.5, 1.0 and 1.3 µrad for Rh, Si and Pt surfaces, the value slightly above specifications for Pt most likely correlated to the high micro-roughness. Meridional bending behavior measured between 5 and 20 km was as favorable as for mirror 1.
XRD-1 and XRD-2 stations with a spot size of about 20 µm × 70 µm (h × v), while the station is intentionally out of focus to minimize beam damage of samples (beam size on sample up to 10 mm × 2 mm, see Figs. 22.4. Diagnostic modules
Diagnostic and beam-defining modules (FMB Oxford) are installed after the portend (DM 1) and after each optical element (DM 2 to DM 4) (Fig. 2, top). Vertically and horizontally defining slits, consisting of four movable tungsten blades, are installed in DM 1, 2 and 3, and as far as they receive white/pink beam, they are water-cooled (DM 1 and 2).
Fluorescent screens are installed in DM 2, 3 and 4. They consist of a 25 µm-thick Y2O3 coating on a water-cooled oxygen-free high-thermal-conductivity copper (OFHC) block in the case of DM 2, and of a 50 µm-thick Y2O3 coating on an (uncooled) aluminium block in the case of DM 3 and 4. They are introduced into the beam via pneumatic actuators, and are remotely viewed by CCD cameras.
Beam position monitors are installed in DM 1, 2 and 4. In DM 1, a blade beam position is installed for permanent position monitoring of the white beam, consisting of four graphite blades mounted onto two water-cooled OFHC copper blocks. In DM 2, the four slit blades have voltage taps to measure drain current to enable the slits to be used as beam monitors. In DM 4, a profile monitor allows to map the beam vertically and laterally with high resolution, using GaAsP photodiodes (Schottky-type) behind two pinholes of 100 and 300 µm diameter.
2.5. Control system
All motorized components of the optics as well as of the experimental stations described below are driven by IcePAP motor controllers, an ESRF inhouse development (http://www.esrf.eu/Instrumentation/DetectorsAndElectronics/icepap), except for the monochromator unit, where Pmac controllers are used (Delta Tau Data Systems Inc.). The main experiment control system is currently migrated from SPEC (Certified Scientific Software) to BLISS (BeamLine Instrumentation Support Software), an inhouse development of ESRF (https://www.esrf.eu/fr/home/UsersAndScience/support-and-infrastructure/software/bliss---beamline-control-software.html). It comes with a Python-based command line interface, a new scanning engine, a web-based configuration interface, and an online (live) and offline data visualization graphical interface. Fast switching between different optical configurations, including different mirror coatings and shapes and different DCM/DMM modes, can be achieved by a macro, which reads stored motor positions for different optical configurations from a table, and then moves all motor axes incrementally to the selected values. Note that all motors are equipped with encoders to obtain highly reproducible positions.
ROBL-II will also implement the ESRF data policy, which means that all data and metadata collected during an experiment will be stored onsite, searchable and accessible online (with a renewable embargo period of three years, then made public). Details on the data policy can be found here (https://www.esrf.eu/datapolicy).
3. Radiochemical safety system
3.1. Common principles
In order to be able to investigate actinoids and other elements with no stable isotopes (Tc, Po, Ra) at the ESRF, the following basic rules have been established (Funke et al., 2001):
The radioprotection design is optimized for prevalently alpha- and beta-emitting radioisotopes; hence, for each of the above given elements, priority is given to the radioisotopes with the longest half-lifes. Exposure levels outside of the glovebox are limited to <0.5 µSv h−1, in contact with the sample enclosure <15 µSv h−1.
For all samples, which remain inside an underpressurized glovebox during measurements (currently only the ).
station), an upper activity limit of 185 MBq is valid (Table 1
|
Experiments can be conducted with three different types of samples: RAD-0 are non-active samples, which do not require any radioprotection treatment. RAD-1 are samples with activities below the threshold limits established for ESRF beamlines without dedicated e.g. 3.7 MBq for a group 1 radionuclide in solid form, less for liquids or powders (https://www.esrf.eu/Infrastructure/Safety/Experiments/RadioactiveSamples). RAD-2 are samples with activities below the 185 MBq limit established for ROBL. RAD-2 experiments require that samples are constantly maintained during the measurement in an underpressurized glovebox with ventilation, filter and radiation monitoring in place.
equipment, which areOtherwise, common principles for the construction and operation of ROBL-II are the multi-barrier containment for the samples, redundancy and automation of safety installations as well as automation of sample exchanges and measurements.
The hutches RCH-1 and RCH-2 do not only act as radiation shields against the synchrotron radiation (walls, doors, and roof covered by Pb shielding) but are in addition equipped as alpha-radionuclide laboratories by fulfilling specific requirements with respect to tightness and ease of decontamination (see Section 3.2 below). They are both accessed through an entrance lock with hand and foot monitors and body decontamination equipment, which serves also for a safe and easy sample transfer between the experimental stations in RCH-1 and RCH-2.
RCH-1 and RCH-2 are furthermore equipped with a redundant ventilation and filter system to establish a pressure gradient from the outside experimental hall of the ESRF to the hutches and then to gloveboxes in order to confine radionuclides in case of an accidental release from sample holders as close as possible to the source point (see Section 3.3 below).
RCH-1 and RCH-2 are finally equipped with a multistep radiation monitoring system and a radiochemical signal system (RCSS), which registers and stores the data provided by the radiation monitors and ventilation (air below).
pressure, temperature). RCSS automatically creates status reports, warnings and alarms in case of equipment failures and alpha/beta/gamma readings above the threshold, which are displayed at the beamline, in ROBL's main office building, in the office of the hall operator and in the ESRF's main control room (see Section 3.4RAD-2 experiments that require that samples are constantly maintained during the measurement in an underpressurized glovebox can currently only be performed in the
station, where bulk measurements commonly require sample masses and activity levels exceeding those of RAD-1. In contrast, experiments at XRD-1 and XRD-2 can commonly be conducted with a microfocused beam and samples can hence be kept small enough to maintain the activity levels of RAD-1. Therefore, there are currently no gloveboxes installed around these three experiments. For all tasks associated with active sample reception, control and shipment, there is a dedicated glovebox available in RCH-2 with gamma-spectrometer and contamination monitors. Sample confinements and handling procedures are further detailed below.The
station in RCH-1 benefits in addition from a high degree of automation for sample exchange and positioning in the beam under both room-temperature and cryogenic conditions.3.2. Experimental hutches
The hutches RCH-1 and RCH-2 (Innospec) are part of the multi-barrier design concept, separating the hutch airspace from the experimental hall of the ESRF. As such, the walls and roof, as well as all openings like doors, windows (including alignment windows) and chicanes for ventilation, fluids and electrical cabling, do not only form a shielding for the synchrotron radiation but are also sufficiently air tight to maintain an underpressure of 50 to 100 Pa against the outside (leakage rate less than 1% of respective hutch volume). This is obtained by covering the walls and the roof with 1 mm-thick steel panels as inner lining, which are sealed by silicon or self-adhesive sealing tape. The removable roof elements and the doors are air-tightened with self-sealing rubber gaskets. The tightness of the fluid and cabling chicanes is realized by RoxTec (certified air and water tightness and fire resistance) (https://www.roxtec.com). The panels are polyester-coated for easy decontamination. All equipment inside the hutches is placed on 25 mm-thick aluminium or stainless steel plates. The remaining floor area is covered by an Epoxy liner, with air- and water-tight sealing against the floor plates and the wall lining.
