Computational methods for automated center determination in electron diffraction patterns

crossmark logo

aInstitute of Scientific Instruments, Kralovopolska 147, Brno, 61200, Czechia, bFaculty of Mechanical Engineering, Brno University of Technology, Technicka 2896/2, Brno, 61669, Czechia, cInstitute of Macromolecular Chemistry, Heyrovskeho nam. 1888/2, Prague 6, 16200, Czechia, and dCEITEC BUT, Brno University of Technology, Purkynova 123, Brno, 61200, Czechia
*Correspondence e-mail: [email protected]

Edited by V. Hathwar, Goa University, India (Received 10 December 2025; accepted 6 March 2026; online 30 April 2026)

Accurate center detection in electron diffraction patterns is critical for all subsequent processing of experimental diffractograms. This study presents and compares several automated approaches – maximum intensity detection, phase cross-correlation, autocorrelation-based detection, pseudo-Voigt profile fitting and Hough-transform-based detection – applied to both polycrystalline diffractograms with characteristic diffraction rings and single-crystal diffractograms showing discrete diffraction spots. The methods were evaluated in terms of accuracy, robustness, speed, preprocessing requirements and applicability across diverse materials that produce a variety of diffraction patterns. Our findings provide practical guidance for selecting center detection techniques in automated diffractogram processing workflows, thus facilitating improved data quality and reliability in crystallographic analyses. Phase cross-correlation has been proven to deliver high performance consistently on polycrystalline diffractograms with diffraction rings, while pseudo-Voigt profile fitting is best suited to monocrystal-like diffractograms with discrete diffraction spots. All the above-mentioned algorithms have been implemented in the recent version of our open-source Python package EDIFF, which now offers a user-friendly, flexible and fully automated solution for center detection in diffractograms. These algorithms determine the center of the individual two-dimensional diffraction patterns, while the processing of complete three-dimensional electron diffraction or four-dimensional scanning transmission electron microscopy datasets often includes accurate center determination as part of structure refinement workflows.

1. Introduction

The two most widely used methods in transmission electron microscopy (TEM) are bright-field imaging and electron diffraction (Fultz & Howe, 2002View full citation). The importance of electron diffraction was further amplified by the development of four-dimensional scanning transmission electron microscopy (4D-STEM) methods, which are available in both transmission (Ophus, 2019View full citation) and scanning electron microscopy (4D-STEM-in-SEM; Slouf et al., 2025View full citation). The typical result of an electron diffraction measurement is a 2D image (diffraction pattern) containing the most intense spot at the center (the spot of the primary beam, which may be hidden below a beamstop or weakened due to dynamic diffraction effects in some special cases) and units to thousands of weaker diffraction spots. Single-crystal diffraction patterns show individual diffraction spots, while polycrystalline diffraction patterns show multiple spots forming typical diffraction rings (Fultz & Howe, 2002View full citation). In each case, the typical analysis of a diffraction pattern starts with the accurate location of the primary beam position, even if it is eclipsed by the beamstop. The primary beam position defines the center of the reciprocal lattice (Gard, 1976View full citation) and acts as the reference point for all diffraction spots.

Fully automated accurate location of the primary beam position in an arbitrary electron diffraction pattern may be surprisingly challenging. As a result, we have not found any simple, lightweight, open-source and truly universal software for reliable center determination in electron diffractograms. Historically, computational center detection has evolved from fully manual to partially automated methods. One of the earliest widely used tools, Process Diffraction (Lábár, 2000View full citation; Lábár, 2005View full citation), determines the diffractogram center via the manual positioning of a model diffraction ring onto the experimental diffraction pattern, which works well for TEM diffractograms but fails for modern high-throughput 4D-STEM data. These methods were subsequently complemented by algorithms focused on patterns with diffuse diffraction rings, which relied on manual pre-selection of a diffuse peak, sector-based integration and nonlinear least-squares fitting (Lábár & Das, 2017View full citation). Modern packages like CrysTBox (Klinger & Jäger, 2015View full citation; Klinger, 2017View full citation) introduced automatic routines for processing diffraction patterns, but CrystTBox remains closed source (MATLAB based) and GUI driven (unsuitable for batch processing). More recent open-source comprehensive Python frameworks that are connected with the recent development of 4D-STEM methods, such as Py4DSTEM (Savitzky et al., 2019View full citation, 2021View full citation), LiberTEM (Clausen et al., 2020View full citation) and pyxem (Cautaerts et al., 2022View full citation), have broad­ened the automation possibilities, but their focus on 4D-STEM datasets limits their application to other types of diffractograms. Their primary algorithms for center detection rely on the presence of a strong central spot. For example, py4DSTEM's center finding routine explicitly requires that the central spot is visible and intense, precluding its use for diffractograms with beamstops that are typical of TEM (Savitzky et al., 2021View full citation). Analogously, pyxem offers basic center-of-mass utilities, supporting masking and thresholding, but lacks a dedicated general-purpose diffraction pattern center detection module (Cautaerts et al., 2022View full citation). Finally, LiberTEM excels at scalable distributed processing of STEM data like the two previously mentioned packages, but it relies on center-of-mass center detection as far as the processing of individual diffractograms is concerned (Clausen et al., 2020View full citation). However, all three of the above-listed packages (pyxem, LiberTEM and py4DSTEM) also contain computationally intensive algorithms that, among other outputs, yield refined center positions for single-crystal diffraction patterns with arbitrary orientation. These algorithms typically operate in three steps: (i) automatic detection of diffraction spots, (ii) matching the detected spots to an expected or precalculated diffraction pattern, and (iii) using the resulting fit for precise determination of pattern parameters, including the position of the central beam. More information on modern software for diffraction pattern processing can be found in Table S1 in the supporting information.

In addition to the software frameworks discussed above, a few special-case focused approaches for center determination exist. In zone-axis single-crystal diffraction, the center can be refined by exploiting the inversion symmetry of ±g (Friedel) reflection pairs, whose midpoints coincide with the true reciprocal-space origin (Fultz & Howe, 2002View full citation; Zuo & Spence, 2017View full citation). In ring-based patterns, center and detector distortions such as ellipticity are commonly refined simultaneously by fitting concentric rings or through geometric calibration (Pekin et al., 2017View full citation; Savitzky et al., 2019View full citation). In 3D electron diffraction, also known as electron diffraction tomography or MicroED), the accurate center position is typically refined as a part of unit-cell indexing and subsequent refinement workflows. In 4D-STEM workflows, the center is often determined as part of a global lattice-fitting procedure, as mentioned at the end of the previous paragraph and discussed elsewhere (Savitzky et al., 2019View full citation).

