The five-analyzer point-to-point scanning crystal spectrometer at ESRF ID26
aESRF - The European Synchrotron, 71 Avenue des Martyres, 38000 Grenoble, France, and bDesign et Mécanique, Les Coings, 38210 Montaud, France
*Correspondence e-mail: glatzel@esrf.fr
X-ray emission spectroscopy in a point-to-point focusing geometry using instruments that employ more than one analyzer crystal poses challenges with respect to mechanical design and performance. This work discusses various options for positioning the components and provides the formulas for calculating their relative placement. Ray-tracing calculations were used to determine the geometrical contributions to the energy broadening including the source volume as given by the beam footprint on the sample. The alignment of the instrument is described and examples are given for the performance.
1. Introduction
Inner-shell photon-in/photon-out spectroscopy provides an element-selective probe of local coordination and electronic structure (Kotani & Shin, 2001; de Groot & Kotani, 2008; Kowalska et al., 2016; Sa, 2014; Bokhoven & Lamberti, 2016). can be defined as the detection of X-rays emitted from a sample following core-hole creation in an analyte atom with an instrumental energy bandwidth that is on the order of the core-hole lifetime broadening. We refer to such an energy resolution as `high' to distinguish it from standard fluorescence measurements using conventional solid-state detectors with a E/ΔE < 50. can be performed by taking advantage of various modes of core-hole creation of which the most common is photo-excitation but also ion/electron bombardment and radioactive isotopes (e.g. K capture decay) have been used (Kavčič, 2012; Glatzel et al., 2001). following photo-excitation is a second-order optical process that is theoretically treated using the Kramers–Heisenberg equation (Kotani & Shin, 2001). An instrument for thus enables high energy resolution spectroscopy (e.g. Kβ spectroscopy), resonant X-ray emission spectroscopy (RXES), resonant inelastic X-ray scattering (RIXS) and high energy resolution fluorescence detected (HERFD) X-ray absorption spectroscopy (Glatzel et al., 2013; Lafuerza, Retegan et al., 2020). X-ray Raman spectroscopy (Rueff & Shukla, 2010) does not require continuously tuneable (i.e. scanning) photon-out spectroscopy and is thus ideally carried out with the spectrometer angle fixed close to backscattering in order to optimize the energy resolution and focusing properties (Sokaras et al., 2012; Huotari et al., 2017, 2011; Fister et al., 2006; Schülke, 2007; Mao et al., 2006; Hiraoka et al., 2013).
In the hard X-ray regime instruments based on perfect crystal Bragg optics are generally used to analyze the emitted X-rays and thus achieve the required energy resolution. Common geometries are Rowland or von Hámos. The energy of the emitted X-rays is around a fluorescence line or an λ = 2dsinθ) the reflection angle θ cannot be chosen freely and is dependent on the and the electronic transition of interest. An alternative to Bragg optics would be energy-dispersive transition-edge sensors with dramatically reduced energy bandwidth compared with conventional solid-state detectors (Li et al., 2018; Doriese et al., 2017).
Following Bragg's law (Hard e.g. FEFF, ORCA, FDMNES, Crispy (Rehr et al., 2010; Neese, 2018; Bunău & Joly, 2009; Retegan, 2020)]. The full potential of this technique thus becomes available to a broad user community with a wealth of applications.
has a long history dating back to Henry Moseley who used the technique to order elements in the periodic table. The technique has experienced a sharp increase in demand in recent years. This is partly due to the larger number of available hard X-ray beamlines at synchrotron radiation sources around the world that are suitable for Even more important is a better understanding of the theory of User-friendly codes to calculate inner-shell spectra now provide the possibility to model data [Non-resonant et al., 2020). High-quality Bragg optics have become available at a reasonable cost which has led to the development of instruments by many research groups. and in the laboratory require very similar instrumentation and recent years have seen a fantastic increase in the application of laboratory (see the review by Zimmermann et al., 2020). finds attractive applications for time-resolved studies at synchrotrons, free-electron lasers and using pulsed laboratory sources (Alonso-Mori et al., 2012, 2015; Szlachetko et al., 2014, 2017; Abraham et al., 2019; Anwar et al., 2019; Wernet, 2019; Tu et al., 2019; Miaja-Avila et al., 2016; Smolentsev et al., 2020; Vankó et al., 2013).
does not require a monochromatic incident beam, which considerably relaxes the constraints on the X-ray source. At a synchrotron radiation source or free-electron laser one can use a `pink' beam with an energy bandwidth much larger than the core-hole lifetime broadening (DikayaThe key components are an X-ray emitting sample that defines the source volume, one or several analyser crystals and a photon detector. Many instruments use bent crystals that are either cylindrical or spherical in shape. The bending radius, together with the θ, defines the relative position between source volume, analyser and detector. In general, a smaller bending radius results in a larger solid angle captured. At the same time, bending of the crystal wafer introduces elastic stress and thus a broader reflectivity curve. Mechanical stress is likely to deform the surface (Honkanen et al., 2014a).
