Resonant inelastic X-ray scattering endstation at the 1C beamline of Pohang Light Source II
aDepartment of Physics, Pohang University of Science and Technology, Pohang 790-784, Republic of Korea, bCenter for Artificial Low Dimensional Electronic Systems, Institute for Basic Science (IBS), 77 Cheongam-Ro, Pohang 37673, Republic of Korea, and cAdvanced Photon Source, Argonne National Laboratory, Lemont, IL 60439, USA
*Correspondence e-mail: bjkim6@postech.ac.kr
An endstation for resonant inelastic X-ray scattering (RIXS), dedicated to operations in the hard X-ray regime, has been constructed at the 1C beamline of Pohang Light Source II. At the Ir L3-edge, a total energy resolution of 34.2 meV was achieved, close to the theoretical estimation of 34.0 meV, which considers factors such as the incident energy bandpass, intrinsic analyzer resolution, geometrical broadening of the spectrometer, finite beam-size effect and Johann aberration. The performance of the RIXS instrument is demonstrated by measuring the RIXS spectra of Sr2IrO4. The endstation can be easily reconfigured to measure energy-integrated intensities with very low background for diffuse scattering and diffraction experiments.
1. Introduction
Resonant inelastic X-ray scattering (RIXS) has emerged in the last decade as a complementary tool to inelastic neutron scattering for measuring momentum- and energy-resolved excitation spectra of magnetic materials, since the observations of two-magnon (Hill et al., 2008) and single-magnon (Braicovich et al., 2009) excitations using RIXS at the Cu K-edge (8.99 keV) and L3-edge (930 eV), respectively. In the hard X-ray regime, single-magnon excitations were first revealed in Sr2IrO4 (Kim et al., 2012a) using a medium-energy-resolution RIXS instrument (Shvyd'ko et al., 2013). An energy resolution better than 100 meV (Shvyd'ko et al., 2013; Gog et al., 2013; Moretti Sala et al., 2018) allowed low-energy features in an to be resolved, including single-magnon excitations. The hard X-ray RIXS spectrometer is based on the Rowland circle geometry in a near-backscattering condition which, with a position-sensitive detector and an improved energy bandpass of the incident X-rays, has seen a dramatic enhancement in the energy resolution and the photon counts (Shvyd'ko et al., 2013; Huotari et al., 2005). Now that the theoretical energy resolution based on this scheme has been achieved, the development of a new type of spectrometer using flat crystal optics is underway (Kim et al., 2018), which is also advantageous for high-pressure experiments (Kim et al., 2020).
RIXS is a photon-in photon-out spectroscopy which can, in principle, analyse all kinds of bosonic excitations from charge, orbital, spin and lattice et al., 2011a) and polarization of the scattered photons (Gao et al., 2016; Ishii et al., 2013). Resonant processes in RIXS can enhance the sensitivities to low-energy excitations of specific elements and orbitals, which are selected by tuning the incident X-ray energies to an of an element in the sample. Advances in the theoretical interpretation of the RIXS process have been made in parallel (Haverkort, 2010; Ament et al., 2011b; Kim & Khaliullin, 2017) and RIXS is now considered a major tool for studying magnetic and electronic excitations of strongly correlated electron systems. For example, RIXS has been applied to studies of the magnetism of the single-layer square-lattice iridate Sr2IrO4 (Kim et al., 2012a), its bilayer counterpart Sr3Ir2O7 (Kim et al., 2012b; Moretti Sala et al., 2015) and honeycomb iridates (Chun et al., 2015).
by measuring the changes in energy, momentum (AmentIn this paper, we report the successful commissioning of the RIXS endstation at the 1C beamline of Pohang Light Source II (PLS-II). We describe the main components of the endstation, which include monochromators, focusing mirrors and a spectrometer, and show their characteristics and performances. We demonstrate the performance of our RIXS instrument tested by acquiring low-energy excitation spectra of Sr2IrO4. We also demonstrate efficient measurement of X-ray diffuse scattering images by slightly reconfiguring the setup.
