

research papers
A high-energy Laue X-ray emission spectrometer at the FXE instrument at the European XFEL
aEuropean XFEL, Holzkoppel 4, 22869 Schenefeld, Germany, bCenter for Energy and Environmental Sciences, Paul Scherrer Institute, CH-5232 Villigen, Switzerland, cCenter for Photon Science, Paul Scherrer Institute, CH-5232 Villigen, Switzerland, and dDepartment of Modern Physics, University of Science and Technology of China, 230026 Hefei, People's Republic of China
*Correspondence e-mail: xinchao.huang@xfel.eu, christopher.milne@xfel.eu, frederico.lima@xfel.eu
The high-energy-resolution e.g. 4d metals), which thus do not fully profit from the high-photon-energy capabilities of the European XFEL. Here we present the design, implementation and performance of a novel high-resolution spectrometer operating in Laue (transmission) geometry optimized for measurements at high photon energies (>15 keV). The High-Energy Laue X-ray emIssiOn Spectrometer (HELIOS) operates mainly in dispersive mode by placing the crystal analyzer inside or outside the Rowland circle. The Laue spectrometer performance in terms of energy resolution and efficiency is presented and discussed. Two Laue analyzers, silicon and quartz, have been tested at SuperXAS of the Swiss Light Source and at FXE of the European XFEL. The quartz analyzer was found to be about 2.7 times more efficient than the silicon one. The Laue spectrometer energy resolution (ΔE/E) reached at the FXE instrument is around 1.2 × 10−4. Depending on different user requests, the resolution can be further increased by using higher diffraction orders. The new Laue spectrometer increases the existing portfolio of spectrometers at FXE, enabling efficient implementation of ultrafast X-ray spectroscopies with high energy resolution at photon energies above 15 keV. This spectrometer will allow the expansion of studies in the field of ultrafast sciences, particularly including investigation of 4d elements using hard X-rays.
spectrometers available at the Femtosecond X-ray Experiment (FXE) instrument of the European XFEL operate in Bragg (reflective) geometry, with optimum performance in the range between 5 and 15 keV. However, they quickly lose efficiency above around 15 keV due to the decrease in reflectivity of the crystal analyzers at such high photon energies. This hampers high-energy-resolution spectroscopy experiments on heavy elements (Keywords: X-ray free-electron lasers; X-ray spectrometers; X-ray emission spectroscopy; DuMond geometry; ultrafast science; Laue crystal analyzers.
1. Introduction
The European X-ray K edges of 4d elements and L edges of 5f elements (Chen et al., 2021). Accessing those edges results in distinctive opportunities for applying X-ray core level spectroscopies in the investigation of the temporal evolution of physical, chemical and biological processes in condensed matter. X-ray spectroscopies have the advantage of being sensitive to the electronic configurations and local structures around the absorbing element, and have routinely been employed in the study of photo-excited processes of materials at synchrotron radiation and XFEL facilities (Milne et al., 2014
; Gawelda et al., 2016
; de Groot et al., 2024
; Bergmann et al., 2021
). To date, the majority of ultrafast spectroscopic experiments conducted on free-electron lasers in the hard X-ray regime have been limited to ∼4.5–10 keV, with time-resolved (t.r.) X-ray absorption and emission (XAS and respectively) being well established and broadly used for systems containing 3d (K edge) and 5d (L edge) metals (Zhang et al., 2014
; Canton et al., 2015
; Levantino et al., 2015
; Mara et al., 2017
; Lemke et al., 2017
; Katayama et al., 2019
; Kunnus et al., 2020
; Smolentsev et al., 2020
; Bacellar et al., 2023
; Sension et al., 2023
; Nowakowski et al., 2024
; Naumova et al., 2024
). Ultrafast t.r. X-ray spectroscopies have rarely been employed in very hard X-ray regimes, mainly due to the unavailability of free-electron laser sources producing sufficiently intense pulses above 15 keV and efficient high-resolution X-ray spectrometers for such experiments. In fact, high-resolution spectrometers optimized for high photon energies are rare even on synchrotron sources (Ravel et al., 2018
; Jagodziński et al., 2019
).
Static high-energy-resolution X-ray spectroscopies on systems containing 3d (K edges) and 5d (L edges) metals are becoming routine and are used to address a variety of relevant science questions (Lancaster et al., 2011; Sá et al., 2014
; Torres Deluigi et al., 2014
; Pollock & DeBeer, 2015
; Hunault et al., 2017
; Schuth et al., 2018
; Castillo et al., 2021
; Cutsail & DeBeer, 2022
). On the other hand, only recently have high-resolution and on the K edge of 4d metals started to become more common, indicating the importance of expanding the capabilities for performing such experiments with improved efficiency (Doonan et al., 2005
; Lima et al., 2013
; Bjornsson et al., 2014
; Lezcano–González et al., 2016
; Ravel et al., 2018
; Castillo et al., 2020
). In this sense, extending the high-energy-resolution capabilities at XFELs to high photon energies is a natural development. This would significantly expand the scope of science covered, allowing the investigation of photo-induced processes in materials containing 4d elements at XFELs, e.g. charge-transfer dynamics on dyads containing 4d metals (Ducrot et al., 2016
; Huijser et al., 2018
), the photocatalytic properties of niobium and molybdenum nanoparticles (Su et al., 2021
; Ducrot et al., 2016
) and ruthenium dye-sensitized nanoparticles for solar cell applications (Zhang et al., 2011
), the ultrafast evolution of strongly correlated materials (Witczak-Krempa et al., 2014
) and new candidates for battery cathodes and anodes (Han et al., 2023
), among others. These materials often exhibit unique competition between fundamental interactions. For instance, the dominant role of electronic correlations and spin–orbit coupling may be inverted in 3d and 4d metals, resulting in distinct physical behaviors on materials containing 4d metals compared with their 3d counterparts (Cao & Delong, 2013
). Moreover, spectroscopic studies at such high energies would nicely match the advantages of scattering and diffraction experiments, potentially done in parallel in this regime, in terms of wider coverage of the momentum transfer (Dunne et al., 2023
). Additionally, core-level spectroscopy in a high-photon-energy regime suffers from an intrinsic limitation in the energy resolution due to the large core-hole lifetime broadening (Ogasawara et al., 1994
). This is another motivation for developing high-energy-resolution spectrometers (Hiraoka et al., 2013
) to circumvent this limitation by resonant and high-energy-resolution fluorescence-detected (HERFD-XAS) (Lima et al., 2013
; Hämäläinen et al., 1991
).
Currently, there are two X-ray emission spectrometers installed at the Femtosecond X-ray Experiment (FXE) instrument which have been successfully used in several user experiments since inauguration (Naumova et al., 2020a; Naumova et al., 2020b
; Kinschel et al., 2020
; Bacellar et al., 2020
; Bacellar et al., 2023
; Sension et al., 2023
; Canton et al., 2023
; Naumova et al., 2024
; Nowakowski et al., 2024
; Sension et al., 2024
). Both von Hamos and Johann X-ray spectrometers, previously described by Galler et al. (2019
) and Lima et al. (2023
), operate in reflective Bragg diffraction geometry using silicon and germanium crystal analyzers. They typically provide an energy resolution on the order of 1 eV or even below. However, these Bragg-reflection analyzers present a significant limitation in the overall spectrometer efficiency at high photon energies. The X-ray on the Si and Ge analyzers becomes significant at energies above ∼15 keV, with a consequent quick decrease in the crystal reflectivity as the energy increases. Even though using higher-index diffraction orders can result in an enhancement in the crystal reflectivity to partially compensate this, the corresponding Darwin width gets remarkably narrow, balancing the gain in reflectivity and preventing a reasonable efficiency increase (Shvyd'Ko, 2004
; Szlachetko et al., 2013
; Jagodziński et al., 2019
). Hence, the applicability of those spectrometers in investigations of heavier elements, for example 4d metals, is hampered (Castillo et al., 2020
). This not only represents a significant restriction on the science cases that can be tackled at FXE but also a limit on the exploration of the ultrafast high X-ray photon energy capabilities of the European XFEL.
In order to overcome these limitations and increase the existing portfolio of high-resolution X-ray spectrometers at FXE, we have designed a dedicated high-energy Laue (transmission-type) et al., 2013; Szlachetko et al., 2013
; Ravel et al., 2018
; Jagodziński et al., 2019
), in which the diffracted beam is transmitted through a thin crystal analyzer, results in higher efficiency and comparable energy resolution with that reached using Bragg-reflection geometry. Noteworthy, in Laue geometry the crystal planes are in the same direction as the beam, rather than perpendicular. Herein we present our High-Energy Laue X-ray emIssiOn Spectrometer (HELIOS) with an optimized analyzer design for improved efficiency. An energy resolution (ΔE/E) of ∼1.2 × 10−4 was reached and the improved efficiency is demonstrated by comparing the signal magnitude obtained with the Laue spectrometer with that measured simultaneously with the FXE von Hamos spectrometer and also with the Laue-type spectrometer reported in the work of Jagodziński et al. (2019
). Additionally, the Laue spectrometer reported here provides a less distorted emission image when operating in dispersive mode compared with the Laue spectrometer reported in the work of Ravel et al. (2018
). This is particularly relevant for pump–probe experiments in which consecutive laser-on and laser-off spectra are subtracted and any artifacts arising from image processing should be minimized. The successful inauguration of this new emission spectrometer, the only such equipment at XFELs, should allow major questions in the field of ultrafast science to be addressed.
2. Spectrometer design concepts and optimization
Laue-type analyzers have been used for more than half a century in several spectroscopic applications, including identification of fluorescence lines of very high Z atoms and ions (Borchert et al., 1975; Widmann et al., 1997
), measurements of broad Compton profiles with high energy resolution (Suortti et al., 1999
; Itou et al., 2001
; Hiraoka et al., 2001
; Hiraoka et al., 2005
) and to resolve fluorescence lines and suppress background scattering for detection (Zhong et al., 1999
; Kropf et al., 2003
; Kropf et al., 2005
; Wakisaka et al., 2017
). They have also been used in an X-ray Raman spectrometer utilizing ∼20 keV photons (Hiraoka et al., 2013
). However, the energy resolution achieved in these experiments was usually limited to ΔE/E ≃ 10−3, with the work from Hiraoka et al. reporting ΔE/E ≃ 10−4 by using higher diffraction orders and correcting the scattering intensity variations across the analyzer surface. This is not sufficient for efficient high-energy-resolution experiments which require an energy resolution comparable with or better than the natural linewidths of the emission lines and high count rates. For example, the Nb Kα line at around 16.5 keV has a natural linewidth of ∼6 eV and a spectrometer with ΔE/E ≃ 10−3 would correspond to 16 eV resolution, almost three times larger than the natural linewidth (Campbell & Papp, 2001
). Recently, high-photon-energy measurements using Laue analyzers with high energy resolution were reported by Ravel et al. (2018
) and Jagodziński et al. (2019
). In these works the Laue crystal was bent to a cylindrical shape and the DuMond geometry was used (DuMond, 1947
).