The relation between the energy bandwidth and the spread in ΔE = EΔθcotθ, where Δθ is given by the properties of the components and their geometrical arrangement. There are several possibilities to design an instrument for and some guidelines must be established. First, a spectrometer may be dispersive or scanning (point-to-point). A dispersive instrument covers a range of Bragg angles in a stationary configuration and the energies are spatially separated on a position-resolving detector. Such an instrument is often mechanically simpler and ideally suited for time-resolved studies as all energies of a spectrum are recorded simultaneously, i.e. a spectrum can be measured in a single shot of X-rays. In this case the energy resolution is determined by the properties of the analyzer crystal and the position resolution of the detector (Kuzmenko et al., 2019; Alonso-Mori et al., 2012; Machek et al., 2007; Dickinson et al., 2008; Kavčič et al., 2012). The influence of the source volume on the energy resolution strongly depends on the chosen geometry. A scanning instrument attempts to use the entire available solid angle, i.e. the crystal analyzer surface, in an energy bandwidth that is sufficiently small to result in spectra that allow for chemical analysis. Here, Δθ is given by the source volume, the properties of the crystal and geometric contributions that strongly depend on the precise positioning of the analyzer crystals relative to the source volume.
is given by the differential form of Bragg's lawScanning instruments are mechanically challenging particularly when more than one analyzer crystal is employed as the relative displacement between the crystals must be very precise (Sokaras et al., 2013; Uruga et al., 2019; Ablett et al., 2019; Duan et al., 2017; Moretti Sala et al., 2018; Llorens et al., 2012). They do not allow measurement of an entire spectrum in a single exposure but a spectrum is recorded by moving the crystal(s) and detector along a given trajectory. The use of the available solid angle in one energy point and the small detector surface required compared with dispersive setups makes scanning instruments more suitable for measurements of very weak fluorescence signals. Also, is most efficiently carried out with a scanning instrument as only the intensity in the maximum of the fluorescence line is of interest which should be recorded with the entire available solid angle. We discuss here a scanning multi-crystal X-ray emission spectrometer on beamline ID26 of the European Synchrotron Radiation Facility that was commissioned in 2008. The mechanical design was presented by Ducotté et al. (2010). The concept of the instrument has since then been transferred to other beamlines at the ESRF (BM23, ROBL) and elsewhere (Kleymenov et al., 2011). This paper reports the experiences and lessons learned from working with this instrument. The ray-tracing code was written in MATLAB and is available upon request.
2. Description of geometry
The source position (0, 0, 0) of the spectrometer is defined by the position where the incoming X-ray beam impinges on the sample. In most cases, it is necessary to keep this position fixed, for example when an ion, electron or X-ray beam is used to create the inner-shell vacancy in the sample and the beam cannot be easily moved. If a radioactive isotope is used as an X-ray emitter, this constraint can be relaxed. The source volume of the spectrometer is given by the incident beam size, the orientation of the sample surface with respect to the incoming beam direction and the attenuation length of the emitted X-rays. The crystal analyzer has a spherical shape with radius RP. The position of the centre of crystal A, the detector D and the Rowland circle K are given by
An index n (n = −2, −1, 0, 1, 2) is introduced to distinguish between the five analyzer crystals. Each crystal An describes a Rowland circle Kn with its detector Dn. The Rowland circle is defined by its radius RR that is half the bending radius RP = 2RR of the reflecting crystal planes. In Johann geometry, the crystal optical surface has the same radius as the planes of the chosen crystal reflection. The bending radius is assumed to be identical for all five crystals. We note that the manufacturing process may give the best energy resolution for slightly different values for the bending radii of the analyzers (±5 mm). However, it is undesirable to have all components move on Rowland circles with different diameters because the foci would not coincide on the detector (see discussion of detector below). The instrument thus approximates all circle diameters to identical values. Recent improvements in the crystal analyser manufacturing process result in more and more homogeneous analyser properties.
We define the position Kn of the Rowland circle by the coordinates where a beam from the source position hits the centre of the crystal in a symmetric configuration, i.e. the crystal asymmetry angle α = 0. We furthermore define a starting configuration for n = 0 where the crystal is oriented such that the normal on the crystal planes in Kn is parallel to the x axis. The configuration is shown in Fig. 1 for a θB = 75°.
There are several reasons to modify this starting configuration. In case of a crystal asymmetry angle α between the optical surface and the crystal planes, the crystal must be oriented such that the normal on the crystal planes is rotated by −α in the (x, z) plane relative to the starting position in Fig. 1, i.e. counter-clockwise around the y axis. This is to ensure the same as in the symmetric configuration. The source-to-crystal and detector-to-crystal distances are Rpsin(ϑ + α) and Rpsin(ϑ − α), respectively (Suortti et al., 1999). The crystal must be offset from the centre position Kn by
and
followed by a counter-clockwise rotation of the crystal (i.e. the optical surface) around the y axis by 2α.
The positions of the crystal, the Rowland circle and the detector including a possible asymmetry are thus
The energy bandwidth is given by intrinsic contributions of the bent crystal and geometric contributions are given by the geometry between the source and crystal. The intrinsic contributions are calculated by considering the elastic bending and the mechanical stress of a crystal wafer with a certain thickness and radius (Honkanen et al., 2014a). Formulas for geometric contributions have been given by several authors (Bergmann & Cramer, 1998; Moretti Sala et al., 2018). In general, all geometric contributions increase with decreasing and therefore crystal material and are chosen such that the is close to 90°. Here we determine the bandwidth by means of ray-tracing simulations and do not attempt to obtain the energy bandwidth using analytical formulas. We only note that the Johann error in energy scales with cot2(ϑB + α) (Suortti et al., 1999) and it is thus interesting to study how the Johann error for a given changes with the crystal asymmetry. Fig. 2 shows the deviation (Δθ) from the of the central beam θB across the surface of the analyzer crystal. In the absence of asymmetry the deviation shows the typical dependence for the Johann error in the vertical An asymmetry of 15° eliminates the Johann error as θB + α = 90° (Moretti Sala et al., 2018), but the contribution from the crystal surface outside the (i.e. the sagittal direction) increases by the same amount as the decrease in Johann error (Fig. 2). If the asymmetry is chosen as (90 − θB)/2 we obtain a radial distribution of Δθ on the surface. The energy bandwidth calculated by including 80% of all rays around θB is 1.14 eV for the Si(440) (RP = 500 mm) reflection at 75° (6684.9 eV) for α = 0° and 15° and 0.94 eV for α = 7.5°. The figure nicely shows how placing a mask in front of the crystal improves the energy resolution. We note that the figure errors of the crystal surface may make a significant contribution to the geometrical energy broadening and they may increase with increasing radius r. In this case, a mask will help to select the part of the crystal surface with the smallest figure error. A toroidal analyser shape could reduce or eliminate the energy broadening outside the (Bryzgunov, 2000; Jahrman et al., 2019). However, the required bending radius in the sagittal plan depends on the (see below).
