research papers
Multifunctional bending magnet beamline with a capillary optic for X-ray fluorescence studies of metals in tissue sections
aApplied Physics Program, Northwestern University, Evanston, IL 60208, USA, bDepartment of Microbiology, Genetics and Immunology, Michigan State University, East Lansing, USA, cDepartment of Chemistry, Michigan State University, East Lansing, MI 48824, USA, dDepartment of Physics and Astronomy, Northwestern University, Evanston, IL 60208, USA, eChemistry of Life Processes Institute, Northwestern University, Evanston, IL 60208, USA, fX-ray Science Division, Advanced Photon Source, Argonne National Laboratory, Lemont, IL 60439, USA, gDepartment of Biochemistry and Molecular Biology, Bloomberg School of Public Health, Johns Hopkins University, Baltimore, MD 21205, USA, and hElemental Health Institute, Michigan State University, East Lansing, MI 48824, USA
*Correspondence e-mail: [email protected], [email protected]
Scanning fluorescence X-ray microscopy lets one non-destructively and quantitatively map the distribution of most biologically important metals in cells and tissues. For studies on large-scale tissues and organs, a spatial resolution of several micrometres is often sufficient; in this case, bending magnets at synchrotron light sources provide abundant X-ray flux. We describe here the use of bending magnet beamline 8-BM-B at the Advanced Photon Source with two distinct microscopy stations: a pre-existing one with Kirkpatrick–Baez (KB) mirror optics for slightly higher throughput and the ability to accommodate samples tens of centimetres across, and a new prototype station with an axially symmetric, single-bounce, capillary optic with slightly less flux, but finer resolution at similar fluence per time. The KB station provides δres = 10.5 µm spatial resolution at a per-pixel exposure time of tdwell = 100 ms and a fluence per time of 5.8 × 107 photons µm−2 s−1, while the prototype capillary station provides δres = 6.5 µm at tdwell = 50 ms and a fluence per time of 5.6 × 107 photons µm−2 s−1. We used image power spectral density to estimate the achieved spatial resolution δres from individually acquired images, with δres depending on the optic, the fluorescence signal strength of the sample being imaged, and the method used to process raw fluorescence spectral data.
Keywords: X-ray microscopy; X-ray fluorescence; capillary optic; elemental analysis.
1. Introduction
Quantitative methods for mapping elemental distributions are essential for addressing a wide range of problems in the life sciences (McRae et al., 2009
; Zee et al., 2022
). To understand the function of metals in thin sections from tissues and organs, quantitative imaging at a spatial resolution of several micrometres can provide information on cell-to-cell metal variations, while a field of view of several millimetres allows one to image large, representative regions of organs from small animals. For elements with atomic numbers above about Z = 14, scanning fluorescence X-ray microscopy (SFXM) offers a very useful combination of high sensitivity with relatively low beam damage (Kirz, 1980
; Sparks, 1980
; Pushie et al., 2022
). In this approach, a small specimen is raster-scanned through an X-ray beam spot, and an energy-dispersive detector is used to record the emission at characteristic X-ray fluorescence lines (Horowitz & Howell, 1972
; Sparks, 1980
; Jones et al., 1984
). The recorded signal includes background from elastic and Compton X-ray scattering, as well as other factors such as incomplete charge collection in energy-dispersive detectors (Van Grieken & Markowicz, 2002
). A variety of analysis software packages can be used to separate signal from background, delivering quantitative measures of elemental concentration (Ryan, 2000
; Vogt, 2003
; Solé et al., 2007
; Schoonjans et al., 2012
; Schoonjans et al., 2013
; Crawford et al., 2019
).
Obtaining high resolution in SFXM requires an X-ray source providing significant flux within a small area with narrow solid angle; that is, a source with high spectral brightness (Jacobsen, 2020
). Emission from such a source can then be demagnified by an optic to a small spot through which the sample is scanned. For nanoscale imaging within single cells, undulator sources in straight sections of low-emittance storage rings provide the highest possible brightness (Eriksson et al., 2014
) outside of free-electron lasers. When evaluating overall elemental distributions in biological tissues or organs, micrometre-scale imaging can provide megapixel images of samples that are millimetres in size. In this case, the somewhat larger source size and divergence of bending magnet (dipole) sources at synchrotron light sources can provide quite satisfactory performance, often with easier access for imaging many samples.
We describe here the use of bending magnet beamline 8-BM-B at the Advanced Photon Source (APS) at Argonne National Laboratory, USA. This beamline has a pre-existing SFXM station that uses a Kirkpatrick–Baez (KB) mirror (Kirkpatrick & Baez, 1948
) originally designed for use at a separate synchrotron beamline. As part of a project for Quantitative Elemental Mapping for the Life Sciences (QE-MAP), which is an NIH-supported Biomedical Technology Research Resource based at Michigan State University, we have added a second prototype SFXM station just downstream. It is equipped with a capillary focusing optic for higher spatial resolution, and a fluorescence detector with greater signal collection capability. These two stations will be described in more detail in Section 1.2
.
1.1. Example: metals in a mouse kidney
As an example of SFXM capabilities provided at beamline 8-BM-B, Fig. 1
shows the distribution of three important metals in a mouse kidney partial section. This particular mouse was infected with Candida albicans clinical isolate SC5314 via tail vein injections as part of a larger study aimed at understanding how mammalian hosts attempt to starve invading pathogens of essential nutrient metals such as Fe, Cu, Zn, and Mn (Wildeman et al., 2023
). At 72 h post-infection, the kidney was extracted, embedded in an optimum cutting temperature (OCT) compound (Tissue-Tek), and frozen in an isopentane bath chilled with liquid N2-cooled isopentane. We cut the frozen block into sections of 10 µm thickness while using a cryostat (CM3050S, Leica) that maintained them at −15 to 18°C during the process. The sections were then transferred to Si3N4 chips (NX5200, Norcada) previously affixed to glass slides by Kapton tape. We stored the sections in airtight slide boxes in a −80°C freezer until retrieval for SFXM analysis. Upon thawing the original airtight slide boxes at room temperature, the chips containing the kidney sections from the glass slides were removed and placed onto a sample mount specifically designed for 8-BM-B (at which point most of the water in the sections evaporated). We scanned those sections at 8-BM-B and used the known areal mass concentrations of calcium, iron, and copper in an AXO 10X thin film standard (RF8-200-S2454, Applied X-ray Optics GmbH), which was scanned separately, to calibrate their metal concentrations. Two-dimensional elemental maps of each section were obtained by performing fluorescence spectrum fitting with MAPS (Vogt, 2003
) and M-BLANK (Crawford et al., 2019
) software as is described in Section 1.5
. In Fig. 1
(a), elemental maps of iron, calcium, and zinc are shown as obtained using the KB mirror station with a scan step size of Δx = Δy = 25 µm and a per-pixel exposure time tdwell = 50 ms. Also shown in Fig. 1
(a) is a three-color composite map of these three elements. The ring-like features in the Ca distribution indicate possible localized infection sites. Figs. 1
(b) and 1(c) show finer fields of view of localized Ca hotspots at different length scales using (b) the KB mirror with Δx = Δy = 5 µm and tdwell = 100 ms and (c) the capillary optics with Δx = 1 µm, Δy = 2 µm, and tdwell = 50 ms. As can be seen, the capillary optic station yields a higher resolution image.
