research papers
accessCalculation of effective pump dose in X-ray-pump/X-ray-probe experiments
aCenter for Free-Electron Laser Science CFEL, Deutsches Elektronen-Synchrotron DESY, Notkestr. 85, 22607 Hamburg, Germany, bInstitute of Nuclear Physics, Polish Academy of Sciences, Radzikowskiego 152, 31-342 Kraków, Poland, cEuropean XFEL, Holzkoppel 4, 22869 Schenefeld, Germany, dInstitute of Physics, Czech Academy of Sciences, Na Slovance 2, 182 21 Prague 8, Czech Republic, eRIKEN SPring-8 Center, 1-1-1 Kouto, Sayo, Hyogo 679-5148, Japan, fUniversity of Hamburg, Institute for Experimental Physics/CFEL, Luruper Chaussee 149, 22761 Hamburg, Germany, gLinac Coherent Light Source, SLAC National Accelerator Laboratory, 2575 Sand Hill Road, Menlo Park, CA 94025, USA, and hInstitute of Spintronics and Quantum Information, Faculty of Physics and Astronomy, Adam Mickiewicz University in Poznań, Uniwersytetu Poznańskiego 2, PL-61614 Poznań, Poland
*Correspondence e-mail: [email protected], [email protected]
In pump–probe experiments on solid materials performed within ultrafast X-ray science, the energy deposited by an X-ray pump pulse in the sample has a non-uniform spatial distribution. The following X-ray probe pulse then measures a volume-integrated average of contributions from the differently irradiated regions of the sample. Here we propose a scheme to calculate an effective fluence of the pump pulse such that the observable of interest calculated with the effective fluence is very close to the volume-integrated observable. This approach simplifies computational simulations of X-ray irradiated solids, which typically use periodic boundary conditions and assume a uniformly irradiated simulation box. Obtaining a prediction on a volume-integrated observable requires a significant computational effort, as it is necessary to run multiple simulations for the different exposure conditions and then perform their volume integration. The proposed scheme reduces this effort to a single calculation with the effective fluence.
Keywords: femtosecond studies; free-electron laser; dynamical studies; pump–probe; XFEL.
1. Introduction
X-ray-pump/X-ray-probe experiments enable time-resolved studies of electronic and structural changes in materials following X-ray irradiation (Inoue et al., 2021
; Inoue et al., 2022
; Inoue et al., 2023
; Opara et al., 2018
; Pardini et al., 2014
; Inoue et al., 2016
; Ferguson et al., 2016
; Nass et al., 2020
; Hartley et al., 2021
; Inoue et al., 2024
). One of the bottlenecks in the interpretation of such measurements is the spatial non-uniformity of the pulses, usually assumed to have a Gaussian shape. This results in different regions of the investigated material being exposed to different pulse fluences. The following X-ray probe pulse then measures a volume-integrated average of the contributions originating from the differently irradiated regions of the sample (Tkachenko et al., 2021
; Heimann et al., 2023
).
Computational simulations of the X-ray irradiated samples are frequently performed in order to interpret experimental results. Such schemes typically use periodic boundary conditions. This implies that the simulation box – representing a small fraction of the irradiated material volume – is assumed to be uniformly irradiated (Medvedev et al., 2013
). A correct interpretation of the measurement results would require obtaining a prediction on a volume-integrated observable. This usually entails a significant computational effort, as it is necessary to run multiple simulations for the different exposure conditions and then perform their volume integration.
Here we propose an analytical scheme to calculate an effective fluence such that the observable of interest calculated for the effective fluence is very close to the volume-integrated observable. This approach requires pre-knowledge of how the observable in question depends on X-ray fluence. We demonstrate the effectiveness of the scheme on a study case: simulations performed with the code XTANT (Medvedev et al., 2013
) for a silicon face-centred-cubic crystal pumped with 50 eV photons. We obtain volume-integrated predictions for average atomic displacement and intensities of various Bragg diffraction reflections at different times. After applying a dedicated analytical scheme, a comparison of volume-integrated results with the effective fluence results is made, with very satisfactory results.
Application prospects and limitations of the scheme are discussed in the last section. The latter are illustrated with an example of a concrete experimental case (diamond pumped and probed with hard X-rays) reported by Heimann et al. (2023
).
2. Definition and calculation of effective fluence
In a pump–probe experiment, a sample is irradiated first by a pump pulse, which initiates the desired electronic or structural transition in the material. The probe pulse arrives at different time instants, testing the actual state of the pump-irradiated sample. We will assume henceforth that the probing is non-invasive for the irradiated crystal (Chapman et al., 2014
; Yoon et al., 2014
; Inoue et al., 2022
). For simplicity, we assume that the pump and probe pulses have a similar Gaussian profile, with the same spatial spread but different peak energies. The respective fluences can be written as follows,
where E0 is the pump pulse energy energy and sE0 is the probe pulse energy, expressed as a fraction of the pump energy. The spatial spread is defined by σ, for both the pump and the probe pulses.
