research papers
In situ/operando plug-flow fixed-bed cell for synchrotron PXRD and investigations at high temperature, pressure, controlled gas atmosphere and ultra-fast heating
aTechnische Universität Berlin, Faculty III Process Sciences, Institute of Materials Science and Technology, Chair of Advanced Ceramic Materials, Straße des 17 Juni 135, 10623 Berlin, Germany, bTechnische Universität Berlin, Faculty III Process Sciences, Institute for Optic and Atomic Physics, Straße des 17 Juni 135, 10623 Berlin, Germany, cTechnische Universität Berlin, Faculty II Mathematik und Naturwissenschaften, Institut für Chemie, Straße des 17 Juni 135, 10623 Berlin, Germany, dDeutsches Elektronen-Synchrotron (DESY), Notkestrasse 85, 22607 Hamburg, Germany, and eHelmholtz-Zentrum Berlin für Materialien und Energie, 14109 Berlin, Germany
*Correspondence e-mail: maged.bekheet@ceramics.tu-berlin.de
A plug-flow fixed-bed cell for synchrotron powder X-ray diffraction (PXRD) and X-ray absorption fine structure (XAFS) idoneous for the study of heterogeneous catalysts at high temperature, pressure and under gas flow is designed, constructed and demonstrated. The operating conditions up to 1000°C and 50 bar are ensured by a set of mass flow controllers, pressure regulators and two infra-red lamps that constitute a robust and ultra-fast heating and cooling method. The performance of the system and cell for carbon dioxide hydrogenation reactions under specified temperatures, gas flows and pressures is demonstrated both for PXRD and
at the P02.1 (PXRD) and the P64 (XAFS) beamlines of the Deutsches Elektronen-Synchrotron (DESY).Keywords: plug-flow fixed-bed cell; ultra-fast infra-red heating; gas–solid interactions; in situ X-ray diffraction; in situ X-ray absorption fine structure.
1. Introduction
The study of heterogeneous catalysts using synchrotron-based in situ techniques is at the core of rational catalyst design. A detailed picture of a catalytic system can be provided by a combination of electron-based (mainly electron microscopy) and photon-based methods like X-ray diffraction (XRD), (XPS) and X-ray absorption fine structure (XAFS) (Bergmann & Roldan Cuenya, 2019). Comparing physical and chemical data with the performance in terms of reaction rates, selectivity and stability allows for elucidating reaction mechanisms, the active site and the deactivation mechanisms (Gili et al., 2018; Bonmassar et al., 2020). Establishing the structure–activity correlations is impossible without accessing structural information of the working catalysts under real operating conditions.
In order to perform in situ investigations at a beamline, one must ensure proper interaction of the target materials with the reactive atmosphere at a controlled and homogeneous temperature and pressure while the materials are exposed to a beam of photons. Thus, the design and operation of such tools is not trivial. The choice of materials is of uttermost importance: they must be capable of holding the desired reaction conditions while being inert and allowing the transmission of photons.
Several cells for powder X-ray diffraction (PXRD) or et al., 2014; Hansen et al., 2015; Jensen et al., 2010; Schlicker et al., 2018a; Pandit et al., 2022). Static gas and dead-end configurations (Schlicker et al., 2018a) do not grant realistic gas–solid interactions as usually occur on plug-flow fixed-bed cells (Clausen et al., 1991). Moreover, it is highly desirable to achieve elevated pressures, especially for CO2 hydrogenation reactions.
have been developed at different synchrotrons and allow for a wide variety of testing conditions (AndrieuxThe current work describes the design, construction and demonstration of a plug-flow fixed-bed setup and cell for the study of heterogeneous catalysts at elevated temperature, pressure and gas atmosphere. This cell is a continuation of our work done previously at the Advanced Light Source of the LBNL, Berkeley, USA (Schlicker et al., 2018a; Doran et al., 2017). It is upgraded in terms of the flow model (plug-flow) and the reachable pressure. The heating system, performed with a SiC tube acting as a furnace and two infra-red (IR) lamps, provides an extremely fast and robust heating method (Doran et al., 2017). The cell is demonstrated regarding (i) temperature using Pt and Al2O3 calibrants, (ii) pressure using a Mg(OH)2 dehydroxylation and carbonation reaction and (iii) for both in situ and PXRD following structural changes of a PdO/rh-In2O3 system under hydrogen flow (where rh indicates the rhombohedral form). The main goal of the current cell is to provide in situ structural data of heterogeneous catalysts in general: we have selected CO2 hydrogenation reactions as case examples due to their importance for the energy transition. The cell has been operated at two different beamlines of the PETRA III synchrotron at DESY, Hamburg, Germany: the P02.1 (PXRD) (Dippel et al., 2015) and the P64 (XAFS) (Caliebe et al., 2019) beamlines.