|
|
Figure 1
Scanning fluorescence X-ray microscopy (SFXM) images elemental distributions at different length scales. Shown here are trace essential metals inside a sample of a mouse kidney partial section infected with Candida albicans, with Ca being strongly elevated at sites of infection. We obtained the images in (a) using the KB mirror scanning station at 8-BM with a 25 µm step size at tdwell = 50 ms. These metals are also displayed in a color composite map, allowing one to more easily see how the elements are differentially distributed in the specimen. The images in (b) and (c) correspond to regions we selected from (a) using the KB mirror and capillary optics, respectively. In (b), we used a Δx = Δy = 5 µm step size at tdwell = 100 ms. In (c), we used step sizes of Δx = 1 µm and Δy = 2 µm in y. The spatial resolution improved from 10.5 µm to 6.3 µm when transitioning from the KB optic to the capillary (as described in Section 2.2 |
1.2. Beamline 8-BM-B at the APS
Beamline 8-BM-B at the APS has had an evolving mission, with corresponding changes in its instrumentation. This bending magnet beamline was built in the mid-1990s as a general purpose beamline. Around 2010, it was repurposed for micrometre-scale X-ray microprobe studies. The beamline optical layout at the time of our experiments is shown in Fig. 2
. A double-multilayer monochromator (DMM) located 25.25 m from the source was used to deliver a spectral bandwidth of about Δλ/λ = 0.0109 at a photon energy of 10 keV. While our experiments were carried out at 10 keV photon energy, the beamline can operate over a range 9–18 keV. A set of secondary source apertures were located 48.15 m from the primary source, just inside the experimental enclosure (the `hutch'). A vertically deflecting, bendable toroidal mirror (originally designed for a different beamline) was located 30.9 m from the source; it delivered an intermediate focus spot located significantly upstream of the beam-defining aperture (BDA). Within the hutch, the secondary source apertures were located 0.2 m from the upstream end of a 2.4 m long optical table. At the table's downstream end was an X-ray beam camera consisting of a 1 mm thick cadmium tungstate scintillator (10 mm side length), a 5× microscope objective, and a visible light camera. This camera was used for optic and specimen alignment. The BDA consisted of a 250 µm diameter hole in a 0.25 mm thick platinum plate in front of a 1 mm diameter hole, which was in turn located inside a 12.7 mm thick aluminium plate. At the time of our measurements, the beam profile was roughly 1 mm × 2 mm at the input of the pinhole. There is the potential to enhance beamline performance by upgrading the toroidal mirror, and the source characteristics are being improved by the realization of the APS-Upgrade (APS-U) (Kerby, 2023
).
|
Figure 2
Layout of the 8-BM-B beamline when used with both a pre-existing KB mirror/large-area instrument setup (a) and a new, additional capillary/small-area instrument (b) located in the experimental enclosure. Both schematics show the location of the double multilayer monochromator (DMM), and the toroidal mirror intended to image the X-ray beam from the storage ring to the exit slit position. The beam size is then set using a pinhole as a beam-defining aperture (BDA). The KB mirror setup is for scanning sample fields up to several centimetres across at about δres = 10.5 µm spatial resolution, and the capillary optic setup is for δres = 6.5 µm spatial resolution over smaller fields of view (Section 2.1 |
This beamline already featured a large-scanning-area SFXM setup employing a KB focusing optic mounted in air, with a helium enclosure for the specimen and the detector window. The sample scanning setup used with the KB optic had a large working distance of about 50 mm and a maximum scanning field of view of 250 mm and 150 mm in the horizontal and vertical directions, respectively. To allow for clearance for very large specimens and good X-ray fluorescence signal collection efficiency, the sample was mounted at 45° relative to the incident beam normal. The system was equipped with a four-element silicon drift detector (267-VTX-ME4, Hitachi) with xMAP (XIA) detector readout electronics. Because of the imperfect match between 8-BM-B beamline optics and the KB mirror, the achieved resolution was reduced, as will be seen in Section 2.1
. Nevertheless, the KB scan system had proven to be quite successful in studies involving large-area samples (Kirker et al., 2017
; Zelinka et al., 2018
; Copeland-Hardin et al., 2023
; Broda et al., 2024
).
As noted in Section 1
, we enhanced the beamline by adding a second SFXM setup just downstream of the KB setup. This second scanning station, which is presently in prototype status rather than in final status, uses an axially symmetric, single-bounce, capillary focusing optic (Sigray, Inc.) designed to deliver a smaller focus with the optical layout of the 8-BM beamline. This glass capillary was pulled under heating conditions by Sigray with the goal of imaging a source located 1.982 m upstream to a focus located 45 mm downstream of the capillary center while maintaining a working distance of 20 mm. The capillary has an entrance diameter of 0.785 mm and an exit diameter of 0.425 mm; with a 30 nm coating of Pt, it is designed to work at photon energies up to 20 keV with no lower photon energy bound. We evaluated the performance of this optic using the 28-ID-B undulator beamline at the APS so as to provide a well collimated X-ray source; with this source, the optic produced a focal spot with a full width at half-maximum (FWHM) size of δres,x = 2.9 µm and δres,y = 3.0 µm at E = 13 keV (see Section S1 of the supporting information for more details).