3.3. Ventilation system
The ventilation system (see schematic view in Fig. 4) is designed to protect personnel and users working at the beamline, at the ESRF and on the EPN campus from a potential caused by a failure of the various sample enclosures during and measurements. The ventilation system has therefore several tasks: (1) Produce a pressure gradient from the experimental hall (0 Pa) to the interior of the two hutches RCH-1 and RCH-2 (−5 Pa) and then to the gloveboxes (−200 Pa), in order to confine a potential contamination first to the gloveboxes and then to the hutches. (2) Keep the exhaust air contamination-free, first by filters at the exit of the gloveboxes in order to prevent a spreading of the contamination as close as possible to the source, second by filters at the exit of the hutches, and third by a filter at the combined exhaust in order to prevent a release of radioactivity into the environment. (3) Monitor a potential release of inside the gloveboxes, i.e. as close as possible to the samples, inside the hutches to protect staff and users during sample exchanges, and finally before the total filter at the common exhaust of both hutches. (4) The two separate air intake units for RCH-1 and RCH-2 filter and temperature-condition the incoming air; that of RCH-2 also dries the incoming air. (5) The ventilation system is redundant: in case of a failure of one of the ventilators or of one of the glovebox and hutch filters, the system switches automatically to a parallel ventilator and filter set in order to maintain the pressure gradients and the safety of users. (6) The ventilation system is connected to an uninterruptable power supply (UPS) in order to ensure that the ventilation works even after a power outage for at least 30 min. (7) In case of fire inside one of the hutches, air intake and exhaust valves are immediately closed by automatic fire flaps. (8) To maintain the pressure gradient between hall and hutches, only one of the three doors, connecting hall and entrance lock, entrance lock and RCH-1, and entrance lock and RCH-2, can be opened at the same time. (9) Inside the hutches, the air is exchanged within the recommended bracket of 6 to 15 changes per hour, namely 750 m3 h−1 for RCH-1 and 1050 m3 h−1 for RCH-2. This is controlled by ultrasonic air flow meters in the exhaust lines.
Fig. 4 shows a simplified scheme of the ventilation systems, the filters and the radiation monitors, e.g. without redundant components.
A UPS provides power to the ventilation system long enough to enable personnel to leave RCH-1 and RCH-2 and to shut down the ventilation system in a controlled way. After a fault of the electrical power network, the UPS will first continue the operation of the ventilation system without the very power-consuming heaters of the air intake CTAs (RCH1 and RCH2). After the temperature of the heater unit has been sufficiently reduced by the ventilated air flow, the CTA ventilators will be stopped and the bypass of the air intake will be opened. The air flow will pass through the CTA without any ventilator running, and the underpressure in RCH1, RCH2 and the gloveboxes will be maintained solely by the air extractors located in the technical room (air extractor of the RCH1, RCH2 rooms, gloveboxes). About 45 minutes after the start of the UPS, the extractors will be stopped and all valves of the extraction network of the hutches will be closed to isolate the hutches from the outside.
3.4. Radiochemical signal system (RCSS)
RCSS supervises the correct functioning of the ventilation system including the integrated monitors and – in case of an eventual failure – produces adequate signals to indicate the kind and seriousness of the failure. In detail, it takes care of the following tasks: (1) Collecting and analyzing the status data from all monitoring systems (normal, maintenance, warning, failure, emergency) (Fig. 5, top). (2) Displaying these status signals in the experimental hutches and the control room of ROBL as well as in the control room of the ESRF (Fig. 5, bottom). (3) Logging and storing the data in order to visualize trends for maintenance purposes.
The following readings are displayed by RCSS (Fig. 5, bottom). AIR: final air exhaust monitoring system. AIR-MON α/β: radiation monitoring system. GAS: P10 counting gas supply of the AIR-MON α/β monitor. Havary, Failure, Warning, Maintenance, OK, and UPS Emergency Power: system status. TEST: system test button. QUIT: quit alarm sound button.
In the deactivated state all the displays are shown in gray. When activated, the display fields are displayed in the colors blue (for maintenance), green (for ok), yellow (for warning), orange (for failure), and red (for emergency). In the example shown in Fig. 5, RCH-1 is in green, i.e. OK condition, RCH-2 is in blue, i.e. maintenance condition. The combined exhaust ventilation system is running ok, but the P10 counting gas symbol (GAS) shows a warning (low gas pressure).
Signal changes from green to yellow or orange are accompanied by a short signal tone, those to the red level by a continuous signal tone, which can be stopped only by pushing the QUIT button for at least 3 s. When a system is in MAINTENANCE status, the acoustic signals are suppressed for this system. One can use the TEST button to perform a signal and display test on the respective tablet. During this test, the acoustic signal is activated when the button is actuated and all display areas are displayed in gray.
3.5. Radioprotection procedures during experiments
For the four experiments established inside these two hutches, there are two different operation modes foreseen, either with or without glovebox. In the first mode, the samples are confined by a double-wall container and remain during the experiment inside an underpressurized (−200 Pa) glovebox (Fig. 6, left). In this mode, the full sample inventory of up to 185 MBq can be measured. In the second mode, the samples are confined by a triple-wall container and can be measured without glovebox (Fig. 6, right). In this mode, samples must remain within the radioactivity limits set up for other beamlines at the ESRF (e.g. 3.7 MBq for a group 1 radionuclide in solid form, less for liquids or powders).
A range of double- and triple-confinement sample holders certified by the ESRF Radioprotection Group and the corresponding HZDR radiation protection officer are made available to users. These account for different sample states (liquids, solids, wet pastes), temperatures (cryostat or room temperature) and techniques.
