- 1. Introduction
- 2. Theoretical total neutron cross sections of SDSS and PE
- 3. Monte Carlo simulations
- 4. Neutron imaging experimental setup and data processing
- 5. NCrystal-based least-squares fitting
- 6. Conclusions
- A1. Validation of OpenMC+NCrystal simulations
- A2. Simulations for more materials in different sample-to-detector distances
- A3. Discussion regarding the forward-scattering effects across different ROIs
- A4. Discussion regarding the OA correction
- A5. Parameter determination of Jorgensen resolution function for IMAT
- A6. Discussion regarding the H/C ratios for the thick PE sheet measured on IMAT
- A7. Fitting the PE thickness without using a reference
- References
- 1. Introduction
- 2. Theoretical total neutron cross sections of SDSS and PE
- 3. Monte Carlo simulations
- 4. Neutron imaging experimental setup and data processing
- 5. NCrystal-based least-squares fitting
- 6. Conclusions
- A1. Validation of OpenMC+NCrystal simulations
- A2. Simulations for more materials in different sample-to-detector distances
- A3. Discussion regarding the forward-scattering effects across different ROIs
- A4. Discussion regarding the OA correction
- A5. Parameter determination of Jorgensen resolution function for IMAT
- A6. Discussion regarding the H/C ratios for the thick PE sheet measured on IMAT
- A7. Fitting the PE thickness without using a reference
- References
research papers
accessQuantification in wavelength-resolved neutron imaging: exemplified using polyethylene to mimic hydrogen in stainless steel
aEuropean Spallation Source ERIC, Partikelgatan 5, SE-224 84 Lund, Sweden, bDivision of Synchrotron Radiation Research, Department of Physics, Lund University, Box 118, SE-221 00 Lund, Sweden, cDivision of Surface and Corrosion Science, Department of Chemistry, KTH Royal Institute of Technology, Drottning Kristinas väg. 51, SE-100 44 Stockholm, Sweden, dDepartment of Engineering Mechanics, KTH Royal Institute of Technology, Teknikringen 8D, SE-100 44 Stockholm, Sweden, eESS Data Management and Scientific Computing Division, Lyngby, Denmark, fForschungs-Neutronenquelle Heinz Maier-Leibnitz, 85748 Garching, Germany, and gISIS Facility, Rutherford Appleton Laboratory, Chilton, Didcot OX11 0QX, UK
*Correspondence e-mail: [email protected], [email protected]
We present a systematic and improved quantitative methodology for analysing wavelength-resolved neutron transmission spectra in metallic alloys. The approach is demonstrated using data acquired on both the IMAT beamline at ISIS and the RADEN beamline at J-PARC on Super Duplex stainless steel samples partially covered with thin polyethylene layers to mimic small controlled hydrogen concentrations. Initially, we employed Monte Carlo particle transport simulations to assess the effect of forward-scattered neutrons on the recorded transmission signal as a function of sample-to-detector distance and neutron beam size. Guided by this simulation framework, we define an experimental configuration and data processing workflow that enables the reliable quantification of very small transmission differences, reaching the sub-percent regime. An NCrystal-based least-squares fitting routine is developed to resolve multiple microstructural parameters simultaneously, including unit-cell parameters, phase fractions, element composition and extinction-related parameters. The model quantitatively reproduces the experimental transmission spectra and determines polyethylene layer thicknesses with micrometre-level precision. This work establishes a robust microstructure-sensitive framework for quantitative neutron transmission analysis. The approach enables the quantification of trace elements that produce measurable contrasts in neutron transmission, exemplified here for hydrogen in steel with a demonstrated detection limit of 10 wt p.p.m. The presented workflow provides a broadly applicable framework for improving quantitative neutron imaging across diverse materials and applications.
Keywords: materials science; computational modelling; neutron imaging; hydrogen quantification; hydrogen detection limit.
1. Introduction
Wavelength-resolved neutron transmission imaging is a powerful nondestructive technique for characterizing the microstructural properties of materials, with diverse applications in materials science and engineering research domains, such as energy materials and cultural heritage studies (Strobl & Lehmann, 2024
; Aswal et al., 2022
; Woracek et al., 2018
; Kardjilov et al., 2018
; Tran et al., 2021
). When an incident neutron beam traverses a sample, its intensity is attenuated due to nuclear interactions, primarily scattering and absorption, as governed by the Beer–Lambert law. The resulting wavelength-dependent transmission spectrum encodes detailed information about the material's structural and compositional characteristics, including elemental composition and crystallographic properties. For polycrystalline materials, Bragg edge analysis provides access to phase fractions, unit-cell parameters, strain states, preferred orientation (texture) and other microstructural effects that cause, for example, extinction.
This study is motivated by the need for quantitative determination of very low hydrogen concentrations in metallic materials. While the quantification of trace elements is relevant in many fields, this challenge is particularly critical in the study of hydrogen embrittlement (HE), where even trace amounts of hydrogen can lead to significant degradation of mechanical properties. HE is a complex multiscale phenomenon involving hydrogen absorption, diffusion, trapping and interactions with microstructural features, ultimately compromising the ductility, toughness and fatigue life in high-strength steels and Ni-based alloys (Yu et al., 2024
). Despite decades of research, the underlying failure mechanisms, such as hydrogen-enhanced decohesion (HEDE), hydrogen-enhanced localized plasticity (HELP) and hydrogen-enhanced strain-induced vacancy (HESIV), remain difficult to validate experimentally, in part because hydrogen concentrations in engineering alloys are typically very low.
While several experimental techniques exist for hydrogen characterization, quantitative and spatially resolved measurements in structural alloys at technologically relevant concentrations remain highly challenging. Conventional methods such as thermal desorption spectroscopy (TDS) or hot gas extraction (HGE) lack spatial resolution, while high-resolution techniques such as secondary ion mass spectrometry (SIMS) and atom probe tomography (APT) are limited by small sample volumes, demanding sample preparation, and challenges associated with the reliable detection and quantification of hydrogen. Neutron imaging, by contrast, has long been recognized as a promising technique due to its high penetration depth and intrinsic sensitivity to hydrogen, offering the potential for spatially resolved measurements in bulk materials. Although hydrogen possesses an exceptionally large incoherent neutron scattering cross section, extracting reliable quantitative information from transmission measurements has remained challenging. Notably, Buitrago et al. (2018
) demonstrated the advantages of wavelength-resolved neutron transmission analysis for determining low hydrogen concentrations in Zr-based alloys (where stable hydrides form).
While it is possible to manufacture calibrated specimens of Zr-based alloys with well defined hydrogen concentrations, this is far more challenging for steels, from which hydrogen may effuse readily. In an ongoing research project, we are investigating the hydrogen embrittlement behaviour of stainless steels. Previously, we successfully examined delayed cracking in high-strength martensitic steel using in situ neutron radiography on the NeXT beamline at ILL (Lindblom et al., 2024
; Tengattini et al., 2020
). However, these studies relied on wavelength-integrated imaging and did not enable direct detection or quantification of hydrogen. This limitation arose because the hydrogen concentrations were below the detection sensitivity of the employed imaging configuration and/or localized within regions smaller than the achievable spatial resolution.
To address this challenge, we moved towards wavelength-resolved neutron imaging, which provides enhanced sensitivity to small transmission changes. In the present work, we establish a robust analysis workflow that enables the quantification of small hydrogen-equivalent signals from wavelength-resolved spectra. The methodology is demonstrated using Super Duplex stainless steel (SDSS) samples partially covered with thin polyethylene (PE) layers as a controlled proxy for low hydrogen content. While PE serves as a convenient validation case, the underlying workflow is broadly applicable to other systems requiring precise quantification of trace element contributions. We therefore expect that the methodology presented here will support future advances in quantitative neutron imaging at pulsed neutron sources in general (Shinohara et al., 2020
; Kockelmann et al., 2018
; Bilheux et al., 2023
; Strobl, 2015
; Nelson et al., 2018
).
In addition to sample-related scattering effects, quantitative wavelength-resolved neutron imaging requires careful control of systematic effects arising from the experimental configuration. Detector-generated background is commonly assumed to remain unchanged between the open-beam and sample measurements used for transmission normalization. However, changes in the detector irradiation conditions caused by the introduction of a sample can produce wavelength-dependent systematic offsets in the normalized transmission that should be considered for highly quantitative measurements.
To address these systematic effects, our experimental strategy begins with Monte Carlo particle transport simulations to quantify the potential impact of neutron scattering processes, particularly forward scattering, on the measured transmission spectra. These simulations effectively act as a digital twin of the experiment, enabling us to assess and optimize the experimental configuration prior to measurement. Performing these simulations beforehand is an important step in designing an appropriate experimental configuration that ensures systematic effects remain sufficiently small for quantitative analysis.
Forward scattering refers to neutrons being scattered in the downstream direction. When the detector is positioned close to the sample, a fraction of these scattered neutrons still reach regions of the detector that are `covered' by the sample, leading to an overestimation of the transmission signal. This effect is especially relevant in neutron imaging, where short sample-to-detector distances are commonly employed to minimize geometric blurring. The influence of forward scattering on quantitative transmission measurements has been well studied and several correction strategies have been proposed. For example, Kardjilov et al. (2005
) developed a point-scattered function approach to correct for scattering contributions in neutron radiography. More recently, Boillat et al. (2018
) introduced the black-body grid correction, which effectively compensates for systematic detector-related and geometry-induced artefacts – even beyond those arising solely from forward scattering. For a more comprehensive overview, we refer readers to Boillat et al. (2018
) and the references therein.
While the impact of forward scattering has been previously investigated for iron samples using the Monte Carlo ray-tracing code McStas (Boin, 2011
; Willendrup & Lefmann, 2020
; Willendrup & Lefmann, 2021
) and more comprehensively by Sato and co-workers using the PHITS+RITS framework (Sato et al., 2017
; Sato et al., 2013
; Sato et al., 2011
), we make use of recent advancements in open-source simulation tools. Specifically, we utilize the NCrystal toolkit (Cai et al., 2019
; Cai & Kittelmann, 2020
; Kittelmann & Cai, 2021
), which can serve as a scattering kernel backend for Monte Carlo codes including OpenMC (Romano et al., 2015
) and McStas. NCrystal provides the scattering physics of low-energy neutrons (including both coherent and incoherent, elastic and inelastic scattering for various materials; complementary physics such as magnetic neutron scattering is also supported via the plugin feature) while the particle transport code OpenMC handles complex geometry, multiple scattering and other neutron transport physics. Recent work has demonstrated the promising potential of the combination of OpenMC and NCrystal for quantitative neutron experiment modelling (Xu et al., 2024
). Consequently, we performed such simulations using OpenMC and NCrystal to account for forward-scattering effects accurately, thereby enabling improved experimental design and quantitative spectra analysis.