There are two working geometries for Laue analyzers, the Cauchois type (Cauchois, 1932) for extended sources and the DuMond type for point-like sources. The DuMond geometry can be regarded as an inverse Johann geometry, as shown in Fig. 1
, where the diffracted light passing through the crystal does not converge but diverges onto the detector. Therefore, specific characteristics of the Laue analyzer should be considered as they can induce distortions on the spectral image on the detector. Compared with reflection geometry, the transmission geometry requires an open aperture on the crystal support, which can induce a saddle-like distortion in the analyzer surface due to anisotropic Poisson-ratio effects (Lethbridge et al., 2010
) and worse surface distortion on the edge of the open area. Furthermore, the emission image on the detector is sensitive to distortions in the analyzer surface due to the unfocused characteristic of the DuMond geometry (see Fig. 1
). Those distortions can result in complex emission images with a consequent effect in final energy resolution. Both Laue spectrometers described by Ravel et al. (2018
) and Jagodziński et al. (2019
) operate in the off-Rowland condition of the DuMond geometry to mitigate the effects arising from the analyzer surface distortion. However, different working modes are employed in the measurements: Jagodzinksi et al. used linked scans and Ravel et al. employed dispersion mode.
![]() | Figure 1 Schematic of the DuMond geometry. The Laue analyzer is bent to a cylindrical shape with curvature of RC. The light is diffracted by crystal planes oriented in the same direction as the incoming beam, instead of perpendicular planes as in Bragg geometry. The emission is collected from a point-like source and imaged on a detector. When operating in on-Rowland geometry the source and the Laue crystal are located exactly on the Rowland circle and the distance is RC. When operating in off-Rowland geometry, the Laue analyzer has an offset from the Rowland circle. The detector is behind the analyzer. |
The development of the high-energy Laue emission spectrometer at FXE started with an investigation of the optimum analyzer thickness and the effects of asymmetric angles on the diffraction efficiency. Different geometries were evaluated through X-ray tracing simulations, which were carried out using SHADOWOUI of the OASYS suite (Rebuffi & Rio, 2017; Rebuffi & Sanchez del Rio, 2016
). This allowed additional optimization of key design parameters to improve the spectrometer efficiency and simplify the image processing, while providing routes for a final energy resolution compatible with the emission measurement requirements. Guided by our X-ray tracing simulations and considering the pump–probe character of the experiments at FXE and specifics of XFEL operation, the energy-dispersive mode proved to be more appropriate and was chosen for the Laue spectrometer presented here.
2.1. X-ray transmission optimization
The Laue spectrometer is expected to operate in the >15 keV range and, given that the standard minimum thickness of commercial perfect crystals is usually limited to several hundreds of micrometres, the X-ray transmission of Laue analyzers can be a limiting factor. For example, Jagodziński et al. (2019) employed a silicon crystal with 0.5 mm thickness, which translates into ∼20% diffraction efficiency at ∼19 keV. Here, we have chosen silicon (111) and quartz (110) crystals as the Laue analyzers for the spectrometer, with thicknesses of 300 µm and 400 µm, respectively. One way to increase the diffraction efficiency given a fixed crystal thickness is to use asymmetric cuts. Fig. 2
(a) illustrates the asymmetric cut in Laue analyzers, referring to the presence of an angle α between the physical plane and the crystal planes. In general, this can tune the effective diffraction efficiency of the crystal and also the angular acceptance. In this work, the definition of α for Laue crystals differs from that for Bragg crystals, with a 90° rotation offset between them (see Fig. 2
). In general, both the diffraction intensity and the Darwin width increase with increasing α for Laue crystals, meaning that a careful choice of the asymmetric angle is required to prevent degradation of the energy resolution.
![]() | Figure 2 (a) Schematic of asymmetric cutting for the Laue crystal. The asymmetric angle α is defined as the cross angle between the crystal and planes. When an asymmetric angle is present, the incidence angle θin = θB + α, where θB is the θin + θout is still 2θB, so the asymmetric cutting will not affect the geometry for the detector. Calculated rocking curves at different asymmetric cutting angles for (b) Si(333) and (c) SiO2(330). The integrated area and FWHM are shown in the insets. An example energy of 18.625 keV is considered in the calculations. |
The evaluation of the influence of varying crystal thickness and asymmetric cutting angles was performed using dynamic diffraction calculations carried out using the XOP package (Sanchez del Rio & Dejus, 2011). Figs. 2(b) and 2(c) show the calculated rocking curves for the third-order diffraction of silicon (111) and quartz (110) crystals, respectively. These could be used for measuring niobium Kβ (∼18.62 keV). As expected, compared with the symmetrically cut case (α = 0°) the diffraction intensity increases with increasing α. For the Si(333) diffraction, the X-ray tracing simulations show that as α increases from 0° to 2°, the diffraction intensity increases from around 0.25 to 0.28 with a negligible increase in the full width at half-maximum (FWHM) of the rocking curve, going from 3.3 µrad to 3.4 µrad with α = 0° and 2°, respectively. As α increases further, this intensity enhancement becomes less significant and the main effect is a further broadening in the Darwin width, as shown by the similar Si(333) rocking curves in Fig. 2
(b) with α = 4° or 5°. A similar behavior is also observed for the SiO2(330) diffraction, as shown in Fig. 2
(c). Compared with Si, SiO2 exhibits a more pronounced sensitivity to the asymmetric cut. This is likely due to the SiO2 crystal having more scattering centers within its contributing to the diffraction, resulting in stronger interference effects. These results indicate that using relatively small asymmetric angles α results in a noticeable increase in the diffraction efficiency. The associated increase in the Darwin widths can further improve the diffraction efficiency. On the other hand, if the Darwin width becomes too large the energy resolution can be affected, so a compromise needs to be found. This increase in Darwin width due to asymmetric angles is more significant for higher-order diffraction indices, e.g. fifth order for Si(111) and fourth order for SiO2(110) (Shvyd'Ko, 2004
). We have chosen α = 2.5° and 3.0° for the silicon and quartz analyzers, respectively. The Darwin widths for both analyzers are <5 µrad, which, based on an estimation for the energy resolution of ΔE =
, with ΔθD being the Darwin width, leads to a contribution of less than 0.3 eV to the energy resolution at the Nb Kβ energy (18.62 keV).
2.2. Laue geometry and X-ray tracing simulations
The Laue analyzer can be bent into either a logarithmic spiral or a simple cylindrical shape. These are used to address the mismatch between the narrow Darwin width of a perfect crystal and the intrinsic nature of the 4π divergence of emission. The analyzer with a logarithmic spiral shape operating in Rowland geometry can, in principle, collect a single photon energy emitted from a point source across the entire area of the analyzer and minimize the background scattering. However, the energy resolution is affected by the commonly observed distorted diffraction images. X-rays with different energies originating from the same area on the crystal may overlap in the detector, leading to insufficient energy resolution for measurements. Thus, analyzers with a logarithmic spiral are typically employed for detection in fluorescence geometry (Zhong et al., 1999; Kropf et al., 2003
; Kropf et al., 2005
; Wakisaka et al., 2017
), rather than measurement. The complexity of those diffraction images may be attributed to pronounced distortions caused by variations in the local bending radius from one analyzer end to the other. The simpler cylindrical shape likely results in less distorted surfaces. However, since the Laue analyzer operates in defocused geometry, the diffraction image is more sensitive to surface distortions and aberrations, in comparison with the Bragg crystals. The challenges posed by complex diffraction images cannot be solved by simply modifying the bending geometry. Instead, a suitable working mode is required. For example, the off-Rowland condition was employed on other spectrometers using Laue analyzers (Ravel et al., 2018
; Jagodziński et al., 2019
) in order to have a more uniform image. On the other hand, the magnitude of the bending curvature can also significantly affect the image. For instance, a complex `S'-shaped stripe was observed with a bending curvature of 0.5 m by Ravel et al. (2018
), while a simpler single line stripe was observed in the detector with a bending curvature of 2.0 m by Jagodziński et al. (2019
).
Laue analyzers with a cylindrical shape are able to produce dispersion behavior even when working in exact Rowland geometry. This dispersive ability arises from analyzer aberrations both in the bent and the non-bent directions, as illustrated in Fig. 3. In the non-bent direction there are different angles and distances between the (point) source and analyzer surface on both the upper and lower analyzer parts with respect to the center region. For the bent direction, the emitted X-rays also hit the crystal surface at different angles because of the mismatch between the bent curvature and the Rowland circle. A similar concept was also realized using Bragg analyzers in combination with position-sensitive pixel detectors to improve the energy resolution (Huotari et al., 2005
; Szlachetko et al., 2012
; Alonso-Mori et al., 2012
; Moretti Sala et al., 2018
; Huotari et al., 2017
). Given the well determined geometry of the relative positions of the crystal surface and the point source, the angular difference across the whole analyzer area can be calculated, providing a map of the different analyzer regions collecting different energies. To illustrate this we have performed X-ray tracing simulations of 18625 eV (Nb Kβ1) photons diffracting from a Si(333) Laue analyzer with 155 cm curvature in on-Rowland geometry, converting the angular difference to energy difference, as shown in the 2D mesh in Fig. 3
(c). The central region analyzes energies with the smallest difference with respect to the central energy. On the other hand, the edge regions in both bent and non-bent directions exhibit larger dispersion, with a maximum energy difference of about 20 eV if considering an analyzer with dimensions of 5 cm × 10 cm. This would be insufficient to measure a full Kβ1,3 of a 4d metal, which spans around 100 eV. Moreover, the areas corresponding to different analyzed energies are not uniform, resulting in varying intensities for different energies. This inconsistency makes emission spectra measured using Laue analyzers in on-Rowland geometry challenging to analyze and unsuitable for applications envisioned at the FXE instrument.
![]() | Figure 3 Calculated dispersion of the cylindrical shape analyzer working on the Rowland circle condition. Panels (a) and (b) illustrate the schematics of the aberrations in the bent and non-bent directions. In (c) the 2D dispersion map after converting the angular difference to the energy difference is shown. |
A simpler dispersive behavior can be obtained for Laue analyzers when they are located in an off-Rowland circle geometry. This approach has been previously used to collect emission spectra for niobium compounds (Ravel et al., 2018). Fig. 4
shows the calculated emission images corresponding to different energies when a Laue analyzer with 155 cm radius is placed 5 cm off from the Rowland circle. Each emission image corresponding to a single energy is a unique bent stripe that moves along the bent direction and slightly changes its curvature with changing photon energy. The non-bent plane also exhibits energy dispersion, as indicated by the curved emission image. However, for a realistic Laue analyzer with limited size, e.g. 8 cm × 3 cm (bent and non-bent directions, respectively) the emission stripe becomes simpler, as shown by the signal inside red boxes in Fig. 4
. Within the analyzer area, the shape of the emission stripe approaches a single line and moves along the bent direction. Additionally, in this configuration the energy bandpass at about 18.9 keV is close to 80 eV, being large enough to allow dispersive measurements of the complete Based on the tracing results of on- and off-Rowland conditions, one can imagine that if the analyzer is positioned on the Rowland circle, the image becomes diffuse and is more significantly affected by surface distortions. Conversely, if the analyzer is positioned off the Rowland circle, the image becomes more localized and is less affected by surface distortion. Therefore, the off-Rowland geometry provides a reasonable solution to mitigate the effects arising from surface distortion which is typically a problem of Laue analyzers.
