A low-background setup for in situ X-ray total scattering combined with fast scanning calorimetry

crossmark logo

aDipartimento di Fisica e Astronomia `Galileo Galilei', Università degli Studi di Padova, Via F. Marzolo 8, 35131 Padova, Italy, and bESRF – The European Synchrotron, 71 Avenue des Martyrs, CS40220, 38043 Grenoble Cedex 9, France
*Correspondence e-mail: [email protected]

Edited by M. Yabashi, RIKEN SPring-8 Center, Japan (Received 12 March 2025; accepted 20 June 2025; online 23 July 2025)

We demonstrate a setup, combining fast scanning calorimetry with X-ray total scattering at a synchrotron beamline, allowing for in situ characterizations of the nano-scale structure of samples during and after temperature scans. The setup features a portable vacuum chamber providing a high signal-to-background ratio even on amorphous samples, which enables the observation of detailed structural changes between different sample states. We present three use cases, including one that leverages the high cooling rate of 104 K s−1 achievable by this setup. Our demonstration opens the door to various applications in materials science where understanding the interplay between structure and thermodynamics is crucial.

1. Introduction

Obtaining atomic scale structural information of materials under different thermodynamic conditions is crucial to the understanding of material properties and the physical mechanism governing their behaviors. This is particularly important for systems like glasses whose properties have a sensitive dependence on their thermal history (for example, the cooling rate at which they are quenched down from the liquid state) (Debenedetti & Stillinger, 2001View full citation). Therefore, the ability to perform in situ structural measurements on samples during and after controlled thermal programs is highly valuable for studies in the field of physics and materials science.

The temperature control of samples for in situ measurements is typically realized through the use of conventional furnaces based on radiative or conductive heating. These devices often suffer from serious inaccuracies in the determination of sample temperature or from very limited temperature ramping rates due to large thermal masses (Doran et al., 2017View full citation). More recent works have improved the design, resulting in furnaces that allow for in situ structural characterizations with higher temperature precision and higher ramping rates (Doran et al., 2017View full citation; Chakraborty et al., 2015View full citation; Marshall et al., 2023View full citation). Still, the heating and cooling rates are limited to ∼100 K s−1 and more often to ∼5 K s−1, and in some designs the need for a gaseous environment for efficient thermal equilibration leads to significant additional background in structural measurement techniques including X-ray diffraction.

Nanocalorimetry, including fast scanning calorimetry (FSC) (Mathot et al., 2023View full citation; Yi & LaVan, 2019View full citation), offers a promising alternative for in situ structural and thermodynamic studies of various systems. Due to the small sample mass on the microgram or even nanogram level, nanocalorimeters can achieve heating and cooling rates in excess of 106 K s−1 (Mathot et al., 2023View full citation; Yi & LaVan, 2019View full citation). In addition, microelectromechanical systems (MEMS)-based calorimeters (Mathot et al., 2023View full citation; Mathot et al., 2011View full citation; Zhuravlev & Schick, 2010View full citation), where the sample is mounted on a sensor that can be placed externally to the bulk of the main device, provide a versatile solution that can be combined with X-ray scattering instruments at synchrotron radiation facilities (Rosenthal et al., 2014View full citation; Narayanan et al., 2017View full citation). For example, in previous work, we demonstrated the combination of FSC with X-ray photon correlation spectroscopy and showcased, among other applications, the use of FSC as a fast-responding furnace for dynamic scattering measurements (Martinelli et al., 2024View full citation).

The small sample size for FSC, usually no more than a few tens of micrometres and sometimes even smaller, could in principle present a challenge for nano-scale structural characterizations. However, thanks to recent advancements in synchrotron radiation facilities, it is now possible to have high-flux, high-photon-energy X-ray beams with a small focus, thereby enabling high-quality X-ray total scattering measurements to retrieve the atomic pair distribution function (PDF), providing crucial insight into the atomic-scale structural properties of various materials (Billinge, 2019View full citation). For example, after the upgrade to the Extremely Brilliant Source (EBS) at beamline ID15A of The European Synchrotron (ESRF), it is possible to have 69 keV X-ray beams focused down to 0.3 µm × 0.3 µm with a flux well exceeding 1012 photons s−1 (Vaughan et al., 2020View full citation). This makes it possible to perform high-speed operando X-ray total scattering experiments on small samples, including those for chip-based FSC.