We have developed an analysis routine based on NCrystal and conducted neutron transmission experiments on both the IMAT (Kockelmann et al., 2018
) imaging instrument located at Target Station 2 of the ISIS facility and the RADEN (Shinohara et al., 2020
) imaging beamline at J-PARC. Recent years have seen growing adoption of NCrystal for transmission or imaging experiment analysis (Strobl et al., 2025
; Sebold et al., 2024
; Simoni et al., 2023
; Sumarli et al., 2024
; Minniti et al., 2025
; Hirsh et al., 2025a
; Xu et al., 2025
), including work combining NCrystal with least-squares fitting to characterize extinction effects in beryllium metal (Xu et al., 2025
). Building upon these developments, we present an extended analysis pipeline for quantitative analysis of neutron transmission spectra.
This work is organized as follows. First, we present the theoretical background for the low-energy neutron cross sections of SDSS and PE. Next, we present the Monte Carlo simulations used to study the impact of forward scattering under different experimental configurations, which are then used to determine the optimal experimental configuration. We then describe the utilized neutron imaging experimental setup on IMAT and RADEN and the data processing methodology used to obtain the transmission spectra. Finally, we detail the NCrystal-based least-squares fitting procedure and demonstrate how small signals from thin PE layers can be reliably extracted, where the detection limit of hydrogen is derived, followed by a discussion of the implications for future quantitative neutron imaging.
2. Theoretical total neutron cross sections of SDSS and PE
The total neutron cross section σ(λ) (with λ being the neutron wavelength) is a wavelength-dependent measure of the probability of a neutron interacting with atoms in a material. For the non-fissionable materials SDSS and PE, σ(λ) is composed of the scattering cross section σscat(λ) and the absorption cross section σabs(λ),
The physics of low-energy neutron scattering is detailed in the literature (Squires, 2012
; Schober, 2014
; Carpenter & Loong, 2015
). Under the incoherent approximation, σscat(λ) can be calculated by
where ,
and σinel(λ) represent the coherent elastic, incoherent elastic and inelastic scattering cross sections, respectively.
In this work, the open-source toolkit NCrystal is used to calculate the wavelength-dependent neutron scattering and absorption cross sections for SDSS and PE, taking advantage of the extensive NCrystal database, which includes crystalline and amorphous solids, liquids and gases. Microstructural parameters, including elemental composition, phase fractions and unit-cell parameters, are specified via NCMAT files and built-in configuration string functionalities.
The material investigated in this work is Super Duplex stainless steel (SDSS) grade 2507 (UNS S32750). Its properties are detailed by AZo Materials (2025
). SDSS is composed of approximately 50% ferrite (body-centred cubic, b.c.c.) and 50% austenite (face-centred cubic, f.c.c.). The nominal element compositions are summarized in Table 1
. By incorporating these properties into NCrystal, the scattering and absorption cross sections were obtained [Fig. 1
(a)].
|
|||||||||||||||||||||||||||||
|
|
Figure 1
(a) Neutron cross sections per atom for the SDSS sample. The cross-section components defined in equations (1) |
It should be noted that we initially defaulted to the phonon density of states (PDOS) of pure iron in the ferrite and austenite phases to calculate the scattering cross sections of the SDSS sample. While phonon softening is a common signature in alloys compared with their pure forms, as shown, for example, by Beskrovni et al. (1999
) and Yang et al. (2022
), we found the effect to be negligible here. We used the PDOS of f.c.c. Fe–Cr–Ni–N austenite alloys with varying chromium contents from Beskrovni et al. (1999
) to compute the cross sections. The resulting cross sections are nearly identical to those computed using the pure iron PDOS. Consequently, we proceeded with the pure iron PDOS for all subsequent SDSS cross-section calculations.
Fig. 1
(b) shows the total cross section for amorphous PE (chemical formula CH2) σPE(λ), calculated using NCrystal. This value is composed of the cross section of carbon bound in CH2, , plus twice the cross section of hydrogen bound in CH2,
. These calculations use the NCMAT file of PE available in the NCrystal database, which has been benchmarked against experimental data from Ramić et al. (2018
). For comparison, the total cross sections for H bound in CH2 and free H are also plotted. These differences originate from the chemical bonding and dynamic state of hydrogen in the material, on one hand highlighting the importance of accounting for hydrogen dynamics for accurate quantification, and on the other hand opening new avenues for experimental investigations. Although the dynamics of hydrogen bound in PE cannot be directly used to model those in stainless steels, the workflow presented hereafter remains valid, provided the hydrogen dynamics are determined from complementary studies, such as vibrational spectroscopy measurements, density functional theory (DFT) calculations or molecular dynamics (MD) simulations.
In the present case, we investigate a Super Duplex stainless steel consisting of approximately equal fractions of ferrite (b.c.c. form) and austenite (f.c.c. form). Previous studies have shown that hydrogen is primarily trapped in austenite for ferrite–austenite dual-phase materials (Silverstein et al., 2017
; Zou et al., 2026
). Danilkin et al. (2003
) conducted neutron vibrational spectroscopy measurements on two f.c.c. Fe-based austenite alloys doped with hydrogen. Based on their study, we extracted the PDOS of hydrogen and used it to compute the neutron total cross section. The resulting cross section shows a noticeable difference compared with that of free hydrogen [Fig. 1
(b)]. These results demonstrate the sensitivity of the neutron signal to hydrogen dynamics, providing a framework that can be readily applied to quantify hydrogen concentrations in Fe-based alloys.
3. Monte Carlo simulations
Since it is well known that forward-scattered neutrons can introduce systematic artefacts in transmission measurements, we first evaluated the influence of the experimental geometry using Monte Carlo particle transport simulations, which serve as a digital twin of the experimental setup. A comprehensive investigation of such effects was conducted by Sato et al. (2017
), who studied how parameters such as sample-to-detector distance, detection area, detection position and sample thickness influence transmission spectra. In their work, the Monte Carlo particle transport code PHITS was used to obtain simulated transmission spectra, while the RITS code was employed to compute the corresponding theoretical transmission intensities for comparison.
In the present study, we focus specifically on the impact of the sample-to-detector distance and the slit/beam size on the transmission spectrum using OpenMC+NCrystal simulations. The sample-to-detector distance is defined as the distance from the end of the sample to the detector sensor. We used the established work of Sato et al. (2017
) as a benchmark for validating our Monte Carlo simulations, as detailed in Appendix A
(Fig. 12), ensuring the reliability of our calculations. We then conducted a systematic study to evaluate forward-scattering effects and their impact on the accuracy of transmission-based quantification. The investigation considered several experimental configurations, of which three representative cases (Cases 1–3) are presented here, as illustrated in Figs. 2
(b)–2
(d). While Case 3 was ultimately adopted for the experimental measurements on IMAT, Cases 1 and 2 illustrate the strong influence of slit size on the transmission spectrum.
|
Figure 2
(a) Schematic presentation of the neutron transmission experiment configuration in OpenMC+NCrystal simulations. The incident neutron distribution follows the open-beam spectrum of IMAT, with beam divergence determined by the pinhole diameter and the distance between the pinhole and the sample. The incident neutron wavelength/energy distribution, beam divergence, sample geometry, region of interest (ROI) or detector region and sample-to-detector distance (L) can be flexibly adjusted in the simulations. (b) and (c) Two simulation cases where the sample dimensions are both 10 × 10 × 0.8 cm (width × height × thickness) and the ROIs are both 8 × 3 mm, but with different slit sizes. (d) The setup adopted for the IMAT experiment. See text for more details. |
The open-beam spectrum was taken from a previous experiment and used as input for the simulations. The beam divergence θdiv was determined from the experimental pinhole size Dph and the pinhole-to-sample distance, Lp–s = 10 m, using the relationship =
reported by Burca et al. (2013
). With Dph set to 60 mm – a value typically chosen to optimize the detector count rate – the resulting beam divergence was θdiv = 0.344°. These are key parameters set for the OpenMC+NCrystal simulations. It should be noted that neutron divergence at the guide exit is not accounted for in our simulations. In principle, more detailed beamline descriptions, using for example McStas or VITESS (Zendler et al., 2014
; Robledo et al., 2026
), could be employed to achieve a more accurate representation of the beam characteristics. However, for the purposes of this study, the divergence defined by the pinhole aperture and pinhole-to-sample distance provides a sufficiently realistic and practical basis for experimental design.
The simulated beam profile with divergence is shown in Fig. 2
(a). Regarding the theoretical transmission intensity Tth(λ) for a sample composed of an ensemble of materials with each component indexed by i, Tth(λ) can be calculated from each individual total neutron cross section σi(λ), number density Ni and sample thickness li,
Fig. 3
compares simulated transmission spectra, at both a close distance (0 cm) and a sufficiently large sample-to-detector distance that yields data close to the theoretical neutron transmission predictions, for three different configurations, Cases 1, 2 and 3. The theoretical transmission spectra were derived from total cross sections calculated via NCrystal (Fig. 1
) and equation (3
), using nominal densities of 7.8 g cm−3 and 0.92 g cm−3 for SDSS and PE, respectively. Our results demonstrate that for an 8 mm thick SDSS sample, sample-to-detector distances of 50 cm, 5 cm and 2 cm are required for effective suppression of forward-scattering artefacts in Cases 1, 2 and 3, respectively. For comparison, we also considered 8 mm thick PE samples to represent materials with high hydrogen content. These simulations show that significantly larger sample-to-detector distances would be required in each case. A comparison of Cases 1 and 2 shows that a smaller slit size can dramatically reduce the required sample-to-detector distance (from 50 cm to 5 cm for the 8 mm thick SDSS) to mitigate forward-scattering effects.
|
Figure 3
Comparison of simulated and theoretical transmission spectra (solid black lines) for 8 mm thick SDSS and PE across Cases 1, 2 and 3. Results indicate that sample-to-detector distances of 50 cm, 5 cm and 2 cm are required to mitigate forward-scattering artefacts for Cases 1, 2 and 3, respectively. In contrast, PE – a material with a high hydrogen content – requires significantly larger distances (150 cm, 15 cm and 5 cm) to eliminate these impacts. Additional spectra for other materials are provided in Figs. 13 and 14 in Appendix A |
Guided by these simulations, we selected an experimental configuration that minimizes forward-scattering effects while maintaining a sample geometry representative of conventional mechanical tensile testing. The investigated material consists of an SDSS specimen (gauge length 25 mm, thickness 8 mm, width 3 mm), partially covered with thin polyethylene (PE) sheets (length 2 mm, width 3 mm) of two different thicknesses (45 µm and 250 µm) to mimic hydrogen signals. The width of the specimen (3 mm) ensures that part of the neutron beam (with a 6.8 mm slit) remains unobstructed, providing an internal reference region for normalization of the transmission data. Theoretical transmission spectra for the SDSS specimen with and without PE attachments were calculated using equation (3
).