The capillary optic and its supplied mount (Sigray, Inc.) is placed on a two-angle tilt stage (8807, New Focus) for angular alignment to the incident beam, with a two-axis Aerotech ATS50 stepper motor stage stack used for optic translation transverse to the beam direction. The alignment of the capillary was adjusted and evaluated by imaging a test pattern. Following the capillary is a micrometre-precision, 50 mm × 50 mm maximum field of view xyz scanning stage assembly (x and y stages: ATS50-50-U-NC, Aerotech; z stage: UTM50PP1HL, Newport), with a scanning direction oriented at 15° relative to the incident beam normal (to reduce horizontal beam spreading on the sample; see Fig. 3
). Fluorescence X-ray detection is provided by a seven-element energy-dispersive detector (HHS5700-VTX-ME7, Hitachi) with electronics capable of handling higher count rates (Xspress3X, Quantum Detectors). Because of the smaller working distance of this optic and the 15° scan direction, this setup is better suited to use with samples with no upstream clearance problems. Sections from biological tissues and organs usually satisfy these requirements. With the shorter working distance of the capillary optic, the sample and detector were in air with no helium chamber. We did not conduct any mechanical vibration evaluations; however, at few-micrometre spatial resolution, no vibration effects were observed in images.
|
Figure 3
Diagram of how a planar sample and its scanning stage are typically mounted at an angle θ relative to normal incidence for scanning X-ray fluorescence microscopy. This means that the beam width Wbeam as seen by a sample at normal incidence is broadened to |
With these two SFXM stations, the beamline 8-BM now has two scanning options. One is for larger samples at modest resolution, but with somewhat higher flux, using the KB optic with larger working distance. The other is for smaller samples using the capillary optic with reduced working distance and flux, but higher spatial resolution. In Section 3
, we will see that the fluence per time (photons per area per time) is similar between the two scanning stations. The capillary station is presently in prototype status, with further improvements planned. Biomedical and other users are able to choose which of these two scanning stations are best suited to their research interests without significant downtime for reconfiguring the SFXM system. This is quite useful at a facility where nanofocusing beamlines (which are often oversubscribed) prioritize users needing to collect data on biological samples at submicrometre resolution.
The achieved spatial resolution in SFXM depends not only on the properties of the optic but also on the signal and background level for detecting particular chemical elements. While this has been illustrated previously for SFXM imaging (Deng et al., 2015
) and image deconvolution (Deng et al., 2017
), in Section 2
we explore in greater detail the nature of spatial resolution determination using power spectral densities from X-ray fluorescence images. It is shown there that the spatial resolution at a given signal-to-noise ratio (SNR) shows some dependence on how one processes the signal and background.
1.3. Absolute photon flux in the two scanning stations
One aspect of characterizing the two SFXM stations was to measure the absolute photon flux Φ(E). The APS ran in top-up mode with a constant 100 mA of electron beam current in the storage ring, and we adjusted the beamline optics to maximize the flux observed via an ion chamber just downstream of the beam-defining pinhole at 48.15 m from the source. We carried out absolute photon flux measurements at 10 keV photon energy using a calibrated silicon PIN photodiode (S3590-06, Hamamatsu) immediately downstream of each optic's focus. These signals were fed into a low-noise current preamplifier (SR570, Stanford Instruments); the measurement scale was confirmed with a voltmeter while the signal was passed to a voltage-to-frequency converter (N101VTF, NOVA) for time-dependent measurements using 8-BM-B's data acquisition system. To get the most reliable readings, the gain sensitivity gs of the preamplifier was tuned to 2 µA V−1 for the KB mirror and 1 µA V−1 for the capillary. For this particular photodiode, 1 pA through the device corresponded to about 2382 photons s−1 at E = 10 keV, giving a photon-flux-to-photocurrent conversion factor of K(E) = 2382 photons pA−1 s−1. Since the dark current voltage was less than 1% of the beam signal levels, we were able to determine the absolute incident photon flux via
In this way, we found that the KB mirror station, at E = 10 keV, delivered a focused flux of Φ = 2.1 × 1010 photons s−1, while the capillary optic delivered a flux of Φ = 7.7 × 109 photons s−1 (in each case, the absolute uncertainty was estimated to be approximately ±5% based on slow variations in the measurements). While for diffraction-limited optics there is a tradeoff between smaller focus and less flux [see, for example, Section 4.4.6 of Jacobsen (2020
)], in this case we expected that the lower flux from the capillary optic was due to its lower aperture with the non-optimized beamline optics as existed for 8-BM (see Section 1.2
).
1.4. Fluorescence data collection
While there are a variety of sample orientations and detector configurations that can be used in SFXM (Sparks, 1980
; Ryan et al., 2010
; Sun et al., 2015
), most experiments mount the X-ray fluorescence detector at 90° to the incident beam direction so as to minimize elastic and inelastic scattering from horizontally polarized beams at synchrotron light sources (Dzubay et al., 1974
). In addition, the sample and scanning stage motions are typically inclined to the normal of the incident beam by 15° or 45° for a compromise between minimizing X-ray fluorescence self-absorption in planar samples while not introducing too much effective beam broadening in the scanning direction, as illustrated in Fig. 3
. The effective beam width along the scanning stage direction, relative to beam width Wbeam on a sample oriented at normal incidence to the beam, is given by
As noted above, we used θ = 15° and θ = 45° for the capillary and KB optics, respectively.
All data shown here were taken with continuous scanning, or `fly scan' mode, where the specimen was moved at near-constant velocity along the fast axis of the scan raster, with data recording advanced over position increments. With the KB scanning system, we frequently examined a large sample area using Δx = Δy = 25 µm, and then acquired higher resolution scans using Δx = Δy = 5–25 µm. With the capillary optic, Δx = Δy = 1–20 µm was typically used. In both cases, per-pixel imaging times of tdwell = 20–100 ms were chosen. Both scanning stations have multi-element fluorescence detectors with collimators to minimize the sensitivity to scattering and stray fluorescence from materials other than the sample area in the beam focus. The sum of the signal from all active detector elements was used for data processing. With an incident X-ray beam energy of 10 keV, we were able to estimate the spatially resolved concentrations of biologically important elements including phosphorus, sulfur, potassium, calcium, iron, nickel, copper, and zinc.