In this work, we demonstrate the combination of X-ray total scattering and FSC for in situ measurements. In particular, we present a setup including a custom-designed low-background sample chamber that enables the collection of X-ray diffraction data with a high signal-to-background ratio. We show that, with this setup, it is possible to obtain high-quality PDFs revealing detailed differences in the sample structure during and in between FSC scans. The rest of this article is organized as follows: Section 2[link] describes our experimental setup with a quantitative analysis of the background level; and Sections 3–5 give three example applications of this setup: measurements of detailed structural changes during temperature ramps, quantification of X-ray beam-induced heating effects and characterizations of atomic structure with different cooling rates spanning four orders of magnitude.

2. A low-background sample chamber for in situ measurements

Fig. 1[link] shows a schematic of the setup, including several details of the sample chamber. This setup is realized at beamline ID15A of the ESRF, and a commercial Mettler-Toledo Flash DSC 2+ calorimeter is used. X-ray scattering patterns are collected by a hybrid photon-counting pixel detector Pilatus3 X CdTe (Vaughan et al., 2020View full citation). Although the sample chamber is designed to be compatible with ID15A, it is made to be stand-alone and can be installed at other beamlines for X-ray diffraction measurements. The chamber is vacuum-compatible, but it can also be filled with gas if necessary (albeit with an increased background level for X-ray diffraction measurements). Because we have demonstrated before that FSC scans can be performed with the sample in vacuum achieving high heating and cooling rates (Martinelli et al., 2024View full citation), in this work the FSC scans and X-ray scattering measurements are done with the chamber under vacuum.

[Figure 1]
Figure 1
(a) Schematic of the low-background sample chamber (not to scale). The sample is mounted on an FSC chip inside the chamber. A mask made of tungsten alloy is placed before the chip and acts as a pinhole. The chamber features a large exit window of Kapton foil allowing measurements up to high Q. (b) Section view of the chamber showing various components along the X-ray beam path.

The design of the chamber is similar to that in our previous work (Martinelli et al., 2024View full citation), although with several important modifications to make it a versatile sample environment for implementation at high-energy X-ray scattering beamlines. Firstly, it features a large exit window to allow measurements up to high values of the momentum transfer, Q = (4π/λ)sinθ, where λ is the X-ray wavelength and θ is half of the scattering angle. For the demonstrations in this work, the X-ray photon energy is fixed at 67.808 keV (with 0.25 keV bandwidth) and the valid Q-range is from 0.5 Å−1 to 19 Å−1, with the lower limit given by the 2 mm-diameter beamstop placed close to the exit window. The beamstop is mounted externally to the chamber and is aligned independently. Secondly, for the chamber to be stand-alone (i.e. not directly connected to the beamline's X-ray pipeline), it features a small window (10 mm in diameter) for incoming X-rays, as indicated in Fig. 1[link]. In order to reduce the background that is inevitably created by this window and the air path before it, we have manufactured a mask out of a 3 mm-thick tungsten alloy sheet (composition W80Cu20 by wt%), with a hole of diameter ∼400 µm acting as a pinhole, referred to as `WCu pinhole mask' hereafter. This takes advantage of the fact that the calorimeter chips have small active areas (no more than a few hundred micrometres laterally) at fixed positions where the sample can be mounted. Another 50 mm × 50 mm square mask with a 5 mm-diameter hole [`WCu screen' in Fig. 1[link](b)], made of the same tungsten alloy, is placed between the entry window and the sample to further reduce the background.