![]() | Figure 4 (a) Calculated emission images of a Laue analyzer for different photon energies when working under the off-Rowland condition. A point-like source with beam size of 100 µm × 100 µm in FWHM, energy resolution of 1.4 × 10−4 (ΔE/E, where ΔE is defined by FWHM) and varying incoming beam central energies were considered as the emitted source. A Si(333) Laue analyzer with a cylindrical shape and a bent curvature of 155 cm was used. Surface distortions were omitted here to simplify the X-ray tracing. The analyzer is positioned 5 cm farther from the Rowland circle. The red boxes in the figures represent a realistic size of a Laue analyzer. (b) Energy resolution evaluation using a beam size of 144 µm. A 15 mm-wide region of interest (ROI) at the central position is selected to project the spectrum. A series of Gaussian fittings are performed to extract the peak positions, followed by polynomial fitting to determine the position-to-energy calibration. The energy resolution of the analyzer is then obtained by deconvoluting the Gaussian functions. (c) The energy resolution as a function of beam size is determined using the same procedure described in (b) for various beam sizes. |
When X-ray optics work in dispersive mode, a compromise needs to be found between their efficiency and energy resolution. In principle, the shorter bent curvature leads to a larger solid angle of collection, resulting in better efficiency. However, it also results in worse energy resolution because the emitted source cannot be a point without dimensions. In pump–probe experiments at the FXE instrument of the European XFEL, the desired beamsize varies from tens of micrometres to hundreds of micrometres depending on different sample and experimental conditions. In this range, the ratio of source size to distance is typically the primary contributor to the energy resolution. Therefore, the bent curvature needs to be carefully evaluated to ensure compatibility with the d metals with the K edge in the 15–26 keV range, where the natural linewidths of emission lines vary from about 5 eV to 10 eV. The example tracing images are shown in Fig. 4(a); a beam size of l00 µm was considered in the tracing simulation presented here. The photon energy was set to 18953.4 eV, corresponding to the Nb Kβ2 emission and an energy resolution of 1.4 × 10−4 was chosen to create a Gaussian-distributed monochromatic beam from a Si(111) double crystal. A Si(333) Laue analyzer with a cylindrical curvature and a bent radius of 155 cm was chosen for X-ray tracing. Distortions on the analyzer surface were not considered in order to simplify the X-ray tracing. The analyzer is positioned 5 cm away from the Rowland circle to produce sufficient energy dispersion to cover the Nb Kβ2 emission. The theoretically expected energy resolution is depicted in Fig. 4
(b). The central region is selected to project the spectrum. A series of Gaussian fittings are performed to extract the peak positions, followed by polynomial fitting to determine the position-to-energy calibration. The energy resolution of the analyzer is then obtained by deconvoluting the Gaussian functions. For example, the tracing results in 6.2 eV FWHM when using a 144 µm point-like source (to account for the 45° projection in the sample of an initial 100 µm beam). In the experimental tests (Section 4
), a slightly worse resolution (∼6.7 eV) was obtained when using a similar beam size. We attribute this to contributions from analyzer surface distortions, crystal strain and effects of the detector pixel size, all of which are omitted in the simulations. When the beam size is decreased to 100 µm, the expected energy resolution goes down to 4.7 eV FWHM, which is sufficient for measuring high-resolution niobium As shown in Fig. 4
(c), the resolution can ideally be improved to better than 2 eV by reducing the beam size to below 10 µm. The energy resolution can be further improved by using higher Bragg angles, i.e. higher diffraction indices. However, the narrower Darwin widths associated with higher diffraction indices will also result in lower efficiency. In the X-ray tracing simulations we have chosen relatively moderate diffraction indices, e.g. Si(333) and SiO2(330). This allowed us to first optimize the bent curvature and carry out an initial evaluation of the practical aspects of operation.
2.3. Analyzer design
The X-ray tracing simulations guided the choice of the most appropriate analyzers for the Laue spectrometer. The analyzer design profits from advances in manufacturing technologies that allowed the use of curvatures of relatively short radius and a fixed radius by using a newly engineered support frame. A compromise was found by using a ∼150 cm bending curve, which proved to be sufficiently large to not induce noticeable surface distortion as observed when using a 50 cm radius analyzer (Ravel et al., 2018). Reducing the radius from 2 m used in the dynamically bent analyzers (Jagodziński et al., 2019
) to the chosen ∼150 cm results in an increase in by a factor of about 1.8. The analyzer support frame was engineered to have thin walls (∼11 mm), providing a large clear aperture of 80 cm × 30 cm in the bent and non-bent directions, respectively. Note that the crystal will be subjected to saddle-like distortions, even though the frame is cylindrical only in one direction. In the spectrometer reported by Jagodzinksi et al., a large area of analyzer was shadowed by the crystal bender; thus our design results in a usable area comparable with that of the much larger dynamically bent analyzer (Jagodziński et al., 2019
). The thinner support walls also enable working at larger diffraction angles (∼50°) needed for higher diffraction indices [e.g. (777) order in the case of Si and (440) order for quartz]. A common issue observed on dynamic benders is the variation in the bending radius due to non-uniform forces on the crystal surface, particularly on the edges. Thanks to its optimized design combining a compact crystal in a sturdy frame and large opening angles, this effect is not observed in our static bender. Additionally, operating analyzers with fixed radius simplifies the spectrometer alignment and operation, while also reducing costs and risks (the analyzer can break if bent excessively) associated with the dynamic bender mechanism. Fig. 5
shows photographs of our silicon and quartz Laue analyzers using the same design for the bender frame with different curvatures. The Si analyzer has a radius of 150 cm and the quartz one 155 cm.
![]() | Figure 5 Top and side views of the photographs of Laue analyzers. The top panel depicts silicon (111) and the middle panel depicts quartz (110). The aperture size is 80 mm in the bent direction and 30 mm in the non-bent direction. The bottom panel shows a side view of the bender frame. The crystal curvature of ∼150 cm is visible. |
3. The Laue spectrometer
The Laue spectrometer was designed to operate in the horizontal ). However, the backward scattering geometry was chosen during the test at SuperXAS due to lateral space constraints in the experimental hutch. The JUNGFRAU detector was used for the spectrometer commissioning and mounted on the 2θ arm. In this geometry, the free space in the direction leaves the possibility for other equipment to perform parallel measurements, e.g. inelastic X-ray scattering (Ament et al., 2011
), high-resolution Compton scattering (Suortti et al., 1999
; Itou et al., 2001
; Hiraoka et al., 2001
; Hiraoka et al., 2005
), or X-ray diffraction and scattering using the LPD-1M detector (Galler et al., 2019
; Khakhulin et al., 2020
). The latter is particularly advantageous for combining high-energy scattering and spectroscopy measurements simultaneously. The Laue spectrometer can also operate simultaneously with the von Hamos spectrometer, enabling multi-elemental spectroscopic studies on materials containing e.g. 3d and 4d elements, as well as providing a direct efficiency comparison between measurements taken with Laue and Bragg analyzers.
![]() | Figure 6 (a) 3D drawing of the motorized stages for the Laue spectrometer installed at the FXE instrument. The Laue analyzer is mounted at the center of the θ stage and the detector is placed in the 2θ arm. (b) The layout of the Laue spectrometer, SMS and von Hamos (VH) spectrometer used in measurements at FXE is shown in the view along the beam direction. |
3.1. Spectrometer stage
The dispersive direction of our Laue analyzer was chosen to be on the vertical plane to minimize degradation of energy resolution caused by practical operation considerations, e.g. the beam footprint in grazing-incidence experiments and the horizontal beam jitter lead to a large effective beam size, increasing their contribution to the energy resolution. A custom-made set of stages from Huber is used as the spectrometer platform. A drawing of the Laue spectrometer and all its stages and is shown in Fig. 6(a). The θ and 2θ stages are used to rotate the analyzer and detector arm, respectively. The θ motor can operate over a full 360° angular range, and the rotation motors are elevated on top of linear motors to allow a wide working range of −60°/+240°. Three orthogonal linear motors along X, Y and Z directions are used to precisely align the analyzer. A long rail on the 2θ arm with a total length of 440 mm is equipped with a motorized stage with a travel range of ±30 mm. It is used to control the distance from the analyzer to the detector. The complete stage is placed over a long rail sitting on a 1.6 m-long table, allowing large variations in the sample–analyzer distance. Absolute encoders for precise alignment and scan are used on the X, Z, θ and 2θ movements.
4. Commissioning at the synchrotron
The first Si(111) Laue analyzer prototype was tested on a commissioning campaign conducted at the SuperXAS beamline at the Swiss Light Source (SLS) in Switzerland, where a Laue spectrometer with a dynamically bent analyzer is installed (Jagodziński et al., 2019). The measurements aimed at determining the improvements in overall efficiency and the energy resolution, while also investigating the possibility of using the Laue analyzer out of the Rowland circle to explore the dispersive capabilities. For this we have chosen to measure non-resonant Nb Kβ and benchmark our results against those reported previously (Ravel et al., 2018
). Space constraints in the SuperXAS hutch limited the tests of our Laue analyzer to either forward or backward scattering geometries. Considering the influence of strong Compton scattering at high energies on the X-ray emission signals, the backward scattering geometry is the best choice under the given conditions at SuperXAS, where the Compton peak shifts to lower energies at large scattering angles. In these measurements a scattering angle of ∼167° was used. In this setup the θ–2θ rotations are on the horizontal plane, meaning that the dispersion axis of the analyzer must also be on the horizontal plane. A Pilatus 100K-S detector was used to detect the emission signals. The 1 mm-thick silicon sensor improves the detector efficiency at high energies. Further details on the spectrometer stage and detector can be found in Jagodziński et al. (2019
).
Niobium Kβ and valence-to-core (VtC) emission at about 18.62 keV and 18.96 keV, respectively, were chosen for the test measurements. These emission lines can be analyzed using the (333) diffraction of a Laue silicon crystal placed at a of about 18.6° (Kβ1) and 18.2° (VtC). Once the Laue spectrometer was set to these conditions and the analyzer placed on the Rowland circle, the incoming beam was tuned to 18.6 keV and the from an aluminium target was used to align the spectrometer. The analyzer was then moved approximately +16 mm away from the Rowland circle position in order to verify the energy dispersion capabilities. This was done by using a 250 µm-thick Nb foil as sample and setting the incoming energy to 19.1 keV, i.e. above the Nb K edge.
Fig. 7 shows an example emission image from these measurements. Two well separated structures corresponding to the Nb Kβ1 and Kβ3 lines can be clearly identified, suggesting that in this configuration the energy resolution is sufficient to resolve the two peaks which are separated by about 15.5 eV (Elam et al., 2002
). The energy dispersion is mainly along the bent axis, with the energy increasing from right to left in Fig. 7
. Note that the separation between the Kβ1 and Kβ3 emission lines in Fig. 7
is different in the top and bottom regions on the image, indicating that the energy dispersion varies across different non-bent regions of the analyzer. Both stripes are curved, consistent with the prediction from the X-ray tracing simulation. The data in Fig. 7
demonstrate two basic characteristics of the Laue analyzers operating on the off-Rowland circle condition: the energy dispersion along the bent direction and the defocused geometry in the non-bent axis due to crystal surface distortions. These result in rather complex images and special image processing procedures are needed to convert the images into spectra. The simple projection along the dispersive axis and linear energy calibration procedures commonly used in data from von Hamos spectrometers (Hoszowska et al., 1996
) are not adequate to process the detector images of a Laue analyzer in DuMond geometry.