Let us now consider the observable of interest, O, at a given time, t. We assume that it depends on the pump fluence Fpp as
This way, we performed a standard polynomial fit to estimate the observable dependence on Fpp. The polynomial order M, at which the series is cut, is determined by the fitting procedure described in the Methods
section.
For argument's sake, we will now fix the time to be the final time recorded in the experiment, tfinal, i.e. O(Fpp) ≡ O(Fpp, tfinal) and am ≡ am(tfinal). The observable is probed by a radially symmetric Gaussian beam Fpb(r). The resulting volume-integrated observable 〈O〉vol then reads
In our analysis, we assumed a uniform in-depth distribution of the absorbed beam energy in the material which, in fact, reduces the volume integration to the area integration. This uniformity assumption is only correct if the irradiated sample is thin or its thickness is comparable with the X-ray photon attenuation length (or penetration depth). The extension of our formalism, taking into account the in-depth absorption of energy (according to the Beer–Lambert law) is possible; however, it is beyond the scope of this introductory work. Performing the integration yields
In order to determine an effective fluence value, Feff, we then equate
Inserting equation (4)
into equation (5)
yields
where Fpeak = E0/(2πσ2). This equation can be solved analytically or numerically depending on the specific fluence dependence. From the multiple roots of the equation we will choose the one that is real and lies within the dose interval, [0, Fpeak]. For the observables we studied here, this criterion proved sufficient to identify a unique value of Feff.
We will now present how the scheme works by performing dedicated simulations of soft X-ray irradiated crystals with the XTANT code (Medvedev et al., 2018
). Beforehand, we will make a useful conversion from fluence to absorbed dose. While in experiments fluence is a natural parameter describing the integrated strength of the irradiation, theoretical simulations in general, and XTANT specifically, use an average absorbed dose per atom, D, expressed in eV atom−1. As discussed previously (Medvedev et al., 2013
; Medvedev et al., 2018
; Tkachenko et al., 2016a
), the absorbed dose is a more universal parameter in ultrafast X-ray science, as X-ray induced transitions directly depend on the absorbed dose per atom. If the sample thickness is comparable with the attenuation length of the X-ray radiation (what we for argument's sake assume here), the angle of X-ray incidence is normal, electron escape and surface reflectivity are negligible (cf. Follath et al., 2019
), the conversion can be readily made as
where e is Euler's number, F is the fluence in eV cm−2, na is the number density of the irradiated sample in atoms cm−3 and λatt (in cm) is the photon attenuation length estimated for the photon energy of the pump. We can now convert fluence to dose in equation (6)
,
where Deff refers to the effective dose, and Dpeak refers to the peak dose, converted from Fpeak, using equation (7)
.
3. Calculation of effective fluence for X-ray irradiated silicon crystal
Below we will present the study case for the effective fluence method. Namely, we will simulate irradiation of a silicon crystal with a pump pulse of 6 fs FWHM duration and 50 eV photon energy with our code XTANT (Medvedev et al., 2013
; Medvedev et al., 2015
; Medvedev et al., 2017
; Medvedev et al., 2018
). Similar simulations were performed to describe experimental data by Tkachenko et al. (2016b
) and Medvedev et al. (2018
), therefore, we can safely assume that the predictions are realistic enough. We will analyze structural transition in silicon initiated by a pulse of Gaussian spatial profile with σ = 63.7 nm. The peak dose deposited by the pulse is 8 eV atom−1 (see Fig. 1
). The number of atoms in the simulation box (with periodic boundary conditions assumed) is 64. This number of atoms was chosen for expediency of the numerical simulations and should not affect conclusions regarding the correctness of the method in this purely numerical study. Since both the volume-integration and the effective-dose simulations are run for the same box size, the eventual finite size effects would act as a systematic error, affecting both simulations equally. However, if the method should be applied to describe realistic phase transitions, a convergence study with respect to the size of the simulation box should be performed. The simulation was run at times ranging from −200 fs to 100 fs, where time zero corresponds to the temporal maximum of the X-ray pump pulse.
|
Figure 1
Radial distribution of the dose deposited by pump pulse (red line). Points used to run the XTANT simulation are marked in blue. |
The choice of this timescale was arbitrary as the goal of this study was to demonstrate that the effective-fluence method works, and not to study the details of phase transition in Si. The timescale chosen does not impact the conclusion that the method is sound, since both the effective-dose and volume-integrated simulations were performed for the same run time. Furthermore, as an intense pump pulse was used, 100 fs is sufficient to observe dislocations of the atoms from their equilibrium positions on the order of 10% of the interatomic distance in silicon. Concerning the divergence between the effective-dose and volume-integrated calculations at later times (close to 100 fs), the reason for it is the decreasing quality of the polynomial fit in the regime of quickly changing observables. This could be improved by extending the simulation time and fitting the coefficients once at the new final time point of the simulation (not considered here).
