A high-energy-resolution resonant inelastic X-ray scattering spectrometer at ID20 of the European Synchrotron Radiation Facility
aESRF, The European Synchrotron, 71 Avenue des Martyrs, F-38000 Grenoble, France, bDepartment of Physics, Faculty of Science, Beirut Arab University, Beirut 11072809, Lebanon, cALBA Synchrotron Light Facility, Carrer de la Llum 2-26, 08290 Cerdanyola del Valleés, Barcelona, Spain, dDepartment of Physics, University of Helsinki, PO Box 64, FI-00014 Helsinki, Finland, and eDipartimento di Fisica, Università di Trento, via Sommarive 14, 38123 Povo (TN), Italy
*Correspondence e-mail: marco.moretti@esrf.fr
An end-station for resonant inelastic X-ray scattering and (resonant) X-ray emission spectroscopy at beamline ID20 of ESRF – The European Synchrotron is presented. The spectrometer hosts five crystal analysers in Rowland geometry for large solid angle collection and is mounted on a rotatable arm for scattering in both the horizontal and vertical planes. The spectrometer is optimized for high-energy-resolution applications, including partial
or high-energy-resolution fluorescence detected X-ray absorption spectroscopy and the study of elementary electronic excitations in solids. In addition, it can be used for non-resonant inelastic X-ray scattering measurements of valence electron excitations.Keywords: resonant inelastic X-ray scattering; X-ray emission; ID20 beamline; RIXS spectrometer; ESRF.
1. Introduction
Beamline ID20 is the state-of-the-art et al., 2017). In this article, we describe a high-energy-resolution spectrometer that is mostly dedicated to resonant (RIXS) and (resonant) X-ray emission spectroscopy [(R)XES], but it can also be employed in non-resonant experiments of valence excitations requiring high energy resolution (Larson et al., 2007; Cudazzo et al., 2014). The spectrometer is named RIXS, but formally there is no distinction between RIXS and RXES, and sometimes the two terms are used interchangeably.
beamline of ESRF – The European Synchrotron (Grenoble, France). This beamline evolved from one of the two branches of the previous ID16 beamline as the UPBL6 project in the frame of the ESRF Upgrade – Phase I. ID20 is today fully dedicated to the study of the electronic structure of materials and provides two branches, optimized for high and high-energy-resolution applications, respectively. The beamline optics common to all its branches will be described elsewhere. The high-efficiency medium-energy-resolution spectrometer that has been conceived for non-resonant from core levels has been described elsewhere (HuotariRIXS is a photon-in–photon-out technique for the study of the electronic structure of materials as it probes transitions involving lattice, charge, spin and/or orbital ; Ament et al., 2011a). In the specific case of low-energy elementary excitations, recent advances in the instrumentation, both in the soft (Ghiringhelli et al., 2006, 2012) and hard X-ray regimes (Shvyd'ko et al., 2013; Ishii et al., 2013; Moretti Sala et al., 2013) and in the theoretical interpretation of RIXS spectra (Haverkort, 2010; Ament et al., 2011b,c; Igarashi & Nagao, 2012; Marra et al., 2012, 2013, 2016; van den Brink & Veenendaal, 2012; Jia et al., 2014, 2016; Moretti Sala et al., 2014a; Kim & van den Brink, 2015; Devereaux et al., 2016) boosted the technique enormously. The most prominent example is the study of magnetic excitations in Cu (Braicovich et al., 2010; Guarise et al., 2010; Le Tacon et al., 2011; Ghiringhelli & Braicovich, 2013; Dean et al., 2012, 2013a,b, 2014; Ishii et al., 2014; Guarise et al., 2014; Lee et al., 2014; Peng et al., 2015; Moser et al., 2015; Minola et al., 2015; Monney et al., 2016; Meyers et al., 2017) and Ir (Kim et al., 2012a,b; Moretti Sala et al., 2014b, 2015; Donnerer et al., 2016; Gretarsson et al., 2016; Clancy et al., 2016; Calder et al., 2016; Liu et al., 2016; Lu et al., 2017; Cao et al., 2017) oxides. As a matter of fact, RIXS can be considered nowadays as a mature technique for studying magnetism in strongly correlated electron systems in a way complementary to other spectroscopies, such as inelastic neutron scattering (INS), X-ray resonant magnetic scattering (XRMS), X-ray (XMCD) or optical Raman spectroscopy. Other applications of RIXS comprise the study of phonons (Yavaş et al., 2010; Lee et al., 2013; Moser et al., 2015), crystal and excitations (Ghiringhelli et al., 2004; Moretti Sala et al., 2011), including orbitons (Ulrich et al., 2009; Benckiser et al., 2013; Kim et al., 2014), and charge-transfer transitions (Kim et al., 2004; Lu et al., 2005). RIXS spectrometers are also employed to isolate element-specific X-ray emission lines in (R)XES, including partial (PFY) or high-energy-resolution fluorescence-detected (HERFD) studies (Glatzel & Bergmann, 2005; Glatzel et al., 2009; Bauer, 2014). Finally, RIXS spectrometers can be used to measure resonant elastic X-ray scattering (REXS) with energy resolution of the scattered radiation in order to improve the sensitivity of the technique with respect to conventional energy-integrated detection (Chun et al., 2015).
(Kotani & Shin, 2001In a RIXS experiment, an incoming photon is prepared in a state of known energy , momentum and polarization . Upon absorption of the photon, the system undergoes a transition from an initial state, usually the ground state, to a highly excited intermediate state with a hole in a core level; within the lifetime of the core hole, then, the system decays to a lower energy state by emitting a photon. A RIXS spectrometer is then used to analyse the energy and momentum of the scattered photons. Also the scattered-photon polarization can be measured, both in the soft (Braicovich et al., 2014) and hard X-ray regimes (Ishii et al., 2011; Gao et al., 2011). The specific advantages of RIXS reside in the resonant nature of this technique, which naturally provides chemical, site and orbital selectivity. Bulk sensitivity and the possibility to exchange a finite momentum are ensured by the use of X-rays, in particular in the hard energy-range. In core-to-core RXES the intermediate state deep core-hole is filled by the radiative decay of a shallow core-electron under the emission of a photon with characteristic energy. A shallow core-hole is present in the final state of the system and detection of the scattered X-rays should be performed with an energy resolution below its intrinsic lifetime broadening. When the intermediate-state deep core-hole is filled by a valence electron, instead, valence-electron excitations can be studied in the energy-loss range of a few eV and the process is usually named valence-to-core RXES or sometimes simply RIXS. Energy resolution is therefore a crucial figure of merit for a spectrometer as it defines the energy scale of the lowest accessible excitations.