![]() | Figure 7 Representative Pilatus 100K-S detector image of the Nb Kβ1 and Kβ3 In this measurement the incoming X-ray beam was set to 19.1 keV to excite the emission of a 250 µm-thick niobium foil. An integration time of 180 s was used. |
4.1. Energy calibration
Kβ emission and varied in steps of 2 eV. Each data point required about 30 s of integration time to provide sufficient signal levels for performing the energy calibration. Once a complete data set of was collected, different image processing methods were used to generate the spectra. These circumvented the non-linearity issues in the calibration curves and provided a correction for the distorted emission image shapes.
from a monochromatic beam is often used to calibrate the energy axis of spectrometers operating in dispersive mode. This same concept was employed in the measurements using our Laue analyzer at the SuperXAS beamline. Briefly, data at different monochromatic energies are collected using the sample itself as the scattering source. This has the advantage of keeping the spectrometer source fixed and acquiring for each sample measured. The incoming energy was set to the vicinity of the Nb4.1.1. Algorithm 1: calibration mask
One method to perform the spectrometer energy calibration is referred to as the `mask' algorithm. This method, first proposed by Ravel et al. (2018), uses the data set to create a series of image masks for each energy. The masks can then be applied to the data and used to extract emission intensities at each unique energy, which are finally merged to form the Fig. 8
(a) illustrates the basic steps needed to prepare the image mask starting from the data. First, an appropriate intensity threshold is set to distinguish between signal and background in the raw image (top panel), resulting in a cleaner image (middle panel); however, some noisy pixels remain due to the long integration times needed to acquire the elastic data. In a next step, a median filtering is applied to further reduce the residual noise and improve the signal-to-background ratio. This is followed by a Gaussian filtering to further remove the noise and improve the continuity of the signal area. After the threshold subtraction and filtering are applied, each pixel on the detector image corresponding to a given energy is assigned a binary value depending on whether it contains signal (set to 1, or True) or not (set to 0, or False). This Boolean logic 2D array is the so-called `mask' and the final result for a single energy is shown in the bottom panel of Fig. 8
(a). This procedure is applied to all images making up the energy calibration data set, creating a set of masks, one for each energy. Fig. 8
(b) shows the example masks at different energies. The distorted stripe is clearly shown for example in the bottom panel. The behavior of the is consistent with the X-ray tracing simulations and that observed in the emission images: the stripes shift from right to left in the bent direction with increasing incident energy, while their shape and curvature also vary slightly. The complete set of masks is finally used to construct the spectrum by multiplying it by the emission image and summing the intensity of all pixels at a given energy.
![]() | Figure 8 (a) Mask preparation from the raw image of (top panel). A threshold is applied to obtain the middle panel, then median and Gaussian filtering are applied to get the Boolean matrix (bottom panel), i.e. the mask image. (b) Example masks for different energies collected from elastic scattering. |
Fig. 9 depicts the Kβ1 and Kβ3 spectrum of a 250 µm-thick niobium foil converted from the emission image using the mask algorithm (blue dots). This attests to the effectiveness of the mask algorithm in reconstructing spectra from emission images, even for distorted image shapes. Importantly, the energy step in the final spectrum is limited by the sparsity used in the calibration data set. In the measurements presented here, an energy stepping of 2 eV was used, which directly reflects the relatively coarsely sampled spectrum shown in Fig. 9
. Note that due to the better design and surface quality of our Laue analyzers, the emission images are less complex than the `S' shape obtained by Ravel et al. (2018
), enabling more straightforward image processing methods.
![]() | Figure 9 The final Kβ1 and Kβ3 spectrum of a 250 µm niobium foil. The background is removed for better comparison. The comparison between the mask and slicing/re-binning algorithms shows similar results. |
4.1.2. Algorithm 2: slicing and re-binning
An alternative method for calibrating the energy axis is a modified approach of projection along the non-dispersion axis. If the analyzer is properly aligned, the energy dispersion in the bent direction can exhibit monotonic behavior, and each narrow vertical region on the analyzer can be approximated as a straight line. Thus, each of these vertical slices can be projected creating a pixel-to-energy mapping for the different vertical regions. The narrowest slice possible, with a height of only one pixel, is ideal for this procedure. However, using a wider slicing region and summing the intensities inside it was necessary to increase the elastic signal intensity. Slices with a height of ten pixels were used in the energy calibration. An illustration of such slices taken on the emission image corresponding to 18603 eV is shown in Fig. 10(a), together with the associated projection and Gaussian fit to the elastic peak in Fig. 10
(b). A batch of slicing and fitting was applied to all images corresponding to the different energies and the pixel-to-energy correspondences were obtained for every slice. Subsequently, an interpolation in the per pixel step was applied for each slice in order to account for the different dispersion along the non-bent direction. Examples of these pixel-to-energy calibration curves for three different vertical slices are shown in Fig. 10
(c).
![]() | Figure 10 Conceptual steps of the slicing algorithm for energy calibration. Panel (a) shows an example image. The rectangular bars represent the ROIs used in the projection and fitting. A Gaussian fitting is employed to extract the peak position of the pixels (b), ROI 1 is used as an example of projection. In batch fitting, the initial parameters of peak position are prepared by using a peak-finding function from the filtered image. This allows the ROI to be selected as narrowly as possible, despite the pronounced noise level. (c) Representative pixel–energy relation for different ROIs along the non-bent axis. The colors correspond to the ROIs used in panel (a). |
The pixel-to-energy correspondence is clearly non-linear, with notable differences across the different slice regions along the non-bent direction. This results in non-uniform energy steps across the spectrum, which can be corrected by re-binning. After re-binning, the emission spectra from different slices are summed to generate the final
spectrum. In this method, the energy sampling only depends on the detector pixel size and avoids the need to collect a large data set with small energy intervals. The final energy calibration yields a relation of ∼0.5–1.0 eV per pixel across the emission spectrum.Fig. 9 shows the Kβ1 and Kβ3 from a niobium foil using the slicing algorithm (solid line). This spectrum was obtained from the same emission image used in the mask algorithm (dots). The spectra obtained using both algorithms are comparable, demonstrating the robustness of both methods of image processing. Their good agreement suggests that the emission energy in each image could also be calibrated using a set of referenced spectra, thus avoiding the need to collect monochromatic data.
4.2. Energy resolution and efficiency at SuperXAS
The Laue spectrometer energy resolution was evaluated using the Kβ1,3 emission line. The approaches resulted in comparable results. A representative measurement of the at 18625 eV is shown in Fig. 11(a) together with a fit using a Gaussian profile. The fitted FWHM of this Gaussian curve was 7.2 eV. Considering the intrinsic energy resolution of Si(111) monochromators at this energy (about 1.4 × 10−4) is on the order of 2.6 eV, a deconvolved spectrometer resolution of 6.7 eV is obtained. This value is slightly worse than the 5.18 eV obtained in the work of Jagodzinski et al., likely due to the longer analyzer radius (2 m) used in their setup (Jagodziński et al., 2019
). Fig. 11
(b) presents the Kβ1,3 of a Nb foil and the fitted pseudo-Voigt functions used in the deconvolution. The instrument response is described by the Gaussian component and the Lorentzian one accounts for the 6.2 eV of the natural linewidth of the Nb Kβ1 emission (Campbell & Papp, 2001
). Considering these, the final spectrometer resolution is about 6.9 eV. The slight difference of about 3% in the resolution obtained using the two evaluation methods is attributed to inaccuracies in the monochromatic bandwidth and the theoretical Nb Kβ natural linewidth. This indicates that collecting a complete data set is not strictly necessary to evaluate the spectrometer energy resolution. The energy resolution was limited mainly by contributions due to the horizontal beam size (∼140 µm, already taking into account the projection footprint on the sample at ∼45°) used at SuperXAS. The measured energy resolution is comparable with the calculated value using X-ray tracing as shown in Fig. 4
(c). This excludes significant effects of analyzer surface distortion, image processing and detector pixel size on the Laue spectrometer energy resolution. A better energy resolution could be expected at FXE and other beamlines when using smaller beam sizes (∼20 µm). In this limit, using detectors with a smaller pixel size can also result in improved energy resolution.
![]() | Figure 11 Evaluation of energy resolution for the Laue analyzer using experimental data from the mask algorithm. (a) The of Nb foil, the incident energy is at 18625 eV, the intrinsic energy resolution of the silicon (111) monochromator is 2.6 eV. The resolution of the Laue spectrometer is obtained by deconvolving the two Gaussian functions. (b) The Kβ1,3 emission spectra of the Nb foil, the resolution of the Laue spectrometer is obtained by deconvolving the Lorentzian and Gaussian functions. |
We have used the Laue spectrometer to measure the Kβ1,3 of a series of solid niobium compounds with varied and ligand environment. The spectra of these reference compounds are shown in Fig. 12. The dispersed image of the Nb Kβ1,3 emission was clearly visible on the detector in less than 100 s (∼1011 photons s−1 incoming at SuperXAS), contrasting with around 1 h in total (45 s per point in scanning mode) used to measure the Ru Kβ1,3 at the same beamline with the spectrometer described by Jagodzinski et al. We have also measured the weaker VtC emission of the same Nb reference [see Fig. 12
(b)]. Each measurement detecting the Kβ2, Kβ′′ and Kβ4 lines could be completed in ∼30 min, while similar data collection on ruthenium oxides took 12 min per data point, resulting in about 10 h of total measurement. As shown in Table 1
, the separations between the peaks of Kβ2 and VtC features of Nb2O5 and NbF5 samples agree well with the values reported by Ravel et al. (2018
). Importantly, the subtle Kβ′′ and Kβ4 on the Ru references were not clearly visible in the measurements using the previous Laue spectrometer from SuperXAS.
|
![]() | Figure 12 XES spectra of different niobium compounds with different electronic states: NbIIICl3, NbIVO2, Nb |
In terms of efficiency, our spectrometer outperforms the previous implementation at SuperXAS using a dynamically bent Laue analyzer. The analyzer optimization, in particular the choice of thinner crystals and an asymmetric cut, combined with a new design of the holding frame resulted in an increased spectrometer efficiency without a significant impact on the energy resolution. Moreover, operation in dispersive mode was successfully demonstrated. These results are encouraging, when considering the implementation of high-resolution ultrafast pump–probe spectroscopies at high X-ray photon energies at XFELs using Laue spectrometers.
5. HELIOS implementation at FXE
After the tests at a synchrotron beamline, the Laue spectrometer was fully assembled, installed and commissioned at the FXE instrument. Two Laue analyzers, one silicon and one quartz, were tested profiting from the strong pulse intensities at high photon energies provided by the European XFEL operation at 16.3 GeV. The Laue spectrometer as discussed in Section 3 was designed to be docked into the SMS [see Fig. 6
(b)]. This allows the scattering angle to be varied in future experiments exploiting momentum-transfer dependence. In the measurements discussed here the spectrometer was placed perpendicular to the beam propagation direction where the elastic and Compton scattering intensities are weakest. This geometry allowed simultaneous operation of the Laue and von Hamos spectrometers at FXE as shown in Fig. 6
(b), providing an opportunity for a direct comparison of signal levels, efficiency and resolution of Laue and Bragg analyzers within a single measurement.