2. Beamline overview
Fig. 1 shows the schematic layout of the 1C beamline with the components relevant to the RIXS instrument. The source (U20) consists of a 20 mm-period in-vacuum Sm2Co17 undulator (total 70 periods, 1.4 m long) with an average effective peak field of 0.81 T at a 5 mm gap. The first major optical component is the nitrogen-cooled Si(111) double-crystal monochromator (DCM), which provides the accessible energy range 2.3 to 23 keV. The vertical position of the second crystal is adjusted when changing the energy to fix the height of the exit beam. For RIXS at the Ir L3-edge (11.215 keV), the DCM provides X-rays with a bandwidth around 1 eV using the fifth-order harmonic. After the DCM, the beam is defined in real space by a four-blade slit (HS1) before entering the experimental hutch. A high-resolution monochromator (HRM), a single channel-cut, is inserted to further narrow the bandwidth. For the Ir L3-edge, an Si(844) channel-cut is used to narrow the energy bandwidth down to ∼15.8 meV (Gog et al., 2013). A pair of Kirkpatrick–Baez (KB) mirrors (Kirkpatrick & Baez, 1948) followed by another four-blade slit (HS2) that defines the beam position after reflections then focus the monochromatic beam on the sample position. The sample is mounted on a closed-cycle helium cryostat which covers the temperature range 4 K to 350 K. The energies and momenta of scattered photons from the sample are analyzed by a RIXS spectrometer based on the Rowland circle geometry (Shvyd'ko et al., 2013; Rueff et al., 2015; Moretti Sala et al., 2018). The RIXS spectrometer is installed on a four-circle kappa diffractometer that provides a wide range of the chi (χ) motion of the sample. The advantages of using a kappa-type diffractometer will be discussed in more detail in Section 4. The at the Ir L3-edge is 1.5 × 1011 photons s−1 after the DCM and 7.1 × 108 photons s−1 after the HRM. The was measured at a storage ring current of 400 mA using a calibrated 300 µm-thick silicon photodiode with corrections including detector efficiency (Owen et al., 2009). Although the beam limited by the 3 GeV ring is two orders of magnitude lower compared with that at Sector 27 of the Advanced Photon Source (Shvyd'ko et al., 2013; Gog et al., 2018), the count rate at the detector is partially compensated by the larger solid angle of acceptance of the 1 m Rowland circle geometry in our RIXS setup. The accessible absorption edges with current in-stock analyzers and channel-cuts are tabulated in Table 1 with the estimated total energy resolutions, which take into account the contributions from the incident energy bandwidth, Darwin width, geometrical broadening, finite beam-size effect and Johann aberration.
|
3. Focusing
A pair of KB mirrors from Thales SESO focuses the monochromatic beam after the HRM. The horizontal focusing mirror (HFM) has a platinum coating with a thickness of 50 nm and a tangential slope error of 2 µrad RMS. The vertical focusing mirror (VFM), on the other hand, has a unique coating consisting of platinum and palladium, each with a thickness of 50 nm, separated into different areas and has a tangential slope error of 1 µrad RMS. The platinum-coated area is typically used at the Ir L3-edge, whereas the palladium-coated area is used for lower energies to achieve higher reflection efficiency.
The incident beam with dimensions 3.5 mm × 1.0 mm (H × V) is first reflected by the 800 mm-long HFM at an incidence angle of 5 mrad, and then by the 150 mm-long VFM at an incidence angle of 3 mrad. The KB mirrors are capable of accepting a beam size of 4 mm × 0.5 mm, which can fully cover the profile in the horizontal direction but only half of the beam in the vertical direction. Due to limited space inside the hutch, the vertical mirror had to be designed to be shorter than necessary to accept the full vertical profile of X-rays. The X-rays that spill over the vertical mirror are blocked by a four-blade slit (HS2) after the mirrors. The total efficiency of the KB mirrors is approximately 40%, primarily due to the short VFM (50%) and the reflection efficiency of platinum coatings (∼80%).