Each experiment is assigned to one of three risk groups, green for no risk, yellow for moderately radioactive samples (typically Tc, Th, U), and red for samples with substantial radiotoxicity (typically transuranium elements). For red experiments, requirements include the 24 h presence of one user at the beamline, and the presence of two persons, one of them a ROBL staff member, for sample exchanges. At the end of each ).
experiment, the samples and experimental stations are checked for contaminations by the ESRF Radioprotection Group, and eventually decontaminated to ensure a contamination-free environment for the following experiments. The RCSS system ensures that ROBL-II users and staff as well as ESRF hall operators are always informed about the correct functioning of all radioprotection systems, and are warned in case of failures or contaminations by the RCSS systems, giving three levels from warning to failure to emergency (Fig. 54. XAFS
X-ray absorption fine-structure (XAFS) spectroscopy, including (conventional) X-ray absorption near-edge structure (XANES) and extended X-ray absorption fine-structure (EXAFS) spectroscopies, has been since the early 1990s one of the most universal methods to study actinide speciation in liquids, solids and at interfaces, with applications including fundamental, coordination, environmental and materials chemistry of actinides (Nitsche, 1995; Denecke, 2006; Geckeis et al., 2013; Maher et al., 2013; Epifano et al., 2019). In fact, the first version of ROBL built in 1997 was dedicated to as a sole method to study actinide chemistry (Reich et al., 2000). While additional methods are now provided as detailed in the following sections, will remain a fundamental method, while we expand its limits by novel technology. Premises for the new station were: (1) keep a relatively large beam footprint on the sample to minimize beam damage of sensitive samples (containing redox sensitive elements as well as water and organics with a tendency to produce radicals) during the often long exposure times of several hours; (2) keep the samples during measurements enclosed in an alpha-glovebox in order to profit of the higher activity limit (185 MBq, RAD-2, see Section 3 above); (3) provide routinely measurements in a closed-cycle He cryostat to block chemical reactions possibly induced by time, exposure to atmospheric oxygen and high (4) minimize user intervention by using automatized sample changes not only for room temperature but also for cryogenic measurements; (5) shift the lower as far as much into the sub-p.p.m. concentration range as is currently achievable.
A large (2 mm × 10 mm) beam is obtained by placing the ). A glovebox has been designed, where the samples are exposed to the monochromatic synchrotron beam passing through a back-side extension with Kapton windows, while ionization and fluorescence detectors are at the outside (Fig. 7).
station far out of focus of the toroidal mirror (Fig. 2The glovebox contains two automated sample stages, one for room-temperature (RT) ). The RT stage consists of a holder for up to eight samples, which can be moved vertically by a motor to bring each sample into the incident beam. The cryogenic sample stage consists of a custom-made closed-cycle He cryostat (CryoVac). Similar to the RT stage, a motorized vertical translation moves a six-sample stage inside the cryostat in order to bring the individual samples into the incident beam. Each of the two stages can be moved manually on a tray system between a front position for sample loading, and the back position for the measurement.
measurements (multi-sample RT holder) and one for cryogenic measurements (multi-sample cryostat) (Fig. 8Finally, two electrically cooled, Ge or Si multi-element solid state detectors are available, with individual elements and CMOS-based charge-sensitive amplifiers named CUBE (XGLab) for increased count-rate throughput, and Falcon-X read-out spectrometers (XIA). The first one is an electrically cooled (Cryo-Pulse 5 Plus) 18-element germanium detector (Ultra-LEGe, GUL0055, Mirion Technologies) with 50 µm-thick elements providing a spectral resolution of ∼150 eV at 100 kcounts s−1. The gain is ∼3.9 mV keV−1, reset time <1 µs, and rise time <40 ns. The second one is a seven-element silicon DRIFT detector (SDD) with 500 µm-thick Si elements and equipped with a 1.5 W pulse tube cooler (PIPS model SXD7X50M-500-CM-CC_01, Mirion Technologies). The best spectral resolution is <130 eV, which raises to <160 eV at 1 Mcount s−1. While the Si detector has a higher performance at energies below 12 keV and the Ge detector at energies above, the Si detector nozzle is arranged off-center so that both nozzles can be mounted as close as possible to each other for simultaneous use. Due to their advanced (CUBE-CMOS) electronics and fast readout electronics (XIA Falcon-X), both detectors can handle about 106 counts element−1 s−1 with reasonable dead-time (<30%), and are hence able to collect spectra of samples with concentrations below 1 p.p.m. in reasonable time.
The high et al., 2005, 2007), iterative transformation factor analysis (ITFA) (Rossberg et al., 2003, 2009; Yalçıntaş et al., 2016), also coupled with Monte Carlo (MC) simulations (Rossberg & Scheinost, 2005; Kirsch et al., 2011), the direct derivation of the radial pair distribution by Landweber iteration (Rossberg & Funke, 2010), and artificial intelligence approaches (Domaschke et al., 2014), in order to derive as much information as possible from this versatile technique. Fig. 9 shows as flagship example the derivation of the structure of Pu(III)-sorbed to magnetite by MC simulations of the Pu EXAFS signal at the magnetite surface [Figs. 9(a) and 9(c)], resulting in the pair distribution function [Fig. 9(b)] and the refined structure of the highly specific, triple-edge sharing surface complex at the oxygen-terminated magnetite {111} face.
performance on the detector side is supplemented by a unique set of inhouse data analysis tools, like wavelet plot analysis using Cauchy or FEFF-derived wavelets (Funke5. XES
X-ray emission spectrometers are used for several techniques, including ; Van Bokhoven & Lamberti, 2016; Glatzel & Bergmann, 2005). Two spectrometers are available at ROBL: a one-crystal and a five-crystal spectrometer. The one-crystal spectrometer is used only for specific experimental arrangements, which require detection of emitted X-rays in particular scattering geometries (Kvashnina & Scheinost, 2016). The five-crystal spectrometer is the principal instrumentation for standard HERFD-XANES, RIXS and RXES experiments (Fig. 10). The spherically bent crystal analyzers, the sample and the detector are mounted in vertical Rowland geometry. Each of the five crystals monochromatizes the fluorescence according to Bragg's law and focuses the reflected monochromatic photons into the detector (Kleymenov et al., 2011; Hazemann et al., 2009; Huotari et al., 2006; Sokaras et al., 2013; Rovezzi et al., 2020; Kavčič et al., 2012). The resolution of the instrument depends on the selected of the analyzer and is thus linked to the selected energy. The best resolution is obtained in the backscattering geometry at Bragg angles close to 90° due to the geometrical effect of the Johann geometry (Kleymenov et al., 2011; Sokaras et al., 2013). The resolution is further improved by a motorized slit (typically 1 mm) in front of the detector. ROBL's five-crystal spectrometer covers the Bragg angles 65–89°. For concentrated samples an (APD) with an area of 10 mm × 10 mm, and for dilute samples a Ketek detector with 10 mm diameter and 450 µm Si thickness are available.
high-energy-resolution fluorescence-detection X-ray absorption near-edge structure spectroscopy (HERFD-XANES), resonant inelastic X-ray scattering (RIXS) and/or resonant inelastic X-ray emission spectroscopy (RXES) (De Groot & Kotani, 2008The five-crystal spectrometer can be operated with crystal analyzers with 0.5 m and 1 m bending radius (Rovezzi et al., 2017). Available spherically bent striped crystal analyzers (∼100 mm diameter) cover most elements of the periodic table in the energy range 3–20 keV (Table 2). Note that the whole spectrometer setup (sample, analyzer crystals and detector) can be placed into a He-filled bag to obtain sufficient also at the lowest energies.
|
Energy scans are based on the ). Each crystal analyzer is mounted on four motorized translation stages: two linear stages for the up/down and horizontal (closer or farther from the sample) motions, a goniometer to focus the fluorescence on to the detector, and a cradle for selection. The detector is mounted on two perpendicular, linear stages, and a goniometer to follow the reflection and to use different detector angles.