3.1. Analysis of atomic displacements
At first, we have analyzed atomic displacements. In the periodic boundary condition framework applied here, the displacement of a single atom in a simulation box at time t can be defined as
with x1, x2, x3 being the three Cartesian coordinates of an atom in the simulation box at time t and Lbox being the cubic box size. The time instant tinit corresponds to the initial time at which the simulation was started, i.e. when silicon was in the ambient state (T = 300 K) with all of its atoms in their crystal lattice points. From the output of the XTANT code (for a fixed dose), we obtain the displacement as a function of time for each atom in the simulation box. Summing up the displacements of all atoms in the simulation box and dividing the sum by the number of atoms, we obtain the average atomic displacement as a function of time. Fig. 2
shows an example of average displacement obtained for a fixed dose of 3.7 eV atom−1.
|
Figure 2
Example of average atomic displacement 〈Δd(t)〉 for an XTANT simulation performed for silicon with 64 atoms in the simulation box. The other parameters were: D = 3.7 eV atom−1, pulse duration of 6 fs FWHM, and photon energy of 50 eV. The atomic displacement shows thermal oscillation before irradiation. After irradiation it starts to increase quickly. |
Our in-house code VOLINT [similar to XSINC (Abdullah et al., 2016
; Abdullah et al., 2018
)] was used to predict the volume-integrated displacement, calculated for the simulation results obtained with the code XTANT. The simulations were performed for different dose values spanning between 0 and the peak dose of 8 eV atom−1. In the present case (σ = 63.7 nm), we resolved the observable spatially up to a cut-off value of
Rmax = 200 nm. We divided the range [0, 200] nm into 11 rings and considered the central dose in each ring as the input for the XTANT run. This is illustrated in Fig. 1
. VOLINT then integrates the different spatial zone results for each time step and obtains 〈Δd(t)〉vol, according to equation (3)
.
Since XTANT includes stochastic molecular dynamics simulations, it is necessary to average the simulation results over a few realizations. Previous calculations with XTANT (Tkachenko et al., 2021
; Medvedev et al., 2013
) showed that several realizations are enough, as averaging over electron kinetics is intrinsically performed in the code (Medvedev et al., 2013
; Medvedev et al., 2018
). Here, we performed ten realizations and averaged over them. The final output of the volume integration is presented in Fig. 3
. The XTANT simulation starts with atoms placed in their equilibrium positions in the crystal (at zero temperature), with random velocities chosen such that the atomic temperature after thermalization should be equal to 300 K. This evolution stage can be clearly identified in Fig. 3
, where, after initialization, the average atomic displacement increases up to 0.1 Å (reflecting the movement of atoms from their equilibrium positions at 0 K). Afterwards, the displacement reduces to a standard thermal atomic displacement at 300 K. After X-ray pumping, the heated crystal continuously loses its structure due to the progressing non-thermal melting process. This is reflected by an increase of atomic displacement (and the corresponding decrease of average Bragg reflection intensities).
|
Figure 3
Volume-integrated and realization averaged atomic displacement 〈Δd(t)〉vol,real. The integration was performed using the XTANT simulations performed for the doses marked in blue in Fig. 1 |
For the calculation of the effective dose, Deff, from equation (8)
, it is necessary to obtain the expansion coefficients am. As stated previously, we perform the estimation only for a fixed time, t = 100 fs, which is the final time for these simulations, in order to gain an idea about the time-dependence of the coefficients am. A respective fitting procedure is required, with details provided in the Methods section
. The optimal order for the fitting polynomial was estimated to be 5. With the polynomial coefficients calculated, we numerically solve equation (7)
. The solution found within our dose domain, [0, 8] eV atom−1, is Deff = 4.20 eV atom−1.
In Fig. 4
we present the average atomic displacement within the simulation box, averaged over ten different XTANT realizations performed for the effective dose of Deff = 4.20 V atom−1. It is compared with the respective volume-integrated atomic displacement from Fig. 3
. The agreement between the two curves is good, which proves the correctness of the effective dose scheme. It is to be emphasized that, although the estimation of Deff was performed at t = tfinal, there is good agreement between the average atomic displacement calculated for the effective dose and the volume-integrated one for all times. This indicates that the time-dependence of the coefficients am(t) is practically negligible, i.e. the dependence of our observable on dose (fluence) is the same for all time steps.