2. Equipment design and implementation
2.1. General technical description
The overall system is composed of (i) the upstream (gas flow control system), (ii) the reaction cell and (iii) the downstream. The pipes and instrumentation diagram (P&ID) can be seen in Fig. 1.
The upstream (i) contains three EL-FLOW Prestige mass flow controllers (MFCs 01–03; Bronkhorst, The Netherlands) connected to gas cylinders that enable the precise control of reactant gas composition and flow rate. Check-valves (CVs 01–03; Swagelok, USA) downstream of the MFCs prevent backward diffusion of gas. A digital pressure indicator (P-01; Omega, UK) allows monitoring of the pressure during the experiments. A stainless-steel high-pressure proportional relief valve (SV-01; Swagelok, USA) set to a burst pressure of 65 bar upstream of the plug-flow unit enables safe ventilation of the gas mixture to the exhaust should the system exceed its maximum pressure (e.g. through a sudden increase in reaction pressure due to chemical reaction). The reaction cell (ii) is described below in detail. The downstream (iii) of the system is composed of a particle filter with a grit of 40 µm (PF-01; FTSS series tee-type, Fitok, Germany), which prevents particles from moving downstream, potentially damaging the pressure regulator. Pressure control is enabled by a combination of an EL-PRESS pressure regulator (PIC P-02; Bronkhorst, The Netherlands) and an Equilibar research series back-pressure regulator (E-01; Equilibar, USA) limited to a pressure of 103.4 bar (1500 psi) and 150°C. The reason for using this combination for pressure regulation is to be able to heat up (∼105°C) the piping between the cell and the (MS), thereby avoiding/reducing condensation of light compounds, such as methanol or water. It is important to note that P-02 does not measure or control (directly) the pressure inside the cell, it only measures and controls the pressure that E-01 receives, thus, the use of PI P-01 was necessary to read the real pressure in the cell and monitor potential leaks. An MS or another analytic device can be connected to obtain quantitative data on the performance of the catalysts. A video-surveyed bubble counter (B-01) provides visual confirmation of the gas flow. Both the analytics and bubbler outlets are directed toward the exhaust. All system parameters are controlled and recorded using a self-made Labview program (Bitter et al., 2006).
2.2. Plug-flow fixed-bed reactor cell
The cell design was similar to some previously applied for diffraction studies (Andrieux et al., 2014). The sample is placed inside a fused silica tube [part 6 in Fig. 2(d), quartz glass; Hilgenberg, Germany] or a sapphire single-crystal tube (kindly provided by Crytur, Czech Republic). Both materials are reported to be extremely inert for the reactions of interest: amorphous quartz tubes are preferred as they do not add the typical diffraction spot of sapphire single crystals on the detector during PXRD. The tubes are opened at both ends, enabling a plug-flow mode inside the reactor cell. Dimensions of the tubing are 100 ± 0.5 mm in length, 1 ± 0.1 mm inner diameter and 1.5 ± 0.1 mm outer diameter (see §2.4 for an explanation of the sizing). The positioning of the sample inside the tube can be schematically seen in Fig. 2(e). At the exact center of the length of the tube, a 1 mm-long quartz wool segment is placed. Upstream of this quartz wool segment, 1–2 mm of the sample is placed, followed by a second 1 mm quartz wool segment. Downstream of this `quartz wool – sample – quartz wool' assembly, a 0.5 mm-diameter thin K-type thermocouple (TC; Reckmann, Germany) is placed [5 in Fig. 2(d)]. The symmetry of this design places the TC tip and the sample at the exact same distance from the center of the tube, minimizing temperature differences. The hot TC can catalytically induce gas phase composition changes; to ensure controlled gas composition exposure of the sample, the TC is placed downstream of the sample. Finally, the TC also helps to hold the sample in position, although sample displacement can be avoided by mitigating sudden increases in gas flows. Mixed vespel/graphite ferrules [3 in Fig. 2(d); Mascom, Germany) are drilled to enclose the tube and seal it using Swagelok nuts [4 in Fig. 2(d)] and two T-pieces from Swagelok [2 in Fig. 2(d)]. These are plugged into Swagelok quick connects [1 in Fig. 2(d)], which allow the cell to be connected to the setup's piping without screwing, preventing the tube breaking by torque. The total weight of the cell [with the support plate, Fig. 2(b), bottom big metal place with `p02.1 coma' inscription] is 1.5 kg. Sample changes of the cell assembly are made by detaching the transport plate [Fig. 2(a), see arrow; 9 in Fig. 2(d)]. Two small metal plates [8 in Fig. 2(d)] fix the montage on the transport plate and prevent the tube from breaking during transport.