In this paper, we will present the guidelines that we followed in the design of the hard X-ray RIXS spectrometer installed on beamline ID20 of the ESRF, we will describe in detail its main characteristics and highlight the technical solutions we adopted. Key performance figures are also reported. The reader should be aware that, while most of the figures refer to the most challenging configuration of the spectrometer with the best energy resolution, we will not forget to discuss features that are relevant for other applications, including core-to-core (R)XES.
2. The multiple-element RIXS spectrometer
In the hard X-ray energy range, analysis of the scattered photon energy is performed using a Bragg crystal, which typically operates in the so-called Rowland geometry. On ID20, in particular, we adopted a solution based on Johann-type spherical crystal analysers (Johann, 1931), as in the case of several other spectrometers present at the ESRF (Glatzel et al., 2009; Llorens et al., 2012; Kvashnina & Scheinost, 2016) and elsewhere (Sokaras et al., 2013; Shvyd'ko et al., 2013; Ishii et al., 2013; Duan et al., 2017; Rovezzi et al., 2017). Nevertheless, the new ID20 spectrometer is a rather innovative instrument in several respects. It is installed on a dedicated beamline and has been conceived aiming at (i) maximizing the collection solid angle; (ii) optimizing the energy resolution; (iii) allowing various scattering geometries; in addition to stringent requirements in terms of precision and stability of the mechanical components. In the next subsections, we will detail the points above, while the ID20 beamline will be described elsewhere. It here suffices to mention that the monochromatic beam is focused at the sample position by a Kirkpatrick–Baez (KB) (Kirkpatrick & Baez, 1948) system down to less than ∼20 µm × 10 µm (H × V). The spectrometer assembly with its KB mirrors can be translated into a stand-by position when experiments are to be performed in the downstream hutch with its non-resonant (X-ray Raman spectroscopy) instrument (Huotari et al., 2017). The X-rays are detected by a 5 × 1 Maxipix (Ponchut et al., 2011) hybrid pixel detector (500 µm sensor thickness) with 55 µm × 55 µm pixel size and 14 mm × 70 mm active surface area.
2.1. Collection solid angle
In the Rowland geometry, diffraction (monochromatization) and focusing of hard X-rays are combined in a single optical element, a spherical crystal analyser. The source (beam spot on the sample), crystal analyser and detector must lie on a Rowland circle and their position is i.e. scattered photon energy , dependent. For convenience, the source of the spectrometer remains fixed in space, while the crystal analysers and detector are displaced to track .
,The solid angle of the spectrometer depends on the source-to-crystal analyser distance, the active surface of the crystal analyser and their number. On beamline ID20, the RIXS spectrometer is equipped with five crystal analysers operating on Rowland circles of either R = 1 or 2 m diameter, hosted in dedicated chambers operated in He atmosphere in order to reduce air absorption and scattering, as shown in Fig. 1. The crystal analysers and detector must simultaneously fulfil the Rowland condition and are therefore equipped with the necessary Fig. 2 shows the crystal analysers and detector assembly. The ith crystal analyser unit (i = 0, ±1, ±2) is equipped with two translations (xi and zi) and two rotations ( and , around the yi and zi axes, respectively) as shown in the inset of Fig. 2. The xi translation stages are directed towards the source and adjacent translations form an angle β of 7° (3.5°) in the R = 1 m (2 m) RIXS chamber. The detector unit is mounted on two translation stages (xD and zD) inclined by 21° with respect to the x and z axes, and can be rotated () around the yD y axis. The angle of inclination has been chosen in order to minimize the stroke of the xD translation stage. Details of the rotation and translation stages are given in Table 1. In Appendix A, we describe how to calculate the trajectories of the crystal analysers and detector in order to fulfil the Rowland condition as a function of the Bragg angle.
|
The signals from the five crystal analysers are collected simultaneously. In general, separating the foci of the crystal analysers on the detector is recommended as they may have slightly different energy resolutions and/or calibration, in which case merging them would cause a broadening of the overall energy resolution. For this reason, a two-dimensional detector is used on ID20. Also, a tight selection of the detector `region of interest' allows reduction of stray-scattering background and the cross-talk between foci that may leak into adjacent chips, as is the case in Fig. 3(c), and, depending on the geometry and scattering angle, discrimination of the signals of the sample and its environment (Huotari et al., 2011). On the other hand, the images of the different crystal analysers can intentionally be superimposed in order to minimize the detector exposure area and thus the stray-scattering background even further. Another case where the use of a two-dimensional detector is necessary is in experiments that require momentum resolution, as is often the case when studying electronic elementary excitations in single-crystal samples. Fig. 3 shows typical detector images obtained in various spectrometer configurations, where the signals from the various crystal analysers are focused at the centre of each chip. It should be noted that the use of a two-dimensional detector allows the crystal analyser focusing properties to be assessed; the irregular shape of the foci in Fig. 3(c) is indicative of a poor bent crystal analyers quality.
2.2. Energy resolution
Energy resolution is one of the most important figures of merit of a RIXS spectrometer, as it sets the lowest energy scale of measurable excitations. A number of contributions determines the actual energy resolution of a RIXS spectrometer (Huotari et al., 2006; Moretti Sala et al., 2013), which will be considered in the following.
2.2.1. Analyser contribution
The first contribution to energy resolution is the crystal analyser contribution . For high-resolution applications, spherical diced crystal analysers are used (Verbeni et al., 2005). In this case is well approximated by the intrinsic energy resolution of the crystal,
where is the crystal analyser α is the asymmetry angle. For lower-energy-resolution applications, spherically bent crystals are used. The calculation of is more complicated in this case. The energy resolution can be calculated according to
corresponding to the selected photon energy , is the Darwin width of the reflection andwhere t is the absorption length of the material (usually Si or Ge) and ν is its Poisson's ratio ( = 0.22 for Si and 0.27 for Ge). In the symmetric case ( = 0), it reduces to = . Additional details on the calculation of the energy resolution contribution of bent crystal analysers are given by Honkanen et al. (2014a, 2017).