A finite source size will contribute to the geometrical energy bandwidth depending on the ). In standard configuration on ID26 the incoming beam with size 0.1 mm (vertical) × 0.7 mm (horizontal) (values before the upgrade of ESRF to the extremely brilliant source) impinges on the sample under 45° and the horizontal scattering angle with the central crystal is 90°. In this case, the geometrical contribution increases from 0.3 eV to 0.5 eV for the Si (440) reflection at 6684.9 eV (RP = 1000 mm). When reducing the bending radius to RP = 500 mm the bandwidth increases to 1.1 eV and 1.6 eV for the two cases. The chosen beam size and bending radius are often a compromise between spectrometer energy resolution and possible radiation damage in the sample. We note that the source size contribution can be minimized in some dispersive setups with 2D detectors (Holden et al., 2017; Nowak et al., 2020; Kavčič et al., 2012)
(Bergmann & Cramer, 1998In order to cover a large solid angle it is necessary to use more than one analyzer crystal. One could position several crystals on one Rowland circle each with an appropriate asymmetry angle corresponding to its offset az,α. We note that placing a crystal off the Rowland centre K without an asymmetric cut increases the energy bandwidth. At θB = 75° moving the crystal up (down) from the symmetric position by 50 mm with the crystal surface centre on the Rowland circle increases the geometrical energy bandwidth from 0.3 eV to 1.2 eV (1.6 eV) for the Si (440) reflection at 6684.9 eV (RP = 1000 mm).
The least complex way to increase the solid angle is to add crystals in the sagittal plane (Fig. 3). We use a rotation around the z axis to achieve an offset between the crystals along the y axis. Each crystal defines a Rowland circle. The five circles intersect at two points. One intersection point coincides with the source volume. The detector may be placed at the second intersection point. We would like the sagittal offset between the crystals to be as small as possible and constant for all Bragg angles. This is desirable because in situ and high-pressure cells often have limited opening windows. The ratio between the solid angle covered by the analyzer crystals and the solid angle given by the required window opening such that the source volume sees all crystals should be as large as possible. The rotation angle is given by
The spectrometer on ID26 uses ay,off = 120 mm with a 100 mm diameter for the analyzer crystals.
Finally, we would like to have the central analyzer position at az,off = 0 for all Bragg angles. This is again desirable for sample environments with limited opening windows but also to ensure that the scattering geometry varies as little as possible over a scan range of the This is also important for isotropic as a strong background may arise from Compton scattering that is minimized at 90° scattering angle for linear polarized incoming light. We thus rotate all analyzer crystals around the y axis () by an angle of
where we give the general expression for any position az,off of the centre of the crystal analyzer along the z axis.
The detector positioned on the Rowland circles provides the smallest focal spot size in the direction of the Rsinϑ for the image-to-mirror distance for a spherical mirror). For θB = 75° the focus is 13 mm-wide in the y direction for α = 0° and 20 mm for α = 15°. The sagittal spot size decreases with increasing but does not depend on the crystal bending radius RP.
but a large spot size in the sagittal direction (the condition for meridional focusing isThe condition for sagittal focusing is (RPsinϑ)/(−cos2ϑ) for the distance between the crystal and detector. We can use this in case we would like the analyzer crystals to give distinct foci on the detector. Images of the foci for a five-crystal spectrometer are shown in Fig. 4. Moving the detector out of the Rowland circle along the direction defined by A0 and the Rowland circle intersection point separates the foci from the five crystals and changes the focal shape. This requires a larger detector area or five separate detectors (Moretti Sala et al., 2018). The images also show that the position resolution of the detector can be used to correct the energy dispersion and thus improve the resolution. For example, this was done by Huotari et al. (2005) and Honkanen et al. (2014b ) for a Johann spectrometer.
The spectrometer on ESRF ID26 currently uses a single-element detector (avalanche photo diode, silicon drift diode) such that continuous scans with acquisition times down to 10 ms per datapoint are easily achievable and the data acquisition chain is simple. The rapid development of pixel detectors make them a very attractive alternative also for continuous scans in spectroscopy and ID26 may implement one in the future. We note that it is beneficiary to have only a small detector surface (or a carefully chosen region of interest on the 2D detector) exposed if the background must be kept low. Stray X-rays in the experimental hutch may give rise to a signal as strong as or stronger than weak fluorescence lines. This is particularly important if the analyte atom is low in concentration or in a matrix of high-Z elements that strongly fluoresce. In this case or when a lower-order crystal analyzer reflection is relevant, energy resolution of the spectrometer detector (e.g. using a silicon drift diode detector) may help to further reduce the background. A 2D detector also provides the possibility to record the X-ray background outside the focus region. However, the background underneath the fluorescence line of interest may be different inside the focus than outside the focus. The reason is that some of the background is Bragg reflected (e.g. Compton scattering or tails of strong fluorescence lines emitted from the matrix) and is thus only observed in the focus.