To quantify the background level, we perform X-ray scattering measurements on and off a metallic glass sample of composition Pd42.5Cu30Ni7.5P20 (at.%) and thickness on the order of ∼20 µm. The sample is mounted on a Mettler–Toledo UFH1 chip, whose active area consists of gold and SiNx layers (Mathot et al., 2023View full citation) and has a diameter on the order of 90 µm. Further details of the chips available with Mettler–Toledo Flash DSC machines can be found in previous reports (Mathot et al., 2023View full citation; Mathot et al., 2011View full citation). In our setup, the chip is fixed together with the WCu pinhole mask onto the sample frame by an aluminium cover [see Fig. 1[link](b)]. Both the cover and the frame have clearances with large diameters (>15 mm) at the center for X-rays to pass through. For consistency, the same sample is used here and in the rest of this work, and the X-ray spot size at the sample position is about 1 µm × 2 µm. Azimuthal integration of the diffraction patterns is performed using the pyFAI library (Kieffer et al., 2025View full citation), which takes into account the X-ray polarization factor and the solid angle of each detector pixel. The experimental geometry is calibrated using a Cr2O3 powder fixed onto the chip, and the sample-to-detector distance is determined to be 199.6 mm. We note that, while the detector is mounted on a motor and can be moved if needed, the same detector position is used for all measurements presented below.

At ID15A, the incoming X-ray flux at the sample position can be attenuated by inserting an SiO2 wedge in the beam path upstream from the sample in experimental hutch EH2 (Vaughan et al., 2020View full citation). The wedge is mounted on a motorized linear stage and can be moved to change its thickness in the beam path, thereby achieving the desired attenuation. The correspondence between the motor position and the X-ray flux onto the sample is calibrated using a beam intensity monitor (PIN diode). The attenuation does not change the beam shape at the sample position since the homogeneous chemical composition and optically polished surfaces of the wedge do not alter the shape of the X-ray beam wavefront.

Fig. 2[link](a) shows an example image of the diffraction pattern on the sample, where the color map is set on a log scale in order to show different features in the pattern. The incoming X-ray flux is about 4.5 × 1011 photons s−1 in this measurement. Scattering signal up to at least the third diffraction peak can be identified in the image, while the large circular contour corresponds to the edge of the exit window. In Fig. 2[link](b), we show the background scattering taken at a position away from the sample, with the same incoming X-ray flux and using the same color map. It is evident that the intensity here is orders of magnitude lower than the signal from the sample. To show this more clearly, we plot the integrated intensity profiles in Fig. 2[link](c). It can be seen that the background is negligible compared with the signal from the sample, even in the low-Q region where the structure factor of the sample is expected to be very small. The profile subtracting the background from the signal is not shown here because it visibly overlaps with the total signal shown in the plot (more information is provided in Appendix A[link]). The two broad peaks in the background centered around 2 Å−1 and 4.6 Å−1, as can also be seen in Fig. 2[link](b), are consistent with the structure factor of amorphous silicon nitride (Aiyama et al., 1979View full citation) and arise from the silicon nitride substrate of the FSC chip. In comparison, the black dashed line in Fig. 2[link](c) indicates the background level after the chamber was opened to ambient pressure and the sample assembly, including the WCu pinhole mask, was removed. Note that for this measurement, it was necessary to move the chamber upstream by 11 mm to prevent the Kapton exit window from touching the beamstop when it expands upon ventilating the chamber. It is clear that the background level becomes much higher, particularly in the low-Q region. This kind of background can be difficult to account for accurately, as it contains contributions from different sources (for example, air scattering before and after the sample) that will be attenuated differently after insertion of the sample.

[Figure 2]
Figure 2
Background characterization. (a) Example diffraction pattern from a Pd42.5Cu30Ni7.5P20 metallic glass sample, roughly 20 µm thick, mounted on a Mettler–Toledo UFH1 chip. (b) Background diffraction pattern taken at another position on the chip away from the sample, plotted with the same color map. (c) Azimuthally integrated scattering intensity profiles for the sample (red) and background (blue), in units of photons per second per unit solid angle. The dashed black line shows the intensity profile with the chamber in air and without the sample assembly including the WCu pinhole mask (see text). Dashed and dotted vertical lines indicate the positions of Bragg reflections of W and Au, respectively, that are visible in the intensity profiles.