To determine the distance from the sample to the X-ray fluorescence detector and thus determine the effective solid angle of collection Ωeff for each detector, we collected average fluorescence intensities Iavg(Δd) at three different detector displacements Δd of 0, 3, and 6 mm along the detection axis. Because the solid angle, and therefore the detected intensity, should vary according to an inverse square law, we were able to determine the effective sample-to-detector plane distance deff using
where α is a fitting constant. The strong calcium fluorescence signal from the sample shown in Fig. 1
was used as the signal, and we obtained effective detector distances of deff = 31.2 mm for the KB system and deff = 14.4 mm in the capillary system. Only Nd = 3 of the four detector elements in the KB system's detector were functioning, with each element n having active area Aact,n = Aact = 42.5 mm2. With the Nd = 7 element detector used in the capillary system, the active area for each detector element was 40 mm2 as limited by the collimator system. From these values, we determined the effective solid angles of collection Ωeff via
Using this approach, Ωeff = 0.13 sr for the KB system, and Ωeff = 1.35 sr for the capillary system. While these values represent the effective solid angles based on the distances deff determined from equation (3)
, the actual distances to the detector center involve considering the distances to each detector element their corresponding obliquity factors (see Section S3 of the supporting Information for more details).
For ease of data collection, we developed a graphical user interface (GUI) for the 8-BM EPICS control system used at the APS. This allowed the user to set up a batch of scans for acquisition. The software prompted the user for scan parameters such as scan height and width about a desired center position, pixel size, and the per-pixel exposure time (dwell time). The GUI then calculated the number of pixels per line, and number of scan rows. The software then verified that the parameters for each scan did not exceed motor position or velocity limits, after which it returned an estimated completion time. At the start of each scan in the queue, the software would reset the acquisition hardware and ready the controls hardware such that every nth motor pulse produced a single trigger event based on the ratio of the desired pixel size to the motor's minimum step size. With the XIA xMAP fluorescence detector readout system for the KB mirror scanning station, the full fluorescence spectrum from each element was acquired at each pixel interval. The Xspress3X detector readout electronics used with the capillary optic scanning station provided additional functionality, which required modifications to the data acquisition software. The Xspress3X system was able to internally store full fluorescence spectra for each detector element at each pixel in one scan line, with transfer to the scan control system at the end of each scan line. It also included parameter import/export functionality, event logging, parameter validation, and periodic visual updates indicating that scans were progressing. We calibrated the Xspress3X system as well to ensure accurate and consistent measurement across each of the detector elements and electronics channels prior to commissioning.
1.5. Fluorescence data processing
Energy-dispersive X-ray detectors collect all the charge from single above-threshold photon detection events and calculate the photon energy using the electron–hole separation energy of the semiconductor detection surface (for silicon-based detectors, that energy is 3.65 eV) (Shockley, 1961
; Lowe & Sareen, 2007
; Mazziotta, 2008
). However, the collected signal also includes photons that are elastically and inelastically scattered from the sample and also from instrument materials, incomplete charge detection, and escape peaks in the detector material (Van Grieken & Markowicz, 2002
). Therefore, a variety of analysis programs have been developed to process the as-recorded signal, separating X-ray fluorescence photon events from the background (Ryan, 2000
; Vogt, 2003
; Solé et al., 2007
; Schoonjans et al., 2012
; Schoonjans et al., 2013
; Crawford et al., 2019
). These programs return a map of concentration of each element Z in units of mass per area. In our case, we used two programs: MAPS (Vogt, 2003
) and M-BLANK (Crawford et al., 2019
). MAPS offers both region-of-interest (ROI) selection (sometimes called `spectral binning'), and full-spectrum fitting algorithms. M-BLANK offers full-spectrum fitting, while ROI selection is still used in some cases for expediency. Because background subtraction affects both signal and noise levels, one obtains somewhat different resolution estimates for ROI versus full-spectrum fitting when using the power spectral density method described in Section 2
. This is discussed in Section S4 of the supporting information.
When obtaining elemental concentration maps using background subtraction, pixels that are absent of any concentration of a particular element will have Poisson fluctuations in the number of detected fluorescent photons. If the background is measured using strong photon statistics, subtraction of this high-statistics background from low-statistics per-pixel fluorescence measurements when there is little to no elemental content is likely to produce concentration maps
with values fluctuating on either side of zero. This is in fact the case when using analysis programs like M-BLANK (Crawford et al., 2019
).
1.5.1. Limit of detection
Careful determination of limits of detection requires simulation studies of the behavior of full-spectrum fitting methods (like the one employed by M-BLANK) with known low- and zero-concentration pixels and the incorporation of Poisson statistics, as well as choices for false-positive and false-negative tolerances (Currie, 1968
). All of that is well beyond the scope of the present study. Instead, we took an empirical approach where we examined a histogram of low-calculated-Ca mass concentration pixels from the kidney partial section shown in Fig. 1
(c) (see Section 2.1
for why we chose to look at the Ca map), as well as a histogram of an experimental scan we performed of an empty Si3N4 window presumably free of calcium (at least as far as this comparison is concerned). Because the Si3N4 scan utilized a dwell time of tdwell = 200 ms as opposed to tdwell = 50 ms as used for the kidney specimen scan, we expanded the width of the Si3N4 window scan by a factor of (200 ms/50 ms)1/2 = 2 to correct for differing photon statistics. The resulting histograms shown in Fig. 4
indicated that at a concentration of = 0.05 µg cm-2, we had very few `false-positive' pixels in the empty Si3N4 scan; we therefore used this as an estimate of the minimum detectable mass concentration of Ca in our example measurement for tdwell = 50 ms (thus, a fluence of
= 2.8 × 106 photons µm−2; see Section 3
for more information).