|
Figure 4
Comparison between the volume-integrated and realization averaged atomic displacement 〈Δd(t)〉vol,real (blue line) and the atomic displacement calculated for the effective dose and averaged over ten XTANT realizations (red line). The volume integration was performed for a silicon sample with 64 atoms in the simulation box, using the XTANT results for the doses marked in blue in Fig. 1 |
3.2. Analysis of Bragg reflection intensities
We will now apply the effective dose (fluence) scheme for the intensities of various Bragg diffraction reflections. Bragg reflection intensities are directly observed in experiments and can be used, in particular, to estimate atomic displacements. The Bragg diffraction condition is: Δk = q, where Δk is the difference between the incoming and outgoing momentum and q is a reciprocal lattice vector described by Miller indices (Kittel, 2004
). When this condition is fulfilled, the resulting structure factor associated with a given reciprocal lattice vector can be expressed as (Kittel, 2004
)
where fj is the so-called atomic form factor, and is the position of the jth atom in the unit cell of the crystal. The scattered (Bragg) intensity (Iq) is then proportional to |Fq|2.
VOLINT was used to predict the scattering intensity for a given Bragg reflection. VOLINT proceeds in the following steps: (1) performs several XTANT simulations; (2) from each XTANT simulation it obtains the actual position of each atom in the simulation box at each time step as well as the average electronic population for each atomic subshell; (3) it then distributes the number of holes randomly over the atoms in the box to obtain the average electronic population in each subshell as determined by XTANT; (4) afterwards, it uses the XATOM code (Inhester et al., 2023
; Jurek et al., 2016
; Son et al., 2011
) to calculate atomic form factors for all atomic species and electronic configurations; (5) it calculates Fq(t) with equation (10)
; (6) finally, it performs realization averaging and then volume integration of Iq(t) to obtain the respective (relative) Bragg signal, according to equation (3)
. An example of volume-integrated Bragg reflection intensity obtained with VOLINT is presented in Fig. 5
.
|
Figure 5
Volume-integrated 111 Bragg reflection intensity |
The same procedure as above was applied to calculate the respective Deff for the intensity of the 111 reflection (for details see the Methods section
). A comparison between the 111 reflection obtained for the effective dose and the volume-integrated 111 reflection is presented in Fig. 6
. Again, we can see that the two curves resemble one another closely, validating our approach. An increase of the realization number to 50 realizations was carried out in order to improve the statistics. Both curves reflect thermal oscillations within silicon crystal before time zero and rapid atomic dislocation after time zero. The latter results in the Bragg intensity dropping.
|
Figure 6
Comparison between the volume-integrated Bragg reflection intensity 〈I111(t)〉vol,real (blue line), and the Bragg reflection intensity obtained for the effective dose 〈I111(t)〉rel averaged over 50 XTANT realizations (red line). The volume integration was performed for a silicon sample with 64 atoms in the simulation box, using the XTANT results for the doses marked in blue in Fig. 1 |
Again, we note that, although the fitting of Deff was performed at t = tfinal, there is good agreement between the effective dose and the volume-integrated curves at all times. This indicates that the time-dependence of the polynomial coefficients is practically negligible, i.e. the dependence of our observable on dose (fluence) is the same for all time steps.
The same search procedure for an effective dose was repeated for other diffraction reflections, namely for the 311, 220, 400 and 331 reflections (see Fig. 7
). The procedure needs to be repeated since different Bragg reflections have different sensitivity to atomic displacements. They will all decay as the atoms move away from their equilibrium positions, however, not in the same way, because they probe different directions of local order. As the crystal melts, it is expected that some bonds will break faster than others, creating anisotropies which make different Bragg reflections respond differently. Table 1
summarizes the predicted effective doses, both for all the Bragg reflections analyzed as well as for the average atomic displacement.
|
||||||||||||||||||
|
Figure 7
Comparison between volume-integrated Bragg reflection intensities (blue) and the Bragg reflection intensity for the appropriate effective dose, averaged over ten XTANT realizations (red line) for (a) 〈I220〉, (b) 〈I311〉, (c) 〈I400〉 and (d) 〈I331〉. Volume integration was performed for a silicon sample with 64 atoms in the simulation box, using the XTANT results for the doses marked in blue in Fig. 1 |
All effective doses are found within a narrow range of 4 to 4.5 eV atom−1, with a maximum variation of approximately 12%. This consistency indicates that the range of effective doses accurately captures the dynamics occurring in the sample during the structural transition. Conversely, if we were to calculate the dose from the so called `average' fluence obtained by dividing the total beam energy over the focal area (up to the FWHM), we would obtain a value of 11.5 eV atom−1 for the same pump profile (see Fig. 1
). Following the above-described procedure, since the full beam energy is assumed to be deposited in the FWHM focus, the relation 〈F〉 = 1.442Fpeak will hold for any Gaussian pulse. If this average value were used in constant dose simulations, the results would differ significantly from the experimental observations.