We also performed OpenMC+NCrystal simulations to quantify potential forward-scattering effects for our specific experimental setup on IMAT (Case 3), which utilized a slit size of 21 × 6.8 mm. With a pinhole size of Dph = 60 mm, the calculated divergence was θdiv = 0.344°. The regions of interest (ROIs) were configured to match the maximum regions usable for experimental data analysis: 8 × 3 mm for the SDSS grade without PE and 2 × 3 mm for the PE-attached regions. In practice, we restrict the ROIs to dimensions smaller than the sample to avoid neutron imaging edge enhancement effects due to different neutron refractive indices at the interfaces (Strobl et al., 2009
; Mirashi et al., 2026
); thus, the simulated transmission spectra represent the upper limits of potential forward-scattering artefacts. The results of the combined SDSS + PE simulations (Fig. 4
) show that having the detector positioned 20 mm from the sample is still sufficient to mitigate forward-scattering artefacts.
|
Figure 4
Comparison of simulated (OpenMC+NCrystal) versus theoretical transmission spectra for the front configuration (PE attachments facing away from the detector). For the Case 3 experimental setup, a sample-to-detector distance of 20 mm is sufficient to mitigate forward-scattering artefacts for an SDSS sample (25 mm in length of gauge, 8 mm thickness, 3 mm width), both with and without PE attachments, within the selected ROIs. Smaller sample-to-detector distances would introduce biases in the measured transmission spectra and thus hinder quantitative analysis. These data define the expected experimental results for the quantitative transmission measurements explained in this work. |
Consequently, we adopted this configuration (Case 3 with a sample-to-detector distance of 20 mm) for our experimental setup on IMAT. Further discussion regarding the forward-scattering effects across the selected ROIs is detailed in Appendix A3
.
4. Neutron imaging experimental setup and data processing
The methodology presented in this paper is established using experiments performed on the IMAT beamline at ISIS. The workflow was subsequently implemented on the RADEN beamline at J-PARC for the investigation of hydrogen-charged SDSS specimens. While the full RADEN study will be reported separately, reference measurements of the SDSS specimen with PE attachments are included here to give further validation of the sensitivity and reproducibility of the proposed methodology.
4.1. Wavelength-resolved imaging experiments on IMAT
Neutron transmission measurements were performed on the IMAT beamline on an SDSS specimen. The sample was machined from bulk material (provided by the industrial collaborator Alleima) into a standard tensile geometry (Fig. 5
), with a total length of 149 mm, a gauge length of 25 mm and a cross section of 8 × 3 mm. The nominal element compositions for the SDSS grade are given in Table 1
. Two PE sheets with nominal thicknesses of 45 µm and 250 µm were attached to one surface using aluminium tape. The 250 µm sheet was sourced from Goodfellow (https://www.goodfellow.com) and had been previously characterized on the VESUVIO spectrometer at ISIS. Cadmium markers were used to indicate the positions of both PE sheets. As hydrogen is commonly quantified in parts per million (wt p.p.m.), the two PE layers (density 0.92 g cm−3) correspond to hydrogen-equivalent concentrations of approximately 95 wt p.p.m. (45 µm sheet) and 530 wt p.p.m. (250 µm sheet) relative to the underlying steel volume.
|
Figure 5
SDSS sample used in the transmission measurements conducted on the IMAT beamline at ISIS, (a) viewed perpendicular to the beam direction and (b) viewed along the beam direction. The dimensions of the SDSS sample (length of gauge, thickness, width = 25 × 8 × 3 mm) and the sample-to-detector distance (20 mm), as determined from Monte Carlo simulations, are indicated. Two PE sheets were attached to one surface of the SDSS sample using aluminium tape, to mimic known hydrogen concentrations, as shown in the enlarged image inset in panel (a). Neutron-absorbing cadmium markers were placed on the sides of the sample to aid subsequent image analysis. Not shown is that the entire sample was rotated by 180° for an additional measurement, placing the PE layers on the side facing the detector. |
The experimental setup employed a LumaCam detector system (Losko et al., 2021
; Wolfertz et al., 2024a
) positioned downstream of the neutron beam path. Using optical lenses, the system allows for an adaptable field of view (active detector area). As the experiment targeted small changes in the transmission signal, it was essential to use a detector with low gamma sensitivity and high dynamic range to minimize uncertainties in the measurements. The LumaCam fulfils these requirements through on-line single neutron event reconstruction in counting mode, enabling the discrimination of individual neutron interactions within the scintillator. This approach suppresses scintillator afterglow and background contributions, resulting in a detector system with high spatiotemporal resolution suitable for a wide range of neutron imaging applications (Wolfertz et al., 2024b
; Jäger et al., 2024
; Jäger et al., 2025
; Mirashi et al., 2025
; Hirsh et al., 2025b
; Khaplanov et al., 2025
; Hirsh et al., 2025a
). Individual neutron events can be identified through the nanosecond resolution of their time of arrival. This facilitates the separation of two neutron events that have different arrival times. A limitation of this method is that the neutron intensity must not exceed a threshold where individual neutron events can no longer be temporally resolved. An alternative detector technology with high spatiotemporal resolution is the neutron-sensitive microchannel plate (MCP) detector (Tremsin & Vallerga, 2020
).
Using the LumaCam detector in this experiment, a field of view of 28 × 28 mm was configured and binned into 512 × 512 pixels in space (55 µm effective pixel size). The time-of-flight (TOF) binning was set to 25 µs bin width, resulting in 4000 individual images across the wavelength spectrum for each individual measurement. The pinhole size of IMAT was set to 60 mm, corresponding to an L/D ratio of 935. Initial calibration measurements were performed with 15 min acquisition time for both the open-beam and an iron powder reference sample, corresponding to proton charges of 10.1023 µAh and 10.0602 µAh, respectively. The TOF–wavelength relationship is given by
where t is the TOF, h is Planck's constant, m is the neutron mass, L0 is the flight path length and t0 is the time delay of the source trigger.
The determined flight path length is L0 = 56.1 (3) m and the time delay is 3.2 µs. Using the TOF–wavelength relationship, the measured open-beam spectrum can be determined.
The subsequent wavelength-resolved transmission data fitting excluded image frames of neutron wavelength λ ≤ 1.6 Å, due to a poor signal-to-noise ratio, and those of λ ≥ 7 Å, where the neutron flux diminishes significantly.
Guided by the preceding OpenMC+NCrystal simulations, we adopted the Case 3 configuration for the experiment, positioning the SDSS specimen (attached to two PE sheets of differing thicknesses) at a distance of 20 mm from the detector. The distance is defined from the end of the sample to the detector sensor. Two complementary measurements were performed: (i) a `front' configuration with the PE sheets facing the beam (Fig. 5
) and (ii) a `back' configuration with the sample flipped 180°, placing the PE sheets downstream. Acquisition times were 46 min for the front configuration and 61.5 min for the back configuration, corresponding to integrated proton charges of 30.258 and 40.119 µAh, respectively. Fig. 6
(a) displays the front radiograph integrated across neutron wavelengths of 1.6–7.0 Å, confirming proper sample alignment within the 21 × 6.8 mm beam profile. Four rectangular ROIs, delineated in red (thin PE), blue (without PE), green (thick PE) and purple [open area (OA)], were defined based on cadmium marker positions. The OA serves as a reference to monitor neutron intensity fluctuations or other systematic variations.
|
Figure 6
(a) Integrated neutron radiograph of the front SDSS sample configuration, summed over the wavelength range of 1.6–7.0 Å. The 21 × 6.8 mm rectangular beam profile, the sample and three cadmium markers are clearly visible within the 512 × 512 pixel detection area. (b) Normalized transmission image obtained by dividing the image in panel (a) by the open-beam reference (integrated over the same wavelength range) with proton-charge normalization. Four ROIs [coloured red, blue, green and purple, also indicated in (a)] were defined based on cadmium marker positions: thin PE attachment (left), without PE attachments (bottom centre), with thick PE attachment (right) and OA (top centre). (c) Front neutron transmission spectra for the OA (top panel) and the PE-free SDSS ROI (lower panel). The OA correction compensates for the wavelength-dependent offset observed in the OA transmission, resulting in a non-negligible upward shift of the transmission spectrum. (d) Transmission spectra for all ROIs after OA correction. Primary hkl reflections for both the ferrite and austenite phases are labelled (austenite indices in italics). Back spectra are omitted due to their overlap with front measurements. The lower panel displays the front-to-back transmission ratios, showing a consistent average ratio of unity across all ROIs. |
Following the same integration procedure for open-beam measurements (155 min, 101.603 µAh), we obtained a normalized transmission image by dividing the sample radiograph by the open-beam data, with both datasets normalized by proton charge. Fig. 6
(b) presents the resulting transmission image; the distinct transmission intensities across the three ROIs validate the selection of these regions. To derive wavelength-dependent spectra, detector counts within each ROI were summed per time frame and normalized by the corresponding open-beam counts and proton charges. The upper panel of Fig. 6
(c) displays the resulting spectrum in the OA region for the front configuration. The observed transmission is up to 7% below unity, although it should theoretically be 1, indicating an additional source of intensity fluctuation. Because the measured transmission is below unity, scattering from the sample into the OA can be excluded, as such scattering would instead increase the detected intensity. A more plausible explanation is detector-related parasitic scattering that contributes to the open-beam measurement. During the sample measurement, part of the direct neutron beam that contributes to this detector-related background is attenuated by the sample. Consequently, the detector-generated background is reduced in the sample measurement compared with the open-beam measurement, leading to an apparent transmission below unity after normalization.
Although this interpretation cannot be proven directly from the present measurements, it is supported by the comparison with the RADEN measurements presented below, where a different detector configuration exhibits OA transmission much closer to unity. To compensate for this wavelength-dependent offset, we implemented an OA-based correction by dividing each transmission spectrum by the wavelength-dependent OA intensity.
The corrected transmission spectra are shown in [Fig. 6
(d)]. They allow for the clear identification of Bragg edges corresponding to the ferrite (b.c.c.) and austenite (f.c.c.) phases. As expected, the PE-attached regions show reduced transmission compared with PE-free areas. This attenuation scales with PE thickness and a detailed quantitative analysis is presented later in this work. Despite the smaller ROIs for the PE-covered regions, the propagated statistical uncertainties remain sufficiently low. The back transmission spectra overlap with the front spectra. The ratios between them are near to unity, as shown in the lower panel.