and trigonometric equations that follow the Rowland circle geometry (Kvashnina & Scheinost, 2016HERFD-XANES measurements are performed by scanning the incident energy across the PyXES software has been developed, which allows to select the the (K, L, M), the emission line, and the crystal analyzer set, and then based on this input moves all motors positions into position (Fig. 11). Energy ranges for HERFD-XANES, and RIXS are proposed automatically and can then be adjusted by the user. Additionally, batch scripts can be written in order to set up automatic measurements for several samples and/or techniques. The current position of all spectrometer motors and their expected position upon the selection of a particular experiment are displayed.
of the selected element at the maximum of the X-ray emission line. spectra are measured by scanning the emitted energy with the crystal analyzers, while keeping the incident energy fixed. If the incident energy is selected above the X-ray non-resonant is recorded. If the incident energy is selected below or near the RXES or/and RIXS is recorded. In order to minimize the down-time between the different techniques and to increase the user-friendliness of the spectrometer operation in general, theWe demonstrate here the different techniques using ThO2, prepared as a powder sample sealed between two layers of Kapton foil of 25 µm thickness (Amidani et al., 2019). Fig. 12(a) shows the conventional Th L3-edge XANES (collected in normal, total mode without crystal analyzer, `TFY') in comparison with the HERFD-XANES spectrum recorded at the maximum of the Th Lα1 (3d5/2–2p/2) emission line (∼12968 eV) as a function of the incident energy (collected in about 5 min). The emission energy was selected using the (880) reflections of five spherically bent striped Si crystal analyzers with 0.5 m bending radius aligned at 84.4° The intensity was normalized to the incident A combined (incident convoluted with emitted) energy resolution of 2.5 eV was determined by measuring the full width at half-maximum (FWHM) of the elastic peak of tungsten foil. The full RIXS map [Fig. 12 (c)] established by scanning the incident energy at different emission energies was recorded in about 30 min. The shown in Fig. 12(d) was obtained by scanning the X-ray emission energy at fixed incident energy above the Th L3 edge (∼16.5 keV). Following the dipole selection rules, the XANES spectrum arises from electron excitations from the ground 2p level to the unoccupied 6d level [Fig. 12(b)]. The core hole created by this process is very unstable and quickly filled in by an electron from the higher levels. The X-ray photons emitted during the process 3d → 2p are measured by at the Th Lα1 emission line.
6. XRD-1 (six-circle diffractometer)
The XRD-1 endstation with a six-circle diffractometer (Huber Diffraktionstechnik GmbH) is adapted to analyze well defined single reflexes with high resolution, and is hence suited for high-resolution powder diffraction as well as for surface diffraction. The diffractometer is equipped with a rather small detector mounted on a rigid detector arm, which can freely rotate in relation to the sample mount (Fig. 13). The diffraction angle 2θ is defined by the mechanical angle between the detector position and the primary beam. The construction is based on the classical Eulerian cradle geometry. The detector arm can be moved in the horizontal (nu) and in the vertical (del) as well as in any direction combing the horizontal and vertical movements (Fig. 13, right).
Two rotation axes of the sample goniometers (mu and eta) are aligned parallel to the rotation axes of the detector (nu and del) to allow a precise θ/2θ scattering geometry in any spatial direction. The Eulerian cradle with an inner diameter of 500 mm carries the sample support, which consists of an independent phi-rotation axis and an x/y/z table. The axis of the Eulerian cradle (chi) is aligned perpendicular to the two detector axes. The complete diffractometer can be moved horizontally (thoriz) and vertically (tvert) to place the rotation center in the geometrical focus of the X-ray beam. All rotational axes are equipped with encoders.
The incoming X-ray beam is shaped and monitored with the beam conditioning unit, which is placed in front of the diffractometer (Fig. 13, left). Its components are modular and easily exchangeable. The intensity of the incoming X-ray beam is counted by a monitor, which consists of a 45°-tilted, 20 µm Kapton scattering foil and an LaCl3 (FMB Oxford/Cyberstar). The second module allows to reduce the primary intensity by remote-controlled pneumatic insertion of different metal foils. A third module suppresses parasitic scattering with a vacuum-tight slit (IB-C22-HV, JJ-X-ray). The fourth module is a fast shutter (RI Research Instruments GmbH) to protect the final detector from too intense reflections during goniometer and detector movements. Module 5 is an (IC Plus 50, FMB Oxford) to monitor the true incoming for normalization. Module 6 is a pinhole, which cuts off the remaining parasitic beam scattering off the tungsten blades of the slit. The unit is under vacuum except for the last two components.
Two detection modules are available, one for surface diffraction and one for high-resolution powder diffraction (Fig. 14). Both modules use a Pilatus 100k detector with a 450 µm Si sensor (Dectris Ltd). The beam reflected from the sample hits the module for surface diffraction (Fig. 14, top) first through an in-vacuum tube slit (Huber Diffraktionstechnik GmbH) to reduce air scattering after the sample. Another vacuum-tight slit (IB-C22-HV, JJ-X-ray) allows further reduction of the scattering window. Then follows a purpose-built with 30 mm distance between the HV electrodes; this large distance is necessary to illuminate the whole detector area. This is used to align the sample in the direct beam. A pneumatic protection module with two 0.5 mm-thick tungsten blades closes automatically, when the detector arm moves close to the primary beam. It is controlled by SPEC-based software, which allows setting of the limits for closing the tungsten blades as well as deliberately opening them for detector alignment with intensity-reduced primary beam. The module for surface diffraction can be evacuated to <10−3 mbar. SPEC macros to operate the diffractometer for surface diffraction and for data registration have been provided by Peter Eng (GSECARS, 13 IDC, Advanced Photon Source). A long-distance microscope (K2, Infinity Photo-Optical GmbH) at 800 mm distance from the diffractometer center permits optical sample alignment.
While the highest resolution in powder diffraction can be achieved with secondary analyzer crystals (Dejoie et al., 2018; Hodeau et al., 1998), we refrained from this approach, since installation and alignement of the analyzer crystals turned out to be too time consuming for user operation with experiments changing typically from week to week. Instead, as a compromise between short installation time and reasonable resolution, we developed a module for the small Pilatus 100k detector also used for surface diffraction, with a sample–detector distance of 800 mm (Fig. 14, bottom). A detector with smaller pixel size will be procured in the future. The resolution of this setup depends on the used sample capillary diameter, which is typically 300 µm. The capillary is rotated with up to 200 r.p.m. The entrance slit for the powder diffraction module is a vacuum-tight IB-C30-HV model (JJ-X-ray), connected to an anti-scattering tube. The module contains also a pneumatic detector protection unit as described above. The entrance and the exit of the module is sealed by Kapton windows and can be evacuated to <10−3 mbar to reduce scattering from air.