Fig. 2 shows the 25.3 µm × 2.6 µm (H × V) focused beam size at the sample position, measured by knife-edge scans. The beam size is estimated by fitting the derivative of the knife-edge scans in both the horizontal and the vertical directions using a pseudo-Voigt function. The measured beam size is close to the theoretically estimated size of 28.4 µm × 1.9 µm calculated only by the demagnification factors from the focal lengths. The focal lengths of the HFM and the VFM are 1975 mm and 1400 mm located at 34.725 m and 36.7 m from the source, respectively. The source size, 500 µm × 50 µm (H × V), then provides the theoretical values for the focused beam size at the sample position, 28.4 µm × 1.9 µm. A vertical beam size of 2.6 µm is notably smaller than those of other hard X-ray RIXS beamlines (Shvyd'ko et al., 2013; Moretti Sala et al., 2018), which prevents a potential broadening of energy resolution by a finite beam size.
4. Spectrometer
For the energy and momentum analyses, we installed a RIXS spectrometer based on the Rowland circle geometry on a four-circle kappa diffractometer as shown in Fig. 3. A kappa-type diffractometer implements the chi (χ) rotation in the Euler geometry through the combination of kappa (κ), theta (θ) and phi (ϕ) rotations. The advantage of using a kappa-type diffractometer is that it provides a wide range of the χ rotation without a χ cradle, which interferes with the detector located right above the sample in the Rowland circle geometry [Figs. 3 and 4(a)]. We note that most RIXS spectrometers based on the Eulerian diffractometer employ an arc segment for the χ rotation with a limited range (Shvyd'ko et al., 2013; Moretti Sala et al., 2018). On the contrary, the κ angle of our diffractometer (50°) guarantees the χ range ±100°. Also, in a horizontal scattering geometry where RIXS measurements are conducted, χ is at 90° and the ϕ rotation provides a full azimuthal angle range.
These large et al., 2019; Bertinshaw et al., 2020). However, with a limited χ range, it is hard to control the incident angle close to 0° or 90°. For example, Q = (3 2 28.2) was used to make a grazing incidence geometry in previous work (Porras et al., 2019), where χ ≃ 6.7° and the incident angle is maintained around 14.5°. In contrast, with our diffractometer, an incdient angle below 1° can be achieved with Q = (5 2 21.5) which requires a high value of χ ≃ 25°. Also, it would be useful to carry out thin-film measurements which require a grazing incident geometry to increase the footprint of the X-rays.
in rotations of the sample allow us to explore more and measure in a more extreme scattering geometry. It has been shown in RIXS measurements of a square-lattice iridate that RIXS in normal or grazing incident geometry – where the incident angle is close to 0° or 90°, respectively – enhances or suppresses one of the polarization components (PorrasThe introduction of the Rowland circle geometry (Johann, 1931), the advances in fabrication technology for diced spherical analyzers (Masciovecchio et al., 1996; Verbeni et al., 2005; Said et al., 2011) and the use of a position-sensitive detector (Huotari et al., 2005) significantly enhance the energy resolution and the photon counts simultaneously (Shvyd'ko et al., 2013). In this geometry, the scattered photons from the sample are analyzed by a diced spherical analyzer and focused onto a position-sensitive detector [Fig. 4(a)]. The analyzer and the detector are placed on the Rowland circle with a radius RA = 1 m [Fig. 4(a)] and a flight path filled with helium gas is installed on the beam path to reduce air scattering and maximize photon counts (Fig. 3). At the Ir L3-edge, an Si(844) diced spherical analyzer is used made of a 3 mm-thick Si(844) wafer and glued on top of a concave glass to create a curvature radius of 1 m. It has a 10 cm diameter and has square pixels (or dices) with dimensions 1 mm × 1 mm. A manual iris in front of the analyzer allows for continuous adjustment of its active area, enabling improvement of the momentum resolution in exchange for photon counts. The flat surface of the dices with a 1 mm vertical size accepts a finite energy window, 0.84 eV, of the scattered X-rays from the sample since the incident angle to the analyzer varies at each vertical position of a dice. The analyzer then disperses these X-rays according to their energies following Bragg's law and maps them onto a position-sensitive Dectris MYTHEN2 detector which has 640 pixels with a size of 50 µm [Fig. 4(a)]. The use of this position-sensitive detector dramatically suppresses the geometrical broadening contribution to the energy resolution of the spectrometer (Huotari et al., 2005). The resulting geometrical broadening ΔEg of the spectrometer is determined by the pixel size of the detector ΔdD (Shvyd'ko et al., 2013) as
where θB is the of the analyzer reflection. This geometrical broadening of our spectrometer (RA = 1 m and ΔdD = 50 µm) at the Ir L3-edge is calculated as ΔEg = 20.94 meV.