It can be seen in Fig. 2[link] that some sharp Bragg peaks appear in the diffraction images. As shown in Fig. 2[link](c), these Bragg peaks can be attributed to tungsten and gold. The tungsten peaks arise from the tungsten alloy mask and are negligible compared with the scattering signal from the sample, as shown in Appendix A[link]. The gold peaks arise from the active area and from the wiring of the UFH1 chip, so they are not intrinsic to the chamber design but are specific to the chip we used. These peaks are stronger at the sample position where the X-ray beam directly hits the active area, and their intensity varies with position due to the changing texture and thickness of the gold layer. During sample alignment, we typically adjust the beam position on the sample to minimize the intensity of these peaks. Then, in data processing, we identify the remaining peaks and mask the surrounding pixels; the procedure is described in Appendix A[link].

3. In situ X-ray total scattering during temperature ramps

As a first demonstration, we show that the setup can be used as a fast-responding furnace that allows for quick and accurate measurements of the atomic structure at various temperatures. To do this, we set up the following temperature scan program on the calorimeter: a temperature ramp up from 298 K to 523 K at 2 K s−1, a 10 s isotherm at 523 K, and a symmetric ramp down to 298 K at 2 K s−1. The scan is padded with isotherms at 298 K at the beginning and the end of the temperature ramps to allow for possible mismatches in launching the X-ray measurements. The highest temperature is kept well below the glass transition temperature Tg ≃ 580 K (Georgarakis et al., 2011View full citation) in order to avoid aging effects above Tg that can obscure the reproducibility between scans. A visual representation of the temperature program is shown as a red line in Fig. 3[link](b) (right y axis). Then, we collect X-ray diffraction patterns during the scan. Here, the incoming X-ray flux is about 4.4 × 1010 photons s−1, which does not cause a significant temperature rise as demonstrated in the next section. The frame rate is 1 Hz and the exposure time is 0.99 s per frame. We repeat the scan twice to check the reproducibility of the results.

[Figure 3]
Figure 3
Demonstration of in situ X-ray diffraction during an FSC scan. (a) Intensity profile I(Q) normalized by its maximum value, focusing on the vicinity of the first sharp diffraction peak to show its temperature dependence during the heating and cooling ramps of a scan. From dark blue to dark red, the curves show increasing temperatures, as indicated by the color bar. (b) Blue and yellow dots show the center position of the first sharp diffraction peaks, Qm1, as a function of time during two different scans. The temperature program is shown as the red line (right y axis).

A sensitive and direct observable that reflects temperature change in this sample is the position of the first sharp diffraction peak, Qm1. This can be seen in Fig. 3[link](a), which shows the azimuthally integrated intensity profile I(Q) normalized by its maximum value I(Qm1) during the heating and cooling ramps of a scan. Each curve represents an average over five consecutive images. When extracting Qm1 from these data, since the curves do not appear to have an exact functional shape (e.g. Gaussian) and the differences between them are small, the absolute values of Qm1 may depend somewhat on the model used to extract them. Thus, to avoid ambiguity, we define Qm1 as the first moment (center of mass) of I(Q) in the range where it is above 95% of its maximum value, as shown in Fig. 3[link](a), which we find to be a definition that is robust and easy to implement. We then plot Qm1 as a function of time during the scans, shown as dots in Fig. 3[link](b). The two colors represent the two different scans, and the good agreement between them demonstrates a high level of reproducibility. Furthermore, the shift in Qm1 overlaps very well with the temperature profile, showing the possibility to use the calorimeter as a fast-responding and accurate furnace.

One of the main advantages of our setup is its capability of covering a wide Q-range with a low background, which enables the reconstruction of the atomic PDF of the sample in situ during temperature scans. Therefore, we perform PDF analysis on the diffraction data using the software PDFgetX3 (Juhás et al., 2013View full citation), and the results are shown in Fig. 4[link](a) (for clarity, only data from the temperature ramp up of one of the scans is shown in this panel). Even though the changes are small as expected for this temperature range (Georgarakis et al., 2011View full citation), they are nonetheless clearly visible. For example, as shown in the insets in Fig. 4[link](a), with increasing temperature, the first peak (centered around 2.8 Å) lowers in amplitude and becomes broader without apparently shifting its center position, while the third peak (centered around 7.0 Å) shifts to higher r. Higher order peaks behave similarly as the third peak, so here we focus on the third peak due to its larger amplitude giving a higher signal-to-noise ratio.