The Pilatus 100k detector has an active area of 33.5 mm × 83.8 mm (height × width) and contains 195 × 487 pixels with a size of 172 µm × 172 µm. A powder diffraction measurement is performed by a stepwise movement of the detector, e.g. with 80 overlapping shots at steps of 0.5° as shown for LaB6 (Fig. 15). The obtained pattern requires a radial integration along the intersections of the Debye–Scherrer cones with the detector surface, which can be done using the pyFAI software (Ashiotis et al., 2015). This process includes corrections for small tilting errors of the detector. The detector calibration requires the measurement of standards (so-called calibrants). An independent determination of the wavelength is necessary due to its strong correlation with the sample–detector distance. As the detector surface of the Pilatus 100k is relatively small, the calibration needs to be done iteratively, starting with the measurement of the Debye–Scherrer cones at small 2θ angles using silver behenate [CH3(CH2)20COO·Ag] as first calibrant. Subsequently, the calibration is extended to detector positions at higher 2θ values using LaB6 as second calibrant. The extracted point of normal incidence (PONI) parameters are required for radial integration of the 2D powder pattern. The PONI parameters are: the sample–detector distance (dist), two orthogonal x- and y-dimensions in the detector plane (poni1, poni2), and three rotation parameters around the vertical axis (rot1), around the horizontal axis (rot2) and around the incoming beam (rot3). The calibration procedure is described in more detail by Kieffer et al. (2020). The obtained resolution of 0.013° is one order of magnitude better than that of typical laboratory diffractometers, and much faster (typically 5 to 10 min for an excellent signal-to-noise ratio). To further improve the resolution, we consider the acquisition of a 2D detector with smaller pixel size than that provided by the Pilatus 100k.
Interface specific X-ray diffraction techniques are a relatively new addition to the methods suitable for the characterization of interfacial processes. Through crystal truncation rod (CTR) diffraction and resonant anomalous X-ray reflectivity (RAXR) it is possible to selectively determine 3D interfacial structures with sub-Å resolution and elemental specificity. CTR diffraction is an extension of single-crystal X-ray diffraction focusing on a single crystal's surface. Here, the translational symmetry breaks down, which means the Bragg conditions do not need to be met in the surface normal direction. Consequently, constructive interference is not limited to zero-dimensional Bragg peaks, but instead one-dimensional rods of intensity can be detected connecting these peaks in ; Fenter, 2002). The measured intensity is extremely sensitive to changes in the structure of the interface region, which reaches from the bulk crystal to the undisturbed solution and includes any relaxed surface layers of the crystal, ordered absorbed water, and of course ions or particles registered at the interface. Due to the (Als-Nielsen & McMorrow, 2011), it is typically not possible to derive a structural model (electron density as a function of distance from the interface) from measured data. Instead, structural models are provided, and the corresponding CTR intensity is calculated. An interfacial structure can then be obtained by fitting the structural model to reproduce the measured data (Fenter, 2002).
albeit several orders of magnitude weaker than bulk Bragg peaks (Robinson & Tweet, 1992More recently, CTR has frequently been combined with RAXR, which allows to determine the contribution of one or several resonant atoms to the total electron density as determined by CTR (Park et al., 2005). The method is based on the dependence of the scattered or reflected intensity on the X-ray energy, close to the of an element, if it is part of the interfacial structure. For RAXR the incident X-ray energy E is scanned around the E0 of the resonant element at constant momentum transfer q. If the element is coherently registered at the interface, the energy scan will show modulations at E0 as superpositions of the element's terms f ′ and f ′′ (see also Fig. 18), whose amplitude and phase will depend on the elements adsorbed quantity and average distance from the interface, respectively (Park & Fenter, 2007). Based on the CTR data, a model for the distribution of the resonant atom can then be constructed using similar approaches as for the CTR modeling. There are a number of related techniques based on the same physical effects, such as coherent Bragg rod analysis (COBRA) or diffraction anomalous fine structure (DAFS), all of which are in principle feasible at ROBL, but have not been implemented yet.
When hard X-rays (>7 keV) are applied, the method is suited for in situ investigations, e.g. of mineral interfaces in contact with a solution (Eng et al., 2000; Fenter, 2002; Trainor et al., 2004, 2006). The type of information obtainable through CTR/RAXR is valuable in the context of nuclear waste management, in order to characterize the interaction of radionuclides with mineral phases. As heavy scatterers with a high number of electrons, actinides are ideally suited elements, yet the publications remain relatively few, due to the lack of accessible beam time for this type of research, amongst other factors (Catalano et al., 2005; Schmidt et al., 2013). ROBL's diffraction end-station is the only end-station purpose built for studying interfacial reactions of radioactive samples using surface X-ray diffraction techniques, such as CTR diffraction and RAXR. Several thin-film cell sample holders are available at the beamline, including triple confinement holders for radioactive (or very sensitive) samples based on more complex earlier designs (Schmidt et al., 2011).
As an example, Fig. 16 shows representative data and the corresponding electron densities at the mica/water interface, after reacting muscovite with Zr for 24 h in two different background electrolytes (LiCl and CsCl) (Qiu et al., 2018). The CTR-derived electron density changes abruptly at the mica surface (0 Å, defined as the position of the top-most oxygen layer in the mica structure) from the crystalline structure left to the aqueous phase right. The peak at about 2.5 Å corresponds to the first adsorbed water layer. The RAXR-derived electron density of Zr (filled area) is broadly distributed over 20–25 Å from the surface, indicating the presence of Zr-oxo-nanoparticles. From the CsCl medium, less Zr adsorbs (which is directly evident from the smaller amplitude in the RAXR modulations (Fig. 16, bottom left) and the formed nanoparticles are smaller. The result can be interpreted as an effect of the stronger competition by the more strongly adsorbing Cs+ relative to Li+, which in turn is caused by the weaker hydration of the large Cs+cation.
7. XRD-2 (Pilatus diffractometer)
Recent detector developments resulted in very large single photon-counting 2D detectors, which can be used for a complete registration of all reflexes in the in situ measurements. While the large sensitive surface area of these detectors allows a simultaneous registration of a wide 2θ range, their heavy weight interferes with an accurate and reproducible movement relative to the sample using a six-circle diffractometer. For this reason, we decided to build a second diffractometer station, with a Pilatus 2M detector placed in a sturdy support frame which is mounted together with the sample goniometer on a common granite table. The design was developed by Instrument Design Technology Ltd together with SNBL/ESRF (Dyadkin et al., 2016). The diffractometer is equipped with a Pilatus3 X 2M detector (Dectris Ltd), with a photon-sensitive layer of 450 µm-thick Si of an area of 253.7 mm × 288.8 mm (width × height), and a pixel size of 172 µm × 172 µm. The detector is mounted in a support frame, which allows lateral and vertical movements as well as rotation (see Fig. 17).