4. Discussion
4.1. Non-Gaussian beams
The current analysis does not explore a general case of any (non-Gaussian) beam profile, as the scope of the paper is to introduce the idea of an effective fluence and to illustrate it on a typical example of a beam profile. The assumption that an X-ray beam has a spatial Gaussian profile is a frequent approximation applied during the analysis of various X-ray free-electron laser pump–probe experiments (e.g. Abdullah et al., 2016
). One-dimensional analysis can then be sufficient to obtain effective fluence. However, the effective-fluence methodology itself is not limited to the Gaussian case. If we can perform the fluence scan as proposed by, for example, Chalupský et al. (2013
), the effective fluence can still be evaluated numerically by solving equation (5)
. This analysis will then be much more complex (multi-dimensional). However, strong shot-to-shot variations of the beam profile cannot be treated accurately. Predictions for the shot-averaged (Gaussian) pulse profile can then only be obtained.
4.2. Scaling properties
An interesting property of equation (8)
is that it only depends on the peak dose, Fpeak, and the fitting coefficients, am. If both E0 and σ would be rescaled to keep Fpeak unchanged (and, as a consequence, Dpeak as well), the effective dose value Deff should remain the same.
Following this line of reasoning, we performed a simple check of the rescaling property. Namely, we reduced the total energy, E0, by half and changed σ to . The peak fluence is then the same as before. In Fig. 8
the new radial profile of the pump pulse is shown, along with the discrete points used to perform volume integration.
|
Figure 8
Radial distribution of the `rescaled' pump pulse (red line). The points used for XTANT simulations are marked with blue dots. |
In Fig. 9
a comparison between the volume-integrated average atomic displacement calculated at the new irradiation conditions and the results for Deff = 4.20 eV atom−1 averaged over ten XTANT realizations are shown. The agreement between both predictions is quite good, of a quality similar to that obtained in Fig. 4
, i.e. for the same Dpeak but at different irradiation conditions. We also analyzed the time-dependent intensity of the 111 Bragg reflection presented in Fig. 10
with the same conclusion. These findings confirm the scaling feature of equation (8)
.
|
Figure 9
Comparison between the volume-integrated atomic displacement 〈Δd(t)〉vol,real (blue line) and the atomic displacement calculated for the effective fluence averaged over 50 XTANT realizations (red line). The volume integration was performed for a silicon sample with 64 atoms in the simulation box, using the XTANT results for the doses marked in blue in Fig. 8 |
|
Figure 10
Comparison between the volume-integrated Bragg reflection intensity 〈I111(t)〉vol,real (blue line) and the Bragg reflection intensity calculated for an effective dose averaged over 50 XTANT realizations (red line). The volume integration was performed for a silicon sample with 64 atoms in the simulation box, using the XTANT results for the doses marked in blue in Fig. 8 |
4.3. Effect of ballistic electrons
There are some limitations of the method; for example, it is assumed that the energy deposited by the X-ray pulse stays within the beam focus. This is true if the range of ballistic electrons released by the X-ray pump pulse is much smaller than the beam focus. Unfortunately, this is not the case when tightly focused hard X-rays are used for pumping, as the example presented below shows. Therein, experimental data are used to demonstrate the failure of the effective fluence method in this particular case. Still, a possible extension of the method by taking ballistic transport into account when estimating the effective fluence can be considered.
Heimann et al. (2023
) reported results from a hard X-ray pump/hard X-ray probe experiment. The target was solid-density diamond that underwent partial graphitization on an ultrafast time scale. To track the transition, time-resolved intensities of several Bragg reflections were measured. The experiment was performed under different irradiation conditions. To illustrate the limitations of our method we will use one specific irradiation case for which the parameters are summarized in Table 2
. Therein, E0 is the total pump pulse energy, wx/y the 1/e width of the pulse, and λe the range of electrons released by 7 keV pump pulse photons. Note that in the paper by Heimann et al. (2023
) a mistake was made and the range of electrons was quoted for an electronic cascade triggered by 8 keV and not by 7 keV photons.
|
||||||||||||||
To calculate the average absorbed dose, the following formula was proposed, correcting for fast electronic escape from the focal region,
where ρA is the atomic density, and λx is the X-ray penetration depth for the pump's specific photon energy. In our calculations, we assume that the pump profile is a radially symmetric Gaussian. Therefore, we apply here the approximation
, and with it calculate Dpeak ∝ E0/(2πσ2) for the radial Gaussian profile.