In this work, we implemented and tested a range of promising, easy-to-use, fully automated and universal algorithms for reliable center detection in arbitrary 2D electron diffraction patterns. These algorithms do not require symmetric diffraction patterns or computationally intensive detection and fitting of all diffraction spots. These algorithms have been implemented and integrated into our open-source Python package for processing electron diffraction patterns, which is named EDIFF (https://pypi.org/project/ediff). We employed both traditional well established approaches and less common techniques that have not been designed specifically for diffraction center location.

The traditional center-of-mass method estimates the center from the intensity distribution under the assumption that the direct beam or low-angle scattering dominates near the true center (Wang et al., 2022View full citation; Thronsen et al., 2024View full citation). An extension of the center-of-mass method is the fitting of a central spot by means of a two-dimensional pseudo-Voigt function (Xu, 2009View full citation; Zhang et al., 2022View full citation), and the fitted parameters may provide a better estimate of the central peak position (Abbondanza et al., 2021View full citation). Another traditional approach, based on manual center estimation by a three-point method combined with automatic center refinement, is available in our package as a control if some of the fully automated methods should fail (Slouf et al., 2025View full citation). Cross-correlation methods exploit the inherent symmetry of diffraction patterns, as explored by Pekin et al. (2017View full citation) and René De Cotret et al. (2018View full citation). The Hough transform, specifically in its circle-detection variant, has been applied to detect concentric ring patterns in powder diffraction data (Du & Yang, 2009View full citation; Adatrao & Mittal, 2016View full citation). All six methods (center-of-mass, pseudo-Voigt fitting, manual three-point, autocorrelation, phase cross-correlation and Hough transform) were applied to tens of diffraction patterns from real TEM and 4D-STEM experiments, which yielded both mono- and polycrystalline diffractograms with or without a beamstop. The position of the primary beam could be located in each of the tested diffractograms by at least one of the automatic center detection methods.

2. Materials and methods

2.1. Experimental electron diffraction patterns

All diffraction patterns analyzed in this study come from experiments conducted in the laboratory of the second and last authors of this article (MS and VK). Measurements were carried out either in a standard transmission electron microscope using selected-area electron diffraction (TEM/SAED) or in a high-resolution scanning electron microscope equipped with an array detector of transmitted electrons by means of the 4D-STEM method [4D-STEM-in-SEM, as described in our previous studies (Slouf et al., 2021aView full citation, 2021bView full citation, 2025View full citation)]. For the present analysis, the data were partitioned into two datasets, as summarized in Table 1[link].

Table 1
Datasets for testing center detection methods

Dataset No. Description Details
D1 15 Polycrystalline diffraction patterns exhibiting circular patterns, acquired either via TEM/SAED or 4D-STEM-in-SEM Table S2
D2 75 Single-crystal diffraction patterns, acquired via 4D-STEM-in-SEM Table S3

2.2. Software for automatic location of primary beam position

The position of the primary beam is readily identified when no beamstop is used but becomes more challenging when the beamstop obscures the central spot. To handle both scenarios, we have implemented multiple algorithms for primary beam location in our open-source EDIFF package (Section 2.2.1[link]), which provides end-to-end processing of electron diffraction patterns. In particular, we have substantially extended the ediff.center module of the package (Section 2.2.2[link]).

2.2.1. EDIFF package for processing of electron diffraction patterns

EDIFF is an open-source Python package (https://pypi.org/project/ediff). It processes experimental electron diffraction (ED) patterns, computes the theoretical X-ray diffraction (XRD) pattern for a candidate crystal and compares ED with XRD to confirm the crystal structure. The package has evolved from a set of Python scripts (Slouf et al., 2021aView full citation, 2021bView full citation) through a standard Python module focused on powder diffractograms (Slouf et al., 2025View full citation) to the current version that supports processing of both polycrystalline and single-crystal diffractograms (as detailed below). The first step towards handling arbitrary diffraction patterns is a more general module for accurate location of the primary beam position, which is the main subject of this work.

2.2.2. ediff.center module for accurate location of primary beam position

The ediff.center module (https://mirekslouf.github.io/ediff/docs/pdoc.html/ediff/center.html) focuses on locating the primary beam position. The overall workflow – center location – comprises center detection and center refinement. The module implements six detection algorithms and three refinement algorithms; each algorithm is exposed as a method of the CenterLocator class (Fig. 1[link]).

[Figure 1]
Figure 1
Jupyter Notebook interface for the ediff.center module. (a) Example code snippet for automated center detection of an ED pattern. The workflow demonstrates how to initialize the CenterLocator, specify the detection method (here phase cross-correlation) and refinement strategy (here intensity sum), and obtain the numerical output for both the initially detected and refined center positions. (b) Visualization of the results, showing the diffraction pattern with overlaid markers: the yellow circle and cross mark the detected center, while the red circle and cross indicate the refined center after intensity-based adjustment. Reported coordinates and integrated intensities allow users to evaluate the accuracy of the location.

The six detection algorithms are summarized in Table 2[link]. Only the manual method requires user input; all others are fully automated. Several center detection methods benefit from operating on edge-enhanced versions of the input image, as reported previously (Pekin et al., 2017View full citation). Accordingly, the ediff.center module includes optional preprocessing using the Sobel edge-detection operator (Pekin et al., 2017View full citation).

Table 2
List of center detection methods

Method Brief description
manual User clicks three ring points and the center is triangulated
intensity Calculates maximum intensity within a central region
curvefit Fits a pseudo-Voigt profile to the central peak
phase Phase cross-correlation detection of inversion symmetry
ccor Maximum autocorrelation peak
hough Detection of circular features using Hough transform
none No center detection, the center coordinates are inserted or read from a file
†Detailed descriptions of all methods are given in the supporting information, Section S1.2.

The three refinement algorithms are summarized in Table 3[link]. Again, only the manual method requires user input; the other two are fully automated. All three algorithms are based on small shifts of a diffraction ring that passes through a set of strong, approximately equidistant, reflections. The optimized variables are the center coordinates and the ring radius. Fig. 1[link] illustrates the approach: rings from the detection and refinement stages are shown in yellow and red, respectively, on the diffractogram.