[Figure 4]
Figure 4
PDF analysis as a function of temperature during the heating ramp. (a) Full G(r) profile. The insets show magnified views of the first and third peaks. The lower panels show, as a function of temperature in two different scans; (b) the mean position of the first peak, r1; (c) the standard deviation of the first peak, σ1; and (d) the mean position of the third peak, r3. The red dashed line in (d) shows a linear fit of the data.

To quantify the observations above, we extract and plot the following quantities from the G(r) data: the first moment of the main first peak, r1 [Fig. 4[link](b)]; the square root of the second moment of the first peak (i.e. its standard deviation), σ1 [Fig. 4[link](c)]; and the first moment of the third peak, r3 [Fig. 4[link](d)]. We choose the following range to calculate these quantities: the range for the first peak of G(r) is determined by the two near-isosbsetic points at around 2.60 Å and 2.95 Å, while the range for the third peak is determined by the roots of G(r) around 6.5 Å and 7.5 Å (varying slightly with temperature). Note that the results depend only sightly on the exact range chosen. We can see that Figs. 4[link](b)–4[link](d) quantitatively confirm the observations above. In particular, r3 appears to expand linearly with temperature, as shown by the fit (red dashed line) in Fig. 4[link](d). The slope suggests a linear expansion coefficient of 1.8 × 10−5 K−1, or a volumetric expansion coefficient of 5.3 × 10−5 K−1, consistent with values reported in the literature (Lu et al., 2002View full citation; Georgarakis et al., 2011View full citation). We note that similar values can be obtained from the slope of Qm1 shown in Fig. 3[link] above. We also note in passing that the much smaller changes (if any) in r1 and the increase in σ1 suggest that the overall thermal expansion is not related to changes in the average nearest-neighbor distance, but to an increase in the width of its distribution. In other words, in this temperature range, the system appears to expand in the long range accompanied by an increase in local disorder.

4. Quantifying X-ray beam-induced heating

As an application of this setup, we characterize the effective heating due to the X-ray beam, which is highly relevant for measurements on temperature-sensitive samples. To do this, we use the attenuator available at the beamline to tune the incoming flux, I0, from full beam (∼1.4 × 1012 photons s−1) to 0.3% transmission (∼4.2 × 109 photons s−1). During this experiment, the synchrotron was operating under the uniform filling mode. Then, we take X-ray diffraction images at each flux value, from which we can extract the position of the first sharp diffraction peak, Qm1. Since we have shown in the previous section that Qm1 shifts linearly with temperature at least up to 523 K, we may infer the effective temperature rise from the Qm1 position at each I0 value. Specifically, we obtain the difference between the measured Qm1 at each I0 and that measured at the lowest flux (0.3% transmission), and calculate the corresponding temperature difference ΔT using the measured slope from Fig. 3[link], ΔQm1/ΔT = −4.0 × 10−5 Å−1 K−1. The results are plotted in Fig. 5[link](a). We can see that ΔT scales linearly with the incoming flux, I0, as expected. Importantly, the effective temperature rise with full beam appears to be more than 60 K, which may lead to rather serious consequences if not considered carefully (for example, it may cause effective shifts in the apparent glass transition temperature of the sample). This significant temperature rise is likely an effect of the relatively high absorption by the sample (Z = 46 for Pd) and the small beam size (even though the effective area of energy deposition is larger due to the diffusion of secondary electrons). Nonetheless, the values obtained here may offer as a guide for measurements under other conditions.

[Figure 5]
Figure 5
(a) Effective temperature rise due to X-ray beam. Data points (black circles) are calculated by extracting the difference of Qm1 at each incoming flux with respect to Qm1 at the lowest flux (0.3% transmission, about ∼4.2 × 109 photons s−1), and converted to ΔT using the linear relation obtained from Fig. 3[link]. The red dashed line shows a linear fit. (bQm1 position (black circles, left y axis) and r1 position (orange crosses, right y axis) as a function of time under full beam. No significant changes are observed.

