Atomic resolution cryo-EM at 200 keV

crossmark logo

aGraduate School of Medicine, The University of Tokyo, 7-3-1 Hongo, Bunkyo-ku, Tokyo, 113-0033, Japan, bResearch Center for Advanced Science and Technology, The University of Tokyo, 4-6-1 Komaba, Meguro-ku, Tokyo, 153-8904, Japan, and cDECTRIS Ltd, Taefernweg 1, 5405 Baden-Daettwil, Switzerland
*Correspondence e-mail: [email protected]

Edited by W. Kühlbrandt, Max Planck Institute of Biophysics, Germany (Received 16 January 2026; accepted 18 April 2026; online 10 June 2026)

This article is part of the special issue CryoEM in the Fast Lane of Structural Biology.

Atomic resolution in cryo-electron microscopy was first demonstrated six years ago. This was accom­plished using 300 kV electron microscopes equipped with new hardware that provided narrower energy spread, aberration correction and energy filtering. Here, we report the achievement of 1.24 Å atomic resolution on an upgraded 200 kV electron microscope featuring a cold field emission gun, a high-resolution objective lens polepiece and an energy filter. These com­ponents transform the instrument into a cost-effective single-particle cryo-EM platform with performance com­parable to that of significantly more expensive 300 kV systems. The microscope can also be operated at 100 kV and by using a high-speed hybrid-pixel detector we were able to reach sub-2 Å resolution.

1. Introduction

The cryo-EM `resolution revolution' has been ongoing for more than a decade. It reached a major milestone in 2020 with the demonstration of atomic resolution by two independent teams (Nakane et al., 2020View full citation; Yip et al., 2020View full citation). According to the current consensus in X-ray crystallography, `atomic resolution' is defined as 1.2 Å or better (Dauter, 2003View full citation), based on a criterion of non-overlapping atomic densities proposed by Sheldrick (1990View full citation).

The first atomic resolution cryo-EM results were obtained using well-behaved highly symmetric heavy-chain apoferritin, which has served as the gold-standard test specimen for cryo-EM single-particle performance since its introduction by Russo & Passmore (2014View full citation) and its subsequent improvement to a heavy-chain variant (Danev et al., 2019View full citation). The two atomic resolution teams used upgraded 300 kV microscopes: one featured a cold field emission gun (CFEG) and a newly developed energy filter (Nakane et al., 2020View full citation), and the other a monochromator and a spherical aberration corrector (Yip et al., 2020View full citation). Since these reports, other groups have also pre­sent­ed atomic resolution results on 300 kV instruments, almost all equipped with CFEG sources (Zhang et al., 2020View full citation; Fujita et al., 2023View full citation; Maki-Yonekura et al., 2023View full citation; Küçükoğlu et al., 2024View full citation; Croy et al., 2025View full citation). In 2020, we tested the newly installed Titan Krios G4 (Thermo Fisher Scientific, Waltham, USA) 300 kV electron microscope at the University of Tokyo, equipped with a Schottky FEG (SFEG), and the dataset reached 1.31 Å (Danev et al., 2021View full citation). At present, 300 kV microscopes offer the best cryo-EM performance for both single-particle analysis and cryo-tomography. However, they are expensive to purchase, costly to maintain, heavy in electricity consumption and require a large installation space.

The cryo-EM capabilities of 200 kV microscopes have been extensively evaluated over the years with remarkable results (Herzik et al., 2017View full citation; Herzik et al., 2019View full citation; Wu et al., 2020View full citation; Merk et al., 2020View full citation; Hamdi et al., 2020View full citation; Kayama et al., 2021View full citation; Feathers et al., 2021View full citation; Gerle et al., 2022View full citation; Thangaratnarajah et al., 2022View full citation; Koh et al., 2022View full citation; Jia et al., 2024View full citation; Cushing et al., 2024View full citation). Surprisingly, despite their excellent performance, only ∼10% of deposited maps with resolutions better than 6 Å in the Electron Microscopy Data Bank (EMDB; https://www.ebi.ac.uk/emdb/) come from 200 kV microscopes. Currently, 200 kV machines are widely used as a cost-effective alternative to 300 kV instruments for sample screening and data collection at universities, research institutions and industrial facilities, with the understanding that the achievable resolution may not be as high but is often sufficient to answer most research questions.

In the past few years, 100 kV microscopes were shown to hold substantial promise as `people's cryo-microscopes', offering sufficient performance for sample screening and exploratory data collection at an affordable cost and with minimal installation requirements (Naydenova et al., 2019View full citation; McMullan et al., 2023View full citation; Chan et al., 2024View full citation; Venugopal et al., 2025View full citation; Karia et al., 2025View full citation). However, achieving high-resolution results on current commercially available 100 kV instruments remains challenging, as evidenced by the fact that there are fewer than 50 maps at resolutions better than 6 Å in the EMDB, the majority of which are test samples.

Here, we evaluated the performance of an upgraded 200 kV microscope equipped with a CFEG, a narrow gap objective lens polepiece and an omega-type energy filter. We collected apoferritin test datasets at accelerating voltages of 200 and 100 kV using a latest generation direct detection camera and a high-speed hybrid-pixel detector, respectively. The results are pre­sent­ed below.

2. Materials and methods

2.1. Microscope configuration

The experiments were performed on a CRYO ARM 200 II (JEM-Z200CA, JEOL Ltd, Tokyo, Japan) electron cryo-microscope upgraded with a narrow-gap high-resolution objective lens polepiece. This reduced the spherical and chromatic aberration coefficients from the regular CS = 2.7 mm and CC = 2.8 mm to CS = 1.5 mm and CC = 1.8 mm. The microscope is equipped as standard with a CFEG, a three-lens condenser system and an omega-type energy filter. The system also featured a Gatan K3 (Gatan, Pleasanton, USA) direct electron detector and an electrostatic dose modulator (EDM), which was used as a fast electrostatic shutter. For the 100 kV experiments, the camera was replaced with a DECTRIS SINGLA (DECTRIS Ltd, Baden-Daettwil, Switzerland) high-speed hybrid-pixel detector.