Table 3
List of center refinement methods

Method Brief description
manual Manual refinement – shifts of overlaid diffraction ring using keyboard controls
sum Radial intensity sum maximization – auto-shifts of overlaid diffraction ring
var Radial intensity variance minimization – auto-shifts of overlaid diffraction ring
none No refinement – the coordinates from the center detection step are unchanged

ED datasets may exhibit elliptical distortion due to external fields or detector imperfections (Hou & Li, 2008View full citation). Although such distortion is typically minimal in most modern devices, the ediff.center module offers a simple correction method that can be applied as part of the center detection step.

2.2.3. Center detection testing workflow

Each method was applied to both test datasets (Table 1[link]). Reference center coordinates were obtained using the manual method followed by automatic refinement (sum). All other methods were evaluated without additional refinement to isolate their intrinsic performance. Detection was performed on both raw images and their edge-enhanced counterparts (as described in Section 2.2.2[link]).

After center coordinates had been determined for all images in both datasets, we used another EDIFF tool to compute one-dimensional radially averaged intensity profiles from the diffraction patterns – a procedure that is highly sensitive to center location accuracy. Errors in the coordinates directly affect the sharpness, symmetry, peak positions and peak heights of the resulting profiles and thus propagate into subsequent analyses.

Although the width and height of diffraction peaks are primarily governed by the accuracy of center location, image distortions such as the aforementioned ellipticity (Section 2.2.2[link]) may also contribute. However, in modern datasets acquired with calibrated detectors, significant elliptical distortion is uncommon. The data analyzed in this study do not exhibit significant distortions that would affect the comparative evaluation of the tested methods.

2.3. Electron diffractogram classification

We divided the dataset into polycrystalline (Fig. 2[link]) and single-crystal (Fig. 3[link]) diffractograms. The polycrystalline diffraction patterns show characteristic rings, while the single-crystal diffraction patterns show diffraction spots. Within each group, every pattern was annotated using morphology- and artifact-based metrics derived from radial and azimuthal intensity distributions, for example ring completeness, eccentricity, spot density and signal-to-noise ratio (SNR). The resulting heterogeneity precluded a single uniform processing pipeline; accordingly, all subsequent analyses are stratified by class to ensure a robust and comparable evaluation. Representative examples of each category are shown in Fig. 2[link], Fig. 3[link] and Table 4[link].

Table 4
Classification of ED patterns

Class Description Sample
BB Beamstop blocks primary beam Polycrystalline (rings)
BR Broken rings (incomplete Debye–Scherrer rings) Polycrystalline (rings)
RA Reconstruction of artifacts from deconvolution Polycrystalline (rings)
SNB Strong noise and background Polycrystalline (rings)
FB Flat (ideal) background Polycrystalline (rings)
P Perfect quality Single crystal (spots)
G Good quality Single crystal (spots)
I Intermediate quality Single crystal (spots)
B Bad quality Single crystal (spots)
W Worst quality Single crystal (spots)
BB, BR, RA, SNB and FB are diffractogram classifications based on image features. P, G, I, B and W are diffractogram classifications based on image quality in terms of noise and background levels. In Fig. 2 there are typical diffractograms from each category.
[Figure 2]
Figure 2
Classification of polycrystalline diffraction patterns. We categorize them on the basis of center detection challenges: (a), (b) and (d) with beamstops, (b) broken rings, (c) reconstruction artifacts, (c) and (d) strong noise/background, and (e) flat background. Samples: (a) and (e) Au nano-islands, (b) Ag crystals, (c) Fe3O4 nanoclusters, and (d) ZnO–CuO nanoparticles.
[Figure 3]
Figure 3
Classification of single-crystal diffraction patterns. We categorize them on the basis of their quality (SNR, background intensity), sorted from (a) best quality to (e) worst quality. Samples: (a) and (b) Au nano-islands, (c) TbF3 nanocrystalline aggregates, (d) Fe3O4 nanoclusters, and (e) LaF3 nanoparticles in amorphous biological tissue.

There is a specific case of single-crystal diffraction patterns with a beamstop. They come from traditional transmission electron microscopes where a beamstop was present to protect the camera for both single-crystal and polycrystalline diffractograms. With the advent of modern pixelated detectors, the dynamic range is sufficient to record the central beam without the need for a beamstop. Nevertheless, for completeness, the performance of the center-detection methods is also demonstrated on single-crystal data containing a beamstop (Fig. S1).

Importantly, diffraction patterns cannot be classified solely on the basis of the sample material. The difficulty of diffraction pattern processing depends on multiple factors, as described above. While different materials may statistically predispose diffraction patterns to stronger or weaker scattering under given experimental conditions, the classification used here is based on the actual properties of the recorded diffraction patterns and on the characteristics of the resulting reconstructed powder diffractograms.

3. Results

3.1. Center detection in polycrystalline diffraction patterns

In polycrystalline diffraction patterns, automated center detection tends to be easy and reliable. Polycrystalline diffraction patterns exhibit approximate centrosymmetry due to concentric Debye–Scherrer rings. This centrosymmetry, which is absent from the great majority of single-crystal diffraction patterns, provides strong constraints for automated center detection.

For all polycrystalline diffractograms, the manual center detection method (Table 2[link]), followed by refinement with the sum method (Table 3[link]), yielded the center coordinates, which we then adopted as the reference standard. We selected this protocol for its simplicity, reliability and direct user control. In addition, the subsequent automatic refinement mitigates possible user-placement errors introduced during the manual step. Finally, the manual method itself is not included among the fully automated approaches under comparison.

The comparison of the performance of all six center detection algorithms is summarized in Fig. 4[link]. The phase center detection method (Section S1.2.3) emerged as the most robust, accurate and generalizable. It performed well regardless of the presence of the beamstop, delivered sub-pixel accuracy and outperformed the reference method in approximately 67% of cases, as shown by sharper and higher peaks in the resulting 1D profiles (Fig. 5[link]).