2.2.2. Johann aberration
Another contribution to the energy resolution comes from the so-called Johann aberration, which causes the et al., 1999), where zi is the coordinate along the dispersive direction. The actual Johann aberration contribution to energy resolution is obtained from the histogram of the corresponding energy variations. For a crystal analyser with a circular mask of radius A the resulting energy distribution is shown in Fig. 4 to be rather sharp, but with a long asymmetric tail. In this case, an estimate of the Johann aberration is given by
to continuously vary on the surface of the crystal analyser as (Suorttiwhere = is the weighted average of the ). The Johann aberration contribution is negligible when working close to backscattering geometry. In addition, it can be eliminated either by choosing α such that = 90° or by using Johansson crystal analysers (Johansson, 1933). In the latter case, the reflecting lattice planes are bent to twice the Rowland circle radius, while the crystal analyser surface has the curvature of the Rowland circle. Their fabrication is, however, more complicated and Johansson spherical crystal analysers are not widely utilized yet. Otherwise, the Johann aberration can be minimized by limiting the active surface of the crystal analyser in the dispersive direction at the expense of a reduced solid angle.
variation across the crystal analyser surface (inset of Fig. 42.2.3. Pixel-size contribution
When diced crystal analysers are used, the system becomes partially dispersive and photons with different energies are separated in angle/space. The image on the surface of the detector has the dimension of twice the cube size and features an energy gradient along the detector's vertical coordinate (Huotari et al., 2006). The energy resolution, then, depends on how finely this distribution of photon energies is sampled on the detector surface and a pixel size contribution is introduced,
where = , with p the detector pixel size. When bent crystal analysers are used, instead, no exploitable energy gradient is produced on the detector surface. However, the spatial resolution of the detector can be used to enhance the energy resolution of a bent crystal analyser when it operates out-of-focus, as explained by Honkanen et al. (2014b).
2.2.4. Source-size contributions
Finally, one has to consider the finite source size, which introduces a distribution of point-like sources for the spectrometer. The three spatial dimensions sx, sy and sz are determined by the focal spot size at the sample position, the beam incidence angle on the sample surface and the X-ray probing depth. We then define sxi, syi and szi, the projections of the source in each crystal analyser's local coordinate system (top of Fig. 2), with sxi and szi lying in the crystal analyser dispersive plane and syi perpendicular to it. szi and syi induce a variation, which is first order in szi, = , and second order in syi, = . Accordingly, the energy resolution is broadened by
Finally, sxi introduces a little off-Rowland contribution, which is typically negligible. In order to minimize source size contributions, the X-ray spot size on the sample has to be made as small as possible. At ID20, the source size contribution is negligible for most of the applications.
In the case of our multiple-element RIXS spectrometer, these contributions are almost identical for all the crystal analysers and so is their energy resolution. As an example, we consider the specific case of Ir L3-edge RIXS experiments ( = 11.215 keV) employing five R = 2 m Si(844) diced crystal analysers ( = 85.7°), for which the various contributions to the energy resolution are = 14.7, = 7.1 (for a mask radius A = 30 mm), = 11.5 and = 4.2 meV. The calculated overall energy resolution, including the incident photon energy bandwidth = 14.7 meV determined by a Si(844) backscattering channel-cut, is 25.2 meV. Fig. 3(a) shows the foci of the five crystal analysers, which can be distinguished with the use of a two-dimensional detector. The displacement of the foci on the detector as a function of the incident photon energy is shown in the inset of Fig. 5. The slope of 4.7 µm meV−1 is related to the pixel size contribution to the energy resolution, p/4.7 = 11.7 meV pixel−1, in agreement with the estimate of given above. The measured elastic lines are shown in Fig. 5 to be 25 ± 3 meV wide, again in line with the estimate above.
In order to optimize the energy resolution at all scattered photon energies, ID20 is equipped with a collection of crystal analysers (and post-monochromators). For convenience, we report in Table 2 the calculated and measured energy resolutions at a few selected absorption edges.
|
3. Energy calibration and stability
With the setup for Ir L3-edge RIXS, we checked the reliability on the energy scale of the RIXS spectrometer by measuring the well known dispersion of longitudinal acoustic phonons in diamond along the (H,0,0) direction. This is shown in Fig. 6 (open blue circles) to overlap with literature data (red line) (Warren et al., 1967). Finally, we also tested the stability of the energy scale in the long term; the inset of Fig. 6 shows the change in the phonon energy at (3,0,0) momentum transfer over several hours. A small drift is observed, which can be attributed to heat-load effects of the beamline optics and/or temperature changes in the experimental hutch. Indeed, heat-load effects of the beamline optics cause a small change in the energy of the incident photons, as well as in the beam position on the sample. Note that a height change of the beam by 1 µm corresponds to approximately 1 meV in this case. Also, a temperature change of 0.1 K induces a 3 meV shift in the scattered photon energy, given the linear coefficient of Si at room temperature of 2.6 × 10−6 K−1. For this reason the temperature of the experimental hutch is stabilized within better than 0.5 K over 24 h. Considering the effects above, an (in)stability of ∼1 meV h−1 over several hours is remarkably good, given that these measurements were performed in the 16-bunch mode of ESRF, which has the largest ring current changes associated with storage-ring refills out of all the current-filling modes practical for RIXS at ESRF. It should be mentioned, however, that experiments are performed for operation modes of the storage ring with smaller relative changes of the ring current upon refilling and energy drifts are usually negligible. Moreover, the future implementation of top-up at the ESRF is expected to mitigate these undesired effects related to heat-load variations for all operation modes of the storage ring.
3.1. Scattering geometry
Beside energy loss, momentum transfer is another important observable in RIXS experiments. It is defined as = and it is varied by changing the scattering geometry, i.e. by tuning , while keeping fixed. The RIXS spectrometer on beamline ID20 is capable of rotating the spectrometer arm in both the horizontal and vertical directions: as shown in Figs. 7(a)–7(d), the rotation around the z axis at the bottom of the spectrometer carries the rotation around the y axis, thus allowing different scattering geometries.
In addition to providing full flexibility in the choice of the momentum transfer, the RIXS spectrometer also offers the possibility to study effects related to changes in the incident photon polarization. Indeed, the photon polarization in the horizontal laboratory plane corresponds to π (σ) incident photon polarization when scattering in the horizontal (vertical) This way polarization effects can partially be investigated, despite the polarization of the scattered photons not being analysed. However, in order to minimize the source size contribution to the energy resolution, the and the dispersive plane of the crystal analysers should be orthogonal. For this reason the RIXS chamber can be rotated around the source-to-middle crystal analyser axis, as shown in Figs. 7(b) and 7(c).
4. Sample goniometer and momentum transfer
A typical application of the RIXS spectrometer is the study of the momentum dependence of electronic and magnetic excitations in single crystals that need to be carefully aligned and oriented in order to reach specific points in , the translation stages are mounted on top of the rotation stages. The ω rotation around the z axis at the bottom carries, in order, the χ and the φ rotations, around the x and y axes, respectively. The sphere of confusion of the circles is less than 50 µm. Details of the rotation and translation stages are given in Table 3.