|
Figure 4
Limit of detection estimation at 8-BM using the capillary optic. We inspected histograms of low-mass concentration pixels |
2. Resolution determination via power spectral density
One commonly used method for evaluating the spatial resolution of images is to use 3D Fourier shell correlations (or 2D Fourier ring correlations) between two independent images of the same specimen as a function of spatial frequency u = (Saxton & Baumeister, 1982
; Heel & Schatz, 2005
). However, expediency in data collection, as well as radiation dose minimization, make it useful to evaluate spatial resolution from single images. With low-exposure images, this can be carried out using power spectral density (PSD) analysis. In two dimensions, we calculated the PSD S(ux, uy) for each element using
where the spatial frequency representation Ψ(ux, uy) is found in transmission imaging from the complex specimen transmittance ψ(x, y) via a Fourier transform, or
In the discrete Fourier transform with an even number of pixels Nx, the spatial frequency ux runs from −1/(2Δx) to +bx/(2Δx), with bx = (Nx/2 − 1)/(Nx/2) (a similar result holds for uy). When processing images with an odd number of pixels, each image was padded with a row and/or column of the average low signal level. With the exception of the approach described in Section 2.3
, we show azimuthally averaged power spectral densities S(ur), where
ur =
(ux2+uy2)1/2 is the radial spatial frequency. We also binned ur to smaller numbers of pixels to obtain smoother representations.
For objects with no phase contrast, the complex transmittance ψ(x, y) can be found from the square root of image intensity, or ψ(x, y) = [I(x, y)]1/2. Fluorescence is an incoherent process depending only on sample absorbance, and furthermore the fluorescence spectrum is analyzed to yield a concentration map per element Z of as described in Section 1.5
. Normally, we have ∝ I(x, y), so we should use
in equations (5)
and (6)
for power spectral density analysis. However, as noted in Section 1.5
, pixels that are absent of any concentration of a particular element will have Poisson fluctuations in the number of detected fluorescent photons, which should lead to fluctuations of about a value of zero due to subtraction of a background measured with strong photon statistics. The presence of some pixels in
having weak negative values in fact arises when using analysis programs like M-BLANK (Crawford et al., 2019
). Therefore, for PSD analysis using equations (5)
and (6)
, we used ψ(x, y) = for pixels with positive values of
and ψ(x, y) =
for pixels with negative values of
.
When shown on a logarithmic scale for power spectral density versus spatial frequency, S(ur) typically declines linearly with ur for many types of images (Porod, 1982
; Jacobsen, 2020
). This corresponds to a signal trend of
where a < 0. When images are obtained using a finite number of photons with associated Poisson noise statistics, it is common for S(ur) to decrease until it reaches a constant `noise floor' value Snf. This is due to Poisson fluctuations being uncorrelated between pixels. As a result, Poisson noise is like a delta function in real space with a `flat' power spectral density that is uniform over all pixels in Fourier space. One can exploit these trends in signal and noise for Wiener noise suppression in Fourier plane representations of images, where the Wiener filter W(ur) is given by (Press et al., 1986
)
Because W(ur) ≃ 1 at low spatial frequencies where S(ur) dominates, and because W(ur) ≃ 0 at high spatial frequencies where the noise floor Snf dominates, it is only in the approximate `knee' region ∼uknee [where S(ur) ≳ Snf] that one requires an estimate of the trend of S(ur) for Wiener filtering. In this spirit, one can use a fit of
to find the power law decrease slope a for the signal. That is, one can use data points from spatial frequencies ur somewhat lower than the approximate `knee' ∼uknee to find the signal trend and thus determine P and a. One can also use the average of data points at spatial frequencies ur above the approximate `knee' ∼uknee to find
With P, a, and Snf now determined, the actual spatial frequency uknee of the `knee' is the value of ur for which Pur a = Snf, or
For evaluating the effective spatial resolution of an image, it is best not to have the signal be equal to the noise (the condition of the `knee') but to have the signal be higher than the noise. It is common to use the Rose criterion (Rose, 1946
) of requiring a signal-to-noise ratio of SNRres = 5 for high-quality images (lower SNR values are often deemed acceptable in specific contexts). Thus, we can obtain an estimate of the maximum spatial frequency ures corresponding to a high-quality image from the case where the resolution-defining power spectral density is a multiple SNRres of the noise floor Snf, or
This leads to a spatial resolution estimate of
based only upon the power spectral density of a single image.
We developed a Python computer program (using the PyQt graphical user interface) called PyXRFPower to carry out power spectral density analysis on X-ray fluorescence images. The program allows one to either estimate the spatial resolution assuming a symmetric beam profile or separate resolution in the horizontal and vertical directions (Section 2.3
). After reading in a scanning fluorescence X-ray microscope dataset, one can select images corresponding to emission at one or several elemental X-ray fluorescence lines and carry out power spectral density fitting separately for each element. The selection of the data points to use for the signal trend S(ur) [equation (8)
] and noise floor Snf [equation (11)
] is accomplished by dragging a cursor through spatial frequency regions on either side of the approximate `knee' ∼uknee. Spatial frequencies ures [equation (13)
] resulting from a user-defined value of SNRres and corresponding spatial resolutions are reported alongside data trend fit slopes a as provided by equation (8)
. PyXRFPower can currently be found on Github at https://github.com/bwr0835/pyxrfpower.
Using the approaches described above, we characterized the scanning fluorescence X-ray microscopy performance of both the KB mirror and capillary optic experimental stations. For the KB mirror, we acquired images with scan step sizes of Δx = Δy = 5 µm and per-pixel acquisition times of tdwell = 25 ms (97.7% live time) and 100 ms (99.6% live time). The scan step size for the capillary optic was Δx = 1 µm and Δy = 2 µm, and tdwell = 20 ms (100.0% live time) and 50 ms (99.6% live time). Unless stated otherwise, results shown below were obtained using the M-BLANK program (Crawford et al., 2019
) for fluorescence spectrum analysis as described in Section 1.5
.
2.1. Spatial resolution versus fluorescing element
As noted in Section 1
, spatial resolution actually achieved in an image depends on the optic but also on the signal-to-noise characteristics of the X-ray fluorescence image (Deng et al., 2015
; Deng et al., 2017
). We therefore evaluated SFXM images separately for each element using the methods described in Section 2
to find the power spectral density trends for the signal
S(ur) ≃
Pur a of equation (8)
and the noise floor Snf of equation (11)
. From that, spatial resolution estimates δres of equation (14)
for a signal-to-noise ratio of SNRres = 5 were obtained.