External to the reactor tube, a SiC tube acting as a furnace is assembled [Fig. 2(c), 7 in Fig. 2(d), original unmodified tubing kindly provided by Fraunhofer IKTS]. Note that this tube does not withstand any gas pressure from the cell. Two holes (X-ray entrance 1 mm × 1 mm and exit 2 mm × 2 mm) were cut with diamond cutting disks with sizes [see Fig. 2(c)] designed to allow for the X-ray beam to interact with the sample and travel toward the detector without interacting with the SiC tube. Two metallic U-shaped pieces [Fig. 2(b)] hold the SiC tubing, and rotation of the tube is prevented by softly placing the tip of a screw on top of a flat indent made on the SiC tubing of 8 mm in length [Fig. 2(c)]. SiC is the ideal material to act as a furnace due to its hardness and (Schlicker et al., 2018a; Doran et al., 2017). It heats up by being illuminated with two 64635 HLX halogen lamps (Osram, Germany). The wall of the SiC is placed at the focal point of the lamp (19.5 mm), and the TC is used for the control loop. A schematic sketch of the cell can be seen in Fig. 2(a), and in the explosion views in Figs. S1 and S2 of the supporting information.
2.3. Temperature measurements at the sample position
Fig. 3(a) shows the temperature measured by a TC placed at the sample position in the cell as a function of time at 50 bar (abs) and a constant N2 flow rate of 5 Nml min−1. A set point of 1000°C was selected, and the initial heating rate [for the room-temperature (rt) to 500°C range] shows the maximum heating rate of the setup, approximately 20°C s−1. As the temperature of the cell increases, so does the heat loss, and the heating rate progressively decreases until the cell reaches 965°C. Repeated use of the SiC tube progressively oxidizes the SiC, decreasing its Using a new SiC tube allows a temperature of 1000°C to be reached (see Fig. S3 of the supporting information). After a few seconds, the set point was selected to be 25°C and the temperature was recorded. Similarly, the maximum cooling rate is 20°C s−1 for the initial period. As no external cooling is applied, the cooling speed decreases as the temperature gradient between the cell and the environment decreases. This figure proves the ultra-fast heating and cooling concept of the cell. Fig. 3(b) shows the T-profile in the axial direction of the tube. It was achieved by displacing a second TC while maintaining the first TC (used for the set-point) in the center of the tube. We assume that these T-profiles are symmetric in the other axial direction.
2.4. Safety considerations
The fused silica tube sizing and wall thickness were calculated by applying the von Mises criterion using the same assumptions as Andrieux et al. (2014) and applying a safety factor of ×10. The robustness of several tubes was assessed at different temperatures, pressures, heating and cooling rates, and pressurizing rates before using them in the synchrotron experiments.