2.2. Sample preparation

Mouse heavy-chain apoferritin was expressed and purified as described previously (Danev et al., 2021View full citation). Cryo-EM samples were prepared by applying 3 µl of 1.9 mg ml−1 sample solution on UltrAuFoil R0.6/1 300 mesh (200 kV experiments) or UltrAuFoil R1.2/1.3 300 mesh (100 kV experiments) grids (Russo & Passmore, 2016View full citation) (Qu­anti­foil Micro Tools GmbH, Jena, Germany) and plunge-freezing in liquid ethane on a Vitrobot Mark IV (Thermo Fisher Scientific, Waltham, USA), with blot time 20 s (R0.6/1.0 grids) or 10 s (R1.2/1.3 grids), chamber tem­per­a­ture 4 °C and 100% humidity.

2.3. Data collection

The datasets were collected automatically by SerialEM software (Schorb et al., 2019View full citation) using its built-in single-particle automation routines. The main acquisition parameters are summarized in Table S1 in the supporting information.

For the 200 kV dataset, the microscope was set up at an indicated magnification of 150000×, calibrated pixel size 0.3056 Å pixel−1, spot size 3, convergence angle 3, condenser aperture 100 µm, beam diameter ∼0.95 µm, no objective aperture, zero-loss energy filtering with a 20 eV slit and target defocus −0.3 to −0.5 µm. The detector was operated in counting (non-super-resolution) correlated double sampling (CDS) mode, with an exposure time of 1.51 s, an exposure rate of 3.3 e pixel−1 s−1, a total exposure of 53.4 e Å−2, a frame time of 0.0185 s, 81 frames and an exposure per frame of 0.66 e Å−2. In total, 13654 movies were collected in 56 h, with an average throughput of 244 movies per hour using a beam-tilt com­pensated 3 × 3 × 1 image-shift acquisition pattern with a single image in the centre of each hole.

For the 100 kV dataset, the microscope was set up at an indicated magnification of 500000×, calibrated pixel size 1.17 Å pixel−1, spot size 6, convergence angle 3, condenser aperture 70 µm, beam diameter ∼0.53 µm, objective aperture 250 µm, zero-loss energy filtering with a 20 eV slit and target defocus −0.45 to −0.55 µm. The movies were recorded in HDF5 format at the raw frame rate of the detector of 4500 frames s−1, with an exposure time of 3.0 s, an exposure rate of 23.8 e pixel−1 s−1, a total exposure of 52.2 e Å−2, 13500 total frames and an exposure per frame of 0.0039 e Å−2. In total, 17424 movies were collected in 49.5 h, with an average throughput of 352 movies per hour using a beam-tilt com­pensated 3 × 3 × 4 image-shift acquisition pattern with four images in each hole.

2.4. Data processing

The data were processed with CryoSPARC (Punjani et al., 2017View full citation) (Structura Biotechnology, Toronto, Canada).

The 200 kV dataset processing workflow is summarized in Fig. S1 in the supporting information. Briefly, 13654 movies were subjected to patch motion correction and patch CTF estimation, followed by exposure curation with a CTF fit resolution selection range between 2 and 4 Å, retaining 10189 micrographs. The micrographs were split into nine exposure groups based on their image shift. Particles were picked with 20 Å low-pass filtered templates generated from a previous 3D apoferritin map, resulting in 927k picks after selection, with an NCC score of greater than 0.4 and local power between 600 and 900. Extraction with a 160 pixel box at 1.57 Å pixel−1 produced 652k particles that were subjected to 2D classification with 100 classes. Manual 2D class selection left 635k particles that were run through ab initio 3D reconstruction with two classes. The 3D reference and 622k particles from the larger class were used for further processing. Initial 3D homogeneous refinement hit Nyquist at 3.23 Å. The particles were re-extracted with a 400 pixel box at 0.627 Å pixel−1 and another 3D homogeneous refinement reached 1.56 Å. The particles were then split into hourwise exposure groups for a total of 504 exposure groups, followed by a 3D homogeneous refinement which reached 1.35 Å. A round of reference-based motion correction and 3D homogeneous refinement improved the map to 1.26 Å. A 3D heterogeneous refinement with three classes reduced the particle stack to 615k, followed by another round of reference-based motion correction and 3D homogeneous refinement to produce the final map at 1.24 Å.

The 100 kV dataset processing workflow is summarized in Fig. S2. Briefly, the 17654 HDF5 hardware frame stacks were subjected to 2× super-resolution electron counting (Zambon, 2023View full citation) with a GPU-accelerated software tool provided by DECTRIS (DECTRIS Ltd, Baden-Daettwil, Switzerland) and were fractionated with 225 hardware frames per fraction into 60 frame MRC movies. The movies were renamed with a Python script according to image-shift group specifications in summed MRC images that were saved separately by SerialEM during acquisition. The renamed super-resolution MRC movies were imported into CryoSPARC and were subjected to patch motion correction with 3/4 Fourier crop and patch CTF estimation, followed by exposure curation with a CTF fit resolution selection range between 2.5 and 5 Å, retaining 9123 micrographs. Particles were picked with 20 Å low-pass filtered templates generated from a previous 3D apoferritin map, resulting in 477k picks after selection, with an NCC score of greater than 0.4 and local power between 8 × 106 and 13 × 106. Extraction with a 150 pixel box at 1.4 Å pixel−1 produced 350k particles that were subjected to 2D classification with 200 classes. Manual 2D class selection left 322k particles that were run through ab initio 3D reconstruction with three classes. The 3D reference and 321k particles from the largest class were used for further processing. Initial 3D homogeneous refinement reached 2.9 Å. The particles were re-extracted with a 250 pixel box at 0.936 Å pixel−1 and split into 36 exposure groups according to image shift. Another 3D homogeneous refinement reached 2.01 Å. A reference-based motion correction was attempted but did not improve the resolution while exhibiting unexpectedly smooth particle tracks and an anomalous overweighting of high-resolution spectral com­ponents in later movie frames. To circumvent that, patch motion correction of the movies was performed again by using only the first 30 frames. Particles were re-extracted from the limited-frame micrographs and a 3D homogeneous refinement reached 2.0 Å. A round of reference-based motion correction and 3D homogeneous refinement improved the map resolution to 1.93 Å. Splitting the particles into hourwise exposure groups and further exposure curation reduced the particle stack to 250k particles and improved the 3D homogeneous refinement resolution to 1.91 Å.