For the capillary optic, we imaged the specimen described in Section 1.1
with 339 × 242 pixels of size Δx = 1 µm and Δy = 2 µm at a per-pixel imaging time of tdwell = 50 ms [see Fig. S2(a) of Section S2 in the supporting information for the integrated spectrum we collected using this optic]. Of all the elemental images obtained from this sample, the calcium image shown in Fig. 5
(a), taken from Fig. 1
(c), appeared to have the most favorable signal-to-noise ratio. With a stronger signal, one can come close to the optic-limited resolution that would be observed at low noise conditions. Therefore, we carried out power spectral density analyses on that Ca image, along with images of Fe and Zn fluorescence. The results displayed in Fig. 5
(b) show resolution results of δres = 6.5 µm for Ca, δres = 14.4 µm for Fe, and δres = 19.1 µm for Zn [see Fig. S2(b) in Section S2 of the supporting information for the Fe and Zn maps alongside the Ca map]. The spatial resolution estimate for the Ca image was in good agreement with a rough estimation obtained using images of a 1951 United States Air Force (USAF) test pattern (R3L3S1P, Thorlabs).
|
|
Figure 5
Image of a Candida albicans-infected mouse kidney partial section obtained using the capillary optic scan station and a per-pixel scan time of tdwell = 50 ms. At left (a) is the calcium fluorescence image from Fig. 1 |
We also scanned a larger area of a region in the same specimen, containing most of the capillary field of view, using the KB mirror scanning station. This scan was over 187 × 75 pixels of size Δx = Δy = 5 µm with a per-pixel imaging time of tdwell = 100 ms. This pixel size was just small enough for meeting the Nyquist criterion, and Fig. S4 in Section S4 of the supporting information shows that it was small enough to allow the noise floor to be seen. Using the same analysis approach as shown in Fig. 5
, we obtained resolution estimates of δres = 10.5 µm for Ca, δres = 12.8 µm for Fe, and δres = 12.9 µm for Zn. The weaker Fe and Zn signals came closer to the optic limit in the longer-dwell time KB station measurement than they did in the shorter-dwell time capillary station measurement (which, again, was a different field of view that had smaller concentrations of those two elements), illustrating the point that the achieved spatial resolution depends both on the optic and on signal levels.
When spatial resolution is set by diffraction from an optic's aperture, the depth of focus is approximated by DOF = (Jacobsen, 2020
). In that case, focusing to δres = 6.5 µm by the capillary optic would correspond to DOF = 1.8 mm, while δres = 10.5 µm would give DOF = 6.8 mm for the KB mirror. However, we did not test for the actual depth of focus in experiments.
2.2. Achieved resolution versus scan time
As one increases the per-pixel exposure time tdwell, the signal S(ur) of equation (8)
should increase relative to the noise floor Snf of equation (11)
, meaning that the obtained spatial resolution δres of equation (14)
should improve as one approaches the high-signal-level limit of the optic. For the Ca image obtained with the KB mirror optic, the resolution at a shortened per-pixel time of tdwell = 25 ms was δres = 15.6 µm, corresponding to about a 52% worsening of spatial resolution compared with the result of δres = 10.2 µm at tdwell = 100 ms. Because the spatial resolution limit involves both the intrinsic resolution of the optic and the signal-to-noise ratio, it does not scale simply with exposure time alone. For the Ca image obtained with the capillary optic, decreasing the per-pixel exposure time to tdwell = 20 ms gave δres = 9.7 µm compared with δres = 6.5 µm at tdwell = 50 ms, or about a 32% worsening of spatial resolution. These results clearly show how the achieved spatial resolution depends strongly on exposure time for photon statistics-limited images such as are obtained in scanning fluorescence X-ray microscopy of intrinsic concentrations of metals in biological tissues. Fig. 6
shows the two S(ur) profiles we obtained for the capillary at the two values of tdwell corresponding to that optic.
|
Figure 6
Comparison of isotropic capillary spatial resolutions δres when increasing per-pixel exposure time tdwell. The resolution improved from δres = 9.7 µm at tdwell = 20 ms to δres = 6.5 µm at tdwell = 50 ms. The incident photon fluence |
2.3. Non-azimuthally symmetric spatial resolution estimate
Up to this point, we have assumed that the spatial resolution is the same in the horizontal and vertical directions. However, it is not uncommon at synchrotron light source beamlines to have slight spatial resolution asymmetries due to both beamline optics, and tilt of the sample and scanning stage motion as shown in Fig. 3
. One can therefore modify the above analysis so that it is carried out in two distinct azimuthal angle ranges as shown in Fig. 7
, with azimuthal angle ranges of ±30° about each direction representing one reasonable choice. This allows for resolution estimates of δres,x and δres,y, respectively.
|
Figure 7
When images have asymmetry in their spatial resolution, one can define azimuthal angle ranges about the x and y axes (shown here as blue and red regions, respectively) to obtain separate estimates of spatial resolution of δres,x and δres,y. We used azimuthal angle ranges of ±30° about the x and y axes as a compromise between the mixing of ux and uy components and having too few `spatial frequency pixels' in each Fourier direction for each circle being averaged over. |
For the Ca image shown in Fig. 5
(a) for which the azimuthally averaged resolution was δres = 6.5 µm, we used the approach shown in Fig. 7
to obtain separate estimates for the spatial resolution in the x and y directions, yielding δres,x = 6.9 µm and δres,y = 6.4 µm as shown in Fig. 8
. Because the maximum value of uy decreased by a factor of two due to using Δy = 2 µm step size (versus Δx = 1 µm), Sx(ur) did not extend out to the same value of ur as Sy(ur). However, in both cases the step size was smaller than half the spatial resolution, thus satisfying the Nyquist sampling criterion. For the KB mirror, where the azimuthally averaged Ca image resolution was δres = 10.5 µm, the resolution in the two directions was δres,x = 9.5 µm and δres,y = 12.1 µm.