Briefly, the Kα1 and Kβ1,3 spectra of a metallic foil and different oxides (NbO2 and Nb2O5) were analyzed using the third-order reflection of Si(111) and SiO2(110) Laue crystals. All samples were mounted in the center of the SMS at a distance of about 1.4 m from the Laue analyzer. The for measuring Nb Kα1 was 20.9° and 27.1° when using Si(333) and SiO2(330), respectively, and for measuring Kβ1,3 it was 18.6° and 24.0° when using Si(333) and SiO2(330), respectively. Initially, the strong Nb Kα1 emission line was used to calibrate the spectrometer motor stages and optimize the signal levels. The Laue spectrometer was set for dispersive operation with the analyzer placed around −8 mm out of the Rowland circle. Possible damage to the solid samples induced by the intense focused XFEL beam was avoided by attenuating the incoming beam to ∼40% total transmission during Kα data collection and ∼20% for Kβ. The combined beamline transmission including the focusing lenses, attenuators, beam imaging screens and windows was estimated to be 10%–20%. A total of 10–20 pulses per train at a 282 kHz intra-train repetition rate (100–200 pulses per second) were used during the measurements, resulting in an incoming comparable with that of SuperXAS. In regular operation conditions up to around 200 pulses per train at 0.5 MHz (equivalent to 2000 pulses per second) can be used at FXE. A total of eight Si(111) cylindrical analyzers with 0.5 m radius were installed in the von Hamos spectrometer and placed at about 73° to analyze the Nb Kα emission. Interestingly, the Bragg angles for Nb Kα1 (16.61 keV, 72.16°), Kα2 (16.52 keV, 73.21°) and Kβ (18.62 keV, 72.84°) when using Si(111) crystals are very similar, with the eighth diffraction order used for the former and ninth used for the latter. One cylindrical crystal used in the von Hamos spectrometer has a size of 11 cm × 3 cm, which is comparable with the Laue crystal. The same detector type, namely the JUNGFRAU, was used on the von Hamos (JF-1M) and Laue (JF-500 K) spectrometers. Benefiting from the modest energy resolution of the JF detector (1–2 keV), a lower signal threshold was set to remove the electronic noise (<2.5 ADU) and secondary scattering (e.g. background fluorescence from the lead shielding around the detector, ∼13 ADU) during the image processing, which helped to improve the signal-to-background ratio.
measurements at FXE using the Laue spectrometer were done using a SASE (self-amplified spontaneous emission) beam of 19.2 keV photon energy with a pulse intensity of ∼550 µJ, as measured by an X-ray gas monitor detector installed in the XFEL tunnel. The X-ray beam was focused down to ∼20 µm × 20 µm using beryllium lenses. Niobium5.1. Energy calibration at XFELs: self-calibration method
The use of Kα implies a change of more than ∼2.4 keV. Such large energy changes on XFEL undulators operating in SASE mode are not straightforward. Additionally, the use of a monochromator can limit the total number of X-ray pulses in each burst train at the European XFEL, translating into a lower overall output. Combined, these factors make the collection of data sets at XFELs significantly more cumbersome and time-consuming when compared with synchrotrons.
to calibrate the energy axis on measurements using the Laue analyzer at XFELs is not convenient because of the following limitations: (i) the Laue spectrometer is installed perpendicular to the beam propagation direction where is weakest, and (ii) measurements require the monochromator to be tuned to an energy close to that of the analyzed fluorescence line, which in the case of NbAn alternative approach to the energy calibration is proposed here based on the basic concept of dispersive optics. The dispersion property of Laue analyzers arises from the fact that different energies originating at the source have different incident angular offsets along the analyzer bent axis. This is illustrated in the scheme of Fig. 3(b). This behavior is equivalent to changing the angular offset by rotating the analyzer while keeping the detector position fixed when a single photon energy is emitted from the source. Fig. 13
illustrates this effect in the measurement of Nb Kα1 emission collected at different angles. The signal, in the form of a vertical stripe, moves from right to left along the bent axis with decreasing incident angle on the analyzer, exhibiting a behavior similar to that observed during an energy scan in [Fig. 8
(b)]. Similar to the slicing and re-binning algorithm discussed in Section 4
, a 2D calibration map using angles instead of energies can be prepared and used for the energy calibration. For any given emission feature, e.g. the single peak of Kα1 or double peaks of Kβ1,3, the pixel coordinate of each peak position can be extracted by slicing the image along the non-bent direction and batch fitting. The emission image is then re-binned into a spectrum and the angles can be converted into energies using the Bragg equation and compared with a reference This method is referred to as self-calibration.
![]() | Figure 13 Example emission stripes collected at different incident angles. The stripe moves from the right side to the left side as the incident angle decreases. The signal is from a metallic Nb foil. |
Fig. 14 illustrates the steps necessary to convert the emission images to a spectrum using the self-calibration method. The data were collected using a 250 µm-thick Nb metallic foil as sample. Slices with a height of 30 pixels along the non-bent axis were sufficient to fulfill the assumption that within that small ROI, the emission stripe can be regarded as a straight line while providing sufficient intensity for an accurate determination of the peak positions. A similar batch fitting as used for the slicing algorithm (illustrated in Fig. 10
) is then applied to extract the angle-to-pixel calibration relation for all slices within a given emission image. As expected, a non-linear and non-uniform relation between angle and pixel is also observed. The results of these procedures are summarized in the top and middle panels of Fig. 14
. Finally, all spectra are re-binned and summed to form the as a function of angle. The last step is to identify the peak angular position and convert it into energy using the Bragg equation and the reference energy value for Nb Kα1 (16615 eV). The inter-planar d-spacing values from the silicon (111) and quartz (110) analyzers were taken from the XOP package (Sanchez del Rio & Dejus, 2011
). Similar values were reported by the analyzer manufacturer (Saint-Gobain). The energy step per pixel varied from 0.2 eV to 0.5 eV, depending on the different dispersive regions along the non-bent direction. One additional effect to consider is the non-uniform background. Specifically, the low-energy region (larger angles) has a larger background than the high-energy one. This is due to the geometry used: photons scattered by the sample and transmitted through the Laue analyzer and its supporting frame contribute more to the background for the higher angular region than for the lower angular region. This effect could be minimized by the use of a trapezoidal shaped flight path between the sample and Laue analyzer [see Fig. 6
(b)] and additional shielding around the detector and along the beam path before and after the sample.
![]() | Figure 14 The basic concept to convert the emission image into a spectrum by using the self-calibration method. Top panel is the sliced spectra after pixel calibration based on the angular scan showing good overlap of the peak position. Middle panel is the pixel–angle calibration at different slice regions; non-linear and non-uniform relations are shown. Bottom panel is the emission spectrum of Nb Kα1 after re-binning and transforming angles to energies. The signal is from Nb foil sample. |
The accuracy of the self-calibration method was evaluated by comparing the Nb Kβ1,3 collected at FXE with that collected at SuperXAS, which was calibrated using The result of this comparison is shown in Fig. 15. In the FXE data, the energy is converted from angle using only one energy position of Kβ1 peak position. The absolute energy of the Kβ1 and Kβ3 peaks and their energy separation are similar in both data sets. A slight difference of around 3% in the energy separation was observed, likely due to inaccuracies in the elastic scan calibration method and worse energy resolution in the synchrotron data set. Importantly, this difference is smaller than the Si(111) monochromator bandwidth used in the elastic scan calibration. The data can be further re-scaled and corrected by using additional energy positions on the reference spectrum. Moreover, there is a large difference in the due to the better energy resolution of the FXE data.
![]() | Figure 15 Comparison of the Kβ1,3 of Nb foil collected at SuperXAS and FXE. The background for each spectrum was subtracted for better comparison. The SuperXAS spectrum is calibrated by the scan and the FXE data are calibrated by the angular scan in the self-calibration method. |
5.2. Energy resolution and efficiency at FXE
The comparison between the two Nb Kβ spectra using the same Si(111) analyzer shown in Fig. 15 shows the improved energy resolution obtained in the measurements at FXE. The main factors contributing to a better energy resolution are the smaller beam size of 20 µm (compared with 100 µm at SuperXAS) and the smaller detector pixel size (75 µm for the JUNGFRAU at FXE versus 172 µm for the Pilatus at SuperXAS), combined with setting the dispersion on the vertical axis (implying the beam footprint on the sample at 45° does not affect the beam size in the dispersion direction).
Given the absence of Kβ using the same procedure employed in the SuperXAS data. The Nb Kβ spectrum measured using the Si(111) Laue analyzer in the third diffraction order was fitted using two pseudo-Voigt functions describing the Kβ1 and Kβ3 peaks and the results are shown in Fig. 16. The instrument response is deconvoluted from the fit considering a Lorentzian function with a natural width of either 6.2 eV according to the calculated value reported by Campbell & Papp (2001
) or 6.4 eV obtained from the measurement in Fig. 11
. These resulted in a spectrometer resolution of 2.4 eV or 2.1 eV, depending on the value used for the Nb Kβ linewidth. That corresponds to a ΔE/E ≃ 1.1 × 10−4 to 1.3 × 10−4, which is approximately three times better than the value obtained at SuperXAS.
![]() | Figure 16 Evaluation of energy resolution of the Laue spectrometer using a Si(333) analyzer. The Kβ1,3 is collected from the Nb foil. The resolution of the Laue spectrometer is obtained by deconvolving the Lorentzian and Gaussian functions. A resolution of 2.4 eV is obtained by using the Lorentzian function to account for the natural linewidth of 6.2 eV, as reported by Campbell & Papp (2001 ![]() |
Fig. 17 presents niobium comparing different performance aspects of these measurements: in panel (a) a comparison of the Kα1 emission collected using the von Hamos and Laue analyzers is shown, in panel (b) the Kβ emission measured using von Hamos and Laue analyzers is compared, and in panel (c) the Kβ emission measured using silicon and quartz Laue crystals is compared. Using the (330) reflection of a quartz Laue analyzer with similar radius for measuring the same Nb Kβ resulted in a spectrum with slightly better resolution compared with the data collected using the silicon analyzer. The comparison of the two data sets is shown in Fig. 17
(c). Further improvements in energy resolution can be achieved by using higher diffraction orders; however, the narrower Darwin width associated with high diffraction orders leads to lower integrated signal efficiency. An improvement factor of around 1.7 is predicted from the X-ray tracing when using Si(555) instead of Si(333), while giving about a factor 1.9 lower efficiency.