To generate the Rosenthal–Henderson plot (Rosenthal & Henderson, 2003View full citation) for the 200 kV dataset, random particle subsets from the final particle stack were subjected to 3D homogeneous refinement independently and the results were plotted in coordinates of reciprocal squared resolution versus logarithm of the number of particles. The B factor was calculated as 2 times (slope of linear fit)−1 from the linear regression through the data points.

Evaluation of the effect of radiation damage by pre-exposure on the achievable resolution was performed by using the first 1000 movies containing 58k particles from the 200 kV dataset, running motion correction jobs with omission of a varying number of initial movie frames (Table S2 in the supporting information), extracting the particles from the aligned micrographs and performing a 3D homogeneous refinement with defocus, nine image-shift exposure group beam tilt refinement and Ewald sphere correction. Reference-based motion correction and higher-order aberration refinements were not used in this processing.

2.5. Beam tilt stability measurement

To investigate the reason for the beam tilt drift that was observed in the 200 kV dataset, a long-term measurement of beam tilt was performed over a week. The sample stage and sample loader were at room tem­per­a­ture. The microscope was set at a magnification of 80000×, 0.568 Å pixel−1, spot size 4, convergence angle 3 and exposure rate 10.5 e pixel−1 s−1. The condenser and objective lenses were normalized, followed by coma-free alignment in SerialEM. A SerialEM script was used to automatically measure the beam tilt every 10 minutes using the built-in five image Zemlin tableau method (`FixComaByCTF 1' command). The script also performed a CFEG `high flash' every six hours.

2.6. Numerical estimation of the effect of pre-exposure on resolution

To qu­antify the resolution decrease as a function of pre-exposure, we must estimate the spectral signal-to-noise ratio (SNR) and dampen it with a resolution-dependent radiation damage model. It is not necessary to know the absolute but rather the relative spectral SNR, which can be calculated from the B factor fit in the Rosenthal–Henderson plot. Assuming that particles represent independent measurements, the SNR is inversely proportional to the square root of the number of particles necessary to reach a given resolution. Therefore, having a 3D refinement without pre-exposure, the relative SNR of lower resolution shells can be calculated as the square root of the particle number ratio. Then, by applying a resolution-dependent radiation damage model, the exposure needed to attenuate the relative SNR of a given resolution shell to 1 can be estimated. This will correspond to the exposure that will limit the 3D reconstruction to the resolution of that shell.

Using this approach and the empirical radiation damage model by Grant & Grigorieff (2015View full citation) adjusted for 200 kV by reducing the critical exposure by 25% (as suggested in the article), we estimated the resolution versus pre-exposure. The calculations were performed in Wolfram Mathematica (Wolfram Research, Champaign, USA).

2.7. Model refinement and visualization

The initial model (PDB-7A4M) was rigid-body fitted in Coot (Casañal et al., 2020View full citation), followed by reciprocal-space refinement using Servalcat (Yamashita et al., 2021View full citation) against unsharpened half maps. During Servalcat refinement, H atoms were placed strictly based on stereochemical restraints and did not participate in the map-based gradient calculation. The model was then inter­actively adjusted in Coot using the sharpened map and FoFc difference map. The hy­dro­gen-omit FoFc map was used to position rotatable H atoms and to determine histidine tautomerization states. Finally, a hy­dro­gen-omit FoFc difference map was calculated using Servalcat. The models were validated using MolProbity (Chen et al., 2010View full citation).

3D images of the maps and model for the figures were created using UCSF Chimera (Pettersen et al., 2004View full citation) and UCSF Chimera X (Meng et al., 2023View full citation).

3. Results

Fig. 1[link] shows normalized histograms of the resolution of Electron Microscopy Data Bank (EMDB) depositions from 300 and 200 kV microscopes in the last six years. The ratio of 200 to 300 kV depositions is approximately 1:9 (3058 versus 27825). Lognormal fits of the histograms indicate a 0.25 Å higher median resolution at 300 kV. This rather moderate resolution advantage cannot explain the big disparity in the number of depositions. Nevertheless, a closer look at the histograms shows that 300 kV is ∼1.7 times more prolific in the 2.5–3.0 Å range and ∼3.3 times more prolific in the 2.0–2.5 Å range. These resolution ranges, and especially the latter, are of increasing importance in cryo-EM studies because they enable more accurate modelling of side chains, identification and pose assignment of small mol­ecules, and high-confidence detection of water mol­ecules.

[Figure 1]
Figure 1
Resolution statistics of cryo-EM maps with resolution below 6 Å from 300 and 200 kV microscopes deposited in the Electron Microscopy Data Bank (EMDB) in the last six years (2020–2025). The resolution histograms (pink and light-blue bars) were normalized by the number of depositions. Lognormal fits of the histograms are shown with red and blue lines. The number of depositions and median resolution from the lognormal fits are listed in the text boxes.

Fig. 2[link] contains theoretical plots of the temporal coherence envelope of the contrast transfer function (CTF) [Equation (6.42) in Reimer & Kohl (2010View full citation)] for a number of cryo-microscope configurations. We used the general energy spread values provided by microscope manufacturers for the two FEG source types (Hamaguchi et al., 2019View full citation; Fislage et al., 2020View full citation; Nakane et al., 2020View full citation) and a theoretical formula assuming a Gaussian energy distribution. However, the actual energy distribution is not strictly Gaussian and the energy spread depends on the age, condition and emission parameters of each emitter (Ross & Walther, 2012View full citation). Therefore, Fig. 2[link] is just a basic qu­anti­tative com­parison. The top-of-the-line CRYO ARM 300 II (JEOL Ltd, Tokyo, Japan) and Krios 5 (Thermo Fisher Scientific, Waltham, USA) 300 kV machines (green solid line) have demonstrated atomic resolution in the past (Nakane et al., 2020View full citation; Yip et al., 2020View full citation; Maki-Yonekura et al., 2023View full citation) and are currently the highest optical performance commercially available cryo-microscopes. The CRYO ARM 200 II (pink solid line) used in this study is a close second and is presently the top performing 200 kV instrument. Other CFEG 200 kV microscopes, such as the CRYO ARM 200 (JEOL Ltd, Tokyo, Japan) and Glacios 2 (Thermo Fisher Scientific, Waltham, USA) (blue solid line), follow next. Remarkably, the recently pre­sent­ed `Dublin lens' design for 100 keV (Alves et al., 2025View full citation) (grey solid line), with a very low chromatic aberration coefficient of CC = 1.0 mm, if coupled with a CFEG will perform on a par with the CRYO ARM 200 and Glacios 2. SFEG-equipped instruments, such as the Krios, Glacios (Thermo Fisher Scientific, Waltham, USA), the `Dublin lens' and Tundra (Thermo Fisher Scientific, Waltham, USA) come next in terms of theoretical optical quality.