The sample stage therefore features three rotational and three translational As shown in Fig. 8
|
Given the rotation of the spectrometer arm and for the crystal analyser i = 0, we obtain the angles
where Ai,x, Ai,y and Ai,z are the positions of the ith crystal analyser (see Appendix A). With the convention used in Fig. 7, the momentum transfer in the laboratory reference frame is given by
where = are the transformation matrices describing the rotation of the -scattering arm. One then needs to project onto the sample via
where , and are transformation matrices: = describes the sample rotations; links the laboratory and the sample reference frames. It depends on how the sample is mounted on the sample stage and is found through the sample alignment procedure; finally, links the sample real space to its B.
reference frame and so it contains the dependence on the lattice parameters. Knowledge of these matrices allows one to easily obtain the momentum transfer associated with the various crystal analysers. Similarly, they can be used to estimate the momentum resolution. Details of the formalism are given in Appendix5. Conclusions and outlook
We have presented a multiple-element RIXS spectrometer installed at beamline ID20 at the ESRF. It has been conceived in order to maximize the collection solid angle, optimize the energy resolution and allow a flexible choice of the scattering geometry, for momentum and/or polarization resolved studies of the electronic structure of materials. It is mainly dedicated to the investigation of low-energy electronic excitation, but it can be exploited as well to the study of higher energy excitations or emission lines.
Future developments may include improvements of the energy resolution via the use of crystal analysers of different materials, such as lithium niobate (LiNbO3), sapphire (Al2O3) and α-quartz (SiO2). These materials have crystal structures of lower symmetry and thus offer more possible reflections than Si or Ge, with numerous choices of intrinsic resolution and throughput (Gog et al., 2013; Ketenoglu et al., 2015; Yavaş et al., 2017). Another option is to adopt different optical schemes than that of a conventional Rowland spectrometer, such as for example those based on the CDW design and its variants introduced by Shvyd'ko et al. (2006, 2011, 2014). In addition, future developments may comprise a smaller focal spot size, which will facilitate high-pressure experiments and high-energy-resolution applications. Also beneficial to energy resolution could be the advent of two-dimensional detectors with higher spatial resolution.
APPENDIX A
Multiple Rowland circles
We consider here the case of five identical crystal analysers and explain how the position (linear and angular) of the various crystal analysers and detector are determined as a function of the i = 0, ±1, ±2 labels the five crystal analysers. Similarly, the centre of the Rowland circle is located at . By introducing a reference system with the origin at the source and oriented as in Fig. 9(a), the one-crystal analyser case (i = 0) is described by S0,x = S0,y = S0,z = 0 for the source, A0,x = , A0,y = A0,z = 0 for the crystal analyser, D0,x = , D0,y = 0, D0,z = for the detector and R0,x = , R0,y = 0, R0,z = for the centre of the Rowland circle, where α allows for an asymmetric reflection.
. The position of the source, crystal analyser and its focus on the surface of the detector is described by the vectors , and , respectively, whereThe position of the generic ith crystal analyser should be calculated taking into account that (i) the Rowland circles intersect at the common source position and that choices in the mechanical implementation of the multiple crystal analysers spectrometer impose that (ii) the foci of the various crystal analysers should lie on, but not necessarily coincide with, the detector surface; (iii) the crystal analysers move on a translation stage directed towards the source. These stages are fixed on the spectrometer assembly and form an angle with respect to the central analyser translation.
For the generic ith crystal analyser case, constraint (i) imposes that = = 0. and are then determined by applying two consecutive rotations to the reference Rowland circle (i = 0) as explained in the following. Initially = and = . The first rotation is around the axis by an angle , and affects the position of the focus on the detector,
while leaving the position of the crystal analyser unchanged, = = , as is shown in Fig. 9(b). Constraint (ii) requires that
where di is the distance between the foci on the surface of the detector from its centre. Typically, di = id, with 14 mm, the lateral size of the individual chips. For d = 0 all the crystal analysers focus at the centre of the detector, while for d 0 each individual contribution can be separated. It is easy to see that equation (12) is fulfilled for = -id/D0,z, while equation (11) is naturally satisfied as does not affect the Di,x components.
The final crystal analyser position is obtained by applying a second rotation around by an angle ,
Obviously, this rotation does not alter the position of the focus on the detector, = , as is shown in Fig. 9(c). Constraint (iii) imposes that
where is the angle between the translation of the ith crystal analyser and the central one. By mechanical construction, = , where = 7° and 3.5° for the R = 1 m and 2 m case, respectively. Equation (14) is fulfilled for that satisfies
Finally, the centre of the Rowland circles is given by
and their orientation is such to intersect the corresponding source, crystal analyser and detector, as shown in Fig. 9(d). In addition, in order to correctly orient the crystal analysers and the detector, it is useful to consider the normal to the crystal analysers surface,
where = , and the normal to the detector surface,
where = . These are also shown in Fig. 9(d). Ultimately, a variation of the spectrometer energy involves two rotations (, ) and two translations (xi, zi) of the crystal analysers, and one rotation () and two translations (xD, zD) of the detector, which are calculated from and , respectively. Finally, Figs. 9(d) and 9(f) show two spectrometer configurations depending on the sign of d: for d > 0 the foci of the crystal analysers do not cross on the detector, while they do for d < 0.
APPENDIX B
, matrices, momentum transfer and momentum resolution
As mentioned above, the calculation of the momentum transfer in the sample ), the definition of is reported here for convenience,
requires knowledge of the and matrices. The former is related to how the sample is mounted in the laboratory reference frame, namely it defines the sample orientation before any rotation is applied. The latter defines a Cartesian reference frame in the sample and therefore it contains the dependence on the lattice constants. Following the convention of Busing & Levy (1967with the usual definitions of the primitive and
parameters.The momentum transfer given by equation (9) depends on the matrix, which needs to be determined prior to the experiment. The typical procedure is to find two non-collinear Bragg reflections and use them to derive the sample orientation in the laboratory reference frame. Ideally, the matrix satisfies the equations
simultaneously, where () is the first (second) Bragg reflection expressed in the sample ). By defining
while () and () contain information on the scattering arm and sample angles for which the first (second) Bragg reflection is found. However, because of experimental errors in the angle measurements and/or uncertainties in the cell parameters, it is not in general possible to find an orthogonal matrix which satisfies both conditions. In order to avoid this difficulty it is generally required that is parallel to , and is only constrained to lie in the plane of and (Busing & Levy, 1967and
then the matrix can be found as
where = and = (). If the and matrices are known, the sample (ω, χ and φ) and the scattering arm ( and ) angles can be calculated in order to reach a well defined momentum transfer .