![]() | Figure 17 Performance comparisons between the von Hamos spectrometer with eight Si(111) Bragg crystals and the single Si(333) Laue crystal and between the Si(333) and SiO2(330) analyzers. (a) Kα1, and (b) and (c) Kβ1,3 spectra from a Nb foil. The background in each data set was kept here to provide a realistic performance comparison, and the background-subtracted spectra are shown in the inset of (c). To carefully compare the efficiency at the Kβ energy, a larger beam size (∼40 µm), i.e. worse energy resolution for the Laue analyzer, was used to prevent sample damage, as acquiring the Kβ signal takes longer compared with the Kα signal. The inset of (b) shows the full spectra measured using the von Hamos spectrometer, where Kα and Kβ lines are collected by (888) and (999) reflections, respectively. |
As stated before, the von Hamos spectrometer was equipped with eight Si(111) cylindrical analyzers, each with a similar size to the Laue analyzers. The emission signals collected with either spectrometer were normalized to the incoming Kα energy (∼16.6 keV), and comparable with eight von Hamos analyzers at the Kβ energy (∼18.6 keV). It can be observed that the efficiency of the Bragg crystal gradually decreases with increasing energy with respect to that of the Laue crystal. It is important to note the measurements with both spectrometers shown in Figs. 17(a) and 17
(b) were collected simultaneously. Comparing the spectra measured with the two different Laue analyzers indicated a ∼2.7 times better efficiency when using the quartz analyzer instead of the silicon, which agrees with the rocking-curve calculations shown in Section 2.2
. The Laue analyzer manufacturing technology is progressing in the direction of reducing the bending radius to below 1 m, which should result in an additional twofold increase in efficiency. The results shown here demonstrate the superior efficiency of the Laue analyzer with respect to Bragg crystals at energies >15 keV. At even higher photon energies, e.g. when measuring Kα or Kβ emission lines of Pd or Ag (∼21.1–24.9 keV), the improved efficiency of Laue analyzers will become even more pronounced.
Regarding energy resolution, using the Laue analyzers resulted in spectra with resolution comparable with that measured with the von Hamos spectrometer. The pixel-to-energy positions in the von Hamos data were calibrated using two peak positions of either Kα1, 2 lines [Fig. 17(a)] or Kβ1,3 lines [Fig. 17
(b)] assuming a The energy resolution in the von Hamos data is worse than expected without considering the much larger dispersion window and other effects due to the larger penetration of the harder X-rays into the analyzer crystals before being diffracted, e.g. strain in the crystal planes, deformations in the crystal–substrate interface etc. The spectra measured with the Laue quartz analyzer have slightly better energy resolution than those using the Laue Si one, being 2.2 eV and 2.4 eV, respectively. This difference is likely due to the different Bragg angles needed for the same measurement using different analyzers – for SiO2(330) a larger angle is needed than for Si(333). At the Nb Kβ1 energy the angles are 24.0° and 18.6°, respectively. The comparisons between the von Hamos and Laue spectrometers are summarized in Table 2
.
|
6. Conclusions and future perspectives
We have presented the design and first commissioning results of a new X-ray spectrometer for high X-ray photon energies using Laue analyzers. The High-Energy Laue X-ray emIssiOn Spectrometer installed at the FXE instrument at the European XFEL was optimized for improved efficiency and ease of operation, while maintaining a total energy resolution below the core-hole lifetime of 4d elements. This spectrometer significantly increases the X-ray spectroscopic capabilities of FXE by enabling unique high-resolution experiments at high photon energies. To the best of our knowledge, this is the only spectrometer of this kind at any XFEL facility worldwide.
Two Laue analyzers, a silicon (333) and a quartz (330), with bending radius of ∼1.5 m were used in the measurements. A silicon analyzer was initially tested in dispersive mode at the SuperXAS beamline of the Swiss Light Source, providing an energy resolution of 3.6 × 10−4. This limitation was mainly due to the relatively large beam and detector pixel sizes at SuperXAS. An improved energy resolution of ∼1.2 × 10−4 was obtained in the measurements at FXE using a quartz analyzer. The smaller beam (20 µm) and detector pixel sizes (75 µm) were the main factors contributing to this improvement. The energy dispersion capabilities of our Laue spectrometer were shown to be a convenient and effective way of measuring spectra at high photon energies. Our Laue spectrometer has demonstrated the ability to resolve the K emission spectra of 4d metals at high X-ray energies, detecting an energy interval of ∼100 eV, sufficient for covering a large portion of the Kβ emission at once.
Several image processing methods have been developed to convert the emission images to spectra. Distortions in the emission images were observed due to the nature of Laue crystal surfaces and the defocused geometry. Despite this, effective algorithms for data reduction have been developed for reconstructing the spectra. Using either
data sets or analyzer angular scans allows for accurate corrections and conversions of the emission images to spectra in a robust manner. In particular, the self-calibration method using angular scans can also be used to correctly calibrate the spectrometer, thereby circumventing the need for measurements and addressing operational inconveniences of XFELs operating in SASE mode.The Laue spectrometer reported here has demonstrated a significant improvement in efficiency when compared with the (Bragg) von Hamos spectrometer at high photon energies. This was shown by simultaneous measurements using both spectrometers, which indicated an increase of a factor of ∼4–22 in the signal strength measured with the Laue spectrometer, depending on the analyzer material, emission line and number of analyzers used. In fact, similar conclusions regarding the higher efficiency of Laue analyzers at high photon energies have been recently reported by Hiraoka (2025). The first Nb Kβ results collected at SuperXAS also indicate a significant efficiency improvement compared with their previous spectrometer equipped with a dynamic bent analyzer with larger radius and operated in scanning mode. Clear features in VtC and Kβ regions of the spectra could be measured in about half an hour and just several minutes, respectively. Moreover, the HELIOS spectrometer is equipped with a thin analyzer frame and a large aperture, which combined with motorization of all analyzer and detector offer the ability to choose different diffraction planes without shadowing the detector. This allows the measurements to be tuned for a compromise between energy resolution and efficiency, depending on different measurement scenarios. For instance, valence-to-core emission spectra can be measured using modest resolution with lower diffraction orders for higher efficiency, while Kβ lines, X-ray Raman scattering and other emission lines requiring higher energy resolution can benefit from a better resolution scheme. By utilizing the high X-ray energy capabilities and the improved energy resolution and efficiency of the Laue spectrometer, FXE will enable several novel spectroscopic capabilities, e.g. non-resonant and resonant and HERFD-XAS at high photon energies. These advancements will inspire new ultrafast pump–probe spectroscopic studies in the 15–20 keV range, covering the K lines of 4d elements and L lines of 5f elements.
Footnotes
‡Present address: Attosecond and Ultrafast Optics Group, ICFO – The Institute of Photonics Sciences, Barcelona, Spain.
Acknowledgements
The authors acknowledge the European X-ray Free-Electron Laser GmbH in Germany for providing access to XFEL beamtime at the FXE scientific instrument (proposal No. 4936) and the support from all groups at European XFEL. Data recorded for the experiment at the European XFEL are available at https://doi.org/10.22003/XFEL.EU-DATA-004936-00. We thank in particular the accelerator and undulator groups for delivering the high-photon-energy SASE beam to the FXE instrument. We acknowledge the Swiss Light Source in Switzerland for providing access to beamtime at SuperXAS beamline (proposal No. 20211827). We acknowledge Dr P. Jagodzinski, Dr F. Otte, Dr X. Pan and Professor K. Yang for the fruitful discussions about X-ray tracing simulations. We thank Dr B. Ravel for the discussions about the calibration mask algorithm. We thank Saint-Gobain S. A. (now Luxium Solutions) for the discussions, developments and manufacture of the Laue crystal analyzers used in this work. This work is supported by the research and development project at European XFEL (R&D project `HELIOS: High-Energy Laue X-ray emIssiOn Spectrometer' at FXE). Open access funding enabled and organized by Projekt DEAL.
References