Knowing Dpeak and σ we can perform a set of simulations needed for volume integration, similarly to what we did in the previous sections. The results of these efforts are shown in Fig. 11
for the tightly focused case and the diffraction reflections 111, 220 and 311. The results with the peak dose (calculated with the method described above) are in green. As can be seen, they underestimate the Bragg reflection decay in all cases. As an improvement, the peak dose was recalculated without correcting for electronic escape (λe = 0; blue line). In this case, the Bragg reflection decay was overestimated. Finally, a peak dose was found such that the experimental and theoretical predictions would fall in the same range (orange line).
|
Figure 11
Volume-integrated Bragg reflection intensity for the irradiation conditions from Table 2 |
The volume integration fails to properly describe the observable and an ad hoc fitting procedure is required (still volume-integrated but with a peak dose not directly related to the experimental parameters). This discrepancy is probably related to the electron cascade size being comparable with the focal spot size, which also introduces a time dependence into the problem. As energy is deposited according to the pump profile, it also starts to spread out due to high-energy `ballistic' electrons. Both effects happen simultaneously during the pump pulse. The electronic transport continues for a few femtoseconds afterwards.
This hypothesis is supported by the fact that the peak dose calculated with λe (representing an immediate escape of the hot electrons) gives us the upper limit for the Bragg reflection decay while the peak dose calculated without λe (the other extreme, representing the case when electrons never escape from the focus) gives us the lower limit. Since none of the limiting cases, based on actual experimental parameters, provide a good fit to the data, no effective dose calculation was attempted.
4.4. Summary
Our results indicate that the effective dose method can be used for simulations of pump–probe experiments, replacing the costly volume-integration scheme. Here, we tested this method on atomic displacements and Bragg reflection intensities. However, the method can also be applied to other volume-integrated observables, provided their fluence (dose) dependence can be approximated with a polynomial fit. We believe that this study offers a computationally efficient and rigorous method to estimate the effective fluence (dose) for volume-integrated experimental observables. Importantly, we find that the effective dose better represents the sample dynamics than the average dose – total beam energy divided by the focal area (up to the FWHM) – commonly used in experiments. Using the average dose in simulations can lead to significant discrepancies from experimental results, underscoring the relevance of the effective dose approach introduced here. Nonetheless, a full exploration of the prospective applications of this method goes beyond the scope of this introductory work.
5. Methods
5.1. Atomic displacement as a function of dose
As explained before, in what follows we fix time to tfinal = 100 fs. The fitting procedure is applied to the pairs , where N is the number of different XTANT realizations being averaged over (here, N = 10). In what follows we will study the average atomic displacement 〈Δd(tfinal)〉real.
Since we do not know a priori the order of the polynomial dependence of 〈Δd(tfinal)〉real, the fits are performed for polynomials of increasing order. To avoid overfitting, we also calculate , i.e. the reduced χ2, and
, i.e. the probability of obtaining a χ2 value at least as high as the observed.
Following Taylor (1997
), we define
where yi are data points used for the fit with the uncertainty =
, where μ is the average data value obtained from all XTANT realizations for a given dose. f(xi) is the value predicted by the fit, and ν is the number of degrees of freedom. For linear models, ν = N − (n + 1), with n being the degree of the fitted polynomial. If our data points are independent and normally distributed then
≃
. With the well known probability density function for k degrees of freedom, we can then write
where Γ is the gamma function and , as stated above, is the probability of obtaining a test statistic at least as high as the observed one. This implies that, the higher the probability, the better the fit, and vice-versa.
We performed six fits with increasing polynomial order. They are shown in Fig. 12
. We used the gradient descent method for minimizing equation (12)
. and
were also calculated for all of the fits, according to equations (12)
and (13)
, respectively, and plotted in Figs. 13
and 14
. They consistently indicate that the quality of the fit significantly increases with the increasing polynomial order until we reach fifth order. Afterwards, the fitting quality stabilizes. It does not change when we continue to increase the number of degrees of freedom. This is a signature of overfitting (Bishop, 2006
).
|
Figure 12
Polynomial fits of various orders applied to the data set |
|
Figure 13
|
|
Figure 14
|
Note that the fitting in Fig. 12
was performed for all fluence points, since volume integration was performed for comparison, and we had all these points available anyway. In order to test the robustness of the method in respect of the number of fluence points, we redid the fit in Fig. 12
, using only six out of the 11 original fluence points, i.e. omitting every second point. This is the minimal number of data points needed to perform a fifth-order polynomial fit. After solving equation (8)
for the new set of am coefficients, we obtained the effective dose of 4.18 eV atom−1 for the average atomic displacement, whereas the dose obtained from the full set of fluence points was 4.2 eV atom−1, i.e. the minimal set of six fluence points (needed to perform a fifth-order polynomial fit) was sufficient to estimate the effective dose with good accuracy. Also, in the cases discussed in this study, coefficient fitting at each time step was not necessary, as all coefficients turned out to be time-independent. However, this may not always be the case.