[Figure 4]
Figure 4
Heatmap illustrating the performance of center detection methods on the polycrystalline dataset D1. Each row corresponds to a diffractogram (numbered according to Table S2; blue-highlighted samples 8, 9, 13, 14 and 15 are from 4D-STEM-in-SEM, while the rest are from TEM) and each column represents a center detection method (manual = three-point detection, phase = phase cross-correlation, ccorr = autocorrelation, intensity = maximum intensity detection, curvefit = pseudo-Voigt profile fitting and hough = Hough transform). Hatched columns indicate the method's performance on pre-processed diffractograms (edge detection using a Sobel operator). The color bar on the right indicates the performance scale: the lightest color (yellow) indicates the most accurate detection, while the darkest color (black) signifies failure or substantial inaccuracy in center location.
[Figure 5]
Figure 5
Performance of center detection methods on representative polycrystalline diffractograms corresponding to Fig. 2. Column 1: 1D radially averaged profiles computed using each method's center (overlaid, color coded). Column 2: enlargement of the dominant peak (red box in Column 1), showing the top-performing method(s). Column 3: original diffractogram with detected centers (color-coded crosses). Column 4: enlargement of the central region (blue box in Column 3) around the direct (primary) beam and the detections. Samples: (a) and (e) Au nano-islands, (b) Ag crystals, (c) Fe3O4 nanoclusters, and (d) ZnO–CuO nanoparticles.

Three methods – intensity, ccorr and curvefit – worked well if the primary beam spot was present (samples 8, 9, 13, 14 and 15 in Fig. 4[link]), but they failed when the primary beam was blocked by a beamstop [samples (a), (b) and (d) in Fig. 2[link]]. The hough method showed intermediate performance – superior to the three above-mentioned methods relying on the primary beam spot, yet still insufficiently robust, as it remained vulnerable to diffractogram asymmetry introduced by the beamstop.

Regarding pre-processing, generic edge detection provided no consistent benefit, except in the case of the hough method, where it proved almost essential (see the differences highlighted by the hatched columns in the heatmap in Fig. 4[link], or the corresponding 1D profiles labeled with an `-s' suffix in the legend in Fig. 5[link]). In our tests, the Canny edge detector outperformed the Sobel operator for polycrystalline diffractograms.

3.2. Center detection in single-crystal diffraction patterns

Center detection in single-crystal diffraction patterns proved more challenging than in polycrystalline diffractograms. Monocrystalline patterns often lack strong radial symmetry and contain fewer dominant features, reducing the number of reliable cues available to automatic location algorithms.

For all diffractograms, the intensity center detection method (Table 2[link]) followed by refinement with the manual method (Table 3[link]) yielded the center coordinates which we adopted as the reference standard. Unlike the powder diffractograms in Section 3.1[link], automated refinement by means of the sum method could not be used for single-crystal patterns because it assumes a concentric near-symmetric intensity distribution about the true center – a hallmark of polycrystalline rings but not of single-crystal spot patterns. Likewise, manual detection which exploits continuous diffraction rings was inapplicable here, as single-crystal patterns lack sufficient spots (≥3) at identical radii.

The heatmap in Fig. 6[link] shows substantial performance variability; no automatic method outperformed the reference approach on the single-crystal dataset, so manual centering remained necessary for optimal accuracy on individual patterns. For large-scale datasets (e.g. 4D-STEM-in-SEM) automation is still essential; in our 4D-STEM/powder nanobeam diffraction (NBD) pipeline, patterns unsuitable for reconstruction are filtered automatically and several decision criteria depend on the detected center (Slouf et al., 2021aView full citation).

[Figure 6]
Figure 6
Heatmap illustrating the performance of center detection methods on the single-crystal dataset D2. Each row corresponds to an individual diffractogram [15 patterns per material: (1) Au, (2) TbF3, (3) Fe3O4, (4) LaF3 and (5) GdF3]. Each column represents a center detection method (manual = three-point detection, phase = phase cross-correlation, ccorr = autocorrelation, intensity = maximum intensity detection, curvefit = pseudo-Voigt profile fitting, hough = Hough transform). Hatched columns indicate a method's performance on pre-processed diffractograms (edge detection using a Sobel operator). The color bar on the right indicates the performance scale: the lightest color (yellow) indicates the most accurate detection, while the darkest color (black) signifies failure or substantial inaccuracy in center location. The groups (1)–(5) correspond to various diffractogram qualities, as explained in Section 2.3[link], Table 4 and Table S3. The quality is as follows: (1) perfect, (2) good, (3) bad, (4) poor and (5) intermediate.

Method-wise, the phase detector underperformed on the single-crystal dataset, ccorr showed near-total failure across individual patterns, hough failed entirely and edge-detection preprocessing did not measurably improve correlation-based methods. By contrast, intensity and curvefit were the most effective. The curvefit method succeeded across all samples (as indicated by the absence of dark rows in the columns corresponding to this method in the heatmap in Fig. 6[link]). The intensity approach was less reliable for low-quality LaF3 patterns, where no clear Bragg peaks were visible beyond the saturated central beam (note several dark rows in the intensity column for the LaF3 sample in Fig. 6[link]). The performance of center detection methods on representative single-crystal diffractograms is shown in Fig. 7[link].

[Figure 7]
Figure 7
Performance of center detection methods on representative single-crystal diffractograms corresponding to Fig. 3. Column 1: 1D radially averaged profiles computed using each method's center (overlaid, color coded). Column 2: enlargement of the dominant peak (black box in Column 1), marked with black arrows, showing the top-performing method(s). Column 3: original diffractogram with detected centers (color-coded crosses). Column 4: enlargement of the central region (blue box in Column 3) around the direct (primary) beam and the detections. Samples: (a) and (b) Au nano-islands, (c) TbF3 nanocrystalline aggregates, (d) Fe3O4 nanoclusters, and (e) LaF3 nanoparticles in amorphous biological tissue.

Examples of single-crystal diffraction patterns with beamstops (Fig. S1), along with the performance of the corresponding center detection methods, are provided in the supporting information (Section S1.3).

3.3. Computational efficiency

The computational efficiency of all methods was benchmarked for speed (Table S4). The intensity method was the fastest, locating the center within hundredths of a second per pattern, whereas the phase method was slower but consistently delivered the highest accuracy. The curvefit method exhibited dataset-dependent runtimes: on the polycrystalline dataset it was relatively fast, while on the single-crystal diffractograms it was the slowest. Edge-detection preprocessing using the Sobel operator had minimal impact on the runtime of most methods. By contrast, combining the Hough detector with Canny edge preprocessing substantially increased runtime.