Momentum resolution is related to uncertainties in the scattering arm angles ( and ) and ultimately to the finite aperture (A) of the crystal analysers by
where and are both of the order of (for A ). In the case of (quasi)elastic scattering, ≃ = and the expression above simplifies to
where = . As a typical example, we discuss the case of Ir L3-edge RIXS of a Sr3Ir3O7 single crystal ideally oriented with the (−1,−1,0) direction along the x axis and the (0,0,−1) direction along the y axis when all sample rotations are set to zero. We consider the case for = = 90° and = = = 0, for which = (0,0,26.7) r.l.u. Figs. 10(a), 10(c) and 10(e) show the momentum transfer spread across the surface of the crystal analyser (R = 2 m) with a circular aperture of radius A = 30 mm. The corresponding width of the momentum resolution function is shown in Figs. 10(b), 10(d) and 10(f) to be different in the various directions. Note that the shape of the crystal analyser mask could be varied in order to optimize the momentum transfer resolution in a specific direction depending on the experimental requirements. In any case, the finite momentum resolution of the spectrometer should be considered, in particular when interpreting REXS and RIXS results.
Acknowledgements
We kindly acknowledge the ESRF for financing the UPBL6 project as part of the ESRF Upgrade – Phase I. We also would like thank the ESRF for support during the conception, design, construction and commissioning of the spectrometer, specifically the TID and ISDD divisions.
References
Ament, L. J. P., Khaliullin, G. & van den Brink, J. (2011b). Phys. Rev. B, 84, 020403. CrossRef Google Scholar
Ament, L. J. P., van Veenendaal, M., Devereaux, T. P., Hill, J. P. & van den Brink, J. (2011a). Rev. Mod. Phys. 83, 705–767. CrossRef CAS Google Scholar
Ament, L. J. P., van Veenendaal, M. & van den Brink, J. (2011c). Europhys. Lett. 95, 27008. CrossRef Google Scholar
Bauer, M. (2014). Phys. Chem. Chem. Phys. 16, 13827–13837. CrossRef CAS Google Scholar
Benckiser, E., Fels, L., Ghiringhelli, G., Moretti Sala, M., Schmitt, T., Schlappa, J., Strocov, V. N., Mufti, N., Blake, G. R., Nugroho, A. A., Palstra, T. T. M., Haverkort, M. W., Wohlfeld, K. & Grüninger, M. (2013). Phys. Rev. B, 88, 205115. Web of Science CrossRef Google Scholar
Braicovich, L., Minola, M., Dellea, G., Le Tacon, M., Moretti Sala, M., Morawe, C., Peffen, J.-C., Supruangnet, R., Yakhou, F., Ghiringhelli, G. & Brookes, N. B. (2014). Rev. Sci. Instrum. 85, 115104. Web of Science CrossRef PubMed Google Scholar
Braicovich, L., van den Brink, J., Bisogni, V., Sala, M. M., Ament, L. J. P., Brookes, N. B., De Luca, G. M., Salluzzo, M., Schmitt, T., Strocov, V. N. & Ghiringhelli, G. (2010). Phys. Rev. Lett. 104, 077002. CrossRef Google Scholar
Brink, J. van den & van Veenendaal, M. (2012). Synchrotron Radiat. News, 25(4), 29–32. Google Scholar
Busing, W. R. & Levy, H. A. (1967). Acta Cryst. 22, 457–464. CrossRef IUCr Journals Web of Science Google Scholar
Calder, S., Kim, J. W., Taylor, A. E., Upton, M. H., Casa, D., Cao, G., Mandrus, D., Lumsden, M. D. & Christianson, A. D. (2016). Phys. Rev. B, 94, 220407. CrossRef Google Scholar
Cao, Y., Liu, X., Xu, W., Yin, W.-G., Meyers, D., Kim, J., Casa, D., Upton, M. H., Gog, T., Berlijn, T., Alvarez, G., Yuan, S., Terzic, J., Tranquada, J. M., Hill, J. P., Cao, G., Konik, R. M. & Dean, M. P. M. (2017). Phys. Rev. B, 95, 121103. CrossRef Google Scholar
Clancy, J. P., Gretarsson, H., Lee, E. K. H., Tian, D., Kim, J., Upton, M. H., Casa, D., Gog, T., Islam, Z., Jeon, B.-G., Kim, K. H., Desgreniers, S., Kim, Y. B., Julian, S. J. & Kim, Y.-J. (2016). Phys. Rev. B, 94, 024408. CrossRef Google Scholar
Cudazzo, P., Ruotsalainen, K. O., Sahle, C. J., Al-Zein, A., Berger, H., Navarro-Moratalla, E., Huotari, S., Gatti, M. & Rubio, A. (2014). Phys. Rev. B, 90, 125125. CrossRef Google Scholar
Dean, M. P. M., Dellea, G., Minola, M., Wilkins, S. B., Konik, R. M., Gu, G. D., Le Tacon, M., Brookes, N. B., Yakhou-Harris, F., Kummer, K., Hill, J. P., Braicovich, L. & Ghiringhelli, G. (2013a). Phys. Rev. B, 88, 020403. CrossRef Google Scholar
Dean, M. P. M., James, A. J. A., Springell, R. S., Liu, X., Monney, C., Zhou, K. J., Konik, R. M., Wen, J. S., Xu, Z. J., Gu, G. D., Strocov, V. N., Schmitt, T. & Hill, J. P. (2013b). Phys. Rev. Lett. 110, 147001. CrossRef Google Scholar
Dean, M. P. M., James, A. J. A., Walters, A. C., Bisogni, V., Jarrige, I., Hücker, M., Giannini, E., Fujita, M., Pelliciari, J., Huang, Y. B., Konik, R. M., Schmitt, T. & Hill, J. P. (2014). Phys. Rev. B, 90, 220506. CrossRef Google Scholar
Dean, M. P. M., Springell, R. S., Monney, C., Zhou, K. J., Pereiro, J., Božović, I., Dalla Piazza, B., Rønnow, H. M., Morenzoni, E., van den Brink, J., Schmitt, T. & Hill, J. P. (2012). Nat. Mater. 11, 850–854. Web of Science CrossRef CAS PubMed Google Scholar