The total energy resolution of our RIXS spectrometer was determined by measuring 3M Magic Scotch tape as an elastic reference scatterer and fitting the results with a pseudo-Voigt function as shown in Fig. 4(b). Our analysis shows a total resolution of 34.2 meV, which is in close agreement with the theoretically calculated value of 34.0 meV, as presented in Table 1. In order to calculate the total energy resolution at various edges, we considered five different contributions: (i) the energy bandpass of incident X-rays (ΔEi), (ii) the intrinsic resolution (Darwin width) of the analyzer (ΔEa), (iii) the geometrical contribution defined in equation (2) (ΔEg), (iv) the finite beam-size effect (ΔEs) and (v) the Johann aberration (ΔEJ).
The broadening caused by the finite beam size was calculated by estimating the et al. 2018)
variation due to the vertical beam size at the sample position, as described by (Moretti Salawhere sz is the vertical beam size at the sample position. We did not consider the horizontal beam size as it results in a negligible amount of broadening. The contribution from the Johann aberration, which arises from the deviation of the curvature of the spherical analyzer from the ideal angle for focusing the scattered photons, was estimated by the weighted average of the variation across the circular analyzer, as given by (Moretti Sala et al., 2018)
where A is the analyzer iris radius. The energy resolutions of the various accessible absorption edges, using our in-stock optical components, are summarized in Table 1. We note that the contribution from the Johann aberration (ΔEJ) usually dominates the total energy resolution, except for the Ir L3-edge where the reflection is close to back-scattering. However, the Johann aberration can be mitigated by reducing the active analyzer area. For the W L3-edge, reducing the diameter of the analyzer iris to half its original size results in a reduction of the Johann contribution to 79 meV, leading to a significantly improved total energy resolution of 129 meV.
Fig. 5 shows the RIXS spectra of Sr2IrO4 measured at (0, 0) and (π/2, π/2) using our endstation. Both are measured at L = 32.5 to minimize with 2θ ≃ 90°. We note that the low-energy excitations of Sr2IrO4 are almost two-dimensional and independent of a momentum transfer along the c axis (Kim et al., 2012a; Ishii et al., 2011). The analyzer iris was fully open to 9 cm diameter which produces a momentum resolution of ∼0.32 r.l.u. (in-plane units) in the H or K direction. The (0, 0) spectrum in Fig. 5(a) shows the out-of-plane magnon mode at 40 meV (Pincini et al., 2017; Porras et al., 2019) clearly resolved from the elastic peak. Other features such as the spin–orbit excitons at 350 meV and dd-excitations around 600 meV are also well reproduced (Kim et al., 2012a, 2014; Gretarsson et al., 2016; Bertinshaw et al., 2020). The (π/2, π/2) spectrum in Fig. 5(b) exhibits a resolution-limited single magnon around 100 meV with a count rate of approximately 45 counts per minute at a ring current of 400 mA. Overall, the RIXS spectra in Fig. 5 are in excellent agreement with previously reported results, demonstrating the performance of the RIXS spectrometer constructed at the 1C beamline of the PLS-II.