5.2. Bragg reflection intensity as a function of dose
The same method as above was used to study the dose dependence of the 111 Bragg reflection intensity. We performed polynomial fits of increasing order to the respective data set. In Fig. 15
the data set and their fits are shown. Note that the error bars for the Bragg signal are significantly larger than for the atomic displacement case (cf. Fig. 12
).
|
Figure 15
Polynomial fits of different orders performed for the data set [ |
and
for all the fits are depicted in Figs. 16
and 17
, respectively, as a function of polynomial order. They indicate that the optimal fit to represent the dose dependence of our observable is of sixth order.
|
Figure 16
|
|
Figure 17
Parameter |
As before, we use the estimated polynomial coefficients to solve equation (8)
. In the desired dose range of the [0, 8] eV atom−1 range, we obtain the solution Deff = 4.51 eV atom−1. The same analysis was performed for other Bragg reflection intensities discussed in this study.
Acknowledgements
The authors are grateful for fruitful discussions with Zoltan Jurek and Vladimir Lipp. Open access funding enabled and organized by Projekt DEAL.
Funding information
KJK thanks the Polish National Agency for Academic Exchange for funding in the frame of the National Component of the Mieczysław Bekker program (2020 edition) (BPN/BKK/2022/1/00011). The participation of VT in this project was supported in part by a grant of the Polish Ministry of Science and Higher Education – decision No. 2022/WK/13.
References
Abdullah, M. M., Jurek, Z., Son, S.-K. & Santra, R. (2016). Struct. Dyn. 3, 054101.
Web of Science
CrossRef
PubMed
Google Scholar
Abdullah, M. M., Son, S.-K., Jurek, Z. & Santra, R. (2018). IUCrJ 5, 699–705.
CrossRef
CAS
PubMed
IUCr Journals
Google Scholar
Bishop, C. M. (2006). Pattern Recognition and Machine Learning (Information Science and Statistics). Berlin, Heidelberg: Springer-Verlag.
Google Scholar
Chalupský, J., Burian, T., Hájková, V., Juha, L., Polcar, T., Gaudin, J., Nagasono, M., Sobierajski, R., Yabashi, M. & Krzywinski, J. (2013). Opt. Express 21, 26363–26375.
PubMed
Google Scholar
Chapman, H. N., Caleman, C. & Timneanu, N. (2014). Philos. Trans. R. Soc. B 369, 20130313.
CrossRef
Google Scholar
Ferguson, K. R., Bucher, M., Gorkhover, T., Boutet, S., Fukuzawa, H., Koglin, J. E., Kumagai, Y., Lutman, A., Marinelli, A., Messerschmidt, M., Nagaya, K., Turner, J., Ueda, K., Williams, G. J., Bucksbaum, P. H. & Bostedt, C. (2016). Sci. Adv. 2, e1500837.
Web of Science
CrossRef
PubMed
Google Scholar
Follath, R., Koyama, T., Lipp, V., Medvedev, N., Tono, K., Ohashi, H., Patthey, L., Yabashi, M. & Ziaja, B. (2019). Sci. Rep. 9, 2029.
Web of Science
CrossRef
PubMed
Google Scholar
Hartley, N. J., Grenzer, J., Huang, L., Inubushi, Y., Kamimura, N., Katagiri, K., Kodama, R., Kon, A., Lu, W., Makita, M., Matsuoka, T., Nakajima, S., Ozaki, N., Pikuz, T., Rode, A. V., Sagae, D., Schuster, A. K., Tono, K., Voigt, K., Vorberger, J., Yabuuchi, T., McBride, E. E. & Kraus, D. (2021). Phys. Rev. Lett. 126, 015703.
CrossRef
PubMed
Google Scholar
Heimann, P., Hartley, N. J., Inoue, I., Tkachenko, V., Antoine, A., Dorchies, F., Falcone, R., Gaudin, J., Höppner, H., Inubushi, Y., Kapcia, K. J., Lee, H. J., Lipp, V., Martinez, P., Medvedev, N., Tavella, F., Toleikis, S., Yabashi, M., Yabuuchi, T., Yamada, J. & Ziaja, B. (2023). Struct. Dyn. 10, 054502.
CrossRef
PubMed
Google Scholar
Inhester, L., Jurek, Z., Son, S.-K. & Santra, R. (2023). Xraypac – a software package for modeling x-ray-induced dynamics of matter, https://www.desy.de/~xraypac/index.html.
Google Scholar
Inoue, I., Deguchi, Y., Ziaja, B., Osaka, T., Abdullah, M. M., Jurek, Z., Medvedev, N., Tkachenko, V., Inubushi, Y., Kasai, H., Tamasaku, K., Hara, T., Nishibori, E. & Yabashi, M. (2021). Phys. Rev. Lett. 126, 117403.
CrossRef
PubMed
Google Scholar
Inoue, I., Inubushi, Y., Sato, T., Tono, K., Katayama, T., Kameshima, T., Ogawa, K., Togashi, T., Owada, S., Amemiya, Y., Tanaka, T., Hara, T. & Yabashi, M. (2016). Proc. Natl Acad. Sci. USA 113, 1492–1497.