4. Discussion

4.1. Diffractogram processing challenges

4.1.1. Polycrystalline diffractograms

Guided by the classification in Section 3.1[link] (Fig. 2[link], Fig. 3[link] and Table 4[link]), we identified four recurrent challenges that impede robust automated center detection in polycrystalline diffractograms exhibiting Debye–Scherrer rings:

(i) Beamstop blocked (BB) diffractograms [Fig. 2[link](a)]. Typically in TEM/SAED, a beamstop hides the primary beam spot and covers some part of the diffraction rings. The resulting absence of a central spot, an asymmetrical diffractogram and abrupt intensity drops around the beamstop distort intensity-based and pixel-wise detection algorithms.

(ii) Diffractograms with broken rings (BR) [Fig. 2[link](b)]. Common in samples containing mixed micro- and nano­crystals, broken rings are formed by discrete spots at similar radial distances from the center. Although they may appear continuous to the eye, their segmented nature disrupts the rotational symmetry and weakens the radial periodicity, often leading automated methods to lock onto local intensity maxima.

(iii) Reconstruction artifacts (RA) [Fig. 2[link](c)]. Cross-like patterns centered on the beam spot may arise in 2D diffractograms reconstructed using point spread function deconvolution (e.g. from the STEMDIFF library; https://pypi.org/project/stemdiff/). While they do not compromise radial averaging or crystallographic interpretation, they bias algorithms that rely on intensity gradients or symmetry, impairing automated detection.

(iv) Strong noise and background (SNB) [Fig. 2[link](d)]. Aggregating individual patterns introduces a Gaussian-like background with oversaturation and elevated noise near the center, the most critical region for location.

We also identify one further group that is straightforward for center detection, which is rare but close to ideal for 1D profile calculation:

(v) Flat-background (FB) diffractograms [Fig. 2[link](e)]. There is no Gaussian-like background and diffraction features remain sharp with minimal noise.

We emphasize that these categories are not mutually exclusive. For example, TEM/SAED diffractograms usually include a beamstop and may also exhibit strong noise (Table S2).

4.1.2. Single-crystal diffractograms

In addition to polycrystalline patterns, we evaluated single-crystal patterns because their quality governs the reconstructed composite 2D diffractogram (with Debye–Scherrer rings) and, consequently, the downstream classification. We therefore categorize them as perfect (P) > good (G) > intermediate (I) > bad (B) > worst (W) according to their background/noise levels and peak clarity. Perfect patterns yield clean reconstructions, whereas the worst ones produce noisy diffractograms with barely visible rings (Fig. 3[link]). As described at the end of Section 2.3[link], patterns must be categorized on the basis of their observable diffraction characteristics and processing challenges, rather than on sample material alone.

A special subset of single-crystal diffractograms contains a beamstop (Fig. S1). When the spot pattern is approximately centrosymmetric our algorithms can still succeed, but these cases are generally challenging. The beamstop hides the direct beam and its high-intensity neighborhood, as well as disrupting azimuthal information so that the already sparse Bragg spots (compared with continuous rings) exhibit stronger asymmetry and broken periodicity.

4.2. Assessment of centering accuracy through radial profiles

All methods were applied to the available 2D diffraction patterns and their performance could be assessed by analyzing the resulting 1D radially averaged distribution profiles, as illustrated in Fig. 8[link]. The initial pass – deciding whether a method clearly failed or succeeded – is typically straight­forward: incorrect center location yields visibly distorted or flattened profiles with poorly resolved peaks. A typical example of a poorly suited method for a given diffractogram type is the hough method in Fig. 8[link](a), which yielded a completely wrong 1D profile.

[Figure 8]
Figure 8
Comparison of detection methods in diffractograms. (a) and (e) One-dimensional radially averaged profiles of diffractograms calculated from center locations obtained by all detection methods (manual = three-point detection, phase = phase cross-correlation, ccorr = autocorrelation, intensity = maximum intensity detection, curvefit = pseudo-Voigt profile fitting and hough = Hough transform; methods with an `-s' suffix were applied to edge-detected diffractograms). (b) and (f) Close-up views of the most intense diffraction peak [highlighted by the red rectangle in panels (a) and (e), respectively], illustrating how inaccuracies in center location affect the peak's intensity, location and shape. (c) and (g) Diffractograms with detected center locations. (d) and (h) Close-up views of the detected center positions. The samples consisted of (top row) Fe3O4 nanoclusters with an organic shell using 4D-STEM-in-SEM and (bottom row) Au nanoclusters.

The comparison of the accuracy of those methods that seemed to succeed requires closer inspection. Subtle deviations in center coordinates still degrade peak symmetry, sharpness etc. Differences that may seem negligible in the 2D diffractogram [Figs. 8[link](c), 8[link](d), 8[link](g) and 8[link](h)] become evident in selected parts of the corresponding radial profiles [Figs. 8[link](b) and 8[link](f)]: exact peak intensities, sharpness and positions may change, producing a distorted and/or unreliable basis for subsequent crystallographic analysis.

Although discrete well separated diffraction peaks are relatively tolerant to minor center-position errors, the impact is pronounced in regions where weaker peaks are closely spaced or even overlapping. In such cases, especially in the presence of background and noise, peaks may broaden, merge or vanish altogether, undermining the reliability of peak identification.

4.3. Detailed comparison of the implemented center detection methods

4.3.1. Polycrystalline diffractograms

For polycrystalline diffraction patterns, the phase method proved to be the most universal. The superiority of the phase method stems from the fact it operates in the frequency domain and exploits structural symmetry and spatial correlation rather than absolute intensity. The method is tolerant to moderate asymmetry of a diffraction pattern due to a beamstop and is broadly generalizable across acquisition modalities, because the center is inferred from relative pattern similarity rather than local peak characteristics. While other robust spatial-domain approaches, such as sector-wise nonlinear least-squares fitting for low SNR data (Lábár & Das, 2017View full citation), have been successfully applied to particularly challenging diffraction patterns, the phase method provides a computationally efficient alternative in the frequency domain that effectively exploits intrinsic structural symmetry.

In contrast, the intensity and curvefit methods rely on the presence of a discernible central peak; when the region is masked or distorted by a beamstop, the methods are prone to failure. The ccorr method approach relies on relatively intact rotational symmetry and periodicity, so broken or discontinuous rings reduce its power. The hough method benefits from high-quality continuous ring arcs but remains sensitive to asymmetry and to the missing data blocked by the beamstop.