Alonso-Mori, R., Kern, J., Sokaras, D., Weng, T.-C., Nordlund, D., Tran, R., Montanez, P., Delor, J., Yachandra, V. K., Yano, J. & Bergmann, U. (2012). Rev. Sci. Instrum. 83, 073114. Web of Science PubMed Google Scholar
Ament, L. J. P., van Veenendaal, M., Devereaux, T. P., Hill, J. P. & van den Brink, J. (2011). Rev. Mod. Phys. 83, 705–767. Web of Science CrossRef CAS Google Scholar
Bacellar, C., Kinschel, D., Mancini, G. F., Ingle, R. A., Rouxel, J., Cannelli, O., Cirelli, C., Knopp, G., Szlachetko, J., Lima, F. A., Menzi, S., Pamfilidis, G., Kubicek, K., Khakhulin, D., Gawelda, W., Rodriguez-Fernandez, A., Biednov, M., Bressler, C., Arrell, C. A., Johnson, P. J. M., Milne, C. J. & Chergui, M. (2020). Proc. Natl Acad. Sci. USA, 117, 21914–21920. CrossRef CAS PubMed Google Scholar
Bacellar, C., Rouxel, J. R., Ingle, R. A., Mancini, G. F., Kinschel, D., Cannelli, O., Zhao, Y., Cirelli, C., Knopp, G., Szlachetko, J., Lima, F. A., Menzi, S., Ozerov, D., Pamfilidis, G., Kubicek, K., Khakhulin, D., Gawelda, W., Rodriguez-Fernandez, A., Biednov, M., Bressler, C., Arrell, C. A., Johnson, P. J. M., Milne, C. J. & Chergui, M. (2023). J. Phys. Chem. Lett. 14, 2425–2432. Web of Science CrossRef CAS PubMed Google Scholar
Bergmann, U., Kern, J., Schoenlein, R. W., Wernet, P., Yachandra, V. K. & Yano, J. (2021). Nat. Rev. Phys. 3, 264–282. Web of Science CrossRef CAS PubMed Google Scholar
Bjornsson, R., Lima, F. A., Spatzal, T., Weyhermüller, T., Glatzel, P., Bill, E., Einsle, O., Neese, F. & DeBeer, S. (2014). Chem. Sci. 5, 3096–3103. Web of Science CrossRef CAS Google Scholar
Borchert, G., Scheck, W. & Schult, O. (1975). Nucl. Instrum. Methods, 124, 107–117. CrossRef CAS Google Scholar
Campbell, J. L. & Papp, T. (2001). At. Data Nucl. Data Tables, 77, 1–56. Web of Science CrossRef CAS Google Scholar
Canton, S. E., Biednov, M., Pápai, M., Lima, F. A., Choi, T.-K., Otte, F., Jiang, Y., Frankenberger, P., Knoll, M., Zalden, P., Gawelda, W., Rahaman, A., Møller, K. B., Milne, C., Gosztola, D. J., Zheng, K., Retegan, M. & Khakhulin, D. (2023). Adv. Sci. 10, 2206880. Web of Science CrossRef Google Scholar
Canton, S. E., Kjaer, K. S., Vankó, G., van Driel, T. B., Adachi, S., Bordage, A., Bressler, C., Chabera, P., Christensen, M., Dohn, A. O., Galler, A., Gawelda, W., Gosztola, D., Haldrup, K., Harlang, T., Liu, Y., Møller, K. B., Németh, Z., Nozawa, S., Pápai, M., Sato, T., Sato, T., Suarez-Alcantara, K., Togashi, T., Tono, K., Uhlig, J., Vithanage, D. A., Wärnmark, K., Yabashi, M., Zhang, J., Sundström, V. & Nielsen, M. M. (2015). Nat. Commun. 6, 6359. Web of Science CrossRef PubMed Google Scholar
Cao, G. & Delong, L. E. (2013). Editors. Frontiers of 4d- and 5d-Transition Metal Oxides. World Scientific. Google Scholar
Castillo, R. G., Hahn, A. W., Van Kuiken, B. E., Henthorn, J. T., McGale, J. & DeBeer, S. (2021). Angew. Chem. Int. Ed. 60, 10112–10121. Web of Science CSD CrossRef CAS Google Scholar
Castillo, R. G., Henthorn, J. T., McGale, J., Maganas, D. & DeBeer, S. (2020). Angew. Chem. Int. Ed. 59, 12965–12975. Web of Science CrossRef CAS Google Scholar
Cauchois, Y. (1932). J. Phys. Radium, 3, 320–336. CrossRef CAS Google Scholar
Chen, Y., Brinker, F., Decking, W., Scholz, M. & Winkelmann, L. (2021). Appl. Sci. 11, 10768. Web of Science CrossRef Google Scholar
Cutsail, G. E. III & DeBeer, S. (2022). ACS Catal. 12, 5864–5886. Google Scholar
Doonan, C. J., Zhang, L., Young, C. G., George, S. J., Deb, A., Bergmann, U., George, G. N. & Cramer, S. P. (2005). Inorg. Chem. 44, 2579–2581. CrossRef PubMed CAS Google Scholar
Ducrot, A. B., Coulson, B. A., Perutz, R. N. & Duhme-Klair, A.-K. (2016). Inorg. Chem. 55, 12583–12594. CrossRef CAS PubMed Google Scholar
DuMond, J. W. M. (1947). Rev. Sci. Instrum. 18, 626–638. CrossRef CAS PubMed Web of Science Google Scholar
Dunne, M., Schoenlein, R. W., Cryan, J. P. & Wolf, T. J. A. (2023). Structural Dynamics with X-ray and Electron Scattering, ch. 8, pp. 301–343. Royal Society of Chemistry. Google Scholar
Elam, W., Ravel, B. & Sieber, J. (2002). Radiat. Phys. Chem. 63, 121–128. Web of Science CrossRef CAS Google Scholar
Galler, A., Gawelda, W., Biednov, M., Bomer, C., Britz, A., Brockhauser, S., Choi, T.-K., Diez, M., Frankenberger, P., French, M., Görries, D., Hart, M., Hauf, S., Khakhulin, D., Knoll, M., Korsch, T., Kubicek, K., Kuster, M., Lang, P., Alves Lima, F., Otte, F., Schulz, S., Zalden, P. & Bressler, C. (2019). J. Synchrotron Rad. 26, 1432–1447. Web of Science CrossRef CAS IUCr Journals Google Scholar
Gawelda, W., Szlachetko, J. & Milne, C. J. (2016). X-ray Spectroscopy at Free Electron Lasers, ch. 22, pp. 637–669. John Wiley and Sons. Google Scholar
Groot, F. M. F. de, Haverkort, M. W., Elnaggar, H., Juhin, A., Zhou, K.-J. & Glatzel, P. (2024). Nat. Rev. Methods Primers, 4, 45. Google Scholar
Hämäläinen, K., Siddons, D. P., Hastings, J. B. & Berman, L. E. (1991). Phys. Rev. Lett. 67, 2850–2853. CrossRef PubMed CAS Web of Science Google Scholar
Han, H., Jacquet, Q., Jiang, Z., Sayed, F. N., Jeon, J.-C., Sharma, A., Schankler, A. M., Kakekhani, A., Meyerheim, H. L., Park, J., Nam, S. Y., Griffith, K. J., Simonelli, L., Rappe, A. M., Grey, C. P. & Parkin, S. S. P. (2023). Nat. Mater. 22, 1128–1135. CrossRef CAS PubMed Google Scholar
Hiraoka, N. (2025). J. Synchrotron Rad. 32, 109–117. CrossRef CAS IUCr Journals Google Scholar
Hiraoka, N., Buslaps, T., Honkimäki, V. & Suortti, P. (2005). J. Synchrotron Rad. 12, 670–674. Web of Science CrossRef CAS IUCr Journals Google Scholar
Hiraoka, N., Fukui, H., Tanida, H., Toyokawa, H., Cai, Y. Q. & Tsuei, K. D. (2013). J. Synchrotron Rad. 20, 266–271. Web of Science CrossRef CAS IUCr Journals Google Scholar
Hiraoka, N., Itou, M., Ohata, T., Mizumaki, M., Sakurai, Y. & Sakai, N. (2001). J. Synchrotron Rad. 8, 26–32. Web of Science CrossRef CAS IUCr Journals Google Scholar
Hoszowska, J., Dousse, J.-C., Kern, J. & Rhême, C. (1996). Nucl. Instrum. Methods Phys. Res. A, 376, 129–138. CrossRef CAS Web of Science Google Scholar
Huijser, A., Pan, Q., van Duinen, D., Laursen, M. G., El Nahhas, A., Chabera, P., Freitag, L., González, L., Kong, Q., Zhang, X., Haldrup, K., Browne, W. R., Smolentsev, G. & Uhlig, J. (2018). J. Phys. Chem. A, 122, 6396–6406. CrossRef CAS PubMed Google Scholar
Hunault, M. O. J. Y., Khan, W., Minár, J., Kroll, T., Sokaras, D., Zimmermann, P., Delgado-Jaime, M. U. & de Groot, F. M. F. (2017). Inorg. Chem. 56, 10882–10892. CrossRef CAS PubMed Google Scholar
Huotari, S., Sahle, C. J., Henriquet, C., Al-Zein, A., Martel, K., Simonelli, L., Verbeni, R., Gonzalez, H., Lagier, M.-C., Ponchut, C., Moretti Sala, M., Krisch, M. & Monaco, G. (2017). J. Synchrotron Rad. 24, 521–530. Web of Science CrossRef CAS IUCr Journals Google Scholar
Huotari, S., Vankó, Gy., Albergamo, F., Ponchut, C., Graafsma, H., Henriquet, C., Verbeni, R. & Monaco, G. (2005). J. Synchrotron Rad. 12, 467–472. Web of Science CrossRef CAS IUCr Journals Google Scholar
Itou, M., Hiraoka, N., Ohata, T., Mizumaki, M., Deb, A., Sakurai, Y. & Sakai, N. (2001). Nucl. Instrum. Methods Phys. Res. A, 467–468, 1109–1112. Web of Science CrossRef CAS Google Scholar
Jagodziński, P., Szlachetko, J., Dousse, J. C., Hoszowska, J., Szlachetko, M., Vogelsang, U., Banaś, D., Pakendorf, T., Meents, A., van Bokhoven, J. A., Kubala-Kukuś, A., Pajek, M. & Nachtegaal, M. (2019). Rev. Sci. Instrum. 90, 063106. Web of Science PubMed Google Scholar
Katayama, T., Northey, T., Gawelda, W., Milne, C. J., Vankó, G., Lima, F. A., Bohinc, R., Németh, Z., Nozawa, S., Sato, T., Khakhulin, D., Szlachetko, J., Togashi, T., Owada, S., Adachi, S., Bressler, C., Yabashi, M. & Penfold, T. J. (2019). Nat. Commun. 10, 3606. Web of Science CrossRef PubMed Google Scholar
Khakhulin, D., Otte, F., Biednov, M., Bömer, C., Choi, T., Diez, M., Galler, A., Jiang, Y., Kubicek, K., Lima, F. A., Rodriguez-Fernandez, A., Zalden, P., Gawelda, W. & Bressler, C. (2020). Appl. Sci. 10, 995. Web of Science CrossRef Google Scholar
Kinschel, D., Bacellar, C., Cannelli, O., Sorokin, B., Katayama, T., Mancini, G. F., Rouxel, J. R., Obara, Y., Nishitani, J., Ito, H., Ito, T., Kurahashi, N., Higashimura, C., Kudo, S., Keane, T., Lima, F. A., Gawelda, W., Zalden, P., Schulz, S., Budarz, J. M., Khakhulin, D., Galler, A., Bressler, C., Milne, C. J., Penfold, T., Yabashi, M., Suzuki, T., Misawa, K. & Chergui, M. (2020). Nat. Commun. 11, 4145. Web of Science CrossRef PubMed Google Scholar
Kropf, A. J., Finch, R. J., Fortner, J. A., Aase, S., Karanfil, C., Segre, C. U., Terry, J., Bunker, G. & Chapman, L. D. (2003). Rev. Sci. Instrum. 74, 4696–4702. Web of Science CrossRef CAS Google Scholar
Kropf, A. J., Fortner, J. A., Finch, R. J., Cunnane, J. C. & Karanfil, C. (2005). Phys. Scr. T115, 998. CrossRef Google Scholar
Kunnus, K., Vacher, M., Harlang, T. C. B., Kjaer, K. S., Haldrup, K., Biasin, E., van Driel, T. B., Pápai, M., Chabera, P., Liu, Y., Tatsuno, H., Timm, C., Källman, E., Delcey, M., Hartsock, R. W., Reinhard, M. E., Koroidov, S., Laursen, M. G., Hansen, F. B., Vester, P., Christensen, M., Sandberg, L., Németh, Z., Szemes, D. S., Bajnóczi, É., Alonso-Mori, R., Glownia, J. M., Nelson, S., Sikorski, M., Sokaras, D., Lemke, H. T., Canton, S. E., Møller, K. B., Nielsen, M. M., Vankó, G., Wärnmark, K., Sundström, V., Persson, P., Lundberg, M., Uhlig, J. & Gaffney, K. J. (2020). Nat. Commun. 11, 634. Web of Science CrossRef PubMed Google Scholar