[Figure 2]
Figure 2
Theoretical temporal coherence envelope for a number of microscope configurations. Solid lines represent cold field emission gun (CFEG) instruments (ΔE = 0.3 eV). Dashed and dashed/dotted lines correspond to Schottky (thermionic) FEG machines (ΔE = 0.7 eV). The CRYO ARM 200 II (pink line) evaluated here is second in optical performance after the top-of-the-line CRYO ARM 300 II and Krios 5.

To experimentally test the overall optical performance of the CRYO ARM 200 II, we collected high-magnification images (200000×, 0.23 Å pixel−1, 101.5 e Å−2) of oriented single-crystal (100) gold-standard test specimen (AGS135, Agar Scientific, Rotherham, UK). The Fourier transform of one of the best images is shown in Fig. 3[link]. It contains spots up to lattice plane (046) with a spacing of 0.566 Å along the sample stage tilt axis and up to lattice plane (044) with a spacing of 0.721 Å perpendicular to the tilt axis. This confirms the excellent optical performance of the microscope, reaching beyond 1 Å in all directions. It also indicates that the performance may have been slightly affected by mechanical disturbances. The instrument is installed in a room that is on the ground floor of a building but was not specially designed for high-resolution electron microscopy with considerations and measures to minimize floor vibrations.

[Figure 3]
Figure 3
2D Fourier transform amplitudes of an oriented single crystal (100) gold sample image from the CRYO ARM 200 II. The visibility of lattice plane (046) spots corresponding to 0.566 Å spacing in the direction of the sample stage tilt axis and lattice plane (044) spots corresponding to 0.721 Å spacing in the perpendicular direction attests to the excellent optical performance of the microscope, reaching beyond 1 Å in all directions.

The data acquisition scheme and exemplary micrograph from the 200 kV apoferritin dataset are pre­sent­ed in Figs. 4[link](a) and 4(b). To minimize beam-induced sample motion, we pre­pared the 200 kV sample on a 0.6 µm hole gold foil grid and col­lected a single image in the centre of each hole with the beam illuminating uniformly the hole edge all-around [Fig. 4[link](a)]. To maximize the camera's detective quantum efficiency (DQE) performance over the target resolution range, the 200 kV data were collected at a relatively high indicated magnification of 150000× with a pixel size of 0.3056 Å pixel−1 [Fig. 4[link](b) and Table S1 in the supporting information]. The resolution of the final 3D map corresponds to ∼49% of the physical Nyquist limit of the camera.

[Figure 4]
Figure 4
Acquisition patterns and exemplary images for the (a)/(b) 200 kV and (c)/(d) 100 kV apoferritin test datasets. (a)/(c) Acquisition pattern schematics with support film holes in light blue, detector size and shape in blue, and beam size and position in pale yellow. Numbers indicate the acquisition order. (b)/(d) Exemplary images from the datasets collected at (b) 200 kV on a Gatan K3 camera and (d) 100 kV on a DECTRIS SINGLA hybrid-pixel detector. The inter­polated inter-module gap of the SINGLA detector is visible as a blurred horizontal stripe through the middle of the image.

The 100 kV dataset was also collected on a gold foil support, but to maximize data acquisition throughput, we used a grid with 1.2 µm holes and acquired four images per hole [Fig. 4[link](c)]. In previous experiments using such an acquisition strategy on this and other microscopes, we have reached resolutions far below 2 Å (Danev et al., 2021View full citation) and therefore did not expect it to be a limiting factor in this test. Due to the much larger physical pixel of the DECTRIS SINGLA camera (75 µm versus 5 µm of the Gatan K3), we had to collect at a much higher magnification of 500000× (Table S1). This pre­sent­ed some practical challenges in terms of microscope alignment and operation in the ultra-high magnification range, which is not typically used in cryo-EM. Fig. 4[link](d) contains an exemplary image from the 100 kV dataset illustrating the gap (horizontal stripe in the middle of the image) between the two modules of the detector which is inter­polated in software to facilitate cryo-EM data processing.

The results from the processing of the 200 kV dataset are pre­sent­ed in Fig. 5[link]. The reconstruction [Fig. 5[link](a)] reached 1.24 Å according to the 0.143 gold-standard Fourier Shell Correlation (FSC) criterion [Fig. 5[link](b)]. The 3D map exhibits features consistent with the estimated resolution, such as atomic bulges at a lower surface threshold and individual blobs for non-H atoms at a higher threshold [Fig. 5[link](c), grey and blue surfaces]. Furthermore, an FoFc difference density map calculated in Servalcat (Yamashita et al., 2021View full citation) revealed H-atom densities at many of their expected positions [Fig. 5[link](c), green surfaces].

[Figure 5]
Figure 5
Atomic resolution cryo-EM reconstruction of the 200 kV apoferritin dataset. (a) Overview of the 3D density map of a monomer. (b) Plot of the gold-standard FSC (GS FSC), indicating a 1.24 Å resolution at the 0.143 level and the map-model FSC indicating a 1.25 Å resolution at the 0.5 level. (c) Side-chain map features. The grey surface is at a lower threshold level typically used to depict the map. The blue surface is at a high threshold level, showcasing the atomic resolution of the map with blobs for individual atoms. The green surface represents an FoFc difference density map (3σ level, normalized within a mask) calculated from a hy­dro­gen-omit model. Positive density peaks are clearly visible at positions corresponding to H atoms in the mol­ecular model.

We qu­anti­fied the overall experimental performance using Rosenthal–Henderson B-factor plots (Rosenthal & Hen­der­son, 2003View full citation). Fig. 6[link] contains B-factor plots for the 200 kV dataset (red symbols and line) and a Krios G4 [set 1 in Danev et al. (2021View full citation)] apoferritin dataset (blue symbols and line). The measured B factors were 39.0 Å2 for the CRYO ARM 200 II and 38.7 Å2 for the Krios G4. In practical terms, they are virtually identical. The CRYO ARM 200 II plot was offset vertically toward higher resolutions, which indicates a slightly higher overall signal-to-noise ratio of the data.