in a short time, necessary for single-crystal diffraction of small- and large-molecule crystallography, as well as for the quick registration of complete scattering patterns duringA total of 15 motorized movements allow the sample and the detector to be precisely positioned and aligned relative to the synchrotron beam. The granite table is a five-degrees-of-freedom system and consists of three 2.5 Mg-carrying high-stiffness jacks. Differential positioning of the custom guided jacks provides actuation of the vertical height (z), pitch and roll of the support frame. The jack assemblies are mounted on motorized horizontal slide units, that can be driven together to achieve a horizontal translation (y), or in opposition to each other to obtain a yaw movement. The detector mount comprises two parallel translational (x) and one vertical linear stage (z). The main linear translation supports the detector support frame. The second translation moves the detector support. The detector support is mounted on a goniometer 420 (Huber diffraction GmbH) which is used to rotate around the y-axis. The detector support unit, consisting of the Pilatus detector and two beam diagnostic devices, is counter-balanced by a 60 kg weight to support the vertical movement of the unit. The two beam diagnostic devices are positioned in the same translational plane as the photosensitive layer of the Pilatus detector and aid in shaping the X-ray beam. The first is an air beam viewer with a magnification of 0.8 (ESRF) and the second a FDS photonic direct beam camera (Photonic Science). The sample–detector distances can be varied between 140 and 600 mm. Three goniometers are available to mount samples: a Huber uniaxial goniometer 410, a Huber Kappa goniometer 512.410, and an Arinax Kappa goniometer MK3. Diffraction measurements can be combined with XANES and XRF spectroscopy using a single-element Si drift detector (Vortex X90 CUBE, 1000 µm SDD, 50 mm2 collimated down to 30 mm2, 25 µm Be window) with a FalconX1 processor. Parasitic scattering can be suppressed with additional collimators (titanium, tantalum or molybdenum) with an entry hole diameter of 2 mm and a total detection angle of 22.6°.
A beam conditioning unit in front of the goniometer (Fig. 18) comprises an entrance slit, a small to monitor the incoming beam intensity, and a collimation unit to suppress parasitic scattering from the slit blades. For optical sample alignment, a long-distance microscope K2 (Infinity Photo-Optical GmbH) is placed at a distance of 180 mm from the sample position. A macro camera is used to align supporting devices like a heat chamber HTK1200 (Anton Paar, up to 1200°C). The setup also includes a cryocooler 800 series (Oxford Cryosystems Ltd) which controls a temperature range of 80–400 K. The diffractometer is operated with the GUI-based software Pylatus (Dyadkin et al., 2016).
Fig. 18 (top) shows the result of a feasibility test with a crystal of UCl2salen(MeOH)2 mounted in a sealed Kapton capillary. The was set to 17038 eV to reduce the sample absorption and to avoid fluorescence arising from the uranium L3 at 17162 eV. The fine structure of the factors f ′ and f ′′ of U(IV) were extracted using the kkcalc code (Watts, 2014) from the U L3 The single-crystal diffraction patterns were registered with the program suite Pylatus (Dyadkin et al., 2016) and the reflections were extracted using the Crysalis software (Rigaku). The structure was solved with Olex2 (Dolomanov et al., 2009) with a R1 of 2.99% (Radoske et al., 2021).
In situ powder XRD is a powerful technique to follow structural modifications as a function of, for example, temperature or gas atmosphere, with various applications in kinetic, catalytic and thermodynamic studies. A hot gas generator from Cyberstar [Fig. 19(a)] is available to heat a sample up to 1000°C in air with a precision of ±1°C controlled by a Eurotherm and a gas flow controller. Alternatively, a heating chamber HTK1200N (Anton Paar) can be mounted at XRD-1 and XRD-2 [Fig. 19(b)]. This furnace can be used for measuring PXRD in reflection and transmission geometries at temperatures up to 1200°C in air, inert atmosphere (O2, N2, Ar, He, and other non-hazardous, non corrosive gasses) or in vacuum (≥10−4 mbar). A capillary spinner enables measurements in transmission geometry for highly air-sensitive materials or samples with strong preferrential orientation.
Fig. 19(c) shows the stepwise transformation of UO2 into U3O8 through oxidation measured with the Pilatus diffractometer. Determining the molar fractions of the individual oxide phases (i.e. UO2, U4O9, U3O7 and U3O8) is essential for the understanding of oxidation reactions and their kinetics. Simultaneous with the PXRD measurements, the oxygen-to-metal ratio was determined by recording in operando the XANES spectra with the Vortex X90 CUBE silicon drift detector. Fig. 19(d) shows the good agreement between the O/M ratio derived from the PXRD oxide phase fractions and from the XANES spectra (De Bona et al., 2021).
8. ROBL-II in comparison with other European radionuclide beamlines
To the best of our knowledge, only five dedicated radionuclide beamlines are currently fully operational, all of them in Europe (Table 3). Although ROBL-II shares many characteristics with the other four beamlines, including a variety of techniques, its four permanently mounted experimental stations provide the possibility to use all available techniques even during a typical six-day run, with easy sample transport between the stations, which are in one controlled area. Other distinguished features are the high-count-rate detector and multisample cryostat for the six-crystal spectrometer with complete crystal sets for 1 and 0.5 m Rowland radii, and finally the surface and single-crystal diffraction stations for radioactive samples offered by XRD-1 and XRD-2.
‡Vertical collimation and dynamic focusing in horizontal. §VFM: vertical focusing mirror. |
The two KARA beamlines are distinguished by their proximity to dedicated alpha labs of INE, where sample preparation can be performed, as well as the possibility for in situ experiments at non-ambient conditions including heating up to 2000 K. Although beam time is no longer provided through the typical proposal review process, it is provided through collaboration with INE.
While these previously described beamlines have been designed with predominately alpha-emitting radionuclides in mind, the experimental stations of the MARS beamline incorporate substantial lead shieldings around samples to enable the measurement of much stronger gamma-emitting samples. Such lead-shieldings can also be (temporarily) established at the microXAS beamlines, which is further discriminated by its high spatial resolution through strong microfocussing down to 1 µm.
Footnotes
‡Current address: Institute of Experimental and Applied Physics, Christian Albrechts Universität, 24098 Kiel, Germany.