|
Figure 8
Using the same Ca image as shown in Fig. 5 |
2.4. Sensitivity in resolution estimation due to selection of data trend points
Our method of evaluating spatial resolution using power spectral density in photon-limited images relies on fits of the `signal' region of the power spectral density using
S(ur) ≃
Pur a from equation (8)
, and the `noise' floor Snf from equation (11)
. As noted above, the `signal' points should be selected over some range of spatial frequencies below the estimated `knee' ∼uknee, while the `noise' floor should be obtained by averaging points above ∼uknee. In order to test the sensitivity of the spatial frequency ures of equation (13)
and the resulting resolution estimate δres of equation (14)
to user selections, we show the results for three different selections of `signal' spatial frequencies and resulting values of slope a and resolution δres in Fig. 9
. For the Ca image of Fig. 5
(a), we obtained slopes of a = −5.23, a = −6.02, and a = −6.59 with respective resolution estimates of δres = 6.51 µm, δres = 6.52 µm, and δres = 6.65 µm. Even though the slope estimates a were quite different from each other, the spatial resolution estimates were consistent within a range of 2.2%. For the KB mirror image with tdwell = 100 ms, a similar analysis yielded δres = 10.30 µm, δres = 10.49 µm, and δres = 10.26 µm, again consistent with a spatial resolution estimate range of 2.2% (the respective slopes were a = −4.65, a = −5.47, and a = −6.32). We therefore infer that our approach allows one to estimate the achieved spatial resolution in these X-ray fluorescence images with a reproducibility of better than 3%.
|
Figure 9
Test of the dependence on the estimated spatial resolution δres based on three different selections of the `signal' trend points to include in the fit of S(ur) |
3. Fluence per time on samples
The achievable spatial resolution in low-photon statistics images is limited in part by the fluence on the sample, or the cumulative photons per area (Schropp & Schroer, 2010
; Deng et al., 2015
; Deng et al., 2017
; Jacobsen, 2020
), as shown in Fig. 6
. We approximated the area of the probe Abeam as
since, for a symmetric Airy probe, the spatial resolution is equal to the radius associated with the first minimum. With that in mind, we can use the absolute photon fluxes measured in Section 1.3
and the best resolution as obtained from Ca images (Section 2.3
) to estimate the fluence per time as
Utilizing this approach, fluence per time estimates of = 5.8 × 107 photons µm−2 s−1 ± 6.6% for the KB mirror optic and
= 5.6 × 107 photons µm−2 s−1 ± 6.6% for the capillary were obtained.
4. Conclusion
We have modified the 8-BM-B beamline at the Advanced Photon Source (APS) at Argonne National Laboratory to provide two different setups for scanning fluorescence X-ray microscopy (SFXM) studies of intrinsic metals in biological tissue sections. The KB mirror station provides a spatial resolution of 10.5 µm (best used for larger area scans), while the prototype capillary optic station provides an improved spatial resolution of 6.5 µm as shown in Section 2.1
. We have described the use of power spectral density analysis of single images to obtain spatial resolution estimates, and we have shown how the achieved spatial resolution is affected by signal strength from different fluorescing elements.
The results reported in this work were for the beamline as it existed right before the shutdown of the original APS storage ring in April 2023. The upgrade of the APS (APS-U) (Kerby, 2023
) should offer a slight improvement in the brightness of bending magnet sources (and 75–100× improvement for undulator sources). Further gains might be possible by addressing limitations in the beamline toroidal mirror discussed in Section 1.2
. Therefore, this beamline should be an even more valuable resource for future studies of elemental distribution in biological tissues, aiding user communities including that of the NIH-supported QE-MAP center at Michigan State University.
5. Related literature
The following reference, not cited in the main body of the paper, has been cited in the supporting information: Qiao et al. (2020
).
Supporting information
Raw spectral data and additional elemental maps for the capillary, capillary beam spot size evaluation at beamline 32-ID at the Advanced Photon Source, detector obliquity and solid angle, and effect of x-ray fluorescence spectrum fitting method on spatial resolution. DOI: https://doi.org/10.1107/S1600577526000925/tol5019sup1.pdf
Footnotes
‡These authors contributed equally to this work.
§Present address: Department of Medicine, Massachusetts General Hospital, Boston, MA 02115, USA.
¶Present address: Department of Biochemistry and Molecular Biology, Robert Wood Johnson Medical School, Rutgers University, Piscataway, NJ 08854, USA.
Acknowledgements
We thank Professor Brendan Cormack (School of Medicine, Johns Hopkins University) for contributions in mouse infection studies.
Conflict of interest
We report no conflicts of interest.
Data availability
Elemental concentration maps as fitted and mass-calibrated by M-BLANK are available on Arch at https://doi.org/10.21985/n2-cx5v-f797. Any spectra of interest are available upon request.
Funding information
Our work was funded by NIGMS at NIH under both the QE-MAP award P41GM135018 to Thomas O'Halloran and the awards R35GM136644 and R21AI54726 to Valeria Culotta. Our research also used resources provided by the Advanced Photon Source, a US Department of Energy (DOE) Office of Science User Facility at Argonne National Laboratory supported by studies performed under the US DOE Office of Science-Basic Energy Sciences Contract No. DE-AC02-06CH11357.
References
Broda, M., Jakes, J. E., Li, L., Antipova, O. A., Maxey, E. & Jin, Q. (2024). Wood Sci. Technol. 58, 649–675.
CrossRef
CAS
PubMed
Google Scholar
Copeland-Hardin, L., Paunesku, T., Murley, J. S., Crentsil, J., Antipova, O., Li, L., Maxey, E., Jin, Q., Hooper, D., Lai, B., Chen, S. & Woloschak, G. E. (2023). Sci. Rep. 13, 7806–7815.
CAS
PubMed
Google Scholar
Crawford, A. M., Deb, A. & Penner-Hahn, J. E. (2019). J. Synchrotron Rad. 26, 497–503.
Web of Science
CrossRef
IUCr Journals
Google Scholar
Currie, L. A. (1968). Anal. Chem. 40, 586–593.
CrossRef
CAS
Web of Science
Google Scholar
Deng, J., Vine, D. J., Chen, S., Jin, Q., Nashed, Y. S. G., Peterka, T., Vogt, S. & Jacobsen, C. (2017). Sci. Rep. 7, 445.
Web of Science
CrossRef
PubMed
Google Scholar
Deng, J., Vine, D. J., Chen, S., Nashed, Y. S. G., Peterka, T., Ross, R., Vogt, S. & Jacobsen, C. (2015). Proc. SPIE 9592, 95920U.