[Figure 6]
Figure 6
Rosenthal–Henderson B-factor plots com­paring the overall cryo-EM performance of the CRYO ARM 200 II (red symbols and line) and the Krios G4 (blue symbols and line) microscopes with apoferritin samples. The slopes of the linear fits (red and blue lines) correspond to B factors of 39.0 and 38.7 Å2, respectively.

During data processing, we observed a 0.2 Å improvement in resolution (from 1.56 to 1.35 Å) after splitting the data into hourwise exposure groups (Fig. S1), indicating that there was a variation over time of the optical parameters. Fig. S3 in the supporting information shows a temporal trend plot of the refined beam tilt for the nine image-shift exposure groups. The beam tilt steadily increased by ∼0.65 mrad over the course of the experiment. The plot also shows that the image-shift beam tilt com­pensation of SerialEM performed well by keeping the spread between the nine acquisition areas within 0.1 mrad. Our original hypothesis for the cause of the beam tilt increase over time was that lens degaussing routines were not used during acquisition to reduce time overhead. During manuscript review, one of the reviewers (Dr Chris Russo) suggested that the reason may be CFEG instability. Consequently, we performed a long-term beam tilt stability measurement over a week. The results are pre­sent­ed in Fig. S4. For the first two days, the beam tilt stayed below 0.2 mrad [Fig. S4(a)]. After that, it began to increase linearly with a slope very similar to that in the 200 kV dataset (Fig. S3). When the beam tilt reached ∼1 mrad, we decided to test whether it would continue to increase when the emission is off. To this end, we switched off the emission for varying inter­vals [Fig. S4(a), greyed-out areas]. The data showed that the beam tilt magnitude did not increase during dark periods and even decreased slightly, followed by a quick rebound. Toward the end of the measurement, the CFEG appeared to have reached a relatively stable condition, but due to time constraints we had to terminate the experiment. The Cartesian plot of the beam tilt in Fig. S4(b) in the supporting information shows clear clustering of points corresponding to CFEG flashes. This is another confirmation that the beam tilt drift is related to the stability of the CFEG.

We were curious about the effect of radiation damage on the achievable resolution. To qu­antify this, we used the first 1000 movies and performed independent reconstructions starting from motion correction by omitting a varying number of initial frames to simulate pre-exposure. The same set of particle coordinates was used to extract particles from the aligned micrographs, followed by a 3D homogeneous refinement. The results are pre­sent­ed in Fig. 7[link]. Surprisingly, the reconstruction reached ∼2.9 Å resolution even after 20 e Å−2 of pre-exposure. The resolution dependence on pre-exposure was fitted very well (R2 = 0.997) by a parabola [Fig. 7[link](a), red line]. Numerical estimation of the resolution versus pre-ex­posure based on applying an empirical radiation damage model (Grant & Grigorieff, 2015View full citation) produced results that underestimated (larger value) the achievable resolution [Fig. 7[link](a), black symbols]. Using a total exposure value that was 82% of the experimentally measured one produced results that were much closer to the experimental ones [Fig. 7[link](a), blue symbols]. The difference in the exposure parameter could be due to various reasons, such as empirical damage model accuracy for this particular sample and support film, variation of the actual exposure during the experiment due to CFEG current decline, or other factors. Fig. 7[link](b) contains pre-exposure series illustrating the effect of radiation damage on 3D map fidelity. Acidic residues, such as glutamate (Glu) and aspartate (Asp), are known to be particularly sensitive to radiation (Fioravanti et al., 2007View full citation; Hattne et al., 2018View full citation). Indeed, the glutamate and aspartate side chains in Fig. 7[link](b) appear to be deca­rboxylated by ∼3 and ∼7 e Å−2, respectively, while nearby aromatic side-chain densities maintain their overall shape beyond 10 e Å−2.

[Figure 7]
Figure 7
Effect of radiation damage on resolution and map detail. (a) Plot of the resolution of reconstructions from the first 1000 movies containing 58k particles versus pre-exposure (red symbols), emulated by omitting a varying number of initial movie frames. The experimental distribution was well fit (R2 = 0.997) by a parabola (red line, coefficients in the legend). The plot also contains numerical estimates of the effect of radiation damage using the measured (black symbols) and 82% of the measured (blue symbols) total exposure. (b) Pre-exposure map series illustrating the effect of radiation damage on side-chain detail. Acidic residues (Glu and Asp) are deca­rboxylated and fade faster than large aromatic residues. The pre-exposure value and map resolution are shown in the top left corner of each panel.

The reconstruction of the 100 kV dataset reached a resolution of 1.91 Å (Fig. 8[link]). This corresponds to 123% of the physical Nyquist frequency of the detector. Map features, such as holes in aromatic side chains, confirm the estimated resolution [Fig. 8[link](a)]. Overall, this is an outstanding performance at 100 kV that closely matches what was achieved recently at 120 kV on a non-standard Glacios microscope with a narrow-gap `SP-Twin' objective lens polepiece (CC = 1.7 mm) and an Alpine (Gatan, Pleasanton, USA) camera (Chan et al., 2024View full citation). This configuration at 120 kV is expected to perform halfway between the Tundra microscope and the `Dublin lens' (SFEG) in Fig. 2[link]. Having a CFEG and a narrow-gap polepiece, at 100 kV the CRYO ARM 200 II is expected to perform better than the Glacios (SFEG) at 200 kV (Fig. 2[link]). However, the dataset did not reach higher resolution, and at this stage it is not clear what limited the practical performance. One possible reason is the relatively large pixel size and the need to rely on super-resolution processing. More advanced algorithms, such as post-acquisition super resolution (PASR) (Burton-Smith & Murata, 2023View full citation), may help to improve the result.

[Figure 8]
Figure 8
Result from the reconstruction of the 100 kV apoferritin dataset collected on the CRYO ARM 200 II with the DECTRIS SINGLA hybrid-pixel detector. The 3D reconstruction reached 1.91 Å resolution. (a) 3D density map showing holes in the aromatic side chains. (b) Gold-standard FSC plot.