The ID26 spectrometer uses a slit in front of the detector with a 1.0–1.5 mm opening in the meridional direction. The slit defines the position of the detector on the Rowland circle. Having all five crystals impinge on the same detector element puts additional constraints on the alignment procedure that we explain in the following. Closing the slits may improve the energy resolution at the expense of intensity. However, if the slits cut into the focus the spectrometer becomes sensitive to imperfect tracking during the scan and a slit size above 1 mm has proven to be a good choice.
3. Spectrometer alignment and scanning
Each analyzer crystal has four X that carries a rotation stage with a vertical axis. On top of this rotation state is a vertical translation to achieve Az that carries a cradle to achieve a rotation with the axis perpendicular to the meridional plane of the analyzer (Fig. 3). The five assemblies for each crystal are separated by ay,off. The stages serve the alignment and scanning of the spectrometer. The motion errors must be kept to a minimum in order to avoid misalignment of the meridional angle between the assemblies. For the design of the instrument the objective was to keep the error in the meridional angle below 10 µrad over a scan range. All 20 crystal stages are mounted on a granite block that can be displaced along X in order to accommodate bending radii between 0.5 m and 2 m. The detector is mounted on translation stages and a rotation stage. It is positioned with a precision of 50 µm and the detector slit always faces the central crystal.
In the bottom is a translation alongThe spatial coordinates of all components and the sample position are determined with a precision of ∼100 µm by an optical alignment. The source volume is defined by aligning the incoming beam in X and Z through a pinhole. The alignment along Y is less critical and is achieved by considering the dimensions of the sample mount. A reference is placed in the source volume and the five analyzer crystals are aligned such that the fluorescence signal is centred on the detector slit, i.e. we do not attempt to maximize the fluorescence signal in this step. Thus, the and therefore the spectrometer energy calibration is defined by the geometrical alignment and not by the presumed maximum of the fluorescence line, the energy of which is not always precisely known. This approach gives the best results for tracking of the focal spot over a given scan range.
In a second step, the elastic peak is measured for all crystals and the Az position for each crystal is adjusted such that the energies of all elastic peaks coincide. Alternatively, the meridional angle for each crystal can be adjusted. There are several reasons why a geometrical alignment does not achieve identical energy calibration between crystals. An error in the optical alignment translates into an error of the and thus the energy. The most sensitive direction is Z. Furthermore, the crystals may have a small asymmetry angle while we assume a symmetric geometry in the alignment. The two errors are related in the Rowland geometry as we discussed before and the two corrections are thus equivalent. An unintended miscut is <0.05° according to the manufacturers and a misalignment in Z will be less than a few hundred micrometres. In both cases, the energy resolution will not suffer. The final step of the alignment is a scan of the detector slit closed to 0.1 mm along Z to determine the shape of the focus. The appropriate slit size is chosen accordingly. An offset in the vertical direction is introduced for the calculation of the detector position to center the slit in the measured position. This offset is typically a few hundred micrometres and arises from the relative energy calibration between the analyzer crystals.
The d-spacing. For pure Si this is known to very high precision but the bonding process of the wafer on the glass substrate may affect the Furthermore, Ge – another important material for analyzer crystals – is not available in such high purity and the manufacturer may not be able to provide the precise value for the lattice parameters. Although it would be possible to determine the lattice parameters, we note that accurate absolute values for the emission energy are hardly ever needed and ID26 decided to keep the energy calibration of the incident beam and of the instrument separate and use the measurement of the elastic peak to link the two.
could be calibrated using the energy calibration of the incident beam by measuring the elastic peak. This has led to poor results with respect to tracking over the whole scan range in our experience. Determination of the via the optically aligned spectrometer components leads to better tracking. One reason for this may be an error in the calibration of the energy of the incoming beam. The generally accepted values for absorption edges are often a convention rather than precisely determined quantities. Furthermore, calculation of the from a given wavelength requires precise knowledge of the crystalA stable relative energy calibration is crucial in order to detect small changes in the samples. To this end, the θ-cradles and Az translations are equipped with encoders and the mechanical design ensures very high repeatability. The main source of error is a vertical beam motion on the sample because it modifies the to first order. Slits placed in front of the sample that define the incoming beam position keep this error below 10 µm. For Fe Kα measured using the Ge(440) reflection at 75° we determined an error for the energy calibration to less than 10 meV (Lafuerza, Carlantuono et al., 2020). When changing samples the correct placement along Y is achieved by scanning the sample in the Y direction while observing the signal on the detector. Such a scan moves the focus across the detector slit in the sagittal direction and allows alignment to within 100 µm.
The alignment steps with X-rays are shown in Fig. 5 for Si(620) crystals. Scans of the elastic peak are shown for further crystal reflections. The top panel shows scans to determine the crystal bending radius RP. We usually take ten scans for different bending radii and use a figure of merit combining energy bandwidth and intensity of the elastic peak to determine the optimal radius. The nominal radius of the glass substrate is RP = 1000 m but the best performance is found for RP = 996 mm in this case. The best performance is typically found in a range ±5 mm around the nominal radius. In general, similar values are found for crystals from the same manufacturing batch and the spectrometer is run using the experimentally determined best radius for all crystals. The elastic peaks differ slightly for each crystal.