Preprocessing effects align with these mechanisms: generic edge maps can suppress informative low-frequency content needed by correlation-based methods, whereas hough explicitly requires clean edge contours. For this task, Canny's double-thresholding with hysteresis and Gaussian smoothing yield more reliable ring edges than the Sobel method (Lynn et al., 2021View full citation).

4.3.2. Single-crystal diffractograms

For single-crystal diffraction patterns the situation is more complex due to their lower symmetry, as already discussed above in Section 3[link]. The phase approach, which excelled on polycrystalline patterns, assumes symmetric shifts and roughly consistent azimuthal intensity around the center, but single-crystal diffractograms are in general asymmetric, often noisy and/or with high background, which leads to center mis-location. The ccorr method depends on recurring structures (such as the concentric rings in powder diffractograms); with isolated diffraction spots it can return false positives because such spots correlate strongly with a central peak template (Liu et al., 2015View full citation; Brewster et al., 2018View full citation). The hough detector is ineffective when circular features are absent or fragmentary. Conversely, intensity and curvefit operate on local intensity structures rather than global symmetry, making them better suited to patterns dominated by the direct beam or discrete Bragg spots. However, the intensity method is particularly vulnerable to detector saturation: when the dynamic range of the detector is exceeded near the central beam, the brightest pixels are clipped to the maximum recordable value. This distortion masks the true intensity maximum and can systematically misplace the estimated center, the extent of which also affects the detection: if the area is too small, the true maximum may fall outside the search window, while if it is too large, spurious local maxima or noise can bias the result. Both methods are also sensitive to a poor SNR, where random noise can generate false local maxima or distort intensity gradients, thereby misleading the algorithm and shifting the detected center away from the true position.

The special case of single-crystal diffraction patterns (usually asymmetric) with a beamstop (hiding the central spot) is the most challenging. Fortunately, it is also the rarest case. In the field of TEM, single-crystal diffraction patterns are often aligned by the user along a defined zone axis, which makes them symmetric, and, as a result, the center can be found using the phase method, regardless of the beamstop. In the field of 4D-STEM (in both transmission and scanning electron microscopes), the individual diffraction patterns are usually asymmetric, but the pixelated detectors employed in 4D-STEM do not require a beamstop, which makes the center detectable with the intensity or curvefit methods.

4.4. Computational efficiency

Computational efficiency is a critical factor for automated center detection on large datasets where manual annotation is impractical. For conventional diffraction techniques such as TEM/SAED or TEM/NBD, only individual diffractograms are acquired, and a difference of a few fractions of a second when finding the diffractogram center is unimportant. In contrast, modern 4D-STEM experiments generate thousands of diffraction patterns within a single dataset. In such cases, the cumulative computational load becomes substantial, as the center must be detected accurately in each pattern.

While faster methods offer clear time savings, accuracy remains the primary criterion. For example, although the intensity method was the fastest, it performed poorly on data featuring beamstops, making slower but more reliable approaches preferable. The phase method, despite being slower, consistently achieved the highest accuracy.

The dataset-dependent behavior of curvefit suggests sensitivity to data quality: on a dataset with powder diffraction patterns (Fig. 2[link]), fast but unreliable performance may reflect convergence to local rather than global optima, while on a dataset with single-crystal diffraction patterns (Fig. 3[link]) the method was slower yet more accurate, probably because clearer peak profiles allowed more effective parameter-space exploration. Similarly, the increase in runtime for the hough method coupled with Canny pre-processing can be attributed to the suppression of weaker edges, reducing the feature density and prolonging voting and peak selection.

4.4.1. Practical guide: when to use which method

We distilled our results into a practical guide for selecting a center-detection method (Table 5[link]). The decision table maps common pattern characteristics to recommend reliable algorithms. Generally, ccorr should be avoided: it failed nearly completely on monocrystals and was inconsistent on polycrystalline diffractograms. phase is the default for 2D polycrystalline diffractograms with Debye–Scherrer rings; curvefit (and, where appropriate, intensity) suits single-crystal frames dominated by discrete diffraction spots.

Table 5
Practical guide to determine when to use which center detection method

Dataset or category Primary method Why Fallback Notes
Polycrystalline (D1) phase Robust to intensity fluctuations and occlusions; uses symmetry and correlation hough or curvefit The best overall on D1, subpixel accuracy (see Figs. 3 and 4)
BB phase Tolerates missing data hough The implementation uses an explicit beamstop mask
BR curvefit Models discrete spots/arc segments phase hough if arcs are long/clean
RA phase Uses symmetry and correlation ccorr Edge detection prior to center determination is beneficial in this case
SNB phase Less reliant on absolute intensity curvefit Requires background normalization
FB phase Strongest/sharpest peaks Any An easy case, multiple methods are applicable
Single-crystal (D2) curvefit Works on local peak/spot structure intensity No automatic method beats the reference; manual is still the best
†Category tags are according to the classifications in Table 4.
4.4.2. Applicability of our software to X-ray, neutron and light diffraction

Our software focuses on electron diffractograms, but the underlying requirement – accurate location of the direct beam in 2D diffraction images – extends well beyond electron diffraction. As summarized in Table 6[link], similar demands are reported in other types of diffraction, where 2D detectors are occasionally used. Regardless of the diffraction type, the diffraction patterns are similar in the sense that they contain diffraction rings for powder-like samples and individual diffraction spots for monocrystal-like samples.

Table 6
Diffraction techniques employing 2D detectors

The final column gives brief comments concerning the 2D detectors employed by the given diffraction method.

Diffraction method Radiation type Typical 2D detectors Comments
Electron diffraction Electrons CCD/CMOS, hybrid-pixel Area detectors are standard in TEM/STEM setups (Heidler et al., 2019View full citation; Slouf et al., 2021aView full citation)
X-ray diffraction X-rays CCD/CMOS, hybrid-pixel Point detectors are perhaps more common, but area detectors for recording 2D diffractograms are used as well (Ashiotis et al., 2015View full citation)
Neutron diffraction Neutrons 2D area detectors (e.g. image plates, pixel detectors) Increasingly employs pixelated detectors, especially in time-of-flight setups (Jäger et al., 2024View full citation), with beam center finding required for data reduction
Light diffraction Visible or NIR light CCD/CMOS, photodiode arrays 2D detectors used in laser diffraction and holographic methods (Siavashani et al., 2021View full citation); 2D fringe or ring patterns require identification of the central spot

5. Conclusions

This work targets accurate fully automatic center location in 2D electron diffraction patterns. We have implemented six center detection and three center refinement algorithms in Python and released them in ediff.center, a module of the open-source EDIFF package (https://pypi.org/project/ediff). All methods were evaluated systematically on a representative suite of diffractograms – single crystal and polycrystalline, with and without a beamstop, and spanning a range of background and/or noise levels. The main conclusions are:

(i) The algorithms in ediff.center can locate the center of all common 2D electron diffractograms in a straightforward automated manner. From the user's perspective, it suffices to select and call a function appropriate to the pattern type.