5. X-ray diffuse scattering and diffraction
The endstation can be easily reconfigured for measurements in which only energy-integrated intensities are needed, such as X-ray diffuse scattering and diffraction [Fig. 6(a)]. X-ray diffuse scattering measures short-range correlations between ionic positions or spins from energy-integrated signals, and thus is useful to reveal the nature of microscopic interactions between them. For example, X-ray thermal diffuse scattering can determine the phonon dispersion relations (Xu & Chiang, 2005) which reveal how atoms interact with each other in solids, and X-ray magnetic diffuse scattering can provide important insights into the anisotropy of spin interactions (Chun et al., 2015).
However, high efficiency and low noise level are essential for this technique since it measures small quasi-elastic signals that are diffuse in momentum space. To this end, our setup utilizes a bent spherical analyzer and the DECTRIS EIGER2 R 1M 2D area detector as shown in Fig. 6(a). The bent analyzer replaces the diced spherical analyzer and the 2D detector is placed before the position-sensitive detector. Unlike the diced analyzer, the bent analyzer preserves the spatial information of scatterings and maps them onto 2D images on the 2D detector, which can then be converted to momentum space. Therefore, this setup can efficiently measure diffuse signals for a finite area in momentum space in a single measurement. In addition, the bent analyzer with a broad energy bandwidth (∼2.7 eV) can accept 100× more photons compared with the diced spherical analyzer.
Fig. 6(b) shows raw images of a charge density wave (CDW) peak of TiSe2 taken by the 2D area detector over 10 s. Despite the short measuring time, our setup successfully measures diffuse signals around the CDW peak across the (Tc = 200 K). Furthermore, the reflection geometry along with a finite energy bandwidth of the analyzer effectively reduce the noise level in measurements, as demonstrated by the impressive signal-to-noise ratio (∼106) observed in the θ–2θ rocking curve of a charge Bragg peak of TiSe2 in Fig. 6(c).
6. Conclusions
We have described and demonstrated the performance of the RIXS endstation installed at the 1C beamline of PLS-II. The main components of the endstation include a high-resolution channel-cut monochromator, a pair of focusing KB mirrors and an RIXS spectrometer. The use of a kappa-type diffractometer for the spectrometer provides a wide range of accessible scattering geometries. We achieved a total energy resolution of 34.2 meV at the Ir L3-edge which is close to the calculated value of 34.0 meV. Our measurement on Sr2IrO4 reproduces all the features observed in prior works, and thus demonstrates that a scan over a 1 eV energy range can be acquired within an hour with good statistics. Further, our X-ray diffuse scattering setup in a reflection geometry demonstrates acquiring high-quality diffuse scattering images in ∼10 s.
Funding information
This work was supported by the Institute for Basic Science (grant no. IBS-R014-A2). Experiments at Pohang Light Source II (PLS-II) 1C beamline were supported in part by the Ministry of Science and ICT (MSICT) of the Korean government. JKK was supported by the Global PhD Fellowship Program from the National Research Foundation of Korea (grant no. 2018H1A2A1059958).