The crystals are aligned to give identical centre energies at the half maxima. The energy bandwidths of the elastic peaks for Si(620) at 74.8° are between 0.7 eV and 0.8 eV using the Si(311) reflection for the incident beam. The
of the Si(311) reflection is ∼35700 and combined with the incident beam vertical divergence we obtain 0.3 eV for the incoming beam energy bandwidth. This gives an energy bandwidth of 0.6 (0.7) eV for the spectrometer taking 0.7 (0.8) eV for the elastic peak. The calculated geometrical contribution is 0.55 eV in this case and the intrinsic contributions to the crystal analyzer energy broadening are thus small. The elastic peaks for Si(620) at 74.8° show a strong asymmetry whereas the elastic peaks for other reflections at higher Bragg angles are more symmetric.We also tested the dependence of the energy bandwidth on the azimuthal rotation [see also the work by Mortensen & Seidler (2017)]. This dependence mainly arises from errors in the bonding process such that the wafer is not bent to a perfect sphere. Azimuthal alignment results are shown in Fig. 6 for Si(440) and Ge(620) crystals. While Si crystals typically show little influence, Ge crystals may show a strong dependence on the azimuthal angle [in the Ge(620) crystal example, azimuthal orientation a2 clearly yields the best performance].
Table 1 lists the average elastic as collected from several experimental sessions. The variation around the given values is large because each crystal is different due to the manufacturing process. In particular, crystals made of Ge show a very large spread with variations up to 40%. The estimated analyzer bandwidth is based on several assumptions (Gaussian shapes, approximate vertical divergence for incoming beam) that have not been verified experimentally. The given values are thus only a very rough guide. The elastic peaks become sharper with increasing and decreasing energy. Otherwise, the bandwidth strongly depends on the crystal quality. The performance of the crystal analyzer has dramatically improved over the past ten years, considerably reducing the energy bandwidth of the spectrometer. This process is ongoing. New fabrication processes for Ge and the use of other materials (e.g. quartz, LiNbO3) will further improve the energy resolution for a given emission energy. Table 1 is therefore only a snapshot of the current situation.
|
The geometrical energy broadening strongly depends on the crystal bending radius. The spectrometer on ID26 can thus accommodate crystals with radii between 0.5 m and 2 m. This allows optimization of either the captured solid angle or the energy resolution. The required energy resolution depends on the natural widths of the fluorescence line as discussed in detail by Rovezzi et al.(2017). Crystals with RP = 500 mm have pushed the for the study of Hg in natural samples below the ppm level (Manceau et al., 2016).
The spectrometer is implemented in the beamline software as a virtual motor and the user only needs to select an emission energy to move all spectrometer motors. RIXS and constant final state scans are executed with single commands after choosing the energy range and step sizes. The
instrument on ID26 is used in almost all experiments. In order to allow fast change between analyzer crystals, the alignment procedure was automated as much as possible reducing the time for aligning five crystals to below 1 h.4. Outlook
The complexity of a point-to-point, scanning multi-analyzer et al. (2020). This new design is based on cylindrical crystals and opens up a path towards increasing the captured solid angle while maintaining the energy resolution. It also allows the use of fewer crystal reflections in the hard X-ray range because it can employ Johansson-type crystals. With a growing interest in the scientific community such instruments may also be developed for use in laboratories.
instrument is manageable. The instrument described in this paper employs spherical Johann crystals that are nowadays available in very high quality at affordable prices from various suppliers. A new concept for a multi-crystal point-to-point spectrometer is presented in a recent paper by RovezziFootnotes
‡Current address: AGH University of Science and Technology, Krakow, Poland.
§Current address: School of Physical Science and Technology, Shanghai Tech University, Shanghai 201210, Popele's Republic of China.
¶Current address: University Grenoble Alpes, CNRS, IRD, Irstea, Météo France, OSUG, FAME, 38000 Grenoble, France.