Lancaster, K. M., Roemelt, M., Ettenhuber, P., Hu, Y., Ribbe, M. W., Neese, F., Bergmann, U. & DeBeer, S. (2011). Science, 334, 974–977. Web of Science CrossRef CAS PubMed Google Scholar
Lemke, H. T., Kjaer, K. S., Hartsock, R., van Driel, T. B., Chollet, M., Glownia, J. M., Song, S., Zhu, D., Pace, E., Matar, S. F., Nielsen, M. M., Benfatto, M., Gaffney, K. J., Collet, E. & Cammarata, M. (2017). Nat. Commun. 8, 15342. Web of Science CrossRef PubMed Google Scholar
Lethbridge, Z. A., Walton, R. I., Marmier, A. S., Smith, C. W. & Evans, K. E. (2010). Acta Mater. 58, 6444–6451. CrossRef CAS Google Scholar
Levantino, M., Lemke, H. T., Schirò, G., Glownia, M., Cupane, A. & Cammarata, M. (2015). Struct. Dyn. 2, 041713. CrossRef PubMed Google Scholar
Lezcano–González, I., Oord, R., Rovezzi, M., Glatzel, P., Botchway, S. W., Weckhuysen, B. M. & Beale, A. M. (2016). Angew. Chem. Int. Ed. 55, 5215–5219. Google Scholar
Lima, F. A., Bjornsson, R., Weyhermüller, T., Chandrasekaran, P., Glatzel, P., Neese, F. & DeBeer, S. (2013). Phys. Chem. Chem. Phys. 15, 20911–20920. Web of Science CrossRef CAS PubMed Google Scholar
Lima, F. A., Otte, F., Vakili, M., Ardana-Lamas, F., Biednov, M., Dall'Antonia, F., Frankenberger, P., Gawelda, W., Gelisio, L., Han, H., Huang, X., Jiang, Y., Kloos, M., Kluyver, T., Knoll, M., Kubicek, K., Bermudez Macias, I. J., Schulz, J., Turkot, O., Uemura, Y., Valerio, J., Wang, H., Yousef, H., Zalden, P., Khakhulin, D., Bressler, C. & Milne, C. (2023). J. Synchrotron Rad. 30, 1168–1182. Web of Science CrossRef CAS IUCr Journals Google Scholar
Mara, M. W., Hadt, R. G., Reinhard, M. E., Kroll, T., Lim, H., Hartsock, R. W., Alonso-Mori, R., Chollet, M., Glownia, J. M., Nelson, S., Sokaras, D., Kunnus, K., Hodgson, K. O., Hedman, B., Bergmann, U., Gaffney, K. J. & Solomon, E. I. (2017). Science, 356, 1276–1280. Web of Science CrossRef CAS PubMed Google Scholar
Milne, C., Penfold, T. & Chergui, M. (2014). Coord. Chem. Rev. 277–278, 44–68. Web of Science CrossRef CAS Google Scholar
Moretti Sala, M., Martel, K., Henriquet, C., Al Zein, A., Simonelli, L., Sahle, C., Gonzalez, H., Lagier, M.-C., Ponchut, C., Huotari, S., Verbeni, R., Krisch, M. & Monaco, G. (2018). J. Synchrotron Rad. 25, 580–591. Web of Science CrossRef CAS IUCr Journals Google Scholar
Naumova, M. A., Kalinko, A., Wong, J. W. L., Abdellah, M., Geng, H., Domenichini, E., Meng, J., Gutierrez, S. A., Mante, P.-A., Lin, W., Zalden, P., Galler, A., Lima, F., Kubicek, K., Biednov, M., Britz, A., Checchia, S., Kabanova, V., Wulff, M., Zimara, J., Schwarzer, D., Demeshko, S., Murzin, V., Gosztola, D., Jarenmark, M., Zhang, J., Bauer, M., Lawson Daku, M. L., Gawelda, W., Khakhulin, D., Bressler, C., Meyer, F., Zheng, K. & Canton, S. E. (2020a). J. Phys. Chem. Lett. 11, 2133–2141. Web of Science CrossRef CAS PubMed Google Scholar
Naumova, M. A., Kalinko, A., Wong, J. W. L., Alvarez Gutierrez, S., Meng, J., Liang, M., Abdellah, M., Geng, H., Lin, W., Kubicek, K., Biednov, M., Lima, F., Galler, A., Zalden, P., Checchia, S., Mante, P.-A., Zimara, J., Schwarzer, D., Demeshko, S., Murzin, V., Gosztola, D., Jarenmark, M., Zhang, J., Bauer, M., Lawson Daku, M. L., Khakhulin, D., Gawelda, W., Bressler, C., Meyer, F., Zheng, K. & Canton, S. E. (2020b). J. Chem. Phys. 152, 214301. Web of Science CrossRef PubMed Google Scholar
Naumova, M. A., Paveliuc, G., Biednov, M., Kubicek, K., Kalinko, A., Meng, J., Liang, M., Rahaman, A., Abdellah, M., Checchia, S., Alves Lima, F., Zalden, P., Gawelda, W., Bressler, C., Geng, H., Lin, W., Liu, Y., Zhao, Q., Pan, Q., Akter, M., Kong, Q., Retegan, M., Gosztola, D. J., Pápai, M., Khakhulin, D., Lawson Daku, M., Zheng, K. & Canton, S. E. (2024). J. Phys. Chem. Lett. 15, 3627–3638. CrossRef CAS PubMed Google Scholar
Nowakowski, M., Huber–Gedert, M., Elgabarty, H., Kalinko, A., Kubicki, J., Kertmen, A., Lindner, N., Khakhulin, D., Lima, F. A., Choi, T.-K., Biednov, M., Schmitz, L., Piergies, N., Zalden, P., Kubicek, K., Rodriguez–Fernandez, A., Salem, M. A., Canton, S. E., Bressler, C., Kühne, T. D., Gawelda, W. & Bauer, M. (2024). Adv. Sci. 11, 2404348. CrossRef Google Scholar
Ogasawara, H., Kotani, A. & Thole, B. T. (1994). Phys. Rev. B, 50, 12332–12341. CrossRef CAS Google Scholar
Pollock, C. J. & DeBeer, S. (2015). Acc. Chem. Res. 48, 2967–2975. Web of Science CrossRef CAS PubMed Google Scholar
Ravel, B., Kropf, A. J., Yang, D., Wang, M., Topsakal, M., Lu, D., Stennett, M. C. & Hyatt, N. C. (2018). Phys. Rev. B, 97, 125139. Web of Science CrossRef Google Scholar
Rebuffi, L. & del Rio, M. S. (2017). Proc. SPIE, 10388, 169–177. Google Scholar
Rebuffi, L. & Sánchez del Río, M. (2016). J. Synchrotron Rad. 23, 1357–1367. Web of Science CrossRef IUCr Journals Google Scholar
Sá, J., Kayser, Y., Milne, C. J., Abreu Fernandes, D. L. & Szlachetko, J. (2014). Phys. Chem. Chem. Phys. 16, 7692. PubMed Google Scholar
Sanchez del Rio, M. & Dejus, R. J. (2011). Proc. SPIE, 8141, 368–372. Google Scholar
Schuth, N., Zaharieva, I., Chernev, P., Berggren, G., Anderlund, M., Styring, S., Dau, H. & Haumann, M. (2018). Inorg. Chem. 57, 10424–10430. CAS PubMed Google Scholar
Sension, R. J., McClain, T. P., Lamb, R. M., Alonso-Mori, R., Lima, F. A., Ardana-Lamas, F., Biednov, M., Chollet, M., Chung, T., Deb, A., Dewan, P. A. J., Gee, L. B., Huang Ze En, J., Jiang, Y., Khakhulin, D., Li, J., Michocki, L. B., Miller, N. A., Otte, F., Uemura, Y., van Driel, T. B. & Penner-Hahn, J. E. (2023). J. Am. Chem. Soc. 145, 14070–14086. Web of Science CrossRef CAS PubMed Google Scholar
Sension, R. J., McClain, T. P., Michocki, L. B., Miller, N. A., Alonso-Mori, R., Lima, F. A., Ardana-Lamas, F., Biednov, M., Chung, T., Deb, A., Jiang, Y., Kaneshiro, A. K., Khakhulin, D., Kubarych, K. J., Lamb, R. M., Meadows, J. H., Otte, F., Sofferman, D. L., Song, S., Uemura, Y., van Driel, T. B. & Penner-Hahn, J. E. (2024). J. Phys. Chem. B, 128, 8131–8144. CAS PubMed Google Scholar
Shvyd'ko, Y. (2004). X-ray Optics: High-Energy-Resolution Applications, Vol. 98. Springer Science & Business Media. Google Scholar
Smolentsev, G., Milne, C. J., Guda, A., Haldrup, K., Szlachetko, J., Azzaroli, N., Cirelli, C., Knopp, G., Bohinc, R., Menzi, S., Pamfilidis, G., Gashi, D., Beck, M., Mozzanica, A., James, D., Bacellar, C., Mancini, G. F., Tereshchenko, A., Shapovalov, V., Kwiatek, W. M., Czapla-Masztafiak, J., Cannizzo, A., Gazzetto, M., Sander, M., Levantino, M., Kabanova, V., Rychagova, E., Ketkov, S., Olaru, M., Beckmann, J. & Vogt, M. (2020). Nat. Commun. 11, 2131. Web of Science CrossRef PubMed Google Scholar
Su, K., Liu, H., Gao, Z., Fornasiero, P. & Wang, F. (2021). Adv. Sci. 8, 2003156. Google Scholar
Suortti, P., Buslaps, T., Fajardo, P., Honkimäki, V., Kretzschmer, M., Lienert, U., McCarthy, J. E., Renier, M., Shukla, A., Tschentscher, T. & Meinander, T. (1999). J. Synchrotron Rad. 6, 69–80. Web of Science CrossRef CAS IUCr Journals Google Scholar
Szlachetko, J., Nachtegaal, M., de Boni, E., Willimann, M., Safonova, O., Sa, J., Smolentsev, G., Szlachetko, M., van Bokhoven, J. A., Dousse, J. C., Hoszowska, J., Kayser, Y., Jagodzinski, P., Bergamaschi, A., Schmitt, B., David, C. & Lücke, A. (2012). Rev. Sci. Instrum. 83, 103105. Web of Science CrossRef PubMed Google Scholar
Szlachetko, M., Berset, M., Dousse, J. C., Hoszowska, J. & Szlachetko, J. (2013). Rev. Sci. Instrum. 84, 093104. Web of Science CrossRef PubMed Google Scholar
Torres Deluigi, M., de Groot, F. M. F., López-Díaz, G., Tirao, G., Stutz, G. & Riveros de la Vega, J. (2014). J. Phys. Chem. C, 118, 22202–22210. Web of Science CrossRef CAS Google Scholar
Wakisaka, Y., Iwasaki, Y., Uehara, H., Mukai, S., Kido, D., Takakusagi, S., Uemura, Y., Wada, T., Yuan, Q., Sekizawa, O., et al. (2017). Hyomen Kagaku, 38, 378–383. CAS Google Scholar
Widmann, K., Beiersdorfer, P., Brown, G. V., Crespo López Urrutia, J. R., Decaux, V. & Savin, D. W. (1997). Rev. Sci. Instrum. 68, 1087–1090. CAS Google Scholar
Witczak-Krempa, W., Chen, G., Kim, Y. B. & Balents, L. (2014). Annu. Rev. Condens. Matter Phys. 5, 57–82. CAS Google Scholar
Zhang, W., Alonso-Mori, R., Bergmann, U., Bressler, C., Chollet, M., Galler, A., Gawelda, W., Hadt, R. G., Hartsock, R. W., Kroll, T., Kjaer, K. S., Kubiček, K., Lemke, H. T., Liang, H. W., Meyer, D. A., Nielsen, M. M., Purser, C., Robinson, J. S., Solomon, E. I., Sun, Z., Sokaras, D., van Driel, T. B., Vankó, G., Weng, T.-C., Zhu, D. & Gaffney, K. J. (2014). Nature, 509, 345–348. Web of Science CrossRef CAS PubMed Google Scholar
Zhang, X., Smolentsev, G., Guo, J., Attenkofer, K., Kurtz, C., Jennings, G., Lockard, J. V., Stickrath, A. B. & Chen, L. X. (2011). J. Phys. Chem. Lett. 2, 628–632. CAS Google Scholar
Zhong, Z., Chapman, L. D., Bunker, B. A., Bunker, G., Fischetti, R. & Segre, C. U. (1999). J. Synchrotron Rad. 6, 212–214. Web of Science CrossRef CAS IUCr Journals Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.