Google Scholar
Dzubay, T. G., Jarrett, B. V. & Jaklevic, J. M. (1974). Nucl. Instrum. Methods 115, 297–299.
CrossRef
CAS
Google Scholar
Eriksson, M., van der Veen, J. F. & Quitmann, C. (2014). J. Synchrotron Rad. 21, 837–842.
Web of Science
CrossRef
CAS
IUCr Journals
Google Scholar
Heel, M. van & Schatz, M. (2005). J. Struct. Biol. 151, 250–262.
Web of Science
PubMed
Google Scholar
Horowitz, P. & Howell, J. A. (1972). Science 178, 608–611.
CrossRef
CAS
PubMed
Google Scholar
Jacobsen, C. (2020). X-ray Microscopy. Cambridge: Cambridge University Press.
Google Scholar
Jones, K. W., Gordon, B. M., Hanson, A. L., Hastings, J. B., Howells, M. R., Kraner, H. W. & Chen, J. R. (1984). Nucl. Instrum. Methods Phys. Res. B 3, 225–231.
CrossRef
Google Scholar
Kerby, J. (2023). Synchrotron Radiat. News 36(4), 26–27.
CrossRef
Google Scholar
Kirker, G., Zelinka, S., Gleber, S.-C., Vine, D. J., Finney, L. A., Chen, S., Hong, Y. P., Uyarte, O., Vogt, S., Jellison, J., Goodell, B. & Jakes, J. E. (2017). Sci. Rep. 7, 41798.
CrossRef
PubMed
Google Scholar
Kirkpatrick, P. & Baez, A. V. (1948). J. Opt. Soc. Am. 38, 766–774.
CrossRef
PubMed
CAS
Web of Science
Google Scholar
Kirz, J. (1980). Scanning Electron Microscopy, Vol. 2, pp. 239–249. Chicago: SEM Inc.
Google Scholar
Lowe, B. G. & Sareen, R. A. (2007). Nucl. Instrum. Methods Phys. Res. A 576, 367–370.
CrossRef
CAS
Google Scholar
Mazziotta, M. N. (2008). Nucl. Instrum. Methods Phys. Res. A 584, 436–439.
CrossRef
CAS
Google Scholar
McRae, R., Bagchi, P., Sumalekshmy, S. & Fahrni, C. J. (2009). Chem. Rev. 109, 4780–4827.
Web of Science
CrossRef
PubMed
CAS
Google Scholar
Porod, G. (1982). Small Angle X-ray Scattering, edited by O. Glatter & O. Krattky, ch. 2, pp. 17–51. London: Academic Press.
Google Scholar
Press, W. H., Teukolsky, S. A., Vetterling, W. T. & Flannery, B. P. (1986). Numerical Recipes: The Art of Scientific Computing. Cambridge University Press.
Google Scholar
Pushie, M. J., Sylvain, N. J., Hou, H., Hackett, M. J., Kelly, M. E. & Webb, S. M. (2022). Metallomics, 14, mfac032.
CrossRef
PubMed
Google Scholar
Qiao, Z., Shi, X., Kenesei, P., Last, A., Assoufid, L. & Islam, Z. (2020). Rev. Sci. Instrum. 91, 113703.
CrossRef
PubMed
Google Scholar
Rose, A. (1946). J. Soc. Motion Picture Eng. 47, 273–294.
CrossRef
Google Scholar
Ryan, C. G. (2000). Int. J. Imaging Syst. Tech. 11, 219–230.
CrossRef
Google Scholar
Ryan, C. G., Kirkham, R., Hough, R. M., Moorhead, G., Siddons, D. P., de Jonge, M. D., Paterson, D. J., De Geronimo, G., Howard, D. L. & Cleverley, J. S. (2010). Nucl. Instrum. Methods Phys. Res. A 619, 37–43.
CrossRef
CAS
Google Scholar
Saxton, W. O. & Baumeister, W. (1982). J. Microsc. 127, 127–138.
CrossRef
CAS
PubMed
Web of Science
Google Scholar
Schoonjans, T., Solé, V. A., Vincze, L., Sanchez del Rio, M., Appel, K. & Ferrero, C. (2013). At. Spectrosc. 82, 36–41.
CrossRef
CAS
Google Scholar
Schoonjans, T., Vincze, L., Solé, V. A., Sanchez del Rio, M., Brondeel, P., Silversmit, G., Appel, K. & Ferrero, C. (2012). At. Spectrosc. 70, 10–23.
Web of Science
CrossRef
CAS
Google Scholar
Schropp, A. & Schroer, C. G. (2010). New J. Phys. 12, 035016.
Web of Science
CrossRef
Google Scholar
Shockley, W. (1961). Solid-State Electron. 2, 35–67.
CrossRef
Google Scholar
Solé, V. A., Papillon, E., Cotte, M., Walter, P. & Susini, J. (2007). At. Spectrosc. 62, 63–68.
Google Scholar
Sparks, C. J. Jr (1980). Synchrotron Radiation Research, edited by H. Winick & S. Doniach, ch. 14, pp. 459–512. New York: Plenum Press.
Google Scholar
Sun, Y., Gleber, S.-C., Jacobsen, C., Kirz, J. & Vogt, S. (2015). Ultramicroscopy 152, 44–56.
CrossRef
CAS
PubMed
Google Scholar
Van Grieken, R. A. & Markowicz, A. A. (2002). Handbook of X-ray Spectrometry, 2nd ed., Vol. 29 of Practical Spectroscopy. New York: Marcel Dekker.
Google Scholar
Vogt, S. (2003). J. Phys. IV (Paris) 104, 635–638.
CAS
Google Scholar
Wildeman, A. S., Patel, N. K., Cormack, B. P. & Culotta, V. C. (2023). PLoS Pathog. 19, e1011478.
CrossRef
PubMed
Google Scholar
Zee, D. Z., MacRenaris, K. W. & O'Halloran, T. V. (2022). Curr. Opin. Chem. Biol. 69, 102152.
CrossRef
PubMed
Google Scholar
Zelinka, S. L., Jakes, J. E., Tang, J., Ohno, K., Bishell, A., Finney, L., Maxey, E., Vogt, S. & Kirker, G. T. (2018). Wood Mater. Sci. Eng. 3, 174–184.
Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.

journal menu
access