Acknowledgements
The authors thank all ID26 staff and users for making excellent use of this instrument. The ESRF crystal analyzer laboratory is acknowledged (R. Verbeni, C. Lapras).
References
Ablett, J. M., Prieur, D., Céolin, D., Lassalle-Kaiser, B., Lebert, B., Sauvage, M., Moreno, T., Bac, S., Balédent, V., Ovono, A., Morand, M., Gélebart, F., Shukla, A. & Rueff, J.-P. (2019). J. Synchrotron Rad. 26, 263–271. Web of Science CrossRef CAS IUCr Journals Google Scholar
Abraham, B., Nowak, S., Weninger, C., Armenta, R., Defever, J., Day, D., Carini, G., Nakahara, K., Gallo, A., Nelson, S., Nordlund, D., Kroll, T., Hunter, M. S., van Driel, T., Zhu, D., Weng, T.-C., Alonso-Mori, R. & Sokaras, D. (2019). J. Synchrotron Rad. 26, 629–634. Web of Science CrossRef CAS IUCr Journals Google Scholar
Alonso-Mori, R., Kern, J., Gildea, R. J., Sokaras, D., Weng, T. C., Lassalle-Kaiser, B., Tran, R., Hattne, J., Laksmono, H., Hellmich, J., Glöckner, C., Echols, N., Sierra, R. G., Schafer, D. W., Sellberg, J., Kenney, C., Herbst, R., Pines, J., Hart, P., Herrmann, S., Grosse-Kunstleve, R. W., Latimer, M. J., Fry, A. R., Messerschmidt, M. M., Miahnahri, A., Seibert, M. M., Zwart, P. H., White, W. E., Adams, P. D., Bogan, M. J., Boutet, S., Williams, G. J., Zouni, A., Messinger, J., Glatzel, P., Sauter, N. K., Yachandra, V. K., Yano, J. & Bergmann, U. (2012). Proc. Natl Acad. Sci. USA, 109, 19103–19107. Web of Science CAS PubMed Google Scholar
Alonso-Mori, R., Sokaras, D., Zhu, D., Kroll, T., Chollet, M., Feng, Y., Glownia, J. M., Kern, J., Lemke, H. T., Nordlund, D., Robert, A., Sikorski, M., Song, S., Weng, T.-C. & Bergmann, U. (2015). J. Synchrotron Rad. 22, 612–620. Web of Science CrossRef CAS IUCr Journals Google Scholar
Anwar, M. I., Ha, S. S., Hwang, B.-J., Han, S., Oh, M. S. H., Faiyaz, M., Noh, D. Y., Kang, H. C. & Kim, S. (2019). J. Korean Phys. Soc. 75, 494–497. Web of Science CrossRef CAS Google Scholar
Bergmann, U. & Cramer, S. P. (1998). Proc. SPIE, 3448, 198. CrossRef Google Scholar
Bokhoven, J. A. van & Lamberti, C. (2016). X-ray Absorption andX-ray Emission Spectroscopy: Theory and Applications. Chichester: Wiley. Google Scholar
Bryzgunov, V. A. (2000). Tech. Phys. 45, 46–49. CrossRef CAS Google Scholar
Bunău, O. & Joly, Y. (2009). J. Phys. Condens. Matter, 21, 345501. Web of Science PubMed Google Scholar
Dickinson, B., Seidler, G. T., Webb, Z. W., Bradley, J. A., Nagle, K. P., Heald, S. M., Gordon, R. A. & Chou, I. M. (2008). Rev. Sci. Instrum. 79, 123112. Web of Science CrossRef PubMed Google Scholar
Dikaya, O., Nachtegaal, M., Szlachetko, J., Ebner, K., Saveleva, V., Weder, N., Probst, B., Alberto, R., Serebrennikov, D., Clementyev, E., Maksimova, K., Goikhman, A. & Smolentsev, G. (2020). Results Phys. 18, 103212. CrossRef Google Scholar
Doriese, W. B., Abbamonte, P., Alpert, B. K., Bennett, D. A., Denison, E. V., Fang, Y., Fischer, D. A., Fitzgerald, C. P., Fowler, J. W., Gard, J. D., Hays-Wehle, J. P., Hilton, G. C., Jaye, C., McChesney, J. L., Miaja-Avila, L., Morgan, K. M., Joe, Y. I., O'Neil, G. C., Reintsema, C. D., Rodolakis, F., Schmidt, D. R., Tatsuno, H., Uhlig, J., Vale, L. R., Ullom, J. N. & Swetz, D. S. (2017). Rev. Sci. Instrum. 88, 053108. Web of Science CrossRef PubMed Google Scholar
Duan, P., Gu, S., Cao, H., Li, J. & Huang, Y. (2017). X-ray Spectrom. 46, 12–18. Web of Science CrossRef CAS Google Scholar
Ducotté, L., Glatzel, P., Marion, P., Lapras, C., Lesourd, M., Harris, A. & Heyman, C. (2010). Diamond Light Source Proc. 1, e26. Google Scholar
Fister, T. T., Seidler, G. T., Wharton, L., Battle, A. R., Ellis, T. B., Cross, J. O., Macrander, A. T., Elam, W. T., Tyson, T. A. & Qian, Q. (2006). Rev. Sci. Instrum. 77, 063901. Web of Science CrossRef Google Scholar
Glatzel, P., Bergmann, U., de Groot, F. M. F. & Cramer, S. P. (2001). Phys. Rev. B, 64, 045109. Web of Science CrossRef Google Scholar
Glatzel, P., Weng, T. C. T.-C., Kvashnina, K., Swarbrick, J., Sikora, M., Gallo, E., Smolentsev, N. & Mori, R. A. (2013). J. Electron Spectrosc. Relat. Phenom. 188, 17–25. CrossRef CAS Google Scholar
Groot, F. M. F. de & Kotani, A. (2008). Core Level Spectroscopy of Solids. New York: Taylor and Francis. Google Scholar
Hiraoka, N., Fukui, H., Tanida, H., Toyokawa, H., Cai, Y. Q. & Tsuei, K. D. (2013). J. Synchrotron Rad. 20, 266–271. Web of Science CrossRef CAS IUCr Journals Google Scholar