4.2. Wavelength-resolved imaging experiments at RADEN
To demonstrate the transferability of the proposed workflow, the methodology developed for the IMAT measurements was directly applied to experiments on the RADEN beamline at J-PARC (Shinohara et al., 2020
). Specimens with dimensions of 30 × 10 × 6 mm from the same SDSS material, covered with three PE sheets of nominal thicknesses 35, 100 and 230 µm, were measured. As in the IMAT study, OpenMC+NCrystal simulations were first performed to optimize the experimental geometry. A sample-to-detector distance of 50 mm was found to be sufficient to eliminate forward-scattering effects across the complete 30 × 10 mm ROI. The experimental setup is shown in Fig. 7
(a). Great care was taken to define the neutron beam upstream of the sample using several neutron-absorbing beam limiters (made from B4C). The sample holder was 3D-printed from neutron-absorbing BNPLA [mixing of hexagonal boron nitride (h-BN) and polylactic acid (PLA)] (Sebold et al., 2024
) and designed to minimize the amount of material surrounding the sample while maximizing the air volume around it, thereby reducing parasitic neutron scattering. The PE sheets (with a density of 0.92 g cm−3) were sourced from Goodfellow (https://www.goodfellow.com) and were in pristine condition, without any previous exposure to neutron radiation.
|
Figure 7
(a) Experimental setup for wavelength-resolved neutron imaging on the RADEN beamline at J-PARC. The sample holder is moved to be irradiated by the neutron beam during measurements. Three PE sheets with different nominal thicknesses of 100, 35 and 230 m (left to right) are attached to the 6 mm thick SDSS sample using aluminium tape. Cadmium markers indicate the PE sheet positions. (b) Neutron transmission images of the SDSS sample with and without the PE sheets. Fifteen selected ROIs are arranged in a 5 column × 3 row grid. Five zones in the second row are labelled with numbers from 1 to 5. (c) Transmission spectra for the OA regions (top) and for the second-row ROIs (Zones 1 to 5) of the SDSS sample with and without PE sheets (middle and bottom). Primary hkl reflections for both the ferrite and austenite phases are labelled (austenite indices in italics). |
The LumaCam detector was configured with a field of view of 47 × 47 mm and operated in event-counting mode, recording 512 × 512 pixels and 2000 individual TOF images per pulse with a time-bin width of 20 µs.
The TOF–wavelength relationship was calibrated using an iron powder measurement, yielding a flight path length of 23.7 (1) m and a time delay of 3.8 µs via equation (4
). Three actual measurements were then conducted: open beam, the SDSS sample with PE sheets and the SDSS sample without PE sheets. The exposure time of each measurement was 20 min, corresponding to trigger counts (used at RADEN instead of accumulated proton charge) of 25828, 25827 and 25827, respectively. Fig. 7
(b) shows the resulting neutron transmission images, obtained by dividing the integrated radiograph of the sample by that of the open beam and normalizing by their respective trigger counts. Fifteen ROIs were selected and arranged in a 5 × 3 (column × row) grid based on the PE sheet positions. The corresponding transmission spectra of the second row of the ROIs, labelled with numbers from 1 to 5, corrected for OA (transmission spectra shown in the top panel), are shown in Fig. 7
(c). The OA transmission spectra are below but much closer to unity than those measured on IMAT. This observation further supports the interpretation that the wavelength-dependent offset originates from a detector-related systematic effect rather than changes in the incident neutron spectrum. While the exact origin of the detector-related background cannot be identified conclusively from the present measurements, the different detector configuration at RADEN evidently reduces this effect, resulting in OA transmission values much closer to unity. These results highlight the importance of carefully controlling detector-related systematic effects for highly quantitative wavelength-resolved neutron transmission measurements.
5. NCrystal-based least-squares fitting
5.1. Data analysis procedure
This section details our least-squares fitting procedure, built upon the NCrystal toolkit, for the quantitative analysis of wavelength-resolved neutron imaging data. The procedure is implemented in a script that has been made publicly available (Xu, 2025
). The overall workflow of the present study is summarized in Fig. 8
. The determination of the experimental setup using Monte Carlo simulations and the subsequent data processing have been detailed in Sections 3
and 4
, respectively. NCrystal calculates the wavelength-dependent total neutron cross section [equation (1
)] by incorporating key microstructural parameters: elemental composition, phase fractions, unit-cell parameters, extinction-related parameters and PDOS. These properties are specified via NCMAT files, built-in configuration strings or the specialized CrysExtn plugin (Xu, 2024
; Xu et al., 2025
) for extinction effects. Notably, the extinction model for spherical crystallites has recently been improved numerically (Kittelmann et al., 2026
) and will be integrated as a built-in functionality within NCrystal.
|
Figure 8
Overall workflow developed in this work, which includes Monte Carlo simulations to determine the experimental configuration, data processing and a least-squares fitting procedure based on the cross section σ(λ) calculated by NCrystal. See text for a detailed explanation. |
The wavelength-dependent neutron cross section calculated by NCrystal provides a forward model for analysing the transmission spectrum. The analytical transmission intensity, T(λ), is calculated by convoluting the theoretical transmission intensity Tth(λ), given in equation (3
), with the resolution function R(λ),
where A is a scale factor (typically near unity) to account for systematic variations such as residual wavelength-independent normalization differences. Even though A is considered flat across the wavelength range used in this study, it can be further refined using a more complex model.
For the IMAT beamline, we found that a Jorgensen-type resolution function (Jorgensen et al., 1978
; Von Dreele et al., 1982
) provides an excellent fit to the Bragg edge transmission spectrum. This function is characterized by seven free parameters: α0, α1, β0, β1, Σ0, Σ1 and Σ2. The parameters α and β (parameterized by α0 and α1, β0 and β1, respectively) are the exponential rise and decay constants that characterize the asymmetric line shape of the neutron moderator pulse. The parameter Σ (parameterized by Σ0, Σ1 and Σ2) represents the standard deviation of the Gaussian component of the line shape, which originates from instrumental (Σ0 and Σ1) and sample strain (Σ2) contributions (Sabine et al., 1988
). Following established conventions (Vogel, 2000
), and considering that our samples are strain-free, α0, Σ0 and Σ2 were set to zero in our analysis. We detail the determination of the Jorgensen function parameters for the IMAT beamline in Appendix A5
. While this function can also model the resolution of the RADEN beamline, determining its parameters is beyond the scope of this work, as the primary objective of the RADEN data is fitting the PE thickness from spectral ratios. This parameter determination for RADEN will be addressed in a future study.
A primary objective of this study is the quantitative detection of trace materials, specifically low-concentration hydrogen within metals, which is simulated using PE layers in the present work. To isolate the attenuation contribution of the PE sheets, the transmission spectrum of the base material TSDSS(λ) is divided by the spectrum of SDSS and PE, TSDSS+PE(λ). According to the Beer–Lambert law, the ratio is obtained by
where σPE(λ) is the total neutron cross section of PE [Fig. 1
(b)], NPE = 0.1185 atoms Å−3 is the number density (calculated from a mass density of 0.92 g cm−3) and lPE is the sample thickness, which is determined by fitting the spectral ratio.
In practice, critical material parameters, such as precise phase fractions and unit-cell parameters, are often not fully known a priori. To address these challenges when analysing full transmission spectra, we developed a nonlinear least-squares fitting routine using the Python package lmfit (Newville et al., 2024
). This routine minimizes the residual between the experimental transmission spectrum Texp(λ) and the analytical spectrum T(λ), or the experimental and analytical ratios [equation (6
)], as calculated by NCrystal. Notably, while fitting the PE thickness is not strictly nonlinear, the overall transmission spectrum fitting remains inherently nonlinear because the underlying parameters (e.g. phase fraction and scale factor) do not share a linear correlation. The subsequent sections detail the fitting procedure and present the results for the SDSS spectra and the PE thickness.
5.2. Results and discussion
5.2.1. IMAT measurements
The transmission spectrum of the SDSS sample without PE attachments, measured in the front configuration, was analysed first. To ensure robust convergence of the fitting procedure, we fixed the values of well defined parameters associated with the transmission calculation. These fixed parameters included the sample thickness (8 mm) and density (7.8 g cm−3), the temperature (296 K) and the element composition specified in Table 1
. Additionally, the flight path length (56.1 m) and time delay (3.2 µs) were fixed using values determined from the iron powder calibration measurements described above in Section 4
.
Unit-cell parameters (2.88 Å for the ferrite phase and 3.61 Å for the austenite phase) were fixed based on the Bragg edge positions in the transmission spectrum. Although dissolved hydrogen can induce lattice expansion in steels, an effect that has been investigated for SDSS in recent modelling and diffraction studies (Lindblom et al., 2026
), the present study focuses on hydrogen-equivalent signals introduced via external PE layers, and the unit-cell parameters were kept fixed during refinement. For the extinction model, following the assumptions of Xu et al. (2025
), the grain size parameter in the model was set equal to the average value measured by electron backscatter diffraction (EBSD). In this work, the average grain sizes measured by EBSD for the ferrite and austenite phases were 17 µm and 12 µm, respectively; these values were kept fixed in the extinction model during fitting.
The Jorgensen-type resolution function parameters were determined as listed in Fig. 17 in Appendix A5
and kept constant. The remaining free parameters for the SDSS grade (front configuration, no PE attachments) and their fitted values are listed in Table 2
.
|
|||||||||||||||||||||||
As shown in Fig. 9
, the quality of the fit improves significantly when the Becker & Coppens extinction model (Becker & Coppens, 1974a
; Becker & Coppens, 1974b
) is applied to both phases, compared with the assumption of an ideal polycrystalline material. Primary and secondary extinction effects were considered for the ferrite phase, while only type-I secondary extinction was applied to the austenite phase. The reduced χ2 value drops from 50 (no extinction) to 3.8 (with extinction in ferrite and austenite). Restricting the extinction model exclusively to the ferrite phase yielded a comparable fit quality (with a reduced χ2 value of 4.3), suggesting that extinction effects are predominant in the ferrite phase of our SDSS sample. The observed extinction in the ferrite phase aligns with observations by Santisteban et al. (2001
), who attributed transmission data discrepancies in b.c.c. iron powder to extinction effects. Notably, mosaic spread parameters cause a rounding of the Bragg edges, which is correlated with the Jorgensen-type resolution function of the instrument. Therefore, prior determination of the resolution parameters (Fig. 17) decoupled these effects, enabling a more precise quantification of the extinction parameters.
|
Figure 9
NCrystal-based least-squares fits of the SDSS grade within the ROI without PE attachments [blue zone in Fig. 6 |
A full quantification of these extinction effects would require complementary diffraction measurements to distinguish them from potential texture, as demonstrated by Xu et al. (2025
). While this suggests an interesting path for future research, the primary objective of this work is to establish a general analysis pipeline for wavelength-resolved neutron imaging. Therefore, a detailed analysis of extinction in iron remains beyond the immediate scope of this paper.