One might be concerned about permanent beam-induced structural modifications which are observed in several network glasses (Ruta et al., 2017View full citation; Baglioni et al., 2024View full citation). Here, we show that if any effect exists, it is negligible in this metallic glass sample. This can already be seen in Fig. 3[link], where the Qm1 returns to its original value after the temperature ramp down and is highly reproducible in a subsequent scan (no other calorimetric scans were performed in between). To further demonstrate the lack of significant beam-induced structural changes, we take diffraction patterns while irradiating the sample with full beam. In Fig. 5[link](b), we plot Qm1 (left y axis) as well as r1 (right y axis, extracted from the PDF) as a function of time. Both quantities appear to remain constant within the noise level (note the small ranges of y axes), and no time-dependent trend can be seen. Therefore, we observe no permanent beam-induced effects beyond noise in this sample with more than 20 s of irradiation by the full beam, despite the apparent ∼70 K temperature rise due to X-ray irradiation.

We note that the flux used both in the previous section and in the next section is about 4.4 × 1010 photons s−1 (about 3% transmission). The results suggest that ΔT is negligible (less than 3 K) under this flux.

5. Cooling-rate dependence of sample structure

In this section, we demonstrate an application leveraging the ability of the flash calorimeter to prepare samples at controlled cooling rates spanning several orders of magnitude. To do this, we launch temperature programs in which the sample is first heated up to 643 K at a rate of 1000 K s−1, then kept at 643 K with an isotherm of 10 s, and finally cooled down at a given cooling rate to 298 K. The cooling rates used here range from 1 K s−1 to 104 K s−1. We note that, because the isotherm temperature 643 K is much higher than the glass transition temperature, Tg ≃ 580 K (Georgarakis et al., 2011View full citation), the sample has reached equilibrium at 643 K before being quenched. Then, we make X-ray total scattering measurements on the sample thus prepared, using an incoming X-ray flux of about 4.4 × 1010 photons s−1. We note in passing that the sequence of the cooling rates used are in a random order, so as to exclude systematic shifts in the structure with time during the measurements.

The PDFs obtained from our measurements are presented in Fig. 6[link], where the curves from dark blue to dark red represent increasing cooling rates. The changes are small but nonetheless visible, as shown in the magnified views in the insets: with higher cooling rates, the first peak appears to shift inward, while the third peak appears to lower in amplitude without shifting significantly in its center position. In Figs. 6[link](b)–6[link](c), we further quantify these observations by extracting the first moments of the first peak, r1, and that of the third peak, r3, in the same way as detailed in Section 3[link]. Notably, we observe that r1 appears to shift logarithmically with the cooling rate. Since the same trend appears to continue to the highest cooling rate, 104 K s−1, it suggests that the cooling rate achievable under vacuum with this sample is at least 104 K s−1.

[Figure 6]
Figure 6
Structural dependence on the cooling rate. (a) PDF of the sample quenched at different rates after equilibration at 634 K for 10 s. From dark blue to dark red, the curves show increasing cooling rates, as indicated by the color bar. The insets show magnified views of the first and third peaks. (b) Mean position of the first peak, r1; and (c) the mean position of the third peak, r3, as a function of the cooling rate.

It is particularly noteworthy that the structural changes observed here are rather different from those measured during the temperature ramp as shown in Fig. 4[link]. For example, whereas r1 does not seem to change significantly with temperature up to 523 K, here it appears to contract with higher cooling rates, while higher order shells do not seem to change in their positions. We note that this curious contraction of the first coordination shell is observed also when the sample is heated above Tg in a previous work (Georgarakis et al., 2011View full citation). A more detailed understanding of the mechanism behind these structural changes with cooling rate is beyond the scope of this work and can be the subject of future studies.

6. Conclusions

We have presented a setup that combines FSC with X-ray total scattering which can enable in situ structural characterizations on the atomic scale during and after FSC scans. Using a low-background vacuum-compatible sample chamber, we are able to collect high-quality diffraction data from a Pd-based metallic glass sample and observe detailed structural changes between different thermodynamic states, including during temperature ramps and after quenching from the supercooled liquid state at different cooling rates ranging from 1 K s−1 to 104 K s−1. The results indicate interesting structural behaviors in this sample which can become the basis for future studies, and the high level of structural details revealed demonstrates the potential of the setup to be applied to a wide range of materials studies.