Acknowledgements
The authors would like to thank: Amparo Vivo and Raymond Barrett (ESRF, France) for the quality control of mirrors; Manuel Sanchez del Rio (ESRF) for ray-tracing calculations; Oleg Tchoubar (Brookhaven National Lab, USA) for wavefront propagation calculations. Pieter Glatzel, Mauro Rovezzi, Olivier Mathon (ESRF) and Catherine Heyman (SASU Grenoble) for design and specification of
Peter Eng and Joanne Stubbs (University of Chicago, USA) for support to develop the surface diffraction module and a bunch of macros to control the measurement. Roman Gumeniuk (TU Bergakademie Freiberg, Germany) for the development of the powder diffraction module, Staffan Ohlsson (ESRF) for the development of the automatic protection module for the six-circle diffractometer. Wouter van Beek and Dmitry Chernyshov (SNBL at ESRF), Atsushi Ikeda-Ohno and Thomas Radoske (HZDR, Germany) for the support to develop the Pilatus diffractometer. Joerg Rothe (KIT, Germany), Pier-Solari Lorenzo, Myrtille Hunault, Denis Menut (Synchrotron Soleil, France), and Daniel Grolimund (PSI, Switzerland) for updated information on the radionuclide beamlines at KARA, Soleil and SLS. Manuel Feig is grateful to DFG (Deutsche Forschungsgemeindschaft) for financial support (grant GU1632/4-1, Number 325295543). Open access funding enabled and organized by Projekt DEAL.References
Als-Nielsen, J. & McMorrow, D. (2011). Elements of Modern X-ray Physics, 2nd ed. Chichester: John Wiley & Sons, Ltd. Google Scholar
Amidani, L., Plakhova, T., Romanchuk, A. Y., Gerber, E., Weiss, S., Efimenko, A., Sahle, C. J., Butorin, S. M., Kalmykov, S. N. & Kvashnina, K. (2019). Phys. Chem. Chem. Phys. 21, 10635–10643. Web of Science CrossRef CAS PubMed Google Scholar
Ashiotis, G., Deschildre, A., Nawaz, Z., Wright, J. P., Karkoulis, D., Picca, F. E. & Kieffer, J. (2015). J. Appl. Cryst. 48, 510–519. Web of Science CrossRef CAS IUCr Journals Google Scholar
Borca, C. N., Grolimund, D., Willimann, M., Meyer, B., Jefimovs, K., Vila-Comamala, J. & David, C. (2009). J. Phys. Conf. Ser. 186, 012003. CrossRef Google Scholar
Catalano, J. G., Trainor, T. P., Eng, P. J., Waychunas, G. A. & Brown, G. E. (2005). Geochim. Cosmochim. Acta, 69, 3555–3572. Web of Science CrossRef CAS Google Scholar
De Bona, E., Cologna, M., Vigier, J.-F., Hennig, C. & Prieur, D. (2021). Inorg. Chem. Submitted. Google Scholar
De Groot, F. M. F. & Kotani, A. (2008). Core Level Spectroscopy of Solids. Boca Raton: CRC Press. Google Scholar
Dejoie, C., Coduri, M., Petitdemange, S., Giacobbe, C., Covacci, E., Grimaldi, O., Autran, P.-O., Mogodi, M. W., Šišak Jung, D. & Fitch, A. N. (2018). J. Appl. Cryst. 51, 1721–1733. Web of Science CrossRef CAS IUCr Journals Google Scholar
Denecke, M. A. (2006). Coord. Chem. Rev. 250, 730–754. Web of Science CrossRef CAS Google Scholar
Dolomanov, O. V., Bourhis, L. J., Gildea, R. J., Howard, J. A. K. & Puschmann, H. (2009). J. Appl. Cryst. 42, 339–341. Web of Science CrossRef CAS IUCr Journals Google Scholar
Domaschke, K., Rossberg, A. & Villmann, T. (2014). European Symposium on Artificial Neural Networks, Computational Intelligence and Machine Learning, Bruges, Belgium, pp. 277–282. Google Scholar
Dyadkin, V., Pattison, P., Dmitriev, V. & Chernyshov, D. (2016). J. Synchrotron Rad. 23, 825–829. Web of Science CrossRef CAS IUCr Journals Google Scholar
Eng, P. J., Trainor, T. P., Brown, G. E., Waychunas, G. A., Newville, M., Sutton, S. R. & Rivers, M. L. (2000). Science, 288, 1029–1033. Web of Science CrossRef PubMed CAS Google Scholar
Epifano, E., Naji, M., Manara, D., Scheinost, A. C., Hennig, C., Lechelle, J., Konings, R. J. M., Guéneau, C., Prieur, D., Vitova, T., Dardenne, K., Rothe, J. & Martin, P. M. (2019). Commun. Chem. 2, 1–11. Web of Science CrossRef Google Scholar
Fenter, P. A. (2002). Rev. Mineral. Geochem. 49, 149–221. Web of Science CrossRef CAS Google Scholar
Funke, H., Bernhard, G., Claussner, J., Jansen, K., Matz, W., Nitsche, H., Oehme, W., Reich, T. & Rollig, D. (2001). Kerntechnik, 66, 195–201. CAS Google Scholar
Funke, H., Chukalina, M. & Scheinost, A. C. (2007). J. Synchrotron Rad. 14, 426–432. Web of Science CrossRef CAS IUCr Journals Google Scholar
Funke, H., Scheinost, A. C. & Chukalina, M. (2005). Phys. Rev. B, 71, 094110. Web of Science CrossRef Google Scholar
Geckeis, H., Lützenkirchen, J., Polly, R., Rabung, T. & Schmidt, M. (2013). Chem. Rev. 113, 1016–1062. Web of Science CrossRef CAS PubMed Google Scholar
Glatzel, P. & Bergmann, U. (2005). Coord. Chem. Rev. 249, 65–95. Web of Science CrossRef CAS Google Scholar
Hazemann, J.-L., Proux, O., Nassif, V., Palancher, H., Lahera, E., Da Silva, C., Braillard, A., Testemale, D., Diot, M.-A., Alliot, I., Del Net, W., Manceau, A., Gélébart, F., Morand, M., Dermigny, Q. & Shukla, A. (2009). J. Synchrotron Rad. 16, 283–292. Web of Science CrossRef CAS IUCr Journals Google Scholar
Hodeau, J. L., Bordet, P., Anne, M., Prat, A., Fitch, A. N., Dooryhee, E., Vaughan, G. & Freund, A. (1998). Proc. SPIE, 3448, 353–361. CrossRef Google Scholar
Huotari, S., Albergamo, F., Vankó, G., Verbeni, R. & Monaco, G. (2006). Rev. Sci. Instrum. 77, 053102. Web of Science CrossRef Google Scholar
Kavčič, M., Budnar, M., Mühleisen, A., Gasser, F., Žitnik, M., Bučar, K. & Bohinc, R. (2012). Rev. Sci. Instrum. 83, 033113. Web of Science PubMed Google Scholar
Kieffer, J., Valls, V., Blanc, N. & Hennig, C. (2020). J. Synchrotron Rad. 27, 558–566. Web of Science CrossRef CAS IUCr Journals Google Scholar