4. Discussion

The EMDB deposition statistics (Fig. 1[link]) indicate that, despite their proven excellent performance, 200 kV cryo-microscopes remain heavily under-utilized. At cryo-EM facilities that also have 300 kV instruments, 200 kV microscopes are used primarily for sample screening and exploratory data collection. Understandably, having the option to use a top-of-the-line 300 kV instrument, researchers will choose to collect their data on it instead of the potentially lower-performing 200 kV machine. Such a choice was indeed justifiable with last-generation 200 kV microscopes equipped with SFEG emitters which limited their performance in the sub-3 Å region, as illustrated by the lower number of 200 kV EMDB depositions in this range (Fig. 1[link]) and the temporal envelope (Fig. 2[link], Glacios).

The introduction and wider adoption of CFEG emitters in recent years dramatically reshaped the cryo-EM performance landscape (Fig. 2[link], solid versus dashed/dotted lines). In particular, it significantly improved 200 kV performance in the 2–3 Å range (Fig. 2[link], Glacios versus CRYO ARM 200/Glacios 2). In addition to their optical advantages, CFEGs are also less costly to maintain because they do not require periodic emitter exchanges. In use, they do exhibit a gradual decline of emission current, necessitating tip flashes every ∼8 h, but this is handled automatically and can be performed during cryogen filling cycles to avoid extra acquisition inter­ruptions. The beam tilt drift observed in the 200 kV apoferritin dataset and during additional testing is worrisome and requires further investigation. It appears to be related to CFEG tip erosion and/or reshaping, and may be an issue specific to our instrument. We will continue to monitor it, and if necessary, request a service action.

The CRYO ARM 200 II microscope tested here introduces yet another performance-enhancing feature in the form of a narrow-gap objective lens polepiece, which reduces chromatic aberration and further expands the resolution envelope (Fig. 2[link], pink line). Its single-particle performance is com­parable to top-of-the-line 300 kV instruments, both in theory and practice, as demonstrated by the results pre­sent­ed above (Figs. 3[link], 5[link] and 6[link]). The only practical limitation of the narrow-gap polepiece is the lack of sample tilting capability. This precludes applications such as cryo-tomography (Nogales & Mahamid, 2024View full citation), MicroED (Clabbers et al., 2025View full citation) or single-particle samples with severe preferred orientation requiring tilted acquisition (Aiyer et al., 2024View full citation). Observations involving thicker samples (≥100 nm), such as cryo-tomography and 2D template matching of cellular samples, single-particle analysis of viruses, liposomes etc., will continue to benefit from higher accelerating voltages (Peet et al., 2019View full citation). However, for the vast majority of single-particle projects, the CRYO ARM 200 II offers uncom­promising performance and we are already using it for real-world samples with remarkable success (Jiang et al., 2025View full citation).

New technological developments, such as narrow-gap lenses and CFEGs will bring 100 kV microscopes closer to becoming the most cost-effective option for sample screening and exploratory data collection. Here, as well as in other recent works (Chan et al., 2024View full citation), the first sub-2 Å cryo-EM test structures from 100 kV instruments were pre­sent­ed. In practice, factors such as sample thickness, higher sensitivity to optical aberrations, electromagnetic disturbances, sample charging and detector performance, must also be considered and tested before declaring success. Overall, the proof will be in the real-world sample results that will be obtained from upgraded 100 kV microscopes.

In conclusion, with the technological advances discussed here, cryo-EM instrumentation could be entering a period of experimental role hand-down, where tasks that were exclusive to 300 kV instruments will become more common on 200 kV, and those from 200 kV will be shifted to 100 kV. The goal of this process is to make cryo-EM more accessible, affordable and prolific.

Acknowledgements

We thank Hiromu Shinmiya (Service), Toshiyuki Yamamoto (Service), Kazuya Omoto (R&D) and Sohei Motoki (R&D) from JEOL Ltd for their help with setting up and aligning the CRYO ARM 200 II microscope for observations at 100 kV. We are grateful to Takeyoshi Taguchi, Matthias Meffert, Saori Kawaguchi, Toshinobu Miyoshi and colleagues from DECTRIS Ltd for their efforts in organizing the loan and the installation of the DECTRIS SINGLA detector. We thank Yoichi Sakamaki for support in the operation and maintenance of the microscope. We thank Chris Russo for his suggestion to investigate CFEG instability as the probable cause of beam tilt drift.

Conflict of interest

Fabian Eisenstein is currently employed by DECTRIS Ltd, Baden-Daettwil, Switzerland, which loaned the SINGLA hybrid-pixel detector for the 100 kV experiments. The other authors declare no com­peting inter­ests.

Data availability

The 100 kV and 200 kV apoferritin datasets were deposited to the Electron Microscopy Public Image Archive (EMPIAR) with accession codes EMPIAR-13203 and EMPIAR-13202, respectively. The 3D map and model from the 200 kV dataset were deposited to the Electron Microscopy Data Bank (EMDB) and the Protein Data Bank (PDB) with accession codes EMD-68251and PDB-22FX, respectively. The 3D map from the 100 kV dataset was deposited to the Electron Microscopy Data Bank (EMDB) with accession code EMD-68204. The full raw datasets and CryoSPARC projects are also accessible on DECTRIS CLOUD (https://www.dectris.cloud), with instant access to the workflows, including the capability to reprocess the data. For access, please contact [email protected].

Funding information

Funding for this research was provided by: Research Support Project for Life Science and Drug Discovery [Basis for Supporting Innovative Drug Discovery and Life Science Research (BINDS)] from AMED under grant No. JP25ama121002j0004 to M. Kikkawa and R. Danev. R. Danev was also supported by the Japan Society for the Promotion of Science grant (KAKENHI, 22H02554).