The system's safety measures include the following. The MFCs are normally closed; in case of malfunction, gas flow into the system would stop. The pressure regulator concept is normally open: in case of malfunction, it would open and prevent a build-up of uncontrolled pressure. Check valves are used to prevent back-mixing of the gas flows. The piping sizing is very small to minimize the total volume of gas inside the system. The total volume between the MFCs and the equilibar is ∼25 ml. The stainless-steel high-pressure proportional relief valve (SV-01) prevents the system from reaching pressures above a safety threshold of 65 bar. All standard system components (except the quartz tubes) are certified for the maximum working pressure multiplied by a safety factor of (at least) 1.3. Prior to the operation, a burst test was performed at 1.3× maximum operating pressure, and a leak test was performed at mild pressures (∼5 bar) prior to each single experiment.
The current cell was designed to perform 2 hydrogenation: the design system operating window is summarized in Table 1.
with a special focus on CO2.5. Beamline description and data analysis (Dioptas, Rietveld refinement)
This system and cell were installed and tested on two different beamlines of the PETRA III synchrotron. All data shown in the present work were obtained using the quartz capillaries. The P02.1 beamline for powder diffraction and total scattering (Dippel et al., 2015) was operated at 60 keV (0.207 Å) with a sample–detector distance of 1200 mm in center ring configuration on a Varex 4343CT area detector. The beam size used was 0.7 mm × 0.7 mm. The cell [Fig. 2(b)] is assembled on the beamline's XYZ Huber stage to allow for fine alignment with respect to the beam and the SiC tube/sample. Note that the current design does not allow for rotation in any directions. The 2D images from the detector were integrated using Dioptas (Prescher & Prakapenka, 2015) software and plotted using Origin (Pro), Version 2022 (OriginLab Corporation, Northampton, MA, USA). was performed using FULLPROF software (Rodriguez-Carvajal, 1990). The instrument broadening parameters and the sample–detector distance were calibrated using an LaB6 NIST 660b standard. All pressures are absolute.
The et al., 2019). An Si(111) monochromator was used for scanning the energy. The measurements were performed in the fluorescence and transmission modes simultaneously. The transmission and fluorescence signals were recorded using an and passivated implanted planar silicon (PIPS) detector. The data displayed in this work were obtained in fluorescence mode. Before normalization measurement artifacts were removed manually. The normalization of the spectra was performed with the flattening algorithm of ATHENA (part of the Demeter software package) (Ravel & Newville, 2005) using a first-degree polynomial for the pre-edge line and a second-degree polynomial for the post-edge line.
measurements were performed at the P64 beamline, DESY (Caliebe3. Experimental results
The applicability of the cell was demonstrated on three different levels: (i) temperature, (ii) pressure and (iii) for both in situ and PXRD.
3.1. of platinum and alumina
The temperature at the sample position in the cell was demonstrated by performing in situ PXRD experiments on two different standard materials, platinum black (≥99.9%; Sigma-Aldrich) and α-Al2O3 corundum NIST (∼99.02% ± 1.11%; Sigma-Aldrich). The lattice parameters of Pt and α-Al2O3 were determined from the of PXRD data and compared with the reported values as a function of temperature in the literature. Fig. 4 shows the lattice parameter of both samples heated from 25°C to 950°C at 50/100°C intervals and under Ar flow of 10 Nml min−1 as a function of the recorded temperature of the TC. For comparison, the lattice parameters of platinum are also calculated from its coefficient as reported by Kirby (1991), while the parameter a of α-Al2O3 is calculated according to first-principle calculations (Reeber & Wang, 2000) and previous PXRD results (Fiquet et al., 1999). As shown in Fig. 4, both platinum and Al2O3 lattice parameters are in excellent agreement with the theoretical lattice parameter values (Kirby, 1991; Reeber & Wang, 2000; Fiquet et al., 1999). The difference between the measured temperature by the TC and the determined sample temperature derived from the platinum lattice parameter, used as internal temperature calibration for the system, is below 20°C in the full temperature range. Thus, the cell and T-control system provides a very accurate temperature in the sample zone.
3.2. Dehydroxylation and carbonation of Mg(OH)2 in pure CO2
To demonstrate the performance of the setup under elevated pressure and temperature, the following dehydroxylation [equation (1)] and carbonation [equation (2)] reactions of Mg(OH)2 were investigated:
These two reactions have been reported to be temperature- and pressure-dependent (Pfeiffer, 2011; Fagerlund et al., 2012).