Kirsch, R., Fellhauer, D., Altmaier, M., Neck, V., Rossberg, A., Fanghänel, T., Charlet, L. & Scheinost, A. C. (2011). Environ. Sci. Technol. 45, 7267–7274. Web of Science CrossRef CAS PubMed Google Scholar
Kleymenov, E., van Bokhoven, J. A., David, C., Glatzel, P., Janousch, M., Alonso-Mori, R., Studer, M., Willimann, M., Bergamaschi, A., Henrich, B. & Nachtegaal, M. (2011). Rev. Sci. Instrum. 82, 065107. Web of Science CrossRef PubMed Google Scholar
Kvashnina, K. O. & Scheinost, A. C. (2016). J. Synchrotron Rad. 23, 836–841. Web of Science CrossRef CAS IUCr Journals Google Scholar
Llorens, I., Solari, P. L., Sitaud, B., Bes, R., Cammelli, S., Hermange, H., Othmane, G., Safi, S., Moisy, P., Wahu, S., Bresson, C., Schlegel, M. L., Menut, D., Bechade, J. L., Martin, P., Hazemann, J. L., Proux, O. & Den Auwer, C. (2014). Radiochim. Acta, 102, 957–972. CrossRef CAS Google Scholar
Maher, K., Bargar, J. R. & Brown, G. E. (2013). Inorg. Chem. 52, 3510–3532. Web of Science CrossRef CAS PubMed Google Scholar
Matz, W., Schell, N., Bernhard, G., Prokert, F., Reich, T., Claußner, J., Oehme, W., Schlenk, R., Dienel, S., Funke, H., Eichhorn, F., Betzl, M., Pröhl, D., Strauch, U., Hüttig, G., Krug, H., Neumann, W., Brendler, V., Reichel, P., Denecke, M. A. & Nitsche, H. (1999). J. Synchrotron Rad. 6, 1076–1085. Web of Science CrossRef CAS IUCr Journals Google Scholar
Nitsche, H. (1995). J. Alloys Compd. 223, 274–279. CrossRef CAS Web of Science Google Scholar
Park, C. & Fenter, P. A. (2007). J. Appl. Cryst. 40, 290–301. Web of Science CrossRef CAS IUCr Journals Google Scholar
Park, C., Fenter, P. A., Sturchio, N. C. & Regalbuto, J. R. (2005). Phys. Rev. Lett. 94, 076104. Web of Science CrossRef PubMed Google Scholar
Qiu, C., Eng, P. J., Hennig, C. & Schmidt, M. (2018). Langmuir, 34, 12270–12278. Web of Science CrossRef CAS PubMed Google Scholar
Radoske, T., Kloditz, R., Fichter, S., Kaden, P., März, J., Stumpf, T. & Schmidt, M. (2021). Inorg. Chem. Submitted. Google Scholar
Reich, T., Bernhard, G., Geipel, G., Funke, H., Hennig, C., Roßberg, A., Matz, W., Schell, N. & Nitsche, H. (2000). Radiochim. Acta, 88, 633–637. Web of Science CrossRef CAS Google Scholar
Robinson, I. K. & Tweet, D. J. (1992). Rep. Prog. Phys. 55, 599–651. CrossRef CAS Web of Science Google Scholar
Rossberg, A. & Funke, H. (2010). J. Synchrotron Rad. 17, 280–288. Web of Science CrossRef CAS IUCr Journals Google Scholar
Rossberg, A., Reich, T. & Bernhard, G. (2003). Anal. Bioanal. Chem. 376, 631–638. Web of Science PubMed CAS Google Scholar
Rossberg, A. & Scheinost, A. C. (2005). Anal. Bioanal. Chem. 383, 56–66. Web of Science CrossRef PubMed CAS Google Scholar
Rossberg, A., Ulrich, K.-U., Weiss, S., Tsushima, S., Hiemstra, T. & Scheinostt, A. C. (2009). Environ. Sci. Technol. 43, 1400–1406. Web of Science CrossRef PubMed CAS Google Scholar
Rothe, J., Butorin, S., Dardenne, K., Denecke, M. A., Kienzler, B., Löble, M., Metz, V., Seibert, A., Steppert, M., Vitova, T., Walther, C. & Geckeis, H. (2012). Rev. Sci. Instrum. 83, 043105. Web of Science CrossRef PubMed Google Scholar
Rovezzi, M., Harris, A., Detlefs, B., Bohdan, T., Svyazhin, A., Santambrogio, A., Degler, D., Baran, R., Reynier, B., Noguera Crespo, P., Heyman, C., Van Der Kleij, H.-P., Van Vaerenbergh, P., Marion, P., Vitoux, H., Lapras, C., Verbeni, R., Kocsis, M. M., Manceau, A. & Glatzel, P. (2020). J. Synchrotron Rad. 27, 813–826. Web of Science CrossRef CAS IUCr Journals Google Scholar
Rovezzi, M., Lapras, C., Manceau, A., Glatzel, P. & Verbeni, R. (2017). Rev. Sci. Instrum. 88, 013108. Web of Science CrossRef PubMed Google Scholar
Schmidt, M., Eng, P. J., Stubbs, J. E., Fenter, P. & Soderholm, L. (2011). Rev. Sci. Instrum. 82, 075105. Web of Science CrossRef PubMed Google Scholar
Schmidt, M., Lee, S. S., Wilson, R. E., Knope, K. E., Bellucci, F., Eng, P. J., Stubbs, J. E., Soderholm, L. & Fenter, P. (2013). Environ. Sci. Technol. 47, 14178–14184. Web of Science CrossRef CAS PubMed Google Scholar
Sitaud, B., Solari, P. L., Schlutig, S., Llorens, I. & Hermange, H. (2012). J. Nucl. Mater. 425, 238–243. Web of Science CrossRef CAS Google Scholar
Sokaras, D., Weng, T. C., Nordlund, D., Alonso-Mori, R., Velikov, P., Wenger, D., Garachtchenko, A., George, M., Borzenets, V., Johnson, B., Rabedeau, T. & Bergmann, U. (2013). Rev. Sci. Instrum. 84, 053102. Web of Science CrossRef PubMed Google Scholar
Trainor, T. P., Chaka, A. M., Eng, P. J., Newville, M., Waychunas, G. A., Catalano, J. G. & Brown, G. E. (2004). Surf. Sci. 573, 204–224. Web of Science CrossRef CAS Google Scholar
Trainor, T. P., Templeton, A. S. & Eng, P. J. (2006). J. Electron Spectrosc. Relat. Phenom. 150, 66–85. Web of Science CrossRef CAS Google Scholar
Van Bokhoven, J. A. & Lamberti, C. (2016). X-ray Absorption and X-ray Emission Spectroscopy: Theory and Applications. Chichester: John Wiley & Sons. Google Scholar
Watts, B. (2014). Opt. Express, 22, 23628–23639. Web of Science CrossRef PubMed Google Scholar
Yalçıntaş, E., Scheinost, A. C., Gaona, X. & Altmaier, M. (2016). Dalton Trans. 45, 17874–17885. Web of Science PubMed Google Scholar
Zimina, A., Dardenne, K., Denecke, M. A., Doronkin, D. E., Huttel, E., Lichtenberg, H., Mangold, S., Pruessmann, T., Rothe, J., Spangenberg, T., Steininger, R., Vitova, T., Geckeis, H. & Grunwaldt, J. D. (2017). Rev. Sci. Instrum. 88, 113113. Web of Science CrossRef PubMed Google Scholar
Zimina, A., Dardenne, K., Denecke, M. A., Grunwaldt, J. D., Huttel, E., Lichtenberg, H., Mangold, S., Pruessmann, T., Rothe, J., Steininger, R. & Vitova, T. (2016). J. Phys. Conf. Ser. 712, 012019. CrossRef Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.