APPENDIX A

Data processing

As mentioned above in the main text, the Bragg peaks arising from the gold layer in the active area of the chip may have an influence on the PDF analysis, but it is possible to identify and mask them. Here we describe our algorithm for peak masking. Firstly, we apply a 2D Gaussian filter on the image with σ = 3 pixels in each dimension. Then, we identify the pixels in which the ratio between the original and the filtered image, Iorig/Ifiltered, is more than a threshold of 1.3. Lastly, we mask all pixels within a 5 × 5 square centered on these identified pixels. We note that the parameters used here are determined for the specific sample in this study; for other samples, it may be necessary to adjust these parameters to optimize the results.

Fig. 7[link] shows an example of the result of peak masking. The original image is the same as the one shown as `sample' in Fig. 2[link] and contains obvious Bragg peaks. These Bragg peaks lead to significant oscillations in the calculated PDF that should not appear for an amorphous sample. After the peaks are masked following the procedure described above, both I(Q) and G(r) become smooth and typical of amorphous materials. We also demonstrate that the subtraction of the background shown in Fig. 2[link] does not lead to significant changes in I(Q) or G(r), as the results visibly overlap with the curves before this background subtraction in Fig. 7[link].

[Figure 7]
Figure 7
Effects of Bragg peak masking and background subtraction. (a) Azimuthally integrated intensity, I(Q), of the original image (red), after peak masking (green) and after background subtraction (black dotted line). The original image is the same as in Fig. 2[link]. (b) Corresponding PDFs.

We note that, for demonstration purposes, the example chosen here is not specifically optimized to reduce the intensity of the Bragg peaks. Nonetheless, the amount of masked pixels not adjacent to detector panel edges or beamstop shadow (both create sharp edges in the images which are identified by the algorithm) is on the order of 1.5 × 104, less than 1.5% of the valid pixels. Therefore, the peak masking does not cause significant degradation to the data quality.

Acknowledgements

We thank Denis Duran and Emmanuel Papillon for assistance with the setup at ID15A. We thank Dr Carlo Scian for technical support in the construction and testing of the vacuum chamber. We thank Professor Eloi Pineda for providing the sample used in this study. Open access publishing facilitated by Universita degli Studi di Padova, as part of the Wiley–CRUI-CARE agreement.

Conflict of interest

The authors declare no conflicts of interest.

Data availability

The data supporting the results reported in the present work can be found within the article. Raw data are obtained at beamline ID15A of the ESRF and can be found at https://doi.org/10.15151/ESRF-ES-2005462101. Further data can be provided by the corresponding author, PS, upon reasonable request.

Funding information

This project is financed by the European Union – NextGenerationEU and by the STARS@UNIPD program. We acknowledge the ESRF for provision of synchrotron radiation facilities under proposal No. SC-5581. DL, LP, FD and GM acknowledge support by the project GLAXES ERC-2021-ADG (grant agreement No. 101053167) funded by the European Research Council, European Union.