Mg(OH)2 powder was filled into a quartz capillary, and the system was first purged with pure CO2 for 10 min to remove any residual air; then, the pressure was increased to the target pressure (i.e. 1.5 bar and 35 bar) using a total gas flow of 25 mL min−1. After pressure stabilization in the system, the temperature was increased to 495°C at 10°C min−1 without gas flow to simulate equilibrium conditions. Temperature and pressure were then kept constant for 1 h while changes in the were monitored by acquiring one pattern every 30 s; the results are displayed in Fig. 5. As shown in Figs. 5(a) and 5(b), although Mg(OH)2 starts transforming first into MgO during heating in both experiments, the dehydroxylation process starts at ∼410°C at 1.5 bar and shifts to a higher temperature (∼450°C) at 35 bar. These results are in line with previous studies that reported the increase in the dehydroxylation temperature of Mg(OH)2 by 40–45°C by increasing the pressure from 1 to 35 bar using N2 as pressurizing gas. However, the dehydroxylation starts at a slightly higher temperature in our experiments, which agrees with previous studies reported that Ca(OH)2 and Mg(OH)2 become stable at significantly higher temperatures in the presence of CO2 (Fagerlund et al., 2012; Materić et al., 2010).
Shortly after reaching 495°C at 1.5 bar, the Mg(OH)2 has fully reacted to MgO, and no other reaction product is formed. In contrast, the formed MgO during heating at 35 bar is immediately reacted with CO2 to form oxymagnesite MgO·2MgCO3, which is usually observed at relatively high pressures during the carbonation of Mg(OH)2 (Fagerlund et al., 2012). This phase disappears after holding the temperature at 495°C for 2 min while the full transformation of the remaining Mg(OH)2 phase into MgO and MgCO3 takes place after 15 min. analysis reveals that the recovered sample at room temperature after carbonation at 495°C and 35 bar for 60 min is composed of 43.6 ± 0.9 wt% MgO and 56.4 ± 1.1 wt% MgCO3. These weight fraction values agree with previously reported values at the same temperature and pressure conditions (Fagerlund et al., 2012), confirming that the pressure in the cell is equal to the set pressure by the pressure control system.
3.3. Combined in situ PXRD and study
We studied the structural and chemical changes of reactive metal-support interaction (RMSI) effects occurring on PdO/rh-In2O3 material during H2 reduction separately with PXRD and using the same cell. The data generated with both techniques can be combined to obtain a more thorough description of a catalyst in situ. This RMSI results in the formation of irreversible intermetallic compounds (IMCs) in the PdO/rh-In2O3 system, influencing its catalytic and gas-sensing properties (Ovsianytskyi et al., 2023). Our previous in situ synchrotron PXRD studies showed that 5 wt% PdO-loaded rh-In2O3 undergoes a sequence of phase transformations upon heating under a pure H2 atmosphere at 1 bar and up to 500°C (Ovsianytskyi et al., 2023; Schlicker et al., 2018b). In this experiment we studied the influence of H2 on the sequence of these phase transformations. Fig. 6 shows the in situ PXRD patterns and In K-edge X-ray absorption near-edge structure (XANES) for PdO/rh-In2O3 material obtained during pressurizing using pure H2 with flow rate of 25 mL min−1 up to 15 bar, followed by heating from room temperature to 450°C at 15 bar after decreasing the flow rate of H2 to 5 mL min−1. In situ PXRD results [Fig. 6(a)] reveal the formation of the PdHx phase with increasing pressure of H2 to 15 bar at rt, which can be explained by the hydrogenation of Pd metal present in the sample. However, no structural and chemical changes are observed for the rh-In2O3 phase at these conditions, as indicated from in situ PXRD and results [Fig. 6(b)]. At 15 bar, the interaction between Pd and rh-In2O3 starts with increasing the temperature, leading to the formation of IMCs such as InPd (at 210°C), In3Pd2 (260°C) and In3Pd (340°C). At 15 bar and 400°C, only the In3Pd phase and liquid In metal (recrystallized during cooling) are present in the sample. It is worth noting that our previous in situ PXRD study on the same material showed that, at 1 bar of pure H2, a sequence of PdO reduction to Pd metal followed by PdH0.706 hydride formation and subsequently InPd and In3Pd2 IMC formation occurred between 30°C and 500°C (Ovsianytskyi et al., 2023; Schlicker et al., 2018b). According to the In–Pd phase diagram, InPd and In3Pd2 are among the thermodynamically most stable compounds, thus it can be expected that these two phases are predominantly formed at 1 bar. The In3Pd phase was not observed at 1 bar in our previous work in the temperature range 30–500°C, suggesting the metastability of this IMC. Thus, increasing the pressure to 15 bar could stabilize the formation of this phase.