So far, the full transmission spectra have been analysed, including contributions from unit-cell parameters, element composition and extinction effects. However, if the primary objective is solely to quantify hydrogen, a simplified and more robust approach can be employed by considering the ratio of transmission spectra, as introduced in equation (6
). In this approach, the signal from a reference region without hydrogen (or without PE) is divided by that of a hydrogen-containing region, effectively isolating the hydrogen contribution while largely cancelling out systematic effects originating from the sample and instrument.
For this analysis, the density of PE was fixed at 0.92 g cm−3 and we used the σPE(λ) data in Fig. 1
(b) as numerical values for each wavelength bin. The resulting transmission ratios are shown in Fig. 10
for both front and back configurations. The corresponding fits show excellent agreement with the experimental data. The fitted PE thicknesses are consistent between both configurations, yielding lthin,fit = 44 (1) µm and lthick,fit = 228 (1) µm for the front, and lthin,fit = 45 (1) µm and lthick,fit = 229 (1) µm for the back configuration.
|
|
Figure 10
Experimental (markers) and fitted (lines) neutron transmission spectra ratios for the SDSS sample measured on IMAT: (a) front and (b) back measurements. The NCrystal-based least-squares fits quantitatively reproduce this attenuation trend, yielding PE thicknesses with 1 µm uncertainty, as listed. See text for further discussion. |
Given the nominal PE sheet thicknesses of 45 µm and 250 µm, our results show excellent agreement for the thin PE but underestimate the thick sheet by approximately 10%. Discussions regarding the discrepancy of the thick PE thickness, which may stem from the reduced hydrogen-to-carbon (H/C) ratio in this particular sample, are detailed in Appendix A6
.
5.2.2. RADEN measurements
Section 5.2.1
details the data analysis procedure for the IMAT experiment, which includes both nonlinear least-squares fitting on wavelength-resolved transmission spectra and spectral ratio fitting to determine PE thickness. In contrast, the subsequent RADEN data analysis focuses specifically on fitting the PE thickness.
Transmission ratios were obtained by dividing the spectra of the SDSS sample without PE by those with PE for Zones 1 to 5 shown in Fig. 7
(c). The resulting ratios for the selected ROIs are shown in Fig. 11
(a). Following the fitting procedure detailed above, the PE thicknesses were fitted across two distinct intervals: the full neutron wavelength range (1.5–6.0 Å) and the range after the Bragg cutoff (4.5–6.0 Å). The fitted thicknesses and their associated uncertainties are listed in the figure. Notably, the fitted PE thicknesses obtained from both wavelength ranges are consistent within their uncertainties.
|
Figure 11
(a) Experimental and fitted neutron transmission spectra ratios for Zones 1–5. Experimental data (binned for clarity) are shown alongside fits over the full wavelength range (1.5–6.0 Å, solid lines) and after the Bragg cutoff (4.5–6.0 Å, dashed lines). Inset values indicate the corresponding fitted PE thicknesses. For comparison, the transmission ratios calculated using nominal (dotted lines) and micrometer-measured (dashed–dotted lines) PE thicknesses are also shown. (b) and (c) Two-dimensional fitted PE thickness maps for the wavelength ranges of 1.5–6.0 Å and 4.5–6.0 Å, respectively. See text for further discussion. |
The reason for performing a fit after the Bragg cutoff stems from the fact that hydrogen can induce lattice expansion in steels, as recently studied by e.g. Lindblom et al. (2026
). Additionally, hydrogen could potentially change the extinction and/or texture effects. These influences might distort the transmission ratio prior to the cutoff. Nevertheless, incorporating the lower-wavelength data remains critical for investigating the dynamics of hydrogen in dual-phase steels.
By applying this analysis systematically, we generated 2D thickness maps fitted for both the 1.5–6.0 Å and 4.5–6.0 Å ranges, as shown in Figs. 11
(b) and 11
(c), respectively. The fitted PE thicknesses in the middle and rightmost regions are in excellent agreement with their nominal values. However, the fitted PE thickness on the left is approximately 30% higher than its nominal value. To investigate this, the PE sheets were measured using a Mitutoyo Digimatic micrometer. As mentioned previously, the nominal thicknesses of the three PE sheets (from left to right) are 100, 35 and 230 µm; the corresponding measured thicknesses are 139, 39 and 241 µm, respectively. While these micrometer measurements are all roughly 10% higher than the fitted values, they do not account for the 30% discrepancies on the left. For a better illustration, the transmission ratios calculated using these PE thicknesses are shown in Figs. 11
(a). Consequently, a plausible explanation for the leftmost discrepancy is that the supplier inadvertently sent a PE sheet with a nominal thickness of 125 or 130 µm instead of the ordered specification. Notably, a pristine PE sheet with a nominal thickness of 230 µm (the same thickness as in the IMAT measurements) was accurately determined with the RADEN measurements. This result strongly supports our previous hypothesis regarding a stoichiometric deviation in the neutron-irradiated PE that was measured on IMAT (see Appendix A6
).
In regions without PE sheets, the fitted PE thicknesses are close to zero, maintaining an uncertainty within 4 µm, which corresponds to 11 wt p.p.m. hydrogen in PE. Given these overall fitting uncertainties and the background values obtained in the bare regions, the methodology developed in this work successfully resolves a hydrogen detection limit down to the 10 wt p.p.m. level.
In this work, quantifying additional hydrogen signals relies heavily on a reference sample or region. For the RADEN measurements, we utilized the exact same SDSS sample for reference. This setup, combined with a precise cross-section model and well controlled experimental instrumentation, enables us to push the hydrogen detection limit down to the 10 wt p.p.m. level. A reference sample is critical when trying to detect such a weak signal, which represents less than a 0.5% change. At this level, omitting the reference sample would compromise the validity of the method; for instance, any inhomogeneity in sample thickness or composition would cause variations in transmission intensity, making it difficult to distinguish whether the signal change is driven by hydrogen or variations in other physical quantities. While this method remains applicable to the IMAT measurements, where we lacked a strict reference and instead used a `similar' SDSS region (in terms of thickness and element composition), the significant contrast (100 wt p.p.m. hydrogen variation) and the homogeneity between these regions made detection feasible.
6. Conclusions
Motivated by the need for quantitative detection of very small hydrogen signals in metallic materials, this work establishes a universal and comprehensive analysis workflow for wavelength-resolved neutron transmission imaging. The workflow is built entirely on open-source components – most critically the NCrystal toolkit – and is designed to enable reliable quantification even when hydrogen-equivalent concentrations are far below levels accessible with conventional imaging.
A series of key contributions have been demonstrated:
(i) A systematic simulation study was performed to quantify forward-scattering artefacts under different sample-to-detector distances. By benchmarking OpenMC+NCrystal against the established work by Sato et al. (2017
) and extending the simulations to a wide range of materials and experimental slit configurations, we provide practical guidance on when forward scattering can be safely neglected and when it becomes significant and requires correction. A few examples for different materials are provided in Appendix A
to offer interested readers practical guidance when planning experiments.
(ii) The developed least-squares fitting routine – based on NCrystal cross sections – accurately reproduced the measured spectra after applying the normalization and correction procedures described in this work. The model effectively incorporated the known chemical compositions while adjusting microstructural parameters such as extinction and lattice constants.
(iii) Using Super Duplex stainless steel with polyethylene attachments as a controlled proxy for low hydrogen-equivalent content, we have demonstrated that the workflow can reliably determine polyethylene thicknesses with micrometre-level uncertainty, validating the hydrogen detection limit down to 10 wt p.p.m. needed for low-level hydrogen quantification. The same workflow is readily applicable to quantifying other elements and is expected to be highly useful for a wide variety of time-of-flight neutron imaging experiments.
(iv) All analysis tools, including the fitting script, have been made publicly available to support reproducibility and to facilitate the wider adoption of this workflow within the neutron imaging community.
(v) More generally, this work highlights that the dynamic state of atoms contributes directly to wavelength-resolved neutron transmission and therefore represents an additional contrast mechanism, namely spectroscopic contrast. While this must be accounted for during quantitative analysis, it also opens new opportunities, e.g. to distinguish different hydrogen trapping states and local atomic environments using neutron imaging.
Overall, the results highlight that well characterized reference states are essential for absolute quantification, particularly when targeting trace hydrogen concentrations. In this context, in situ experiments, where the hydrogen-free initial state of the material serves as an internal reference, are essential for achieving accurate quantitative analysis.
The comparison between the IMAT and RADEN measurements further indicates that detector-related systematic effects can introduce wavelength-dependent offsets in normalized transmission spectra. Although the exact origin of the detector-generated background could not be identified conclusively, the results demonstrate that such systematic effects should be considered in highly quantitative wavelength-resolved neutron imaging, and that the implemented OA-based correction effectively compensates for the resulting wavelength-dependent offset.
More broadly, this study demonstrates that the combination of rigorous Monte Carlo simulations as a digital twin, physically grounded cross-section modelling and well controlled experimental instrumentation enables quantitative neutron imaging at pulsed neutron sources. In the future, more detailed beamline models, such as those implemented in Monte Carlo codes like McStas or VITESS, could be incorporated to improve further the accuracy of the simulations beyond the simplified beam geometry used here. The workflow presented here establishes a practical and transferable approach for routinely detecting and quantifying weak transmission signals across a wide range of materials and applications. In this context, the quantitative detection of low-level hydrogen down to 10 wt p.p.m. (modelled by polyethylene in this work) in Fe-based engineering alloys, central to the study of hydrogen embrittlement, represents a particularly important application.