References
Ament, L. J. P., Khaliullin, G. & van den Brink, J. (2011b). Phys. Rev. B, 84, 020403. Web of Science CrossRef Google Scholar
Ament, L. J. P., van Veenendaal, M., Devereaux, T. P., Hill, J. P. & van den Brink, J. (2011a). Rev. Mod. Phys. 83, 705–767. Web of Science CrossRef CAS Google Scholar
Bertinshaw, J., Kim, J. K., Porras, J., Ueda, K., Sung, N. H., Efimenko, A., Bombardi, A., Kim, J., Keimer, B. & Kim, B. J. (2020). Phys. Rev. B, 101, 094428. Web of Science CrossRef Google Scholar
Braicovich, L., Ament, L. J. P., Bisogni, V., Forte, F., Aruta, C., Balestrino, G., Brookes, N. B., De Luca, G. M., Medaglia, P. G., Granozio, F. M., Radovic, M., Salluzzo, M., van den Brink, J. & Ghiringhelli, G. (2009). Phys. Rev. Lett. 102, 167401. Web of Science CrossRef PubMed Google Scholar
Gao, X., Casa, D., Kim, J., Gog, T., Li, C. & Burns, C. (2016). Rev. Sci. Instrum. 87, 083107. Web of Science CrossRef PubMed Google Scholar
Gog, T., Casa, D. M., Knopp, J., Kim, J., Upton, M. H., Krakora, R., Jaski, A., Said, A., Yavaş, H., Gretarsson, H. & Huang, X. R. (2018). J. Synchrotron Rad. 25, 1030–1035. Web of Science CrossRef CAS IUCr Journals Google Scholar
Gog, T., Casa, D. M., Said, A. H., Upton, M. H., Kim, J., Kuzmenko, I., Huang, X. & Khachatryan, R. (2013). J. Synchrotron Rad. 20, 74–79. Web of Science CrossRef CAS IUCr Journals Google Scholar
Gretarsson, H., Sung, N. H., Porras, J., Bertinshaw, J., Dietl, C., Bruin, J. A. N., Bangura, A. F., Kim, Y. K., Dinnebier, R., Kim, J., Al-Zein, A., Moretti Sala, M., Krisch, M., Le Tacon, M., Keimer, B. & Kim, B. J. (2016). Phys. Rev. Lett. 117, 107001. Web of Science CrossRef PubMed Google Scholar
Haverkort, M. W. (2010). Phys. Rev. Lett. 105, 167404. Web of Science CrossRef PubMed Google Scholar
Hill, J. P., Blumberg, G., Kim, Y.-J., Ellis, D. S., Wakimoto, S., Birgeneau, R. J., Komiya, S., Ando, Y., Liang, B., Greene, R. L., Casa, D. & Gog, T. (2008). Phys. Rev. Lett. 100, 097001. Web of Science CrossRef PubMed Google Scholar
Huotari, S., Vankó, Gy., Albergamo, F., Ponchut, C., Graafsma, H., Henriquet, C., Verbeni, R. & Monaco, G. (2005). J. Synchrotron Rad. 12, 467–472. Web of Science CrossRef CAS IUCr Journals Google Scholar
Hwan Chun, S., Kim, J.-W., Kim, J., Zheng, H., Stoumpos, C. C., Malliakas, C., Mitchell, J., Mehlawat, K., Singh, Y., Choi, Y., Gog, T., Al-Zein, A., Sala, M., Krisch, M., Chaloupka, J., Jackeli, G., Khaliullin, G. & Kim, B. J. (2015). Nat. Phys. 11, 462–466. Web of Science CrossRef Google Scholar
Ishii, K., Jarrige, I., Yoshida, M., Ikeuchi, K., Inami, T., Murakami, Y. & Mizuki, J. (2013). J. Electron Spectrosc. Relat. Phenom. 188, 127–132. Web of Science CrossRef CAS Google Scholar
Ishii, K., Jarrige, I., Yoshida, M., Ikeuchi, K., Mizuki, J., Ohashi, K., Takayama, T., Matsuno, J. & Takagi, H. (2011). Phys. Rev. B, 83, 115121. Web of Science CrossRef Google Scholar
Johann, H. H. (1931). Z. Phys. 69, 185–206. CrossRef CAS Google Scholar
Kim, B. J. & Khaliullin, G. (2017). Phys. Rev. B, 96, 085108. Web of Science CrossRef Google Scholar
Kim, J., Casa, D., Said, A., Krakora, R., Kim, B. J., Kasman, E., Huang, X. & Gog, T. (2018). Sci. Rep. 8, 1958. Web of Science CrossRef PubMed Google Scholar