Devereaux, T. P., Shvaika, A. M., Wu, K., Wohlfeld, K., Jia, C. J., Wang, Y., Moritz, B., Chaix, L., Lee, W.-S., Shen, Z.-X., Ghiringhelli, G. & Braicovich, L. (2016). Phys. Rev. X, 6, 041019. Google Scholar
Donnerer, C., Rahn, M. C., Sala, M. M., Vale, J. G., Pincini, D., Strempfer, J., Krisch, M., Prabhakaran, D., Boothroyd, A. T. & McMorrow, D. F. (2016). Phys. Rev. Lett. 117, 037201. Web of Science CrossRef PubMed Google Scholar
Duan, P., Gu, S., Cao, H., Li, J. & Huang, Y. (2017). X-ray Spectrom. 46, 12–18. CrossRef CAS Google Scholar
Gao, X., Burns, C., Casa, D., Upton, M., Gog, T., Kim, J. & Li, C. (2011). Rev. Sci. Instrum. 82, 113108. CrossRef Google Scholar
Ghiringhelli, G. & Braicovich, L. (2013). J. Electron Spectrosc. Relat. Phenom. 188, 26–31. Web of Science CrossRef CAS Google Scholar
Ghiringhelli, G., Brookes, N. B., Annese, E., Berger, H., Dallera, C., Grioni, M., Perfetti, L., Tagliaferri, A. & Braicovich, L. (2004). Phys. Rev. Lett. 92, 117406. CrossRef Google Scholar
Ghiringhelli, G., Le Tacon, M., Minola, M., Blanco-Canosa, S., Mazzoli, C., Brookes, N. B., De Luca, G. M., Frano, A., Hawthorn, D. G., He, F., Loew, T., Moretti Sala, M., Peets, D. C., Salluzzo, M., Schierle, E., Sutarto, R., Sawatzky, G. A., Weschke, E., Keimer, B. & Braicovich, L. (2012). Science, 337, 821–825. CrossRef CAS Google Scholar
Ghiringhelli, G., Piazzalunga, A., Dallera, C., Trezzi, G., Braicovich, L., Schmitt, T., Strocov, V. N., Betemps, R., Patthey, L., Wang, X. & Grioni, M. (2006). Rev. Sci. Instrum. 77, 113108. Web of Science CrossRef Google Scholar
Glatzel, P. & Bergmann, U. (2005). Coord. Chem. Rev. 249, 65–95. Web of Science CrossRef CAS Google Scholar
Glatzel, P., de Groot, F. M. F. & Bergmann, U. (2009). Synchrotron Radiat. News, 22, 12–16. CrossRef Google Scholar
Gog, T., Casa, D. M., Said, A. H., Upton, M. H., Kim, J., Kuzmenko, I., Huang, X. & Khachatryan, R. (2013). J. Synchrotron Rad. 20, 74–79. Web of Science CrossRef CAS IUCr Journals Google Scholar
Gretarsson, H., Sung, N. H., Porras, J., Bertinshaw, J., Dietl, C., Bruin, J. A. N., Bangura, A. F., Kim, Y. K., Dinnebier, R., Kim, J., Al-Zein, A., Moretti Sala, M., Krisch, M., Le Tacon, M., Keimer, B. & Kim, B. J. (2016). Phys. Rev. Lett. 117, 107001. CrossRef Google Scholar
Guarise, M., Dalla Piazza, B., Berger, H., Giannini, E., Schmitt, T., Rønnow, H. M., Sawatzky, G. A., van den Brink, J., Altenfeld, D., Eremin, I. & Grioni, M. (2014). Nat. Commun. 5, 5760. CrossRef Google Scholar
Guarise, M., Dalla Piazza, B., Moretti Sala, M., Ghiringhelli, G., Braicovich, L., Berger, H., Hancock, J. N., van der Marel, D., Schmitt, T., Strocov, V. N., Ament, L. J. P., van den Brink, J., Lin, P., Xu, P., Rønnow, H. M. & Grioni, M. (2010). Phys. Rev. Lett. 105, 157006. CrossRef Google Scholar
Haverkort, M. W. (2010). Phys. Rev. Lett. 105, 167404. Web of Science CrossRef PubMed Google Scholar
Honkanen, A.-P., Verbeni, R., Simonelli, L., Moretti Sala, M., Al-Zein, A., Krisch, M., Monaco, G. & Huotari, S. (2014b). J. Synchrotron Rad. 21, 762–767. Web of Science CrossRef CAS IUCr Journals Google Scholar
Honkanen, A.-P., Verbeni, R., Simonelli, L., Moretti Sala, M., Monaco, G. & Huotari, S. (2014c). J. Synchrotron Rad. 21, 104–110. CrossRef CAS IUCr Journals Google Scholar
Honkanen, A.-P., Verbeni, R., Simonelli, L., Moretti Sala, M., Monaco, G. & Huotari, S. (2017). J. Synchrotron Rad. 24, 545–546. CrossRef IUCr Journals Google Scholar
Huotari, S., Albergamo, F., Vankó, G., Verbeni, R. & Monaco, G. (2006). Rev. Sci. Instrum. 77, 053102. Web of Science CrossRef Google Scholar
Huotari, S., Pylkkänen, T., Verbeni, R., Monaco, G. & Hämäläinen, K. (2011). Nat. Mater. 10, 489–493. Web of Science CrossRef CAS PubMed Google Scholar
Huotari, S., Sahle, C. J., Henriquet, C., Al-Zein, A., Martel, K., Simonelli, L., Verbeni, R., Gonzalez, H., Lagier, M.-C., Ponchut, C., Moretti Sala, M., Krisch, M. & Monaco, G. (2017). J. Synchrotron Rad. 24, 521–530. CrossRef CAS IUCr Journals Google Scholar
Huotari, S., Vankó, Gy., Albergamo, F., Ponchut, C., Graafsma, H., Henriquet, C., Verbeni, R. & Monaco, G. (2005). J. Synchrotron Rad. 12, 467–472. Web of Science CrossRef CAS IUCr Journals Google Scholar
Hwan Chun, S., Kim, J.-W., Kim, J., Zheng, H., Stoumpos, C. C., Malliakas, C. D., Mitchell, J. F., Mehlawat, K., Singh, Y., Choi, Y., Gog, T., Al-Zein, A., Sala, M. M., Krisch, M., Chaloupka, J., Jackeli, G., Khaliullin, G. & Kim, B. J. (2015). Nat. Phys. 11, 462–466. CrossRef Google Scholar
Igarashi, J. & Nagao, T. (2012). Phys. Rev. B, 85, 064421. CrossRef Google Scholar
Ishii, K., Fujita, M., Sasaki, T., Minola, M., Dellea, G., Mazzoli, C., Kummer, K., Ghiringhelli, G., Braicovich, L., Tohyama, T., Tsutsumi, K., Sato, K., Kajimoto, R., Ikeuchi, K., Yamada, K., Yoshida, M., Kurooka, M. & Mizuki, J. (2014). Nat. Commun. 5, 3714. Web of Science PubMed Google Scholar