APPENDIX A
A1. Validation of OpenMC+NCrystal simulations
Our OpenMC+NCrystal simulations replicated the configuration shown by Sato et al. (2017
), thereby modelling a pulsed neutron source of 10 × 10 cm with zero divergence incident on a b.c.c. iron plate, with a TOF detector positioned a distance L downstream. A schematic presentation of the configuration is illustrated in Fig. 2
(a). To allow for a direct comparison with the results shown by Sato et al. (2017
), a first simulation was performed for a 6 mm thick plate using 2 × 1010 neutrons distributed uniformly between 1 and 6 Å, ensuring sufficient statistics for the transmission spectrum calculations. The detector distance was varied from 0 to 150 cm from the sample, while the ROI was maintained at 10 × 10 cm. The resulting transmission spectra are presented in Fig. 12
(a). The theoretically predicted transmission spectrum is shown as a solid black line. Transmission signal artefacts increase as the sample-to-detector distance decreases. This effect is most pronounced at neutron wavelengths below the 110 Bragg edge, confirming that the increased transmission signal is primarily due to the detection of forward Bragg-scattered neutrons at small sample-to-detector distances. Our simulation results are in good agreement with Fig. 2 of Sato et al. (2017
), thus verifying the capability of OpenMC+NCrystal in modelling transmission experiments.
|
Figure 12
OpenMC+NCrystal simulated transmission spectra for a b.c.c. iron plate illuminated by a 10 × 10 cm parallel neutron beam. The setup replicates the configurations presented by Sato et al. (2017 |
A second simulation was performed for an 18 mm thick plate using the same configuration. The resulting transmission spectra are compared with the theoretical predictions in Fig. 12
(b). As expected, the transmission intensities are lower than those for the 6 mm plate due to increased attenuation. Both simulated spectra exhibit a similar trend of decreasing forward-scattering artefacts with increasing sample-to-detector distance. Additionally, both simulations confirm that the impact of forward scattering can be neglected only at distances greater than 50 cm for this specific configuration.
We investigated the influence of detection area and position on the transmission spectra to compare our results with those presented by Sato et al. (2017
). A third simulation was conducted, identical to the first but employing 2 × 1011 neutrons to improve statistics for the smaller pixel-sized ROIs (0.8 × 0.8 mm). The resulting spectra for a pixel located at the centre and one at the corner (4.2 cm, 4.2 cm) are shown in Figs. 12
(c) and 12
(d), respectively. These results are in very good agreement with Figs. 5 and 6 in Sato et al. (2017
). The central pixel displays more pronounced forward-scattering artefacts compared with the corner pixel.
By employing the same configurations as the reference study, our OpenMC+NCrystal simulation results align closely with those reported by Sato et al. (2017
), validating our simulation framework.
A2. Simulations for more materials in different sample-to-detector distances
Having established a comprehensive OpenMC+NCrystal simulation toolkit in this work, we broadened the scope of the study by performing simulations across several commonly investigated materials, making use of the extensive NCrystal database of crystalline and amorphous solids, liquids and gases. This broader survey aims to compare how strongly forward-scattering artefacts manifest across different classes of materials. Additional results are presented in Figs. 13
and 14
, which may serve as practical reference cases for researchers designing quantitative neutron transmission experiments for different material systems.
|
Figure 13
Comparison of simulated and theoretically predicted transmission spectra (solid black lines) for 8 mm thick Al, Be, Cu and Ni across simulation Cases 1, 2 and 3. The results indicate that the sample-to-detector distances of 50 cm, 5 cm and 2 cm are sufficient to mitigate forward-scattering artefacts in Cases 1, 2 and 3, respectively. |
|
Figure 14
Comparison of simulated and theoretically predicted transmission spectra (solid black lines) for 8 mm thick Al2O3, ZrO2, polylactide (C3H4O2) and polyvinylchloride (PVC; C2H3Cl) across simulation Cases 1, 2 and 3. The results indicate that the sample-to-detector distances of 50 cm, 5 cm and 2 cm are sufficient to mitigate forward-scattering artefacts in Cases 1, 2 and 3 for Al2O3 and ZrO2. However, polylactide and PVC necessitate larger sample-to-detector distances of 150 cm, 15 cm and 5 cm for the three cases, respectively. |
A3. Discussion regarding the forward-scattering effects across different ROIs
As shown in Fig. 4
, Monte Carlo simulations prove that the effects of forward scattering can be safely neglected for the selected ROIs (corresponding to no PE, thin PE and thick PE, respectively) in the IMAT experimental setup at a sample-to-detector distance of 20 mm. Figs. 15
(a) and 15
(b) illustrate these effects at zero sample-to-detector distance. Specifically, Fig. 15
(a) highlights the forward-scattering artefacts for the three ROIs, while Fig. 15
(b) displays the ratio between the simulated transmission Tsim(λ) and the theoretical transmission Ttheo(λ). The ratios demonstrate that the effects are highly similar across these ROIs; because the attached PE sheets are relatively thin (≤ 0.25 mm), any additional scattering effects remain negligible.
|
|
Figure 15
(a) Forward-scattering artefacts at zero sample-to-detector distance for three different ROIs. (b) Ratios of simulated to theoretical transmission spectra, highlighting that the forward-scattering effects remain consistent across these ROIs. The drops in the data near the Bragg cutoff (4.2 Å) are an artefact of the binning in the simulated transmission spectra. |
A4. Discussion regarding the OA correction
In this section, we discuss how the choice of ROI within the OA region impacts the corrected transmission intensity. First, we calculated the wavelength-resolved transmission spectra for two specific ROIs, denoted OA1 and OA2 in Fig. 16
(a) [where OA1 is identical to the ROI in Fig. 6
(a)]. The corresponding neutron transmission spectra are shown in Fig. 16
(b).
|
|
Figure 16
(a) Radiograph measured for the front configurations measured on IMAT. Two ROIs within the OA regions are indicated, denoted OA1 and OA2. (b) Corresponding wavelength-resolved neutron transmission spectra for OA1 and OA2. The spectrum of OA1 is identical to that shown in the upper plot of Fig. 6(c). |
The transmission spectra for both OA regions exhibit a wavelength dependency. Additionally, the transmission intensity of OA1 is consistently higher than that of OA2. By dividing the transmission spectrum of each SDSS ROI by the wavelength-dependent intensity of OA1, we obtained the OA-corrected transmission spectra, as illustrated in Fig. 6
(c). As shown in Fig. 6
(d), using OA1 to evaluate the transmission intensity yields a consistent ratio of unity between the front and back transmission.
Conversely, if OA2 is used for the correction, a consistent ratio of 1.01 is observed between the transmission spectra of the front and back configurations. While this minor deviation does not impact the final PE thickness fits, as it cancels out when dividing the transmission of the no-PE region by that of the PE-attached region, this observation underscores the necessity of using a reference region on the same sample as a baseline. This internal baseline ensures that systematic effects cancel out, thereby enhancing the overall robustness of the methodology and preventing minor experimental variations from compromising the accuracy of the analysis.
Such precision is particularly critical when quantifying signals from weak contributions, where precise baseline subtraction is required to isolate subtle spectral variations. It should be noted that, if left unaccounted for, a 1% deviation in the transmission signal would correspond to approximately 30 wt p.p.m. of hydrogen in the current sample configuration. When aiming to detect such small signal changes, ruling out systematic errors from the experimental setup is of the utmost importance. The observed OA transmission below unity cannot be explained by scattering from the sample into the OA, as such scattering would instead increase the detected intensity. Instead, the observations are consistent with detector-related parasitic scattering. A plausible explanation is that the detector-generated background (generated inside the detector) differs between the open-beam and sample measurements because the presence of the sample modifies the neutron field experienced by the detector system. Consequently, normalization using the open-beam measurement produces a wavelength-dependent offset in the apparent transmission. This interpretation is further supported by the repeat measurements performed on the RADEN beamline using a different detector configuration, where the OA transmission remained much closer to unity over the full wavelength range [top panel of Fig. 7
(c)]. While the exact origin of the detector-related background cannot be identified conclusively from the present measurements, the comparison indicates that the effect is associated with the detector configuration rather than with changes in the incident neutron spectrum.
A5. Parameter determination of Jorgensen resolution function for IMAT
According to Von Dreele et al. (1982
), the resolution can be calculated using
where C is a characteristic constant of the instrument. By parameterizing α = α1/d, β = β0 + β1/d4 and Σ = Σ1d (where d is the interplanar spacing), the resolution function can be written as
Additionally, because the parameter β represents the decay process in the Jorgensen model, a corresponding decay time can be defined as
where C′ is a scaling constant.
The decay time and resolution functions were fitted simultaneously against the wavelength-dependent experimental data digitized from Fig. 6 of Ramadhan et al. (2019
). During the nonlinear least-squares refinement using lmfit, the parameters (Σ1 = 1.43 × 10−3, α1 = 100000 and C′ = 20000) were kept fixed. Specifically, Σ1 was predetermined from the ratio of the moderator thickness (8 cm) of ISIS Target Station 2 and the flight path length (56.1 m) of IMAT. The values of α1 and C′ were fixed, because allowing them to vary freely did not significantly improve the quality of the fit, but instead introduced large uncertainties in the remaining parameters.
As shown in Fig. 17
, the fitted curves are in good agreement with the experimental data, yielding β0 = 45 (2), β1 = 58 (6) and C = 0.0140 (5). Consequently, these refined parameters for the Jorgensen-type resolution function of the IMAT instrument can be fixed for transmission spectrum fittings for the SDSS data (Section 5.2.1
). This fitting approach for the Jorgensen-type resolution function can be readily applied to other pulsed neutron beamlines, such as RADEN.
|
Figure 17
(a) Decay time τ and (b) resolution as a function of neutron wavelength on IMAT. The experimental data shown in both panels were digitized from Ramadhan et al. (2019 |
A6. Discussion regarding the H/C ratios for the thick PE sheet measured on IMAT
The 10% discrepancy of the fitted thickness for the thick PE sample may stem from a reduced hydrogen-to-carbon (H/C) ratio, dropping from the stoichiometric value of 2 to approximately 1.8. This reduction is likely to be caused by radiation-induced material modifications, as this particular sample had previously been exposed to neutrons for extended periods at the ISIS facility.
Under neutron irradiation, the actual H/C ratio can fall below 2 if the material undergoes dehydrogenation: radiation creates radicals, some H is abstracted, H2 can evolve and unsaturated groups such as C=C can form. In irradiated PE, the reported chemistry includes crosslinking, chain scission, oxidation and formation of double bonds, with explicit evidence that some hydrogen atoms important for neutron shielding are released. Neutron interactions in hydrocarbon polymers are also often treated as producing recoil protons, so the radiation chemistry resembles proton irradiation (O'Donnell, 1989
). For polymer neutron shields in storage/transport systems, the literature notes that, in many practical cases, thermal ageing and oxidation dominate over irradiation as degradation mechanisms (Wolff et al., 2019
). Notably, the pristine PE sheets used in the RADEN measurements do not show this effect (see Section 5.2.2
).