Holden, W. M., Hoidn, O. R., Ditter, A. S., Seidler, G. T., Kas, J., Stein, J. L., Cossairt, B. M., Kozimor, S. A., Guo, J., Ye, Y., Marcus, M. A. & Fakra, S. (2017). Rev. Sci. Instrum. 88, 073904. Web of Science CrossRef PubMed Google Scholar
Honkanen, A.-P., Verbeni, R., Simonelli, L., Moretti Sala, M., Monaco, G. & Huotari, S. (2014a). J. Synchrotron Rad. 21, 104–110. Web of Science CrossRef CAS IUCr Journals Google Scholar
Honkanen, A.-P., Verbeni, R., Simonelli, L., Moretti Sala, M., Al-Zein, A., Krisch, M., Monaco, G. & Huotari, S. (2014b). J. Synchrotron Rad. 21, 762–767. CrossRef CAS IUCr Journals Google Scholar
Huotari, S., Pylkkänen, T., Verbeni, R., Monaco, G. & Hämäläinen, K. (2011). Nat. Mater. 10, 489–493. Web of Science CrossRef CAS PubMed Google Scholar
Huotari, S., Sahle, C. J., Henriquet, C., Al-Zein, A., Martel, K., Simonelli, L., Verbeni, R., Gonzalez, H., Lagier, M.-C., Ponchut, C., Moretti Sala, M., Krisch, M. & Monaco, G. (2017). J. Synchrotron Rad. 24, 521–530. Web of Science CrossRef CAS IUCr Journals Google Scholar
Huotari, S., Vankó, Gy., Albergamo, F., Ponchut, C., Graafsma, H., Henriquet, C., Verbeni, R. & Monaco, G. (2005). J. Synchrotron Rad. 12, 467–472. Web of Science CrossRef CAS IUCr Journals Google Scholar
Jahrman, E. P., Holden, W. M., Ditter, A. S., Kozimor, S. A., Kihara, S. L. & Seidler, G. T. (2019). Rev. Sci. Instrum. 90, 013106. Web of Science CrossRef PubMed Google Scholar
Kavčič, M. (2012). X-ray Spectroscopy, edited by S. K. Sharma, pp. 81–98. Rijeka: InTech. Google Scholar
Kavčič, M., Budnar, M., Mühleisen, A., Gasser, F., Žitnik, M., Bučar, K. & Bohinc, R. (2012). Rev. Sci. Instrum. 83, 033113. Web of Science PubMed Google Scholar
Kleymenov, E., Van Bokhoven, J. A. J. A., David, C., Glatzel, P., Janousch, M., Alonso-Mori, R., Studer, M., Willimann, M., Bergamaschi, A., Henrich, B. & Nachtegaal, M. (2011). Rev. Sci. Instrum. 82, 065107. CrossRef PubMed Google Scholar
Kotani, A. & Shin, S. (2001). Rev. Mod. Phys. 73, 203–246. Web of Science CrossRef CAS Google Scholar
Kowalska, J. K., Lima, F. A., Pollock, C. J., Rees, J. A. & DeBeer, S. (2016). Isr. J. Chem. 56, 803–815. CrossRef CAS Google Scholar
Kuzmenko, D., Vogelsang, U., Hitz, S., Müller, D., Clark, A. H., Kinschel, D., Czapla-Masztafiak, J., Milne, C., Szlachetko, J. & Nachtegaal, M. (2019). J. Anal. At. Spectrom. 34, 2105–2111. CrossRef CAS Google Scholar
Lafuerza, S., Carlantuono, A., Retegan, M. & Glatzel, P. (2020). Inorg. Chem. 59, 12518–12535. Web of Science CrossRef CAS PubMed Google Scholar
Lafuerza, S., Retegan, M., Detlefs, B., Chatterjee, R., Yachandra, V., Yano, J. & Glatzel, P. (2020). Nanoscale, 12, 16270–16284. Web of Science CrossRef CAS PubMed Google Scholar
Li, D., Alpert, B. K., Becker, D. T., Bennett, D. A., Carini, G. A., Cho, H. M., Doriese, W. B., Dusatko, J. E., Fowler, J. W., Frisch, J. C., Gard, J. D., Guillet, S., Hilton, G. C., Holmes, M. R., Irwin, K. D., Kotsubo, V., Lee, S. J., Mates, J. A. B., Morgan, K. M., Nakahara, K., Pappas, C. G., Reintsema, C. D., Schmidt, D. R., Smith, S. R., Swetz, D. S., Thayer, J. B., Titus, C. J., Ullom, J. N., Vale, L. R., Van Winkle, D. D., Wessels, A. & Zhang, L. (2018). J. Low Temp. Phys. 193, 1287–1297. CrossRef CAS Google Scholar
Llorens, I., Lahera, E., Delnet, W., Proux, O., Braillard, A., Hazemann, J. L., Prat, A., Testemale, D., Dermigny, Q., Gelebart, F., Morand, M., Shukla, A., Bardou, N., Ulrich, O., Arnaud, S., Berar, J. F., Boudet, N., Caillot, B., Chaurand, P., Rose, J., Doelsch, E., Martin, P. & Solari, P. L. (2012). Rev. Sci. Instrum. 83, 063104. Web of Science CrossRef PubMed Google Scholar
Machek, P., Welter, E., Caliebe, W., Brüggmann, U., Dräger, G. & Fröba, M. (2007). AIP Conf. Proc. 879, 1755–1758. CrossRef CAS Google Scholar
Manceau, A., Enescu, M., Simionovici, A., Lanson, M., Gonzalez-Rey, M., Rovezzi, M., Tucoulou, R., Glatzel, P., Nagy, K. L. & Bourdineaud, J. P. (2016). Environ. Sci. Technol. 50, 10721–10729. CrossRef CAS PubMed Google Scholar
Mao, W. L., Mao, H. K., Meng, Y., Eng, P. J., Hu, M. Y., Chow, P., Cai, Y. Q., Shu, J. & Hemley, R. J. (2006). Science, 314, 636–638. CrossRef PubMed CAS Google Scholar
Miaja-Avila, L., O'Neil, G. C., Joe, Y. I., Alpert, B. K., Damrauer, N. H., Doriese, W. B., Fatur, S. M., Fowler, J. W., Hilton, G. C., Jimenez, R., Reintsema, C. D., Schmidt, D. R., Silverman, K. L., Swetz, D. S., Tatsuno, H. & Ullom, J. N. (2016). Phys. Rev. X, 6, 031047. Google Scholar
Moretti Sala, M., Martel, K., Henriquet, C., Al Zein, A., Simonelli, L., Sahle, C., Gonzalez, H., Lagier, M.-C., Ponchut, C., Huotari, S., Verbeni, R., Krisch, M. & Monaco, G. (2018). J. Synchrotron Rad. 25, 580–591. Web of Science CrossRef CAS IUCr Journals Google Scholar