The in situ PXRD results are in perfect agreement with the results [Fig. 6(d)], which reveal the shift of the In K-edge position to lower energies upon heating due to the reduction of rh-In2O3 and the formation of IMCs. Moreover, the normalized In K-edge XANES spectra at 15 bar and 400°C agree with that of metallic indium (Maeno et al., 2019). These results confirm the applicability of our setup for both in situ synchrotron PXRD and experiments, allowing for investigating the structural and chemical states of several material systems in a single environment for different applications.
4. Conclusion
A setup and cell for the in situ/operando characterization with PXRD and under high temperature, elevated pressure and controllable gas atmosphere has been designed, constructed and demonstrated at two beamlines of the DESY/German synchrotron in Hamburg. The cell acts as plug-flow fixed-bed reactor suitable for monitoring chemical reactions on heterogeneous catalysts and uses two infrared lamps and an SiC tube as an ultra-fast and robust heating system. We demonstrated the functionality of this device in three steps: the heating capability and temperature accuracy based on Pt and Al2O3 standards; a MgO carbonatization experiment performed up to 35 bar showing the capacity of operating at elevated pressure; a Pd/rh-In2O3 system studied both with in situ PXRD and The cell and setup hereby shown are suitable for the characterization of heterogeneous catalysts (with special focus on CO2 hydrogenation).
Supporting information
Supporting Figures S1, S2 and S3. DOI: https://doi.org/10.1107/S1600577523009591/vl5016sup1.pdf
Footnotes
‡These authors contributed equally to this paper.
Acknowledgements
We acknowledge DESY (Hamburg, Germany), a member of the Helmholtz Association HGF, for the provision of experimental facilities. Parts of this research were carried out at PETRA III and we would like to thank Alexander Schökel, Marcel Görlitz and Wolfgang A. Caliebe for assistance in using the P02.1 and P64 beamlines. Beam time was allocated for proposal LTP II-20210010. We kindly thank Delf Kober, Peter Schneppmüller and Kai Jens Weisser. Open access funding enabled and organized by Projekt DEAL.
Funding information
This work was funded by the Deutsche Forschungsgemeinschaft (DFG German Research Foundation) under Germany's Excellence Strategy – EXC2008-390540038 – UniSysCat. A. Gili acknowledges support from the German Federal Ministry of Education and Research in the framework of the project Catlab (03EW0015A/B). Emiliano Dal Molin was funded by the Einstein Foundation Berlin and UniSysCat.