References

Return to citationAiyer, S., Baldwin, P. R., Tan, S. M., Shan, Z., Oh, J., Mehrani, A., Bowman, M. E., Louie, G., Passos, D. O., Đorđević-Marquardt, S., Mietzsch, M., Hull, J. A., Hoshika, S., Barad, B. A., Grotjahn, D. A., McKenna, R., Agbandje-McKenna, M., Benner, S. A., Noel, J. A. P., Wang, D., Tan, Y. Z. & Lyumkis, D. (2024). Nat. Commun. 15, 389.  Web of Science CrossRef PubMed Google Scholar
Return to citationAlves, G. M., Andrews, T., McBean, P., Wells, T., El-Gomati, M. M., Burkhalter, S., Schulze-Briese, C., Zambon, P., McMullan, G., Henderson, R., Russo, C. J. & Jones, L. (2025). arXiv:2505.11458.  Google Scholar
Return to citationBurton-Smith, R. N. & Murata, K. (2023). bioRxiv:2023.06.09.544325.  Google Scholar
Return to citationCasañal, A., Lohkamp, B. & Emsley, P. (2020). Protein Sci. 29, 1055–1064.  Google Scholar
Return to citationChan, L. M., Courteau, B. J., Maker, A., Wu, M., Basanta, B., Mehmood, H., Bulkley, D., Joyce, D., Lee, B. C., Mick, S., Czarnik, C., Gulati, S., Lander, G. C. & Verba, K. A. (2024). J. Struct. Biol. 216, 108108.  Web of Science CrossRef PubMed Google Scholar
Return to citationChen, V. B., Arendall, W. B., Headd, J. J., Keedy, D. A., Immormino, R. M., Kapral, G. J., Murray, L. W., Richardson, J. S. & Richardson, D. C. (2010). Acta Cryst. D66, 12–21.  Web of Science CrossRef CAS IUCr Journals Google Scholar
Return to citationClabbers, M. T. B., Hattne, J., Martynowycz, M. W. & Gonen, T. (2025). Nat. Commun. 16, 2247.  Web of Science CrossRef PubMed Google Scholar
Return to citationCroy, N., Gaullier, G., Saura, P., Masiulis, S., Bhowmick, A., Kern, J., Raschdorf, O., Andersson, I., Kaila, V. R. I., Blikstad, C. & Messinger, J. (2025). bioRxiv:2025.12.22.696010.  Google Scholar
Return to citationCushing, V. I., Koh, A. F., Feng, J., Jurgaityte, K., Bondke, A., Kroll, S. H. B., Barbazanges, M., Scheiper, B., Bahl, A. K., Barrett, A. G. M., Ali, S., Kotecha, A. & Greber, B. J. (2024). Nat. Commun. 15, 2265.  CrossRef PubMed Google Scholar
Return to citationDanev, R., Yanagisawa, H. & Kikkawa, M. (2019). Trends Biochem. Sci. 44, 837–848.  Web of Science CrossRef CAS PubMed Google Scholar
Return to citationDanev, R., Yanagisawa, H. & Kikkawa, M. (2021). Microscopy 70, 487–497.  CrossRef PubMed Google Scholar
Return to citationDauter, Z. (2003). Macromolecular Crystallography (Part C), Vol. 368, Methods in Enzymology, edited by C. W. Carter Jr & R. M. Sweet, pp. 288–337. New York: Academic Press.  Google Scholar
Return to citationFeathers, J. R., Spoth, K. A. & Fromme, J. C. (2021). J. Struct. Biol. X 5, 100047.  PubMed Google Scholar
Return to citationFioravanti, E., Vellieux, F. M. D., Amara, P., Madern, D. & Weik, M. (2007). J. Synchrotron Rad. 14, 84–91.  Web of Science CrossRef CAS IUCr Journals Google Scholar
Return to citationFislage, M., Shkumatov, A. V., Stroobants, A. & Efremov, R. G. (2020). IUCrJ 7, 707–718.  CrossRef CAS PubMed IUCr Journals Google Scholar
Return to citationFujita, J., Makino, F., Asahara, H., Moriguchi, M., Kumano, S., Anzai, I., Kishikawa, J., Matsuura, Y., Kato, T., Namba, K. & Inoue, T. (2023). Sci. Rep. 13, 2279.  Web of Science CrossRef PubMed Google Scholar
Return to citationGerle, C., Kishikawa, J., Yamaguchi, T., Nakanishi, A., Çoruh, O., Makino, F., Miyata, T., Kawamoto, A., Yokoyama, K., Namba, K., Kurisu, G. & Kato, T. (2022). Microscopy 71, 249–261.  CrossRef CAS PubMed Google Scholar
Return to citationGrant, T. & Grigorieff, N. (2015). eLife 4, e06980.  CrossRef PubMed Google Scholar
Return to citationHamaguchi, T., Maki-Yonekura, S., Naitow, H., Matsuura, Y., Ishikawa, T. & Yonekura, K. (2019). J. Struct. Biol. 207, 40–48.  Web of Science CrossRef CAS PubMed Google Scholar
Return to citationHamdi, F., Tüting, C., Semchonok, D. A., Visscher, K. M., Kyrilis, F. L., Meister, A., Skalidis, I., Schmidt, L., Parthier, C., Stubbs, M. T. & Kastritis, P. L. (2020). PLoS One 15, e0232540.  CrossRef PubMed Google Scholar
Return to citationHattne, J., Shi, D., Glynn, C., Zee, C.-T., Gallagher-Jones, M., Martynowycz, M. W., Rodriguez, J. A. & Gonen, T. (2018). Structure 26, 759–766.e4.  CrossRef CAS PubMed Google Scholar
Return to citationHerzik, M. A., Wu, M. & Lander, G. C. (2017). Nat. Methods 14, 1075–1078.  CrossRef CAS PubMed Google Scholar
Return to citationHerzik, M. A., Wu, M. & Lander, G. C. (2019). Nat. Commun. 10, 1032.  Web of Science CrossRef PubMed Google Scholar
Return to citationJia, L., Ruben, E. A., Suarez, H. J., Olsen, S. K. & Wasmuth, E. V. (2024). bioRxiv:2024.05.07.593029.  Google Scholar
Return to citationJiang, X., Danev, R., Ichinose, S., Niu, B., Ohtsuki, S., Yanagisawa, H., Nagatoishi, S., Tsumoto, K., Hirokawa, N. & Kikkawa, M. (2025). Sci. Adv. 11, eady5861.  Web of Science CrossRef PubMed Google Scholar
Return to citationKaria, D., Koh, A. F., Yang, W., Cushing, V. I., Basanta, B., Mihaylov, D. B., Khavnekar, S., Vyroubal, O., Malínský, M., Sháněl, O., Doležal, V., Plitzko, J., Yu, L., Lander, G. C., Aricescu, A. R., Greber, B. J. & Kotecha, A. (2025). Structure 33, 1717–1727.  CrossRef CAS PubMed Google Scholar