(ii) Center detection is generally easier in polycrystalline diffractograms due to their higher radial symmetry (Debye–Scherrer rings). The most universal automatic method is phase (phase cross-correlation exploiting inversion symmetry), followed by hough (circular feature detection using the Hough transform).

(iii) Single-crystal diffractograms are more challenging. In modern 4D-STEM data, a strong direct beam spot often allows accurate location using the curvefit and intensity methods, which target the central peak. In traditional TEM/NBD experiments, diffractograms are usually aligned along some specific user-defined zone axis, which makes them symmetric. The center can be determined using the phase method as in symmetric polycrystalline patterns, even if the central spot is hidden by a beamstop.

Overall, our module ediff.center offers a general modality-agnostic solution for center determination in diffraction data; the same principles extend to X-ray, neutron and light diffraction.

6. Related literature

For further literature related to the supporting information, see Bücker et al. (2021View full citation), Filik et al. (2017View full citation), Hammersley (2016View full citation), Heybrock et al. (2025View full citation), Kieffer & Karkoulis (2013View full citation), Mitchell (2008View full citation) and Prescher & Prakapenka (2015View full citation).

Supporting information


Acknowledgements

Open access publishing facilitated by Ustav pristrojove techniky Akademie ved Ceske republiky, as part of the Wiley–CzechELib agreement.

Conflict of interest

There are no conflicts of interest.

Data availability

All diffraction patterns, calculations and results from this work are available from the first author ([email protected]) upon request. The EDIFF Python package that contains the ediff.center module with all algorithms employed in this study is freely available in the standard PyPI repository (Python Package Index): https://pypi.org/project/ediff.

Funding information

This research was funded by the Technology Agency of the Czech Republic (grant No. TN02000020). We also acknowledge the Core Facility Electron Microscopy and Raman Spectroscopy, supported by the Czech BioImaging Large RI project (grant No. LM2023050 funded by MEYS CR), for providing access to instrumentation and technical support. TM acknowledges the CzechNanoLab Research Infrastructure supported by MEYS CR (grant No. LM2023051) and Brno University of Technology (Central European Institute of Technology project No. VUT-J-24-8667).