Kim, J., Casa, D., Upton, M. H., Gog, T., Kim, Y.-J., Mitchell, J. F., van Veenendaal, M., Daghofer, M., van den Brink, J., Khaliullin, G. & Kim, B. J. (2012a). Phys. Rev. Lett. 108, 177003. Web of Science CrossRef PubMed Google Scholar
Kim, J., Daghofer, M., Said, A., Gog, T., van den Brink, J., Khaliullin, G. & Kim, B. (2014). Nat. Commun. 5, 4453. Web of Science CrossRef PubMed Google Scholar
Kim, J., Said, A. H., Casa, D., Upton, M. H., Gog, T., Daghofer, M., Jackeli, G., van den Brink, J., Khaliullin, G. & Kim, B. J. (2012b). Phys. Rev. Lett. 109, 157402. Web of Science CrossRef PubMed Google Scholar
Kim, J.-K., Casa, D., Huang, X., Gog, T., Kim, B. J. & Kim, J. (2020). J. Synchrotron Rad. 27, 963–969. Web of Science CrossRef CAS IUCr Journals Google Scholar
Kirkpatrick, P. & Baez, A. V. (1948). J. Opt. Soc. Am. 38, 766–774. CrossRef PubMed CAS Web of Science Google Scholar
Masciovecchio, C., Bergmann, U., Krisch, M., Ruocco, G., Sette, F. & Verbeni, R. (1996). Nucl. Instrum. Methods Phys. Res. B, 111, 181–186. CrossRef CAS Web of Science Google Scholar
Moretti Sala, M., Martel, K., Henriquet, C., Al Zein, A., Simonelli, L., Sahle, C., Gonzalez, H., Lagier, M.-C., Ponchut, C., Huotari, S., Verbeni, R., Krisch, M. & Monaco, G. (2018). J. Synchrotron Rad. 25, 580–591. Web of Science CrossRef CAS IUCr Journals Google Scholar
Moretti Sala, M., Schnells, V., Boseggia, S., Simonelli, L., Al-Zein, A., Vale, J. G., Paolasini, L., Hunter, E. C., Perry, R. S., Prabhakaran, D., Boothroyd, A. T., Krisch, M., Monaco, G., Rønnow, H. M., McMorrow, D. F. & Mila, F. (2015). Phys. Rev. B, 92, 024405. Web of Science CrossRef Google Scholar
Owen, R. L., Holton, J. M., Schulze-Briese, C. & Garman, E. F. (2009). J. Synchrotron Rad. 16, 143–151. Web of Science CrossRef CAS IUCr Journals Google Scholar
Pincini, D., Vale, J. G., Donnerer, C., de la Torre, A., Hunter, E. C., Perry, R., Moretti Sala, M., Baumberger, F. & McMorrow, D. F. (2017). Phys. Rev. B, 96, 075162. Web of Science CrossRef Google Scholar
Porras, J., Bertinshaw, J., Liu, H., Khaliullin, G., Sung, N. H., Kim, J.-W., Francoual, S., Steffens, P., Deng, G., Sala, M. M., Efimenko, A., Said, A., Casa, D., Huang, X., Gog, T., Kim, J., Keimer, B. & Kim, B. J. (2019). Phys. Rev. B, 99, 085125. Web of Science CrossRef Google Scholar
Rueff, J.-P., Ablett, J. M., Céolin, D., Prieur, D., Moreno, T., Balédent, V., Lassalle-Kaiser, B., Rault, J. E., Simon, M. & Shukla, A. (2015). J. Synchrotron Rad. 22, 175–179. Web of Science CrossRef CAS IUCr Journals Google Scholar
Said, A. H., Sinn, H. & Divan, R. (2011). J. Synchrotron Rad. 18, 492–496. Web of Science CrossRef IUCr Journals Google Scholar
Shvyd'ko, Y., Hill, J., Burns, C., Coburn, D., Brajuskovic, B., Casa, D., Goetze, K., Gog, T., Khachatryan, R., Kim, J.-H., Kodituwakku, C., Ramanathan, M., Roberts, T., Said, A., Sinn, H., Shu, D., Stoupin, S., Upton, M., Wieczorek, M. & Yavas, H. (2013). J. Electron Spectrosc. Relat. Phenom. 188, 140–149. CAS Google Scholar
Verbeni, R., Kocsis, M., Huotari, S., Krisch, M., Monaco, G., Sette, F. & Vanko, G. (2005). J. Phys. Chem. Solids, 66, 2299–2305. Web of Science CrossRef CAS Google Scholar
Xu, R. & Chiang, T. C. (2005). Z. Kristallogr. 220, 1009–1016. CrossRef CAS Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.