References

Return to citationAiyama, T., Fukunaga, T., Niihara, K., Hirai, T. & Suzuki, K. (1979). J. Non-Cryst. Solids 33, 131–139.  CrossRef CAS Web of Science Google Scholar
Return to citationBaglioni, J., Martinelli, A., Sun, P., Dallari, F., Westermeier, F., Sprung, M., Grübel, G. & Monaco, G. (2024). Rep. Prog. Phys. 87, 120503.  Web of Science CrossRef Google Scholar
Return to citationBillinge, S. J. L. (2019). Philos. Trans. R. Soc. A. 377, 20180413.  Web of Science CrossRef Google Scholar
Return to citationChakraborty, R., Serdy, J., West, B., Stuckelberger, M., Lai, B., Maser, J., Bertoni, M. I., Culpepper, M. L. & Buonassisi, T. (2015). Rev. Sci. Instrum. 86, 113705.  Web of Science CrossRef PubMed Google Scholar
Return to citationDebenedetti, P. G. & Stillinger, F. H. (2001). Nature 410, 259–267.  Web of Science CrossRef PubMed CAS Google Scholar
Return to citationDoran, A., Schlicker, L., Beavers, C. M., Bhat, S., Bekheet, M. F. & Gurlo, A. (2017). Rev. Sci. Instrum. 88, 013903.  Web of Science CrossRef PubMed Google Scholar
Return to citationGeorgarakis, K., Louzguine-Luzgin, D. V., Antonowicz, J., Vaughan, G., Yavari, A. R., Egami, T. & Inoue, A. (2011). Acta Mater. 59, 708–716.  Web of Science CrossRef CAS Google Scholar
Return to citationJuhás, P., Davis, T., Farrow, C. L. & Billinge, S. J. L. (2013). J. Appl. Cryst. 46, 560–566.  Web of Science CrossRef IUCr Journals Google Scholar
Return to citationKieffer, J., Orlans, J., Coquelle, N., Debionne, S., Basu, S., Homs, A., Santoni, G. & De Sanctis, D. (2025). J. Appl. Cryst. 58, 138–153.  Web of Science CrossRef CAS IUCr Journals Google Scholar
Return to citationLu, I.-R., Görler, G., Fecht, H. & Willnecker, R. (2002). J. Non-Cryst. Solids 312–314, 547–551.  Web of Science CrossRef CAS Google Scholar
Return to citationMarshall, K. P., Emerich, H., McMonagle, C. J., Fuller, C. A., Dyadkin, V., Chernyshov, D. & van Beek, W. (2023). J. Synchrotron Rad. 30, 267–272.  Web of Science CrossRef IUCr Journals Google Scholar
Return to citationMartinelli, A., Baglioni, J., Sun, P., Dallari, F., Pineda, E., Duan, Y., Spitzbart-Silberer, T., Westermeier, F., Sprung, M. & Monaco, G. (2024). J. Synchrotron Rad. 31, 557–565.  Web of Science CrossRef CAS IUCr Journals Google Scholar
Return to citationMathot, V., Pyda, M., Pijpers, T., Vanden Poel, G., van de Kerkhof, E., van Herwaarden, S., van Herwaarden, F. & Leenaers, A. (2011). Thermochim. Acta 522, 36–45.  Web of Science CrossRef CAS Google Scholar
Return to citationMathot, V., van Herwaarden, S., Schick, C., van de Kerkhof, E., van Grinsven, P., Splinter, R. & van Wetten, I. (2023). The Handbook of Differential Scanning Calorimetry edited by J. D. Menczel & J. Grebowicz, pp. 221–308. Butterworth–Heinemann.  Google Scholar
Return to citationNarayanan, T., Wacklin, H., Konovalov, O. & Lund, R. (2017). Crystallogr. Rev. 23, 160–226.  Web of Science CrossRef CAS Google Scholar
Return to citationRosenthal, M., Doblas, D., Hernandez, J. J., Odarchenko, Y. I., Burghammer, M., Di Cola, E., Spitzer, D., Antipov, A. E., Aldoshin, L. S. & Ivanov, D. A. (2014). J. Synchrotron Rad. 21, 223–228.  Web of Science CrossRef CAS IUCr Journals Google Scholar
Return to citationRuta, B., Zontone, F., Chushkin, Y., Baldi, G., Pintori, G., Monaco, G., Rufflé, B. & Kob, W. (2017). Sci. Rep. 7, 3962.  Web of Science CrossRef PubMed Google Scholar
Return to citationVaughan, G. B. M., Baker, R., Barret, R., Bonnefoy, J., Buslaps, T., Checchia, S., Duran, D., Fihman, F., Got, P., Kieffer, J., Kimber, S. A. J., Martel, K., Morawe, C., Mottin, D., Papillon, E., Petitdemange, S., Vamvakeros, A., Vieux, J.-P. & Di Michiel, M. (2020). J. Synchrotron Rad. 27, 515–528.  Web of Science CrossRef CAS IUCr Journals Google Scholar
Return to citationYi, F. & LaVan, D. A. (2019). Appl. Phys. Rev. 6, 031302.  Web of Science CrossRef Google Scholar
Return to citationZhuravlev, E. & Schick, C. (2010). Thermochim. Acta 505, 1–13.  Web of Science CrossRef CAS Google Scholar

This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.