A7. Fitting the PE thickness without using a reference
In practice, the PE thickness can be fitted simultaneously with all other free parameters for the PE-covered regions. As an example, Fig. 18
shows the fitting results of the transmission spectrum of Zone 3 measured on RADEN [shown in the middle panel of Fig. 7
(c)].
|
Figure 18
NCrystal-based least-squares fit of the SDSS sample with PE covers, using transmission spectrum data from Zone 3 [indicated in the middle panel of Fig. 7(c)]. |
The fitting transmission shows good agreement with the experimental data, yielding a reduced χ2 value of 2.4. The directly fitted PE thickness is 68 (7) µm, which is nearly twice the nominal value (35 µm). This discrepancy is primarily attributed to a strong correlation (0.97) between the PE thickness and the scale factor; in contrast, the correlation between the thickness and all other free parameters remains below 0.25. Consequently, any systematic error in the scale factor directly propagates into the estimated PE thickness. This highlights the importance of utilizing a reference sample, as dividing the transmission spectra minimizes the impact of the scale factor. Alternatively, one can determine the scale factor and other free parameters from the reference sample and fix them when fitting the PE thickness from the PE-covered region. Although this indirect approach fits the spectrum directly rather than the ratio, it is mathematically equivalent to fitting the spectral ratio shown in Figs. 10
and 11
.
Acknowledgements
The authors express their gratitude to Matthew Krzystyniak from the ISIS Neutron and Muon Source for provision of reference samples and helpful discussions. The authors also thank Javier Santisteban for helpful discussions. The ISIS Neutron and Muon Source is acknowledged for access to neutron beam facilities (DOI: 10.5286/ISIS.E.RB2320208). Support during the preparation and execution of the experiments on IMAT from Armin Halilovic, Xiaoqi Yue, León Zendejas Medina, Mary-Ellen Donnelly and Edvin Lundgren is greatly appreciated. The authors would also like to express their sincere gratitude for the invaluable help with the experiments on the RADEN imaging beamline at J-PARC (Japan Proton Accelerator Research Complex) from Tobias Neuwirth, Clas Dahlin, Premysl Beran, Yuhua Su and Takenao Shinohara. Materials and technical expertise were provided by industrial partners at Alleima, in particular Ulf Kivisäkk and Lisa Lautrup. The neutron experiment at the Materials and Life Science Experimental Facility on the BL22 RADEN beamline at J-PARC was performed under a user program (proposal No. 2024B0313).
Conflict of interest
A. Losko is the director of LoskoVision GmbH and A. Wolfertz received a salary as an employee of LoskoVision GmbH. LoskoVision GmbH is a startup that offers detector solutions using technologies described in this work. All other authors report no competing interests.
Data availability
The data supporting the results reported can be accessed upon request.
Funding information
This work was partially supported by the VR (Vetenskapsrådet) project `Revealing Hydrogen Embrittlement of Alloys in Real Time' (ReHEART) (grant No. 2022-06059) and by the BMBF in the framework of the research project 05K22WO5.
References
Aswal, D. K., Sarkar, P. S. & Kashyap, Y. S. (2022). Editors. Neutron Imaging. Singapore: Springer Nature.
Google Scholar
AZo Materials (2025). Stainless steel grade Super Duplex 2507 (UNS S32750). https://www.azom.com/article.aspx?ArticleID=9185.
Google Scholar
Becker, P. J. & Coppens, P. (1974a). Acta Cryst. A30, 129–147.
CrossRef
IUCr Journals
Web of Science
Google Scholar
Becker, P. J. & Coppens, P. (1974b). Acta Cryst. A30, 148–153.
CrossRef
IUCr Journals
Web of Science
Google Scholar
Beskrovni, A., Danilkin, S., Fuess, H., Jadrowski, E., Neova-Baeva, M. & Wieder, T. (1999). J. Alloys Compd. 291, 262–268.
CrossRef
CAS
Google Scholar
Bilheux, H. Z., Bilheux, J. C., Hanks, R. A., Byrd, A., Jones, A., Skorpenske, H., Donnelly, M. E., Mangus, R., Conner, D., Lessard, T. L., Venkatakrishnan, S., Tang, S., Thomasson, T., Keener, W. S. & Rennich, G. (2023). J. Phys. Conf. Ser. 2605, 012004.
CrossRef
Google Scholar
Boillat, P., Carminati, C., Schmid, F., Grünzweig, C., Hovind, J., Kaestner, A., Mannes, D., Morgano, M., Siegwart, M., Trtik, P., Vontobel, P. & Lehmann, E. (2018). Opt. Express 26, 15769–15784.
CrossRef
CAS
PubMed
Google Scholar
Boin, M. (2011). PhD thesis. The Open University, Milton Keynes, UK.
Google Scholar
Buitrago, N., Santisteban, J., Tartaglione, A., Marín, J., Barrow, L., Daymond, M., Schulz, M., Grosse, M., Tremsin, A., Lehmann, E., Kaestner, A., Kelleher, J. & Kabra, S. (2018). J. Nucl. Mater. 503, 98–109.
CrossRef
CAS
Google Scholar
Burca, G., Kockelmann, W., James, J. A. & Fitzpatrick, M. E. (2013). J. Instrum. 8, P10001.
Google Scholar
Cai, X.-X. & Kittelmann, T. (2020). Comput. Phys. Commun. 246, 106851.
Web of Science
CrossRef
Google Scholar
Cai, X.-X., Kittelmann, T., Klinkby, E. & Márquez Damián, J. (2019). J. Comput. Phys. 380, 400–407.
CrossRef
CAS
Google Scholar
Carpenter, J. M. & Loong, C.-K. (2015). Elements of Slow-Neutron Scattering: Basics, Techniques and Applications. Cambridge University Press.
Google Scholar
Danilkin, S. A., Fuess, H., Wipf, H., Ivanov, A., Gavriljuk, V. G., Delafosse, D. & Magnin, T. (2003). Europhys. Lett. 63, 69–75.
CrossRef
CAS
Google Scholar
Hirsh, T. Y., Leong, A. F. T., Losko, A. S., Wolfertz, A., Savage, D. J., Jäger, T. T., Rakovan, J., Wall, J. J., Long, A. M. & Vogel, S. C. (2025a). Sci. Rep. 15, 12901.
CrossRef
PubMed
Google Scholar
Hirsh, T. Y., Long, A. M., Losko, A. S., Jaeger, T. T., Wolfertz, A. & Vogel, S. C. (2025b). Sci. Rep. 15, 14026.
CrossRef
PubMed
Google Scholar
Jäger, T. T., Hirsh, T. Y., Scheuren, S., Long, A. M., Losko, A. S., Wolfertz, A., Zimmer, M., Roth, M. & Vogel, S. C. (2025). Sci. Rep. 15, 6823.
PubMed
Google Scholar
Jäger, T. T., Losko, A. S., Wolfertz, A., Schmidt, S., Bertelsen, M., Khaplanov, A., Agnew, S. R., Funama, F., Morgano, M., Roth, M., Gochanour, J. R., Long, A. M., Lutterotti, L. & Vogel, S. C. (2024). J. Appl. Cryst. 57, 1107–1114.
Web of Science
CrossRef
IUCr Journals
Google Scholar
Jorgensen, J. D., Johnson, D. H., Mueller, M. H., Peterson, S. W., Worlton, J. G. & Von Dreele, R. B. (1978). Proceedings of the Conference on Diffraction Profile Analysis, Krakow, Poland, pp. 14–15 August 1978, pp. 20–22.
Google Scholar
Kardjilov, N., de Beer, F., Hassanein, R., Lehmann, E. & Vontobel, P. (2005). Nucl. Instrum. Methods Phys. Res. A 542, 336–341.
CrossRef
CAS
Google Scholar
Kardjilov, N., Manke, I., Woracek, R., Hilger, A. & Banhart, J. (2018). Mater. Today 21, 652–672.
CrossRef
CAS
Google Scholar
Khaplanov, A., Ankner, J. F., Hirsh, T. Y., Losko, A. S., Long, A. M., Loyd, M., Sykora, G. J., Vogel, S. C., Watkins, E. B. & Wolfertz, A. (2025). Sci. Rep. 15, 25014.
Web of Science
CrossRef
PubMed
Google Scholar
Kittelmann, T. & Cai, X.-X. (2021). Comput. Phys. Commun. 267, 108082.
Web of Science
CrossRef
Google Scholar
Kittelmann, T., DiJulio, D. D., Xu, S. & Damián, J. I. M. (2026). Acta Cryst. A82, 163–178.
CrossRef
IUCr Journals
Google Scholar
Kockelmann, W., Minniti, T., Pooley, D. E., Burca, G., Ramadhan, R., Akeroyd, F. A., Howells, G. D., Moreton-Smith, C., Keymer, D. P., Kelleher, J., Kabra, S., Lee, T. L., Ziesche, R., Reid, A., Vitucci, G., Gorini, G., Micieli, D., Agostino, R. G., Formoso, V., Aliotta, F., Ponterio, R., Trusso, S., Salvato, G., Vasi, C., Grazzi, F., Watanabe, K., Lee, J. W. L., Tremsin, A. S., McPhate, J. B., Nixon, D., Draper, N., Halcrow, W. & Nightingale, J. (2018). J. Imaging 4, 47.
CrossRef
Google Scholar
Lindblom, D., Halilović, A. E., Woracek, R., Tengattini, A., Helfen, L. & Dahlberg, C. F. (2024). Mater. Sci. Eng. A 895, 146215.
CrossRef
Google Scholar
Lindblom, D., Liu, M., Pan, J., Woracek, R. & Dahlberg, C. F. O. (2026). Int. J. Plast. 196, 104546.
CrossRef
Google Scholar
Losko, A. S., Han, Y., Schillinger, B., Tartaglione, A., Morgano, M., Strobl, M., Long, J., Tremsin, A. S. & Schulz, M. (2021). Sci. Rep. 11, 21360.
Web of Science
CrossRef
PubMed
Google Scholar
Minniti, T., Coppola, R., Kockelmann, W., Tremsin, A. S., Roccella, S., Greuner, H. & You, J.-H. (2025). Nucl. Fusion 65, 026041.
CrossRef
Google Scholar
Mirashi, V. U., Kockelmann, W., Kapur, N., Fedrigo, A., Morgano, M., Chulapakorn, T., Losko, A., Wolfertz, A., Tremsin, A. S. & Matamoros-Veloza, A. (2025). Nucl. Instrum. Methods Phys. Res. A 1073, 170284.
CrossRef
Google Scholar
Mirashi, V. U., Kockelmann, W., Kapur, N. & Matamoros-Veloza, A. (2026). Nucl. Instrum. Methods Phys. Res. A 1088, 171490.