Return to citationKayama, Y., Burton-Smith, R. N., Song, C., Terahara, N., Kato, T. & Murata, K. (2021). Sci. Rep. 11, 8395.  Web of Science CrossRef PubMed Google Scholar
Return to citationKoh, A., Khavnekar, S., Yang, W., Karia, D., Cats, D., van der Ploeg, R., Grollios, F., Raschdorf, O., Kotecha, A. & Němeček, D. (2022). J. Vis. Exp. (JoVE) e63519.  Google Scholar
Return to citationKüçükoğlu, B., Mohammed, I., Guerrero-Ferreira, R. C., Ribet, S. M., Varnavides, G., Leidl, M. L., Lau, K., Nazarov, S., Myasnikov, A., Kube, M., Radecke, J., Sachse, C., Müller-Caspary, K., Ophus, C. & Stahlberg, H. (2024). Nat. Commun. 15, 8062.  Web of Science PubMed Google Scholar
Return to citationMaki-Yonekura, S., Kawakami, K., Takaba, K., Hamaguchi, T. & Yonekura, K. (2023). Commun. Chem. 6, 98.  Web of Science PubMed Google Scholar
Return to citationMcMullan, G., Naydenova, K., Mihaylov, D., Yamashita, K., Peet, M. J., Wilson, H., Dickerson, J. L., Chen, S., Cannone, G., Lee, Y., Hutchings, K. A., Gittins, O., Sobhy, M. A., Wells, T., El-Gomati, M. M., Dalby, J., Meffert, M., Schulze-Briese, C., Henderson, R. & Russo, C. J. (2023). Proc. Natl Acad. Sci. USA 120, e2312905120.  CrossRef PubMed Google Scholar
Return to citationMeng, E. C., Goddard, T. D., Pettersen, E. F., Couch, G. S., Pearson, Z. J., Morris, J. H. & Ferrin, T. E. (2023). Protein Sci. 32, e4792.  Web of Science CrossRef PubMed Google Scholar
Return to citationMerk, A., Fukumura, T., Zhu, X., Darling, J. E., Grisshammer, R., Ognjenovic, J. & Subramaniam, S. (2020). IUCrJ 7, 639–643.  CrossRef CAS PubMed IUCr Journals Google Scholar
Return to citationNakane, T., Kotecha, A., Sente, A., McMullan, G., Masiulis, S., Brown, P. M. G. E., Grigoras, I. T., Malinauskaite, L., Malinauskas, T., Miehling, J., Uchański, T., Yu, L., Karia, D., Pechnikova, E. V., de Jong, E., Keizer, J., Bischoff, M., McCormack, J., Tiemeijer, P., Hardwick, S. W., Chirgadze, D. Y., Murshudov, G., Aricescu, A. R. & Scheres, S. H. W. (2020). Nature 587, 152–156.  Web of Science CrossRef CAS PubMed Google Scholar
Return to citationNaydenova, K., McMullan, G., Peet, M. J., Lee, Y., Edwards, P. C., Chen, S., Leahy, E., Scotcher, S., Henderson, R. & Russo, C. J. (2019). IUCrJ 6, 1086–1098.  Web of Science CrossRef CAS PubMed IUCr Journals Google Scholar
Return to citationNogales, E. & Mahamid, J. (2024). Nature 628, 47–56.  Web of Science CrossRef CAS PubMed Google Scholar
Return to citationPeet, M. J., Henderson, R. & Russo, C. J. (2019). Ultramicroscopy 203, 125–131.  CrossRef CAS PubMed Google Scholar
Return to citationPettersen, E. F., Goddard, T. D., Huang, C. C., Couch, G. S., Greenblatt, D. M., Meng, E. C. & Ferrin, T. E. (2004). J. Comput. Chem. 25, 1605–1612.  Web of Science CrossRef PubMed CAS Google Scholar
Return to citationPunjani, A., Rubinstein, J. L., Fleet, D. J. & Brubaker, M. A. (2017). Nat. Methods 14, 290–296.  Web of Science CrossRef CAS PubMed Google Scholar
Return to citationReimer, L. & Kohl, H. (2010). In Transmission Electron Microscopy: Physics of Image Formation. New York: Springer.  Google Scholar
Return to citationRosenthal, P. B. & Henderson, R. (2003). J. Mol. Biol. 333, 721–745.  Web of Science CrossRef PubMed CAS Google Scholar
Return to citationRoss, I. M. & Walther, T. (2012). J. Phys. Conf. Ser. 371, 012012.  CrossRef Google Scholar
Return to citationRusso, C. J. & Passmore, L. A. (2014). Science 346, 1377–1380.  CrossRef CAS PubMed Google Scholar
Return to citationRusso, C. J. & Passmore, L. A. (2016). Curr. Opin. Struct. Biol. 37, 81–89.  Web of Science CrossRef CAS PubMed Google Scholar
Return to citationSchorb, M., Haberbosch, I., Hagen, W. J. H., Schwab, Y. & Mastronarde, D. N. (2019). Nat. Methods 16, 471–477.  CrossRef CAS PubMed Google Scholar
Return to citationSheldrick, G. M. (1990). Acta Cryst. A46, 467–473.  CrossRef CAS Web of Science IUCr Journals Google Scholar
Return to citationThangaratnarajah, C., Rheinberger, J. & Paulino, C. (2022). Curr. Opin. Struct. Biol. 76, 102440.  CrossRef PubMed Google Scholar
Return to citationVenugopal, H., Mobbs, J., Taveneau, C., Fox, D. R., Vuckovic, Z., Gulati, S., Knott, G., Grinter, R., Thal, D., Mick, S., Czarnik, C. & Ramm, G. (2025). Sci. Adv. 11, eadr0438.  CrossRef PubMed Google Scholar
Return to citationWu, M., Lander, G. C. & Herzik, M. A. (2020). J. Struct. Biol. X 4, 100020.  PubMed Google Scholar
Return to citationYamashita, K., Palmer, C. M., Burnley, T. & Murshudov, G. N. (2021). Acta Cryst. D77, 1282–1291.  Web of Science CrossRef IUCr Journals Google Scholar
Return to citationYip, K. M., Fischer, N., Paknia, E., Chari, A. & Stark, H. (2020). Nature 587, 157–161.  Web of Science CrossRef CAS PubMed Google Scholar
Return to citationZambon, P. (2023). Front. Phys. 11, 1123787.  CrossRef Google Scholar
Return to citationZhang, K., Pintilie, G. D., Li, S., Schmid, M. F. & Chiu, W. (2020). Cell Res. 30, 1136–1139.  Web of Science CrossRef PubMed Google Scholar

This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.

ISSN: 2052-2525