References
Andrieux, J., Chabert, C., Mauro, A., Vitoux, H., Gorges, B., Buslaps, T. & Honkimäki, V. (2014). J. Appl. Cryst. 47, 245–255. Web of Science CrossRef CAS IUCr Journals Google Scholar
Bergmann, A. & Roldan Cuenya, B. (2019). ACS Catal. 9, 10020–10043. Web of Science CrossRef CAS Google Scholar
Bitter, R., Mohiuddin, T. & Nawrocki, M. (2006). LabVIEW: Advanced Programming Techniques. CRC Press. Google Scholar
Bonmassar, N., Bekheet, M. F., Schlicker, L., Gili, A., Gurlo, A., Doran, A., Gao, Y., Heggen, M., Bernardi, J., Klötzer, B. & Penner, S. (2020). ACS Catal. 10, 1102–1112. Web of Science CrossRef CAS Google Scholar
Caliebe, W. A., Murzin, V., Kalinko, A. & Görlitz, M. (2019). AIP Conf. Proc. 2054, 060031. Google Scholar
Clausen, B., Steffensen, G.; Fabius, B., Villadsen, J., Feidenhans'l, R. & Topsøe, H. (1991). J. Catal. 132, 524–535. CrossRef CAS Web of Science Google Scholar
Dippel, A.-C., Liermann, H.-P., Delitz, J. T., Walter, P., Schulte-Schrepping, H., Seeck, O. H. & Franz, H. (2015). J. Synchrotron Rad. 22, 675–687. Web of Science CrossRef CAS IUCr Journals Google Scholar
Doran, A., Schlicker, L., Beavers, C. M., Bhat, S., Bekheet, M. F. & Gurlo, A. (2017). Rev. Sci. Instrum. 88, 013903. Web of Science CrossRef PubMed Google Scholar
Fagerlund, J., Highfield, J. & Zevenhoven, R. (2012). RSC Adv. 2, 10380. Google Scholar
Fiquet, G., Richet, P. & Montagnac, G. (1999). Phys. Chem. Miner. 27, 103–111. Web of Science CrossRef CAS Google Scholar
Gili, A., Schlicker, L., Bekheet, M. F., Görke, O., Penner, S., Grünbacher, M., Götsch, T., Littlewood, P., Marks, T. J., Stair, P. C., Schomäcker, R., Doran, A., Selve, S., Simon, U. & Gurlo, A. (2018). ACS Catal. 8, 8739–8750. Web of Science CrossRef CAS Google Scholar
Hansen, B. R. S., Møller, K. T., Paskevicius, M., Dippel, A.-C., Walter, P., Webb, C. J., Pistidda, C., Bergemann, N., Dornheim, M., Klassen, T., Jørgensen, J.-E. & Jensen, T. R. (2015). J. Appl. Cryst. 48, 1234–1241. Web of Science CrossRef CAS IUCr Journals Google Scholar
Jensen, T. R., Nielsen, T. K., Filinchuk, Y., Jørgensen, J.-E., Cerenius, Y., Gray, E. M. & Webb, C. J. (2010). J. Appl. Cryst. 43, 1456–1463. Web of Science CrossRef CAS IUCr Journals Google Scholar
Kirby, R. K. (1991). Int. J. Thermophys. 12, 679–685. CrossRef CAS Web of Science Google Scholar
Maeno, Z., Yasumura, S., Liu, C., Toyao, T., Kon, K., Nakayama, A., Hasegawa, J. & Shimizu, K. (2019). Phys. Chem. Chem. Phys. 21, 13415–13427. Web of Science CrossRef CAS PubMed Google Scholar
Materić, V., Edwards, S., Smedley, S. I. & Holt, R. (2010). Ind. Eng. Chem. Res. 49, 12429–12434. Google Scholar
Ovsianytskyi, O., Bekheet, M. F., Gili, A. & Gurlo, A. (2023). ChemPhysChem, 24, e202200775. Web of Science CrossRef PubMed Google Scholar
Pandit, L., Serrer, M., Saraci, E., Boubnov, A. & Grunwaldt, J. (2022). Chem. Methods, 2, e202100078. Web of Science CrossRef Google Scholar
Pfeiffer, H. (2011). Thermochim. Acta, 525, 180–182. Web of Science CrossRef CAS Google Scholar
Prescher, C. & Prakapenka, V. B. (2015). High. Press. Res. 35, 223–230. Web of Science CrossRef CAS Google Scholar
Ravel, B. & Newville, M. (2005). J. Synchrotron Rad. 12, 537–541. Web of Science CrossRef CAS IUCr Journals Google Scholar
Reeber, R. R. & Wang, K. (2000). MRS Proc. 622, T6.35.1. Google Scholar
Rodriguez-Carvajal, J. (1990). Abstract of the Satellite Meeting on Powder Diffraction of the XV Congress of the IUCr, Toulouse, France. Google Scholar
Schlicker, L., Bekheet, M. F., Gili, A., Doran, A., Gurlo, A., Ploner, K., Schachinger, T. & Penner, S. (2018b). J. Solid State Chem. 266, 93–99. Web of Science CrossRef CAS Google Scholar
Schlicker, L., Doran, A., Schneppmüller, P., Gili, A., Czasny, M., Penner, S. & Gurlo, A. (2018a). Rev. Sci. Instrum. 89, 033904. Web of Science CrossRef PubMed Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.