CrossRef
Google Scholar
Nelson, R. O., Vogel, S. C., Hunter, J. F., Watkins, E. B., Losko, A. S., Tremsin, A. S., Borges, N. P., Cutler, T. E., Dickman, L. T., Espy, M. A., Gautier, D. C., Madden, A. C., Majewski, J., Malone, M. W., Mayo, D. R., McClellan, K. J., Montgomery, D. S., Mosby, S. M., Nelson, A. T., Ramos, K. J., Schirato, R. C., Schroeder, K., Sevanto, S. A., Swift, A. L., Vo, L. K., Williamson, T. E. & Winch, N. M. (2018). J. Imaging 4, 45.
CrossRef
Google Scholar
Newville, M. et al. (2024). lmfit/lmfit-py. Version 1.3.2. https://doi.org/10.5281/zenodo.12785036.
Google Scholar
O'Donnell, J. H. (1989). The Effects of Radiation on High-Technology Polymers edited by E. Reichmanis & J. H. O'Donnell, ch. 1, pp. 1–13. American Chemical Society Symposium Series, Vol. 381.
Google Scholar
Ramadhan, R. S., Kockelmann, W., Minniti, T., Chen, B., Parfitt, D., Fitzpatrick, M. E. & Tremsin, A. S. (2019). J. Appl. Cryst. 52, 351–368.
Web of Science
CrossRef
CAS
IUCr Journals
Google Scholar
Ramić, K., Wendorff, C., Cheng, Y., Kolesnikov, A. I., Abernathy, D. L., Daemen, L., Arbanas, G., Leal, L., Danon, Y. & Liu, L. E. (2018). Ann. Nucl. Energy 120, 778–787.
Google Scholar
Robledo, J. I., Violini, N., Beule, F., Voigt, J., Zakalek, P. & Lieutenant, K. (2026). J. Appl. Cryst. 59, 678–686.
CrossRef
CAS
IUCr Journals
Google Scholar
Romano, P. K., Horelik, N. E., Herman, B. R., Nelson, A. G., Forget, B. & Smith, K. (2015). Ann. Nucl. Energy 82, 90–97.
CrossRef
CAS
Google Scholar
Sabine, T. M., Von Dreele, R. B. & Jørgensen, J.-E. (1988). Acta Cryst. A44, 374–379.
CrossRef
CAS
Web of Science
IUCr Journals
Google Scholar
Santisteban, J. R., Edwards, L., Steuwer, A. & Withers, P. J. (2001). J. Appl. Cryst. 34, 289–297.
Web of Science
CrossRef
CAS
IUCr Journals
Google Scholar
Sato, H., Kamiyama, T. & Kiyanagi, Y. (2011). Mater. Trans. 52, 1294–1302.
Web of Science
CrossRef
CAS
Google Scholar
Sato, H., Watanabe, K., Kiyokawa, K., Kiyanagi, R., Hara, K., Kamiyama, T., Furusaka, M., Shinohara, T. & Kiyanagi, Y. (2017). Phys. Procedia 88, 322–330.
CrossRef
CAS
Google Scholar
Sato, T., Niita, K., Matsuda, N., Hashimoto, S., Iwamoto, Y., Noda, S., Ogawa, T., Iwase, H., Nakashima, H., Fukahori, T., Okumura, K., Kai, T., Chiba, S., Furuta, T. & Sihver, L. (2013). J. Nucl. Sci. Technol. 50, 913–923.
CrossRef
CAS
Google Scholar
Schober, H. (2014). J. Neutron Res. 17, 109–357.
CrossRef
CAS
Google Scholar
Sebold, S. R., Neuwirth, T., Tengattini, A., Cubitt, R., Gilch, I., Mühlbauer, S. & Schulz, M. (2024). Sci. Rep. 14, 19348.
CrossRef
PubMed
Google Scholar
Shinohara, T., Kai, T., Oikawa, K., Nakatani, T., Segawa, M., Hiroi, K., Su, Y., Ooi, M., Harada, M., Iikura, H., Hayashida, H., Parker, J. D., Matsumoto, Y., Kamiyama, T., Sato, H. & Kiyanagi, Y. (2020). Rev. Sci. Instrum. 91, 043302.
Web of Science
CrossRef
PubMed
Google Scholar
Silverstein, R., Sobol, O., Boellinghaus, T., Unger, W. & Eliezer, D. (2017). J. Alloys Compd. 695, 2689–2695.
CrossRef
CAS
Google Scholar
Simoni, M., Minniti, T., Senesi, R. & Romanelli, G. (2023). Phys. Chem. Chem. Phys. 25, 30821–30831.
CrossRef
CAS
PubMed
Google Scholar
Squires, G. L. (2012). Introduction to the Theory of Thermal Neutron Scattering, 3rd ed. Cambridge University Press.
Google Scholar
Strobl, M. (2015). Phys. Procedia 69, 18–26.
CrossRef
CAS
Google Scholar
Strobl, M., Baur, M. E., Samothrakitis, S., Malamud, F., Zhang, X., Tung, P. K., Schmidt, S., Woracek, R., Lee, J., Kiyanagi, R., Kuhn, L. T., Gavish Segev, I. & Ein–Eli, Y. (2025). Adv. Energy Mater. 15, 2405238.
CrossRef
Google Scholar
Strobl, M., Kardjilov, N., Hilger, A., Kühne, G., Frei, G. & Manke, I. (2009). Nucl. Instrum. Methods Phys. Res. A 604, 640–645.
CrossRef
CAS
Google Scholar
Strobl, M. & Lehmann, E. (2024). Editors. Neutron Imaging: From Applied Materials Science to Industry. IOP Publishing.
Google Scholar
Sumarli, S., Malamud, F., Petegem, S. V., Gaudez, S., Baganis, A., Busi, M., Polatidis, E., Leinenbach, C., Logé, R. E. & Strobl, M. (2024). Virtual Phys. Prototyping 19, e2429132.
CrossRef
Google Scholar
Tengattini, A., Lenoir, N., Andò, E., Giroud, B., Atkins, D., Beaucour, J. & Viggiani, G. (2020). Nucl. Instrum. Methods Phys. Res. A 968, 163939.
CrossRef
Google Scholar
Tran, K., Woracek, R., Kardjilov, N., Markötter, H., Hilger, A., Kockelmann, W., Kelleher, J., Puplampu, S., Penumadu, D., Tremsin, A., Banhart, J. & Manke, I. (2021). Materials Today Advances 9, 100132.
CrossRef
Google Scholar
Tremsin, A. & Vallerga, J. (2020). Radiat. Meas. 130, 106228.
CrossRef
Google Scholar
Vogel, S. (2000). PhD thesis. Christian-Albrechts Universität Kiel, Germany.
Google Scholar
Von Dreele, R. B., Jorgensen, J. D. & Windsor, C. G. (1982). J. Appl. Cryst. 15, 581–589.
CrossRef
CAS
Web of Science
IUCr Journals
Google Scholar
Willendrup, P. K. & Lefmann, K. (2020). J. Neutron Res. 22, 1–16.
Web of Science
CrossRef
Google Scholar
Willendrup, P. K. & Lefmann, K. (2021). J. Neutron Res. 23, 7–27.
Web of Science
CrossRef
Google Scholar
Wolfertz, A., Gustschin, A., Schulz, M., Long, A. M., Khaplanov, A., Hirsh, T. Y., Nomerotski, A., Morgano, M., Tremsin, A., Mauri, G., Sykora, G. J. & Losko, A. (2024a). Sci. Rep. 14, 30495.
CrossRef
PubMed
Google Scholar
Wolfertz, A., Losko, A., Long, A. M., Brodish, S., Craft, A. E., Khaplanov, A., Vogel, S. C., Nelson, R. O., Wender, S. A., Tremsin, A., Hirsh, T. Y., Jäger, T. T., Morgano, M. & Feng, P. (2024b). Sci. Rep. 14, 30487.
CrossRef
PubMed
Google Scholar
Wolff, D., Völzke, H., Bevilacqua, A., Alejano, C., Conde, J. M., Einziger, R. E., Fukuda, S. I., González Espartero, A., Gouzy-Portaix, S., Haddad Andalaf, R. E., Hambley, D., Issard, H., Kessler, J., Kurpaska, L., Legat, A., Qureshi, A. H., Ruiz, J., Shirai, K., Šmaižys, A. & Verrastro, C. A. (2019). Demonstrating Performance of Spent Fuel and Related Storage System Components During Very Long Term Storage – Final Report of a Coordinated Research Project, IAEA TECDOC series (ISSN 1011-4289), Vol. 1878, pp. 1–198. Vienna, Austria; Bundesanstalt für Materialforschung und -prüfung (BAM), Berlin, Germany: IAEA Publishing Section.
Google Scholar
Woracek, R., Santisteban, J., Fedrigo, A. & Strobl, M. (2018). Nucl. Instrum. Methods Phys. Res. A 878, 141–158.
Web of Science
CrossRef
CAS
Google Scholar
Xu, S. (2024). ncplugin-CrysExtn. https://github.com/XuShuqi7/ncplugin-CrysExtn.
Google Scholar
Xu, S. (2025). NCrystal-based Neutron Imaging Fitting Script. https://github.com/XuShuqi7/NCrystal-based_Neutron_Imaging_Fitting_Script.
Google Scholar
Xu, S., DiJulio, D. D., Marquez Damian, J. I., Kittelmann, T., Bernasconi, M., Campi, D., Abou El Kheir, O., Laporte, S. I., Rataj, B., Czamler, V., Zimmer, O., Gorini, G., Santoro, V. & Muhrer, G. (2024). J. Phys. Condens. Matter 36, 385904.
Google Scholar
Xu, S., DiJulio, D. D., Marquez Damian, J. I., Vogel, S. C., Long, A. M., Hirsh, T. Y., Kittelmann, T., Kuksenko, V. & Muhrer, G. (2025). J. Appl. Cryst. 58, 1957–1966.
CrossRef
CAS
IUCr Journals
Google Scholar
Yang, J., Ren, W., Zhao, X., Kikuchi, T., Miao, P., Nakajima, K., Li, B. & Zhang, Z. (2022). J. Mater. Sci. Technol. 99, 55–60.
CrossRef
CAS
Google Scholar
Yu, H., Díaz, A., Lu, X., Sun, B., Ding, Y., Koyama, M., He, J., Zhou, X., Oudriss, A., Feaugas, X. & Zhang, Z. (2024). Chem. Rev. 124, 6271–6392.
CrossRef
CAS
Google Scholar
Zendler, C., Lieutenant, K., Nekrassov, D. & Fromme, M. (2014). J. Phys. Conf. Ser. 528, 012036.
CrossRef
Google Scholar
Zou, S., Chen, H., Zhou, Y., Wang, W., Song, T., Wang, S., Zhu, X. & Sang, B. (2026). Met. Mater. Int. 32, 298–313.
CrossRef
CAS
Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.
access

menu