Mortensen, D. R. & Seidler, G. T. (2017). J. Electron Spectrosc. Relat. Phenom. 215, 8–15. Web of Science CrossRef CAS Google Scholar
Neese, F. (2018). WIREs Comput. Mol. Sci. 8, e1327. Google Scholar
Nowak, S. H., Armenta, R., Schwartz, C. P., Gallo, A., Abraham, B., Garcia-Esparza, A. T., Biasin, E., Prado, A., Maciel, A., Zhang, D., Day, D., Christensen, S., Kroll, T., Alonso-Mori, R., Nordlund, D., Weng, T. C. & Sokaras, D. (2020). Rev. Sci. Instrum. 91, 033101. CrossRef PubMed Google Scholar
Rehr, J. J., Kas, J. J., Vila, F. D., Prange, M. P. & Jorissen, K. (2010). Phys. Chem. Chem. Phys. 12, 5503. Web of Science CrossRef PubMed Google Scholar
Retegan, M. (2020). Crispy v0.7.3, doi:10.5281/zenodo.1008184. Google Scholar
Rovezzi, M., Harris, A., Detlefs, B., Bohdan, T., Svyazhin, A., Santambrogio, A., Degler, D., Baran, R., Reynier, B., Noguera Crespo, P., Heyman, C., Van Der Kleij, H.-P., Van Vaerenbergh, P., Marion, P., Vitoux, H., Lapras, C., Verbeni, R., Kocsis, M. M., Manceau, A. & Glatzel, P. (2020). J. Synchrotron Rad. 27, 813–826. Web of Science CrossRef CAS IUCr Journals Google Scholar
Rovezzi, M., Lapras, C., Manceau, A., Glatzel, P. & Verbeni, R. (2017). Rev. Sci. Instrum. 88, 013108. Web of Science CrossRef PubMed Google Scholar
Rueff, J.-P. P. & Shukla, A. (2010). Rev. Mod. Phys. 82, 847–896. CrossRef CAS Google Scholar
Sa, J. (2014). High-Resolution XAS/XES. Boca Raton: CRC Press. Google Scholar
Schülke, W. (2007). Electron Dynamics by InelasticX-ray Scattering. Oxford University Press. Google Scholar
Smolentsev, G., Milne, C. J., Guda, A., Haldrup, K., Szlachetko, J., Azzaroli, N., Cirelli, C., Knopp, G., Bohinc, R., Menzi, S., Pamfilidis, G., Gashi, D., Beck, M., Mozzanica, A., James, D., Bacellar, C., Mancini, G. F., Tereshchenko, A., Shapovalov, V., Kwiatek, W. M., Czapla-Masztafiak, J., Cannizzo, A., Gazzetto, M., Sander, M., Levantino, M., Kabanova, V., Rychagova, E., Ketkov, S., Olaru, M., Beckmann, J. & Vogt, M. (2020). Nat. Commun. 11, 2131. Web of Science CrossRef PubMed Google Scholar
Sokaras, D., Nordlund, D., Weng, T.-C. C., Mori, R. A., Velikov, P., Wenger, D., Garachtchenko, A., George, M., Borzenets, V., Johnson, B., Qian, Q., Rabedeau, T. & Bergmann, U. (2012). Rev. Sci. Instrum. 83, 043112. CrossRef PubMed Google Scholar
Sokaras, D., Weng, T.-C. C., Nordlund, D., Alonso-Mori, R., Velikov, P., Wenger, D., Garachtchenko, A., George, M., Borzenets, V., Johnson, B., Rabedeau, T. & Bergmann, U. (2013). Rev. Sci. Instrum. 84, 053102. CrossRef PubMed Google Scholar
Suortti, P., Buslaps, T., Fajardo, P., Honkimäki, V., Kretzschmer, M., Lienert, U., McCarthy, J. E., Renier, M., Shukla, A., Tschentscher, T. & Meinander, T. (1999). J. Synchrotron Rad. 6, 69–80. Web of Science CrossRef CAS IUCr Journals Google Scholar
Szlachetko, J., Milne, C. J., Hoszowska, J., Dousse, J.-C., Błachucki, W., Sà, J., Kayser, Y., Messerschmidt, M., Abela, R., Boutet, S., David, C., Williams, G., Pajek, M., Patterson, B. D., Smolentsev, G., van Bokhoven, J. A. & Nachtegaal, M. (2014). Struct. Dyn. 1, 021101. Web of Science CrossRef PubMed Google Scholar
Szlachetko, J., Nachtegaal, M., Grolimund, D., Knopp, G., Peredkov, S., Czapla–Masztafiak, J. & Milne, C. (2017). Appl. Sci. 7, 899. Web of Science CrossRef Google Scholar
Tu, M.-F., Doumy, G., Al Haddad, A., March, A. M., Southworth, S. H., Assoufid, L., Kumagai, Y., Walko, D. A., DiChiara, A. D., Liu, Z., Shi, B., Young, L. & Bostedt, C. (2019). J. Synchrotron Rad. 26, 1956–1966. Web of Science CrossRef CAS IUCr Journals Google Scholar
Uruga, T., Tada, M., Sekizawa, O., Takagi, Y., Yokoyama, T. & Iwasawa, Y. (2019). Chem. Rec. 19, 1444–1456. CrossRef CAS PubMed Google Scholar
Vankó, G., Bordage, A., Glatzel, P., Gallo, E., Rovezzi, M., Gawelda, W., Galler, A., Bressler, C., Doumy, G., March, A. M., Kanter, E. P., Young, L., Southworth, S. H., Canton, S. E., Uhlig, J., Smolentsev, G., Sundström, V., Haldrup, K., van Driel, T. B., Nielsen, M. M., Kjaer, K. S. & Lemke, H. T. (2013). J. Electron Spectrosc. Relat. Phenom. 188, 166–171. Google Scholar
Wernet, P. (2019). Philos. Trans. R. Soc. A. 377, 201704. CrossRef Google Scholar
Zimmermann, P., Peredkov, S., Abdala, P. M., DeBeer, S., Tromp, M., Müller, C. & van Bokhoven, J. A. (2020). Coord. Chem. Rev. 423, 213466. Web of Science CrossRef Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.