CrossRef
CAS
PubMed
Google Scholar
Inoue, I., Tkachenko, V., Kapcia, K. J., Lipp, V., Ziaja, B., Inubushi, Y., Hara, T., Yabashi, M. & Nishibori, E. (2022). Phys. Rev. Lett. 128, 223203.
CrossRef
PubMed
Google Scholar
Inoue, I., Tkachenko, V., Kubota, Y., Dorchies, F., Hara, T., Höppner, H., Inubushi, Y., Kapcia, K. J., Lee, H. J., Lipp, V., Martinez, P., Nishibori, E., Osaka, T., Toleikis, S., Yamada, J., Yabashi, M., Ziaja, B. & Heimann, P. A. (2024). Phys. Rev. B 110, L100102.
CrossRef
Google Scholar
Inoue, I., Yamada, J., Kapcia, K. J., Stransky, M., Tkachenko, V., Jurek, Z., Inoue, T., Osaka, T., Inubushi, Y., Ito, A., Tanaka, Y., Matsuyama, S., Yamauchi, K., Yabashi, M. & Ziaja, B. (2023). Phys. Rev. Lett. 131, 163201.
Web of Science
CrossRef
PubMed
Google Scholar
Jurek, Z., Son, S.-K., Ziaja, B. & Santra, R. (2016). J. Appl. Cryst. 49, 1048–1056.
Web of Science
CrossRef
CAS
IUCr Journals
Google Scholar
Kittel, C. (2004). Introduction to Solid State Physics, 8th ed. Wiley.
Google Scholar
Medvedev, N., Jeschke, H. O. & Ziaja, B. (2013). New J. Phys. 15, 015016.
CrossRef
Google Scholar
Medvedev, N., Li, Z., Tkachenko, V. & Ziaja, B. (2017). Phys. Rev. B 95, 014309.
CrossRef
Google Scholar
Medvedev, N., Li, Z. & Ziaja, B. (2015). Phys. Rev. B 91, 054113.
CrossRef
Google Scholar
Medvedev, N., Tkachenko, V., Lipp, V., Li, Z. & Ziaja, B. (2018). 4open 1, 3.
CrossRef
Google Scholar
Nass, K., Gorel, A., Abdullah, M. M. V., Martin, A., Kloos, M., Marinelli, A., Aquila, A., Barends, T. R., Decker, F.-J., Bruce Doak, R., Foucar, L., Hartmann, E., Hilpert, M., Hunter, M. S., Jurek, Z., Koglin, J. E., Kozlov, A., Lutman, A. A., Kovacs, G. N., Roome, C. M., Shoeman, R. L., Santra, R., Quiney, H. M., Ziaja, B., Boutet, S. & Schlichting, I. (2020). Nat. Commun. 11, 1814.
CrossRef
PubMed
Google Scholar
Opara, N. L., Mohacsi, I., Makita, M., Castano-Diez, D., Diaz, A., Juranić, P., Marsh, M., Meents, A., Milne, C. J., Mozzanica, A., Padeste, C., Panneels, V., Sikorski, M., Song, S., Stahlberg, H., Vartiainen, I., Vera, L., Wang, M., Willmott, P. R. & David, C. (2018). Struct. Dyn. 5, 054303.
CrossRef
PubMed
Google Scholar
Pardini, T., Boutet, S., Bradley, J., Döppner, T., Fletcher, L. B., Gardner, D. F., Hill, R. M., Hunter, M. S., Krzywinski, J., Messerschmidt, M., Pak, A. E., Quirin, F., Sokolowski-Tinten, K., Williams, G. J. & Hau-Riege, S. P. (2014). Phys. Rev. Appl. 1, 044007.
CrossRef
Google Scholar
Son, S.-K., Young, L. & Santra, R. (2011). Phys. Rev. A 83, 033402.
CrossRef
Google Scholar
Taylor, J. (1997). An Introduction to Error Analysis: The Study of Uncertainties in Physical Measurements. University Science Books.
Google Scholar
Tkachenko, V., Abdullah, M. M., Jurek, Z., Medvedev, N., Lipp, V., Makita, M. & Ziaja, B. (2021). Appl. Sci. 11, 5157.
CrossRef
Google Scholar
Tkachenko, V., Medvedev, N., Li, Z., Piekarz, P. & Ziaja, B. (2016b). Phys. Rev. B 93, 144101.
CrossRef
Google Scholar
Tkachenko, V., Medvedev, N. & Ziaja, B. (2016a). Appl. Sci. 6, 238.
CrossRef
Google Scholar
Yoon, C. H., Barthelmess, M., Bean, R. J., Capotondi, F., Kirian, R. A., Kiskinova, M., Pedersoli, E., Raimondi, L., Stellato, F., Wang, F. & Chapman, H. N. (2014). Opt. Express 22, 8085–8093.
CrossRef
PubMed
Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.
access
journal menu