References

Return to citationAbbondanza, G., Larsson, A., Carlá, F., Lundgren, E. & Harlow, G. S. (2021). J. Appl. Cryst. 54, 1140–1152.  Web of Science CrossRef CAS IUCr Journals Google Scholar
Return to citationAdatrao, S. & Mittal, M. (2016). 2nd International Conference on Frontiers of Signal Processing (ICFSP), Warsaw, Poland, pp. 110–115. IEEE.  Google Scholar
Return to citationAshiotis, G., Deschildre, A., Nawaz, Z., Wright, J. P., Karkoulis, D., Picca, F. E. & Kieffer, J. (2015). J. Appl. Cryst. 48, 510–519.  Web of Science CrossRef CAS IUCr Journals Google Scholar
Return to citationBrewster, A. S., Waterman, D. G., Parkhurst, J. M., Gildea, R. J., Young, I. D., O'Riordan, L. J., Yano, J., Winter, G., Evans, G. & Sauter, N. K. (2018). Acta Cryst. D74, 877–894.  Web of Science CrossRef IUCr Journals Google Scholar
Return to citationBücker, R., Hogan-Lamarre, P. & Miller, R. J. D. (2021). Front. Mol. Biosci. 8, 624264.  Web of Science PubMed Google Scholar
Return to citationCautaerts, N., Crout, P., Ånes, H. W., Prestat, E., Jeong, J., Dehm, G. & Liebscher, C. H. (2022). Ultramicroscopy 237, 113517.  Web of Science CrossRef PubMed Google Scholar
Return to citationClausen, A., Weber, D., Ruzaeva, K., Migunov, V., Baburajan, A., Bahuleyan, A., Caron, J., Chandra, R., Halder, S., Nord, M., Müller-Caspary, K. & Dunin-Borkowski, R. (2020). JOSS 5, 2006.  Google Scholar
Return to citationDu, W. & Yang, J. (2009). Phys. Med. Biol. 54, 555–567.  Web of Science CrossRef PubMed Google Scholar
Return to citationFilik, J., Ashton, A. W., Chang, P. C. Y., Chater, P. A., Day, S. J., Drakopoulos, M., Gerring, M. W., Hart, M. L., Magdysyuk, O. V., Michalik, S., Smith, A., Tang, C. C., Terrill, N. J., Wharmby, M. T. & Wilhelm, H. (2017). J. Appl. Cryst. 50, 959–966.  Web of Science CrossRef CAS IUCr Journals Google Scholar
Return to citationFultz, B. & Howe, J. M. (2002). Transmission Electron Microscopy and Diffractometry of Materials. Berlin, Heidelberg: Springer.  Google Scholar
Return to citationGard, J. A. (1976). Electron Microscopy in Mineralogy, edited by H.-R. Wenk, pp. 52–67. Berlin, Heidelberg: Springer.  Google Scholar
Return to citationHammersley, A. P. (2016). J. Appl. Cryst. 49, 646–652.  Web of Science CrossRef CAS IUCr Journals Google Scholar
Return to citationHeidler, J., Pantelic, R., Wennmacher, J. T. C., Zaubitzer, C., Fecteau-Lefebvre, A., Goldie, K. N., Müller, E., Holstein, J. J., van Genderen, E., De Carlo, S. & Gruene, T. (2019). Acta Cryst. D75, 458–466.  Web of Science CrossRef IUCr Journals Google Scholar
Return to citationHeybrock, S., Wynen, J.-L., Vaytet, N., Arnold, O., Gudich, I., Nixon, D., Willemsen, T., Yoo, S., jokasimr, Seth, M., Rozyczko, P., Jones, S., DavidFair, Tucker, G. S., Thibault, mads-bertelsen, McCluskey, A., Jones, M. D., Tasev, D., Behar, D., YXQD, Lund, M., Draper, N. & Peterson, P. (2025). scipp/scipp, https://zenodo.org/records/19045118Google Scholar
Return to citationHou, V. D.-H. & Li, D. (2008). Microsc. Today 16, 36–41.  CrossRef Google Scholar
Return to citationJäger, T. T., Losko, A. S., Wolfertz, A., Schmidt, S., Bertelsen, M., Khaplanov, A., Agnew, S. R., Funama, F., Morgano, M., Roth, M., Gochanour, J. R., Long, A. M., Lutterotti, L. & Vogel, S. C. (2024). J. Appl. Cryst. 57, 1107–1114.  Web of Science CrossRef IUCr Journals Google Scholar
Return to citationKieffer, J. & Karkoulis, D. (2013). J. Phys. Conf. Ser. 425, 202012.  CrossRef Google Scholar
Return to citationKlinger, M. (2017). J. Appl. Cryst. 50, 1226–1234.  Web of Science CrossRef CAS IUCr Journals Google Scholar
Return to citationKlinger, M. & Jäger, A. (2015). J. Appl. Cryst. 48, 2012–2018.  Web of Science CrossRef CAS IUCr Journals Google Scholar
Return to citationLábár, J. L. (2000). Proceedings of EUREM 12, pp. I379–I380. Czechoslovak Society for Electron Microscopy.  Google Scholar
Return to citationLábár, J. L. (2005). Ultramicroscopy 103, 237–249.  Web of Science PubMed Google Scholar
Return to citationLábár, J. L. & Das, P. P. (2017). Microsc. Microanal. 23, 647–660.  Web of Science PubMed Google Scholar
Return to citationLiu, J., Chuang, M., Chisholm, A. & Cosman, P. (2015). 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Milan, Italy, pp. 1975–1980. IEEE.  Google Scholar
Return to citationLynn, N. D., Sourav, A. I. & Santoso, A. J. (2021). IOP Conf. Ser. Mater. Sci. Eng. 1096, 012079.  Google Scholar
Return to citationMitchell, D. R. G. (2008). Ultramicroscopy 108, 367–374.  Web of Science CrossRef PubMed CAS Google Scholar
Return to citationOphus, C. (2019). Microsc. Microanal. 25, 563–582.  Web of Science CrossRef CAS PubMed Google Scholar
Return to citationPekin, T. C., Gammer, C., Ciston, J., Minor, A. M. & Ophus, C. (2017). Ultramicroscopy 176, 170–176.  Web of Science CrossRef CAS PubMed Google Scholar
Return to citationPrescher, C. & Prakapenka, V. B. (2015). High Pressure Res. 35, 223–230.  Web of Science CrossRef CAS Google Scholar
Return to citationRené de Cotret, L. P., Otto, M. R., Stern, M. J. & Siwick, B. J. (2018). Adv. Struct. Chem. Imag 4, 11.  Google Scholar
Return to citationSavitzky, B. H., Hughes, L., Bustillo, K. C., Deng, H. D., Jin, N. L., Lomeli, E. G., Chueh, W. C., Herring, P., Minor, A. & Ophus, C. (2019). Microsc. Microanal. 25, 124–125.  CrossRef Google Scholar
Return to citationSavitzky, B. H., Zeltmann, S. E., Hughes, L. A., Brown, H. G., Zhao, S., Pelz, P. M., Pekin, T. C., Barnard, E. S., Donohue, J., Rangel DaCosta, L., Kennedy, E., Xie, Y., Janish, M. T., Schneider, M. M., Herring, P., Gopal, C., Anapolsky, A., Dhall, R., Bustillo, K. C., Ercius, P., Scott, M. C., Ciston, J., Minor, A. M. & Ophus, C. (2021). Microsc. Microanal. 27, 712–743.  Web of Science CrossRef CAS PubMed Google Scholar
Return to citationSiavashani, M. J., Naghedi, I., Abbasian, V., Akhlaghi, E. A., Charsooghi, M. A., Tavassoly, M. T. & Moradi, A.-R. (2021). Sci. Rep. 11, 482.  Web of Science CrossRef PubMed Google Scholar
Return to citationSlouf, M., Sikorova, P., Pavlova, E., Swietek, M., Lartigue, L., Skoupy, R. & Krzyzanek, V. (2025). Microsc. Microanal. 31, ozaf045.  Web of Science CrossRef PubMed Google Scholar
Return to citationSlouf, M., Skoupy, R., Pavlova, E. & Krzyzanek, V. (2021a). Materials 14, 7550.  Web of Science CrossRef PubMed Google Scholar
Return to citationSlouf, M., Skoupy, R., Pavlova, E. & Krzyzanek, V. (2021b). Nanomaterials 11, 962.  Web of Science CrossRef PubMed Google Scholar
Return to citationThronsen, E., Bergh, T., Thorsen, T. I., Christiansen, E. F., Frafjord, J., Crout, P., van Helvoort, A. T. J., Midgley, P. A. & Holmestad, R. (2024). Ultramicroscopy 255, 113861.  Web of Science CrossRef PubMed Google Scholar
Return to citationWang, S., Eldred, T. B., Smith, J. G. & Gao, W. (2022). Ultramicroscopy 236, 113513.  Web of Science CrossRef PubMed Google Scholar
Return to citationXu, G.-S. (2009). Second International Conference on Information and Computing Science, Manchester, UK, pp. 373–375. IEEE.  Google Scholar
Return to citationZhang, H., Wang, W., Yuan, L., Wei, Z., Zhang, H. & Zhang, W. (2022). Mater. Charact. 187, 111854.  Web of Science CrossRef Google Scholar
Return to citationZuo, J. M. & Spence, J. C. H. (2017). Advanced Transmission Electron Microscopy. New York: Springer.  Google Scholar

This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.

Journal logoJOURNAL OF
APPLIED
CRYSTALLOGRAPHY
ISSN: 1600-5767
Follow J. Appl. Cryst.
Sign up for e-alerts
Follow J. Appl. Cryst. on Twitter
Follow us on facebook
Sign up for RSS feeds