Ishii, K., Ishihara, S., Murakami, Y., Ikeuchi, K., Kuzushita, K., Inami, T., Ohwada, K., Yoshida, M., Jarrige, I., Tatami, N., Niioka, S., Bizen, D., Ando, Y., Mizuki, J., Maekawa, S. & Endoh, Y. (2011). Phys. Rev. B, 83, 241101. Web of Science CrossRef Google Scholar
Ishii, K., Jarrige, I., Yoshida, M., Ikeuchi, K., Inami, T., Murakami, Y. & Mizuki, J. (2013). J. Electron Spectrosc. Relat. Phenom. 188, 127–132. CrossRef CAS Google Scholar
Jia, C. J., Nowadnick, E. A., Wohlfeld, K., Kung, Y. F., Chen, C.-C., Johnston, S., Tohyama, T., Moritz, B. & Devereaux, T. P. (2014). Nat. Commun. 5, 3314. Google Scholar
Jia, C., Wohlfeld, K., Wang, Y., Moritz, B. & Devereaux, T. P. (2016). Phys. Rev. X, 6, 021020. Google Scholar
Johann, H. H. (1931). Z. Phys. 69, 185–206. CrossRef CAS Google Scholar
Johansson, T. (1933). Z. Phys. 82, 507–528. CrossRef CAS Google Scholar
Ketenoglu, D., Harder, M., Klementiev, K., Upton, M., Taherkhani, M., Spiwek, M., Dill, F.-U., Wille, H.-C. & Yavaş, H. (2015). J. Synchrotron Rad. 22, 961–967. Web of Science CrossRef CAS IUCr Journals Google Scholar
Kim, B. H. & van den Brink, J. (2015). Phys. Rev. B, 92, 081105. CrossRef Google Scholar
Kim, J., Casa, D., Upton, M. H., Gog, T., Kim, Y.-J., Mitchell, J. F., van Veenendaal, M., Daghofer, M., van den Brink, J., Khaliullin, G. & Kim, B. J. (2012a). Phys. Rev. Lett. 108, 177003. Web of Science CrossRef PubMed Google Scholar
Kim, J., Daghofer, M., Said, A. H., Gog, T., van den Brink, J., Khaliullin, G. & Kim, B. J. (2014). Nat. Commun. 5, 4453. Web of Science PubMed Google Scholar
Kim, J., Said, A. H., Casa, D., Upton, M. H., Gog, T., Daghofer, M., Jackeli, G., van den Brink, J., Khaliullin, G. & Kim, B. J. (2012b). Phys. Rev. Lett. 109, 157402. Web of Science CrossRef PubMed Google Scholar
Kim, Y.-J., Hill, J. P., Komiya, S., Ando, Y., Casa, D., Gog, T. & Venkataraman, C. T. (2004). Phys. Rev. B, 70, 094524. CrossRef Google Scholar
Kirkpatrick, P. & Baez, A. V. (1948). J. Opt. Soc. Am. 38, 766–774. CrossRef PubMed CAS Web of Science Google Scholar
Kotani, A. & Shin, S. (2001). Rev. Mod. Phys. 73, 203–246. Web of Science CrossRef CAS Google Scholar
Kvashnina, K. O. & Scheinost, A. C. (2016). J. Synchrotron Rad. 23, 836–841. CrossRef CAS IUCr Journals Google Scholar
Larson, B. C., Ku, W., Tischler, J. Z., Lee, C.-C., Restrepo, O. D., Eguiluz, A. G., Zschack, P. & Finkelstein, K. D. (2007). Phys. Rev. Lett. 99, 026401. Web of Science CrossRef PubMed Google Scholar
Lee, W. S., Johnston, S., Moritz, B., Lee, J., Yi, M., Zhou, K. J., Schmitt, T., Patthey, L., Strocov, V., Kudo, K., Koike, Y., van den Brink, J., Devereaux, T. P. & Shen, Z. X. (2013). Phys. Rev. Lett. 110, 265502. CrossRef Google Scholar
Lee, W. S., Lee, J. J., Nowadnick, E. A., Gerber, S., Tabis, W., Huang, S. W., Strocov, V. N., Motoyama, E. M., Yu, G., Moritz, B., Huang, H. Y., Wang, R. P., Huang, Y. B., Wu, W. B., Chen, C. T., Huang, D. J., Greven, M., Schmitt, T., Shen, Z. X. & Devereaux, T. P. (2014). Nat. Phys. 10, 883–889. CrossRef CAS Google Scholar
Le Tacon, M., Ghiringhelli, G., Chaloupka, J., Sala, M. M., Hinkov, V., Haverkort, M. W., Minola, M., Bakr, M., Zhou, K. J., Blanco-Canosa, S., Monney, C., Song, Y. T., Sun, G. L., Lin, C. T., De Luca, G. M., Salluzzo, M., Khaliullin, G., Schmitt, T., Braicovich, L. & Keimer, B. (2011). Nat. Phys. 7, 725–730. CrossRef CAS Google Scholar
Liu, X., Dean, M. P. M., Meng, Z. Y., Upton, M. H., Qi, T., Gog, T., Cao, Y., Lin, J. Q., Meyers, D., Ding, H., Cao, G. & Hill, J. P. (2016). Phys. Rev. B, 93, 241102. CrossRef Google Scholar
Llorens, I., Lahera, E., Delnet, W., Proux, O., Braillard, A., Hazemann, J.-L., Prat, A., Testemale, D., Dermigny, Q., Gelebart, F., Morand, M., Shukla, A., Bardou, N., Ulrich, O., Arnaud, S., Berar, J.-F., Boudet, N., Caillot, B., Chaurand, P., Rose, J., Doelsch, E., Martin, P. & Solari, P. L. (2012). Rev. Sci. Instrum. 83, 063104. Web of Science CrossRef PubMed Google Scholar
Lu, L., Chabot-Couture, G., Zhao, X., Hancock, J. N., Kaneko, N., Vajk, O. P., Yu, G., Grenier, S., Kim, Y. J., Casa, D., Gog, T. & Greven, M. (2005). Phys. Rev. Lett. 95, 217003. CrossRef Google Scholar
Lu, X., McNally, D. E., Moretti Sala, M., Terzic, J., Upton, M. H., Casa, D., Ingold, G., Cao, G. & Schmitt, T. (2017). Phys. Rev. Lett. 118, 027202. CrossRef Google Scholar
Marra, P., Sykora, S., Wohlfeld, K. & van den Brink, J. (2013). Phys. Rev. Lett. 110, 117005. CrossRef Google Scholar
Marra, P., van den Brink, J. & Sykora, S. (2016). Sci. Rep. 6, 25386. CrossRef Google Scholar
Marra, P., Wohlfeld, K. & van den Brink, J. (2012). Phys. Rev. Lett. 109, 117401. Web of Science CrossRef PubMed Google Scholar
Meyers, D., Miao, H., Walters, A. C., Bisogni, V., Springell, R. S., d'Astuto, M., Dantz, M., Pelliciari, J., Huang, H. Y., Okamoto, J., Huang, D. J., Hill, J. P., He, X., Božović, I., Schmitt, T. & Dean, M. P. M. (2017). Phys. Rev. B, 95, 075139. CrossRef Google Scholar
Minola, M., Dellea, G., Gretarsson, H., Peng, Y. Y., Lu, Y., Porras, J., Loew, T., Yakhou, F., Brookes, N. B., Huang, Y. B., Pelliciari, J., Schmitt, T., Ghiringhelli, G., Keimer, B., Braicovich, L. & Le Tacon, M. (2015). Phys. Rev. Lett. 114, 217003. CrossRef Google Scholar
Monney, C., Schmitt, T., Matt, C. E., Mesot, J., Strocov, V. N., Lipscombe, O. J., Hayden, S. M. & Chang, J. (2016). Phys. Rev. B, 93, 075103. CrossRef Google Scholar
Moretti Sala, M., Bisogni, V., Aruta, C., Balestrino, G., Berger, H., Brookes, N. B., de Luca, G. M., Di Castro, D., Grioni, M., Guarise, M., Medaglia, P. G., Miletto Granozio, F., Minola, M., Perna, P., Radovic, M., Salluzzo, M., Schmitt, T., Zhou, K. J., Braicovich, L. & Ghiringhelli, G. (2011). New J. Phys. 13, 043026. CrossRef Google Scholar
Moretti Sala, M., Boseggia, S., McMorrow, D. F. & Monaco, G. (2014a). Phys. Rev. Lett. 112, 026403. CrossRef Google Scholar
Moretti Sala, M., Henriquet, C., Simonelli, L., Verbeni, R. & Monaco, G. (2013). J. Electron Spectrosc. Relat. Phenom. 188, 150–154. Web of Science CrossRef CAS Google Scholar
Moretti Sala, M., Ohgushi, K., Al-Zein, A., Hirata, Y., Monaco, G. & Krisch, M. (2014b). Phys. Rev. Lett. 112, 176402. CrossRef Google Scholar
Moretti Sala, M., Schnells, V., Boseggia, S., Simonelli, L., Al-Zein, A., Vale, J. G., Paolasini, L., Hunter, E. C., Perry, R. S., Prabhakaran, D., Boothroyd, A. T., Krisch, M., Monaco, G., Rønnow, H. M., McMorrow, D. F. & Mila, F. (2015). Phys. Rev. B, 92, 024405. CrossRef Google Scholar
Moser, S., Shaik, N. E., Samal, D., Fatale, S., Dalla Piazza, B., Dantz, M., Pelliciari, J., Olalde-Velasco, P., Schmitt, T., Koster, G., Mila, F., Rønnow, H. M. & Grioni, M. (2015). Phys. Rev. B, 92, 140404. CrossRef Google Scholar
Peng, Y. Y., Hashimoto, M., Sala, M. M., Amorese, A., Brookes, N. B., Dellea, G., Lee, W.-S., Minola, M., Schmitt, T., Yoshida, Y., Zhou, K.-J., Eisaki, H., Devereaux, T. P., Shen, Z.-X., Braicovich, L. & Ghiringhelli, G. (2015). Phys. Rev. B, 92, 064517. CrossRef Google Scholar
Ponchut, C., Rigal, J. M., Clément, J., Papillon, E., Homs, A. & Petitdemange, S. (2011). J. Instrum. 6, C01069. Web of Science CrossRef Google Scholar
Rovezzi, M., Lapras, C., Manceau, A., Glatzel, P. & Verbeni, R. (2017). Rev. Sci. Instrum. 88, 013108. CrossRef Google Scholar
Shvyd'ko, Y., Hill, J., Burns, C., Coburn, D., Brajuskovic, B., Casa, D., Goetze, K., Gog, T., Khachatryan, R., Kim, J.-H., Kodituwakku, C., Ramanathan, M., Roberts, T., Said, A., Sinn, H., Shu, D., Stoupin, S., Upton, M., Wieczorek, M. & Yavas, H. (2013). J. Electron Spectrosc. Relat. Phenom. 188, 140–149. CAS Google Scholar
Shvyd'ko, Y. V., Lerche, M., Kuetgens, U., Rüter, H. D., Alatas, A. & Zhao, J. (2006). Phys. Rev. Lett. 97, 235502. PubMed Google Scholar
Shvyd'ko, Y., Stoupin, S., Shu, D., Collins, S. P., Mundboth, K., Sutter, J. & Tolkiehn, M. (2014). Nat. Commun. 5, 4219. Web of Science PubMed Google Scholar
Shvyd'ko, Y., Stoupin, S., Shu, D. & Khachatryan, R. (2011). Phys. Rev. A, 84, 053823. Google Scholar
Sokaras, D., Weng, T.-C., Nordlund, D., Alonso-Mori, R., Velikov, P., Wenger, D., Garachtchenko, A., George, M., Borzenets, V., Johnson, B., Rabedeau, T. & Bergmann, U. (2013). Rev. Sci. Instrum. 84, 053102. Web of Science CrossRef PubMed Google Scholar
Suortti, P., Buslaps, T., Fajardo, P., Honkimäki, V., Kretzschmer, M., Lienert, U., McCarthy, J. E., Renier, M., Shukla, A., Tschentscher, T. & Meinander, T. (1999). J. Synchrotron Rad. 6, 69–80. Web of Science CrossRef CAS IUCr Journals Google Scholar
Ulrich, C., Ament, L. J. P., Ghiringhelli, G., Braicovich, L., Moretti Sala, M., Pezzotta, N., Schmitt, T., Khaliullin, G., van den Brink, J., Roth, H., Lorenz, T. & Keimer, B. (2009). Phys. Rev. Lett. 103, 107205. Web of Science CrossRef PubMed Google Scholar
Verbeni, R., Kocsis, M., Huotari, S., Krisch, M., Monaco, G., Sette, F. & Vanko, G. (2005). J. Phys. Chem. Solids, 66, 2299–2305. Web of Science CrossRef CAS Google Scholar
Warren, J. L., Yarnell, J. L., Dolling, G. & Cowley, R. A. (1967). Phys. Rev. 158, 805–808. CrossRef CAS Web of Science Google Scholar
Yavaş, H., Sutter, J. P., Gog, T., Wille, H.-C. & Baron, A. Q. (2017). MRS Bull. 42, 424–429. Google Scholar
Yavaş, H., van Veenendaal, M., van den Brink, J., Ament, L. J. P., Alatas, A., Leu, B. M., Apostu, M.-O., Wizent, N., Behr, G., Sturhahn, W., Sinn, H. & Alp, E. E. (2010). J. Phys. Condens. Matter, 22, 485601. Web of Science PubMed Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.